Impact of motion correction on reproducibility and spatial variability of quantitative myocardial T2 mapping

verfasst von: Sébastien Roujol, Tamer A. Basha, Sebastian Weingärtner, Mehmet Akçakaya, Sophie Berg, Warren J. Manning, Reza Nezafat

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2015

Abstract

Background

To evaluate and quantify the impact of a novel image-based motion correction technique in myocardial T2 mapping in terms of measurement reproducibility and spatial variability.

Methods

Twelve healthy adult subjects were imaged using breath-hold (BH), free breathing (FB), and free breathing with respiratory navigator gating (FB + NAV) myocardial T2 mapping sequences. Fifty patients referred for clinical CMR were imaged using the FB + NAV sequence. All sequences used a T2 prepared (T2prep) steady-state free precession acquisition. In-plane myocardial motion was corrected using an adaptive registration of varying contrast-weighted images for improved tissue characterization (ARCTIC). DICE similarity coefficient (DSC) and myocardial boundary errors (MBE) were measured to quantify the motion estimation accuracy in healthy subjects. T2 mapping reproducibility and spatial variability were evaluated in healthy subjects using 5 repetitions of the FB + NAV sequence with either 4 or 20 T2prep echo times (TE). Subjective T2 map quality was assessed in patients by an experienced reader using a 4-point scale (1-non diagnostic, 4-excellent).

Results

ARCTIC led to increased DSC in BH data (0.85 ± 0.08 vs. 0.90 ± 0.02, p = 0.007), FB data (0.78 ± 0.13 vs. 0.90 ± 0.21, p < 0.001), and FB + NAV data (0.86 ± 0.05 vs. 0.90 ± 0.02, p = 0.002), and reduced MBE in BH data (0.90 ± 0.40 vs. 0.64 ± 0.19 mm, p = 0.005), FB data (1.21 ± 0.65 vs. 0.63 ± 0.10 mm, p < 0.001), and FB + NAV data (0.81 ± 0.21 vs. 0.63 ± 0.08 mm, p < 0.001). Improved reproducibility (4TE: 5.3 ± 2.5 ms vs. 4.0 ± 1.5 ms, p = 0.016; 20TE: 3.9 ± 2.3 ms vs. 2.2 ± 0.5 ms, p = 0.002), reduced spatial variability (4TE: 12.8 ± 3.5 ms vs. 10.3 ± 2.5 ms, p < 0.001; 20TE: 9.7 ± 3.5 ms vs. 7.5 ± 1.4 ms) and improved subjective score of T2 map quality (3.43 ± 0.79 vs. 3.69 ± 0.55, p < 0.001) were obtained using ARCTIC.

Conclusions

The ARCTIC technique substantially reduces spatial mis-alignment among T2-weighted images and improves the reproducibility and spatial variability of in-vivo T2 mapping.
Hinweise

Competing interests

SR, WJM and RN have a pending patent for methods for correcting motion for tissue characterization sequences. TB, MA, WJM, and RN have a pending patent for system and method for assessing T2 relaxation times with improved accuracy.

Authors’ contributions

SR participated in the study design and coordination, developed the ARCTIC approach, carried out the motion correction/reconstruction of the data, performed the data analysis and drafted the manuscript. TB developed the prospective T2 mapping sequence and participated in data acquisition. SW developed the T2 mapping reconstruction code. MA participated in the data acquisition. SB was in charge of subject recruitment. WJM helped in revising the manuscript. RN conceived the study, participated in the study design and interpretation of the data. All authors read and approved the final manuscript.
Abkürzungen
T2prep
T2 prepared
ARCTIC
Adaptive registration of varying contrast-weighted images for improved tissue characterization
BH
Breath-hold
FB
Free breathing
FBNAV
Free breathing conditions with respiratory navigator gating
DSC
DICE similarity coefficient
MBE
Myocardial boundary errors (MBE)
TE
Eecho times
SSFP
Steady-state free precession
HIPAA
Health insurance portability and accountability act
ECG
Electrocardiogram
TET2P
T2prep echo times
TR
Repetition time
FOV
Field of view
SENSE
Sensitivity encoding
SNR
Signal-to-noise ratio
GPU
Graphic processing unit
CUDA
Compute unified device architecture

Background

The T2 relaxation time is dependent on the amount of free water [1] and can be exploited as a potential marker of inflammation and edema [27]. In cardiac MR (CMR), T2 changes are generally assessed using a dark blood T2-weighted acquisition [8]. Elevated signal intensity in T2-weighted images have been reported in presence of several cardiomyopathies such as myocarditis [2, 3], Tako-Tsubo [4], and acute myocardial infarction [57]. However, this technique only provides qualitative measurements and image interpretation can be limited by several factors including regional signal variations induced by phased array coil, elevated signal induced by sub-endocardial stagnant blood, and signal loss caused by through-plane motion [9, 10].
Quantitative myocardial T2 mapping [11, 12] is an alternative technique, which shows promise for reducing uncertainties in interpretations of dark blood T2-weighted images. In this technique, several T2-weighted images are acquired, each with a different T2 contrast. The signal intensity obtained from the T2-weighted images is then fit to a physical model of T2 signal decay on a per-pixel basis, leading to the creation of a T2 map. The acquisition of each T2-weighted image was initially performed using either spin echo/fast spin echo acquisitions [1114] with varying echo times (TE) which results in very long scan time. Recently, T2-prepared (T2prep) [15] steady-state free precession (SSFP) acquisitions have been proposed and provide higher imaging efficiency [16]. These sequences can be acquired within a breath-hold [16, 17] or under free breathing conditions with respiratory motion correction techniques [16, 18, 19].
Despite the promise of this technique, its in-vivo reproducibility and precision have not been fully characterized. These two factors play a major role for clinical acceptance of any quantitative myocardial tissue characterization technique [20, 21]. The presence of motion among T2-weighted images is one of the main challenges in T2 mapping and is expected to have important impact on the technique precision and reproducibility.
Breath-hold acquisitions can be used to reduce the impact of respiratory motion. However, some motion can still be detected in 40-60 % of patients due to their limited breath-holding capabilities, as reported by several T1 mapping studies using breath-held acquisitions of ~11-17 heart beats [2225]. The breath-hold approach imposes severe time limitations on the number of acquired T2-weighted images (typically ~3-4) since a rest time of ~4-6 heart beats is required between each acquisition to allow for full longitudinal magnetization recovery. Therefore, the use of a free breathing acquisition is attractive as it enables the acquisition of a larger number of T2-weighted images which may be beneficial to improve precision and reproducibility. On the other hand, free breathing acquisitions require the use of respiratory navigators to account for through plane motion and image registration algorithms to correct for residual in-plane motion [18].
We recently developed a technique for Adaptive Registration of varying Contrast-weighted images for improved TIssue Characterization (ARCTIC) which we have evaluated for myocardial T1 mapping [23]. In this study, we sought to investigate the performance of ARCTIC for T2 mapping and its impact on in-vivo reproducibility and spatial variability of myocardial T2 estimates.

Methods

All subjects were scanned using a 1.5 T Philips Achieva (Philips Healthcare, Best, The Netherlands) scanner with a 32-channel cardiac phased array receiver coil. This study was health insurance portability and accountability act (HIPAA) compliant and the imaging protocol was approved by our institutional review board (Committee on Clinical Investigations (CCI)) at the Beth Israel Deaconess Medical Center. Written informed consent was obtained from each participant.

T2 mapping acquisition scheme

T2 mapping was performed using our recently reported T2 mapping sequence [26] in which multiple T2-weighted images are acquired using an electrocardiogram (ECG)-triggered T2prep steady-state free precession (SSFP) acquisition with different T2prep echo times (TET2P). A rest cycle of 6 s was used between the acquisitions of two successive T2-weighted images to ensure full re-growth of the longitudinal magnetization. The TET2P = 0 image was acquired using 90° pulse followed immediately by a -90° pulse and a crusher gradient to ensure consistency with all other images in term of longitudinal signal reduction induced by imperfect 90° and -90° flip angles used in the T2prep. Finally, to model the signal re-growth induced by the SSFP imaging pulses, an infinitely long T2prep echo time (TET2P = ∞) was simulated by acquiring an image immediately after a saturation pulse. In this study, this sequence has been evaluated with 4 T2prep echo times (T2P4TE: 0, 25, 50, ∞) and 20 T2prep echo times (T2P20TE: 0, 25, 30, 35, …, 95, 100, ∞, ∞, ∞). For free breathing acquisitions, a respiratory navigator positioned immediately prior to the T2prep was used for end expiratory gating (window size = 5 mm). No T2prep or imaging pulses were applied if the navigator signal was outside the gating window to enable the acquisition of undisturbed signal in the next heartbeat.

In-plane motion correction

The ARCTIC approach was used to compensate for in-plane motion between T2-weighted images [23]. In this approach, all images are registered individually to a common reference image, which was chosen as the first image of the series (TET2P = 0). The motion was then estimated in a two-step process. Affine motion descriptors are first estimated over a region of interest surrounding the heart. This global transformation is then provided as input of a more sophisticated local non-rigid motion estimation step using an extended formulation of the optical flow problem which enables the simultaneous estimation of both motion field and intensity variations on a per-pixel basis [27]. An additional term is used to constrain the motion estimates based on prior automatic tracking of specific feature points in the images [2830]. In this algorithm, both motion field and intensity variation map are solved using an iterative scheme. A multi-resolution approach was used for the local non-rigid motion estimation step where the optical flow is initially estimated from first sub-resolution images and then refined using the full resolution images [31]. For each resolution level, the iterative scheme used 100 iterations and was repeated fifty times. These parameters were empirically optimized in this study. Since optical flow algorithms are well suitable for parallelization on graphic processing unit (GPU) [3234], a GPU implementation of the method was used based on the compute unified device architecture (CUDA). More details about the algorithm can be found in [23].

T2 map reconstruction

T2 maps were reconstructed offline using a 3-parameter curve fitting model.
$$ S\left(A,B,{T}_2,{t}_n\right)=A{e}^{-{\mathrm{t}}_n/{T}_2}+B. $$
(1)
where tn is the T2prep echo time of nth T2-weighted image, and A, B, and T2 are the model parameters. A, B, and T2 are estimated independently for each pixel using a Levenberg-Marquard optimizer with the online library provided in [35].

In-vivo study in healthy subjects

Twelve healthy adult subjects (32 ± 16 years, 6 male) without any history of cardiovascular disease underwent CMR examination. Each subject was imaged using eight T2 mapping sequences in the following order:
1.
Breath-held T2P4TE
 
2.
Free breathing T2P4TE without respiratory navigator
 
3.
Free breathing T2P4TE with respiratory navigator
 
4.
Free breathing T2P20TE with respiratory navigator (5 repetitions).
 
All sequences were acquired in the short axis view using a single-shot ECG-triggered acquisition with SSFP imaging readout and the following parameters: field of view = 240 × 240 mm2, in-plane resolution = 2.5 × 2.5 mm2, slice thickness = 8 mm, TR/TE = 2.7 ms/1.35 ms, flip angle = 85°, 10 linear ramp-up pulses, SENSE rate = 2, acquisition window = 138 ms, number of phase encoding lines = 51, linear k-space ordering. All T2 scans were acquired in the same short axis orientation at the mid-diastolic cardiac phase using one single mid-ventricular slice.
Accuracy of motion correction was evaluated in the first three scans (T2p4TE) by quantifying the motion between the T2-weighted images without (uncorrected) and with in-plane motion correction using ARCTIC (motion corrected). Endocardial and epicardial contours were manually drawn in all T2-weighted images of all T2 mapping scans. The two contours were used to create a binary representation of the myocardium for each T2-weighted image. The DICE similarity coefficient (DSC) [36] was then calculated between the myocardial binary mask of the reference image (Mref) and the myocardial binary mask of each kth T2-weighted image (Mk) as follows:
$$ \mathrm{D}\mathrm{S}\mathrm{C}=\frac{2\times \mathrm{area}\left({\mathrm{M}}_{\mathrm{ref}}\cap {\mathrm{M}}_{\mathrm{k}}\right)}{\mathrm{area}\left({\mathrm{M}}_{\mathrm{ref}}\right)+\mathrm{area}\left({\mathrm{M}}_{\mathrm{k}}\right)} $$
(2)
The myocardial boundary error (MBE), which provides a local alignment measure is also reported. MBE was measured as the average distance between the myocardial boundary of each T2-weighted image (boundary of Mk) and the myocardial boundary of the reference image (boundary of M ref ) as follows:
$$ MBE\left({M}_k,{M}_{ref}\right)=\frac{1}{N}{\displaystyle \sum_{i=1}^N\left\Vert {P}_{M_K}^i,-,{P}_{M_{ref}}^{Closest-i}\right\Vert }2 $$
(3)
Where \( {P}_{M_K}^i \)is the ith point along the boundary of M k , \( {P}_{M_{ref}}^{Closest-i} \) is the closest point of \( {P}_{M_K}^i \) located on the boundary of M ref . Since TET2P = ∞ images are very low signal-to-noise ratio (SNR) images, which makes the detection of the myocardial borders very difficult, no DSC/MBE were measured in those images. The statistical significant difference between DSCs (and MBEs) obtained with and without motion correction was evaluated using Wilcoxon signed rank tests. Statistical significance was considered at p < 0.05.
The impact of in-plane motion correction on the reproducibility and spatial variability of T2 mapping was evaluated using the five T2P20TE scans. For each scan, T2 maps were reconstructed without (uncorrected) and with prior in-plane motion correction using ARCTIC (motion corrected). The endocardial and epicardial border of the myocardium and the insertion point were manually drawn on each T2 map. A six myocardial segment model [37] was automatically created for each single slice (1:anterior, 2:anterospetal, 3:inferospetal, 4:inferior, 5:inferolateral, 6:anterolateral). Segment-based analysis of reproducibility and spatial variability of T2 estimates was then performed. Spatial variability was defined as the standard deviation of T2 estimates over a given segment. Reproducibility was defined as the standard deviation over the 5 scans of the spatial average T2 values in one given segment. Both reproducibility and spatial variability are reported in average over all segments for each subject, and in average over all subjects for each segment. To investigate the motion influence in T2 mapping sequences using a limited number of T2prep echo times, this overall analysis was repeated using a subset of the T2-weighted images from each scan (4 T2prep echo times of 0, 25, 50, ∞). The statistical significant difference between uncorrected and ARCTIC motion corrected T2 reproducibility (and spatial variability) measured for each subject (in average over all myocardial segments) was evaluated using Wilcoxon signed rank tests.

In-vivo study in patients

Fifty patients referred for clinical CMR (56 ± 14 y, 29 male) were imaged using the free breathing T2P4TE T2 mapping sequence with respiratory navigator. All sequences were acquired in the short axis view using a single-shot ECG-triggered acquisition with SSFP imaging readout and the following parameters: field of view = 360 × 360 mm2, in-plane resolution = 2 × 2 mm2, slice thickness = 8 mm, slice number = 3, TR/TE = 2.9 ms/1.45 ms, flip angle = 85°, 10 linear ramp-up pulses, SENSE rate = 2, acquisition window = 270 ms, number of phase encoding lines = 93, linear k-space ordering. T2 maps were reconstructed without and with ARCTIC motion correction.
A subjective qualitative analysis was performed by an experienced cardiologist. The initial motion level in uncorrected data was assessed for each slice as “no motion”, “small motion”, or “large motion” by visual inspection of all uncorrected T2-weighted images. Subjective assessment of uncorrected and motion correction T2 maps (150 T2 maps) followed. Each pair of uncorrected and motion correction T2 maps were shown simultaneously to the reader side by side in a random order. The reader was blinded to the reconstruction approach (uncorrected vs. motion corrected). Each map was assessed in term of overall quality (1-non diagnostic/large artifacts/no confidence in interpreting T2 values in more than half of the myocardial segments, 2-fair/moderate artifacts/confidence in interpreting T2 values in more than half of the myocardial segments, 3-good/small motion artifacts/no confidence in interpreting T2 values in at most one myocardial segment, 4-excellent/no motion artifact/confidence in interpreting T2 values in all myocardial segments). Furthermore, for each pair of T2 maps, the reader was asked to evaluate if any of the two T2 map had “1-inferior”, “2-similar”, or “3-superior” quality. Wilcoxon signed rank test was used to test the null hypothesis that the difference of overall T2 map quality scores between uncorrected and motion corrected T2 maps was zero. Statistical significance was considered at p < 0.05.

Results

All scans were successful. The nominal scan time (assuming 100% gating efficiency) corresponded to 13 heart beats for the T2P4TE sequence and to 99 heart beats for the T2P20TE sequence. The employed ARCTIC motion correction and reconstruction of one T2 map with 20 T2prep echo times was 20s.
Figure 1 shows an example of the remaining in-plane motion between T2-weighted images acquired in one healthy subject using the T2P4TE sequence under breath-hold, free breathing, and free breathing with respiratory navigator gating. Motion artifacts can be observed in the reconstructed T2 maps (see white arrows). In-plane motion correction improves the spatial alignment of T2-weighted images and results in visually improved T2 map quality (Figure 1).
Figure  2 shows quantitative metrics of motion accuracy (DSC and MBE) obtained in healthy subjects using the three aforementioned acquisition sequences. Increased DSC and reduced MBE were observed in each of the three acquisition sequences. In the remaining part of this paragraph, DSC and MBE are reported as (uncorrected data vs. motion corrected data using ARCTIC). On average for all subjects, the DSC increased in breath-hold data (0.85 ± 0.08 vs. 0.90 ± 0.02, p = 0.007), free breathing data (0.78 ± 0.13 vs. 0.90 ± 0.21, p < 0.001), and free breathing data with respiratory navigator gating (0.86 ± 0.05 vs. 0.90 ± 0.02, p = 0.002). The MBE decreased in breath-hold data (0.90 ± 0.40 vs. 0.64 ± 0.19 mm, p = 0.005), free breathing data (1.21 ± 0.65 vs. 0.63 ± 0.10 mm, p < 0.001), and free breathing data with respiratory navigator gating (0.81 ± 0.21 vs. 0.63 ± 0.08 mm, p < 0.001).
Figure  3 shows an example of multiple T2 maps obtained in one healthy subject using the T2P20TE sequence acquired under free breathing conditions with respiratory navigator gating. T2 maps are shown when reconstructed from only 4 T2prep echo times and from all 20 T2prep echo times. The level of artifacts in uncorrected T2 maps appears higher than in motion corrected T2 maps (see white arrows). As expected, motion artifact patterns have high spatial variability in uncorrected T2 maps. Furthermore, the spatial variability of the myocardial T2 estimates appears well reduced when using all 20 T2prep echo times compared to only 4 T2prep echo times.
Figures 4 and 5 summarize the reproducibility and spatial variability of T2 measurements obtained in healthy subjects using the T2P20TE sequence. Results are shown for uncorrected and motion corrected T2 maps reconstructed using either 4 T2prep echo times or 20 T2prep echo times. Reproducibility and spatial variability are reported as uncorrected T2 maps vs. motion corrected T2 maps using ARCTIC. Improved reproducibility was observed over all subjects and myocardial segments in T2 maps reconstructed from 4 T2prep echo times (5.3 ± 2.5 ms vs. 4.0 ± 1.5 ms, p = 0.016) and 20 T2prep echo times (3.9 ± 2.3 ms vs. 2.2 ± 0.5 ms, p = 0.002). Similarly, reduced spatial variability was observed over all subjects and myocardial segments in T2 maps reconstructed from 4 T2prep echo times (12.8 ± 3.5 ms vs. 10.3 ± 2.5 ms, p < 0.001) and 20 T2prep echo times (9.7 ± 3.5 ms vs. 7.5 ± 1.4 ms, p = 0.005).
As expected, T2 maps reconstructed using 20 T2prep echo times had better reproducibility than those reconstructed using only 4 T2prep echo times in both uncorrected data (3.9 ± 2.3 ms vs. 5.3 ± 2.5 ms, respectively, p = 0.007) and motion corrected (2.2 ± 0.5 ms vs. 4.0 ± 1.5 ms, respectively, p < 0.001). The spatial variability of myocardial T2 estimates reconstructed using 20 T2prep echo times was also lower than the one obtained with 4 T2prep echo times in both uncorrected data (9.7 ± 3.5 ms vs. 12.8 ± 3.5 ms, respectively, p < 0.001) and motion corrected data (7.5 ± 1.4 ms vs. 10.3 ± 2.5 ms, respectively, p < 0.001).
Figure 6 shows example uncorrected and ARCTIC motion corrected T2 maps obtained in four patients. Large regional variations and artifacts can be observed in uncorrected T2 maps (see white arrows). The proposed ARCTIC motion correction substantially improved the T2 map quality in all 4 patients.
Figure  7 shows the subjective assessment of T2 map quality obtained in 50 patients. Overall (N = 150 T2 maps), ARCTIC motion corrected T2 maps had higher quality score than uncorrected T2 maps (3.69 ± 0.55 vs. 3.43 ± 0.79, p < 0.001). In the relative comparison of T2 map quality, uncorrected T2 maps has superior, similar, and inferior quality than ARCTIC motion corrected T2 maps in 4 maps (3 %), 99 maps (66 %), and 47 maps (31 %), respectively. Furthermore, the motion level was assessed as “no motion” in 35 slices (23%), “small motion” in 69 slices (46%), and “large motion” in 46 slices (30%). In “no motion” data, all ARCTIC motion corrected and uncorrected T2 maps received a subjective quality score of 4.0 and 97 % of them had similar relative quality. In “small motion” data, ARCTIC motion corrected T2 maps had higher subjective quality score (3.71 ± 0.49 vs. 3.61 ± 0.60, p = 0.015) and superior (23%), similar (75%) and inferior (1%) relative quality than uncorrected T2 maps. In “large motion” data, ARCTIC motion corrected T2 maps had higher subjective quality score (3.41 ± 0.69 vs. 2.72 ± 0.83, p < 0.001) and superior (65%), similar (28%) and inferior (6%) relative quality than uncorrected T2 maps.

Discussion

In this study, we demonstrate the benefit of in-vivo in-plane ARCTIC motion correction in myocardial T2 mapping. The method provides improved alignment of the myocardium in T2-weighted images acquired with breath-hold acquisitions and free breathing acquisitions with and without respiratory navigator gating. ARCTIC motion correction improves T2 map quality which results in improved reproducibility and spatial variability of myocardial T2 estimates. Finally, the CPU/GPU implementation of ARCTIC substantially reduces the computation time of the T2 map reconstruction to 20s which is suitable for clinical applicability.
DSCs and MBEs found in this study are in good agreement with previous studies [2325, 38]. As expected higher mis-alignments were observed using free breathing acquisitions without respiratory navigator gating. DSCs/MBEs improvement was obtained in all three types of acquisitions. This confirms the benefit of motion correction, even for data acquired with a breath-hold. This is likely because 40-60% of patients fail to sustain a stable breath-hold in these conditions [2325]. Furthermore, similar DSCs/MBEs were obtained after motion correction using the three acquisition conditions (breath-hold and free breathing with and without respiratory navigator). It is important to note that through-plane motion cannot be compensated when using the free breathing acquisition without respiratory navigator gating. In this case, the efficacy of in-plane motion correction algorithms depends on the subject’s heart orientation in relation to his respiratory movement. The use of respiratory navigator appears thus desirable to enable through plane motion compensation in free breathing acquisitions. The registration accuracy was not evaluated in the TET2P = ∞ images since the contrast is too low to identify the myocardium. Motion correction is expected to have slightly lower accuracy in those images due to the expected limited ability to compensate for complex motion.
The ARCTIC approach successfully corrected the encountered motion in all subjects. In this study, the heart motion patterns were mainly influenced by the breathing activity of the subjects and to lesser extent to their RR-interval variations. However, the motion pattern can be more complex in patients imaged during arrhythmic events. The performance of the method in such conditions was not investigated and should be addressed in future work.
The reproducibility and spatial variability of T2 mapping was improved using ARCTIC. The use of 20 T2-weighted images improved the reproducibility and the spatial variability of T2 mapping (over the use of only 4 T2-weighted images) by a factor of 2 and 1.4, respectively. Therefore, the choice of the number of T2prep echo times depends on the desired trade-off between acquisition time and T2 map quality. Further studies are warranted to determine the clinically relevant threshold providing satisfactory T2 map quality in a reasonable amount of time.
Reproducibility and spatial variability of T2 estimates were found similar in all myocardial segments when using 20 TEs. However, slight differences seemed to be observed when using 4TEs only, especially in the myocardial segment #4 (inferior wall). Several factors could have contributed to this observation including 1) increased sensitivity to cardiac motion and partial voluming in the free wall due to reduced wall thickness, 2) increased field inhomogeneity in myocardial segments located at the heart/lung interface. Future studies are warranted to study the impact of each of these factors.
In this study, the data were acquired using our recently developed T2 mapping sequence. The ARCTIC approach is expected to provide similar motion correction performance using other T2 mapping sequences. Nevertheless, the impact of motion correction on the reproducibility and spatial variability of other T2 mapping sequences may be different and is beyond the scope of this study. Furthermore, all data were acquired in 2D. 3D myocardial T2 mapping may represent a valuable approach for true 3D assessment of pathological tissues [39, 40]. The extension of the ARCTIC approach to 3D is straightforward and is expected to provide similar improvement of the reproducibility and spatial variability of 3D T2 mapping.
There are several limitations in this study. In the in-vivo analysis of reproducibility and spatial variability, the 4TEs data were extracted from the T2P20TE sequence and were thus not acquired using the T2P4TE sequence. However, since the T2P20TE sequence was acquiring with respiratory gating, the potential bias in reproducibility and spatial variability obtained in the 4TEs data should have been kept to the minimum. Finally, the study was only performed in healthy adult subjects with limited sample size. Further studies are warranted to confirm the benefit of the ARCTIC approach to improve the reproducibility and spatial variability of myocardial T2 mapping in patients.

Conclusions

The ARCTIC technique substantially reduces spatial mis-alignment among T2-weighted images. This method improves the reproducibility and reduces the spatial variability of in-vivo T2 mapping. Furthermore, the in-vivo reproducibility and spatial variability of T2 mapping is improved using a higher number of T2prep echo times combined with ARCTIC motion correction.

Acknowledgements

The project described was partially supported by NIH R01EB008743-01A2, and Samsung Electronics, Suwon, South Korea.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.
The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.
The Creative Commons Public Domain Dedication waiver (https://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Competing interests

SR, WJM and RN have a pending patent for methods for correcting motion for tissue characterization sequences. TB, MA, WJM, and RN have a pending patent for system and method for assessing T2 relaxation times with improved accuracy.

Authors’ contributions

SR participated in the study design and coordination, developed the ARCTIC approach, carried out the motion correction/reconstruction of the data, performed the data analysis and drafted the manuscript. TB developed the prospective T2 mapping sequence and participated in data acquisition. SW developed the T2 mapping reconstruction code. MA participated in the data acquisition. SB was in charge of subject recruitment. WJM helped in revising the manuscript. RN conceived the study, participated in the study design and interpretation of the data. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Higgins CB, Herfkens R, Lipton MJ, Sievers R, Sheldon P, Kaufman L, et al. Nuclear magnetic resonance imaging of acute myocardial infarction in dogs: alterations in magnetic relaxation times. Am J Cardiol. 1983;52(1):184–8.PubMedCrossRef Higgins CB, Herfkens R, Lipton MJ, Sievers R, Sheldon P, Kaufman L, et al. Nuclear magnetic resonance imaging of acute myocardial infarction in dogs: alterations in magnetic relaxation times. Am J Cardiol. 1983;52(1):184–8.PubMedCrossRef
2.
Zurück zum Zitat Abdel-Aty H, Boye P, Zagrosek A, Wassmuth R, Kumar A, Messroghli D, et al. Diagnostic performance of cardiovascular magnetic resonance in patients with suspected acute myocarditis: comparison of different approaches. J Am Coll Cardiol. 2005;45(11):1815–22.PubMedCrossRef Abdel-Aty H, Boye P, Zagrosek A, Wassmuth R, Kumar A, Messroghli D, et al. Diagnostic performance of cardiovascular magnetic resonance in patients with suspected acute myocarditis: comparison of different approaches. J Am Coll Cardiol. 2005;45(11):1815–22.PubMedCrossRef
3.
Zurück zum Zitat Friedrich MG, Sechtem U, Schulz-Menger J, Holmvang G, Alakija P, Cooper LT, et al. Cardiovascular magnetic resonance in myocarditis: A JACC White Paper. J Am Coll Cardiol. 2009;53(17):1475–87.PubMedCentralPubMedCrossRef Friedrich MG, Sechtem U, Schulz-Menger J, Holmvang G, Alakija P, Cooper LT, et al. Cardiovascular magnetic resonance in myocarditis: A JACC White Paper. J Am Coll Cardiol. 2009;53(17):1475–87.PubMedCentralPubMedCrossRef
4.
Zurück zum Zitat Abdel-Aty H, Cocker M, Friedrich MG. Myocardial edema is a feature of Tako-Tsubo cardiomyopathy and is related to the severity of systolic dysfunction: insights from T2-weighted cardiovascular magnetic resonance. Int J Cardiol. 2009;132(2):291–3.PubMedCrossRef Abdel-Aty H, Cocker M, Friedrich MG. Myocardial edema is a feature of Tako-Tsubo cardiomyopathy and is related to the severity of systolic dysfunction: insights from T2-weighted cardiovascular magnetic resonance. Int J Cardiol. 2009;132(2):291–3.PubMedCrossRef
5.
Zurück zum Zitat Abdel-Aty H, Zagrosek A, Schulz-Menger J, Taylor AJ, Messroghli D, Kumar A, et al. Delayed enhancement and T2-weighted cardiovascular magnetic resonance imaging differentiate acute from chronic myocardial infarction. Circulation. 2004;109(20):2411–6.PubMedCrossRef Abdel-Aty H, Zagrosek A, Schulz-Menger J, Taylor AJ, Messroghli D, Kumar A, et al. Delayed enhancement and T2-weighted cardiovascular magnetic resonance imaging differentiate acute from chronic myocardial infarction. Circulation. 2004;109(20):2411–6.PubMedCrossRef
6.
Zurück zum Zitat Raman SV, Simonetti OP, Winner 3rd MW, Dickerson JA, He X, Mazzaferri Jr EL, et al. Cardiac magnetic resonance with edema imaging identifies myocardium at risk and predicts worse outcome in patients with non-ST-segment elevation acute coronary syndrome. J Am Coll Cardiol. 2010;55(22):2480–8.PubMedCentralPubMedCrossRef Raman SV, Simonetti OP, Winner 3rd MW, Dickerson JA, He X, Mazzaferri Jr EL, et al. Cardiac magnetic resonance with edema imaging identifies myocardium at risk and predicts worse outcome in patients with non-ST-segment elevation acute coronary syndrome. J Am Coll Cardiol. 2010;55(22):2480–8.PubMedCentralPubMedCrossRef
7.
Zurück zum Zitat Cury RC, Shash K, Nagurney JT, Rosito G, Shapiro MD, Nomura CH, et al. Cardiac magnetic resonance with T2-weighted imaging improves detection of patients with acute coronary syndrome in the emergency department. Circulation. 2008;118(8):837–44.PubMedCrossRef Cury RC, Shash K, Nagurney JT, Rosito G, Shapiro MD, Nomura CH, et al. Cardiac magnetic resonance with T2-weighted imaging improves detection of patients with acute coronary syndrome in the emergency department. Circulation. 2008;118(8):837–44.PubMedCrossRef
8.
Zurück zum Zitat Simonetti OP, Finn JP, White RD, Laub G, Henry DA. "Black blood" T2-weighted inversion-recovery MR imaging of the heart. Radiology. 1996;199(1):49–57.PubMedCrossRef Simonetti OP, Finn JP, White RD, Laub G, Henry DA. "Black blood" T2-weighted inversion-recovery MR imaging of the heart. Radiology. 1996;199(1):49–57.PubMedCrossRef
9.
Zurück zum Zitat Abdel-Aty H, Simonetti O, Friedrich MG. T2-weighted cardiovascular magnetic resonance imaging. J Magn Reson Imaging. 2007;26(3):452–9.PubMedCrossRef Abdel-Aty H, Simonetti O, Friedrich MG. T2-weighted cardiovascular magnetic resonance imaging. J Magn Reson Imaging. 2007;26(3):452–9.PubMedCrossRef
10.
Zurück zum Zitat Arai AE. Using magnetic resonance imaging to characterize recent myocardial injury: utility in acute coronary syndrome and other clinical scenarios. Circulation. 2008;118(8):795–6.PubMedCentralPubMedCrossRef Arai AE. Using magnetic resonance imaging to characterize recent myocardial injury: utility in acute coronary syndrome and other clinical scenarios. Circulation. 2008;118(8):795–6.PubMedCentralPubMedCrossRef
11.
Zurück zum Zitat Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM. A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1-100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med Phys. 1984;11(4):425–48.PubMedCrossRef Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM. A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1-100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med Phys. 1984;11(4):425–48.PubMedCrossRef
12.
Zurück zum Zitat McNamara MT, Higgins CB, Schechtmann N, Botvinick E, Lipton MJ, Chatterjee K, et al. Detection and characterization of acute myocardial infarction in man with use of gated magnetic resonance. Circulation. 1985;71(4):717–24.PubMedCrossRef McNamara MT, Higgins CB, Schechtmann N, Botvinick E, Lipton MJ, Chatterjee K, et al. Detection and characterization of acute myocardial infarction in man with use of gated magnetic resonance. Circulation. 1985;71(4):717–24.PubMedCrossRef
13.
Zurück zum Zitat Foltz WD, Stainsby JA, Wright GA. T2 accuracy on a whole-body imager. Magn Reson Med. 1997;38(5):759–68.PubMedCrossRef Foltz WD, Stainsby JA, Wright GA. T2 accuracy on a whole-body imager. Magn Reson Med. 1997;38(5):759–68.PubMedCrossRef
14.
Zurück zum Zitat He T, Gatehouse PD, Anderson LJ, Tanner M, Keegan J, Pennell DJ, et al. Development of a novel optimized breathhold technique for myocardial T2 measurement in thalassemia. J Magn Reson Imaging. 2006;24(3):580–5.PubMedCrossRef He T, Gatehouse PD, Anderson LJ, Tanner M, Keegan J, Pennell DJ, et al. Development of a novel optimized breathhold technique for myocardial T2 measurement in thalassemia. J Magn Reson Imaging. 2006;24(3):580–5.PubMedCrossRef
15.
Zurück zum Zitat Brittain JH, Hu BS, Wright GA, Meyer CH, Macovski A, Nishimura DG. Coronary angiography with magnetization-prepared T2 contrast. Magn Reson Med. 1995;33(5):689–96.PubMedCrossRef Brittain JH, Hu BS, Wright GA, Meyer CH, Macovski A, Nishimura DG. Coronary angiography with magnetization-prepared T2 contrast. Magn Reson Med. 1995;33(5):689–96.PubMedCrossRef
16.
Zurück zum Zitat Huang TY, Liu YJ, Stemmer A, Poncelet BP. T2 measurement of the human myocardium using a T2-prepared transient-state TrueFISP sequence. Magn Reson Med. 2007;57(5):960–6.PubMedCrossRef Huang TY, Liu YJ, Stemmer A, Poncelet BP. T2 measurement of the human myocardium using a T2-prepared transient-state TrueFISP sequence. Magn Reson Med. 2007;57(5):960–6.PubMedCrossRef
17.
Zurück zum Zitat Giri S, Chung YC, Merchant A, Mihai G, Rajagopalan S, Raman SV, et al. T2 quantification for improved detection of myocardial edema. J Cardiovasc Magn Reson. 2009;11:56.PubMedCentralPubMedCrossRef Giri S, Chung YC, Merchant A, Mihai G, Rajagopalan S, Raman SV, et al. T2 quantification for improved detection of myocardial edema. J Cardiovasc Magn Reson. 2009;11:56.PubMedCentralPubMedCrossRef
18.
Zurück zum Zitat Giri S, Shah S, Xue H, Chung YC, Pennell ML, Guehring J, et al. Myocardial T(2) mapping with respiratory navigator and automatic nonrigid motion correction. Magn Reson Med. 2012;68(5):1570–8.PubMedCrossRef Giri S, Shah S, Xue H, Chung YC, Pennell ML, Guehring J, et al. Myocardial T(2) mapping with respiratory navigator and automatic nonrigid motion correction. Magn Reson Med. 2012;68(5):1570–8.PubMedCrossRef
19.
Zurück zum Zitat Blume U, Lockie T, Stehning C, Sinclair S, Uribe S, Razavi R, et al. Interleaved T(1) and T(2) relaxation time mapping for cardiac applications. J Magn Reson Imaging. 2009;29(2):480–7.PubMedCrossRef Blume U, Lockie T, Stehning C, Sinclair S, Uribe S, Razavi R, et al. Interleaved T(1) and T(2) relaxation time mapping for cardiac applications. J Magn Reson Imaging. 2009;29(2):480–7.PubMedCrossRef
21.
Zurück zum Zitat Piechnik SK, Ferreira VM, Lewandowski AJ, Ntusi NA, Banerjee R, Holloway C, et al. Normal variation of magnetic resonance T1 relaxation times in the human population at 1.5 T using ShMOLLI. J Cardiovasc Magn Reson. 2013;15:13.PubMedCentralPubMedCrossRef Piechnik SK, Ferreira VM, Lewandowski AJ, Ntusi NA, Banerjee R, Holloway C, et al. Normal variation of magnetic resonance T1 relaxation times in the human population at 1.5 T using ShMOLLI. J Cardiovasc Magn Reson. 2013;15:13.PubMedCentralPubMedCrossRef
22.
Zurück zum Zitat Cheng ASH, Pegg TJ, Karamitsos TD, Searle N, Jerosch-Herold M, Choudhury RP, et al. Cardiovascular magnetic resonance perfusion imaging at 3-tesla for the detection of coronary artery disease. J Am Coll Cardiol. 2007;49(25):2440–9.PubMedCrossRef Cheng ASH, Pegg TJ, Karamitsos TD, Searle N, Jerosch-Herold M, Choudhury RP, et al. Cardiovascular magnetic resonance perfusion imaging at 3-tesla for the detection of coronary artery disease. J Am Coll Cardiol. 2007;49(25):2440–9.PubMedCrossRef
23.
Zurück zum Zitat Roujol S, Foppa M, Weingartner S, Manning WJ, Nezafat R. Adaptive registration of varying contrast-weighted images for improved tissue characterization (ARCTIC): application to T1 mapping. Magn Reson Med. 2015;73(4):1469-82. Roujol S, Foppa M, Weingartner S, Manning WJ, Nezafat R. Adaptive registration of varying contrast-weighted images for improved tissue characterization (ARCTIC): application to T1 mapping. Magn Reson Med. 2015;73(4):1469-82.
24.
Zurück zum Zitat Xue H, Greiser A, Zuehlsdorff S, Jolly MP, Guehring J, Arai AE, et al. Phase-sensitive inversion recovery for myocardial T1 mapping with motion correction and parametric fitting. Magn Reson Med. 2013;69(5):1408–20.PubMedCentralPubMedCrossRef Xue H, Greiser A, Zuehlsdorff S, Jolly MP, Guehring J, Arai AE, et al. Phase-sensitive inversion recovery for myocardial T1 mapping with motion correction and parametric fitting. Magn Reson Med. 2013;69(5):1408–20.PubMedCentralPubMedCrossRef
25.
Zurück zum Zitat Xue H, Shah S, Greiser A, Guetter C, Littmann A, Jolly MP, et al. Motion correction for myocardial T1 mapping using image registration with synthetic image estimation. Magn Reson Med. 2012;67(6):1644–55.PubMedCentralPubMedCrossRef Xue H, Shah S, Greiser A, Guetter C, Littmann A, Jolly MP, et al. Motion correction for myocardial T1 mapping using image registration with synthetic image estimation. Magn Reson Med. 2012;67(6):1644–55.PubMedCentralPubMedCrossRef
26.
Zurück zum Zitat Akçakaya M, Basha TA, Weingärtner S, Roujol S, Berg S, Nezafat R. Improved quantitative myocardial T2 mapping. Mag Reson Med 2014, In Press. doi:10.1002/mrm.25377. Akçakaya M, Basha TA, Weingärtner S, Roujol S, Berg S, Nezafat R. Improved quantitative myocardial T2 mapping. Mag Reson Med 2014, In Press. doi:10.​1002/​mrm.​25377.
27.
Zurück zum Zitat Cornelius N, Kanade T. Adapting optical-flow to measure object motion in reflectance and X-ray image sequences. ACM SIGGRAPH Comput Graph. 1984;18:24–5.CrossRef Cornelius N, Kanade T. Adapting optical-flow to measure object motion in reflectance and X-ray image sequences. ACM SIGGRAPH Comput Graph. 1984;18:24–5.CrossRef
28.
Zurück zum Zitat Butler CR, Thompson R, Haykowsky M, Toma M, Paterson I. Cardiovascular magnetic resonance in the diagnosis of acute heart transplant rejection: a review. J Cardiovasc Magn Reson. 2009;11:7.PubMedCentralPubMedCrossRef Butler CR, Thompson R, Haykowsky M, Toma M, Paterson I. Cardiovascular magnetic resonance in the diagnosis of acute heart transplant rejection: a review. J Cardiovasc Magn Reson. 2009;11:7.PubMedCentralPubMedCrossRef
29.
Zurück zum Zitat Roujol S, Benois-Pineau J, de Senneville BD, Quesson B, Ries M, Moonen C. Real time constrained motion estimation for ECG-gated cardiac MRI. 2010. IEEE. p 757-760. Roujol S, Benois-Pineau J, de Senneville BD, Quesson B, Ries M, Moonen C. Real time constrained motion estimation for ECG-gated cardiac MRI. 2010. IEEE. p 757-760.
30.
Zurück zum Zitat Roujol S, Benois-Pineau J, de Senneville BD, Ries M, Quesson B, Moonen CT. Robust real-time-constrained estimation of respiratory motion for interventional MRI on mobile organs. IEEE Trans Inf Technol Biomed. 2012;16(3):365–74.PubMedCrossRef Roujol S, Benois-Pineau J, de Senneville BD, Ries M, Quesson B, Moonen CT. Robust real-time-constrained estimation of respiratory motion for interventional MRI on mobile organs. IEEE Trans Inf Technol Biomed. 2012;16(3):365–74.PubMedCrossRef
31.
Zurück zum Zitat Pratikakis I, Barillot C, Hellier P, Memin E. Robust multiscale deformable registration of 3D ultrasound images. International Journal of Image and Graphics. 2003;3(04):547–65.CrossRef Pratikakis I, Barillot C, Hellier P, Memin E. Robust multiscale deformable registration of 3D ultrasound images. International Journal of Image and Graphics. 2003;3(04):547–65.CrossRef
32.
Zurück zum Zitat de Senneville BD, Noe KO, Ries M, Pedersen M, Moonen CT, Sorensen T. An optimised multi-baseline approach for on-line MR-temperature monitoring on commodity graphics hardware. 2008. IEEE. p 1513-1516. de Senneville BD, Noe KO, Ries M, Pedersen M, Moonen CT, Sorensen T. An optimised multi-baseline approach for on-line MR-temperature monitoring on commodity graphics hardware. 2008. IEEE. p 1513-1516.
33.
Zurück zum Zitat Ostergaard Noe K, De Senneville BD, Elstrom UV, Tanderup K, Sorensen TS. Acceleration and validation of optical flow based deformable registration for image-guided radiotherapy. Acta Oncol. 2008;47(7):1286–93.PubMedCrossRef Ostergaard Noe K, De Senneville BD, Elstrom UV, Tanderup K, Sorensen TS. Acceleration and validation of optical flow based deformable registration for image-guided radiotherapy. Acta Oncol. 2008;47(7):1286–93.PubMedCrossRef
34.
Zurück zum Zitat Roujol S, Ries M, Quesson B, Moonen C, Denis de Senneville B. Real-time MR-thermometry and dosimetry for interventional guidance on abdominal organs. Magn Reson Med. 2010;63(4):1080–7.PubMedCrossRef Roujol S, Ries M, Quesson B, Moonen C, Denis de Senneville B. Real-time MR-thermometry and dosimetry for interventional guidance on abdominal organs. Magn Reson Med. 2010;63(4):1080–7.PubMedCrossRef
36.
Zurück zum Zitat Dice LR. Measures of the amount of ecologic association between species. Ecology. 1945;26(3):297–302.CrossRef Dice LR. Measures of the amount of ecologic association between species. Ecology. 1945;26(3):297–302.CrossRef
37.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association Circulation. 2002;105(4):539–42. Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association Circulation. 2002;105(4):539–42.
38.
Zurück zum Zitat Cheng C, Herfkens R, Taylor C. Inferior vena caval hemody- namics quantified in vivo at rest and during cycling exercise using magnetic resonance imaging. Am J Physiol Heart Circ Physiol. 2003;284(4):H1161–1167.PubMedCrossRef Cheng C, Herfkens R, Taylor C. Inferior vena caval hemody- namics quantified in vivo at rest and during cycling exercise using magnetic resonance imaging. Am J Physiol Heart Circ Physiol. 2003;284(4):H1161–1167.PubMedCrossRef
39.
Zurück zum Zitat Ding H, Schär M, Zviman M, Halperin HR, Beinart R, Herzka DA. High-resolution quantitative 3D T2 mapping allows quantification of changes in edema after myocardial infarction. J Cardiovasc Magn Reson. 2013;15 Suppl 1:181. Ding H, Schär M, Zviman M, Halperin HR, Beinart R, Herzka DA. High-resolution quantitative 3D T2 mapping allows quantification of changes in edema after myocardial infarction. J Cardiovasc Magn Reson. 2013;15 Suppl 1:181.
40.
Zurück zum Zitat Heeswijk RB, Piccini D, Feliciano H, Hullin R, Schwitter J, Stuber M. Self-navigated isotropic three-dimensional cardiac T2 mapping. Magn Reson Med. 2015;73(4):1549-54. Heeswijk RB, Piccini D, Feliciano H, Hullin R, Schwitter J, Stuber M. Self-navigated isotropic three-dimensional cardiac T2 mapping. Magn Reson Med. 2015;73(4):1549-54.
Metadaten
Titel
Impact of motion correction on reproducibility and spatial variability of quantitative myocardial T2 mapping
verfasst von
Sébastien Roujol
Tamer A. Basha
Sebastian Weingärtner
Mehmet Akçakaya
Sophie Berg
Warren J. Manning
Reza Nezafat
Publikationsdatum
01.12.2015
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2015
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-015-0141-1

Weitere Artikel der Ausgabe 1/2015

Journal of Cardiovascular Magnetic Resonance 1/2015 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.