Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
New Bipolar Diathermy Forceps with Automatic Dripping and Flushing —Technical Note—
Tomio OHTAToshihiko KUROIWAYoshinaga KAJIMOTO
Author information
JOURNAL FREE ACCESS

1999 Volume 39 Issue 7 Pages 548-550

Details
Abstract

A new bipolar diathermy forceps system was developed to solve the problems of constant, pressurelimited flow rate, and one-sided irrigation. A roller pump, activated synchronously by pressing a foot switch, feeds dripping and flushing solution to the target tissue via the tip at both ends of the forceps. This system is volume-limited. Continuous compression of the foot switch first activates the flushing function, which continues for less than 1 second, during which time bleeding spots can be detected. The flow then changes automatically to the dripping function to suppress tip burning and prevent damage to the surrounding tissues from heat and current leakage. Repeated pressing of the foot switch initiates the jet irrigation function (continuous high flow rates), allowing irrigation of hematomas and removal of excess debris.

Content from these authors
© The Japan Neurosurgical Society
Previous article
feedback
Top