A Catalyzing Phantom for Reproducible Dynamic Conversion of Hyperpolarized [1-13C]-Pyruvate

  • Christopher M. Walker,

    Affiliation Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America

  • Jaehyuk Lee,

    Affiliation Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America

  • Marc S. Ramirez,

    Affiliation Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America

  • Dawid Schellingerhout,

    Affiliations Department of Cancer Systems Imaging, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America, Department of Diagnostic Radiology, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America

  • Steven Millward,

    Affiliation Department of Cancer Systems Imaging, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America

  • James A. Bankson

    jbankson@mdanderson.org

    Affiliation Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, United States of America

Abstract

In vivo real time spectroscopic imaging of hyperpolarized 13C labeled metabolites shows substantial promise for the assessment of physiological processes that were previously inaccessible. However, reliable and reproducible methods of measurement are necessary to maximize the effectiveness of imaging biomarkers that may one day guide personalized care for diseases such as cancer. Animal models of human disease serve as poor reference standards due to the complexity, heterogeneity, and transient nature of advancing disease. In this study, we describe the reproducible conversion of hyperpolarized [1-13C]-pyruvate to [1-13C]-lactate using a novel synthetic enzyme phantom system. The rate of reaction can be controlled and tuned to mimic normal or pathologic conditions of varying degree. Variations observed in the use of this phantom compare favorably against within-group variations observed in recent animal studies. This novel phantom system provides crucial capabilities as a reference standard for the optimization, comparison, and certification of quantitative imaging strategies for hyperpolarized tracers.

Introduction

Quantitative analysis of tissue structure, function, and composition through imaging and imaging biomarkers has the potential to help transform clinical care by facilitating personalized management of diseases such as cancer. Imaging measurements are particularly attractive because they can provide researchers and clinicians with non-invasive, serial views of heterogeneous tumor tissue in toto, in stark comparison to the limited number of samples that are practically available through traditional biopsy [1]. In order for imaging biomarkers to provide additional information that positively affects clinical decisions, they must be robust, reproducible, and reflective of relatively specific tissue characteristics. Biomarkers that provide insight into dynamic biological processes are particularly challenging due to the difficulty of establishing a reference truth against which comparisons can be made.

Magnetic resonance imaging (MRI), spectroscopy (MRS), and spectroscopic imaging (MRSI) technologies allow interrogation of anatomic, functional, and molecular characteristics of tissue without the use of ionizing radiation. Traditional MRI, which is fundamentally based on the detection of hydrogen nuclei (1H), is sensitive to a wide range of contrast mechanisms that lead to excellent soft-tissue contrast and functional imaging capabilities. Spectroscopic measurements exploit differences in the resonance frequency of nuclei with a nonzero spin, due to changes in the local magnetic field imparted by their molecular environments, to permit quantitative estimation of chemical concentrations in vivo. MRSI provides spectroscopic information at points within a slice or volume of interest with varying degrees of spatial and spectral resolution. A vast array of information is available through such measurements, including the ability to generate chemically selective maps of the spatial distribution of key metabolic compounds by 13C MRSI.

Unfortunately, magnetic resonance remains a noise-limited measurement and the tremendous clinical potential of MRS/MRSI techniques has not been realized largely due to an intrinsically low signal-to-noise ratio (SNR). SNR limitations are pronounced for MR of 13C in vivo due to the low natural abundance and MR sensitivity of the 13C isotope. Dynamic nuclear polarization (DNP) of 13C-enriched tracers enables dramatic increases in the excess spin population and thus the observable 13C MR signal [2], [3]. The hyperpolarized state created by DNP has been shown to increase SNR by approximately 104-fold compared to normal thermal equilibrium for select substrates such as [1-13C]-pyruvate [3]. This increase in signal has made in vivo spectroscopic imaging of hyperpolarized (HP) tracers feasible [4]. However, the signal from HP tracers is non-renewable and short-lived due to intrinsic relaxation mechanisms and signal depletion from necessary radiofrequency (RF) excitations, placing fundamental constraints on the measurement of their spatial and chemical fate. All acquisition protocols reflect a balance of acquisition time, the number and kind of signal excitations, and spatial, spectral, and temporal resolution. A range of data reduction techniques aimed at extracting more information from fewer signal excitations have been employed [5][13]. Optimization of acquisition strategies with respect to these unique constraints is ongoing, but assessment of the accuracy and reproducibility of these methods remains a challenge. Repeatable calibration standards would help characterize the accuracy and reproducibility of measurements under various data reduction strategies, a critical step in the translation of these powerful new biomarkers to clinical use.

13C-labeled pyruvate is the most studied hyperpolarized tracer to date because of its strong polarization, favorable kinetics, and central role in metabolism. The chemical fate of pyruvate is of particular interest in oncology because many cancers display the ‘Warburg effect’ and metabolize glucose, the precursor of pyruvate, using mechanisms that are less efficient for energy production but are thought to provide other advantages for survival and proliferation of disease [14], [15]. Preservation of nuclear spin state through chemical conversion allows us to observe the spatial distribution of HP-pyruvate and of HP metabolites such as lactate, alanine, and bicarbonate that can only arise through interactions between tracer and the relevant metabolic enzymes. Early studies have shown that a decrease in the conversion rate of HP pyruvate to lactate correlates with response to therapy and tumor grade [4], [16][19].

In this work, we demonstrate a phantom system in which the conversion of hyperpolarized pyruvate to lactate can proceed in the controlled, tunable, and reproducible environment of an isolated buffer. Hyperpolarized pyruvate is injected into a vessel containing the reagents that are necessary for reproducible conversion of pyruvate to lactate at a rate that is consistent with in vivo observations. This platform provides a robust tool for evaluation and comparison of spectroscopic and imaging measurements without the added complexity of a biological system. Many sources of variability inherent to the use of HP tracers in vivo are eliminated, allowing focused study of controllable parameters in order to maximize SNR and minimize measurement errors that can lead to artifacts. The ability to carry out these reactions at a known rate over a predetermined spatial distribution will significantly accelerate the optimization and validation of instrumentation and strategies for efficient signal acquisition, reconstruction, and analysis. This work focuses on [1-13C]-pyruvate due to its progress in clinical trials and near-term potential for clinical use [4]. Similar phantoms and calibration standards can be developed for other HP tracers, employing alternative enzyme systems.

Methods

Pyruvate is specifically converted into lactate by the enzyme lactate dehydrogenase (LDH) and the reduced form of the coenzyme nicotinamide adenine dinucleotide (NADH):(1)

This ordered ternary complex can be modeled using classical enzyme kinetics [20][23] to derive velocities (Mol/s) of the reaction as a function of constituent concentrations. The Gibbs free energy for this reaction is large (ΔG’0 = −25.1 kJ/mol), strongly favoring the forward reaction and lactate production; under normal physiological conditions, these reagents are involved in multiple cellular processes that modify their cytosolic concentration [24].

Under normal conditions for MRSI of HP lactate and pyruvate, the signal from 13C at thermal equilibrium is below the threshold of detectability, and only polarized labels can be observed. The rate of chemical conversion for HP tracers depends on the velocities of the chemical reaction and the likelihood that reactants are polarized:(2)(3)

Here, VPL denotes the forward velocity of the reaction, for conversion of pyruvate to lactate, and VLP represents the reverse reaction; L is the concentration of unpolarized lactate, P represents the concentration of unpolarized pyruvate, and the concentration of the polarized spin pools for each metabolite are indicated by asterisk. Witney et. al. [20] presented a model to describe the net conversion of pyruvate to lactate as well as the forward velocity of HP label exchange.

A. Modeling

Equations 23 can be modified to account for spin-lattice (T1) relaxation and losses due to signal excitation:(4)(5)

Where T1,pyr and T1,lac are the spin lattice relaxation time constants for pyruvate and lactate respectively, α is the flip angle of the RF excitation pulse, and TR is the repetition time. Using the models outline by Witney et. al. [20] and equations 4 and 5, a finite-difference approach was implemented in Matlab (The Mathworks, Natick, MA) to calculate conversion and exchange as a function of the total concentrations of pyruvate, lactate, LDH, NADH, and NAD+, and the percent polarization of the tracers. This model was used to determine phantom reagent concentrations that would reduce reaction rate variability.

Kinetic models with multiple compartments can be used to determine the flux of hyperpolarized tracers. To account for the conditions of an in vivo environment, models with two to six chemical and physical pools have been proposed [25][27]. One clear advantage of a single enzyme phantom is that the minimal kinetic model, employing only two chemical and no spatially separate pools within a voxel, can be applied without simplifying assumptions of unknown effect. Briefly, the equilibrium between pyruvate (Pyr) and lactate (Lac) can be described by:(6)

Where kPL and kLP are the unidirectional rate constants of the LDH catalyzed conversions. By defining and , and recognizing that observations are samples of the longitudinal magnetization of each pool, the two-compartment model can be expressed in a more suitable form [25], [28], [29]:(7)(8)

In contrast to the isolated phantom system where reagents can be accurately quantified, in vivo measurements based on this model reflect apparent rate constants due to the uncertainties in the cellular concentration of unlabeled reagents. Leveraging known reagent concentrations, time series of hyperpolarized pyruvate and lactate signal levels measured as described below were fit to this model using custom software developed in Matlab (The Mathworks, Natick, MA). Rate constants were determined by fitting dynamic tracer curves to equations 7 and 8 in the least squares sense, as previously described [25], [28], [29].

B. Phantom Preparation

Phantom concentrations were qualitatively optimized to reduce reaction rate sensitivity to variabilities in the concentrations of its components and ensure the reaction ran to completion before hyperpolarized signal decayed below threshold of detectability and as consistent with previous in vivo observations. Special consideration was given to reduce sensitivity due to pyruvate concentration and LDH activity as these were assumed to be the least reproducible characteristics of the phantom system. A custom phantom container was machined out of cylindrical Ultem resin stock and fitted with a 1 m polyethylene 3.175 mm diameter catheter (Coilhose Pneumatics, East Brunswick, NJ) for remote injection into the cavity when located at the isocenter of the magnet. The rectangular cavity was 1×1×3 cm with the injection catheter connecting to the front as seen in Figure 1. This structure was used for assessment of the enzymatic phantom using dynamic spectroscopy. To test the feasibility for such a reaction to be conducted inside a phantom with spatial details necessary to validate spectroscopic imaging sequences, a standard MRI quality assurance phantom (Aufloesung 30 KIT G, model 1P T58930; Bruker Biospin MRI, Ettlingen, Germany) was drained and refilled with the catalytic mixture described below.

thumbnail
Figure 1. A schematic view of the dynamic chemical phantom structure.

The injection and exhaust ports were fitted with catheters to facilitate rapid mixture of reagents at isocenter. A thin acrylic sheet was attached to the top to seal the fill cavity. This top could be removed to allow cleaning after injection. The phantom rested on a sled that allowed convenient removal and insertion of the phantom and included warm circulating water to maintain constant temperature.

https://doi.org/10.1371/journal.pone.0071274.g001

All reagents save pyruvate were thawed from aliquoted fresh solutions stored at −80°C and mixed in a 5 mL syringe shortly before completion of the polarizing process as described below. Once polarization was complete, HP pyruvate was injected into the phantom and followed by the enzyme substrate mixture to fill the phantom cavity. Nominal final concentrations were 2 mM hyperpolarized 13C-Pyruvate, 40 mM Lactate, 3.92 U/mL LDH (Worthington Biochemical Corp., Lakewood, NJ), and 4 mM Β-NADH (Sigma-Aldrich Corp., St. Louis, MO) in the Tris buffer (81.3 mM trisma preset crystals pH 7.2, 203.3 mM NaCl) (Sigma-Aldrich). This specific configuration places the phantom far from chemical equilibrium, rendering the reverse reaction (VLP, kLP) negligible throughout. The phantom was held at 28°C with a final pH of 7.2 and 3-mL final volume. UV spectrophotometry indicates that enzyme activity remains unchanged for at least two hours at room temperature.

C. Hyperpolarization

13 mg of [1-13C]-pyruvic acid (Sigma-Aldrich) with 15 mM OX063 (GE Healthcare, Fairfield, CT) and 0.0.775 mM Prohance (Bracco Diagnostics) were hyperpolarized in a HyperSense DNP system (Oxford Instruments) as described previously [3], [30]. The sample was dissolved in 4 mL of buffer consisting of 40 mM Trizma pre-set crystals (pH 7.6), 40 mM NaOH, 50 mM NaCl, and 0.1 mg/L EDTA. Once the dissolution process was complete, 0.15 mL of HP [1-13C]pyruvate (nominally 30% polarization) was drawn into a syringe for injection into the phantom.

D. Dynamic Spectroscopy

Dynamic spectroscopy was performed on a 7-T/30-cm Biospec System (Bruker Biospin Corp., Billerica, MA) using B-GA12 gradients (120-mm inner diameter (ID), Gmax = 400 mT/m) and a dual-tuned 1H/13C volume coil (72-mm ID, Bruker Biospin MRI). Dynamic 13C spectra were acquired with a 2.5 kHz bandwidth, 4098 points, 10° excitation using a 1 ms Gaussian pulse, 2-sec repetition time (TR), with 60 repetitions over a 3-min scan time beginning at dissolution and triggered by the HyperSense system. To evaluate performance and repeatability, the measurement was repeated seven times using identical reagent concentration.

Signal from each metabolite was determined by calculating the area of the Lorentzian line shape that fit most closely in the mean square sense. Dynamic signal curves for each tracer were integrated in time to estimate total signal from pyruvate and lactate. Their sum represents the total signal observed from all HP 13C-labeled metabolites in this phantom system. Signal amplitudes were normalized to account for variations in the amount of polarized pyruvate present at the onset of scanning. Three quantitative parameters were used to characterize the reaction rate for each measurement: total lactate signal normalized to total pyruvate, or to the sum of total pyruvate and lactate signal, and kPL, the forward rate constant for the two compartment model described by equations (7) and (8).

E. Spectroscopic Imaging

To demonstrate the usefulness of the enzyme phantom on evaluating spatial sequence performance, a 10-mL standard imaging phantom was drained and fitted with the same injection catheter described above. A slightly lower concentration of NADH (2 mM) was used in an otherwise identical mixture due to the increased phantom volume. A custom built dual-tuned 1H/13C linear birdcage coil with a 35 mm ID was used in conjunction with B-G6 gradients (60-mm ID, Gmax = 1000 mT/m, Bruker Biospin Corp.). The phantom was scanned with a radial echo planar spectral imaging (EPSI) sequence (unpublished). This was a single shot acquisition that expunged the entire hyperpolarized signal to acquire a single set of spectroscopic imaging data. The acquisition was started ∼40 seconds after all components were combined in the phantom, and data was acquired with a repetition time of 60 ms, initial echo time of 5.5 ms, and 1.3 ms echo spacing to form a 32 point echo train. A variable flip angle was used to maintain approximately equal sampling of longitudinal magnetization [31]. Spectral bandwidth was 23.8 kHz with a 744 Hz or 9.85 ppm spectral width. Fifty spatial projections were taken with 32 readout points over a 4 cm by 4 cm field of view and a 2 cm slice thickness.

Results

Isolating the enzymes in a system where initial reagent and product concentrations can be controlled allows the exchange rate to be specifically tuned and modeled without interference from other confounding biological processes. Simulations suggested that a phantom system containing 2 mM hyperpolarized [1-13C]-pyruvate, 40 mM lactate, 3.92 U/mL LDH, and 4 mM Β-NADH would progress at a rate that is consistent with in vivo observations [30] with reduced sensitivity to experimental variations. The phantom system, shown in Figure 1, was assembled and tested (N = 7 replicates), demonstrating reproducible conversion of hyperpolarized tracer as summarized in Figure 2. After a brief delay between start of data acquisition and injection of the polarized tracer, the pyruvate signal peaks quickly as it fills the chamber. A small frequency shift of 0.2 ppm, likely due to changes in susceptibility in the chamber, was observed during injection. Pyruvate signal decays due to relaxation, signal excitation, and chemical conversion to undetectable levels in less than two minutes. The lactate signal rises until the growth of the HP lactate pool is less than losses due to relaxation and signal excitation, at which point it similarly decays. The coefficient of variation for common measures of this reaction, summarized in Table 1, compare favorably to within-group variations reported in 9 recent animal studies [16][19], [32][37] as summarized in Table 2.

thumbnail
Figure 2. Dynamic signal evolution across (N = 7) injections.

The mean signal for lactate and pyruvate, normalized to peak carbon signal for each injection, are displayed with error bars that indicate the minimum and maximum values at each time over all injections. Total HP 13C was estimated by summing signal from HP 13C Lactate and HP 13C Pyruvate. The average linewidth for pyruvate and lactate peaks were 19±5 Hz and 17±5 Hz, respectively.

https://doi.org/10.1371/journal.pone.0071274.g002

thumbnail
Table 1. The mean, standard deviation and coefficient of variation for all repetitions (N = 7) of the dynamic phantom.

https://doi.org/10.1371/journal.pone.0071274.t001

thumbnail
Table 2. Survey of HP parameter variation in recent animal studies.

https://doi.org/10.1371/journal.pone.0071274.t002

Snapshot spectroscopic imaging of the separate imaging phantom shows relatively homogeneous mixture of components and distribution of tracer and metabolite 40 s after initiation of the reaction. Images of HP pyruvate and lactate, alone and in overlay over reference proton images, can be seen in Figure 3. While the resolution of the MRSI sequence is significantly lower than the proton image, it is possible to resolve features within spectroscopic images for both individual metabolites. No significant spatial distortions are seen.

thumbnail
Figure 3. Spectroscopic images of the reaction carried out in a standard imaging phantom.

Proton imaging (top left) shows the phantom structure in high resolution. Spectroscopic imaging data acquired using a radial EPSI sequence allows metabolite-specific visualization of tracer distribution (bottom row). Spectroscopic data can be intrinsically registered to high-resolution proton images (top center and right).

https://doi.org/10.1371/journal.pone.0071274.g003

Discussion

Real time in vivo spectroscopic measurement of labeled metabolites and their products is now possible with the integration of key polarizable tracers and carefully designed strategies for encoding, acquisition, reconstruction, and analysis of signals that are observable through magnetic resonance. Imaging the metabolic conversion of pyruvate into lactate shows promise as a clinically useful biomarker for oncology [4]. Hyperpolarized contrast agents comprise a relatively new and rapidly developing field, and research into best practices for signal acquisition, reconstruction, and analysis is ongoing. This dynamic phantom will provide robust, reproducible, and tunable dynamic processes against which experimental strategies can be compared and optimized.

This phantom system provides new capabilities for experimental development and validation with distinct advantages over single-tracer phantoms, static multi-compartment thermal equilibrium phantoms, and in-vivo models. Static phantoms are useful for confirming some functionality, such as the signal-to-noise ratio and spatial resolution associated with specific measurement settings, but do not create the dynamic conditions that could lead to artifacts in reconstruction algorithms that are based on the assumption of a stationary target. Assessment using in vivo models is challenging because of biological heterogeneity and the evolution of target processes in diseases such as cancer that can progress rapidly and increase within-group variations in a matter of days. With this platform, acquisitions can be readily repeated, at arbitrary intervals, to extract statistical measures of image quality.

To our knowledge, this is the first reproducibility study of a phantom system that provides controlled dynamic evolution of a chemical process that is observable by magnetic resonance using hyperpolarized tracers. This system catalyzes the final step in aerobic glycolysis, the conversion of pyruvate into lactate, without the need for animal subjects, human subjects, or cell suspensions that can increase the cost and the variability of technical measurements. The 12.3–19% variation that we observed is a result of many factors. LDH is sensitive to a range of experimental circumstances [38]; small variations in temperature, pH, or storage conditions can affect enzyme activity and therefore the rate of the reaction. In this pilot work, the injection of a small amount of hyperpolarized pyruvate was performed by hand, potentially leading to unnecessarily high variations in the end concentration of pyruvate. In future investigations, this error will be reduced by utilizing automated injections and a refined mixture and delivery system that reduces local fluctuations in the composition of the system [39].

The compatibility of this concept with a wide range of potential phantom structures provides clear opportunities for future use. Multiple compartments can easily be incorporated into new or existing phantom structures, and reaction rates in distinct regions can be tuned by varying reagent concentrations to simulate different tissues or disease states in parallel. In addition, a separate chamber that is free of enzyme would allow normalization to account for variations in tracer polarization. Because the spatial characteristics of such a phantom would be known a priori, rigorous evaluation and optimization of data encoding, acquisition, and reconstruction algorithms is possible. This is especially important when considering data reduction strategies that are designed to address key limitations in the measurement of hyperpolarized tracers but blur traditional definitions of spatial and temporal resolution in the observation of dynamic processes. Such a platform would be ideal for exploration of thresholds for detectability of pathologies that may not be evident in 1H MRI, early testing of new sequences to ensure preservation of spatial and temporal accuracy, and for quality assurance scans to confirm that similar acquisition, reconstruction, and analysis parameters lead to similar data over time both within and between laboratories and institutions..

Our phantom structure for dynamic spectroscopy (Figure 1) was designed using ultem (χV∼−8.9×10−6) to minimize susceptibility differences with water (χV∼−9.03×106) at interfaces that are normal to the static field. However, air (χV∼0.36×10−6) on one side of the rectangular volume of the chamber led to approximately 0.2 ppm shift in the resonant frequency of metabolites as the chamber was filled. More complex phantom structures will require careful design in order to minimize frequency shifts due to susceptibility differences that are modified as chambers are filled.

A crucial step in the translation of powerful new imaging technologies into routine preclinical and clinical use is the establishment of well-defined reference standards [40] to provide a common reference against which experimental circumstances can be compared. Such a reference can be used to ensure comparable results across platforms, laboratories, and institutions, and aid in study design and execution. The dynamic single enzyme phantom helps fill this critical need by providing reproducible HP tracer evolution that is independent of complex biological barriers and heterogeneity that cannot be strictly controlled. The physical structure of the phantom can be tailored to more closely approximate preclinical or clinical applications, and the rate of the reaction can be controlled through multiple compartments in a spatially dependent manner to simulate a wide range of disease states. This phantom platform represents a flexible and powerful tool to aid in development, optimization, validation, and certification of techniques, processes, and instrumentation that are crucial to ensure the successful and efficient translation of powerful new imaging capabilities afforded by MRSI of hyperpolarized tracers such as [1-13C]-pyruvate.

Acknowledgments

The authors are grateful to Dr. Marites P. Melancon and Dr. Chun Li for generous use of their laboratory equipment.

Author Contributions

Conceived and designed the experiments: CMW DS SM JAB. Performed the experiments: CMW JL MSR. Analyzed the data: CMW MSR DS SM JAB. Contributed reagents/materials/analysis tools: CMW MSR JAB. Wrote the paper: CMW JAB.

References

  1. 1. Morse DL, Gillies RJ (2010) Molecular imaging and targeted therapies. Biochem Pharmacol 80: 731–738.
  2. 2. Abragam A, Goldman M (1978) Principles of Dynamic Nuclear-Polarization. Reports on Progress in Physics 41: 395–467.
  3. 3. Ardenkjaer-Larsen JH, Fridlund B, Gram A, Hansson G, Hansson L, et al. (2003) Increase in signal-to-noise ratio of >10,000 times in liquid-state NMR. Proc Natl Acad Sci U S A 100: 10158–10163.
  4. 4. Kurhanewicz J, Vigneron DB, Brindle K, Chekmenev EY, Comment A, et al. (2011) Analysis of cancer metabolism by imaging hyperpolarized nuclei: prospects for translation to clinical research. Neoplasia 13: 81–97.
  5. 5. Hu S, Lustig M, Chen AP, Crane J, Kerr A, et al. (2008) Compressed sensing for resolution enhancement of hyperpolarized 13C flyback 3D-MRSI. J Magn Reson 192: 258–264.
  6. 6. Larson PE, Hu S, Lustig M, Kerr AB, Nelson SJ, et al. (2011) Fast dynamic 3D MR spectroscopic imaging with compressed sensing and multiband excitation pulses for hyperpolarized 13C studies. Magn Reson Med 65: 610–619.
  7. 7. Ohliger MA, Larson PE, Bok RA, Shin P, Hu S, et al.. (2013) Combined parallel and partial fourier MR reconstruction for accelerated 8-channel hyperpolarized carbon-13 in vivo magnetic resonance spectroscopic imaging (MRSI). J Magn Reson Imaging: In press.
  8. 8. Mayer D, Yen YF, Tropp J, Pfefferbaum A, Hurd RE, et al. (2009) Application of subsecond spiral chemical shift imaging to real-time multislice metabolic imaging of the rat in vivo after injection of hyperpolarized 13C1-pyruvate. Magn Reson Med 62: 557–564.
  9. 9. Levin YS, Mayer D, Yen YF, Hurd RE, Spielman DM (2007) Optimization of fast spiral chemical shift imaging using least squares reconstruction: application for hyperpolarized (13)C metabolic imaging. Magn Reson Med 58: 245–252.
  10. 10. Cunningham CH, Chen AP, Albers MJ, Kurhanewicz J, Hurd RE, et al. (2007) Double spin-echo sequence for rapid spectroscopic imaging of hyperpolarized 13C. J Magn Reson 187: 357–362.
  11. 11. Hu S, Lustig M, Balakrishnan A, Larson PE, Bok R, et al. (2010) 3D compressed sensing for highly accelerated hyperpolarized (13)C MRSI with in vivo applications to transgenic mouse models of cancer. Magn Reson Med 63: 312–321.
  12. 12. Arunachalam A, Whitt D, Fish K, Giaquinto R, Piel J, et al. (2009) Accelerated spectroscopic imaging of hyperpolarized C-13 pyruvate using SENSE parallel imaging. NMR Biomed 22: 867–873.
  13. 13. Larson PE, Bok R, Kerr AB, Lustig M, Hu S, et al. (2010) Investigation of tumor hyperpolarized [1–13C]-pyruvate dynamics using time-resolved multiband RF excitation echo-planar MRSI. Magn Reson Med 63: 582–591.
  14. 14. Vander Heiden MG, Cantley LC, Thompson CB (2009) Understanding the Warburg effect: the metabolic requirements of cell proliferation. Science 324: 1029–1033.
  15. 15. Gatenby RA, Gillies RJ (2004) Why do cancers have high aerobic glycolysis? Nat Rev Cancer 4: 891–899.
  16. 16. Albers MJ, Bok R, Chen AP, Cunningham CH, Zierhut ML, et al. (2008) Hyperpolarized 13C lactate, pyruvate, and alanine: noninvasive biomarkers for prostate cancer detection and grading. Cancer Res 68: 8607–8615.
  17. 17. Day SE, Kettunen MI, Gallagher FA, Hu DE, Lerche M, et al. (2007) Detecting tumor response to treatment using hyperpolarized 13C magnetic resonance imaging and spectroscopy. Nat Med 13: 1382–1387.
  18. 18. Bohndiek SE, Kettunen MI, Hu DE, Brindle KM (2012) Hyperpolarized (13)C spectroscopy detects early changes in tumor vasculature and metabolism after VEGF neutralization. Cancer Res 72: 854–864.
  19. 19. Park I, Bok R, Ozawa T, Phillips JJ, James CD, et al. (2011) Detection of early response to temozolomide treatment in brain tumors using hyperpolarized 13C MR metabolic imaging. J Magn Reson Imaging 33: 1284–1290.
  20. 20. Witney TH, Kettunen MI, Brindle KM (2011) Kinetic modeling of hyperpolarized 13C label exchange between pyruvate and lactate in tumor cells. J Biol Chem 286: 24572–24580.
  21. 21. Yagil G, Hoberman HD (1969) Rate of isotope exchange in enzyme-catalyzed reactions. Biochemistry 8: 352–360.
  22. 22. Zewe V, Fromm HJ (1962) Kinetic studies of rabbit muscle lactate dehydrogenase. J Biol Chem 237: 1668–1675.
  23. 23. Zewe V, Fromm HJ (1965) Kinetic Studies of Rabbit Muscle Lactate Dehydrogenase. Ii. Mechanism of the Reaction. Biochemistry 4: 782–792.
  24. 24. Sun F, Dai C, Xie J, Hu X (2012) Biochemical issues in estimation of cytosolic free NAD/NADH ratio. PLoS One 7: e34525.
  25. 25. Harris T, Eliyahu G, Frydman L, Degani H (2009) Kinetics of hyperpolarized 13C1-pyruvate transport and metabolism in living human breast cancer cells. Proc Natl Acad Sci U S A 106: 18131–18136.
  26. 26. Harrison C, Yang C, Jindal A, DeBerardinis RJ, Hooshyar MA, et al. (2012) Comparison of kinetic models for analysis of pyruvate-to-lactate exchange by hyperpolarized 13 C NMR. NMR Biomed 25: 1286–1294.
  27. 27. Kazan SM, Reynolds S, Kennerley A, Wholey E, Bluff JE, et al.. (2012) Kinetic modeling of hyperpolarized (13) C pyruvate metabolism in tumors using a measured arterial input function. Magn Reson Med: in press.
  28. 28. Brindle KM (1988) NMR Methods for Measuring Enzyme Kinetics in Vivo. Progress in Nuclear Magnetic Resonance Spectroscopy 20: 36.
  29. 29. Ward CS, Venkatesh HS, Chaumeil MM, Brandes AH, Vancriekinge M, et al. (2010) Noninvasive detection of target modulation following phosphatidylinositol 3-kinase inhibition using hyperpolarized 13C magnetic resonance spectroscopy. Cancer Res 70: 1296–1305.
  30. 30. Sandulache VC, Skinner HD, Wang Y, Chen Y, Dodge CT, et al. (2012) Glycolytic inhibition alters anaplastic thyroid carcinoma tumor metabolism and improves response to conventional chemotherapy and radiation. Mol Cancer Ther 11: 1373–1380.
  31. 31. Zhao L, Mulkern R, Tseng CH, Williamson D, Patz S, et al. (1996) Gradient-echo imaging considerations for hyperpolarized 129Xe MR. J Magn Reson B 113: 179–183.
  32. 32. Thind K, Chen A, Friesen-Waldner L, Ouriadov A, Scholl TJ, et al.. (2012) Detection of radiation-induced lung injury using hyperpolarized (13) C magnetic resonance spectroscopy and imaging. Magn Reson Med: in press.
  33. 33. Lodi A, Woods SM, Ronen SM (2013) Treatment with the MEK inhibitor U0126 induces decreased hyperpolarized pyruvate to lactate conversion in breast, but not prostate, cancer cells. NMR Biomed 26: 299–306.
  34. 34. Clatworthy MR, Kettunen MI, Hu DE, Mathews RJ, Witney TH, et al. (2012) Magnetic resonance imaging with hyperpolarized [1,4-(13)C2]fumarate allows detection of early renal acute tubular necrosis. Proc Natl Acad Sci U S A 109: 13374–13379.
  35. 35. Bohndiek SE, Kettunen MI, Hu DE, Witney TH, Kennedy BW, et al. (2010) Detection of tumor response to a vascular disrupting agent by hyperpolarized 13C magnetic resonance spectroscopy. Mol Cancer Ther 9: 3278–3288.
  36. 36. Matsumoto S, Saito K, Yasui H, Morris HD, Munasinghe JP, et al. (2013) EPR oxygen imaging and hyperpolarized 13C MRI of pyruvate metabolism as noninvasive biomarkers of tumor treatment response to a glycolysis inhibitor 3-bromopyruvate. Magn Reson Med 69: 1443–1450.
  37. 37. Lau AZ, Chen AP, Barry J, Graham JJ, Dominguez-Viqueira W, et al. (2013) Reproducibility study for free-breathing measurements of pyruvate metabolism using hyperpolarized (13) C in the heart. Magn Reson Med 69: 1063–1071.
  38. 38. Bergmeyer HU, Gawehn K (1983) Methods of Enzymatic Analysis. Weinheim; Deerfield Beach, FL: Verlag Chemie.
  39. 39. Reynolds S, Kazan S, Bluff J, Port M, Wholey E, et al. (2012) Fully MR-Compatible Syringe Pump for the Controllable Injection of Hyperpolarized Substrate in Animals. Applied Magnetic Resonance 43: 263–273.
  40. 40. Buckler AJ, Bresolin L, Dunnick NR, Sullivan DC, Aerts HJ, et al. (2011) Quantitative imaging test approval and biomarker qualification: interrelated but distinct activities. Radiology 259: 875–884.
  41. 41. Laustsen C, Ostergaard JA, Lauritzen MH, Norregaard R, Bowen S, et al. (2013) Assessment of early diabetic renal changes with hyperpolarized [1-(13) C]pyruvate. Diabetes Metab Res Rev 29: 125–129.