Multiple-animal MR Imaging using a 3T Clinical Scanner and Multi-channel Coil for Volumetric Analysis in a Mouse Tumor Model
Minoru MITSUDAMasayuki YAMAGUCHIToshihiro FURUTAAkira NABETANIAkira HIRAYAMAAtsushi NOZAKIMamoru NIITSUHirofumi FUJII
Author information
JOURNAL OPEN ACCESS

2011 Volume 10 Issue 4 Pages 229-237

Details
Abstract

Purpose: Multiple small-animal magnetic resonance (MR) imaging to measure tumor volume may increase the throughput of preclinical cancer research assessing tumor response to novel therapies. We used a clinical scanner and multi-channel coil to evaluate the usefulness of this imaging to assess experimental tumor volume in mice.
Methods: We performed a phantom study to assess 2-dimensional (2D) geometric distortion using 9-cm spherical and 32-cell (8×4 one-cm2 grids) phantoms using a 3-tesla clinical MR scanner and dedicated multi-channel coil composed of 16 5-cm circular coils. Employing the multi-channel coil, we simultaneously scanned 6 or 8 mice bearing sarcoma 180 tumors. We estimated tumor volume from the sum of the product of tumor area and slice thickness on 2D spin-echo images (repetition time/echo time, 3500/16 ms; in-plane resolution, 0.195×0.195×1 mm3). After MR acquisition, we excised and weighed tumors, calculated reference tumor volumes from actual tumor weight assuming a density of 1.05 g/cm3, and assessed the correlation between the estimated and reference volumes using Pearson's test.
Results: Two-dimensional geometric distortion was acceptable below 5% in the 9-cm spherical phantom and in every cell in the 32-cell phantom. We scanned up to 8 mice simultaneously using the multi-channel coil and found 11 tumors larger than 0.1 g in 12 mice. Tumor volumes were 1.04±0.73 estimated by MR imaging and 1.04±0.80 cm3 by reference volume (average±standard deviation) and highly correlated (correlation coefficient, 0.995; P<0.01, Pearson's test).
Conclusion: Use of multiple small-animal MR imaging employing a clinical scanner and multi-channel coil enabled accurate assessment of experimental tumor volume in a large number of mice and may facilitate high throughput monitoring of tumor response to therapy in preclinical research.

Content from these authors
© 2011 by Japanese Society for Magnetic Resonance in Medicine
Previous article Next article
feedback
Top