Systolic MOLLI T1 mapping with heart-rate-dependent pulse sequence sampling scheme is feasible in patients with atrial fibrillation

verfasst von: Lei Zhao, Songnan Li, Xiaohai Ma, Andreas Greiser, Tianjing Zhang, Jing An, Rong Bai, Jianzeng Dong, Zhanming Fan

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2016

Abstract

Background

T1 mapping enables assessment of myocardial characteristics. As the most common type of arrhythmia, atrial fibrillation (AF) is often accompanied by a variety of cardiac pathologies, whereby the irregular and usually rapid ventricle rate of AF may cause inaccurate T1 estimation due to mis-triggering and inadequate magnetization recovery. We hypothesized that systolic T1 mapping with a heart-rate-dependent (HRD) pulse sequence scheme may overcome this issue.

Methods

30 patients with AF and 13 healthy volunteers were enrolled and underwent cardiovascular magnetic resonance (CMR) at 3 T. CMR was repeated for 3 patients after electric cardioversion and for 2 volunteers after lowering heart rate (HR). A Modified Look-Locker Inversion Recovery (MOLLI) sequence was acquired before and 15 min after administration of 0.1 mmol/kg gadopentetate dimeglumine. For AF patients, both the fixed 5(3)3/4(1)3(1)2 and the HRD sampling scheme were performed at diastole and systole, respectively. The HRD pulse sequence sampling scheme was 5(n)3/4(n)3(n)2, where n was determined by the heart rate to ensure adequate magnetization recovery. Image quality of T1 maps was assessed. T1 times were measured in myocardium and blood. Extracellular volume fraction (ECV) was calculated.

Results

In volunteers with repeated T1 mapping, the myocardial native T1 and ECV generated from the 1st fixed sampling scheme were smaller than from the 1st HRD and 2nd fixed sampling scheme. In healthy volunteers, the overall native T1 times and ECV of the left ventricle (LV) in diastolic T1 maps were greater than in systolic T1 maps (P < 0.01, P < 0.05). In the 3 AF patients that had received electrical cardioversion therapy, the myocardial native T1 times and ECV generated from the fixed sampling scheme were smaller than in the 1st and 2nd HRD sampling scheme (all P < 0.05). In patients with AF (HR: 88 ± 20 bpm, HR fluctuation: 12 ± 9 bpm), more T1 maps with artifact were found in diastole than in systole (P < 0.01). The overall native T1 times and ECV of the left ventricle (LV) in diastolic T1 maps were greater than systolic T1 maps, either with fixed or HRD sampling scheme (all P < 0.05).

Conclusion

Systolic MOLLI T1 mapping with heart-rate-dependent pulse sequence scheme can improve image quality and avoid T1 underestimation. It is feasible and with further validation may extend clinical applicability of T1 mapping to patients with atrial fibrillation.
Hinweise

Competing interests

The author(s) declare that they have no competing interests.

Authors’ contributions

LZ contributed substantially to the conception and design of the study, scanning of participants, data interpretation and drafted the manuscript; SL contributed substantially to the conception and design of the study and enrolling of volunteers and patients; XM contributed to substantially to scanning of volunteers and patients, data processing, analysis and interpretation, participated in the design of the study and critical revision of the manuscript; AG, TZ and JA critically revised the manuscript and supported the MR methods development and interpretation; RB and JD contributed substantially to the conception and design of the study and critical revision of the manuscript; ZF contributed to data processing, analysis and interpretation and critical revision of the manuscript. All authors read and approved the final manuscript.
Abkürzungen
AF
atrial fibrillation
ECV
extracellular volume fraction
MOLLI
Modified Look-Locker Inversion Recovery
HRD
heart-rate-dependent
CMR
cardiovascular magnetic resonance
TD
trigger delay
LGE
late gadolinium enhancement
LV
left-ventricular

Background

Atrial fibrillation (AF) is the most common clinical cardiac arrhythmia. The mechanisms of AF are complex and associated with structural and electrical remodeling in the atrial and ventricular myocardium [1]. Development of atrial fibrosis is the hallmark of structural remodeling in AF and considered the substrate for AF. In contrast, data on ventricular fibrosis of patients with AF are limited, although ventricular fibrosis has a detrimental effect on both systolic and diastolic function in patients with AF [2, 3]. Atrial and ventricular fibrosis in AF is likely to share many common mechanisms, including excessive accumulation of collagen either in regions of myocyte loss or diffusely in areas of extracellular matrix in the interstitium [4]. T1 mapping and derived extracellular volume fraction (ECV) enable quantitative assessment of myocardial characteristics, such as edema, focal and diffuse fibrosis [57]. T1 mapping and ECV may provide useful information for patients with AF. However, the irregular and usually rapid ventricular rate hindered the application of T1 mapping in patients with AF. Although different T1 mapping methods such as saturation-pulse-prepared heart-rate-independent inversion recovery T1 mapping have potential advantages in arrhythmia [8], the Modified Look-Locker Inversion Recovery (MOLLI) sequence is still widely used due to its higher precision and better reproducibility [9, 10]. The heart rate sensitivity of MOLLI has been significantly reduced by modification of sampling schemes (5(3)3 and 4(1)3(1)2 for pre-contrast and post-contrast T1 mapping, respectively) to the point where it is of much less concern [11]. The largest factor that affects the MOLLI heart rate sensitivity was the time between the inversions. It can be mitigated by increasing the time between inversions [12]. But the exact method how to increase the time between inversions is not well established yet. In addition, at higher heart rate, the time window with minimal motion in diastole shortens more than in systole. In patients with AF, although the R-R interval varies in each cardiac cycle, the variation in systole is smaller than in diastole [11]. It is potentially practicable to acquire T1 mapping in systole in patients with AF.
Therefore, we hypothesize that systolic T1 mapping with heart-rate-dependent (HRD) pulse sequence sampling scheme may be feasible in patients with AF. We compared the differences of T1 times and ECV acquired in systolic and diastolic phase with fixed/HRD pulse sequence sampling scheme, and assessed the accuracy of systolic T1 mapping with HRD pulse sequence sampling scheme in patients with AF who underwent cardiovascular magnetic resonance (CMR) twice, before and after electric cardioversion.

Methods

The study was approved by Beijing Anzhen Hospital ethics committee, and written informed consent was obtained from all participants. A detailed explanation concerning contrast agent was given to each participant.

Study population

Thirteen healthy volunteers and thirty patients with “persistent” AF without contraindication to CMR were recruited. None of the healthy volunteers was referred as patient for clinical CMR which then turned out to be normal, all of them had no evidence or risk factors of cardiovascular disease. The AF patients were defined as “persistent”, whose duration of continuous AF episodes lasted > 7 days or required electrical cardioversion. Exclusion criteria were presence of a permanent pacemaker, severe claustrophobia, and severe impairment of renal function (glomerular filtration rate < 30 mL/min/1.73 m2).

HRD pulse sequence sampling scheme determination

In the MOLLI sequence with fixed sampling scheme, such as 5(3)3 or 4(1)3(1), the inversion time between data acquisition is fixed as 3 heartbeats or 1 heartbeat as shown in the parentheses. At higher heart rate, each single heartbeat duration is shorter than at slow heart rate. Under such circumstances, the shorter inversion time will lead to inadequate magnetization recovery before acquiring the next data inversion episode, which results in underestimation of T1 times. To ensure the full recovery of longitudinal magnetization before the following inversion pulse is applied, the following pause in units of heartbeats needs to be increased along with the increase of the heart rate. Theoretically, after a duration of 4 times of T1 time (the maximal T1 time expected for the tissue of interest is approximately 2000 ms), the magnetization of tissue is almost back to equilibrium. For the 5(3)3 scheme, at heart rate of 60 beats per minute (bpm, RR interval = 1000 ms), it takes (5 + 3) × 1000 = 8000 ms (which equals to 4 times of tissue of interest’s T1 time) until the second inversion is played out. At higher heart rate, using the above scheme may lead to inadequate inversion recovery. It would make sense to set up different sampling schemes that adapt to the increased heart rate.
For the consideration of ensuring complete magnetization recovery at higher heart rates, the sampling schemes 5(3)3 and 4(1)3(1)2 were adopted to modify the recovery to be determined by heart rate. Based on the principle that 5(3)3 scheme is optimized for a heart rate of 60 bpm (1000 ms), for the pre-contrast T1 mapping, the heartbeat number of the second T1 inversion recovery is calculated as: n = (5+3)/(heart rate in ms) × 1000-5; for the post-contrast T1 mapping, the heartbeat number of the inversion is calculated as: n = (4+1)/(heart rate in ms) × 1000-4. We set up sampling schemes for various heart rate ranges with bpm increment of 5 from 50 to 150 bpm (Fig. 1a).

Study procedure

For the purpose of testing the feasibility and accuracy of T1 mapping with the HRD pulse sequence sampling scheme, two healthy volunteers in normal sinus rhythm with relatively high heart rate (HR, volunteer 1: 67–69 bpm, volunteer 2: 86–89 bpm) during first scan were asked to undergo two CMRs before and after taking beta-blockers. They underwent T1 mapping with fixed and HRD sampling scheme at first scan, and at the next day, after taking beta-blocker, they underwent the second T1 mapping scan with fixed sampling scheme (HR during scan: 57–59 bpm and 58–60 bpm, respectively). The left-ventricle myocardial per-segment native T1 time and ECV generated from fixed sampling scheme at first scan, HRD sampling scheme at first scan, and fixed sampling scheme at second scan were compared.
For observing the impact of different cardiac phases (systole and diastole) on T1 times, 11 healthy volunteers in normal sinus rhythm (HR: 60 ± 5 bpm, HR fluctuation: 3 ± 3 bpm) underwent T1 mapping with fixed sampling scheme (5(3)3/4(1)3(1)2) at diastole and systole. The number of T1 maps and myocardial segments with artifact generated from different cardiac phases were recorded and compared; the T1 time and ECV generated from different cardiac phases were compared.
For verifying the feasibility of systolic T1 mapping with HRD sampling scheme, patients with AF (HR: 88 ± 20 bpm, HR fluctuation: 12 ± 9 bpm) underwent T1 mapping with fixed and HRD sampling scheme at diastole and systole. The number of T1 maps and myocardial segments with artifact generated from different sampling scheme and cardiac phases combination were recorded and compared; the T1 time and ECV generated from different sampling scheme and cardiac phases combination were compared.
For the purpose of verifying the accuracy and effectiveness of systolic T1 mapping with the HRD pulse sequence sampling scheme, three patients (with AF) who received electric cardioversion, at the next day of cardioversion underwent the second T1 mapping with HRD sampling scheme in sinus rhythm, at diastole and systole. The left-ventricle myocardial per-segment native T1 time and ECV generated from different sampling schemes and cardiac phases combination at first scan, and HRD sampling scheme with different cardiac phases at second scan were compared. The flow chart of study is showed in Fig. 1b.

CMR acquisition

All CMR was performed using a 3 T MR system (MAGNETOM Verio, Siemens Healthcare, Erlangen, Germany) with a 32-channel cardiac coil. Subject-specific, volume-selective first- and second-order B0-shimming based on field maps derived from double-gradient-echo acquisitions was performed to improve static field uniformity.
T1 mapping sequence was obtained using a Siemens prototype (#448B, system software version syngo MR B17A). Data were acquired in basal, mid-ventricular, and apical short-axis planes at diastole and systole before and 15 min after administration of 0.1 mmol/kg gadopentetate dimeglumine (Magnevist, Bayer Healthcare). The diastolic phase was determined by the system with captured trigger delay (TD) according to the ECG gating, the systolic phase was set to TD of 0 ms (the acquisition started after 120 ms due to the TI start time) [13]. For the HRD sampling scheme, the number of inversions was determined by referencing the highest heart rate before the actual scan. Imaging parameters were: TR = 2.6-2.7 ms, TE = 1.0-1.1 ms, FA = 35°, FOV = (270 × 360) mm2, matrix = 256 for heart rate < 90 bpm, 192 for heart rate > 90 bpm, slice thickness = 6 mm, BW= 1045-1028Hz/px, GRAPPA acceleration factor 2, linear phase-encoding ordering, minimum TI of 120 ms. The quality control was performed during scanning by reviewing the “goodness of fit” map and source images to allow an immediate repetition of suboptimal measurements to minimize the respiratory motion and off-resonance effects. The blood sample was taken just before CMR to measure hematocrit for ECV calculation.
For the purpose of clinical diagnosis, all patients also received CMR protocols including cine and late gadolinium enhancement (LGE) imaging as described elsewhere [14].

Image analysis

All image datasets were transferred to a workstation (Viewing and Argus, Siemens Healthcare, Erlangen, Germany) for offline analysis. The T1 maps and source images were assessed, myocardial segments with artifact were excluded for further analysis. Two experienced readers assessed artifacts in consensus. The heart rate of each cardiac cycle during scanning of every image was recorded, the variability in heart rate was calculated as the standard deviation from the mean heart rate.
The left-ventricular (LV) myocardium was delineated by manually contouring the endo-cardial and epi-cardial borders. Care was taken to avoid contamination of signal from blood or epi-cardial fat. T1 time of the blood was measured by manually drawing a region of interest in the LV cavity of T1* map (T1 map based on fitting without Look-Locker correction) taking care to avoid the papillary muscles. When artifact existed in certain segment in image originated from any single diastole/systole and fixed/HRD sampling scheme combination, the patient’s same segment generated from T1 maps with other different cardiac phase and sampling scheme combinations were excluded for per-segment pairwise comparison. But all of the assessable segments generated from any single diastole/systole and fixed/HRD sampling scheme combination were included in calculating the respective average overall LV myocardial T1 times. The overall LV myocardial native T1 time was mean of T1 times of basal, mid-ventricular and apical levels. For the participants with repeated CMRs, all of the pre- and post-contrast per-segment (according to American Heart Association myocardial 17 segments classification, apical segment was excluded in this analysis) myocardial T1 times were drawn in T1 maps generated from different cardiac phases and sampling scheme combinations. ECV was calculated from T1 maps acquired pre- and post-contrast calibrated by blood hematocrit [15]. The ECV was calculated as:
$$ ECV=\left(1- hematocrit\right)\left(\frac{1}{T1myo\kern.3em post}-\frac{1}{T1myo\kern.3em pre}\right)/\left(\frac{1}{T1 blood\kern.3em post}-\frac{1}{T1 blood\kern.3em pre}\right) $$

Statistical analysis

Data are presented as mean and standard deviation. Differences between means were tested by the paired t-test and one-way analysis of variance as appropriate. The assessable images and segments were compared between systole and diastole using χ 2 test. The overall LV myocardial native T1 times and ECV generated from different cardiac phases and sampling scheme combinations were compared in all participants. The per-segment myocardial native T1 times and ECV generated from different cardiac phases and sampling scheme combinations were compared in participants with repeated CMRs. The statistical significance was defined as p < 0.05. Statistical analysis was performed using SPSS software (SPSS Inc., Chicago, IL, USA, version 17.0).

Results

The demographic data, heart rate and heart rate fluctuation during scan, LV functional indices of healthy volunteers and patients with AF are listed in Table 1.
Table 1
Demographic data and left-ventricle functional indices of healthy volunteers and patients with atrial fibrillation
 
Healthy volunteers (n = 13)
Patients with AF (n = 30)
Gender, female
3 (23.08 %)
10 (33.33 %)
Age (years)
53 ± 18
54 ± 13
Height (m)
1.70 ± 0.06
1.69 ± 0.09
Weight (kg)
76.35 ± 14.97
76.20 ± 15.21
BMI (kg/m2)
26.17 ± 3.87
26.46 ± 3.94
Average HR (bpm)
62 ± 9
88 ± 20
HR fluctuation (bpm)
3 ± 3
12 ± 9
Left atrium volume (cm3)
139.03 ± 45.05
158.42 ± 50.55
EF (%)
63.17 ± 7.00
49.19 ± 11.83
EDV (ml)
98.17 ± 19.96
87.81 ± 26.85
ESV (ml)
36.46 ± 10.96
44.94 ± 19.51
SV (ml)
61.71 ± 13.13
42.89 ± 16.61
MM (g)
86.12 ± 24.59
85.94 ± 25.01
CO (l/min)
3.97 ± 1.17
3.58 ± 1.10
LGE positive subject
0
10
Data were presented in mean ± standard deviation. AF atrial fibrillation, BMI body mass index, HR heart rate, EF ejection fraction, EDV end-diastolic volume, ESV end-systolic volume, SV stroke volume, MM left-ventricle myocardial mass, CO cardiac output, LGE late gadolinium enhancement

Comparison between HRD and fixed sampling scheme in normal sinus rhythm with high heart rate

For the two healthy volunteers that underwent two CMRs, volunteer 1 presented HR of 67–69 bpm during the first scan, the HRD sampling scheme was 5(4)3/4(2)3(2)2, volunteer 2 presented a HR of 86–89 bpm during the first scan with HRD sampling scheme of 5(7)3/4(3)3(3)2. In volunteer 1, there were no significant differences of per-segment myocardial native T1 times and ECV among 1st 5(3)3/4(1)3(1)2, 1st 5(4)3/4(2)3(2)2 and 2nd 5(3)3/4(1)3(1)2 either in diastolic or systolic T1 maps (all p > 0.05). In volunteer 2, there were significant differences of per-segment myocardial native T1 times and ECV between 1st 5(3)3/4(1)3(1)2 and 1st 5(7)3/4(3)3(3)2, and between 1st 5(3)3/4(1)3(1)2 and 2nd 5(3)3/4(1)3(1)2 either in diastolic or systolic T1 mappings (all p < 0.05); but no significant differences of per-segment myocardial native T1 times and ECV between 1st 5(7)3/4(3)3(3)2 and 2nd 5(3)3/4(1)3(1)2 either in diastolic or systolic T1 mappings (all p > 0.05).

Comparison between systole and diastole T1 mapping in healthy volunteers

In the 11 healthy volunteers, there were 66 myocardial native T1 maps and 66 myocardial post-contrast T1 maps (11 volunteers × 3 slices × 2 (1 of diastole and 1 of systole)). 6 segments in 2 native T1 maps and 7 segments in 3 post-contrast T1 maps were excluded due to the presence of artifacts. In the native T1 maps with artifacts, 4 segments of 1 T1 map were diastolic images; in the post-contrast T1 maps with artifacts, 5 segments of 2 T1 maps were diastolic images. There were no significant differences of assessable images between systole and diastole (χ 2 = 0.204, p > 0.05).
The LV overall native T1 times, post-contrast T1 times and ECV of myocardium, pre-/post-contrast blood T1 times of diastolic and systolic T1 mapping were compared and are presented in Table 2. There were minor but statistically significant differences in myocardial and blood T1 times and ECV between diastolic and systolic T1 maps.
Table 2
Comparison of pre-/post-contrast myocardium and blood T1 times and ECV between diastolic and systolic T1 maps in healthy volunteers
 
Diastole
Systole
P value
Overall LV myocardium native T1 times (ms)
1247.73 ± 31.86
1231.06 ± 31.45
<0.01
Overall LV myocardium post-contrast T1 times (ms)
544.94 ± 52.80
563.80 ± 56.89
<0.01
Overall LV pre-contrast blood T1 times (ms)
1832.76 ± 61.76
1784.49 ± 82.02
<0.05
Overall LV post-contrast blood T1 times (ms)
352.87 ± 49.15
365.53 ± 51.70
<0.01
Overall LV ECV (%)
25.11 ± 1.63
24.60 ± 1.58
=0.047
Data are presented in mean ± standard deviation. LV left ventricle, ECV extracellular volume

Comparison among different HRD/fixed sampling schemes and systole/diastole combination T1 mapping in patients with AF

For the 30 patients with “persistent” AF, there were 378 native T1 maps and 378 post-contrast T1 maps (27 patients × 3 slices × 4 (diastole/systole + fixed/HRD sampling scheme combination) + 3 patients × 3 slices × 6 (1st CMR: diastole/systole + fixed/HRD sampling scheme combination = 4; 2nd CMR: diastole/systole + HRD sampling scheme combination = 2)). 599 segments in 113 native T1 maps and 592 segments in 111 post-contrast T1 maps were excluded due to the presence of artifacts. The artifacts we encountered mainly were off resonance related dark banding artifact and mistrigger related motion artifact (Fig. 2). The dark banding artifact were presented evenly in diastolic and systolic images, although the local shimming were applied and different center frequencies were tried, 35 segments (0.99 %) were with identified banding artifact ultimately. The mistrigger related motion artifacts were more commonly found in diastolic images than systolic images. In the native T1 maps with motion artifacts, 542 segments (15.36 %) of 100 T1 maps (13.23 %) were diastolic images; in the post-contrast T1 maps with artifacts, 531 segments (15.05 %) of 97 T1 maps (12.83 %) were diastolic images. Overall, there were more assessable T1 maps in systole than diastole with significant differences (χ 2 = 201.003, p < 0.01). Most motion artifacts were presented in patients with moderately rapid heart rate and/or with severe heart rate fluctuation (Fig. 3). In very rapid heart rate (e.g. 130 bpm), the system default diastolic TD was close to the systolic TD of 0 ms, therefore we obtained two series of similar “systolic” T1 maps. Consequently, very fast heart rate was observed with relatively less number of inferior quality images compared to moderate rapid heart rate in diastolic T1 maps (Fig. 4).
The myocardial native T1 times and ECV generated from HRD sampling scheme T1 maps were greater than fixed sampling scheme T1 maps either in diastolic or systolic phase with statistical significance (all p < 0.05). The myocardial native T1 times generated from diastolic T1 maps were longer than systolic T1 maps either in fixed or in HRD scheme T1 maps with statistical significance (all p < 0.05), the ECV generated from diastole was larger than in systole, either in fixed or HRD sampling scheme without statistical significance (all p > 0.05). In the pre-contrast LV blood T1* maps, blood T1 times generated from diastolic phase were longer than systolic phase, while in the post-contrast LV blood T1* maps, the blood T1 times from diastolic phase were shorter than systolic phase. Regarding the comparison between fixed and HRD scheme, blood T1 times generated from fixed scheme were longer than HRD scheme in pre-contrast; In post-contrast CMR, the blood T1 from fixed scheme were shorter than HRD scheme (nearly half of these differences were statistical significant). The LV overall native T1 times, post-contrast T1 times and ECV of myocardium, pre-/post-contrast blood T1 times generated from different diastolic/systolic and fixed/HRD scheme combination were compared and are shown in Table 3.
Table 3
Comparison of pre-/post-contrast myocardium and blood T1 times and ECV between different diastolic/systolic and fixed/HRD scheme T1 maps in patients with persistent AF
 
1
2
3
4
P value
 
Diastole with fixed scheme
Systole with fixed scheme
Diastole with HRD scheme
Systole with HRD scheme
Overall LV myocardium native T1 times (ms)
1272.76 ± 39.30
1263.29 ± 40.71
1311.00 ± 44.51
1304.83 ± 52.01
1 vs. 2: <0.01
3 vs. 4: <0.05
1 vs. 3: <0.01
2 vs. 4: <0.01
Overall LV myocardium post-contrast T1 times (ms)
543.42 ± 57.04
551.07 ± 57.49
571.37 ± 66.16
588.94 ± 64.31
1 vs. 2: <0.05
3 vs. 4: >0.05
1 vs. 3: >0.05
2 vs. 4: <0.05
Overall LV pre-contrast blood T1 times (ms)
1819.40 ± 136.03
1797.49 ± 132.23
1796.32 ± 132.02
1766.30 ± 133.51
1 vs. 2: >0.05
3 vs. 4: >0.05
1 vs. 3: >0.05
2 vs. 4: >0.05
Overall LV post-contrast blood T1 times (ms)
359.61 ± 53.34
366.06 ± 52.73
393.03 ± 61.73
400.18 ± 64.13
1 vs. 2: <0.01
3 vs. 4: >0.05
1 vs. 3: <0.05
2 vs. 4: <0.05
Overall LV ECV (%)
26.70 ± 2.45
26.41 ± 2.73
28.02 ± 3.48
27.18 ± 3.36
1 vs. 2: >0.05
3 vs. 4: >0.05
1 vs. 3: <0.05
2 vs. 4: <0.05
Data are presented in mean ± standard deviation. ECV extracellular volume, HRD heart-rate-dependent, AF atrial fibrillation, LV left ventricle
Three patients underwent CMR 2 times, before and after electric cardioversion. Patient 1’s heart rate during 1st CMR ranged 69–116 bpm, HRD scheme was 5(11)3/4(6)3(6)2; after electric cardioversion, the patient’s heart rate during 2nd CMR ranged 77–80 bpm, scheme was 5(5)3/4(2)3(2)2. Patient 2’s heart rate during 1st CMR ranged 64–87 bpm, HRD scheme was 5(7)3/4(3)3(3)2; the patient’s heart rate during 2nd CMR ranged 62–65 bpm, scheme was 5(4)3/4(1)3(1)2. Patient 3’s heart rate during 1st CMR ranged 80–101 bpm, HRD was 5(9)3/4(5)3(5)2; the patient’s heart rate during 2nd CMR ranged 75–77 bpm, scheme was 5(5)3/4(2)3(2)2. For all the 3 patients, either in the diastolic or systolic T1 maps, there were significant differences of per-segment native T1 times and ECV between 1st CMR fixed scheme and HRD scheme, and between 1st CMR fixed scheme and 2nd CMR HRD scheme (all p <0.01), but no significant differences of per-segment native T1 times and ECV between 1st CMR HRD scheme and 2nd CMR HRD scheme (p >0.05, Table 4).
Table 4
Comparison of per-segment native T1 times and ECV in patients before/after electric cardioversion
   
Patient 1
Patient 2
Patient 3
    
P value
 
P value
 
P value
Diastole
A: fixed
Native T1 (ms)
1294.67 ± 16.50
A vs. B: <0.01
1233.83 ± 25.09
A vs. B: <0.05
1268.80 ± 34.85
A vs. B: <0.01
B: 1st HRD
Native T1 (ms)
1353.30 ± 36.55
B vs. C: >0.05
1241.17 ± 16.08
B vs. C: >0.05
1289.57 ± 13.29
B vs. C: >0.05
C: 2nd HRD
Native T1 (ms)
1342.60 ± 18.12
A vs. C: <0.01
1245.07 ± 11.98
A vs. C: <0.05
1289.07 ± 16.05
A vs. C: <0.01
A: fixed
ECV (%)
23.12 ± 0.22
A vs. B: <0.01
22.28 ± 1.00
A vs. B: <0.01
27.25 ± 0.34
A vs. B: <0.01
B: 1st HRD
ECV (%)
25.51 ± 0.88
B vs. C: >0.05
24.96 ± 0.24
B vs. C: >0.05
28.49 ± 0.37
B vs. C: >0.05
C: 2nd HRD
ECV (%)
25.47 ± 1.14
A vs. C: <0.01
25.60 ± 0.97
A vs. C: <0.01
28.50 ± 0.21
A vs. C: <0.01
Systole
A: fixed
Native T1 (ms)
1283.30 ± 34.76
A vs. B: <0.01
1209.90 ± 33.24
A vs. B: <0.01
1249.73 ± 39.49
A vs. B: <0.01
B: 1st HRD
Native T1 (ms)
1345.1 ± 23.40
B vs. C: >0.05
1246.37 ± 33.92
B vs. C: >0.05
1274.70 ± 17.68
B vs. C: >0.05
C: 2nd HRD
Native T1 (ms)
1339.57 ± 25.75
A vs. C: <0.01
1250.80 ± 36.09
A vs. C: <0.01
1256.10 ± 17.91
A vs. C: <0.05
A: fixed
ECV (%)
23.57 ± 0.62
A vs. B: <0.05
22.86 ± 2.58
A vs. B: <0.05
27.01 ± 1.50
A vs. B: <0.05
B: 1st HRD
ECV (%)
24.41 ± 1.74
B vs. C: >0.05
23.74 ± 0.56
B vs. C: >0.05
27.13 ± 0.65
B vs. C: >0.05
C: 2nd HRD
ECV (%)
24.76 ± 0.52
A vs. C: <0.05
23.56 ± 1.21
A vs. C: <0.05
27.14 ± 0.90
A vs. C: <0.05
Data are presented in mean ± standard deviation. ECV extracellular volume, HRD heart-rate-dependent

Discussion

The main findings of this study include: the HRD sampling scheme can avoid the underestimation of T1 times in the setting of rapid ventricle rate either in sinus rhythm or arrhythmia; although the native T1 times/ECV of myocardium in diastole are greater than in systole, to obtain data at systole can improve the image quality by decreasing the chance of mistriggering in patients with AF. Finally, a subset of patients with repeated T1 mapping demonstrated the effectiveness of systolic HRD sampling scheme in patients with AF. In addition, to the best of our knowledge, this is the first study using native T1 and ECV assessing LV myocardium in patients with AF before catheter ablation. Compared to healthy volunteers, the patients with AF showed greater native T1 times and ECV, which may be an indicator for myocardial fibrosis. Our findings are in line with other studies using other methods or animal study [16, 17].

Validation of T1 mapping in HRD sampling scheme

Incomplete inversion recovery may lead to an underestimation of T1 times. In patients with higher heart rate, the R-R interval shortened, and the fixed sampling scheme of 5(3)3/4(1)3(1)2 does not provide enough time for the magnetization to get back to the equilibrium. Increasing the number of heartbeats after T1 inversion recovery can avoid underestimation of T1 times. In our study, we tested the feasibility and accuracy of HRD sampling scheme in healthy volunteers and patients with AF with repeated CMRs. In volunteers and patients with AF, the T1 times and ECV generated from fixed sampling scheme were smaller than HRD sampling scheme, and repeated T1 mapping demonstrated that the HRD sampling scheme can more accurately estimate the T1 times than the fixed sampling scheme. In volunteer 1, the heartbeats number of T1 inversion recovery increased in a small range (pre-contrast: 3 to 4, post-contrast: 1 to 2), although the myocardial native T1 times and ECV generated from 1st fixed sampling scheme were smaller than 1st HRD and 2nd fixed sampling scheme, the differences did not reach statistical significance. Volunteer 2’s heartbeats number of T1 inversion recovery increased in a larger range (3 to 7 and 1 to 3), the differences of myocardial native T1 times and ECV between 1st fixed sampling scheme and 1st HRD/2nd CMR fixed sampling scheme reached statistical significance. In the 3 patients with AF, all the heartbeats number of T1 inversion recovery increased with a large range (3 to 11, 7, 9 in the pre-contrast, and 1 to 6, 3, 5 in the post-contrast respectively). The myocardial native T1 times and ECV of 1st fixed sampling scheme were smaller than 1st and 2nd HRD sampling scheme with statistical significance. The impact of the HRD sampling scheme was more prominent for higher heart rates. Thereby, HRD sampling scheme is practical for rapid ventricle rate either in sinus rhythm or arrhythmia.

Impact of cardiac cycle on T1 mapping

In healthy volunteers and patients with AF, both for fixed and HRD sampling scheme, the myocardial native T1 times and ECV in diastole were greater than in systole. Our findings concur in general with some previous studies [13, 18, 19]. The differences of myocardial T1 times between diastole and systole can be explained by partial-volume effect on diastole [13] and/or reduced intra-myocardial blood volume contamination in systole [18]. Nevertheless, the mean differences of native T1 times (healthy volunteer: 16.67 ms; patients with AF: 9.47 ms with fixed scheme and 6.17 ms with HRD scheme) and ECV (healthy volunteer: 0.51 %; patients with AF: 0.29 % with fixed scheme and 0.84 % with HRD scheme) were small and might not be clinically relevant on an individual basis. Even so, it is recommended to obtain and compare T1 maps in the same cardiac phase to avoid potential bias [5].
Although apparent myocardial native T1 times and ECV in systole were smaller than in diastole, more images in systole were evaluable than in diastole. In line with other studies [13, 18, 20], our results revealed that T1 mapping acquisition in systole increases the number of evaluable images and segments both in healthy volunteers and patients with AF. At higher heart rates, the motion-free time in diastole shortens more than in systole. In patients with AF, although the R-R interval varies in each cardiac cycle, the variation in systole is smaller than in diastole [11, 13] (Fig. 2). Therefore, the acquisition in systole resulted in more evaluable images. But in case of lower ventricular rate, at mid to end diastole, cardiac motion is relatively minor, and even in patients with AF and low heart rates, diastolic images were presented with less artifacts [13] (Fig. 3).

Diffuse ventricular fibrosis in patients with AF

Except the main finding of our study, our results also revealed that the myocardial native T1 times and ECV of patients with AF were greater than in healthy volunteers. The cardiac profibrotic microenvironment in AF is unlikely to be strictly limited to the atria, and the ventricular myocardium is also likely to be affected, ventricular fibrotic changes are more pronounced in AF patients than in subjects with sinus rhythm [1]. Although the effect of ventricular fibrosis in the pathogenesis of AF is controversial, recent study found that diffuse ventricular fibrosis is an independently predictor of AF recurrence after ablation therapy [3]. T1 mapping and ECV allow quantitative assessment of diffuse myocardial characteristics fibrosis, which was only possible with biopsy in the past. Ling et al. demonstrated diffuse ventricular fibrosis in AF by employing post-contrast T1 mapping [2]. Furthermore, our study demonstrated diffuse ventricular fibrosis in AF using the reliable parameter of ECV. This technique provides a new method for investigating the implication of ventricular fibrosis in AF.

Limitations

The number of participants included in this study is relatively small, especially the number of participants who underwent the CMR twice. Since the repeated CMRs can test the efficiency of the HRD sampling scheme, more participants need to be enrolled in a future study. The HRD sequence used in this study is relatively old version, the inversion time is expressed as heartbeat, now a more advanced sequence with seconds as unit is widely accepted by T1 mapping research community. Due to the sequence version’s limitations, we are not able to test this sequence in our study. However, it’s suggested that we utilize this advanced version of sequence in our future study. The operation of the HRD sampling scheme is relatively complicated, and the impact of HRD sampling scheme is more prominent in higher heart rate, therefore further investigation is needed to simplify the operation under the condition of keeping the feature of HRD sampling scheme. Finally, we did not perform the intra- and inter-observer variability, although a good agreement of intra- and inter-observer variability has been demonstrated by other studies [13, 20].

Conclusions

The heart-rate-dependent sampling scheme is helpful for avoiding underestimation of T1 times, especially in subjects with higher heart rates. Systolic T1 mapping yields shorter myocardial native T1 times and ECV, but the mean differences were small and, importantly, with the benefit of improvement in data quality in subjects with irregular and rapid heart rate. Together with these findings, systolic MOLLI T1 mapping with a heart-rate-dependent sampling scheme is feasible and its clinical application can be extended to patients with atrial fibrillation.

Acknowledgements

The research was supported by the National Natural Science Foundation of China (81300146, 81101173), Beijing Municipal Natural Science Foundation (7144207), Capital Health Research and Development of Special (2014-1-2061), High Levels of Health Technical Personnel in Beijing Municipal Commission of Health and Family Planning (2013-3-005).
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

The author(s) declare that they have no competing interests.

Authors’ contributions

LZ contributed substantially to the conception and design of the study, scanning of participants, data interpretation and drafted the manuscript; SL contributed substantially to the conception and design of the study and enrolling of volunteers and patients; XM contributed to substantially to scanning of volunteers and patients, data processing, analysis and interpretation, participated in the design of the study and critical revision of the manuscript; AG, TZ and JA critically revised the manuscript and supported the MR methods development and interpretation; RB and JD contributed substantially to the conception and design of the study and critical revision of the manuscript; ZF contributed to data processing, analysis and interpretation and critical revision of the manuscript. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Dzeshka MS, Lip GY, Snezhitskiy V, Shantsila E. Cardiac fibrosis in patients with atrial fibrillation: mechanisms and clinical implications. J Am Coll Cardiol. 2015;66(8):943–59.CrossRefPubMed Dzeshka MS, Lip GY, Snezhitskiy V, Shantsila E. Cardiac fibrosis in patients with atrial fibrillation: mechanisms and clinical implications. J Am Coll Cardiol. 2015;66(8):943–59.CrossRefPubMed
2.
Zurück zum Zitat Ling LH, Kistler PM, Ellims AH, Iles LM, Lee G, Hughes GL, et al. Diffuse ventricular fibrosis in atrial fibrillation: noninvasive evaluation and relationships with aging and systolic dysfunction. J Am Coll Cardiol. 2012;60(23):2402–8.CrossRefPubMed Ling LH, Kistler PM, Ellims AH, Iles LM, Lee G, Hughes GL, et al. Diffuse ventricular fibrosis in atrial fibrillation: noninvasive evaluation and relationships with aging and systolic dysfunction. J Am Coll Cardiol. 2012;60(23):2402–8.CrossRefPubMed
3.
Zurück zum Zitat Neilan TG, Mongeon FP, Shah RV, Coelho-Filho O, Abbasi SA, Dodson JA, et al. Myocardial extracellular volume expansion and the risk of recurrent atrial fibrillation after pulmonary vein isolation. JACC Cardiovasc Imaging. 2014;7(1):1–11.CrossRefPubMedPubMedCentral Neilan TG, Mongeon FP, Shah RV, Coelho-Filho O, Abbasi SA, Dodson JA, et al. Myocardial extracellular volume expansion and the risk of recurrent atrial fibrillation after pulmonary vein isolation. JACC Cardiovasc Imaging. 2014;7(1):1–11.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Tarone G, Balligand JL, Bauersachs J, Clerk A, De Windt L, Heymans S, et al. Targeting myocardial remodelling to develop novel therapies for heart failure: a position paper from the Working Group on Myocardial Function of the European Society of Cardiology. Eur J Heart Fail. 2014;16(5):494–508.CrossRefPubMed Tarone G, Balligand JL, Bauersachs J, Clerk A, De Windt L, Heymans S, et al. Targeting myocardial remodelling to develop novel therapies for heart failure: a position paper from the Working Group on Myocardial Function of the European Society of Cardiology. Eur J Heart Fail. 2014;16(5):494–508.CrossRefPubMed
5.
Zurück zum Zitat Moon JC, Messroghli DR, Kellman P, Piechnik SK, Robson MD, Ugander M, et al. Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement. J Cardiovasc Magn Reson. 2013;15:92.CrossRefPubMedPubMedCentral Moon JC, Messroghli DR, Kellman P, Piechnik SK, Robson MD, Ugander M, et al. Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement. J Cardiovasc Magn Reson. 2013;15:92.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Dass S, Suttie JJ, Piechnik SK, Ferreira VM, Holloway CJ, Banerjee R, et al. Myocardial tissue characterization using magnetic resonance noncontrast t1 mapping in hypertrophic and dilated cardiomyopathy. Circ Cardiovasc Imaging. 2012;5(6):726–33.CrossRefPubMed Dass S, Suttie JJ, Piechnik SK, Ferreira VM, Holloway CJ, Banerjee R, et al. Myocardial tissue characterization using magnetic resonance noncontrast t1 mapping in hypertrophic and dilated cardiomyopathy. Circ Cardiovasc Imaging. 2012;5(6):726–33.CrossRefPubMed
7.
Zurück zum Zitat Kellman P, Wilson JR, Xue H, Bandettini WP, Shanbhag SM, Druey KM, et al. Extracellular volume fraction mapping in the myocardium, part 2: initial clinical experience. J Cardiovasc Magn Reson. 2012;14:64.CrossRefPubMedPubMedCentral Kellman P, Wilson JR, Xue H, Bandettini WP, Shanbhag SM, Druey KM, et al. Extracellular volume fraction mapping in the myocardium, part 2: initial clinical experience. J Cardiovasc Magn Reson. 2012;14:64.CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Roujol S, Weingärtner S, Foppa M, Chow K, Kawaji K, Ngo LH, et al. Accuracy, precision, and reproducibility of four T1 mapping sequences: a head-to-head comparison of MOLLI, ShMOLLI, SASHA, and SAPPHIRE. Radiology. 2014;272(3):683–9.CrossRefPubMedPubMedCentral Roujol S, Weingärtner S, Foppa M, Chow K, Kawaji K, Ngo LH, et al. Accuracy, precision, and reproducibility of four T1 mapping sequences: a head-to-head comparison of MOLLI, ShMOLLI, SASHA, and SAPPHIRE. Radiology. 2014;272(3):683–9.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Nacif MS, Turkbey EB, Gai N, Nazarian S, van der Geest RJ, Noureldin RA, et al. Myocardial T1 mapping with MRI: comparison of Look-Locker and MOLLI sequences. J Magn Reson Imaging. 2011;34:1367–73.CrossRefPubMedPubMedCentral Nacif MS, Turkbey EB, Gai N, Nazarian S, van der Geest RJ, Noureldin RA, et al. Myocardial T1 mapping with MRI: comparison of Look-Locker and MOLLI sequences. J Magn Reson Imaging. 2011;34:1367–73.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Raman FS, Kawel-Boehm N, Gai N, Freed M, Han J, Liu CY, et al. Modified Look-Locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners. J Cardiovasc Magn Reson. 2013;15:64.CrossRefPubMedPubMedCentral Raman FS, Kawel-Boehm N, Gai N, Freed M, Han J, Liu CY, et al. Modified Look-Locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners. J Cardiovasc Magn Reson. 2013;15:64.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Wang Y, Vidan E, Bergman GW. Cardiac motion of coronary arteries: variability in the rest period and implications for coronary MR angiography. Radiology. 1999;213:751–8.CrossRefPubMed Wang Y, Vidan E, Bergman GW. Cardiac motion of coronary arteries: variability in the rest period and implications for coronary MR angiography. Radiology. 1999;213:751–8.CrossRefPubMed
13.
Zurück zum Zitat Ferreira VM, Wijesurendra RS, Liu A, Greiser A, Casadei B, Robson MD, et al. Systolic ShMOLLI myocardial T1-mapping for improved robustness to partial-volume effects and applications in tachyarrhythmias. J Cardiovasc Magn Reson. 2015;17(1):77.CrossRefPubMedPubMedCentral Ferreira VM, Wijesurendra RS, Liu A, Greiser A, Casadei B, Robson MD, et al. Systolic ShMOLLI myocardial T1-mapping for improved robustness to partial-volume effects and applications in tachyarrhythmias. J Cardiovasc Magn Reson. 2015;17(1):77.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat von Knobelsdorff-Brenkenhoff F, Prothmann M, Dieringer MA, Wassmuth R, Greiser A, Schwenke C, et al. Myocardial T1 and T2 mapping at 3 T: reference values, influencing factors and implications. J Cardiovasc Magn Reson. 2013;15:53.CrossRef von Knobelsdorff-Brenkenhoff F, Prothmann M, Dieringer MA, Wassmuth R, Greiser A, Schwenke C, et al. Myocardial T1 and T2 mapping at 3 T: reference values, influencing factors and implications. J Cardiovasc Magn Reson. 2013;15:53.CrossRef
15.
Zurück zum Zitat Ugander M, Oki AJ, Hsu LY, Kellman P, Greiser A, Aletras AH. Extracellular volume imaging by magnetic resonance imaging provides insights into overt and sub-clinical myocardial pathology. Eur Heart J. 2012;33(10):1268–78.CrossRefPubMedPubMedCentral Ugander M, Oki AJ, Hsu LY, Kellman P, Greiser A, Aletras AH. Extracellular volume imaging by magnetic resonance imaging provides insights into overt and sub-clinical myocardial pathology. Eur Heart J. 2012;33(10):1268–78.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Sasaki N, Okumura Y, Watanabe I, et al. Transthoracic echocardiographic back scatter based assessment of left atrial remodeling involving left atrial and ventricular fibrosis in patients with atrial fibrillation. Int J Cardiol. 2014;176:1064–6.CrossRefPubMed Sasaki N, Okumura Y, Watanabe I, et al. Transthoracic echocardiographic back scatter based assessment of left atrial remodeling involving left atrial and ventricular fibrosis in patients with atrial fibrillation. Int J Cardiol. 2014;176:1064–6.CrossRefPubMed
17.
Zurück zum Zitat Chrysostomakis SI, Karalis IK, Simantirakis EN, et al. Angiotensin II type 1 receptor inhibition is associated with reduced tachyarrhythmia-induced ventricular interstitial fibrosis in a goat atrial fibrillation model. Cardiovasc Drugs Ther. 2007;21:357–65.CrossRefPubMed Chrysostomakis SI, Karalis IK, Simantirakis EN, et al. Angiotensin II type 1 receptor inhibition is associated with reduced tachyarrhythmia-induced ventricular interstitial fibrosis in a goat atrial fibrillation model. Cardiovasc Drugs Ther. 2007;21:357–65.CrossRefPubMed
18.
Zurück zum Zitat Kawel N, Nacif M, Zavodni A, Jones J, Liu S, Sibley CT, et al. T1 mapping of the myocardium: intra-individual assessment of the effect of field strength, cardiac cycle and variation by myocardial region. J Cardiovasc Magn Reson. 2012;14:27.CrossRefPubMedPubMedCentral Kawel N, Nacif M, Zavodni A, Jones J, Liu S, Sibley CT, et al. T1 mapping of the myocardium: intra-individual assessment of the effect of field strength, cardiac cycle and variation by myocardial region. J Cardiovasc Magn Reson. 2012;14:27.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Reiter U, Reiter G, Dorr K, Greiser A, Maderthaner R, Fuchsjager M. Normal diastolic and systolic myocardial T1 values at 1.5-T MR imaging: correlations and blood normalization. Radiology. 2014;271(2):365–72.CrossRefPubMed Reiter U, Reiter G, Dorr K, Greiser A, Maderthaner R, Fuchsjager M. Normal diastolic and systolic myocardial T1 values at 1.5-T MR imaging: correlations and blood normalization. Radiology. 2014;271(2):365–72.CrossRefPubMed
20.
Zurück zum Zitat Tessa C, Diciotti S, Landini N, Lilli A, Del Meglio J, Salvatori L, et al. Myocardial T1 and T2 mapping in diastolic and systolic phase. Int J Cardiovasc Imaging. 2015;31(5):1001–10.CrossRefPubMed Tessa C, Diciotti S, Landini N, Lilli A, Del Meglio J, Salvatori L, et al. Myocardial T1 and T2 mapping in diastolic and systolic phase. Int J Cardiovasc Imaging. 2015;31(5):1001–10.CrossRefPubMed
Metadaten
Titel
Systolic MOLLI T1 mapping with heart-rate-dependent pulse sequence sampling scheme is feasible in patients with atrial fibrillation
verfasst von
Lei Zhao
Songnan Li
Xiaohai Ma
Andreas Greiser
Tianjing Zhang
Jing An
Rong Bai
Jianzeng Dong
Zhanming Fan
Publikationsdatum
01.12.2016
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2016
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-016-0232-7

Weitere Artikel der Ausgabe 1/2016

Journal of Cardiovascular Magnetic Resonance 1/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.