Impedance and Electrically Evoked Compound Action Potential (ECAP) Drop within 24 Hours after Cochlear Implantation

  • Joshua Kuang-Chao Chen,

    Affiliations Department of Otolaryngology, Cheng Hsin General Hospital, Taipei, Taiwan, Faculty of Medicine, School of Medicine, National Yang-Ming University, Taipei, Taiwan

  • Ann Yi-Chiun Chuang,

    Affiliation Mackay Memorial Hospital, Taipei, Taiwan

  • Georg Mathias Sprinzl,

    Affiliation Department of Oto-Rhino-Laryngology, Medical University Innsbruck, Innsbruck, Austria

  • Tao-Hsin Tung,

    Affiliation Department of Medical Research and Education, Cheng Hsin General Hospital, Taipei, Taiwan

  • Lieber Po-Hung Li

    lieber.li@msa.hinet.net

    Affiliations Department of Otolaryngology, Cheng Hsin General Hospital, Taipei, Taiwan, Faculty of Medicine, School of Medicine, National Yang-Ming University, Taipei, Taiwan, Integrated Brain Research Laboratory, Department of Medical Research and Education, Taipei Veterans General Hospital, Taipei, Taiwan

Abstract

Previous animal study revealed that post-implantation electrical detection levels significantly declined within days. The impact of cochlear implant (CI) insertion on human auditory pathway in terms of impedance and electrically evoked compound action potential (ECAP) variation within hours after surgery remains unclear, since at this time frequency mapping can only commence weeks after implantation due to factors associated with wound conditions. The study presented our experiences with regards to initial switch-on within 24 hours, and thus the findings about the milieus inside cochlea within the first few hours after cochlear implantation in terms of impedance/ECAP fluctuations. The charts of fifty-four subjects with profound hearing impairment were studied. A minimal invasive approach was used for cochlear implantation, characterized by a small skin incision (≈2.5 cm) and soft techniques for cochleostomy. Impedance/ECAP was measured intro-operatively and within 24 hours post-operatively. Initial mapping within 24 hours post-operatively was performed in all patients without major complications. Impedance/ECAP became significantly lower measured within 24 hours post-operatively as compared with intra-operatively (p<0.001). There were no differences between pre-operative and post-operative threshold for air-conduction hearing. A significant drop of impedance/ECAP in one day after cochlear implantation was revealed for the first time in human beings. Mechanisms could be related to the restoration of neuronal sensitivity to the electrical stimulation, and/or the interaction between the matrix enveloping the electrodes and the electrical stimulation of the initial switch-on. Less wound pain/swelling and soft techniques both contributed to the success of immediate initial mapping, which implied a stable micro-environment inside the cochlea despite electrodes insertion. Our research invites further studies to correlate initial impedance/ECAP changes with long-term hearing/speech performance.

Introduction

Previous research revealed little fluctuation of electrically evoked compound action potential (ECAP) thresholds over a period of five to six years after cochlear implantation (CI) [1]. It has been generally accepted that the measurement of CI electrode impedance (and thus ECAP) was not stabilized within the first days or even weeks post-operatively [2], [3]. Mostly this was thought to be due to ongoing physiological processes related to the implantation of a foreign body in the cochlea and the healing from the trauma related to the implant process [4], [5].The impact of cochlear implant insertion on peripheral auditory pathway, however, has never been addressed in terms of impedance and ECAP level variation within the first few hours after operation. The reason resided in the fact that trials to switch on a newly inserted CI device and launch the process of frequency mapping generally could not commence until weeks after cochlear implantation due to factors associated with wound conditions regardless of various surgical approaches currently utilized.

The difficulty in immediate initial mapping arises mainly from the poor compliance of patients caused by postoperative pain. A locally increased impedance of soft tissues due to subcutaneous edema/hematoma following the surgery furthermore hampered the communication between the receiver and electrodes of the CI apparatus. Air bubbles, blood clots, and bone dust that possibly went in the cochlea because of manipulation during operations all added to the intricacy of this procedure. Thus it is obvious that dilemmas raised here could be ascribed at least in part to the surgical procedure itself, including a generally large-scale skin wound and the “hard” technique frequently used for cochleostomy.

Commencement of electrical stimulation within 24 hours after cochlear implantation has long been performed in our department on a routine basis. By means of a modification of the commonly adopted process for CI operation, initial switch-on within 24 hours was made possible. The modified method in turn could promise a better wound condition with respect to having only a tiny amount of local pain/swelling, and very little contamination of perilymph by air bubbles, blood clots and/or bone dust on electrodes insertion.

In the present study, the milieus inside the cochlea within the first few hours after cochlear implantation in terms of impedance/ECAP fluctuations was disclosed for the first time by using the aforementioned minimal invasive approach for CI operation. The charts of fifty-four patients with a profound degree of hearing impairment were reviewed. Impedance and ECAP were evaluated. Effect of gender and age was also addressed.

Materials

Ethics Statement

The study conformed to the Declaration of Helsinki and the guidelines of the Institutional Ethics and Research Committee of the Cheng-Hsin General Hospital. Written informed consent was obtained from each participant or their parents with a protocol approved by the Institutional Ethics and Research Committee of the Cheng Hsin General Hospital.

Subjects (Table 1)

The charts of fifty-four subjects with a profound degree of hearing impairment (twenty-seven males and twenty-seven females; 1∼80 y/o, mean = 16.4) were studied (Table 1). No other neurological deficits were identified. Thirty-two subjects used Freedom (left = 13, right = 20), four subjects used N5 (left = 1, right = 3), and eighteen subjects used Nucleus 24 (left = 5, right = 13)(Cochlear™, Lane Cove NSW, Australia) cochlear implant system.

Surgical procedures (Figure 1)

Under general anesthesia, patients were put in a supine position with their heads turned 45° away from the operated side. The surgical area was well exposed. After draping and sterilization, a 2.5 cm skin incision just anterior to the hair line was made in post-auricular region. An anterior-based skin flap was elevated by entering the areolar tissue plane and dissecting all the way to the level of Henle's spine anteriorly. A posterior-based musculoperiosteal flap for housing the receiver of cochlear implant on skull was then created by entering the subperiosteal space from the incision at a distance of about 2 cm away from the posterior wall of external ear canal. The dissection was done posteriorly from the level of Henle's spine. A well was made only when needed and no sutures were necessary for the fixation, since the tension of the musculoperiosteal flap was enough to secure the implant in position. The channel for the cable, however, was made on a routine basis. To avoid possible tissue injuries, four custom-made hooks were used to expose the cortex of the mastoid bone instead of using a self-retaining instrument (e.g. mastoid retractor). A thin rectangular chip of cortex bone (about 20×10×3 mm in size) was first harvested intact from the mastoid with a chisel, and then a drill was used for mastoidectomy. The size of this chip would be decided on according to the degree of aeration in the mastoid cavity evaluated by pre-operative computed tomography (CT). Electrocautery was seldom used unless it was necessary throughout the whole procedure in order to minimize both pain and swelling caused by burn. In addition, repeated irrigation with normal saline is mandatory, since residual blood clots and/or bone dusts would lead to significant pain and swelling of the wound postoperatively. Soft techniques [6] were employed for the cochleostomy/round window approach after facial nerve was located/well-preserved by using a facial nerve monitor (Medtronics™, Minneapolis, USA) and facial recess was penetrated. Healon (Advanced Medical Optics™, Uppsala, Sweden) was used for the coverage of cochleostomy/round window to prevent possible loss as well as contamination of perilymph by air bubbles, blood clots and/or bone dusts on electrodes insertion. Method of atraumatic insertion including Advance Off-Stylet (AOS; Cochlear™, Lane Cove NSW, Australia) was applied for the introduction of implant into cochlea to prevent excessive trauma of hair cells. The AOS technique has been demonstrated to be atraumatic and is therefore an important element of soft technique [7]. Following electrodes insertion, the thin chip of cortex bone was used as a shielding coverage to shelter the defect of mastoid cavity for the protection of the wire inside. The wound was closed with 3–0 Vicryl (Ethicon™, New Jersey, USA) layer by layer in a primary intention. A pressure mastoid dressing was then applied. The dressing would be removed in the morning of the first postoperative day. After the initial switch-on of newly-implanted cochlear implant and frequency mapping in the afternoon, patients could always be discharged in the evening.

thumbnail
Figure 1. Surgical procedures for a minimal invasive approach of cochlear implantation aiming for immediate initial mapping.

(A) After draping and sterilization, a 2.5 to 3 cm skin incision was made in the post-auricular region at a distance of about 2 cm away from the posterior wall of external ear canal. (B) To avoid possible tissue injuries, four custom-made hooks instead of self-retain instrument (e.g. mastoid retractor) were used to expose the cortex of mastoid bone. (C) A thin piece of cortex bone (about 10×10×3 mm in size) harvested before the mastoidectomy was used as a shielding coverage for the protection of the wire inside mastoid cavity.

https://doi.org/10.1371/journal.pone.0071929.g001

Audiometric and Electrophysiological Exam

Impedance and ECAP were recorded intra-operatively as well as within 24 hours after the surgery, respectively. All participants underwent pure tone audiometry (PTA) with sound field exam in an acoustically-shielded room to determine the hearing threshold before (unaided) as well as after (unaided and CI-aided) the operation, using test frequencies between 250 Hz to 8000 Hz. The maximal output of the Grason-Stadler GS61audiometer (Grason-Stadler™, Minnesota, USA; calibrated in line with American National Standards Institute standards) was 120 dB HL for frequencies 250 to 4000 Hz, and 110 dB HL for 8000 Hz. A value of 5 dB was added to the maximal output threshold of the testing frequency where no responses could be found above the limit [8], [9], [10]. Elapsed time after operation for the postoperative PTA exam was about 6 months.

Data Analysis

Statistical analysis was performed using the software of SAS 8.1 (SAS Institute Inc., Cary, NC, USA). Differences between values of impedance/ECAP measured intra-operatively and within 24 hours post-operatively of channels number 1, 6, 11, 16, and 22 were analyzed using paired t-test. Gender and age (cut-off value at 18 y/o) differences in impedance/ECAP were evaluated using t-test. The threshold for statistical significance was set at P<0.05.

Results

The hookup between the external transmitting coil and the implanted antenna coil was done in all patients without any discomfort right after the compression dressing was removed, which in turn led to a successful switch-on of cochlear implants and thus initial mapping within 24 hours post-operatively. All patients were well-accustomed to their cochlear implants after initial mapping and usually were willing to keep hooking the transmitting coil up in their daily activities. No pre-operative medication except for one dose of prophylactic antibiotics thirty minutes before the operation was prescribed for all patients. None of the postoperative complications such as hematoma, swelling, infection, or flap failure were noted (Table 2). Only one patient asked for oral analgesics three hours after the surgery due to wound pain.

Impedance and ECAP both became significantly lower measured within 24 hours post-operatively (mean  = 5.1 kOhm at channel 1, 5.2 kOhm at channel 6, 5.7 kOhm at channel 11, 6.2 kOhm at channel 16, and 8.0 kOhm at channel 22 for impedance; mean  = 182.3 CL at channel 1, 172.0 CL at channel 6, 178.6 CL at channel 11, 162.2 CL at channel 16, and 152.5 CL at channel 22 for ECAP) than those measured intra-operatively (mean  = 8.3 kOhm at channel 1, 8.6 kOhm at channel 6, 9.6 kOhm at channel 11, 10.0 kOhm at channel 16, and 11.7 kOhm at channel 22 for impedance; mean  = 197.3 CL at channel 1, 195.2 CL at channel 6, 200.7 CL at channel 11, 185.3 CL at channel 16, and 175.2 CL at channel 22 for ECAP). There were no gender and age differences between pre-operative and post-operative impedance and ECAP values (Table 3, 4, Table S1 and S2, and Figure 2).

thumbnail
Figure 2. Variations of impedance (left column) and electrical compound action potential (ECAP; right column) measured intra-operatively and within 24 hours post-operatively.

(A) Impedance and ECAP results of all channels. (B) Impedance and ECAP results of selected channels (no 1, 6, 11, 16, and 22). Impedance and ECAP both became significantly lower measured within 24 hours post-operatively than those measured intra-operatively.

https://doi.org/10.1371/journal.pone.0071929.g002

thumbnail
Table 4. Electrically evoked compound action potential (ECAP) for all participants.

https://doi.org/10.1371/journal.pone.0071929.t004

There were no differences between pre-operative (mean  = 96 dB HL at 500 Hz, 103 dB HL at 1000 Hz, 106 dB HL at 2000 Hz, and 111 dB HL at 4000 Hz) and post-operative (mean  = 112 dB HL at 500 Hz, 118 dB HL at 1000 Hz, 120 dB HL at 2000 Hz, and 121 dB HL at 4000 Hz) threshold for air-conduction hearing (Table 5).

Discussion

One major and novel finding in this study was the significant drop in the level for impedance/ECAP in less than one day after cochlear implantation. Impedance/ECAP measured within 24 hours post-operatively decreased considerably as compared with those measured intra-operatively in our subjects. Our finding was thus in line with one previous study that post-implantation electrical detection levels significantly declined within days in some animals [11]. Due to the aforementioned difficulty of instant switch-on for cochlear implant inserted via a technique of traditionally large-incision approach, issues regarding the levels of electrical thresholds in the first few hours and/or days following electrodes insertion were dealt with only in animal models. To the best of our knowledge, this is the first research yet disclosing results about diminished impedance/ECAP within 24 hours subsequent to cochlear implantation in human beings.

Mechanism(s) underlying the reduced values of impedance/ECAP within one day after the surgery in our subjects remained unclear. One possibility was the interaction between the matrix enveloping the electrodes and the electrical stimulation of the initial switch-on [12]. It has been revealed that a tissue sheath would be created around electrodes shortly after the insertion [4], [13]. This sheath has been assumed to cause an increase in the impedance/ECAP of the individual electrode contacts. Upon the initial switch-on, however, the electric stimulation could lead to the reorganization or “blow-out (possibly due to bubbles formation)” of the thin and newly-formed tissue sheath [12], [14]. It is thus reasonable to conjecture that the changes in the status of tissue sheath in turn would result in a decrease of the impedance/ECAP, since the grade and continuity of tissue growth surrounding the implant have been shown to be proportionately associated with the increase of the impedance [14], [15].

One more plausible explanation is a spontaneous recovery of micro-environment inside cochlea from that disturbed by electrodes insertion [11], which in turn could lead to a restoration of the original level in impedance and thus ECAP. Histological alterations have been shown to be present within hours after chemical and/or physical trauma to the cochlea in animals [16], [17]. The foreign body reaction combined with other physiological responses within the cochlea, which were thought to be related to the process of drilling around the cochlea/round window niche/cochleostomy [18], would furthermore add to the intricacy of impedance/ECAP measurement. Some of these changes such as swelling of auditory nerve fibers, which could cause a decrease in the sensitivity of neurons to electrical signals, were found to be reversible shortly following the insulting event [19]. The reversibility might hence help to resume the sensitivity of neuronal membranes to the electrical stimulation from cochlear implant, as could be indexed by the post-implantation decrease of electrical thresholds observed in the above-mentioned animal study [11]. In the present study, adverse effects on cochlea resulting from extreme heat produced by the drilling process would mostly be averted by using soft techniques during the course of cochleostomy/round window approach. Healon was then used for the coverage of cochleostomy/round window to prevent possible loss as well as contamination of perilymph by air bubbles and blood clots/bone dusts on electrodes insertion. Furthermore, the skill of “atraumatic” insertion including AOS was applied to avoid excessive trauma of hair cells [7]. The employment of soft techniques and of “atraumatic” method for electrodes insertion could therefore promise a less disturbed micro-environment inside cochlea for the monitoring of impedance/ECAP.

Another possibility to consider could be the involvement of reciprocal interactions, similar to those noted in patients with idiopathic sudden sensorineural hearing loss [20], [21], [22], [23], between central and peripheral auditory pathways induced by the restitution of electrical signals from cochlear implant. It has been revealed that plastic changes in central auditory pathway, probably coupled with negative impacts on neurotransmission/neuromodulation, might be instantly provoked by pathological damages (e.g. electrodes insertion) to the cochlea in animals [24], [25]. The effect of intra-operative electrical stimulation from cochlear implant, on the contrary, could launch a corresponding reorganization for the mending in the central auditory pathway [24], [25] and/or tissue development in peripheral auditory pathway (i.e. the implanted cochlea) [4], [11], [26], [27]. Such reorganization and/or tissue formation would help to avoid further deterioration in the auditory pathways, which sequentially could initiate an improvement in the sensitivity of the whole auditory system as was mirrored by the diminished values of impedance and hence ECAP approximately 24 hours later.

In our study, 27 out of 54 patients came from regions outside the country (i.e. 50%, see Table 1). The strategy of immediate initial switch-on in turn shortened the duration of uncertainty/worry for patients as well as families about surgical outcome, since all of our subjects knew they have had a well-functioning cochlear implant in less than one day after the operation. Patients can always be able to go back to school or rehabilitation center of their hometown two days postoperatively with their “new ears”. Our modified techniques for cochlear implantation thus made possible a nonstop rehabilitation programs peri-operatively for these international implantees. Furthermore, earlier initial switch-on did not imply more frequent visits postoperatively. Due to the good wound condition, patients were only required to come back for routine checkup on 28th post-operative day after discharge.

In summary, a significant drop of impedance and ECAP within 24 hours after cochlear implantation was revealed in the present study. A minimal invasive approach for cochlear implantation made initial mapping and thus the measurement of impedance/ECAP within 24 hours after the surgery possible. Whilst the drop was noted for the first time in human beings within less than one day following electrodes insertion, it seemed to imply that the milieus of the disturbed cochlea could recover within 24 hours postoperatively which in turn could justify the rationale for the instant switch-on of a newly implanted CI device. The strategy of initial switch-on within 24 hours post-operatively also led to benefits such as fewer uncertainty/worry for patients/families about surgical outcomes and nonstop rehabilitation programs. Those benefits were especially important for international patients. Our research invites further studies on a larger group of implanted patients to validate the long-term scenario of impedance/ECAP variations. A longitudinal study is also needed to correlate impedance/ECAP changes at the very beginning of electrodes insertion and hearing/speech performance over time.

Supporting Information

Table S2.

Electrically evoked compound action potential (ECAP) for all participants.

https://doi.org/10.1371/journal.pone.0071929.s002

(DOC)

Acknowledgments

Special thanks to Dr. Robert Briggs for the inspiration in the design of the surgical procedures utilized for this study, and Ms Meei-Ling Kuan, Meng-Ju Lien, Mary Huang, Wen-Chen Chiu, Yi-Ping Chang, and Hsiu-Wen Chang for audiological assistance.

References

  1. 1. Brown CJ, Abbas PJ, Etlert CP, O'Brient S, Oleson JJ (2010) Effects of long-term use of a cochlear implant on the electrically evoked compound action potential. J Am Acad Audiol 21: 5–15.
  2. 2. Henkin Y, Kaplan-Neeman R, Muchnik C, Kronenberg J, Hildesheimer M (2003) Changes over time in electrical stimulation levels and electrode impedance values in children using the Nucleus 24M cochlear implant. Int J Pediatr Otorhinolaryngol 67: 873–880.
  3. 3. Henkin Y, Kaplan-Neeman R, Kronenberg J, Migirov L, Hildesheimer M, et al. (2006) A longitudinal study of electrical stimulation levels and electrode impedance in children using the Clarion cochlear implant. Acta Otolaryngol 126: 581–586.
  4. 4. Kawano A, Seldon HL, Clark GM, Ramsden RT, Raine CH (1998) Intracochlear factors contributing to psychophysical percepts following cochlear implantation. Acta Otolaryngol 118: 313–326.
  5. 5. Nadol JB Jr, Eddington DK, Burgess BJ (2008) Foreign body or hypersensitivity granuloma of the inner ear after cochlear implantation: one possible cause of a soft failure? Otol Neurotol 29: 1076–1084.
  6. 6. Friedland DR, Runge-Samuelson C (2009) Soft cochlear implantation: rationale for the surgical approach. Trends Amplif 13: 124–138.
  7. 7. Roland JT Jr (2005) A model for cochlear implant electrode insertion and force evaluation: results with a new electrode design and insertion technique. Laryngoscope 115: 1325–1339.
  8. 8. Balkany TJ, Connell SS, Hodges AV, Payne SL, Telischi FF, et al. (2006) Conservation of residual acoustic hearing after cochlear implantation. Otol Neurotol 27: 1083–1088.
  9. 9. Di Nardo W, Cantore I, Melillo P, Cianfrone F, Scorpecci A, et al. (2007) Residual hearing in cochlear implant patients. Eur Arch Otorhinolaryngol 264: 855–860.
  10. 10. Kiefer J, Gstoettner W, Baumgartner W, Pok SM, Tillein J, et al. (2004) Conservation of low-frequency hearing in cochlear implantation. Acta Otolaryngol 124: 272–280.
  11. 11. Su GL, Colesa DJ, Pfingst BE (2008) Effects of deafening and cochlear implantation procedures on postimplantation psychophysical electrical detection thresholds. Hear Res 241: 64–72.
  12. 12. Newbold C, Richardson R, Huang CQ, Milojevic D, Cowan R, et al. (2004) An in vitro model for investigating impedance changes with cell growth and electrical stimulation: implications for cochlear implants. J Neural Eng 1: 218–227.
  13. 13. Agrawal V, Newbold C (2012) Computer modelling of the cochlea and the cochlear implant: a review. Cochlear Implants Int 13: 113–123.
  14. 14. Charlet de Sauvage R, Lima da Costa D, Erre JP, Aran JM (1997) Electrical and physiological changes during short-term and chronic electrical stimulation of the normal cochlea. Hear Res 110: 119–134.
  15. 15. Clark GM, Shute SA, Shepherd RK, Carter TD (1995) Cochlear implantation: osteoneogenesis, electrode-tissue impedance, and residual hearing. Ann Otol Rhinol Laryngol Suppl 166: 40–42.
  16. 16. Dodson HC (1997) Loss and survival of spiral ganglion neurons in the guinea pig after intracochlear perfusion with aminoglycosides. J Neurocytol 26: 541–556.
  17. 17. Leake-Jones PA, O'Reilly BF, Vivion MC (1980) Neomycin ototoxicity: ultrastructural surface pathology of the organ of Corti. Scan Electron Microsc: 427–434.
  18. 18. Somdas MA, Li PM, Whiten DM, Eddington DK, Nadol JB Jr (2007) Quantitative evaluation of new bone and fibrous tissue in the cochlea following cochlear implantation in the human. Audiol Neurootol 12: 277–284.
  19. 19. Juiz JM, Rueda J, Merchan JA (1988) Reversible damage to the nerve fibres in the organ of Corti after surgical opening of the cochlea in the rat. Acta Otolaryngol 106: 29–33.
  20. 20. Li LPH, Shiao AS, Chen LF, Niddam DM, Chang SY, et al. (2006) Healthy-side dominance of middle- and long-latency neuromagnetic fields in idiopathic sudden sensorineural hearing loss. Eur J Neurosci 24: 937–946.
  21. 21. Po-Hung Li L, Shiao AS, Lin YY, Chen LF, Niddam DM, et al. (2003) Healthy-side dominance of cortical neuromagnetic responses in sudden hearing loss. Ann Neurol 53: 810–815.
  22. 22. Li LP, Chen KC, Lee PL, Niddam DM, Cheng CM, et al. (2013) Neuromagnetic index of hemispheric asymmetry predicting long-term outcome in sudden hearing loss. Neuroimage 64: 356–364.
  23. 23. Li LP, Shiao AS, Chen KC, Lee PL, Niddam DM, et al. (2012) Neuromagnetic index of hemispheric asymmetry prognosticating the outcome of sudden hearing loss. PLoS One 7: e35055.
  24. 24. Durham D, Park DL, Girod DA (2000) Central nervous system plasticity during hair cell loss and regeneration. Hear Res 147: 145–159.
  25. 25. Illing RB, Reisch A (2006) Specific plasticity responses to unilaterally decreased or increased hearing intensity in the adult cochlear nucleus and beyond. Hear Res 216–217: 189–197.
  26. 26. Duckert LG (1983) Morphological changes in the normal and neomycin-perfused guinea pig cochlea following chronic prosthetic implantation. Laryngoscope 93: 841–855.
  27. 27. Pfingst BE, Sutton D, Miller JM, Bohne BA (1981) Relation of psychophysical data to histopathology in monkeys with cochlear implants. Acta Otolaryngol 92: 1–13.