Abstract
The objective of this report is to highlight the potential for false pressure measurements from systems that combine intracranial pressure (ICP) measurement and ventricular drainage. If the ports of the drain become blocked to the extent that they present a high resistance to cerebrospinal fluid flow, then a significant pressure gradient between the inside and outside of the catheter may be established. Thus, any intracatheter transducer will faithfully record a pressure much lower than true ICP. This holds true for catheter tip transducers when the transducer lies inside the catheter. In the absence of flow, however, pressures will equalize; therefore, accurate measurements may be taken if the drain is temporarily closed. We model this situation and provide simulations of expected measurements in such situations; these compare well to observed clinical readings.
Similar content being viewed by others
References
Wilkinson HA, Yarzebski J, Wilkinson EC, Anderson FA. Erroneous measurement of intracranial pressure caused by simultaneous ventricular drainage: a hydrodynamic model study. Neurosurgery 1989;24:348–354.
Piek J, Raes P. Pressure-controlled drainage of cerebrospinal fluid: clinical experience with a new type of ventricular catheter (Ventcontrol MTC) and an integrated piezo-resistive sensor at its tip: technical note. Neurosurgery 1996;38:216–218.
Schley D, Billingham J, Marchbanks RJ. A model of in-vivo hydrocephalus shunt dynamics for blockage and performance diagnostics. Math Med Biol 2004; 21:347–368.
Knowlton FP, Starling EH. The influence of variations in temperature and blood pressure on the performance of the isolated mammalian heart. J Physiol 1912;44:206–219.
Luce JM, Huseby JS, Kirk W, Butler JA. Starling resistor regulates cerebral venous outflow in dogs. J Appl Physiol 1982;53(6):1496–1503.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Birch, A.A., Eynon, C.A. & Schley, D. Erroneous intracranial pressure measurements from simultaneous pressure monitoring and ventricular drainage catheters. Neurocrit Care 5, 51–54 (2006). https://doi.org/10.1385/NCC:5:1:51
Issue Date:
DOI: https://doi.org/10.1385/NCC:5:1:51