Prediction accuracy of intraocular lens power calculation methods after laser refractive surgery

verfasst von: Yubo Wu, Songyu Liu, Rongfeng Liao

Erschienen in: BMC Ophthalmology | Ausgabe 1/2017

Abstract

Background

This study aimed to evaluate the prediction accuracy of postoperative refractions using partial coherence interferometry (IOL-Master) and applanation ultrasound (AL-3000) assisted with corneal topography (TMS-4) in eyes that had undergone myopic laser-assisted in situ keratomileusis (LASIK).

Methods

Haigis-L formula, Koch–Maloney method using Haigis formula, Shammas clinically derived K-value (simulated keratometric value) correction (Shammas c.d.) using Haigis formula, and Shammas post-LASIK (Shammas-PL) formula were used in eyes with myopic LASIK. Constants were derived from the optimized constants in 133 virgin eyes. Refractive outcomes were determined by streak retinoscopy and subjective manifest refraction. Methods and formulas were evaluated by mean error (ME), standard deviation (SD), range of error, mean absolute error (MAE), median absolute error, 95% confidence interval of MAE, and percentage of eyes within ±0.5 diopter (D), ±1.0 D, and ±1.5 D of prediction.

Results

SDs of the Haigis-L, Koch-Maloney method using the Haigis formula, Shammas c.d. using the Haigis formula, and the Shammas-PL formula using IOL-Master were 0.721, 0.695, 0.695, and 0.698; and those using AL-3000 assisted with TMS-4 were 0.782, 0.741, 0.743, and 0.778, respectively.

Conclusions

No-history methods that corrected corneal power with measurements using IOL-Master were promising in myopic post-LASIK eyes, but still a gap in prediction accuracy exists between virgin eyes and post-LASIK eyes.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​s12886-017-0439-x) contains supplementary material, which is available to authorized users.
Abkürzungen
ACD
Anterior chamber depth
AL
Axial length
BCVA
Best-corrected visual acuity
D
Diopter
ELP
Effective lens position
IOL
Intraocular lens
LASIK
laser-assisted in situ keratomileusis
MAE
Mean absolute error
ME
mean error
PCI
Partial coherence interferometry
SD
standard deviation
SE
Spherical equivalent

Background

Refractive outcomes after cataract surgeries and intraocular lens (IOL) implantation have always been a hot topic since the first IOL was implanted in 1949 [1], especially with the increasingly higher expectations of patients in recent years. In spite of its higher resolution, reproducibility, and repeatability of optical biometers [2], an ultrasonic device is irreplaceable in cases with severe opacities (e.g., corneal scar, mature and posterior subcapsular cataract, and vitreous hemorrhage) along the visual axis or fixation instability in most hospitals [3]. It is slightly doubtful that immersion ultrasound is more suitable than contact applanation ultrasound [4]. Ironically, the latter one is more widely used in China and many other countries, leading to increased attention paid to its prediction accuracy in IOL power calculation.
Patients who have undergone myopic laser-assisted in situ keratomileusis (LASIK) tend to have higher expectations regarding the refractive outcome. Intraocular lens calculation for them is much harder mainly because of the following reasons. First, it is hard for most devices to calculate the true post-LASIK corneal power using the corneal radius of curvature. This is due to the change in the relationship between the anterior and posterior curvatures of the cornea, making the standardized keratometric index inappropriate [5]. Moreover, when estimating the effective lens position (ELP), it is not wise to use the post-LASIK K-value (simulated keratometric value), as most of the third-generation formulas do [6].
Dozens of methods have been proposed to solve the problem. However, they can be hardly comparable with the calculation in virgin eyes. Most methods can be classified into two groups: methods that need data before the LASIK that patients underwent and methods that were based only on current measurements. Usually, the data before LASIK are not available or reliable, and literature has shown that no-history approaches are superior [710], which are widely used in clinical practice now. A series of methods that are used for calculating corrected corneal power (Kc) are accepted for their accessibility and convenience. The present study concentrated on the three of them with different formulas: Haigis-L [7] formula; Koch–Maloney method [8], in which Kc = 1.114 × measured K – 6.1, using Haigis formula; Shammas c.d. [11], in which Kc = 1.14 × measured K – 6.8, using Haigis formula; and Shammas-PL formula [9]. Haigis-L is a formula that has been widely used and shown promising results. Shammas c.d. is normally used with the Shammas-PL formula. However, the formula is not routinely preinstalled in an optical device or ultrasonic device, making it a little inconvenient to use. Therefore, the question arose whether the Shammas-PL formula and Haigis formula, which predicted ELP from anterior chamber depth (ACD) and axial length (AL), were comparable.
This study prospectively evaluated three biometers (IOL-Master, AL-3000, and TMS-4) with four methods (Haigis-L formula, Koch-Maloney method using Haigis formula, Shammas c.d. using Haigis formula, and Shammas-PL formula) in post-LASIK eyes after optimization of constants in 133 virgin eyes.

Methods

Inclusion and exclusion criteria

This study recorded data from consecutive Chinese patients with virgin eyes who were hospitalized in the Department of Ophthalmology, The First Affiliated Hospital of Anhui Medical University between January 15, 2016, and May 1, 2016, and post-LASIK patients between July 15, 2015, and May 31, 2016. Patients who had undergone uneventful phacoemulsification cataract surgery by an experienced surgeon (Rongfeng Liao) with a temporal clear corneal incision, 3.0 mm in width, were included in the study. Eyes were included if they received in-the-bag placement of a monofocal IOL (920H, Rayner, Inc., London, UK; Adapt-AO, Bausch & Lomb, Inc., NY, USA; SN60WF, Alcon Laboratories, Inc., Hünenberg, Switzerland; or ZCB00, AMO, Inc., CA, USA). Eyes were excluded if any parameter relative to IOL power calculation (AL, K-value, and ACD) using IOL-Master, AL-3000, or TMS-4 could not be measured reliably. Other exclusion criteria included corneal astigmatism more than 1.5 diopter (D), postoperative best-corrected visual acuity (BCVA) less than 20/40, previous ocular surgeries except myopic LASIK, combined surgery, intraoperative and postoperative complications, active ocular infection, and systemic diseases that might have affected eyes. Patients who had a follow-up time less than 3 months were also excluded. If a patient had both eyes operated, only the first qualified eye was included in the study. This paper was approved by the medical ethics committee of the hospital. The study adhered to the Declaration of Helsinki, and informed consent was obtained from all patients.

Preoperative and postoperative assessment

Preoperative assessment included examinations from IOL-Master (version 5.5, Carl Zeiss Meditec, Inc., Jena, Germany, keratometric index: 1.3375), TMS-4 (Topographic Modeling System, Tomey, Inc., Nagoya, Japan, keratometric index: 1.3375), and AL-3000 (Bio & Pachy Meter AL-3000, Tomey, Inc., Nagoya, Japan). The examination of AL-3000 was performed last to avoid corneal indentation and maintain the integrity of the corneal epithelium when other preoperative measurements were gauged. The sequence between IOL-Master and TMS-4 was random. Patients were asked to blink just before measurements were taken. Pupil sizes during preoperative assessments were normal [12]. All preoperative examinations were carried out by the first author, who was trained and qualified according to the manufacturer’s recommendation. The follow-up points were 1 week, 1 month, and 3 months after the surgery when postoperative assessments using slit-lamp biomicroscopy, auto refractometer, streak retinoscopy, and subjective manifest refraction were implemented. The data in this article were from the 3-month point.

Optimization of constants and prediction error

In virgin eyes, the constants were optimized for each IOL model, formula, and biometer to make the mean error equal to zero. Formulas were programmed in the Excel software (version 12.0, 2007, Microsoft Corp., WA, USA) by the first author and carefully checked against IOL-Master. Hoffer Q [13], Holladay I [14], and SRK/T [15] were optimized by the function of What If Analysis in the Excel software, and a double linear regression analysis was used for Haigis [16]. However, Constants of SRK II [17] were not optimized because it was considered to be an outdated regression formula [18]. Then, the optimized constants were used in post-LASIK eyes. As for C constant in the Shammas-PL formula, the conversion eq. (C = 0.5835 × A 64.40) was used to calculate it from the optimized A constant of SRK/T [9]. The prediction error was defined as the stable postoperative manifest refraction at BCVA in spherical equivalent (SE) minus predicted SE, which means that a positive prediction error indicated a hyperopic shift. Meanwhile, the absolute error was defined as the absolute value of the prediction error.

Statistical analysis

SPSS (version 22.00, SPSS, Inc., IL, USA) was used for statistical analysis. Standard deviations (SDs) were calculated, and F tests were used for determining the significant differences between formulas and methods. P values less than 0.05 were considered to be statistically significant.

Results

Preoperative data of 176 virgin eyes of 176 patients and 10 post-LASIK eyes of 10 post-LASIK patients were collected. Finally, 133 virgin eyes of 133 patients and 10 post-LASIK eyes of 10 post-LASIK patients were analyzed in this study as a result of losing to follow-up (28 eyes), complications (4 eyes), or postoperative BCVA less than 20/40 (11 eyes).
The demographics, mean values (± SDs), and ranges measured by IOL-Master were as follows: virgin eyes—62 males (46.6%), 71 females (53.4%), 63 oculi dexter (47.4%), 70 oculi sinister (52.6%), 36 eyes (27.1%) implanted 920H IOL, 33 eyes (24.8%) implanted Adapt-AO IOL, 33 eyes (24.8%) implanted SN60WF IOL, 31 eyes (23.3%) implanted ZCB00 IOL, age 68.2 ± 10.3 (range 41.0–89.0), AL 24.42 ± 2.40 mm (range 21.11–33.58 mm), average K 43.93 ± 1.72 D (range 39.88–47.68 D), ACD 3.16 ± 0.44 mm (range 1.97–4.05 mm), and IOL power 18.66 ± 5.81 D (range − 5–27 D); post-LASIK eyes—2 males (20.0%), 8 females (80.0%), 5 oculi dexter (50.0%), 5 oculi sinister (50.0%), 4 eyes (40.0%) implanted 920H IOL, 2 eyes (20.0%) implanted Adapt-AO IOL, 1 eye (10.0%) implanted SN60WF IOL, 3 eyes (30.0%) implanted ZCB00 IOL, age 50.3 ± 9.0 (range 40.0–69.0), AL 30.06 ± 2.87 mm (range 25.46–34.08 mm), average K 36.35 ± 0.77 D (range 35.09–37.84 D), ACD 3.34 ± 0.41 mm (range 2.75–3.86 mm), and IOL power 15.70 ± 6.24 D (range 8–26 D).
Table 1 shows the results for post-LASIK eyes measured by IOL-Master and AL-3000 assisted with TMS-4, and Fig. 1 shows the box plot of the prediction errors. Table 2 shows the optimized IOL constants used in this study.
Table 1
Results of 10 post-LASIK eyes
Method
ME
SD
Range
MAE
MedAE
95% CI
± 0.5 D
± 1.0 D
± 1.5 D
IOL-Master
 Haigis-L
0.01
0.721a
– 1.15–1.23
0.604
0.465
0.361–0.847
6
8
10
 Koch–Maloney (Haigis formula)
– 0.15
0.695
– 1.00–1.19
0.557
0.510
0.312–0.842
5
9
10
 Shammas c.d. (Haigis formula)
0.09
0.695
– 0.80–1.41
0.557
0.470
0.283–0.831
6
9
10
 Shammas c.d. (Shammas-PL formula)
– 0.02
0.698a
– 1.19–0.99
0.558
0.519
0.334–0.822
5
9
10
AL-3000 assisted with TMS-4
 Haigis-L
0.45
0.782
– 0.94–1.37
0.774
0.876
0.480–1.067
4
7
10
 Koch–Maloney (Haigis formula)
0.25
0.741
– 0.88–1.24
0.627
0.781
0.324–0.931
4
8
10
 Shammas c.d. (Haigis formula)
0.49
0.743
– 0.67–1.47
0.734
0.708
0.396–1.072
3
6
10
 Shammas c.d. (Shammas-PL formula)
0.03
0.778
– 1.36–1.13
0.653
0.583
0.392–0.914
4
8
10
ME mean error, SD standard deviation, MAE mean absolute error, MedAE median absolute error, 95% CI, 95% confidence interval of mean absolute error; ±0.5 D, ±1.0 D, and ±1.5 D = number of refractions within 0.5 D, 1.0 D, and 1.5 D of prediction
aWorse than Haigis with measurements using IOL-Master in all 133 virgin eyes (P < 0.05)
Table 2
Intraocular lens constants used for formulas
 
Formula
 
Haigis
Shammas-PL
IOL model
a0
a1
a2
C
IOL-Master
 920H
0.154
0.166
0.166
4.612
 Adapt-AO
−1.418
0.629
0.176
4.710
 SN60WF
1.948
0.376
0.076
4.883
 ZCB00
−3.031
0.331
0.309
5.403
AL-3000 assisted with TMS-4
 920H
−1.068
0.209
0.191
4.176
 Adapt-AO
−0.319
0.563
0.106
4.114
 SN60WF
3.091
0.319
0.009
4.432
 ZCB00
0.632
0.614
0.095
4.931
In post-LASIK eyes, the refractive errors of most eyes were within 1.0 D, and all of them were within 1.5 D. The statistics also showed that the Haigis formula and Shammas-PL formula were interchangeable with Shammas clinically derived K-value correction. Moreover, when measurements were performed using IOL-Master, tiny mean errors (MEs) with little clinical significance were observed, with Koch-Maloney method using Haigis formula tended to result in a little myopic shift. However, applanation ultrasound did not result in ignorable hyperopic surprise except when Shammas c.d. and Shammas-PL formula were used. Significant differences of SDs were found between the results of Haigis-L and Shammas c.d. with the Shammas-PL formula and the results of Haigis in virgin eyes with measurements using IOL-Master. In the data of AL-3000 assisted with TMS-4, no significant difference was found between methods in post-LASIK eyes and Haigis in virgin eyes because of the larger SD of Haigis in virgin eyes.

Discussion

Comparing methods with optimized constants in post-LASIK eyes is of great significance. It is a must to show their true prediction ability and avoid other interference factors. This novel study evaluated methods in post-LASIK eyes with optimized constants under certain circumstances. It is widely recognized that AL measured using applanation ultrasound is shorter than that measured using a partial coherence interferometry (PCI) device [19]. Applanation ultrasound achieved fewer number of eyes within 0.5 D and 1.0 D, although the differences of SDs were not so obvious. Hyperopic surprises in eyes after myopic LASIK have been a concern for many years [11], but now, it is possible to solve this problem using modern methods and PCI devices and even without preoperative data, albeit with larger SDs. A possible explanation for this is that all these corneal power correction approaches assume a one-to-one relation between the measured K-value and corrected K-value and do not take the extent of ablation into consideration. Literature shows that the error is directly proportional to the extent of keratectomy [20]. However, when applanation ultrasound was used, no significant difference was found between virgin eyes and post-LASIK eyes because of the larger SD of Haigis in virgin eyes. It might be due to the poorer reproducibility and repeatability of applanation ultrasound while measuring AL, and weaker correlation when constants of Haigis were optimized. Hyperopic shift was most likely to be minimized with Shammas c.d. and Shammas-PL formula. As they were observed in other three methods, it might be more suitable to combine the Haigis formula with a PCI device than applanation ultrasound in post-LASIK eyes. Nevertheless, as all refractive errors are within 1.5 D, most post-LASIK patients are likely to be acceptable with refractive outcomes.
More importantly, the errorless measurement of corneal power is a premise. The present study found that it was likely to get K-values with great variations when measurements were repeated several times in post-LASIK eyes with decentered ablation. The possible reason is that, sometimes, some of the measurement points are within the ablation area, and sometimes not. In these cases, asking patients to focus stably is important, and corneal topography may help a lot. For them, corneal topography can directly show the ablation area and its relationship with the corneal vertex and the pupil. If the ablation area still includes most area of the pupil, and patients can recall that they were relatively satisfied with the LASIK just after the operation, it indicates that their optical axes are still inside the ablation area. Thus, patients can be asked to focus on another point or reduce the radius of measurement, if necessary, to ensure that all measurement points are within the ablation area and achieve the most accurate measurement.
The results of the present study were comparable to those of many previous studies. Haigis [7] found that the ME of the Haigis-L formula was not significantly different from 0. Chen et al. [10] found through meta-analysis that Shammas c.d. with the Shammas-PL formula outperformed the Haigis-L formula in eyes after laser refractive surgery. Jin et al. [21] found that the Koch-Maloney method achieved lower MAE than Shammas c.d. with the Haigis formula, but the difference was not statistically significant. However, the constants in their study were not optimized for each IOL model. Jin et al. [22] found that when the Koch–Maloney method was used in Chinese eyes, an offset value of −6.2 was more appropriate than −6.1. However, in the present study, the results from IOL-Master did not show the same tendency. The possible reason was that the number of cases in this study and previous studies was relatively small, and the difference between two offsets was tiny.
This study had several limitations. First, four IOL models were used instead of one. The main reason was that if only one IOL model were included, it would leave too few post-LASIK eyes to be analyzed, and the influence on clinical decisions was not the intention of this study. Meanwhile, many scholars believe that it is acceptable provided that constants were optimized for each IOL model. Besides, the study group was relatively small.
Nowadays, more and more sophisticated devices that can gauge more parameters of the cornea, such as Pentacam, are continuously being invented. These devices make it possible to measure corneal power directly in post-LASIK eyes. Future studies will explore whether it is possible to achieve more desirable outcomes when the extent of ablation is taken into account and the corneal power is measured more directly using these devices. The prediction accuracy of post-LASIK eyes and virgin eyes may be the same with advanced algorithm and technology.

Conclusions

In post-LASIK eyes, no-history methods that corrected the corneal power with measurements using IOL-Master are promising, but still a gap in prediction accuracy exists between virgin eyes and post-LASIK eyes. However, when the Haigis formula is used with measurements using applanation ultrasound and corneal topography, it tends to result in hyperopic shifts. The results suggest an acceptable refractive outcome can be achieved in most patients.

Acknowledgements

None.

Funding

None.

Availability of data and materials

The complete data set can be found in the supplementary file (see “Additional file 1” section).

Authors’ contribution

YW designed the study, and performed the examinations of the patients. SL interpreted the patient data regarding the IOL power calculation, and was a major contributor in writing the manuscript. RL also designed the study, and supervised the whole process. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.
Not applicable.
This study was approved by the medical ethics committee of the hospital, and the reference number was “Quick-PJ 2016-10-07”. Informed consent was obtained from all participants in a written from.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Sarwar H, Modi N. Sir Harold Ridley: innovator of cataract surgery. J Perioper Pract. 2014;24:210–2.PubMed Sarwar H, Modi N. Sir Harold Ridley: innovator of cataract surgery. J Perioper Pract. 2014;24:210–2.PubMed
2.
Zurück zum Zitat Nemeth J, Fekete O, Pesztenlehrer N. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg. 2003;29:85–8.CrossRefPubMed Nemeth J, Fekete O, Pesztenlehrer N. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg. 2003;29:85–8.CrossRefPubMed
3.
Zurück zum Zitat Rajan MS, Keilhorn I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations. Eye (Lond). 2002;16:552–6.CrossRef Rajan MS, Keilhorn I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations. Eye (Lond). 2002;16:552–6.CrossRef
4.
Zurück zum Zitat Olsen T, Nielsen PJ. Immersion versus contact technique in the measurement of axial length by ultrasound. Acta Ophthalmol. 1989;67:101–2.CrossRef Olsen T, Nielsen PJ. Immersion versus contact technique in the measurement of axial length by ultrasound. Acta Ophthalmol. 1989;67:101–2.CrossRef
5.
Zurück zum Zitat Masket S, Masket SE. Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation. J Cataract Refract Surg. 2006;32:430–4.CrossRefPubMed Masket S, Masket SE. Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation. J Cataract Refract Surg. 2006;32:430–4.CrossRefPubMed
6.
Zurück zum Zitat Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol Scand. 2007;85:472–85.CrossRefPubMed Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol Scand. 2007;85:472–85.CrossRefPubMed
7.
Zurück zum Zitat Haigis W. Intraocular lens calculation after refractive surgery for myopia: Haigis-L formula. J Cataract Refract Surg. 2008;34:1658–63.CrossRefPubMed Haigis W. Intraocular lens calculation after refractive surgery for myopia: Haigis-L formula. J Cataract Refract Surg. 2008;34:1658–63.CrossRefPubMed
8.
Zurück zum Zitat Wang L, Booth MA, Koch DD. Comparison of intraocular lens power calculation methods in eyes that have undergone LASIK. Ophthalmology. 2004;111:1825–31.CrossRefPubMed Wang L, Booth MA, Koch DD. Comparison of intraocular lens power calculation methods in eyes that have undergone LASIK. Ophthalmology. 2004;111:1825–31.CrossRefPubMed
9.
Zurück zum Zitat Shammas HJ, Shammas MC. No-history method of intraocular lens power calculation for cataract surgery after myopic laser in situ keratomileusis. J Cataract Refract Surg. 2007;33:31–6.CrossRefPubMed Shammas HJ, Shammas MC. No-history method of intraocular lens power calculation for cataract surgery after myopic laser in situ keratomileusis. J Cataract Refract Surg. 2007;33:31–6.CrossRefPubMed
10.
Zurück zum Zitat Chen X, Yuan F, Wu L. Metaanalysis of intraocular lens power calculation after laser refractive surgery in myopic eyes. J Cataract Refract Surg. 2016;42:163–70.CrossRefPubMed Chen X, Yuan F, Wu L. Metaanalysis of intraocular lens power calculation after laser refractive surgery in myopic eyes. J Cataract Refract Surg. 2016;42:163–70.CrossRefPubMed
11.
Zurück zum Zitat Shammas HJ, Shammas MC, Garabet A, Kim JH, Shammas A, LaBree L. Correcting the corneal power measurements for intraocular lens power calculations after myopic laser in situ keratomileusis. Am J Ophthalmol. 2003;136:426–32.CrossRefPubMed Shammas HJ, Shammas MC, Garabet A, Kim JH, Shammas A, LaBree L. Correcting the corneal power measurements for intraocular lens power calculations after myopic laser in situ keratomileusis. Am J Ophthalmol. 2003;136:426–32.CrossRefPubMed
12.
Zurück zum Zitat Rodriguez-Raton A, Jimenez-Alvarez M, Arteche-Limousin L, Mediavilla-Pena E, Larrucea-Martinez I. Effect of pupil dilation on biometry measurements with partial coherence interferometry and its effect on IOL power formula calculation. Eur J Ophthalmol. 2015;25:309–14.CrossRefPubMed Rodriguez-Raton A, Jimenez-Alvarez M, Arteche-Limousin L, Mediavilla-Pena E, Larrucea-Martinez I. Effect of pupil dilation on biometry measurements with partial coherence interferometry and its effect on IOL power formula calculation. Eur J Ophthalmol. 2015;25:309–14.CrossRefPubMed
13.
Zurück zum Zitat Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19:700–12.CrossRefPubMed Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19:700–12.CrossRefPubMed
14.
Zurück zum Zitat Holladay JT, Prager TC, Chandler TY, Musgrove KH, Lewis JW, Ruiz RS. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg. 1988;14:17–24.CrossRefPubMed Holladay JT, Prager TC, Chandler TY, Musgrove KH, Lewis JW, Ruiz RS. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg. 1988;14:17–24.CrossRefPubMed
15.
Zurück zum Zitat Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:333–40.CrossRefPubMed Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:333–40.CrossRefPubMed
16.
Zurück zum Zitat Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238:765–73.CrossRefPubMed Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238:765–73.CrossRefPubMed
17.
Zurück zum Zitat Retzlaff J. A new intraocular lens calculation formula. J Am Intraocul Implant Soc. 1980;6:148–52.CrossRefPubMed Retzlaff J. A new intraocular lens calculation formula. J Am Intraocul Implant Soc. 1980;6:148–52.CrossRefPubMed
18.
Zurück zum Zitat Hoffer K, Aramberri J, Haigis W, Olsen T, Savini G, Shammas H, et al. Protocols for studies of intraocular lens formula accuracy. Am J Ophthalmol. 2015;160:403–5.e1.CrossRefPubMed Hoffer K, Aramberri J, Haigis W, Olsen T, Savini G, Shammas H, et al. Protocols for studies of intraocular lens formula accuracy. Am J Ophthalmol. 2015;160:403–5.e1.CrossRefPubMed
19.
Zurück zum Zitat Goyal R, North RV, Morgan JE. Comparison of laser interferometry and ultrasound A-scan in the measurement of axial length. Acta Ophthalmol Scand. 2003;81:331–5.CrossRefPubMed Goyal R, North RV, Morgan JE. Comparison of laser interferometry and ultrasound A-scan in the measurement of axial length. Acta Ophthalmol Scand. 2003;81:331–5.CrossRefPubMed
20.
Zurück zum Zitat Feiz V. Intraocular lens power calculation after corneal refractive surgery. Middle East Afr J Ophthalmol. 2010;17:63–8.PubMedPubMedCentral Feiz V. Intraocular lens power calculation after corneal refractive surgery. Middle East Afr J Ophthalmol. 2010;17:63–8.PubMedPubMedCentral
21.
Zurück zum Zitat Jin H, Holzer MP, Rabsilber T, Borkenstein AF, Limberger IJ, Guo H, et al. Intraocular lens power calculation after laser refractive surgery: corrective algorithm for corneal power estimation. J Cataract Refract Surg. 2010;36:87–96.CrossRefPubMed Jin H, Holzer MP, Rabsilber T, Borkenstein AF, Limberger IJ, Guo H, et al. Intraocular lens power calculation after laser refractive surgery: corrective algorithm for corneal power estimation. J Cataract Refract Surg. 2010;36:87–96.CrossRefPubMed
22.
Zurück zum Zitat Jin H, Auffarth GU, Guo H, Zhao P. Corneal power estimation for intraocular lens power calculation after corneal laser refractive surgery in Chinese eyes. J Cataract Refract Surg. 2012;38:1749–57.CrossRefPubMed Jin H, Auffarth GU, Guo H, Zhao P. Corneal power estimation for intraocular lens power calculation after corneal laser refractive surgery in Chinese eyes. J Cataract Refract Surg. 2012;38:1749–57.CrossRefPubMed
Metadaten
Titel
Prediction accuracy of intraocular lens power calculation methods after laser refractive surgery
verfasst von
Yubo Wu
Songyu Liu
Rongfeng Liao
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2017
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0439-x

Weitere Artikel der Ausgabe 1/2017

BMC Ophthalmology 1/2017 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.