Skip to main content
Erschienen in: BMC Medical Imaging 1/2017

Open Access 01.12.2017 | Research article

Registration error of the liver CT using deformable image registration of MIM Maestro and Velocity AI

verfasst von: Nobuyoshi Fukumitsu, Kazunori Nitta, Toshiyuki Terunuma, Toshiyuki Okumura, Haruko Numajiri, Yoshiko Oshiro, Kayoko Ohnishi, Masashi Mizumoto, Teruhito Aihara, Hitoshi Ishikawa, Koji Tsuboi, Hideyuki Sakurai

Erschienen in: BMC Medical Imaging | Ausgabe 1/2017

Abstract

Background

Understanding the irradiated area and dose correctly is important for the reirradiation of organs that deform after irradiation, such as the liver. We investigated the spatial registration error using the deformable image registration (DIR) software products MIM Maestro (MIM) and Velocity AI (Velocity).

Methods

Image registration of pretreatment computed tomography (CT) and posttreatment CT was performed in 24 patients with liver tumors. All the patients received proton beam therapy, and the follow-up period was 4–14 (median: 10) months. We performed DIR of the pretreatment CT and compared it with that of the posttreatment CT by calculating the dislocation of metallic markers (implanted close to the tumors).

Results

The fiducial registration error was comparable in both products: 0.4–32.9 (9.3 ± 9.9) mm for MIM and 0.5–38.6 (11.0 ± 10.0) mm for Velocity, and correlated with the tumor diameter for MIM (r = 0.69, P = 0.002) and for Velocity (r = 0.68, P = 0.0003). Regarding the enhancement effect, the fiducial registration error was 1.0–24.9 (7.4 ± 7.7) mm for MIM and 0.3–29.6 (8.9 ± 7.2) mm for Velocity, which is shorter than that of plain CT (P = 0.04, for both).

Conclusions

The DIR performance of both MIM and Velocity is comparable with regard to the liver. The fiducial registration error of DIR depends on the tumor diameter. Furthermore, contrast-enhanced CT improves the accuracy of both MIM and Velocity.

Institutional review board approval

H28-102; July 14, 2016 approved.
Abkürzungen
CT
(Computed tomography): image used by computer-processed combinations of many X-ray images
DIR
(Deformable image registration): image registration technique with deformation
HU
(Hounsfield unit): A quantity commonly used in CT to express CT numbers in a standardized and convenient form
MIM
(MIM Maestro): name of software
PBT
(Proton beam therapy): a type of radiotherapy using proton beams
RIR
(Rigid image registration): image registration technique horizontally and rotationally without deformation
Velocity
(Velocity AI): name of software

Background

The various organs of the human body are often deformed by irradiation. Reirradiation is sometimes conducted to treat new lesions that might occur. Before reirradiation is performed, it is vital to confirm the region irradiated by previous radiotherapy to avoid excess irradiation to the normal tissue as this could cause severe adverse effects. Primary liver tumors tend to recur inside the liver after treatment, and metastasis from other organs also makes it highly possible that new lesions will occur inside the liver. A substantial number of patients, therefore, require reirradiation to treat recurrence of cancer in the liver [1].
Thus, image registration is a particularly important issue for treating patients with liver tumors, even though, since the advent of particle beam radiotherapy, irradiated lesions are now well controlled [25]. To conduct image registration, an important problem must be considered: after irradiation, the irradiated area including the tumor is shrunk and the nonirradiated area is shifted and sometimes enlarged or shrunk, thus causing remarkable deformation of the liver in most patients [6]. This issue can be addressed by using deformable image registration (DIR), and a number of software products are now on the market. The use of DIR for applications and assessment of previously delivered irradiation doses is clinically expected to protect the normal liver tissue from receiving harmfully large doses of irradiation [7].
Several DIR algorithms exist, and DIR can be broadly classified into two categories: (1) intensity-based methods, which use a variety of image intensity metrics such as the gray scale, and (2) feature-based methods, which use specific image features such as contours [8]. Transformation models include optical flow-based equations [9], the “Demons” equation [10], B-splines [11], and thin-plate splines [12]. In most registration algorithms, the balance between image similarity and accurate matching of the local features on the one hand and deformation smoothness on the other hand is crucial to accurately measure the deformation [13]. The technique for evaluating the spatial accuracy of DIR involves landmark tracking [14] or contour or structure comparisons [15].
In recent years, advanced software equipped with the function of DIR has been developed for research purposes, and some of them are also available for clinical use. MIM Maestro (MIM Software Inc., OH, USA) (MIM) and Velocity AI (Velocity Medical Solutions, GA, USA) (Velocity) are the two most widely used in Japan and worldwide 2 of the 3 most widely used commercially available software products that can perform DIR and assist radiotherapy planning. In MIM, the DIR algorithm is intensity-based, free form cubic spline interpolation with essentially unlimited degrees of freedom [13]. MIM can dramatically deform the image, while, in cases with little contrast, it might lead to unreasonable deformation of the image. In Velocity, a B-spline deformable model is used [13]. Velocity uses standardized image intensity and has a smoothing and regularization function derived from the B-spline method. Although both software products are relatively new, some reports on their features and differences have been already been published. In brief, these are that MIM has the advantage in terms of small fields but sometimes produces registration error because of image noise, and Velocity has the advantage in terms of large fields. So far, there is no strong evidence for which software is superior [8, 13, 16].
We examined the spatial accuracy of DIR of MIM and Velocity after irradiation in the preliminary stage of deformation of the dose distribution in reirradiated liver tumors.

Methods

We retrospectively reviewed patients who had received proton beam therapy (PBT) at our institute. All the study procedures involving human participants were conducted in accordance with the ethical standards of the institutional research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. All the treatments were discussed at inhospital conferences, and informed consent was obtained from all the individual participants included in the study. The study received institutional review board approval (H28-102). We selected those patients who had a metallic material such as a fiducial marker or surgical clip (herein called metallic marker) already implanted very close to the liver tumor before PBT. We examined 24 consecutive patients treated between 2009 and 2014 (20 men, 4 women; aged 52–84 years). The most common disease was hepatocellular carcinoma (18 patients), followed by liver metastasis (5 patients) and intrahepatic bile duct carcinoma (1 patient). Fiducial markers for previous PBT were present in 21 patients, and surgical clips, in 3 patients. At our institute, abdominal computed tomography (CT) for diagnosis is usually not taken after metallic marker implantation, so these 21 patients had come to our hospital to receive PBT for new lesions in the liver. Twenty-two patients underwent irradiation for single lesions, and 2 patients, for 2 lesions. The total tumor diameter was 10-69 (median: 35) mm. The tumor was located in the left lobe in 5 patients, in the right lobe in 15 patients, and in both lobes in 4 patients. The distance between the tumor and the metallic marker was 5-33.7 (median: 12.0) mm. The total irradiation dose was 50-74 GyE in 22-37 fractions (Table 1).
Table 1
Characteristic of the patients
 
Dislocation
 
Number
Disease
materials
MIM (plain)
MIM (enhance)
Velocity (plain)
Velocity (enhance)
Tumor location
Tumor diameter
tumor-material distance
Dose (GyE/fr)
Duration (months)
Remark
1
HCC
marker
0.40
1.00
0.50
1.00
S8
35
11.5
72.6/22
9.4
 
2
IHBDC
clip
3.00
2.30
18.00
14.30
S3
45
14.2
74/37
14.2
 
3
meta
marker
6.70
8.70
9.20
7.70
S7
30
16.1
72.6/22
10.9
 
4
HCC
clip
4.70
2.10
4.00
3.10
S4
20
5.2
72.6/22
10.3
 
5
HCC
marker
1.80
2.10
4.60
0.30
S4/8
30
7.2
50/25
5.5
 
6
meta
marker
15.20
8.20
20.50
4.50
S8
60
8.7
72.6/22
7.2
at
7
HCC
marker
8.20
5.40
22.30
20.40
S4/8
50
10.8
72.6/22
6.2
pl
8
meta
marker
1.30
1.60
4.00
2.20
S6
25
13.2
60/30
10.0
 
9
HCC
marker
4.70
4.70
4.80
7.30
S7
25
5.2
50/25
10.9
 
10
HCC
marker
1.80
2.20
1.90
2.00
S4
24
5.4
72.6/22
7.8
 
11
HCC
marker
0.40
1.20
6.30
9.7
S5
35
5.7
50/25
12.1
 
12
HCC
marker
18.10
19.10
4.70
16.60
S7
56
8.3
72.6/22
8.9
as
13
HCC
marker
22.90
5.90
18.30
12.50
S8
30
9.5
50/25
11.2
as
14
meta
marker
11.10
14.60
5.50
5.20
S4/8
47
9.6
60/30
7.2
pl
15
HCC
marker
32.90
18.90
21.40
13.80
S8
60
10.3
72.6/22
11.9
as
16
HCC
marker
11.30
2.30
10.70
7.70
S7
15
12.4
60/30
10.3
 
17
HCC
marker
2.10
3.10
1.30
2.60
S1/4
35
14.9
60/30
8.0
as
18
HCC
marker
15.50
20.00
12.50
10.80
S7/8
38
16.1
72.6/22
6.5
 
19
HCC
marker
0.80
1.20
3.40
3.10
S5/8
35
16.5
74/37
10.1
at
20
HCC
marker
0.70
2.00
9.70
9.00
S8
10
19.1
60/15
6.8
 
21
HCC
marker
2.60
2.90
6.80
6.30
S1
24
30.9
72.6/22
8.7
 
22
HCC
marker
29.60
24.90
38.60
S5/8
68
68
32.8
52.8/16
4.2
as
23
HCC
marker
24.60
20.80
30.70
19.10
S6,S8
69
33.0
70/35,66/10
12.7
as
24
meta
clip
3.00
2.50
3.20
4.60
S1,S5/8
40
33.7
45/15,72.6/22
13.4
 
Abbreviations: HCC hepatocellular carcinoma, IHBDC intrahepatic bile duct carcinoma, GyE/fr Gray equivaent/fractions, at; atelectasis appearance during follow-up, pl pleural effusion appearance during follow-up, as ascites appearance or increased during follow-up
Duration means the interval to the post treatment CT. All unit of length is mm
Contrast-enhanced CT with the breath-holding technique was taken before and after treatment. The duration between PBT and posttreatment CT was 4-14 (median: 10) months. CT with a matrix resolution of 512*512 and a slice thickness of 5 mm was used. We used both plain and contrast-enhanced CT in our image analysis. For the patients who had dynamic contrast-enhanced CT, we used portal venous-phase CT. We performed rigid image registration (RIR) and then DIR of the pretreatment CT images. During the registration process, the priority area of calculation was manually designated to cover the whole liver. The fiducial registration error was assessed by examining the dislocation of the metallic marker from its position in the posttreatment CT and that in the deformed pretreatment CT images. We used the point at which the metallic density was the highest as the position of the metallic marker. The same process was performed using MIM (version 6.5.2) and Velocity (version 3.1.0).
The fiducial registration errors of MIM and Velocity were compared using both plain and contrast-enhanced CT and a paired t test. The Pearson product moment correlation was performed to examine the correlation between the fiducial registration error of RIR and DIR. Simple linear regression analysis was performed to examine the correlation between tumor diameter and fiducial registration error. Probability values below 0.05 were considered significant.

Results

In the plain CT, the fiducial registration error was 0.4–32.9 (9.3 ± 9.9) mm for MIM and 0.5–38.6 (11.0 ± 10.0) mm for Velocity. The fiducial registration error was less for MIM in 16 patients and for Velocity in 8 patients; overall however, the results for both MIM and Velocity were similar (P = 0.18). In the contrast-enhanced CT, the fiducial registration errors for MIM (1.0–24.9 [7.4 ± 7.7] mm) and Velocity (0.3–29.6 [8.9 ± 7.1] mm) were also similar (P = 0.22) (Fig. 1a). As for the enhancement effect, the fiducial registration errors for MIM and Velocity were significantly shorter than they were in the plain CT (P = 0.04, for both). (Fig. 1b).
With regard to the fiducial registration error, DIR was significantly correlated with RIR for both MIM (r = 0.62, P = 0.001) and Velocity (r = 0.9, P = 3.3×10-9) in the plain CT (Fig. 2a). In the contrast-enhanced CT, DIR was also significantly correlated with RIR for both MIM (r =0.66, P = 0.0004) and Velocity (r = 0.84, P = 3.6 × 10-7). The fiducial registration error was significantly correlated with the tumor diameter for both MIM (r = 0.69, P = 0.002) and Velocity (r =0.68, P =0.0003) in the plain CT. In the contrast-enhanced CT, the fiducial registration error was also significantly correlated with the tumor diameter for both MIM (r =0.75, P = 2.8 × 10-5) and Velocity (r =0.63, P = 9 × 10-5). The tumor diameter predicted to produce a 10-mm fiducial registration error was 39.4 mm for MIM and 35.5 mm for Velocity in the plain CT and 45.6 mm for MIM and 42 mm for Velocity in the contrast-enhanced CT (Fig. 2b).
The tumor-marker distance did not differ according to the classification of the tumor location. However, cases in which the tumor in the right lobe had a trend toward the tumor-marker distance became large (Table 2).
Table 2
Cases with large tumor marker distance
MIM (plain)
MIM (enhance)
Velocity (palin)
Velocity (enhance)
T. M. D. (mm)
T. L.
T. D. (mm)
Dis. (mm)
Loc.
T. D. (mm)
Dis. (mm)
Loc.
T. D. (mm)
Dis. (mm)
Loc.
T. D. (mm)
22.9
S8
30
20.0
S7/8
38
21.4
S8
60
20.4
S4/8
50
24.6
S6,S8
69
20.8
S6, S8
69
22.3
S4/8
50
29.6
S5/8
68
29.6
S5/8
68
24.9
S5/8
68
30.7
S6, S8
69
   
32.9
S8
60
   
38.6
S5/8
68
68
  
Abbreviations: T. M. D. tumor marker distance, T. L tumor location, T. D tumor diameter
Figure 3 shows a case in which the fiducial registration error was small and similar for both MIM and Velocity. Figure 4 shows the cases that showed the biggest discrepancies between MIM and Velocity in terms of the fiducial registration error. Figure 5 shows the cases in which the DIR results were greatly affected by the contrast enhancement effect. All the deformed images showed slice levels that corresponded to the location of the metallic marker on the posttreatment CT.

Discussion

We compared the capabilities of MIM and Velocity. At first, we considered the advantages of MIM. Although the fiducial registration error of DIR correlated with that of RIR both for MIM and for Velocity, there were some differences in the degree of fiducial registration error. As shown in Fig. 2(a), the correlation coefficient for Velocity was 0.9, which is an extremely strong correlation, whereas for MIM it was 0.62, which is only a moderate correlation. Figure 4(a) shows the case with the largest difference in fiducial registration error between Velocity and MIM. The metallic marker could not be found in the axial, coronal, or sagittal image in the RIR process of both MIM and Velocity. However, during the DIR process, MIM could correctly identify the location in which the metallic marker was observed in any of the directional images. In contrast, the DIR process of Velocity shifted the location to a position in which the metallic marker was not observed in any of the directional images. Yeo et al reported that the magnitude of deformation has a much larger effect on the accuracy of registration than does the complexity of deformation [17]. Our results suggest that the DIR process in Velocity seems to be more dependent on the RIR, while MIM has a greater ability to modify during the process of DIR. Next, we considered the advantages of Velocity. As shown in Fig. 4(b) of an irradiated tumor in the right lobe, MIM showed an unnaturally deformed low-density tumor, especially in the coronal image, whereas Velocity showed the natural shape of the tumor. Previous studies have shown that MIM produces beautiful image similarity but may produce nonphysical deformation fields, while Velocity produces smooth, physically plausible deformation fields [18]. It seems that MIM has higher flexibility and causes excellent overall spatial accuracy; although it tends to deform forcibly. This could sometimes cause unreasonable DIR and diminish the ability to transfer contours.
As shown in Fig. 1(a), although the fiducial registration errors of MIM and Velocity were similar overall, they were not necessary closely correlated, and for some patients either MIM or Velocity had a distinct advantage. Some previous studies reported that it is hard to decide which software is impartially superior to others in terms of DIR accuracy [8, 13, 18]. Our results also demonstrate that the superiority of DIR accuracy for the liver varies by patient, making it difficult to state which software is better. It is generally understood that DIR will work well with feature-rich images in which there is little or no ambiguity between corresponding points in the source and target images. The liver is one of the organs that have a relatively homogeneous Hounsfield unit (HU) and a lack of morphological characteristics. Therefore, one important objective of this study was to determine how well these software products can perform DIR in low-contrast organs such as the liver. We analyzed the portal vein phase image, which enhances a greater number of vessels. As we expected, contrast-enhanced CT could accomplish on average 1.9-mm less fiducial registration error than could plain CT in MIM. Moreover, Velocity could also, on average, accomplish a 2.1-mm enhancement effect, as shown in Fig 1(b). In addition, the enhancement effect could change the deformation pattern. As shown in Fig. 5, the vector went toward the right posterior direction in the plain CT; by contrast, the vector circled in a clockwise direction in the contrast-enhanced CT of MIM. Similarly, in the plain CT, the vector moved slightly backwards only in the peripheral region of the liver, whereas in the contrast-enhanced CT of Velocity, the vector moved forward through most of the regions of the liver. We are convinced that in the DIR of both MIM and Velocity, enhancement-derived contrast not only works toward spatial accuracy but also changes the deformation pattern, such as the linear and rotational directions.
We used a metallic marker to calculate the image registration accuracy because it is difficult to measure the registration error of the tumor itself and because precise contouring of the whole liver by distinguishing the liver from the porta hepatis is not completely reproducible in each study. It is feasible to calculate the fiducial registration error at multiple points in each patient. However, in the daily clinical setting, the number of implanted metallic markers is usually 1 or 2. Therefore, we selected only the patients whose metallic materials were close to the tumor. In this study, the metallic markers were implanted close to the tumor, with a range of 5–33.7 (median: 12.0) mm. Thus, we consider the metallic markers to act as surrogates of the fiducial registration error in DIR. There may be criticism that the high HU of the metallic markers could have affected the image registration. However, in all the patients, the metallic artifact was so small when compared with the large volume of the liver, and therefore, we think that the accuracy of the image registration is seldom affected by the artifact. Moreover, we routinely compare and conduct image registration of CTs with tiny metallic markers already implanted. Therefore, this analysis reflects the conditions of the daily clinical setting, and we consider that analysis using CT with implantation of tiny metallic markers is clinically allowable.
Both MIM and Velocity have rapidly expanded the market for such types of software, and new software products have been manufactured or are planned for manufacture to meet the demand for high-precision radiotherapy. It is expected that several types of commercial-based software products equipped with the DIR function will be developed, not only for examination of previous radiotherapy planning, but also for use in adaptive radiotherapy. In this study, we investigated the registration error by using clinical data. However, it is also important to investigate and validate the registration error by using phantoms. We are considering a phantom study as the next step to prove the registration error that we concluded in this clinical study.

Conclusion

For image registration of the liver, the DIR performances of MIM and Velocity are similar overall. However, which software is the better option varies according to the patient. The spatial accuracy of DIR depends on the accuracy of RIR and also on the tumor diameter. Finally, contrast-enhanced CT improves the accuracy of both MIM and Velocity.

Acknowledgements

We really appreciate Thomas Mayers and Flaminia Miyamasu, Medical English Communications Center, University of Tsukuba, for grammatical revision of the manuscript.

Funding

None.

Author’s contributions

NF carried out manuscript writing. KN carried out data analysis of Velocity. TT carried out data analysis of MIM. TO supported data analysis. HN carried out clinical examinations. YO participated in the design of the study. KO carried out clinical examinations. MM participated in the design of the study. TA carried out clinical examinations. HI supported statistical analysis. KT supported to draft the manuscript. HS supported data analysis. All authors read and approved the final version of the manuscript.

Competing interests

The authors declare that they have no competing interests.
Not appricable.
H28-102, 2016/7/14 IRB approved in Tsukuba University Hospital, Japan. Consent was obtained with written form when patients come to our hospital. It was also obtained by air-mail for some patients who ended to come to our hospital. All consent was obtained before starting data analysis.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Hashimoto T, Tokuuye K, Fukumitsu N, Igaki H, Hata M, Kagei K, Sugahara S, Ohara K, Matsuzaki Y, Akine Y. Repeated proton beam therapy for hepatocellular carcinoma. Int J Radiat Oncol Biol Phys. 2006;65(1):196–202.CrossRefPubMed Hashimoto T, Tokuuye K, Fukumitsu N, Igaki H, Hata M, Kagei K, Sugahara S, Ohara K, Matsuzaki Y, Akine Y. Repeated proton beam therapy for hepatocellular carcinoma. Int J Radiat Oncol Biol Phys. 2006;65(1):196–202.CrossRefPubMed
2.
Zurück zum Zitat Bush DA, Kayali Z, Grove R, Slater JD. The safety and efficacy of high-dose proton beam radiotherapy for hepatocellular carcinoma: a phase 2 prospective trial. Cancer. 2011;117(13):3053–9.CrossRefPubMed Bush DA, Kayali Z, Grove R, Slater JD. The safety and efficacy of high-dose proton beam radiotherapy for hepatocellular carcinoma: a phase 2 prospective trial. Cancer. 2011;117(13):3053–9.CrossRefPubMed
3.
Zurück zum Zitat Fukumitsu N, Sugahara S, Nakayama H, Fukuda K, Mizumoto M, Abei M, Shoda J, Thono E, Tsuboi K, Tokuuye K. A prospective study of hypofractionated proton beam therapy for patients with hepatocellular carcinoma. Int J Radiat Oncol Biol Phys. 2009;74(3):831–6.CrossRefPubMed Fukumitsu N, Sugahara S, Nakayama H, Fukuda K, Mizumoto M, Abei M, Shoda J, Thono E, Tsuboi K, Tokuuye K. A prospective study of hypofractionated proton beam therapy for patients with hepatocellular carcinoma. Int J Radiat Oncol Biol Phys. 2009;74(3):831–6.CrossRefPubMed
4.
Zurück zum Zitat Mizumoto M, Okumura T, Hashimoto T, Fukuda K, Oshiro Y, Fukumitsu N, Abei M, Kawaguchi A, Hayashi Y, Ookawa A, et al. Proton beam therapy for hepatocellular carcinoma: a comparison of three treatment protocols. Int J Radiat Oncol Biol Phys. 2011;81(4):1039–45.CrossRefPubMed Mizumoto M, Okumura T, Hashimoto T, Fukuda K, Oshiro Y, Fukumitsu N, Abei M, Kawaguchi A, Hayashi Y, Ookawa A, et al. Proton beam therapy for hepatocellular carcinoma: a comparison of three treatment protocols. Int J Radiat Oncol Biol Phys. 2011;81(4):1039–45.CrossRefPubMed
5.
Zurück zum Zitat Qi WX, Fu S, Zhang Q, Guo XM. Charged particle therapy versus photon therapy for patients with hepatocellular carcinoma: a systematic review and meta-analysis. Radiother Oncol. 2015;114(3):289–95.CrossRefPubMed Qi WX, Fu S, Zhang Q, Guo XM. Charged particle therapy versus photon therapy for patients with hepatocellular carcinoma: a systematic review and meta-analysis. Radiother Oncol. 2015;114(3):289–95.CrossRefPubMed
6.
Zurück zum Zitat Ahmadi T, Itai Y, Onaya H, Yoshioka H, Okumura T, Akine Y. CT evaluation of hepatic injury following proton beam irradiation: appearance, enhancement, and 3D size reduction pattern. J Comput Assist Tomogr. 1999;23(5):655–63.CrossRefPubMed Ahmadi T, Itai Y, Onaya H, Yoshioka H, Okumura T, Akine Y. CT evaluation of hepatic injury following proton beam irradiation: appearance, enhancement, and 3D size reduction pattern. J Comput Assist Tomogr. 1999;23(5):655–63.CrossRefPubMed
7.
Zurück zum Zitat Fukumitsu N, Hashimoto T, Okumura T, Mizumoto M, Tohno E, Fukuda K, Abei M, Sakae T, Sakurai H. Investigation of the geometric accuracy of proton beam irradiation in the liver. Int J Radiat Oncol Biol Phys. 2012;82(2):826–33.CrossRefPubMed Fukumitsu N, Hashimoto T, Okumura T, Mizumoto M, Tohno E, Fukuda K, Abei M, Sakae T, Sakurai H. Investigation of the geometric accuracy of proton beam irradiation in the liver. Int J Radiat Oncol Biol Phys. 2012;82(2):826–33.CrossRefPubMed
8.
Zurück zum Zitat Wognum S, Heethuis SE, Rosario T, Hoogeman MS, Bel A. Validation of deformable image registration algorithms on CT images of ex vivo porcine bladders with fiducial markers. Med Phys. 2014;41(7):071916.CrossRefPubMed Wognum S, Heethuis SE, Rosario T, Hoogeman MS, Bel A. Validation of deformable image registration algorithms on CT images of ex vivo porcine bladders with fiducial markers. Med Phys. 2014;41(7):071916.CrossRefPubMed
9.
Zurück zum Zitat Yang D, Chaudhari SR, Goddu SM, Pratt D, Khullar D, Deasy JO, El Naqa I. Deformable registration of abdominal kilovoltage treatment planning CT and tomotherapy daily megavoltage CT for treatment adaptation. Med Phys. 2009;36(2):329–38.CrossRefPubMedPubMedCentral Yang D, Chaudhari SR, Goddu SM, Pratt D, Khullar D, Deasy JO, El Naqa I. Deformable registration of abdominal kilovoltage treatment planning CT and tomotherapy daily megavoltage CT for treatment adaptation. Med Phys. 2009;36(2):329–38.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Wang H, Dong L, Lii MF, Lee AL, de Crevoisier R, Mohan R, Cox JD, Kuban DA, Cheung R. Implementation and validation of a three-dimensional deformable registration algorithm for targeted prostate cancer radiotherapy. Int J Radiat Oncol Biol Phys. 2005;61(3):725–35.CrossRefPubMed Wang H, Dong L, Lii MF, Lee AL, de Crevoisier R, Mohan R, Cox JD, Kuban DA, Cheung R. Implementation and validation of a three-dimensional deformable registration algorithm for targeted prostate cancer radiotherapy. Int J Radiat Oncol Biol Phys. 2005;61(3):725–35.CrossRefPubMed
11.
Zurück zum Zitat Wen N, Glide-Hurst C, Nurushev T, Xing L, Kim J, Zhong H, Liu D, Liu M, Burmeister J, Movsas B, et al. Evaluation of the deformation and corresponding dosimetric implications in prostate cancer treatment. Phys Med Biol. 2012;57(17):5361–79.CrossRefPubMed Wen N, Glide-Hurst C, Nurushev T, Xing L, Kim J, Zhong H, Liu D, Liu M, Burmeister J, Movsas B, et al. Evaluation of the deformation and corresponding dosimetric implications in prostate cancer treatment. Phys Med Biol. 2012;57(17):5361–79.CrossRefPubMed
12.
Zurück zum Zitat Vasquez Osorio EM, Hoogeman MS, Bondar L, Levendag PC, Heijmen BJ. A novel flexible framework with automatic feature correspondence optimization for nonrigid registration in radiotherapy. Med Phys. 2009;36(7):2848–59.CrossRefPubMed Vasquez Osorio EM, Hoogeman MS, Bondar L, Levendag PC, Heijmen BJ. A novel flexible framework with automatic feature correspondence optimization for nonrigid registration in radiotherapy. Med Phys. 2009;36(7):2848–59.CrossRefPubMed
13.
Zurück zum Zitat Kirby N, Chuang C, Ueda U, Pouliot J. The need for application-based adaptation of deformable image registration. Med Phys. 2013;40(1):011702.CrossRefPubMed Kirby N, Chuang C, Ueda U, Pouliot J. The need for application-based adaptation of deformable image registration. Med Phys. 2013;40(1):011702.CrossRefPubMed
14.
Zurück zum Zitat Castillo R, Castillo E, Guerra R, Johnson VE, McPhail T, Garg AK, Guerrero T. A framework for evaluation of deformable image registration spatial accuracy using large landmark point sets. Phys Med Biol. 2009;54(7):1849–70.CrossRefPubMed Castillo R, Castillo E, Guerra R, Johnson VE, McPhail T, Garg AK, Guerrero T. A framework for evaluation of deformable image registration spatial accuracy using large landmark point sets. Phys Med Biol. 2009;54(7):1849–70.CrossRefPubMed
15.
Zurück zum Zitat Wognum S, Bondar L, Zolnay AG, Chai X, Hulshof MC, Hoogeman MS, Bel A. Control over structure-specific flexibility improves anatomical accuracy for point-based deformable registration in bladder cancer radiotherapy. Med Phys. 2013;40(2):021702.CrossRefPubMed Wognum S, Bondar L, Zolnay AG, Chai X, Hulshof MC, Hoogeman MS, Bel A. Control over structure-specific flexibility improves anatomical accuracy for point-based deformable registration in bladder cancer radiotherapy. Med Phys. 2013;40(2):021702.CrossRefPubMed
16.
Zurück zum Zitat Nie K, Chuang C, Kirby N, Braunstein S, Pouliot J. Site-specific deformable imaging registration algorithm selection using patient-based simulated deformations. Med Phys. 2013;40(4):041911.CrossRefPubMed Nie K, Chuang C, Kirby N, Braunstein S, Pouliot J. Site-specific deformable imaging registration algorithm selection using patient-based simulated deformations. Med Phys. 2013;40(4):041911.CrossRefPubMed
17.
Zurück zum Zitat Yeo UJ, Supple JR, Taylor ML, Smith R, Kron T, Franich RD. Performance of 12 DIR algorithms in low-contrast regions for mass and density conserving deformation. Med Phys. 2013;40(10):101701.CrossRefPubMed Yeo UJ, Supple JR, Taylor ML, Smith R, Kron T, Franich RD. Performance of 12 DIR algorithms in low-contrast regions for mass and density conserving deformation. Med Phys. 2013;40(10):101701.CrossRefPubMed
18.
Zurück zum Zitat Singhrao K, Kirby N, Pouliot J. A three-dimensional head-and-neck phantom for validation of multimodality deformable image registration for adaptive radiotherapy. Med Phys. 2014;41(12):121709.CrossRefPubMed Singhrao K, Kirby N, Pouliot J. A three-dimensional head-and-neck phantom for validation of multimodality deformable image registration for adaptive radiotherapy. Med Phys. 2014;41(12):121709.CrossRefPubMed
Metadaten
Titel
Registration error of the liver CT using deformable image registration of MIM Maestro and Velocity AI
verfasst von
Nobuyoshi Fukumitsu
Kazunori Nitta
Toshiyuki Terunuma
Toshiyuki Okumura
Haruko Numajiri
Yoshiko Oshiro
Kayoko Ohnishi
Masashi Mizumoto
Teruhito Aihara
Hitoshi Ishikawa
Koji Tsuboi
Hideyuki Sakurai
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
BMC Medical Imaging / Ausgabe 1/2017
Elektronische ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-017-0202-z

Weitere Artikel der Ausgabe 1/2017

BMC Medical Imaging 1/2017 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.