Optimization of [18F]PSMA-1007 PET-CT using regularized reconstruction in patients with prostate cancer

Abstract

Background

Prostate-specific membrane antigen (PSMA) radiotracers such as [18F]PSMA-1007 used with positron emission tomography-computed tomography (PET-CT) is promising for initial staging and detection of recurrent disease in prostate cancer patients. The block-sequential regularization expectation maximization algorithm (BSREM) is a new PET reconstruction algorithm, which provides higher image contrast while also reducing noise. The aim of the present study was to evaluate the influence of different acquisition times and different noise-suppressing factors in BSREM (β values) in [18F]PSMA-1007 PET-CT regarding quantitative data as well as a visual image quality assessment.

We included 35 patients referred for clinical [18F]PSMA-1007 PET-CT. Four megabecquerels per kilogramme were administered and imaging was performed after 120 min. Eighty-four image series per patient were created with combinations of acquisition times of 1–4 min/bed position and β values of 300–1400. The noise level in normal tissue and the contrast-to-noise ratio (CNR) of pathological uptakes versus the local background were calculated. Image quality was assessed by experienced nuclear medicine physicians.

Results

The noise level in the liver, spleen, and muscle was higher for low β values and low acquisition times (written as activity time products (ATs = administered activity × acquisition time)) and was minimized at maximum AT (16 MBq/kg min) and maximum β (1400). There was only a small decrease above AT 10. The median CNR increased slowly with AT from approximately 6 to 12 and was substantially lower at AT 4 and higher at AT 14–16. At AT 4–6, many images were regarded as being of unacceptable quality. For AT 8, β values of 700–900 were considered of acceptable quality.

Conclusions

An AT of 8 (for example as in our study, 4 MB/kg with an acquisition time of 2 min) with a β value of 700 performs well regarding noise level, CNR, and visual image quality assessment.

Background

Prostate cancer remains one of the most common malignancies affecting men worldwide [1, 2]. Positron emission tomography-computed tomography (PET-CT) is a widely accepted imaging technique for initial staging and for detecting sites of disease recurrence in patients with prostate cancer; this has particularly been the case since the introduction of prostate-specific membrane antigen (PSMA) radiotracers [3]. PSMA has emerged as one of the most favourable targets for PET imaging, as it is a transmembrane protein that is highly overexpressed in prostate cancer cells. The most commonly used PSMA tracer in PET imaging is [68Ga]Ga-PSMA-11 but there is a growing interest in 18F-labelled PSMA agents, which offer advantages with respect to image resolution and production amount. One of the promising 18F-labelled PSMA radiotracers is [18F]PSMA-1007 [4].

A novel generation of PET scanners based on silicon (Si)-photomultiplier (PM) technology that was introduced to the market a few years ago can potentially improve pathology detection, due primarily to its higher sensitivity [5,6,7]. Concurrently, improved reconstruction algorithms such as the block-sequential regularization expectation maximization algorithm (BSREM) [8, 9], commercially known as Q.Clear (GE Healthcare, Milwaukee, WI, USA) [10], have been described. Compared to ordered subset expectation maximization (OSEM), BSREM can maintain a low noise level as the number of iterations increases. This reduces the need of “stopping early” as a strategy for noise control and allow for more iterations while keeping the noise level at clinical acceptable levels. The BSREM algorithm suppresses noise via a penalty factor β, in which higher values not only suppress noise more but also reduce resolution. Lesion detectability has been found to be equal or higher with BSREM as compared to OSEM [11, 12]. BSREM have also been shown to increase the quantitative accuracy of standardized uptake value (SUV) measurements, particularly in small lesions [9]..

To the best of our knowledge, the impact of the β factor, acquisition time, and administered activity for image quality has never been investigated for [18F]PSMA-1007. The aim of the present study was to evaluate the influence of different activity time products (ATs = administered activity × acquisition time; in our study we have kept administered activity fixed) and different β values (range 300–1400) on quantitative data as well as visually assessed image quality in [18F]PSMA-1007 SiPM-based PET-CT. A secondary aim was to find the lowest possible AT with adequate image quality.

Methods

PET-CT system

Four Discovery MI (GE Healthcare, Milwaukee, WI, USA) PET-CT systems were used for image acquisition at the Department of Clinical Physiology and Nuclear Medicine, Skåne University Hospital, Lund or Malmö, Sweden. The PET-CT systems were equipped with time-of-flight and configured with four rings of detector blocks with lutetium yttrium oxyorthosilicate crystals coupled to an array of SiPM. The crystal size was 4.0 × 5.3 × 25 mm3 and the PET detector had a transaxial field of view of 70 cm, an axial field of view of 20 cm, and an overlap of 24% between multi-bed positions (per vendor recommendation). The sensitivity was 13 cps/kBq according to NEMA standards. The PET-CT systems were cross-calibrated to a dose calibrator; the calibration is validated monthly using an SUV control with phantoms. The PET systems were combined with a 128-slice CT.

Patients and imaging

We included 35 patients aged 18 years or older, referred for clinical [18F]PSMA-1007 PET-CT due to initial staging (n = 14, all high-risk prostate cancer based on clinical T stage, prostate-specific antigen level and/or Gleason score) or biochemical recurrence (n = 21, mean prostate-specific antigen 3.6, range 0.2–16) of prostate cancer, between 16 September and 7 October 2019. The mean (± standard deviation, SD) weight was 93 ± 15 kg (range 62–124 kg), the mean BMI was 29 ± 4 (range 21–38), and the mean age was 70 ± 7 years (range 52–78 years).

Imaging was performed 120 min after radiotracer administration, and the patients were asked to empty the bladder before the image acquisition. The patients were scanned from the mid-thigh to the base of the skull. The mean administered [18F]PSMA-1007 activity was 4.0 ± 0.1 MBq/kg (range 4.3–3.8), while the mean accumulation time was 121 ± 3 min (range 115–129 min). The PET images were reconstructed using the commercially available BSREM algorithm Q.Clear (GE Healthcare, Milwaukee, WI, USA) including the time-of-flight and point spread function modelling with a 256 × 256 matrix (pixel size 2.7 × 2.7 mm2, slice thickness 2.8 mm). Images were acquired for 4 min/bed position in list mode. Sinograms with acquisition times of 1.0, 1.5, 2, 2.5, 3, 3.5, and 4 min/bed were created from the list files. These were reconstructed with 12 different β values, 300–1400 (increments of 100), thus yielding 84 image series per patient.

Administered activity and acquisition times are, to a close approximation, interchangeable so long as the count rate is within the linear part of the noise-equivalent count rate curve, which can be assumed in the case of clinical [18F]PSMA studies: 2 MBq/kg with an acquisition time of 4 min/bed is equivalent to 4 MBq/kg and 2 min/bed, assuming the same accumulation time between administration and scan time (2 h in this study). Therefore, going forward, we will use AT to refer to the product of the administered activity per unit of body weight and the acquisition time (MBq/kg × min/bed), to emphasize that the results do not depend on the acquisition time alone but, rather, on the combination of time and administered activity. However, it should be remembered that in this study, we used a fixed administered activity of 4 MBq/kg.

CT images were acquired for attenuation correction of the PET images and for anatomic correlation. A diagnostic CT with intravenous and oral contrast was performed. The tube current modulation was applied by adjusting the tube current for each individual with a noise index of 37.5 and a tube voltage of 100 kV. The slice thickness was 0.625 mm. The CT used for attenuation correction was acquired in the late venous phase. The adaptive statistical iterative reconstruction technique was applied.

This study was approved by the Regional Ethical Review Board (#2016/417) and was performed in accordance with the Declaration of Helsinki. All patients provided written informed consent.

Image analysis

Images were evaluated quantitatively (image noise, contrast-to-noise ratio (CNR), and SUVmax) and qualitatively (visual image assessment), and both the quantitative and qualitative analysis were taken into account for the subsequent overall recommendation on what AT and β value to choose.

Quantitative analysis

Two image quality metrics were used for quantitative analysis of the reconstructed PET images: image noise in normal tissue (liver, spleen, and muscle) and the CNR of pathological uptake sites versus the local background. In addition, the SUV in the liver, spleen, and muscle was calculated for all patients at AT 16 and β 500. Two patients were excluded from the quantitative analysis due to a lack of pathological uptakes.

The image noise level was calculated from regions of interest (ROIs) in the liver, spleen, and muscle, drawn on transaxial images using Hermes 2.0.0 (Hermes Medical Solutions, Stockholm, Sweden). Three 5-cm2 ROIs were drawn in transaxial slices with one slice in between; these measurements were averaged. None of the ROIs were placed where liver metastases or large vessels were seen. The ROIs were drawn in the 4 min image series and copied to the other series. The noise level was quantified as the coefficient of variation (COV), i.e. the ratio between the SD and the SUVmean.

CNR was calculated for up to two pathological uptakes per patient. If available, suspected lymph node metastases were chosen first. Otherwise, pathologic uptakes in the prostate or skeleton were selected. A volume of interest (VOI) was drawn over each lesion, covering the whole lesion and some margin, and a spherical local background VOI with a diameter of 1.5 cm was drawn next to the hotspot. CNR was calculated as the difference between SUVpeak in the lesion VOI and the SUVmean in the local background VOI, divided by the SD in the latter. SUVpeak in each lesion VOI was calculated as the average SUV in a 3 × 3 × 3 pixel cube, centred on the SUVmax in the VOI.

Qualitative analysis

Pilot study

To reduce the number of images to be visually evaluated for image quality, a pilot study was performed. Four patients were randomly selected for visual assessment of image quality. This evaluation was performed by two experienced nuclear medicine physicians in consensus. All 84 image series for each patient were viewed and the two to four best images from each acquisition time were selected. This assessment was performed only by visual analysis, not knowing the results from the quantitative analysis. By combining the results from all four patients, the images that were considered best from each acquisition time were chosen for further evaluation (AT of 4 with β 1000 and 1100; AT of 6 with β 800, 900, and 1000; AT of 8 with β 700, 800, and 900), with one image as a reference (AT of 16 with β 500). Images with acquisition times of over 2 min/bed position (AT >8) were, after this pilot study, not regarded as relevant for the further analysis aimed at finding the lower limit regarding AT.

Assessment of image quality

The eight image series from the pilot study were evaluated for image quality along with the images obtained with AT 16 and β 500 (regarded as a reference). Two nuclear medicine physicians individually evaluated the remaining 31 patients in a blinded fashion. The physicians were not aware of the reconstruction parameters or acquisition times used for each patient and were randomly given the nine image series from each patient. The images regarded as being of poor quality were marked, as was the image regarded as having the highest quality. Poor image quality was defined as either the presence of hotspots in the images due to noise, mimicking pathologic uptakes, or pathologic uptakes detected on the reference image (which was correctly identified for all patients by the physicians) but not detected on images with lower AT. The overall image quality was visually assessed, taking the noise level, sharpness, and lesion detectability into account.

Statistical analysis

Continuous patient parameters are presented as mean ± SD and categorical variables as a percent (%). Noise level, SUVmax, and SUVpeak were tested for normality using the Shapiro-Wilks test. Because the variables were not normally distributed, all quantitative PET data are shown as a median ± interquartile range at the 25th and 75th percentiles.

Results

Quantitative assessment

Figure 1 shows noise levels (COV) in the liver, spleen, and muscle. The COVs in the spleen and liver were of a similar magnitude and showed an essentially equal dependence on β and AT. The COV measured in the muscle was higher due to a lower uptake of [18F]PSMA-1007 and, therefore, had a stronger dependence on β and AT. In general, the noise level had a stronger dependence on β than on AT. As expected, image noise was minimized at maximum AT (16 MBq/kg min) and maximum β (1400) but there was only a small decrease above AT 10.

Fig. 1
figure 1

Image noise levels. Coefficient of variation (COV) in liver, spleen, and muscle ROIs, as a function of β value and activity time product (AT). The solid black line with circles represents the median of all patients (n = 33) and the grey shaded area shows the 25th–75th percentile range. The inserts show detailed views for AT = 8 MBq/kg

Figure 2 shows the CNR for all pathological lesions included in the analysis (n = 51). The median CNR increased slowly with AT from approximately 6 to 12, was substantially lower at 4, and was higher at AT 14–16. The 25th percentile CNR was relatively constant with increasing AT, implying that there was a subset of pathological lesions with low CNR which did not improve with increasing AT.

Fig. 2
figure 2

Contrast-to-noise ratio measured in pathological lesions. The median value (solid black line with circles) and 25th–75th percentile ranges (shaded grey area) are shown. The highest CNR for each AT is obtained at βmax, shown in the figure

The CNR curves for each AT had an optimal β value, which gave the highest median CNR. The optimal β appears to decrease with increasing AT, from 900 at AT = 4 to 400–500 at AT 12–16.

The median (25th–75th percentile range) SUVmean in the liver, spleen, and muscle was 13.1 (11.3–14.5), 10.6 (8.4–15.0), and 0.4 (0.4–0.5), respectively. As expected, no dependence on either AT or β could be noted for SUVmean.

Qualitative assessment

Pilot study

The best image series for each AT is shown in Table 1. In general, the 1-min acquisition time (AT of 4) was regarded as being of too poor image quality. However, two of these image series were selected for further evaluation because we wanted to find the lowest possible AT.

Table 1 Image quality from the pilot study. The β values for the image series that were ranked as the best in terms of overall image quality for each AT based on the pilot study of four patients. The image series highlighted in grey were further evaluated, in a blinded fashion, for image quality. Note that all 84 image series for each patient was visually assessed

Assessment of image quality

The two observers both ranked all the AT of 16 images as having the best image quality. The largest number of patients with unacceptable image quality was found for AT of 4, then gradually decreasing for AT of 6, AT of 8, and AT of 16 (Table 2). Figure 3 shows a patient example with images of all β values (300–1400 with increments of 100) for AT of 8. The SUVmax varied considerably between different β values. Figure 4 shows a patient example with selected ATs and β values. The PSMA-uptake in the prostate is clearly seen on the AT of 16, β 500 image. It is hardly visible on the AT of 4-images. Although the SUVmax does not differ much between the different images shown for AT of 4, 6, and 8, the lesion is more visible for AT of 6 and AT of 8 due to the lower noise level and higher CNR.

Table 2 Assessment of image quality. The number of patients with unacceptable image quality in [18F]PSMA-1007 PET-CT examinations assessed by two observers
Fig. 3
figure 3

Example of images with different β values for AT 8. The SUVmax stated for a small perirectal lymph node metastasis

Fig. 4
figure 4

Example of images with different AT and different β values. An uptake in the prostate is indicated, and the SUVmax stated

Discussion

The quantitative image analysis shows that, as expected, longer acquisition time (higher AT) resulted in higher image quality, i.e. lower noise levels and higher contrast-to-noise ratios. In terms of image noise, there was not much improvement from increasing the AT above 10 MBq/kg min (Fig. 1). This holds true even for tissues with low uptake, as shown by data from muscle ROIs. The median COV decreased rapidly with increasing β, especially for low acquisition times and low β values. As β increases, the slope of the COV curve decreases. The inflexion point appears around β 500–700, meaning that a change of β at a level above this value leads to a smaller difference in the COV than an equal change of β at a level below this value. This behaviour is illustrated in the inserts of Fig. 1, where it is seen that (for AT = 8), the COV in, e.g. muscle decreases by 13 % (0.26 → 0.22) for β 300 → 400, but only by 5 % (0.135 → 0.128) for β 1100 → 1200.

The median CNR for the pathological lesions improved slightly with AT, and some improvement was noted even when AT was increased from levels above 10. On the other hand, the 25th percentile CNR did not improve with AT, indicating that there was a subset of low-uptake lesions whose CNR did not increase with longer acquisition times. The median CNR curves for each value of AT had an optimal β, which decreased with an increasing acquisition time. Intuitively, this can be explained as follows: The CNR is decreased by high noise at the lowest β values and by a decreased SUVpeak at the highest β values. Figure 3 shows a clear example of decreasing lesion SUVmax with increasing β, from 16.5 at β = 300 to 3.1 at β = 1400. At the lowest β values, CNR is however decreased because of the high standard deviation in the local background (high noise). Since the noise decreases with increasing β, there is an optimum β somewhere in between where CNR is maximized. As AT increases, image noise decreases but SUVpeak or SUVmax is less affected. Thus, less noise regularization (lower β) is required to reach optimal CNR.

Considering the results of the quantitative analysis alone, it appears that an AT of 10 and a β value of approximately 600 is a reasonable choice of parameters.

One should aim for images of high quality but also consider the possibility of either decreasing administered activity or increasing patient throughput through a shorter acquisition time. Our qualitative analysis suggests using an AT of 8, as the number of unacceptable images was low. β values of 700–900 appear to perform equally well. In general, there was only a slight difference in the images analysed after the pilot study (Fig. 4). Highly avid tumours were seen regardless of reconstruction parameters but faint uptakes in small anatomical structures (such as lymph nodes or physiologic uptakes in ganglia) were less visible for low ATs.

When one combines the results from the quantitative and qualitative analyses, the use of an AT of 8 with a β value of 700 seems reasonable.

We found that the optimal β value for different ATs for CNR was lower compared with the preferred β value in the visual analysis. This has also been shown in a previous study for [18F]fluorocholine [13]. Thus, it seems that CNR alone cannot replace visual analysis by physicians.

Comparison with different β values and acquisition times for the BSREM reconstruction for [68Ga]Ga-PSMA-11 PET-CT has been investigated before [14, 15]. Ter Voert et al. [15] found that the best β value for pelvic [68Ga]Ga-PSMA-11 PET-CT was in the range of 400–550 for a 2 min acquisition time and a median injected dose of 128 MBq. Lindstrom et al. [14] found a slight overall preference for a β value of 900 after administration of 2 MBq/kg and an acquisition time of 2 min, but also a high interobserver variability. In both studies, the accumulation time was 60 min, as is recommended for [68Ga]Ga-PSMA-11 but not for [18F]PSMA-1007. Except for the current study, no similar studies for [18F]PSMA-1007 PET exist. A recent study [16] compared BSREM with OSEM for [18F]PSMA-1007, but only used a single β value and acquisition time after an accumulation time of 60 min. That study used a PM-based PET-CT system and a default producer setting for the BSREM with a β value of 350 and time-of-flight disabled. This was compared to OSEM using 3 iterations/18 subsets, a 5.5-mm filter and the time-of-flight was enabled. Under these conditions, they found higher SUV and tumour-to-background ratios for BSREM compared to OSEM, especially for small and highly avid lesions. Optimization of BSREM has also been published on other radiopharmaceuticals, which include [18F]fluorodeoxyglucose [17,18,19,20], [18F]fluorocholine [13], and [18F]fluciclovine [21], with different optimal β values and acquisition times. The differences depend on different bio-kinetics of the radiopharmaceutical, and different amounts of administered activity, which cause different amounts of signal in the images.

These findings should be viewed in light of some limitations. First, only a few patients were analysed in the pilot study and not all image series were individually assessed for image quality. The goal here was to test many reconstructions, therefore, it was not possible to assess all 84 image series per patient. Second, due to the large number of images, only two nuclear medicine physicians were available to interpret the images. It should also be noted that the results obtained in the present study are only valid for 2-h accumulation time and for [18F]PSMA-1007.

Conclusion

An AT of 8 (for example 4 MB/kg with an acquisition time of 2 min) with a β value of 700 performs well with respect to noise level, CNR and visual image quality assessments. If the AT is increased, a lower β value can be used.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AT:

Administered activity × acquisition time

BSREM:

Block-sequential regularization expectation maximization algorithm

CNR:

Contrast-to-noise ratio

COV:

Coefficient of variation

OSEM:

Ordered subset expectation maximization

PET-CT:

Positron emission tomography with computed tomography

PSMA:

Prostate-specific membrane antigen

ROI:

Region of interest

SiPM:

Silicon photomultiplier

SD:

Standard deviation

SUV:

Standardized uptake value

VOI:

Volume of interest

References

  1. Cornford P, Bellmunt J, Bolla M, Briers E, De Santis M, Gross T, et al. EAU-ESTRO-SIOG guidelines on prostate cancer. Part II: treatment of relapsing, metastatic, and castration-resistant prostate cancer. Eur Urol. 2017 Apr;71(4):630–42.

    Article  Google Scholar 

  2. Mottet N, Bellmunt J, Bolla M, Briers E, Cumberbatch MG, De Santis M, et al. EAU-ESTRO-SIOG guidelines on prostate cancer. Part 1: screening, diagnosis, and local treatment with curative intent. Eur Urol. 2017 Apr;71(4):618–29.

    Article  Google Scholar 

  3. Perera M, Papa N, Christidis D, Wetherell D, Hofman MS, Murphy DG, et al. Sensitivity, specificity, and predictors of positive (68)Ga-prostate-specific membrane antigen positron emission tomography in advanced prostate cancer: a systematic review and meta-analysis. Eur Urol. 2016 Dec;70(6):926–37.

    Article  Google Scholar 

  4. Giesel FL, Hadaschik B, Cardinale J, Radtke J, Vinsensia M, Lehnert W, et al. F-18 labelled PSMA-1007: biodistribution, radiation dosimetry and histopathological validation of tumor lesions in prostate cancer patients. Eur J Nucl Med Mol Imaging. 2017 Apr;44(4):678–88.

    Article  CAS  Google Scholar 

  5. Hsu DF. Ilan E. Uribe J, Lubberink M, Levin CS. Studies of a next generation silicon-photomultiplier-based time-of-flight PET/CT system. J Nucl Med: Peterson WT; 2017 Apr 27.

    Google Scholar 

  6. Wagatsuma K, Miwa K, Sakata M, Oda K, Ono H, Kameyama M, et al. Comparison between new-generation SiPM-based and conventional PMT-based TOF-PET/CT. Phys Med. 2017 Oct;42:203–10.

    Article  Google Scholar 

  7. Economou Lundeberg J, Oddstig J, Bitzen U, Tragardh E. Comparison between silicon photomultiplier-based and conventional PET/CT in patients with suspected lung cancer-a pilot study. EJNMMI Res. 2019 Sep 9;9(1):35.

    Article  Google Scholar 

  8. de Pierro AR, Beleza Yamagishi ME. Fast EM-like methods for maximum "a posteriori" estimates in emission tomography. IEEE Trans Med Imaging. 2001 Apr;20(4):280–8.

    Article  Google Scholar 

  9. Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, Gleeson FV. Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT System. J Nucl Med. 2015 Sep;56(9):1447–52.

    Article  CAS  Google Scholar 

  10. Ross SQ. Clear, GE Healthcare. White paper. 2014.

  11. Wangerin KA, Ahn S, Wollenweber S, Ross SG, Kinahan PE, Manjeshwar RM. Evaluation of lesion detectability in positron emission tomography when using a convergent penalized likelihood image reconstruction method. J Med Imaging (Bellingham). 2017 Jan;4(1):011002.

    Article  Google Scholar 

  12. Parvizi N, Franklin JM, McGowan DR, Teoh EJ, Bradley KM, Gleeson FV. Does a novel penalized likelihood reconstruction of 18F-FDG PET-CT improve signal-to-background in colorectal liver metastases? Eur J Radiol. 2015 Oct;84(10):1873–8.

    Article  Google Scholar 

  13. Bjoersdorff M, Oddstig J, Karindotter-Borgendahl N, Almquist H, Zackrisson S, Minarik D, et al. Impact of penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm for (18)F-fluorocholine PET-CT regarding image quality and interpretation. EJNMMI Phys. 2019 Mar 21;6(1):5.

    Article  Google Scholar 

  14. Lindstrom E, Velikyan I, Regula N, Alhuseinalkhudhur A, Sundin A, Sorensen J, et al. Regularized reconstruction of digital time-of-flight (68)Ga-PSMA-11 PET/CT for the detection of recurrent disease in prostate cancer patients. Theranostics. 2019;9(12):3476–84.

    Article  CAS  Google Scholar 

  15. Ter Voert E, Muehlematter UJ, Delso G, Pizzuto DA, Muller J, Nagel HW, et al. Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical (68)Ga-PSMA PET/MR. EJNMMI Res. 2018 Jul 27;8(1):70.

    Article  Google Scholar 

  16. Witkowska-Patena E, Budzynska A, Gizewska A, Dziuk M, Walecka-Mazur A. Ordered subset expectation maximisation vs Bayesian penalised likelihood reconstruction algorithm in 18F-PSMA-1007 PET/CT. Ann Nucl Med. 2020 Jan;4.

  17. Howard BA, Morgan R, Thorpe MP, Turkington TG, Oldan J, James OG, et al. Comparison of Bayesian penalized likelihood reconstruction versus OS-EM for characterization of small pulmonary nodules in oncologic PET/CT. Ann Nucl Med. 2017 Oct;31(8):623–8.

    Article  CAS  Google Scholar 

  18. Lindstrom E, Sundin A, Trampal C, Lindsjo L, Ilan E, Danfors T, et al. Evaluation of penalized-likelihood estimation reconstruction on a digital time-of-flight PET/CT scanner for (18)F-FDG whole-body examinations. J Nucl Med. 2018 Jul;59(7):1152–8.

    Article  Google Scholar 

  19. Sah BR, Stolzmann P, Delso G, Wollenweber SD, Hullner M, Hakami YA, et al. Clinical evaluation of a block sequential regularized expectation maximization reconstruction algorithm in 18F-FDG PET/CT studies. Nucl Med Commun. 2017 Jan;38(1):57–66.

    Article  Google Scholar 

  20. Tragardh E, Minarik D, Almquist H, Bitzen U, Garpered S, Hvittfelt E, et al. Impact of acquisition time and penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm on a Si-photomultiplier-based PET-CT system for (18)F-FDG. EJNMMI Res. 2019 Jul 24;9(1):64.

    Article  Google Scholar 

  21. Teoh EJ, McGowan DR, Schuster DM, Tsakok MT, Gleeson FV, Bradley KM. Bayesian penalised likelihood reconstruction (Q.Clear) of (18)F-fluciclovine PET for imaging of recurrent prostate cancer: semi-quantitative and clinical evaluation. Br J Radiol. 2018 May;91(1085):20170727.

Download references

Acknowledgements

The staff of Clinical Physiology and Nuclear Medicine at Skåne University Hospital who performed the image examinations are greatly appreciated.

Funding

The Knut and Alice Wallenberg foundation, the Medical Faculty at Lund University, Region Skåne, and the Swedish Prostate Cancer foundation are acknowledged for their generous financial support.

Open access funding provided by Lund University.

Author information

Authors and Affiliations

Authors

Contributions

All authors made substantial contributions to the conception and design of the work and interpretation of data. ET and UB assessed image quality. JO, DM, GB, and BO obtained quantitative data. ET, DM, GB, and JO were major contributors in writing the manuscript. UB and BO revised the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Elin Trägårdh.

Ethics declarations

Ethics approval and consent to participate

These studies were approved by the Regional Ethical Review Board (#2016/417) and was performed in accordance with the Declaration of Helsinki. All patients provided written informed consent.

Consent for publication

All patients provided written informed consent.

Competing interests

The authors declare that they have no conflict of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Trägårdh, E., Minarik, D., Brolin, G. et al. Optimization of [18F]PSMA-1007 PET-CT using regularized reconstruction in patients with prostate cancer. EJNMMI Phys 7, 31 (2020). https://doi.org/10.1186/s40658-020-00298-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-020-00298-8

Keywords