Journal of Oral Science
Online ISSN : 1880-4926
Print ISSN : 1343-4934
ISSN-L : 1343-4934
Original
Comparison of image performance between cone-beam computed tomography for dental use and four-row multidetector helical CT
Koji HashimotoShoji KawashimaMasao ArakiKazuo IwaiKunihiko SawadaYutaka Akiyama
Author information
JOURNAL FREE ACCESS

2006 Volume 48 Issue 1 Pages 27-34

Details
Abstract

The authors evaluated the imaging performance of cone-beam computed tomography (CBCT) for dental use using 3DX multi-image micro-CT (Morita Co., Kyoto, Japan) and four-row multi-detector helical computed tomography (MDCT) using an Asteion (Toshiba, Tokyo, Japan). A dried right maxillary bone was cut into eight slices 2 mm thick toward the zygomatico-palate and used as a phantom. Images of the phantom were then taken using 3DX and MDCT. The images of two bone slices were evaluated by five dentists for image quality and reproducibility of cancellous bone, as well as enamel, dentin, pulp cavity, periodontal ligament space, lamina dura and the overall image. Using the MDCT images as the standard, the 3DX images were evaluated with a subjective 5-level scale: 3 for an image equal to the MDCT image, 4 or greater for better, and 2 or lower for worse. The scores for all parameters exceeded 4 points. Maximum mean score was 4.8 for the lamina dura. Statistically significant differences were found for all items (P < 0.01). Our subjective evaluation of imaging performance clarified that 3DX was superior to MDCT. The results of this study suggest that 3DX is useful for imaging in the dental field. (J. Oral Sci. 48, 27-34, 2006)

Content from these authors
© 2006 by Nihon University School of Dentistry
Previous article Next article
feedback
Top