header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Research

Contamination during removal of cement in revision hip arthroplasty

A CADAVER STUDY USING ULTRASOUND AND HIGH-SPEED CUTTERS



Download PDF

Abstract

Instruments used in surgery which rotate or vibrate at a high frequency can produce potentially contaminated aerosols. Such tools are in use in cemented hip revision arthroplasties. We aimed to measure the extent of the environmental and body contamination caused by an ultrasound device and a high-speed cutter.

On a human cadaver we carried out a complete surgical procedure including draping and simulated blood flow contaminated with Staphylococcus aureus (ATCC 12600). After cemented total hip arthroplasty, we undertook repeated extractions of cement using either an ultrasound device or a high-speed cutter. Surveillance cultures detected any environmental and body contamination of the surgical team.

Environmental contamination was present in an area of 6 x 8 m for both devices. The concentration of contamination was lower for the ultrasound device. Both the ultrasound and the high-speed cutter contaminated all members of the surgical team. The devices tested produced aerosols which covered the whole operating theatre and all personnel present during the procedure. In contaminated and infected patients, infectious agents may be present in these aerosols. We therefore recommend the introduction of effective measures to control infection and thorough disinfection of the operating theatre after such procedures.


Correspondence should be sent to Dr M. Nogler.

For access options please click here