header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Upper Limb

A multicentre study of the long-term results of using a flat-back polyethylene glenoid component in shoulder replacement for primary osteoarthritis



Download PDF

Abstract

We report the long-term clinical and radiological outcomes of the Aequalis total shoulder replacement with a cemented all-polyethylene flat-back keeled glenoid component implanted for primary osteoarthritis between 1991 and 2003 in nine European centres. A total of 226 shoulders in 210 patients were retrospectively reviewed at a mean of 122.7 months (61 to 219) or at revision. Clinical outcome was assessed using the Constant score, patient satisfaction score and range of movement. Kaplan-Meier survivorship analysis was performed with glenoid revision for loosening and radiological glenoid loosening (sd) as endpoints. The Constant score was found to improve from a mean of 26.8 (sd 10.3) pre-operatively to 57.6 (sd 20.0) post-operatively (p < 0.001). Active forward flexion improved from a mean of 85.3° (sd 27.4) pre-operatively to 125° (sd 37.3) postoperatively (p < 0.001). External rotation improved from a mean of 7° (sd 6.5) pre-operatively to 30.3° (sd 21.8°) post-operatively (p < 0.001). Survivorship with revision of the glenoid component as the endpoint was 99.1% at five years, 94.5% at ten years and 79.4% at 15 years. Survivorship with radiological loosening as the endpoint was 99.1% at five years, 80.3% at ten years and 33.6% at 15 years.

Younger patient age and the curettage technique for glenoid preparation correlated with loosening. The rate of glenoid revision and radiological loosening increased with duration of follow-up, but not until a follow-up of five years. Therefore, we recommend that future studies reporting radiological outcomes of new glenoid designs should report follow-up of at least five to ten years.


Correspondence should be sent to Dr A. Young; e-mail: drallanyoung@hotmail.com

For access options please click here