The IRIDICA PCR/Electrospray Ionization–Mass Spectrometry Assay on Bronchoalveolar Lavage for Bacterial Etiology in Mechanically Ventilated Patients with Suspected Pneumonia

  • Kristoffer Strålin ,

    kristoffer.stralin@ki.se

    Affiliations Department of Infectious Diseases, Karolinska University Hospital, Stockholm, Sweden, Unit of Infectious Diseases, Department of Medicine Huddinge, Karolinska Institutet, Stockholm, Sweden

  • Fredrik Ehn,

    Affiliation Department of Infectious Diseases, Karolinska University Hospital, Stockholm, Sweden

  • Christian G. Giske,

    Affiliations Department of Clinical Microbiology, Karolinska University Hospital, Stockholm, Sweden, Division of Clinical Microbiology, Department of Laboratory Medicine, Karolinska Institutet, Stockholm, Sweden

  • Måns Ullberg,

    Affiliations Department of Clinical Microbiology, Karolinska University Hospital, Stockholm, Sweden, Division of Clinical Microbiology, Department of Laboratory Medicine, Karolinska Institutet, Stockholm, Sweden

  • Jonas Hedlund,

    Affiliations Department of Infectious Diseases, Karolinska University Hospital, Stockholm, Sweden, Unit of Infectious Diseases, Department of Medicine Solna, Karolinska Institutet, Stockholm, Sweden

  • Johan Petersson,

    Affiliations Department of Anesthesiology and Intensive Care, Karolinska University Hospital Solna, Stockholm, Sweden, Section of Anaesthesiology and Intensive Care Medicine, Department of Physiology and Pharmacology, Karolinska Institutet, Stockholm, Sweden

  • Carl Spindler,

    Affiliations Department of Infectious Diseases, Karolinska University Hospital, Stockholm, Sweden, Unit of Infectious Diseases, Department of Medicine Solna, Karolinska Institutet, Stockholm, Sweden

  • Volkan Özenci

    Affiliations Department of Clinical Microbiology, Karolinska University Hospital, Stockholm, Sweden, Division of Clinical Microbiology, Department of Laboratory Medicine, Karolinska Institutet, Stockholm, Sweden

Abstract

We studied the diagnostic performance of the IRIDICA PCR/electrospray ionization–mass spectrometry (PCR/ESI-MS) assay applied on bronchoalveolar lavage (BAL) samples, from 51 mechanically ventilated patients with suspected pneumonia, in a prospective study. In 32 patients with X-ray verified pneumonia, PCR/ESI-MS was positive in 66% and BAL culture was positive in 38% (p = 0.045), and either of the methods was positive in 69%. The following BAL result combinations were noted: PCR/ESI-MS+/culture+, 34%; PCR/ESI-MS+/culture-, 31%; PCR/ESI-MS-/culture+, 3.1%; PCR/ESI-MS-/culture-, 31%; kappa 0.36 (95% confidence interval (CI), 0.10–0.63). In pneumonia patients without prior antibiotic treatment, optimal agreement was noted with 88% PCR/ESI-MS+/culture+ and 12% PCR/ESI-MS-/culture- (kappa 1.0). However, in patients with prior antibiotic treatment, the test agreement was poor (kappa 0.16; 95% CI, -0.10–0.44), as 10 patients were PCR/ESI-MS+/culture-. In 8/10 patients the pathogens detected by PCR/ESI-MS could be detected by other conventional tests or PCR tests on BAL. Compared with BAL culture, PCR/ESI-MS showed specificities and negative predictive values of ≥87% for all individual pathogens, an overall sensitivity of 77% and positive predictive value (PPV) of 42%. When other conventional tests and PCR tests were added to the reference standard, the overall PPV increased to 87%. The PCR/ESI-MS semi-quantitative level tended to be higher for PCR/ESI-MS positive cases with pneumonia compared with cases without pneumonia (p = 0.074). In conclusion, PCR/ESI-MS applied on BAL showed a promising performance and has potential to be clinically useful in mechanically ventilated patients with suspected pneumonia. The usefulness of the method for establishment of pneumonia etiology and selection of antibiotic therapy should be further studied.

Introduction

In severe pneumonia, selection of appropriate antibiotic therapy is essential for survival [1]. However, antibiotic selection is challenging, as pneumonia can be caused by a wide range of different pathogens [2, 3]. If the microbiological cause is detected, the patient can be given optimal, targeted antibiotic therapy. Gram stain of lower respiratory tract secretions may provide a rapid microbiological diagnosis, but this method has been associated with considerable intra- and inter-reader variability [4]. Thus, culture of lower respiratory tract secretions is currently the method of choice for detection of pneumonia etiology in most institutions [5, 6]. This method has important disadvantages, such as an analytic time of at least 24 hours, inability to detect atypical pathogens, and risk of false negativity due to antibiotic pre-treatment [5, 7]. Molecular methods are promising alternative diagnostic methods, as they may provide results within a few hours, and as they have been found to identify more microorganisms than conventional culture [8, 9].

PCR/electrospray ionization–mass spectrometry (PCR/ESI-MS) is a very interesting new molecular method, as it enables detection of DNA from >750 different bacterial species in a single test [8, 9]. The first PCR/ESI-MS system for bacterial DNA developed by Abbott, PLEX-ID, was evaluated on bronchoalveolar lavage (BAL) samples, but showed suboptimal performance [10]. Subsequently, the system was redesigned [11], and it showed improved performance when it was recently tested on lower respiratory tract secretions with a research-use instrument [12]. Compared with culture of the same samples in that study, PCR/ESI-MS showed sensitivity 84%, specificity 53%, positive predictive value (PPV) 58%, and negative predictive value (NPV) 81%. Culture was positive in 68/117 PCR/ESI-MS positive samples. For the 49 PCR/ESI-MS+/culture- cases, no additional microbiological data was provided for further evaluation [12].

In the end of 2014, Abbott´s new PCR/ESI-MS system, IRIDICA, was CE marked and became commercially available for in-vitro diagnostics in Europe. The aim of the present study was to evaluate the IRIDICA PCR/ESI-MS on BAL samples from consecutive patients with suspected pneumonia in the intensive care units (ICU) of a University Hospital. For evaluation of mismatching results, we compared the results with other conventional diagnostic tests and ran 16S PCR and species-specific PCR on the study BAL samples in selected cases. In addition, we performed analyses of cases with and without prior antibiotic therapy.

Materials and Methods

Patients

Samples were obtained from mechanically ventilated patients treated in the Central ICU, Thoracic ICU, Neurologic ICU, and the Extracorporeal Membrane Oxygenation Unit, of Karolinska University Hospital, Solna, Stockholm, Sweden, between 1 February and 31 May 2012. In consecutive patients who were subjected to bronchoscopy with low-volume BAL, based on the clinicians own decisions, the BAL samples were cultured and were then frozen at -80°C for future molecular diagnostics.

X-ray and/or computer tomography (CT) examinations were performed when clinically indicated. Patients were considered to have pneumonia if they had new or progressive pulmonary infiltrates on X-ray or CT, combined with clinical signs/symptoms of pneumonia, i.e. worsening of respiration and either increased volume or increased purulence of respiratory secretions. For the present study, we included pneumonia patients with BAL samples collected ± 48 h from the positive X-ray or CT examination. A patient was considered to have community-acquired pneumonia (CAP) if the clinical symptoms developed before hospital admission or within 48 h of hospital admission, and hospital-acquired pneumonia (HAP) if the clinical symptoms developed ≥48 h after hospital admission.

Patients subjected to bronchoscopy with BAL, who did not have new/progressive infiltrate combined with clinical signs of pneumonia, were considered not to have pneumonia. If repeated BAL samples were collected in those cases, only the first collected BAL sample was included in the present study.

Prior antibiotic treatment was defined as any antibiotics administered within the period 72 h–12 h prior to study bronchoscopy.

Bronchoalveolar lavage procedure

As per routine practice, the bronchoscope was inserted through the endotracheal tube and was wedged into a bronchus, most often the bronchus of the most affected pulmonary lobe. A low-volume BAL of 10–20 mL of physiologic saline solution was injected and was subsequently aspirated.

Microbiological analyses

All microbiological analyses were run at the Department of Clinical Microbiology, Karolinska University Laboratory, Karolinska University Hospital, Stockholm.

Culture of bronchoalveolar lavage.

The BAL samples were routinely cultured on two regular blood agar plates, blood agar containing gentiana violet, chocolate agar, and cysteine lactose electrolyte-deficient agar. All agar plates were incubated at 37°C for 48 h. Bacterial concentration, i.e. colony-forming units (CFU)/mL was determined at 48 h. The cut-off for a positive BAL culture was 102 CFU/mL.

Other conventional microbiological analyses.

From the microbiological records, we recorded results of additional lower respiratory tract cultures, blood culture (run with a Bact/ALERT blood culturing system (BioMérieux, Durham, NC, USA)), and the Streptococcus pneumoniae and Legionella pneumophila urinary antigen tests (BinaxNOW, Alere, Waltham, USA), performed ± 5 days from the study bronchoscopy.

IRIDICA PCR/electrospray ionization–mass spectrometry.

PCR/ESI-MS was performed from November 2014 to February 2015, with an IRIDICA BAC LRT system (Part Number: 08N40-010, Ibis Biosciences, Abbott, Des Plaines, USA), according to the manufacturer's instructions. This assay was designed to identify DNA from >750 different bacterial species, including the antibiotic resistance markers mecA, vanA, vanB, and kpc. Briefly, 0.1 mL of BAL fluid was diluted with Tris-EDTA buffered solution to a sample volume of 5 mL, and the sample was lysed using the IRIDICA bead-beater. DNA was extracted with the IRIDICA DNA Prep Kit, using the automated extraction system. Purified DNA in buffer was automatically distributed by the IRIDICA sample prep into the 16-well IRIDICA BAC LRT Assay Strips containing PCR reagents and primers for 16 PCR reactions. PCR was performed on the IRIDICA Thermal Cycler using a preloaded PCR amplification protocol. After PCR amplification, the IRIDICA BAC LRT Assay Stips were loaded onto the IRIDICA desalter, which purified DNA to remove substances which may interfere with mass spectrometry. Following desalting, plates were loaded onto the IRIDICA mass spectrometer. Purified amplicons were injected one well at a time into an electrospray ionization time-of-flight mass spectrometer for determination of the molecular mass of the amplicons. The resulting information was used for species identification by automated database comparison. The IRIDICA system had an analytic time from original sample to result of 6 h.

Semi-quantitative level determined by PCR/electrospray ionization–mass spectrometry.

Each well of the IRIDICA BAC LRT Assay Strip contains an internal control template at a known concentration. This template generates an amplicon with a known signature. Comparison of the relative concentrations of amplicons from this template and from detected organisms is reported as a “level”, which represents a semi-quantitative marker of the DNA content of the sample. For each positive PCR/ESI-MS result in the study, such a level was generated.

16S PCR and species-specific PCR.

After analysis with PCR/ESI-MS, BAL samples positive by PCR/ESI-MS or culture without support from other conventional diagnostic tests were subjected to direct 16S rRNA gene sequencing, with special focus on polybacterial samples and interpretation of mixed DNA chromatograms [13]. In patients with positive PCR/ESI-MS findings of S. pneumoniae, Staphylococcus aureus, and/or Haemophilus influenzae, with no support from conventional diagnostic tests, species-specific PCR [14, 15] was run.

Normal respiratory flora by culture, PCR/electrospray ionization–mass spectrometry, and 16S PCR.

In culture, PCR/ESI-MS, and 16S PCR of BAL, the following species were considered to belong to the normal respiratory flora; alpha-hemolytic streptococci, coagulase-negative staphylococci, Enterococcus spp., Neisseria spp., Corynebacterium spp., and Candida spp.

Normal respiratory flora was considered as negative in the analyses of the study.

Data and statistical analysis

Correlation and significance of relationships between the studied parameters was assessed using Chi-square test, Fisher´s exact test, and Mann-Whitney U test. p-values below 0.05 were considered statistically significant. For test agreement, a κ value with a 95% confidence interval (CI) was determined.

Ethical considerations

The Regional Ethics Committee in Stockholm approved the study and approved inclusion of patients in the study without written informed consent, as no intervention or additional procedures or samplings were performed (approval number, 2011/1774-31/2).

Results

Patients

Fifty-eight mechanically ventilated patients were subjected to bronchoscopy and 51 patients were included (Fig 1), including 32 patients with X-ray verified pneumonia (15 patients with CAP and 17 patients with HAP), and 19 patients without pneumonia. Table 1 shows patient characteristics. All CAP patients and 9 HAP patients (53%) were treated with antibiotics prior to study bronchoscopy.

thumbnail
Fig 1. Flow shart of the study patients.

ICU, intensive care unit; BAL, bronchoalveolar lavage.

https://doi.org/10.1371/journal.pone.0159694.g001

thumbnail
Table 1. Characteristics of included patients.

All patients were mechanically ventilated.

https://doi.org/10.1371/journal.pone.0159694.t001

BAL culture and PCR/electrospray ionization–mass spectrometry in pneumonia patients

Among 32 pneumonia patients, either BAL culture or PCR/ESI-MS was positive in 22 patients (69%), as BAL culture was positive in 12 patients (38%) and PCR/ESI-MS was positive in 21 patients (66%; p = 0.045). The result combinations of PCR/ESI-MS and BAL culture are shown in Table 2. PCR/ESI-MS+/culture- results were noted in 7/15 CAP cases (47%) and in 3/17 HAP cases (18%).

thumbnail
Table 2. Positive and negative results of IRIDICA PCR/electrospray ionization–mass spectrometry (PCR/ESI-MS) and culture of bronchoalveolar lavage in mechanically ventilated patients with pneumonia.

https://doi.org/10.1371/journal.pone.0159694.t002

Table 2 also shows the result combinations of patients with and without prior antibiotic treatment. The test agreement between PCR/ESI-MS and culture was high in patients without, and poor in patients with prior antibiotic treatment. Among cases without and with prior antibiotic treatment, BAL culture was positive in 88% (7/8) and 21% (5/24; p = 0.0016) and PCR/ESI-MS was positive in 88% (7/8) and 58% (14/24; p = 0.21), respectively. All 10 cases with PCR/ESI-MS+/culture- results had received prior antibiotic treatment.

Tables 3 and 4 show individual results of CAP and HAP patients. In 8/10 patients with PCR/ESI-MS+/culture- results (6 CAP and 2 HAP cases), the detected pathogens were also detected by conventional tests or PCR tests.

thumbnail
Table 3. Individual mechanically ventilated patients with community-acquired pneumonia with positive result obtained with either IRIDICA PCR/electrospray ionization–mass spectrometry (PCR/ESI-MS) or culture on bronchoalveolar lavage (BAL).

Support for positive PCR/ESI-MS results is indicated in bold.

https://doi.org/10.1371/journal.pone.0159694.t003

thumbnail
Table 4. Individual mechanically ventilated patients with hospital-acquired pneumonia with positive results obtained with either IRIDICA PCR/electrospray ionization–mass spectrometry (PCR/ESI-MS) or culture on bronchoalveolar lavage (BAL).

Support for positive PCR/ESI-MS results is indicated in bold.

https://doi.org/10.1371/journal.pone.0159694.t004

Compared with BAL culture, PCR/ESI-MS showed specificities of ≥87% and negative predictive values of ≥90% for all individual pathogens. The overall sensitivity was acceptable, 77% (10/13), but it should be noted that false-negative PCR/ESI-MS results were noted only for 3 cases of Enterobacter cloacae. The overall PPV was low, 42% (10/24). However, with an expanded reference standard, it increased. The PPV increased to 67% (16/24) when results of conventional diagnostic tests were added to the reference standard, and to 87% (21/24) when results of PCR tests were also added.

Table 5 shows combined results of culture and PCR/ESI-MS on BAL for individual pathogens.

thumbnail
Table 5. Combined results of IRIDICA PCR/electrospray ioization-mass spectrometry (PCR/ESI-MS) and culture on bronchoalveolar lavage (BAL) for individual pathogens (with > 1 positive result), in mechanically ventilated patients with and without pneumonia.

https://doi.org/10.1371/journal.pone.0159694.t005

Culture and PCR/electrospray ionization–mass spectrometry in patients without pneumonia

In 19 patients without pneumonia, BAL culture was positive in 5 patients (26%) and PCR/ESI-MS was positive in 10 patients (53%; p = 0.18), and either of the methods was positive in 10 patients (53%). The following combined test results were noted: PCR/ESI-MS+/culture+, n = 5 (25%); PCR/ESI-MS+/culture-, n = 5 (25%); PCR/ESI-MS-/culture+, n = 0; PCR/ESI-MS-/culture-, n = 9 (47%); κ for test agreement, 0.49 (95% CI, 0.15–0.82).

Individual microbiological results of patients without pneumonia are shown in Table 6. Among 5 patients with PCR/ESI-MS+/culture- results, the detected pathogen was supported by positive PCR tests in 2 cases. Table 5 shows combined results of PCR/ESI-MS and culture for different species.

thumbnail
Table 6. Individual mechanically ventilated patients without pneumonia with positive result obtained with either IRIDICA PCR/electrospray ionization–mass spectrometry (PCR/ESI-MS) or culture on bronchoalveolar lavage (BAL).

Support for positive PCR/ESI-MS results is indicated in bold.

https://doi.org/10.1371/journal.pone.0159694.t006

Semi-quantitative level of PCR/electrospray ionization–mass spectrometry

Semi-quantitative levels of PCR/ESI-MS are shown in Tables 34 and 6. The levels tended to be higher for 24 positive results of pneumonia patients than for 12 positive results of patients without pneumonia, p = 0.074 (Fig 2A). However, no correlation with semi-quantitative culture results or other microbiological results could be noted (Fig 2B).

thumbnail
Fig 2. Semi-quantitative levels of IRIDICA PCR/electrospray ionization-mass spectrometry (PCR/ESI-MS) on bronchoalveolar lavage (BAL).

(A) Results of cases with and without pneumonia; (B) Results related to other microbiological results. Black triangles, pneumonia; white triangles, not pneumonia. CFU, colony-forming units.

https://doi.org/10.1371/journal.pone.0159694.g002

Resistance genes detected by PCR/electrospray ionization–mass spectrometry

PCR/ESI-MS was positive for mecA in 19 study samples, including 5 samples PCR/ESI-MS positive for S. aureus and 14 samples with PCR/ESI-MS showing normal respiratory flora. Among 5 samples PCR/ESI-MS positive for S. aureus and mecA, BAL culture was positive for methicillin-sensitive S. aureus in 2 cases. A third patient was positive for methicillin-sensitive S. aureus in another lower respiratory tract culture. No methicillin-resistant S. aureus was isolated from any study patient.

No study sample was PCR/ESI-MS positive for vanA, vanB, or kpc, or culture positive for vancomycin resistant enterococci or carbapenemase secreting Gramnegatives.

Discussion

In this study of mechanically ventilated patients with suspected pneumonia, PCR/ESI-MS on BAL showed a promising performance for detection of bacterial etiology.

BAL is a suitable sample type for detection of microbiological cause in pneumonia, as BAL fluid covers respiratory epithelium at or close to the site of the infection. BAL culture has been found to reliably identify qualitatively and quantitatively microorganisms present in lung segments [16]. However, a problem in clinical practice is that antibiotics taken prior to sampling may produce false negative culture results. When Prats et al. [17] performed repeated bronchoscopy on culture positive patients with ventilator-associated pneumonia, the culture positivity rate decreased to 54% after 24 h and to 20% after 48 h of antibiotic treatment. Accordingly, the present study showed that BAL culture was significantly less often positive in patients with than in patients without prior antibiotic therapy.

Our group [18] previously found that PCR applied on BAL was significantly more often positive for S. pneumoniae than BAL culture (31% vs. 2.9%), in patients with prior antibiotic treatment. Baudel et al. [19] found that the multiplex PCR test LightCycler SeptiFast (Roche Diagnostics) on BAL was significantly more often positive than BAL culture in both antibiotic treated and un-treated patients. PCR was positive in 66% of their overall study and in 64% of antibiotic pre-treated patients. In the study by Vincent et al. [12], 75% of the patients were treated with antibiotics at enrolment, and PCR/ESI-MS was more often positive than culture on lower respiratory tract secretions (63% versus 44%).

Accordingly, the present study showed that PCR/ESI-MS was positive for bacterial pathogens more often than culture. When PCR/ESI-MS was added to culture, the frequency of pneumonia patients with a bacterial finding in BAL increased from 38% to 69%. As the positivity rate of culture and PCR/ESI-MS was identical in patients without prior antibiotic treatment, the improved positivity rate resulted exclusively from patients with prior antibiotic therapy (Table 2). The fact that the detected pathogen was identified by other diagnostic methods in 8/10 PCR/ESI-MS+/culture- pneumonia cases indicates that PCR/ESI-MS detected a pathogen that was or had recently been present in the lower respiratory tract. However, the study was not designed to evaluate if the detected pathogen represented the pathogen that caused the patient´s infection or if it was just a colonizer.

Colonization with respiratory pathogens is common in mechanically ventilated patients. Berdal et al. [20] collected repeated BAL samples from mechanically ventilated ICU patients who had any diagnosis other than pneumonia, and found pneumonia pathogens in BAL samples at ≥ 104 CFU/mL in 57% of the patients, including 36% of patients who did not develop X-ray infiltrates. Accordingly, many patients without pneumonia in the present study had positive culture and PCR/ESI-MS on BAL (Table 6).

In order to differentiate between infection and colonization, a cut-off limit of 104 CFU/mL has been proposed for BAL culture [16]. However, as rapid decreases of bacterial counts have been found during antibiotic treatment [17], it has been suggested that the cut-off limit should be decreased in antibiotic treated patients [21]. In addition, a recent meta-analysis did not find evidence that the use of quantitative cultures of respiratory secretions resulted in reduced mortality, reduced time in ICU and on mechanical ventilation, or higher rates of antibiotic changes, when compared to qualitative cultures in patients with ventilator-associated pneumonia [22]. When we studied semi-quantitative data of PCR/ESI-MS on BAL in the present study, there was only a tendency that pneumonia patients had higher semi-quantitative levels than cases without pneumonia (Fig 2A). No correlation with other microbiological results was noted (Fig 2B). Thus, the usefulness of semi-quantitative data of PCR/ESI-MS on BAL appears to be limited.

An interesting finding from this study was that PCR/ESI-MS was positive for S. aureus and mecA in 2 samples that were culture positive for methicillin-sensitive S. aureus. The most probable explanation is that the mecA genes were harboured by coagulase-negative staphylococci (i.e. normal flora) in the samples. As it is difficult to link resistance markers to detected bacterial pathogens by currently available molecular methods, combined detection of S. aureus and mecA in BAL samples by PCR/ESI-MS should be interpreted with caution.

An advantage with the PCR/ESI-MS technique compared with multiplex PCR with species-specific primers is that it can detect a large number of different pathogens, including unexpected pathogens. In this study, L. pneumophila and Tropheryma whipplei were detected by PCR/ESI-MS. Both pathogens were also detected by other diagnostic methods (Table 3). T. whipplei is an uncommon cause of pneumonia [23].

The major limitation of this study was its small sample size. Thus, we could not evaluate the performance of PCR/ESI-MS for detection of etiology in CAP or HAP, or for detection of individual pathogens.

We conclude that PCR/ESI-MS showed a promising performance and has potential to be clinically useful in mechanically ventilated patients with suspected pneumonia. The usefulness of the method for establishment of pneumonia etiology and selection of antibiotic therapy should be further studied.

Acknowledgments

Data from the manuscript was presented at the 55th Interscience Conference of Antimicrobial Agents and Chemotherapy (ICAAC), San Diego, USA, September 17–21, 2015 (Abstract D-227).

Author Contributions

Conceived and designed the experiments: KS FE CG JH JP CS VÖ. Performed the experiments: FE CG MU JP CS VÖ. Analyzed the data: KS FE CS VÖ. Contributed reagents/materials/analysis tools: CG MU VÖ. Wrote the paper: KS.

References

  1. 1. Iregui M, Ward S, Sherman G, Fraser VJ, Kollef MH. Clinical importance of delays in the initiation of appropriate antibiotic treatment for ventilator-associated pneumonia. Chest. 2002; 122:262–8. pmid:12114368
  2. 2. Jones RN. Microbial etiologies of hospital-acquired bacterial pneumonia and ventilator-associated bacterial pneumonia. Clin Infect Dis. 2010; 51 Suppl 1:S81–7. pmid:20597676
  3. 3. Musher DM, Thorner AR. Community-acquired pneumonia. N Engl J Med. 2014; 371:1619–28. pmid:25337751
  4. 4. Cooper GM, Jones JJ, Arbique JC, Flowerdew GJ, Forward KR. Intra and inter technologist variability in the quality assessment of respiratory tract specimens. Diagn Microbiol Infect Dis. 2000; 37:231–5. pmid:10974573
  5. 5. Guidelines for the management of adults with hospital-acquired, ventilator-associated, and healthcare-associated pneumonia. Am J Respir Crit Care Med. 2005; 171:388–416. pmid:15699079
  6. 6. Torres A, Ewig S, Lode H, Carlet J. Defining, treating and preventing hospital acquired pneumonia: European perspective. Intensive Care Med. 2009; 35:9–29. pmid:18989656
  7. 7. Cook DJ, Fitzgerald JM, Guyatt GH, Walter S. Evaluation of the protected brush catheter and bronchoalveolar lavage in the diagnosis of nosocomial pneumonia. J Intensive Care Med. 1991; 6:196–205. pmid:10147949
  8. 8. Endimiani A, Hujer KM, Hujer AM, Kurz S, Jacobs MR, Perlin DS, et al. Are we ready for novel detection methods to treat respiratory pathogens in hospital-acquired pneumonia? Clin Infect Dis. 2011; 52 Suppl 4:S373–83. pmid:21460299
  9. 9. Lung M, Codina G. Molecular diagnosis in HAP/VAP. Curr Opin Crit Care. 2012; 18:487–94. pmid:22889869
  10. 10. Huttner A, Emonet S, Harbarth S, Renzi G, Kaiser L, Schrenzel J. Polymerase-chain reaction/electrospray ionization-mass spectrometry for the detection of bacteria and fungi in bronchoalveolar lavage fluids: a prospective observational study. Clin Microbiol Infect. 2014; 20:O1059–66. pmid:25529027
  11. 11. Jordana-Lluch E, Gimenez M, Quesada MD, Ausina V, Martro E. Improving the diagnosis of bloodstream infections: PCR coupled with mass spectrometry. BioMed Res Int. 2014; 2014:501214. pmid:24818144
  12. 12. Vincent JL, Brealey D, Libert N, Abidi NE, O'Dwyer M, Zacharowski K, et al. Rapid Diagnosis of Infection in the Critically Ill, a Multicenter Study of Molecular Detection in Bloodstream Infections, Pneumonia, and Sterile Site Infections. Crit Care Med. 2015; 43:2283–91. pmid:26327198
  13. 13. Kommedal O, Kvello K, Skjastad R, Langeland N, Wiker HG. Direct 16S rRNA gene sequencing from clinical specimens, with special focus on polybacterial samples and interpretation of mixed DNA chromatograms. J Clin Microbiol. 2009; 47:3562–8. pmid:19741089
  14. 14. Hedberg ST, Olcen P, Fredlund H, Mölling P. Real-time PCR detection of five prevalent bacteria causing acute meningitis. APMIS. 2009; 117:856–60. pmid:19845537
  15. 15. Fang H, Hedin G. Rapid screening and identification of methicillin-resistant Staphylococcus aureus from clinical samples by selective-broth and real-time PCR assay. J Clin Microbiol. 2003; 41:2894–9. pmid:12843018
  16. 16. Chastre J, Fagon JY, Bornet-Lecso M, Calvat S, Dombret MC, al Khani R, et al. Evaluation of bronchoscopic techniques for the diagnosis of nosocomial pneumonia. Am J Respir Crit Care Med. 1995; 152:231–40. pmid:7599829
  17. 17. Prats E, Dorca J, Pujol M, Garcia L, Barreiro B, Verdaguer R, et al. Effects of antibiotics on protected specimen brush sampling in ventilator-associated pneumonia. Eur Respir J. 2002; 19:944–51. pmid:12030737
  18. 18. Strålin K, Korsgaard J, Olcén P. Evaluation of a multiplex PCR for bacterial pathogens applied to bronchoalveolar lavage. Eur Respir J. 2006; 28:568–75. pmid:16737990
  19. 19. Baudel JL, Tankovic J, Dahoumane R, Carrat F, Galbois A, Ait-Oufella H, et al. Multiplex PCR performed of bronchoalveolar lavage fluid increases pathogen identification rate in critically ill patients with pneumonia: a pilot study. Ann Intensive Care. 2014; 4:35. pmid:25593751
  20. 20. Berdal JE, Bjornholt J, Blomfeldt A, Smith-Erichsen N, Bukholm G. Patterns and dynamics of airway colonisation in mechanically-ventilated patients. Clin Microbiol Infect. 2007; 13:476–80. pmid:17430338
  21. 21. Souweine B, Veber B, Bedos JP, Gachot B, Dombret MC, Regnier B, et al. Diagnostic accuracy of protected specimen brush and bronchoalveolar lavage in nosocomial pneumonia: impact of previous antimicrobial treatments. Crit Care Med. 1998; 26:236–44. pmid:9468159
  22. 22. Berton DC, Kalil AC, Teixeira PJ. Quantitative versus qualitative cultures of respiratory secretions for clinical outcomes in patients with ventilator-associated pneumonia. Cochrane Database Syst Rev. 2014; 10:Cd006482. pmid:25354013
  23. 23. Bousbia S, Papazian L, Auffray JP, Fenollar F, Martin C, Li W, et al. Tropheryma whipplei in patients with pneumonia. Emer Infect Dis. 2010; 16:258–63.