Comparison of Glidescope® Go, King Vision, Dahlhausen VL, I‑View and Macintosh laryngoscope use during difficult airway management simulation by experienced and inexperienced emergency medical staff: A randomized crossover manikin study

  • Andreas Moritz ,

    Roles Conceptualization, Data curation, Formal analysis, Methodology, Writing – original draft

    andreas.moritz@uk-erlangen.de

    Affiliation Department of Anesthesiology, University Hospital Erlangen, Faculty of Medicine, Friedrich-Alexander-Universität (FAU) Erlangen-Nürnberg, Erlangen, Germany

  • Veronika Leonhardt,

    Roles Data curation, Formal analysis, Writing – review & editing

    Affiliation Department of Anesthesiology, University Hospital Erlangen, Faculty of Medicine, Friedrich-Alexander-Universität (FAU) Erlangen-Nürnberg, Erlangen, Germany

  • Johannes Prottengeier,

    Roles Supervision, Writing – review & editing

    Affiliation Department of Anesthesiology, University Hospital Erlangen, Faculty of Medicine, Friedrich-Alexander-Universität (FAU) Erlangen-Nürnberg, Erlangen, Germany

  • Torsten Birkholz,

    Roles Methodology, Supervision, Writing – review & editing

    Affiliation Department of Anesthesiology, University Hospital Erlangen, Faculty of Medicine, Friedrich-Alexander-Universität (FAU) Erlangen-Nürnberg, Erlangen, Germany

  • Joachim Schmidt,

    Roles Conceptualization, Methodology, Supervision, Writing – review & editing

    Affiliation Department of Anesthesiology, University Hospital Erlangen, Faculty of Medicine, Friedrich-Alexander-Universität (FAU) Erlangen-Nürnberg, Erlangen, Germany

  • Andrea Irouschek

    Roles Conceptualization, Data curation, Methodology, Supervision, Writing – review & editing

    Affiliation Department of Anesthesiology, University Hospital Erlangen, Faculty of Medicine, Friedrich-Alexander-Universität (FAU) Erlangen-Nürnberg, Erlangen, Germany

Abstract

Background

In pre-hospital emergency care, video laryngoscopes (VLs) with disposable blades are preferably used due to hygienic reasons. However, there is limited existing data on the use of VLs with disposable blades by emergency medical staff. Therefore, the aim of this study was to compare the efficacy of four different VLs with disposable blades and the conventional standard Macintosh laryngoscope, when used by anesthetists with extensive previous experience and paramedics with little previous experience in endotracheal intubation (ETI) in a simulated difficult airway.

Methods

Fifty-eight anesthetists and fifty-four paramedics participated in our randomized crossover manikin trial. Each performed ETI with the new Glidescope® Go, the Dahlhausen VL, the King Vision, the I-View and the Macintosh laryngoscope. “Time to intubate” was the primary endpoint. Secondary endpoints were “time to vocal cords”, “time to ventilate”, overall success rate, number of intubation attempts and optimization maneuvers, Cormack-Lehane score, severity of dental compression and subjective impressions.

Results

The Glidescope® Go, the Dahlhausen VL and the King Vision provided superior intubation conditions in both groups without affecting the number of intubation attempts or the time required for successful intubation. When used by anesthetists with extensive experience in ETI, the use of VLs did not affect the overall success rate. In the hands of paramedics with little previous experience in ETI, the failure rate with the Macintosh laryngoscope (14.8%) decreased to 3.7% using the Glidescope® Go and the Dahlhausen VL. Despite the advantages of hyperangulated video laryngoscopes, the I-View performed worst.

Conclusions

VLs with hyperangulated blades facilitated ETI in both groups and decreased the failure rate by an absolute 11.1% when used by paramedics with little previous experience in ETI. Our results therefore suggest that hyperangulated VLs could be beneficial and might be the method of choice in comparable settings, especially for emergency medical staff with less experience in ETI.

Introduction

Airway management is one of the major challenges of pre-hospital emergency medicine. However, various parameters relating to the patient (anatomical or physiological factors), environment (confined space, poor lighting conditions), out-of-hospital team work and the need for immediate decision-making may contribute to potential difficulties [1, 2]. Thus, in pre-hospital emergency settings, advanced airway management of critically ill or injured patients still remains a challenge, even in the hands of experienced emergency physicians. A difficult airway has been reported in up to 14.8% of patients managed by anesthesia trained physicians in a pre-hospital setting [3]. Several studies demonstrated an incidence of failed pre-hospital endotracheal intubation (ETI) in physician-staffed emergency medical systems varying from 1 to 2% [35]. In non-physician-manned emergency medical services, ETI failure rates of even 15–23% have been reported [4, 6]. Difficult or failed tracheal intubation is well-recognized as an important cause of morbidity and mortality associated with anesthesia and emergency medicine [7]. In addition, multiple intubation attempts have been shown to correlate with an increased incidence of adverse events [8], delay in transport, prolonged hospitalization, worse neurologic outcomes [9] and increased mortality [10].

In recent years, the implementation of video laryngoscopy has improved the in-hospital management of the difficult airway. In a Cochrane systematic review of video laryngoscope (VL) use in patients during general anesthesia, Lewis and colleagues reported reduced numbers of failed intubations, improved laryngeal view, facilitated handling and reduced airway trauma. However, not all VLs perform equally and, in addition, the intubation success also depends on the experience of the user with the respective device [11].

In pre-hospital emergency care, laryngoscopes with disposable blades are preferably used due to hygienic reasons [1215]. Although the equipment for emergency intubation of critically ill patients should meet the highest requirements [2], no study has yet evaluated the performance of video laryngoscopes with disposable blades when used by experienced and inexperienced emergency medical staff. Thus, the aim of our prospective randomized crossover study was to compare the efficacy of four different VLs with disposable blades (the new Glidescope® Go, the Dahlhausen VL, the King Vision aBlade, the I-View) and the conventional standard Macintosh laryngoscope, when used by anesthetists with extensive previous experience in conventional and video laryngoscopy and paramedics with little previous experience in conventional laryngoscopy and even less previous experience in the use of video laryngoscopy in a simulated difficult airway scenario.

Materials and methods

Study design and setting

The manikin study (randomized crossover design) was evaluated and approved by the institutional ethics committee (Ethics Committee of the Friedrich-Alexander-Universität Erlangen-Nürnberg; reference number: 408_18 B). Following written informed consent, fifty-eight anesthetists and fifty-four paramedics were recruited to the study. Data were anonymized and information on the performance of individual participants was not made available to anybody outside the research team.

Each participant performed ETI with the Glidescope® Go VL, a new portable high-resolution video laryngoscopy system, with a LoPro S3 single-use blade (Verathon Inc., Bothell, WA, USA), the Dahlhausen VL for single-use blades with a MAC DX3 single-use blade (P. J. Dahlhasuen & Co. GmbH, Köln, Germany), the King Vision aBlade VL with a non-channeled aBlade single-use blade size 3 (Ambu A/S, Ballerup, Denmark), the I-View one-size single-use VL (Intersurgical Ltd., Wokingham, Berkshire, UK) (Fig 1) and a conventional standard single-use Macintosh laryngoscope blade size 3 (PROACT HPC Green System Laryngoscope Blade-MAC 3, PROACT Medical Ltd., Corby, Northamptonshire, UK; Heine Standard F.O. 4 LED NT metallic laryngoscope handle, Heine Optotechnik GmbH & Co. KG, Herrsching, Germany) in a difficult airway manikin (Laerdal Airway Management Trainer, Laerdal Medical AS, Stavanger, Norway) (Fig 2). The neck of the manikin was fixed in a neutral position by a rigid cervical collar. Thus, the distance between the free edge of the upper and lower incisors (interdental distance) was limited, fulfilling conditions for a difficult intubation model [16, 17].

thumbnail
Fig 1. The four different video laryngoscopes (VLs) used in this study.

From left to right: I-View one-size single-use VL (Intersurgical Ltd., Wokingham, Berkshire, UK); King Vision aBlade VL with a non-channeled aBlade single-use blade size 3 (Ambu A/S, Ballerup, Denmark); Glidescope® Go VL with a LoPro S3 single-use blade (Verathon Inc., Bothell, WA, USA); Dahlhausen VL for single-use blades (Dahlhausen VL) with a MAC DX3 single-use blade (P. J. Dahlhasuen & Co. GmbH, Köln, Germany).

https://doi.org/10.1371/journal.pone.0236474.g001

thumbnail
Fig 2. Laerdal airway management trainer.

The difficult airway is simulated by cervical immobilization applying a cervical collar.

https://doi.org/10.1371/journal.pone.0236474.g002

Randomized order of application was provided by five sealed opaque envelopes each containing a different device name.

All intubations were performed with a 7.0 mm cuffed endotracheal tube (ETT; Rüsch Super Safety Clear endotracheal tube, I.D. 7.0 mm, Teleflex Medical GmbH, Fellbach, Germany). The cuff was lubricated with a silicone spray and was inflated and deflated with a 10 ml syringe. A reusable intubation stylet (Teleflex medical GmbH, Fellbach, Germany) was used to facilitate intubation with the Macintosh laryngoscope. For ETI with the Glidescope® Go, the Dahlhausen VL, the King Vision and the I-View a rigid intubation stylet (GlideRite Rigid Stylet, Verathon Inc., Bothell, WA, USA) was used.

Measurements

Objective findings.

The primary endpoint was the “time to intubate”. Esophageal intubations, attempts requiring more than 120 seconds or more than two attempts (withdrawal of the device from the mouth followed by repositioning) were recorded as failure to intubate. Stopwatch studies were made by a single person having direct observation to avoid interobserver error.

In order to compare the different intubation devices, the intubation process was divided into different time episodes:

  1. The time to visualization of the glottis (“time to vocal cords”) was defined as the time from insertion of the blade between the teeth until the glottis was visualized.
  2. The time to tracheal intubation (“time to intubate”) was defined as the time from insertion of the blade between the teeth until the ETT was deemed to be positioned correctly by each participant.
  3. The duration of a successful intubation attempt was defined as the time from insertion of the blade between the teeth until the ETT was connected to a self-inflating resuscitation bag and lung inflation was confirmed (“time to ventilate”).

We recorded the rate of successful intubation, the number of intubation attempts and the laryngeal view according to the Cormack-Lehane score [18]. The number of optimization maneuvers (readjustment of the head position, application of external laryngeal pressure and the need for assistance by a second person) and the number of audible dental click sounds indicating dental damage were recorded as 0, 1, and ≥ 2 times during ETI.

Subjective findings.

After completing the procedure, each anesthetist was asked to score

  • the view
  • the handling
  • the stability
  • the force applied during tracheal intubation
  • the difficulty of tracheal intubation
using a numeric rating scale (NRS) (0 to 100 mm, from excellent/very easy to poor/very difficult). Finally, the participants were asked to indicate their preferred intubation device.

Data analysis

The sample size was calculated using G*Power (Version 3.1.9.4, Faul F., 2019, Germany). Based on “time to intubate”-duration compiled in a pilot study, an effect size f of 0.39 was anticipated. Considering an α error of 0.05 and β error of 0.1, sample size calculation results that at least 50 experienced and 50 inexperienced participants would be required.

All the statistical analyses were performed using SPSS software (IBM® SPSS® Statistics, version 26.0, IBM Corp., Armonk, NY, USA). Data for the rate of successful intubation were analyzed using the Cochran`s Q test followed by Dunn-Bonferroni post-hoc tests. Data for the “time to vocal cords”, the “time to intubate”, the “time to ventilate”, the number of intubation attempts, the number of optimization maneuvers, the number of audible dental clicks, the Cormack-Lehane score, the view, the handling, the stability, the force applied during tracheal intubation and the difficulty of tracheal intubation were analyzed using paired non-parametric tests. The Friedman and Wilcoxon signed-rank tests with Bonferroni correction were used for multiple and post-hoc comparisons respectively. Statistical significance was accepted at p < 0.05.

Results

Participant characteristics

Fifty-eight anesthetists (43 [74%] senior house officers, 11 [19%] specialist registrars, 4 [7%] consultants) and fifty-four paramedics participated in this randomized crossover trial. The anesthetists included in this study also work as emergency physicians as part of their work in the Department of Anesthesiology at the University Hospital Erlangen. The participant characteristics and the self-reported experience in video and direct laryngoscopy are summarized in Table 1. Data in the result section are presented as mean difference with 95% confidence interval, MD [95% CI].

thumbnail
Table 1. Participant characteristics and self-reported estimates of previous experience in conventional and video laryngoscopy.

https://doi.org/10.1371/journal.pone.0236474.t001

Anesthetists with extensive previous experience in ETI

Regarding the primary endpoint (“time to intubate”), the “time to ventilate” and the overall success rate, no significant differences were observed between the conventional Macintosh laryngoscope, the Glidescope® Go, the Dahlhausen VL and the King Vision. However, the Glidescope® Go, the Dahlhausen VL and the King Vision enabled a significantly shorter “time to vocal cords” (4.1 s [2.0-6.2 s], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 14 s [10.3-17.7 s], p < 0.001, Glidescope® Go vs. I-View; 5.2 s [2.7-7.7 s], p < 0.001, Dahlhausen VL vs. Macintosh laryngoscope; 14.7 s [11.4-18.1 s], p < 0.001, Dahlhausen VL vs. I-View; 5.4 s [3-7.8 s], p < 0.001, King Vision vs. Macintosh laryngoscope; 14.5 s [11.2-17.7 s], p < 0.001, King Vision vs. I-View) and significantly improved the Cormack and Lehane grade (1 [0.8-1.2], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 0.9 [0.7-1.2], p < 0.001, Glidescope® Go vs. I-View; 0.8 [0.6-1], p < 0.001, Dahlhausen VL vs. Macintosh laryngoscope; 0.7 [0.5-1], p < 0.001, Dahlhausen VL vs. I-View; 0.9 [0.7-1.1], p < 0.001, King Vision vs. Macintosh laryngoscope; 0.8 [0.6-1.1], p < 0.001, King Vision vs. I-View) compared to the Macintosh laryngoscope and the I-View, when used by anesthetists with extensive previous experience in direct and in video laryngoscopy. In addition, the severity of dental compression was lower with the Glidescope® Go, the Dahlhausen VL and the King Vision (0.5 [0.3–0.7], p < 0.001, King Vision vs. Macintosh laryngoscope; 0.7 [0.4-0.9], p < 0.001, King Vision vs. I-View; 0.4 [0.2-0.6], p < 0.01, Glidescope® Go vs. Macintosh laryngoscope; 0.6 [0.4-0.8], p < 0.001, Glidescope® Go vs. I-View; 0.4 [0.1-0.6], p < 0.05, Dahlhausen VL vs. Macintosh laryngoscope; 0.6 [0.3-0.8], p < 0.01, Dahlhausen VL vs. I-View). Compared to the other devices, the I-View VL performed significantly worse. The recorded intubation times were significantly increased (“time to vocal cords”: 14 s [10.3-17.7 s], p < 0.001, Glidescope® Go vs. I-View; 14.5 s [11.2-17.7 s], p < 0.001, King Vision vs. I-View; 14.7 s [11.4-18.1 s], p < 0.001, Dahlhausen VL vs. I-View; 10.5 s [6.6-14.3 s], p < 0.001, Macintosh laryngoscope vs. I-View; “time to intubate”: 13.2 s [8.2-18.1 s], p < 0.001, Glidescope® Go vs. I-View; 13 s [8-17.9 s], p < 0.001, King Vision vs. I-View; 10.2 s [4.3-16.1 s], p < 0.01, Dahlhausen VL vs. I-View; 11 s [5.8-16.2 s], p < 0.01, Macintosh laryngoscope vs. I-View; “time to ventilate”: 12 s [6.9-17 s], p < 0.001, Glidescope® Go vs. I-View; 13.4 s [8.4-18.4 s], p < 0.001, King Vision vs. I-View; 11.8 s [6.7-16.9 s], p < 0.01, Macintosh laryngoscope vs. I-View; 9.6 s [3.1-16.2 s], p < 0.05, Dahlhausen VL vs. I-View) and tracheal intubation was less successful (n/N (%): 46/58 (79.3), p < 0.001) compared to the other intubation devices. Although the experienced participants needed more attempts for tracheal intubation using the I-View, this reached statistical significance only when compared to the Dahlhausen VL (0.3 [0.1–0.5], p < 0.05). Post-hoc comparison revealed no significant differences for the number of optimization maneuvers.

Regarding the subjective values, the Glidescope® Go, the Dahlhausen VL and the King Vision enabled a significant better view (3.4 cm [2.6-4.1 cm], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 5.3 cm [4.4-6.2 cm], p < 0.001, Glidescope® Go vs. I-View; 2.8 cm [2-3.6 cm], p < 0.001, Dahlhausen VL vs. Macintosh laryngoscope; 4.7 cm [3.8-5.6 cm], p < 0.001, Dahlhausen VL vs. I-View; 2.4 cm [1.4-3.4 cm], p < 0.001, King Vision vs. Macintosh laryngoscope; 4.3 cm [3.5-5.1 cm], p < 0.001, King Vision vs. I-View) were considered to be more stable (1.7 cm [1-2.4 cm], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 1.9 cm [1.1-2.6 cm], p < 0.001, Glidescope® Go vs. I-View; 1.4 cm [0.7-2.1 cm], p < 0.01, Dahlhausen VL vs. Macintosh laryngoscope; 1.6 cm [0.9-2.3 cm], p < 0.01, Dahlhausen VL vs. I-View; 1.1 cm [0.4-1.8 cm], p < 0.05, King Vision vs. Macintosh laryngoscope; 1.3 cm [0.6-2 cm], p < 0.05, King Vision vs. I-View) and required less force during tracheal intubation (3.3 cm [2.7-3.9 cm], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 4.2 cm [3.6-4.8 cm], p < 0.001, Glidescope® Go vs. I-View; 3.1 cm [2.5-3.8 cm], p < 0.001, Dahlhausen VL vs. Macintosh laryngoscope; 4 cm [3.4-4.6 cm], p < 0.001, Dahlhausen VL vs. I-View; 3.2 cm [2.6-3.9 cm], p < 0.001, King Vision vs. Macintosh laryngoscope; 4.1 cm [3.4-4.8 cm], p < 0.001, King Vision vs. I-View) compared to the Macintosh laryngoscope and the I-View. Furthermore, the ETI was judged to be easier using the Glidescope® Go, the Dahlhausen VL and the King Vision (2.3 cm [1.6-3.1 cm], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 4 cm [3.4-4.7 cm], p < 0.001, Glidescope® Go vs. I-View; 2.3 cm [1.6-3 cm], p < 0.001, Dahlhausen VL vs. Macintosh laryngoscope; 4 cm [3.3-4.8 cm], p < 0.001, Dahlhausen VL vs. I-View; 2.4 cm [1.7-3.1 cm], p < 0.001, King Vision vs. Macintosh laryngoscope; 4.1 cm [3.4-4.7 cm], p < 0.001, King Vision vs. I-View). The I-View VL was most difficult to handle compared to the other intubation devices (4.4 cm [3.6-5.1 cm], p < 0.001, Glidescope® Go vs. I-View; 3.6 cm [2.8-4.4 cm], p < 0.001, King Vision vs. I-View; 3.7 cm [2.9-4.5 cm], p < 0.001, Dahlhausen VL vs. I-View; 4.4 cm [3.6-5.3 cm], p < 0.001, Macintosh laryngoscope vs. I-View). Regarding view (2 cm [1-2.9 cm], p < 0.01) and difficulty of tracheal intubation (1.7 cm [0.9-2.5 cm], p < 0.001), the I-View was rated even worse than the Macintosh laryngoscope. The Glidescope® Go was rated best and offered an even better view than the King Vision (1 cm [0.3-1.6 cm], p < 0.05).

Forty percent of the anesthetists with extensive previous experience in ETI preferred the Glidescope® Go, followed by the Dahlhausen VL (31%) and the King Vision (17%). Only 9% preferred the standard Macintosh laryngoscope and none of the experienced participants would use the I-View.

The recorded intubation times are shown graphically in Fig 3. Objective and subjective findings are summarized in Table 2.

thumbnail
Fig 3. Recorded intubation times of the anesthetists with extensive previous experience in conventional and video laryngoscopy.

Data are presented as boxplots that indicate the median (middle line), the 25th and the 75th percentile (box boundaries), the upper and lower limits within 1.5 times inter-quartile range (whiskers) and outliers (circles). The blue boxplots represent the “time to vocal cords”, the green boxplots represent the “time to intubate” and the red boxplots represent the “time to ventilate”.

https://doi.org/10.1371/journal.pone.0236474.g003

thumbnail
Table 2. Intubation data of anesthetists with extensive previous experience in conventional and video laryngoscopy.

https://doi.org/10.1371/journal.pone.0236474.t002

Paramedics with little previous experience in ETI

We could not find a significant difference in “the time to intubate” between the conventional standard Macintosh laryngoscope, the Dahlhausen VL, the King Vision and the I-View. However, the Glidescope® Go enabled a significantly shorter “time to intubate” compared to the King Vision (7.6 s [2.5-12.7 s], p < 0.05). The “time to vocal cords” was significantly increased using the I-View compared to the other VLs (9.7 s [6.3-13.2 s], p < 0.001, King Vision vs. I-View; 7.9 s [3.8-11.9 s], p < 0.05, Glidescope® Go vs. I-View; 7.1 s [3.2-10.9 s], p < 0.05, Dahlhausen VL vs. I-View). Although, post-hoc analysis revealed no differences in “the time to ventilate” and the number of intubation attempts, the failure rate was 35.2% with the I-View, 14.8% with the Macintosh laryngoscope, 9.3% with the King Vision and only 3.7% with the Glidescope® Go and the Dahlhausen VL. However, post-hoc analysis revealed statistical significance only for the I-View (n/N (%):35/54 (64.8), p < 0.001, Glidescope® Go, Dahlhausen VL and King Vision vs. I-View, p < 0.01 Macintosh laryngoscope vs. I-View). The severity of dental compression was lower with the Glidescope® Go and the King Vision compared to the I-View (0.8 [0.4-1.1], p < 0.001, King Vision vs. I-View; 0.6 [0.4-0.9], p < 0.01, Glidescope® Go vs. I-View). The King Vision also caused less dental trauma than the use of the Macintosh laryngoscope and the Dahlhausen VL (0.4 [0.2-0.7], p < 0.01, King Vision vs. Macintosh laryngoscope; 0.6 [0.3-0.8], p < 0.01, King Vision vs. Dahlhausen VL). In addition, the number of optimization maneuvers was lower when using the Glidescope® Go compared to the I-View (0.2 [0.1-0.4], p < 0.05). The Glidescope® Go, the Dahlhausen VL and the King Vision significantly improved the Cormack and Lehane grade compared to the Macintosh laryngoscope and the I-View, when used by paramedics with low experience in video laryngoscopy (1 [0.6-1.4], p < 0.001, Glidescope® Go vs. I-View; 0.6 [0.3-0.8], p < 0.01, Glidescope® Go vs. Macintosh laryngoscope; 1 [0.7-1.4], p < 0.001, Dahlhausen VL vs. I-View; 0.6 [0.4-0.9], p < 0.001, Dahlhausen VL vs. Macintosh laryngoscope; 1 [0.6-1.3], p < 0.001, King Vision vs. I-View; 0.5 [0.3-0.8], p < 0.01, King Vision vs. Macintosh laryngoscope).

Regarding the subjective values, the Glidescope® Go, the Dahlhausen VL and the King Vision enabled a significant better view (2.7 cm [1.8-3.6 cm], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 5.4 cm [4.5-6.4 cm], p < 0.001, Glidescope® Go vs. I-View; 2.4 cm [1.6-3.2 cm], p < 0.001, Dahlhausen VL vs. Macintosh laryngoscope; 5.2 cm [4.4-6 cm], p < 0.001, Dahlhausen VL vs. I-View; 2.6 cm [1.8-3.4 cm], p < 0.001, King Vision vs. Macintosh laryngoscope; 5.4 cm [4.5-6.3 cm], p < 0.001, King Vision vs. I-View) and required less force during tracheal intubation (2.2 cm [1.4-3.1 cm], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 2.8 cm [1.9-3.7 cm], p < 0.001, Glidescope® Go vs. I-View; 2.2 cm [1.4-3.1 cm], p < 0.001, King Vision vs. Macintosh laryngoscope; 2.8 cm [2-3.6 cm], p < 0.001, King Vision vs. I-View; 1.9 cm [1-2.9 cm], p < 0.01, Dahlhausen VL vs. I-View; 1.4 cm [0.5-2.2 cm], p < 0.05, Dahlhausen VL vs. Macintosh laryngoscope) compared to the Macintosh laryngoscope and the I-View. In addition, the ETI was judged to be easier using the Glidescope® Go, the Dahlhausen VL and the King Vision (2.1 cm [1.3-2.9 cm], p < 0.001, Glidescope® Go vs. Macintosh laryngoscope; 4.3 cm [3.3-5.2 cm], p < 0.001, Glidescope® Go vs. I-View; 3.7 cm [2.8-4.6 cm], p < 0.001, Dahlhausen VL vs. I-View; 1.5 cm [0.7-2.4 cm], p < 0.05, Dahlhausen VL vs. Macintosh laryngoscope; 3.6 cm [2.6-4.5 cm], p < 0.001, King Vision vs. I-View; 1.4 cm [0.5-2.3 cm], p < 0.05, King Vision vs. Macintosh laryngoscope). The I-View VL was considered to be less stable (2 cm [1.1-2.8 cm], p < 0.001, Glidescope® Go vs. I-View; 1.9 cm [1-2.7 cm], p < 0.001, Dahlhausen VL vs. I-View; 1.8 cm [1.1-2.5 cm], p < 0.001, King Vision vs. I-View; 1.6 cm [0.7-2.5 cm], p < 0.01, Macintosh laryngoscope vs. I-View) and more difficult to handle (p < 0.001) (3.7 cm [2.7-4.6 cm], p < 0.001, Glidescope® Go vs. I-View; 3.1 cm [2.2-4.1 cm], p < 0.001, Dahlhausen VL vs. I-View; 2.9 cm [1.9-3.8 cm], p < 0.001, King Vision vs. I-View; 2.3 cm [1.3-3.4 cm], p < 0.001, Macintosh laryngoscope vs. I-View) when compared to the other intubation devices. Regarding the view (2.8 cm [1.7-3.8 cm], p < 0.001) and the difficulty of tracheal intubation (2.2 cm [1-3.3 cm], p < 0.01), the I-View was rated even worse than the Macintosh laryngoscope. The Glidescope® Go was rated best and even easier to handle than the conventional Macintosh laryngoscope (1.3 cm [0.4-2.3 cm], p < 0.05).

Thirty-nine percent of the paramedics preferred the Glidescope® Go, followed by the King Vision (26%) and the Dahlhausen VL (22%). The standard Macintosh laryngoscope was preferred by 11% and only 2% of the paramedics with little previous experience in ETI would use the I-View VL.

The recorded intubation times are shown graphically in Fig 4. Objective and subjective findings are summarized in Table 3.

thumbnail
Fig 4. Recorded intubation times of the paramedics with little previous experience in conventional and even less previous experience in video laryngoscopy.

Data are presented as boxplots that indicate the median (middle line), the 25th and the 75th percentile (box boundaries), the upper and lower limits within 1.5 times inter-quartile range (whiskers) and outliers (circles). The blue boxplots represent the “time to vocal cords”, the green boxplots represent the “time to intubate” and the red boxplots represent the “time to ventilate”.

https://doi.org/10.1371/journal.pone.0236474.g004

thumbnail
Table 3. Intubation data of paramedics with little previous experience in conventional and even less previous experience in video laryngoscopy.

https://doi.org/10.1371/journal.pone.0236474.t003

Discussion

In the in-hospital setting, video laryngoscopy has been shown to reduce the number of failed intubations, to improve the glottic view and to reduce airway trauma [11]. However, there are only a few and heterogeneous studies comparing video laryngoscopy and direct laryngoscopy in the pre-hospital setting [19]. In addition, in the pre-hospital emergency setting, the intubation success depends not only on the laryngoscope type used, but also on provider training and its experience with the device. Bernhard and colleagues reviewed that paramedics tend to have lower success rates than physicians [20]. Thus, the aim of our prospective randomized crossover study was to compare the efficacy of different VLs with disposable blades and the conventional standard Macintosh laryngoscope, when used by anesthetists with extensive previous experienced in conventional and video laryngoscopy and paramedics with little previous experience in conventional laryngoscopy and even less previous experience in the use of video laryngoscopy in a simulated difficult airway scenario.

The results of our study suggest that in the hands of both anesthetists with extensive previous experience in ETI and paramedics with little experience in ETI, the Glidescope® Go, the Dahlhausen VL and the King Vision provide superior intubation conditions and can facilitate ETI in our simulated difficult airway scenario. In addition, the Glidescope® Go, the Dahlhausen VL and the King Vision improved the Cormack and Lehane grade when compared with the Macintosh laryngoscope. The design of these VLs offers a ‘view around the corner’ and provide an optimal glottis visualization via the camera, without the need to align the oral, pharyngeal and tracheal axes [21]. Despite the subjective advantages of the Glidescope® Go, the Dahlhausen VL and the King Vision we could not find a significant difference in the “time to intubate”, the “time to ventilate”, the number of intubation attempts and the number of optimization manoeuvers compared to the standard Macintosh laryngoscope with single-use blade when used by both anesthetists and paramedics. This is in line with the findings of Lewis and colleagues. In a Cochrane review including 64 studies with 7044 participants they found no evidence to indicate that the use of a VL would result in fewer intubation attempts. In addition, due to an extremely high level of statistical heterogeneity in terms of intubation times, Lewis and colleagues also found no evidence that using a VL would affect the time required for intubation [11].

In the hands of anesthetists with extensive previous experience in ETI, the Glidescope® Go, the Dahlhausen VL and the King Vision enabled a significantly shorter “time to vocal cords” and provided improved glottis exposure. However, we did not observe a statistically significant difference between the use of video laryngoscopy and direct laryngoscopy with regard to the overall success rate. This may be due to the anesthetists’ extensive previous experience in direct laryngoscopy with the conventional standard Macintosh laryngoscope in daily practice. Anesthetists experienced in direct laryngoscopy can identify landmarks of the anatomy and may be able to perform successful tracheal intubation even in case of worse Cormack and Lehane grade. Nonetheless, using the Macintosh laryngoscope, a greater peak force is required to align the axes and to visualize the glottis in a difficult airway [22, 23]. This is consistent with our subjective findings and may explain the lower rate of dental trauma when using the Glidescope® Go, the Dahlhausen VL and the King Vision.

In the hands of paramedics with little previous experience in ETI, video laryngoscopy decreased the failure rate for tracheal intubation from 14.8% with the Macintosh laryngoscope to 3.7% with the Glidescope® Go and the Dahlhausen VL. This corresponds to an absolute risk reduction of 11.1%. Although, this did not reach statistical significance, this result is in line with the findings of Savino and colleagues. In a systematic review and meta-analysis, including eight pre-hospital studies, they conclude that video laryngoscopy may lead to increased overall success rate in those emergency medical systems in which providers have less experience with ETI [19].

In our study, however, not all VLs performed equally. In the hands of both anesthetists with extensive previous experience and paramedics with little previous experience in ETI, the I-View VL was considered less stable and the most difficult to handle compared to the other intubation devices. The view was described as significantly worse and the intubation was rated more difficult. The worse performance may be due to the design of the I-View VL. In contrast to the Glidescope® Go, the Dahlhausen VL and the King Vision with hyperangulated disposable blades, the blade of the I-View one-size single-use VL has a standard Macintosh shape and is significantly longer. In a randomized crossover manikin study, Eismann and colleagues demonstrated, that the hyperangulated blade geometries of VLs provided a better view in difficult airway than the standard geometry of the Macintosh-type blade [24]. In addition, Schieren and colleagues showed that VLs with hyperangulated blades needed less force for intubation in difficult airways compared to Macintosh shaped laryngoscope blades [25]. The oversized length and the Macintosh shape of the blade could explain the significantly decreased overall success rate and higher rate of dental trauma when using the I-View VL.

In our prospective randomized crossover study, the Glidescope® Go was rated best. It was considered to be even easier to handle than the Macintosh laryngoscope and enabled a significantly shorter “time to intubate” compared to the King Vision when used by paramedics with little previous experience in ETI. In the hands of experienced anesthetists, the Glidescope® Go offered an even better view than the King Vision. The overall preference of the participants confirmed these results. Forty percent of the anesthetists and thirty-nine percent of the paramedics preferred the Glidescope® Go.

Limitations

The results of our study must be interpreted with consideration of several limitations. First, this study was performed using a high fidelity simulator and not on patients. Despite of their quality, manikins do not fully resemble human structures and are limited by their gross mechanics and tactile and textural differences. However, the use of anatomically correct manikins has proven to be a reliable surrogate for the clinical context [26]. In our difficult airway scenario, we observed a failure rate of 14.8% for paramedics using the Macintosh laryngoscope. This is in line with the findings of Lossius and colleagues. In their comprehensive meta-analysis they reported a failure rate of 15% in emergency services manned by non-physicians [4]. This indicates a good comparability and validity of our manikin. Second, some potential for bias exists, as it was not possible to blind the participants or the assessors to the device used in the difficult airway scenario. Third, the participants were aware that their performance was being assessed, which could lead to an altered performance due to the Hawthorne effect [27]. Fourth, certain measurements used in this study, such as grading the difficulty of tracheal intubation, have a subjective nature. Fifth, we examined only one difficult intubation scenario. The performance of the devices might be different in other scenarios (e.g. bloody/secretion-filled airway scenario, orofacial trauma scenario, obstructive oropharyngeal mass/tissue scenario). Sixth, a malleable intubation stylet was used to facilitate intubation with the Macintosh laryngoscope. Although, a malleable stylet allows to shape the endotracheal tube into an ideal position for direct laryngoscopy, the flexibility of the stylet might affect the performance of the Macintosh laryngoscope to some extent in the simulated difficult airway scenario. Seventh, although the hyperangulated VLs were associated with a statistically significant reduction in the “time to vocal cords” when used by experienced anesthetists, the clinical impact of this time difference remains uncertain. Eighth, the gender disparity in our study might have influenced the results. However, Waddington et al. could demonstrate that female and male intubators did not differ in their ability to intubate a Laerdal airway management trainer [28]. Finally, we compared only four different VLs with disposable blades and one standard Macintosh laryngoscope. There are other types of VLs with disposable blades and their utility in the management of the difficult airway in comparable settings might be different and should be investigated. In addition, Macintosh blades for direct laryngoscopy can vary substantially and therefore external validity may be limited to some extent.

Conclusions

The use of video laryngoscopy did not affect the number of intubation attempts or the time required for successful ETI compared to direct laryngoscopy with a conventional standard Macintosh blade in a simulated difficult airway scenario. However, the Glidescope® Go, the Dahlhausen VL and the King Vision provided superior intubation conditions, enabled a better visualization of the glottis and thus facilitated ETI when used by both anesthetists with extensive previous experience in conventional and video laryngoscopy and paramedics with little previous experience in conventional and even less previous experience in video laryngoscopy. Although, the use of VLs did not affect the overall success rate of the anesthetists, video laryngoscopy in the hands of paramedics with little previous experience in ETI decreased the failure rate from 14.8% with the conventional standard Macintosh laryngoscope to 3.7% with the Glidescope® Go and the Dahlhausen VL. In our manikin model of difficult airway, however, not all VLs performed equally well. The hyperangulated VLs performed significantly better than the Macintosh shaped I-View, with the Glidescope® Go being the most suitable VL.

Our results therefore suggest that hyperangulated VLs could be beneficial and might be the method of choice in comparable settings, especially for users with little previous experience in ETI. Further prospective randomized studies comparing direct laryngoscopy and video laryngoscopy in pre-hospital setting are necessary to confirm our findings.

Acknowledgments

The authors wish to thank the Bayerische Rote Kreuz Kreisverband Erlangen-Höchstadt and the entire anesthesiology staff of the University Hospital Erlangen for the support and the participation in this manikin study.

References

  1. 1. Haerkens MH, Jenkins DH, van der Hoeven JG. Crew resource management in the ICU: the need for culture change. Ann Intensive Care. 2012;2(1):39. Epub 2012/08/22. pmid:22913855; PubMed Central PMCID: PMC3488012.
  2. 2. Higgs A, McGrath BA, Goddard C, Rangasami J, Suntharalingam G, Gale R, et al. Guidelines for the management of tracheal intubation in critically ill adults. Br J Anaesth. 2018;120(2):323–52. Epub 2017/11/26. pmid:29406182.
  3. 3. Timmermann A, Eich C, Russo SG, Natge U, Bräuer A, Rosenblatt WH, et al. Prehospital airway management: a prospective evaluation of anaesthesia trained emergency physicians. Resuscitation. 2006;70(2):179–85. pmid:16828956.
  4. 4. Lossius HM, Røislien J, Lockey DJ. Patient safety in pre-hospital emergency tracheal intubation: a comprehensive meta-analysis of the intubation success rates of EMS providers. Crit Care. 2012;16(1):R24. Epub 2012/02/11. pmid:22325973; PubMed Central PMCID: PMC3396268.
  5. 5. Sunde GA, Heltne JK, Lockey D, Burns B, Sandberg M, Fredriksen K, et al. Airway management by physician-staffed Helicopter Emergency Medical Services—a prospective, multicentre, observational study of 2,327 patients. Scand J Trauma Resusc Emerg Med. 2015;23:57. Epub 2015/08/07. pmid:26250700; PubMed Central PMCID: PMC4528299.
  6. 6. Wang HE, Mann NC, Mears G, Jacobson K, Yealy DM. Out-of-hospital airway management in the United States. Resuscitation. 2011;82(4):378–85. Epub 2011/02/01. pmid:21288624.
  7. 7. Cook TM, MacDougall-Davis SR. Complications and failure of airway management. Br J Anaesth. 2012;109 Suppl 1:i68–i85. pmid:23242753.
  8. 8. Sakles JC, Chiu S, Mosier J, Walker C, Stolz U. The importance of first pass success when performing orotracheal intubation in the emergency department. Acad Emerg Med. 2013;20(1):71–8. pmid:23574475; PubMed Central PMCID: PMC4530518.
  9. 9. Ehrlich PF, Seidman PS, Atallah O, Haque A, Helmkamp J. Endotracheal intubations in rural pediatric trauma patients. J Pediatr Surg. 2004;39(9):1376–80. pmid:15359393.
  10. 10. Mort TC. Emergency tracheal intubation: complications associated with repeated laryngoscopic attempts. Anesth Analg. 2004;99(2):607–13, table of contents. pmid:15271750.
  11. 11. Lewis SR, Butler AR, Parker J, Cook TM, Schofield-Robinson OJ, Smith AF. Videolaryngoscopy versus direct laryngoscopy for adult patients requiring tracheal intubation: a Cochrane Systematic Review. Br J Anaesth. 2017;119(3):369–83. pmid:28969318.
  12. 12. Hirsch N, Beckett A, Collinge J, Scaravilli F, Tabrizi S, Berry S. Lymphocyte contamination of laryngoscope blades—a possible vector for transmission of variant Creutzfeldt-Jakob disease. Anaesthesia. 2005;60(7):664–7. pmid:15960716.
  13. 13. Miller DM, Youkhana I, Karunaratne WU, Pearce A. Presence of protein deposits on 'cleaned' re-usable anaesthetic equipment. Anaesthesia. 2001;56(11):1069–72. pmid:11703239.
  14. 14. Phillips RA, Monaghan WP. Incidence of visible and occult blood on laryngoscope blades and handles. AANA J. 1997;65(3):241–6. pmid:9233093.
  15. 15. Van Wicklin SA. Contamination and Disinfection of Rigid Laryngoscopes: A Literature Review. AORN J. 2019;110(1):49–59. pmid:31246299.
  16. 16. Beleña JM, Gasco C, Polo CE, Vidal A, Núñez M, Lopez-Timoneda F. Laryngeal mask, laryngeal tube, and Frova introducer in simulated difficult airway. J Emerg Med. 2015;48(2):254–9. pmid:25453860.
  17. 17. Moritz A, Prottengeier J, Schmidt J. Comparison of Xenon with LED illuminant in difficult and inhalation injury airway scenario: A randomized crossover manikin study. Am J Emerg Med. 2017. Epub 2017/05/08. pmid:28527873.
  18. 18. Cormack RS, Lehane J. Difficult tracheal intubation in obstetrics. Anaesthesia. 1984;39(11):1105–11. pmid:6507827.
  19. 19. Savino PB, Reichelderfer S, Mercer MP, Wang RC, Sporer KA. Direct Versus Video Laryngoscopy for Prehospital Intubation: A Systematic Review and Meta-analysis. Acad Emerg Med. 2017;24(8):1018–26. Epub 2017/07/25. pmid:28370736.
  20. 20. Bernhard M, Becker TK, Gries A, Knapp J, Wenzel V. The First Shot Is Often the Best Shot: First-Pass Intubation Success in Emergency Airway Management. Anesth Analg. 2015;121(5):1389–93. pmid:26484464.
  21. 21. Sun DA, Warriner CB, Parsons DG, Klein R, Umedaly HS, Moult M. The GlideScope Video Laryngoscope: randomized clinical trial in 200 patients. Br J Anaesth. 2005;94(3):381–4. Epub 2004/11/26. pmid:15567809.
  22. 22. Pieters B, Maassen R, Van Eig E, Maathuis B, Van Den Dobbelsteen J, Van Zundert A. Indirect videolaryngoscopy using Macintosh blades in patients with non-anticipated difficult airways results in significantly lower forces exerted on teeth relative to classic direct laryngoscopy: a randomized crossover trial. Minerva Anestesiol. 2015;81(8):846–54. Epub 2014/10/14. pmid:25311949.
  23. 23. Lee RA, van Zundert AA, Maassen RL, Wieringa PA. Forces applied to the maxillary incisors by video laryngoscopes and the Macintosh laryngoscope. Acta Anaesthesiol Scand. 2012;56(2):224–9. Epub 2011/10/14. pmid:22091734.
  24. 24. Eismann H, Sieg L, Etti N, Friedrich L, Schröter C, Mommsen P, et al. Improved success rates using videolaryngoscopy in unexperienced users: a randomized crossover study in airway manikins. Eur J Med Res. 2017;22(1):27. Epub 2017/08/10. pmid:28797305; PubMed Central PMCID: PMC5553664.
  25. 25. Schieren M, Kleinschmidt J, Schmutz A, Loop T, Staat M, Gatzweiler KH, et al. Comparison of forces acting on maxillary incisors during tracheal intubation with different laryngoscopy techniques: a blinded manikin study. Anaesthesia. 2019;74(12):1563–71. Epub 2019/08/25. pmid:31448404.
  26. 26. Dyson A, Harris J, Bhatia K. Rapidity and accuracy of tracheal intubation in a mannequin: comparison of the fibreoptic with the Bullard laryngoscope. Br J Anaesth. 1990;65(2):268–70. pmid:2223350.
  27. 27. Holden JD. Hawthorne effects and research into professional practice. J Eval Clin Pract. 2001;7(1):65–70. pmid:11240840.
  28. 28. Waddington MS, Paech MJ, Kurowski IH, Reed CJ, Nicholls GJ, Guy DT, et al. The influence of gender and experience on intubation ability and technique: a manikin study. Anaesth Intensive Care. 2009;37(5):791–801. pmid:19775044.