Ultrasound-guided peripheral venous cannulation in critically ill patients: a practical guideline

verfasst von: Pablo Blanco

Erschienen in: The Ultrasound Journal | Ausgabe 1/2019

Abstract

Background

Up to one-third of critically ill patients have difficult intravenous access (DIVA). This occurs often in obese patients, those with generalized edemas or in patients with previous venous cannulations. In DIVA patients, the conventional technique often fails. In contrast, ultrasound-guided cannulation has demonstrated a high success rate, improving patient satisfaction and even a reduction in the need of central venous lines. However, a high rate of premature catheter failure has been shown in cannulations performed by ultrasound guidance and thus a comprehensive knowledge of several aspects related to this procedure is mandatory to improve cannulation success, avoid complications and lengthen the survival of the catheter.

Main text

Several practical issues related to peripheral venous cannulation are described: peripheral venous anatomy, vein size and catheter selection, distance from skin to vein, insertion angle and selection of the catheter length, cannulation technique itself (out-of-plane or in-plane) and checking catheter position.

Conclusion

Key concepts regarding ultrasound-guided peripheral vein cannulation should be well known for practitioners, aiding in improving cannulation success and catheter dwell time, and avoiding complications.
Begleitmaterial
Hinweise

Supplementary information

Supplementary information accompanies this paper at https://​doi.​org/​10.​1186/​s13089-019-0144-5.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
DIVA
difficult intravenous access
PCF
premature catheter failure
US
ultrasound
SLC
standard-length catheters
ULPC
ultra-long peripheral catheters

Background

Peripheral venous cannulation is essential to provide care for the patients in the emergency department or critical care unit. While most intravenous catheters are placed using the conventional technique (i.e., seeing and/or palpating the vein), up to one-third of the patients have difficult intravenous access (DIVA) [1]. This group often involves patients with generalized edemas, obese, those with multiple previous cannulations or intravenous drug users [25]. For patients who have DIVA, ultrasound (US)-guided cannulation has shown an overall success rate higher than 90%, compared to 25–30% using the conventional technique [24, 6], and also aids in reducing the need for central venous lines [4, 5]. Patient satisfaction also improves using US guidance [2]. In spite of that, the rate of premature catheter failure (PCF), which may account for as high as 50% within 24 h of catheter placement, is higher with US-guided cannulations (45–56%) compared to the conventional technique (19–25%) [1, 6, 7]. Infiltration is the leading cause of catheter failure; catheter dislodgement and thrombophlebitis are also common [1, 6]. Key concepts regarding veins, catheters and the technique itself should be considered by practitioners to improve success, reduce complications and improve dwell times in US-guided peripheral intravenous cannulation, and these are provided in this article, and are summarized in Table 1 as well. Basic general knowledge for US-guided vascular cannulation is shown in Fig. 1.
Table 1
Key concepts for ultrasound-guided peripheral venous cannulation
Key concept
Description
Considerations
Must know
Basic general knowledge for ultrasound-guided vascular cannulation
Selection of the transducer and preset; image orientation; basic image optimization; distinguishing veins versus arteries; managing techniques of cannulation (out-of-plane and in-plane technique)
1
Select superficial veins (i.e., epifascial)
Superficial veins: short pathway to reach the vein; high probability that a great proportion of the catheter will dwell in the vein
Deep veins: inherent risks of needle-stick injury of the artery or the nerve; frequent catheter dislodgment
2
Patent veins
Anechoic lumen; fully compressible
Do not misinterpret stagnant blood in the vein lumen with thrombus (distal compression is useful)
3
Vein size: anteroposterior diameter ≥ 4 mm
AP diameter (mm) = maximum Fr catheter size
(e.g., 4 mm = up to 4-Fr catheter)
4
Vein depth: up to 16 mm (short axis)
Real distance to reach the vein (45° insertion angle)*: 1.4 × vertical distance
*Ideally measure real distance in long axis and select the best angle of insertion
≥ 2.75 cm of the catheter must dwell in the vein
Consider using ultra-long peripheral (6.3 cm) and midline (8–20 cm) catheters to minimize catheter failure
5
Select the technique: in-plane or out-of-plane technique
Both are useful, although it seems to be a greater success rate with the out-of-plane technique
Learning and using both techniques is encouraged
6
Checking catheter position
Direct: double hyperechoic line into the lumen vein
Indirect: saline flush test (two-dimensional and/or color Doppler)

Key concept 1: Select superficial veins

Anatomically, peripheral veins may be superficial or epifascial (i.e., above the fascia), and deep or subfascial (i.e., below the fascia). Superficial veins are found close to the skin and travel without an accompanying artery or nerve. In contrast, deep veins (which may be paired, as seen in brachial veins) are found at the neurovascular bundle, and thus are accompanied by an artery and a nerve (Fig. 2 and Additional file 1: Video S1). From these anatomical points, some key concepts advocate for the use of superficial veins instead of deep veins. Using superficial veins provides a short pathway to cannulation, leads to dwell a higher proportion of the catheter inside the vein (an issue intimately related to PCF, see below) and as a safe issue, avoids needle-stick injury of the artery or nerve. Cannulation of deep veins is also associated with a greater risk of catheter dislodgment when compared to cannulation of superficial veins [3, 8]. The superficial veins of the upper limbs which can be cannulated are the basilic (found medial, in the upper arm and in the forearm), and the cephalic (found lateral, in the upper arm and in the forearm).The median vein of the forearm and the median cubital vein are other veins which often have an adequate size for US-guided cannulation (Fig. 3).

Key concept 2: Select patent veins

A fully patent vein is sine qua non for cannulation. This is demonstrated by applying slight compression forces over the skin with the transducer and observing veins that collapse easily (Additional file 2: Video S2). In contrast, a thrombosed vein is partially or totally non-compressible, is filled with thrombotic material (Additional file 3: Video S3) and thus is discarded for cannulation. After applying a tourniquet, a stagnant blood flow may be observed within the vein lumen in two-dimensional imaging, and this should not be confused with a thrombus (Additional file 4: Video S4). Distal compression aids in clearing this stagnant blood from the vein and ruling out a thrombus when vein patency is in doubt. Since peripheral veins have low blood flow velocity, spontaneous signal may not be observed in color Doppler. In these cases, distal compression allows to squeeze the blood from the vein, elevate blood flow velocity, thus aiding in the demonstration of flow in patent veins.

Key concept 3: Determine vein size–catheter size

An optimal vein size is required to improve cannulation, and the size recommended in the literature is at least 4 mm in anteroposterior (AP) diameter [9] (Fig. 4). This suggested vein diameter, although important, should not be used in practice in a strict manner, since smaller veins can still be cannulated successfully, as seen in some studies [4, 6]. The vein size fulfills not only an established role in cannulation success (i.e., large veins are easily observed, as well as the needle within the vessel), but also aids in guiding catheter selection. As a rule of thumb, AP diameter indicates the upper limit of the external diameter of the catheter which can be used, considering that up to one-third of the vein lumen should be occupied by the catheter [10]. Thus, for example, a 4-Fr catheter (with an external diameter of 1.3 mm) is the maximum size for a 4-mm vein.

Key concept 4: Determine vein depth–insertion angle–catheter length

As mentioned before, apart from avoiding the injury of the artery or nerve, practitioners should select superficial veins to guarantee a short pathway to cannulate the vein. The maximum suggested distance from skin to vein is < 16 mm [9, 11], while < 12 mm can be considered ideal [1] (Fig. 5). This distance supposes a 90° needle insertion related to the vessel, and thus, the real or “corrected” distance of the needle traveling to reach the vein can be approached performing Pythagorean assumptions, which are entirely true for a needle insertion angle of 45°. This distance is equal to 1.4 multiplied by the vertical distance (Fig. 6). For example, a vertical distance of 12 mm equals 16.8 mm using a 45° insertion. However, in practice, this length varies with the use of shallower (increased distance) or more sloped (decreased distance) needle insertions. Without the need to make calculations, practitioners can directly get this distance in the long axis, offering a big picture regarding the real distance to reach the vein when using different angles of insertion (Fig. 7). Of note, knowing this distance is of paramount importance to minimize PCF, given that a large proportion of the catheter must dwell in the vein [1, 6], and thus, a large distance to vein will result in a large proportion of the catheter outside the vein using standard-length catheters (SLC). A recent investigation using SLC showed that when < 30% (or one-third) of the catheter resides in the vein, all catheters failed. On the other hand, when > 65% (or two-thirds) of the catheter resides in the vein, none of the catheters failed. When 30–64% of the catheter was in the vein, 32.4% of intravenous catheters were lost [6]. More recently, comparing standard 4.78-cm-long catheters versus 6.35-cm-long catheters, Bahl et al. showed a significantly increased catheter survival when > 2.75 cm of the catheter resided in the vessel [12] (Fig. 8). This is coherent with the previous study, since 2.75 cm is closest to the 65% of an SLC. Thus, as a rule of thumb, achieving at least 2.75 cm of the catheter dwelling in the vein should be the cut-off used to mitigate catheter failure. This means, for example, that for an SLC of 4.78 cm, the “real” distance to reach the vein must be lower than 2 cm. To achieve this, several strategies can be used, for example, selecting vessels at the lowest possible depth, using sloped insertion angles, and inserting catheters which are longer than usual, such as ultra-long peripheral catheters (ULPC, 18-20G, 6.35 cm in length) and midline catheters (8–20 cm in length) [1, 8, 12, 13]. Using catheters which are longer than the standard size aids in minimizing PFC, allowing operators to use shallower insertion angles to improve needle visualization and also to select veins which are even deeper than 16 mm. Advantages of ULPC over midline are its low costs and the fact that they do not require advanced skills such as managing the Seldinger technique, so they can be inserted by nurses or technicians.

Key concept 5: Select the cannulation technique: out-of-plane or in-plane technique

Practitioners should remember that each technique has advantages and disadvantages and thus learning and using both techniques are encouraged (Additional file 5: Video S5 and Additional file 6: Video S6), since they can select one or the other based on the situation [14, 15]. In a recent systematic review and meta-analysis, greater success has been shown with the out-of-plane technique compared to the in-plane technique [16]. However, for the out-of-plane technique, the visualization of the needle tip is an important limitation, having shown a higher rate of posterior wall perforations, compared to in-plane technique, which shows a clear delineation of the needle shaft and needle tip as it is advanced from superficial tissues into the vein [17, 18]. Using the “walk-down” maneuver (i.e., “follow the tip technique”) improves visualization of the needle tip when using out-of-plane insertions [14] and should be considered for using in practice. Side-lobe artifact is common when performing the in-plane technique, which simulates that the needle is inserted into the vein lumen, when is in fact close to it [14]. The learning curve for the in-plane technique seems to be longer compared to the out-of-plane technique [15].

Key concept 6: Demonstrate the catheter is in the vein lumen and perform a saline flush test

After cannulating the vein, it is useful to check if the catheter is in the vein lumen, since is not infrequent that the infused solution passes easily to the subcutaneous tissue without any warning signs, thus delaying the institution of intravenous therapies. The catheter is observed as two parallel hyperechoic lines in the short, the long or both axes (Fig. 9a, b). In midline catheter insertions, the guidewire, seen as a hyperechoic linear structure, should be demonstrated before inserting the catheter (Fig. 9c). Finally, a saline flush test may be performed through the catheter, observing bubbles in the lumen vein in correctly positioned catheters (Additional file 7: Video S7); color Doppler can be used for this purpose as well [19].

Conclusions

Practitioners should consider several issues when inserting intravenous peripheral catheters under ultrasound guidance, aiming to improve success rate, avoid complications and lengthen the survival of the catheter. Based on available data and everyday practice, all indicate that catheters longer than standard size are needed for US-guided peripheral venous cannulation, with the purpose of minimizing premature catheter failure. This is a call for attention to catheter manufacturers, since a more affordable solution at hand is expected from them shortly.

Supplementary information

Supplementary information accompanies this paper at https://​doi.​org/​10.​1186/​s13089-019-0144-5.

Acknowledgements

None.
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Not applicable.

Competing interests

The author declares there are no competing interests
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Anhänge

Supplementary information

Literatur
1.
Zurück zum Zitat Bahl A, Hang B, Brackney A, Joseph S, Karabon P, Mohammad A, Nnanabu I, Shotkin P (2019) Standard long IV catheters versus extended dwell catheters: a randomized comparison of ultrasound-guided catheter survival. Am J Emerg Med 37(4):715–721CrossRefPubMed Bahl A, Hang B, Brackney A, Joseph S, Karabon P, Mohammad A, Nnanabu I, Shotkin P (2019) Standard long IV catheters versus extended dwell catheters: a randomized comparison of ultrasound-guided catheter survival. Am J Emerg Med 37(4):715–721CrossRefPubMed
2.
Zurück zum Zitat Costantino TG, Parikh AK, Satz WA, Fojtik JP (2005) Ultrasonography-guided peripheral intravenous access versus traditional approaches in patients with difficult intravenous access. Ann Emerg Med 46(5):456–461CrossRefPubMed Costantino TG, Parikh AK, Satz WA, Fojtik JP (2005) Ultrasonography-guided peripheral intravenous access versus traditional approaches in patients with difficult intravenous access. Ann Emerg Med 46(5):456–461CrossRefPubMed
3.
Zurück zum Zitat Keyes LE, Frazee BW, Snoey ER, Simon BC, Christy D (1999) Ultrasound-guided brachial and basilic vein cannulation in emergency department patients with difficult intravenous access. Ann Emerg Med 34(6):711–714CrossRefPubMed Keyes LE, Frazee BW, Snoey ER, Simon BC, Christy D (1999) Ultrasound-guided brachial and basilic vein cannulation in emergency department patients with difficult intravenous access. Ann Emerg Med 34(6):711–714CrossRefPubMed
4.
Zurück zum Zitat Gregg SC, Murthi SB, Sisley AC, Stein DM, Scalea TM (2010) Ultrasound-guided peripheral intravenous access in the intensive care unit. J Crit Care 25(3):514–519CrossRefPubMed Gregg SC, Murthi SB, Sisley AC, Stein DM, Scalea TM (2010) Ultrasound-guided peripheral intravenous access in the intensive care unit. J Crit Care 25(3):514–519CrossRefPubMed
5.
Zurück zum Zitat Shokoohi H, Boniface K, McCarthy M, Khedir Al-tiae T, Sattarian M, Ding R, Liu YT, Pourmand A, Schoenfeld E, Scott J, Shesser R, Yadav K (2013) Ultrasound-guided peripheral intravenous access program is associated with a marked reduction in central venous catheter use in noncritically ill emergency department patients. Ann Emerg Med 61(2):198–203CrossRefPubMed Shokoohi H, Boniface K, McCarthy M, Khedir Al-tiae T, Sattarian M, Ding R, Liu YT, Pourmand A, Schoenfeld E, Scott J, Shesser R, Yadav K (2013) Ultrasound-guided peripheral intravenous access program is associated with a marked reduction in central venous catheter use in noncritically ill emergency department patients. Ann Emerg Med 61(2):198–203CrossRefPubMed
6.
Zurück zum Zitat Pandurangadu AV, Tucker JM, Brackney AR, Bahl A (2018) Ultrasound-guided intravenous catheter survival impacted by amount of catheter residing in the vein. Emerg Med J 35(9):550–555CrossRefPubMed Pandurangadu AV, Tucker JM, Brackney AR, Bahl A (2018) Ultrasound-guided intravenous catheter survival impacted by amount of catheter residing in the vein. Emerg Med J 35(9):550–555CrossRefPubMed
7.
Zurück zum Zitat Dargin JM, Rebholz CM, Lowenstein RA, Mitchell PM, Feldman JA (2010) Ultrasonography-guided peripheral intravenous catheter survival in ED patients with difficult access. Am J Emerg Med 28(1):1–7CrossRefPubMed Dargin JM, Rebholz CM, Lowenstein RA, Mitchell PM, Feldman JA (2010) Ultrasonography-guided peripheral intravenous catheter survival in ED patients with difficult access. Am J Emerg Med 28(1):1–7CrossRefPubMed
8.
Zurück zum Zitat Mills CN, Liebmann O, Stone MB, Frazee BW (2007) Ultrasonographically guided insertion of a 15-cm catheter into the deep brachial or basilic vein in patients with difficult intravenous access. Ann Emerg Med 50(1):68–72CrossRefPubMed Mills CN, Liebmann O, Stone MB, Frazee BW (2007) Ultrasonographically guided insertion of a 15-cm catheter into the deep brachial or basilic vein in patients with difficult intravenous access. Ann Emerg Med 50(1):68–72CrossRefPubMed
9.
Zurück zum Zitat Witting MD, Schenkel SM, Lawner BJ, Euerle BD (2010) Effects of vein width and depth on ultrasound-guided peripheral intravenous success rates. J Emerg Med 39(1):70–75CrossRefPubMed Witting MD, Schenkel SM, Lawner BJ, Euerle BD (2010) Effects of vein width and depth on ultrasound-guided peripheral intravenous success rates. J Emerg Med 39(1):70–75CrossRefPubMed
10.
Zurück zum Zitat Tanabe H, Takahashi T, Murayama R, Yabunaka K, Oe M, Matsui Y, Arai R, Uchida M, Komiyama C, Sanada H (2016) Using ultrasonography for vessel diameter assessment to prevent infiltration. J Infus Nurs 39(2):105–111CrossRefPubMed Tanabe H, Takahashi T, Murayama R, Yabunaka K, Oe M, Matsui Y, Arai R, Uchida M, Komiyama C, Sanada H (2016) Using ultrasonography for vessel diameter assessment to prevent infiltration. J Infus Nurs 39(2):105–111CrossRefPubMed
11.
Zurück zum Zitat Panebianco NL, Fredette JM, Szyld D, Sagalyn EB, Pines JM, Dean AJ (2009) What you see (sonographically) is what you get: vein and patient characteristics associated with successful ultrasound-guided peripheral intravenous placement in patients with difficult access. Acad Emerg Med 16(12):1298–1303CrossRefPubMed Panebianco NL, Fredette JM, Szyld D, Sagalyn EB, Pines JM, Dean AJ (2009) What you see (sonographically) is what you get: vein and patient characteristics associated with successful ultrasound-guided peripheral intravenous placement in patients with difficult access. Acad Emerg Med 16(12):1298–1303CrossRefPubMed
12.
Zurück zum Zitat Bahl A, Hijazi M, Chen N-W, Chavette-Lachapelle L, Price J. Ultra long versus standard long peripheral intravenous catheters: a randomized controlled trial of ultrasound-guided catheter survival. SSRN: https://ssrn.com/abstract=3401988. Accessed 05 July 2019 Bahl A, Hijazi M, Chen N-W, Chavette-Lachapelle L, Price J. Ultra long versus standard long peripheral intravenous catheters: a randomized controlled trial of ultrasound-guided catheter survival. SSRN: https://​ssrn.​com/​abstract=​3401988. Accessed 05 July 2019
13.
Zurück zum Zitat Elia F, Ferrari G, Molino P, Converso M, De Filippi G, Milan A, Aprà F (2012) Standard-length catheters vs long catheters in ultrasound-guided peripheral vein cannulation. Am J Emerg Med 30(5):712–716CrossRefPubMed Elia F, Ferrari G, Molino P, Converso M, De Filippi G, Milan A, Aprà F (2012) Standard-length catheters vs long catheters in ultrasound-guided peripheral vein cannulation. Am J Emerg Med 30(5):712–716CrossRefPubMed
14.
Zurück zum Zitat Blanco P (2016) Ultrasound-guided vascular cannulation in critical care patients: a practical review. Med Intensiva 40(9):560–571CrossRefPubMed Blanco P (2016) Ultrasound-guided vascular cannulation in critical care patients: a practical review. Med Intensiva 40(9):560–571CrossRefPubMed
15.
Zurück zum Zitat Lamperti M, Bodenham AR, Pittiruti M, Blaivas M, Augoustides JG, Elbarbary M, Pirotte T, Karakitsos D, Ledonne J, Doniger S, Scoppettuolo G, Feller-Kopman D, Schummer W, Biffi R, Desruennes E, Melniker LA, Verghese ST (2012) International evidence-based recommendations on ultrasound-guided vascular access. Intensive Care Med 38(7):1105–1117CrossRefPubMed Lamperti M, Bodenham AR, Pittiruti M, Blaivas M, Augoustides JG, Elbarbary M, Pirotte T, Karakitsos D, Ledonne J, Doniger S, Scoppettuolo G, Feller-Kopman D, Schummer W, Biffi R, Desruennes E, Melniker LA, Verghese ST (2012) International evidence-based recommendations on ultrasound-guided vascular access. Intensive Care Med 38(7):1105–1117CrossRefPubMed
16.
Zurück zum Zitat Gottlieb M, Holladay D, Peksa GD (2018) Comparison of short- vs long-axis technique for ultrasound-guided peripheral line placement: a systematic review and meta-analysis. Cureus 10(5):e2718PubMedPubMedCentral Gottlieb M, Holladay D, Peksa GD (2018) Comparison of short- vs long-axis technique for ultrasound-guided peripheral line placement: a systematic review and meta-analysis. Cureus 10(5):e2718PubMedPubMedCentral
17.
Zurück zum Zitat Stone MB, Moon C, Sutijono D, Blaivas M (2010) Needle tip visualization during ultrasound-guided vascular access: short-axis vs long-axis approach. Am J Emerg Med 28(3):343–347CrossRefPubMed Stone MB, Moon C, Sutijono D, Blaivas M (2010) Needle tip visualization during ultrasound-guided vascular access: short-axis vs long-axis approach. Am J Emerg Med 28(3):343–347CrossRefPubMed
18.
Zurück zum Zitat Moon CH, Blehar D, Shear MA, Uyehara P, Gaspari RJ, Arnold J, Cukor J (2010) Incidence of posterior vessel wall puncture during ultrasound-guided vessel cannulation in a simulated model. Acad Emerg Med 17(10):1138–1141CrossRefPubMed Moon CH, Blehar D, Shear MA, Uyehara P, Gaspari RJ, Arnold J, Cukor J (2010) Incidence of posterior vessel wall puncture during ultrasound-guided vessel cannulation in a simulated model. Acad Emerg Med 17(10):1138–1141CrossRefPubMed
19.
Zurück zum Zitat Gottlieb M, Sundaram T, Holladay D, Nakitende D (2017) Ultrasound-guided peripheral intravenous line placement: a narrative review of evidence-based best practices. West J Emerg Med 18(6):1047–1054CrossRefPubMedCentralPubMed Gottlieb M, Sundaram T, Holladay D, Nakitende D (2017) Ultrasound-guided peripheral intravenous line placement: a narrative review of evidence-based best practices. West J Emerg Med 18(6):1047–1054CrossRefPubMedCentralPubMed
Metadaten
Titel
Ultrasound-guided peripheral venous cannulation in critically ill patients: a practical guideline
verfasst von
Pablo Blanco
Publikationsdatum
01.12.2019
Verlag
Springer Milan
Erschienen in
The Ultrasound Journal / Ausgabe 1/2019
Print ISSN: 2036-3176
Elektronische ISSN: 2524-8987
DOI
https://doi.org/10.1186/s13089-019-0144-5

Weitere Artikel der Ausgabe 1/2019

The Ultrasound Journal 1/2019 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.