J Adv Prosthodont. 2016 Oct;8(5):354-362. English.
Published online Oct 21, 2016.
© 2016 The Korean Academy of Prosthodontics
Original Article

Comparative analysis on reproducibility among 5 intraoral scanners: sectional analysis according to restoration type and preparation outline form

Ji-Man Park
    • Department of Prosthodontics, Seoul National University Gwanak Dental Hospital, Seoul, Republic of Korea.
Received February 05, 2016; Revised June 24, 2016; Accepted August 08, 2016.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

PURPOSE

The trueness and precision of acquired images of intraoral digital scanners could be influenced by restoration type, preparation outline form, scanning technology and the application of power. The aim of this study is to perform the comparative evaluation of the 3-dimensional reproducibility of intraoral scanners (IOSs).

MATERIALS AND METHODS

The phantom containing five prepared teeth was scanned by the reference scanner (Dental Wings) and 5 test IOSs (E4D dentist, Fastscan, iTero, Trios and Zfx Intrascan). The acquired images of the scanner groups were compared with the image from the reference scanner (trueness) and within each scanner groups (precision). Statistical analysis was performed using independent two-samples t-test and analysis of variance (α=.05).

RESULTS

The average deviations of trueness and precision of Fastscan, iTero and Trios were significantly lower than the other scanners. According to the restoration type, significantly higher trueness was observed in crown and inlay than in bridge. However, no significant difference was observed among four sites of preparation outline form. If compared by the characteristics of IOS, high trueness was observed in the group adopting the active triangulation and using powder. However, there was no significant difference between the still image acquisition and video acquisition groups.

CONCLUSION

Except for two intraoral scanners, Fastscan, iTero and Trios displayed comparable levels of trueness and precision values in tested phantom model. Difference in trueness was observed depending on the restoration type, the preparation outline form and characteristics of IOS, which should be taken into consideration when the intraoral scanning data are utilized.

Keywords
Intraoral scanner; Digital impression; Reproducibility of intraoral scanner; Trueness; Precision

INTRODUCTION

With the recent acceleration of the application of digital technology in the dental medicine, various intraoral scanners (IOSs) that can acquire intraoral images without using the impression material have been introduced in addition to the existing CAD/CAM system.1

According to Nedelcu et al. and Schaefer et al.,2, 3 IOS can be classified into the still image acquisition and video acquisition methods according to the image recombination. Of the IOS distributed currently in use, CEREC AC Bluecam, E4D dentist and iTero (first generation) adopted the former method while the latter is adopted by E4D NEVO, Fastscan, Lava COS, Trios and Zfx Intrascan. Due to its nature that the scanner itself has to move in the narrow oral cavity as compared with the desktop scanner, there are various scanning technologies. For instance, E4D is operated by optical coherence tomography. Also, active triangulation is applied for CEREC AC Bluecam, Fastscan and Lava COS while confocal microscopic technology is adopted in iTero, Trios and Zfx Intrascan. The depth of field depends on these scanning technologies and has an influence on the learning curve of the device. The powder is used to enhance scanner's recognition rate and shorten scan time by reducing reflection on the tooth surface with various materials. CEREC AC Bluecam, Fastscan and Lava COS always require powder. On the other hand, E4D dentist does not require powder in most cases and iTero, PlanScan, Trios and Zfx Intrascan do not require the application of reflective agent.4 The literatures comparing the accuracy of IOSs largely concerned two subjects; the measurement of the internal gap of fabricated crowns and the measurement of the deviations after superposition of two images. There are various studies on these. In particular, the former internal gap study includes Seelbach et al.'s studyon the single crown and Schaefer et al.'s on the partial crown while the latter image superposition study includes Nedelcu et al.'s study on simplified and standardized scan body, Ender et al.'s on full arch, and Patzelt et al.'s on edentulous jaws.2, 3, 5, 6, 7, 8, 9, 10 Papaspyridakos and Wismeijer reported the investigation on the accuracy of digital implant impressions with IOS.11, 12, 13 In the study on the measurement of the gap of crowns, there is a limitation in determining the performance of the scanner itself due to the uncontrollable factors occurring in the fabrication of crowns while the image superposition study suggests neither the relationship with the actual crown nor the data for the specific important sites such as margins since this is a mere comparison of images. If the data on each section of the cavity form according to various restorations are measured and compared, it will be possible to compare the 3-dimensional reproduction capability of digital IOSs that expresses the crucial parts of the abutment affecting the fits of crowns.

The aim of this study is to perform the comparative evaluation of the 3-dimensional reproducibility of several intraoral scanners (IOS) by the restoration type. For this purpose, each intraoral scanner's precision and trueness were investigated by superposing 3-dimensioanl images and measuring the deviation at the main sites of the abutment in each section. Additionally, the differences depending on the scanning technology and characteristics of the scanner were compared and analyzed. The primary null hypothesis was that the 3-dimensional reproducibility of intraoral scanners did not differ by the restoration type and preparation location. The secondary null hypothesis was that the trueness of 3-dimensional images were not influenced by the scanning technology and characteristics of the scanners.

MATERIALS AND METHODS

In this study, a phantom model containing prefabricated abutment teeth was prepared. Right maxillary incisor and canine were selected as abutment teeth for a 3-unit fixed dental prosthesis. Right maxillary second molar was designated as an abutment for MO inlay, and the right mandibular second molar as that for a crown. Artificial tooth with an ideal shape were arranged at the left dentition for the occlusal stability during the digital buccal bite registration (Fig. 1). In order to obtain standardized preparations and to avoid an arbitrary preparation by a practitioner, the abutment teeth that had been prepared in precision processing by the manufacturer (A50H-set Prepared teeth assortment; Nissin) were used. To keep the reflectivity similar to the natural tooth, the metal tooth was not used. The setscrews which fixate the individual teeth were tightened firmly and cyanoacrylate cement was applied to prevent unwanted screw loosening. The tooth on the phantom was not removed or added and no external force was applied during the experiment. All the experiments were carried out at 23 ± 1℃ and 50 ± 5% relative humidity.

Fig. 1
Dental models with various preparation designs. Right maxillary incisor and canine (#11, 13); 3-unit fixed dental prosthesis, right maxillary second molar (#17); MO inlay, and right mandibular second molar (#47); crown.

In order to obtain the reference STL file, a desktop scanner (7 Series; Dental Wings Inc.) with the trueness of 15 µm was used. This scanner uses a laser light source with 5 axes of freedom, and thus is favorable for reproduction of an undercut or complicated shapes. Also, it has a high scanning capacity (140 × 140 × 100 mm), which enables scanning of a full arch restoration. 5 types of IOSs were used in this study: iTero (1st generation; ALIGN TECHNOLOGY, INC.), E4D dentist (initial version; E4D Technologies), Zfx Intrascan (Zfx GmhH), Trios (2nd generation; 3shape A/S) and Fastscan (IOS Technologies, Inc.). These were classified by the scanning technology such as Fastscan with the active triangulation, iTero, Trios and Zfx Intrascan with the confocal microscopy technology, and E4D dentist with the optical coherence tomography (Table 1).

A phantom containing prepared teeth was scanned using 5 different IOSs, and the respective test group was scanned 4 times. Scanning was performed according to the instruction of each manufacturer. In iTero and E4D groups, scanning was guided according to software instructions and abutment scan was finished first, followed by remaining teeth scanning. When scanning with Zfx Intrascan and Trios, occlusal sweep was done first and was followed by buccal and lingual sweeps in case of upper arch. Buccal and lingual sweeps were changed for lower arch scan. In case of Fastscan, posterior sections on both sides and anterior section were scanned separately and stitched together afterwards. Since Fastscan required the use of spray, its scan was performed at the last turn. 4 pairs of 3D data were acquired for each IOS. Data were obtained directly from the scanner of Fastscan and Zfx Intrascan while files were obtained from the manufacturer of E4D dentist, iTero and Trios.

According to ISO 12836, the deviation of measurements between the reference model and the intraoral scan model was termed the "trueness" of impression technique, and the deviation of measurements between digital models of the same intraoral scanner was "precision" of impression technique.

For the image superposition, a reverse engineering program (Rapidform; INUS Technology Inc.) was used. The reference STL file and intraoral scan STL files were superposed. Superposition was performed as follows: three points A, B and C were spotted on images, and while matching these, the two data were moved closer to each other, then, automerging was performed by computation using a command 'fine' (Fig. 2A). After superposition, the tooth model was divided in the mesiodistal, buccolingual, and right and left diagonal directions into a total of 8 sections. For the trueness investigation, the distance between the reference data and intraoral scan data were measured at 4 points (marginal, axial, line-angle, occlusal) on each section (Fig. 2B). In case of precision measurement, the distance between two intraoral scans within the same group at the aforementioned points on each section. For the mesiodistal and buccolingual comparison, mesial, diatal, buccal, and lingual sections were investigated and diagonal sections were excluded in the measurement.

Fig. 2
Illustration of the sectioning of the superposed datasets and selected sites from preparation ouline form. (A) section lines of crown and bridge groups (sections 1, 3, 5, and 7 were investigated for the mesiodistal and buccolingual comparison). (B) measured sites for crown and bridge groups. (C) section line of inlay group (sections a and b were inspected for the box and non-box comparision). (D) sites from preparation outline form for inlay group.

The inlay was measured likewise. After superposition, section was created in the mesiodistal and buccolingual direction of the tooth model, and the distance between the two data were measured at the corresponding 4 points in these sections. For the comparison of box and non-box regions, sections in the mesiodistal directions were investigated (Fig. 2C and Fig. 2D).

The difference in trueness and precision of intraoral scanners, and trueness according to preparation design, preparation site, and scanning technology were verified by analysis of variance (ANOVA). When the difference was significant, Tukey HSD test was used in multiple comparisons. In addition, independent two-samples t-test was used to compare the trueness according to image acquisition and necessity of coating. Data were analyzed using a statistical program IBM SPSS version 18.0 (IBM Corporation). The significance level was set at P = .05.

RESULTS

Each scanner had a different resolution, leading to the difference in the details of the model, and the number of polygones were in the order of iTero (42,687), Trios (28,280), Fastscan (20,210), E4D dentist (17,607), and Zfx intrascan (10,432) (Fig. 3). Scanners used in this study showed the deviations in trueness and precision of 70.1 µm and 58.9 µm on average, respectively. The average deviations in trueness and precision of Fastscan, iTero and Trios were in the range of 40 to 60 µm and 10 to 30 µm, respectively, while those of Zfx Intrascan and E4D dentist were in the range of 80 to 120 µm and 90 to 140 µm, respectively, which were significantly higher than the other three scanners (Table 2 and Fig. 4). In comparison by the restoration type, significantly higher trueness was observed in crown and inlay than in bridge irrespective of intraoral scanners. While no significant difference in overall mean trueness was observed among margin, axial wall, line-angle, and occlusal site, the trueness at line-angle was higher than those at the others in E4D dentist group. In comparison by the section, the mesiodistal direction showed significantly higher deviation than the buccolingual direction. Also, while the trueness is lower in the box overall, there was no significant difference in the trueness between the box and the non-box with Fastscan and iTero (Table 3 and Table 4). Trueness was observed to be in the increasing order of Fastscan with the active triangulation, iTero, Trios and Zfx Intrascan with the parallel confocal technology, and E4D dentist with the optical coherence tomography. In addition, the group using powder displayed lower trueness which means higher accuracy. On the other hand, no significant difference in trueness was observed between the still image acquisition and the video acquisition groups (Table 5).

Fig. 3
Number of polygons for selected maxillary incisor for the comparison of the each IOS's resolution.

Fig. 4
Illustration of absolute mean trueness and precision values of intraoral scanners. Same letters denote significant differences in between the groups at the 5% significance level.

Table 2
Trueness and precision of each intraoral scanner

Table 3
Trueness according to the preparation design

Table 4
Trueness according to the preparation site and in the inlay box

Table 5
Trueness according to the characteristics of IOS

DISCUSSION

The trueness and precision of IOS are among the most fundamental evaluation elements for the clinical application of the digital impression method. Although the evaluation method via image superposition has a fundamental error and limitation in the selection and superposition of random points on the program, and does not necessarily suggest the relationship with the actual restoration, it can still play an important role in the comparative evaluation of each IOS. In this in vitro study, the first null hypothesis was rejected because in general, the intraoral scanner had lower trueness in the inlay box, and it is considered that this was potentially caused by the limitation in the scan depth of IOS and the adjacent teeth undercut that hampered the image acquisition. For the trueness depending on the scanning technology and characteristics of scanners, the second null hypothesis was rejected because the acquired image displayed deviations in trueness by the scanning technology and characteristics of scanners.

A desktop scanner, Dental Wings 3D scanner, was used as a reference scanner as in previous studies,14, 15, 16, 17 and the reason that an IOS was not selected as a reference scanner was that the accumulation of errors caused by sudden actions during scanning due to the high degree of freedom could lead to the bending of the model. A trueness of up to 15 µm was reported on the reference scanner for a single measurement case of the plaster model,18 and it was shown that the internal and marginal gaps of the crown on the experimental model fabricated using this scanner were up to the range of 23.45 to 42.43 µm.19 In this study, the precision of this reference scanner was measured, which was 35.3 ± 13.9 µm.

There have been many studies which evaluated the performance of intraoral scanners by measuring fit of restoration or superimposing scan data. Seelbach et al. fabricated a single crown using CEREC, iTero and Lava COS, and measured the internal gap and marginal fit, which were 29 to 88 µm and 30 to 41 µm, respectively, in agreement with the values measured on the crown fabricated by the conventional method.5 In a study of van der Meer et al. using the implant scan body, the distance errors between the scan bodies measured by CEREC, iTero and Lava COS were 79.6 to 81.6 µm um, 61.1 to 70.5 µm and 14.6 to 23.5 µm, respectively.13 Mehl et al. analyzed the accuracy by superposing the model data acquired from a desktop scanner and CEREC AC Bluecam, and reported that the deviation was 19.2 µm for a single tooth and 35 µm for a quadrant.20 Kim et al. scanned a polyurethane model fabricated by milling process using scan data of iTero and a traditional plaster model with a desktop scanner, and compared the data.21 They reported that the deviation was 23.9 ± 17.6 µm for a single tooth and the model fabricated using IOS showed a significantly higher deviation than the traditionally fabricated model, which, however, was clinically acceptable.

Ender et al. reported that a deviation of 58.6 µm was observed in the comparison of accuracy in the full arch models between CEREC AC system and a reference scanner, and a deviation of 32.4 µm was observed in the analysis of precision by comparing between the CEREC data.22 They concluded that this had a significantly lower accuracy than the traditional method. In their further study, it was reported that the deviations in accuracy and precision for CEREC Bluecam, iTero, Lava COS and Omnicam were up to 29.4 to 44.9 µm um and 19.5 to 63.0 µm, respectively,8 which leads to a conclusion that these IOSs could be applied to full mouth scan cases within the limitations of in-vitro study. Patzelt et al. also reported studies on the accuracy of full arch scan and fully edentulous jaw digitization.9, 10 As the intraoral scanner is a handheld device, the image stitching error from the uneven and abrupt movement of scanner wand might cause the deformation of scan data. These studies have a great significance because they were focused on the deformation of entire arch. This study was conducted to investigate another characteristic of intraoral scanner. The depth of field is related to not only the difficulty of scanning but also the accuracy of scan data. Learning curve of intraoral scanners with narrow depth of field is very long because the operator has to keep distance from patient's teeth while watching computer display. When the maximum range that intraoral scanner can reach is short, the image acquisition could not be possible at the long tooth with deep preparation enclosed by tall adjacent teeth. In this study, the accuracy of various intraoral scanners was evaluated at the microscopic level of the abutment teeth by sectioning and picking up the anatomic structures which might be critical for the adaptation of restorations.

In studies that compared the accuracy of IOSs, Nedelcu et al. analyzed the accuracy and precision of 4 IOSs by the superposition method based on a desktop scanner as a reference, showing that the accuracy and precision of CEREC were better than those of E4D dentist.2 They suggested that this was because E4D dentist did not have sufficient transition area that was relatively clearly defined, and much noise appeared on the surface. In this study, iTero displayed better results compared to E4D dentist and Zfx Intrascan, which was in agreement with previous studies.

Schaefer et al. reported that marginal discrepancy of a partial crown (MOD inlay) fabricated using iTero was up to 90 µm, while An et al. analyzed the marginal fits of zirconia coping fabricated using iTero by a replica technique, which was shown to be up to 103.05 µm for a single tooth.3, 23 Also, Keul et al. performed experiments similarly and reported that the marginal opening of 4-unit zirconia framework was up to 127.23 µm.24 In our study, an abutment scanned by iTero displayed higher deviations in the increasing order of the inlay, crown and bridge, and this trend is consistent with preceding studies. Also, in general, the trueness of IOS was lower in the fixed dental prosthesis, which can be attributed to the errors in the image acquisition and stitching processes due to the presence of the glossy pontic space with less surface characteristics.

Brawek et al. measured the fits of the crown fabricated using Lava COS and CEREC AC, and showed that the internal gap was larger than the marginal gap.25 They suggested that this larger internal gap could be attributed to the fact that there was a lower limit in the size of the milling tool for processing of the interior of the restoration, and thus the software relief had to be made to process the angled corner of the interior of the restoration, which was compounded by the errors in replica technique.26 In this study, the deviations between the marginal and internal gaps did not show any significant difference unlike previous studies, which can be attributed to the difference in the experimental method since this study only compared the abutment forms observed on the scan data. Also, in comparison of sections, the trueness in the mesiodistal section was lower than in the buccolingual section, and it is considered that this was because multiple scans were necessary at the mesiodistal site due to the undercut by adjacent teeth, and subsequently the amount of data to be stitched was larger.

There are various factors that affect the reproducibility of an IOS, including the scanning technology, data processing algorithm, whether or not to use powder and image acquisition method. Active triangulation is a traditional scanning technology and has been frequently applied in the desktop scanner, which offers the highest trueness if the condition is right. In comparison, the parallel confocal technology does not require a certain distance for focusing, and thus images can be acquired regardless of whether the scanner tip is attached to the teeth when the oral cavity is scanned. On the other hand, the optical coherence tomography has a high resolution that can create an image of the micromorphology of the abutment by combining the optical interference phenomenon and the confocal microscopy technology. In this study, the deviations in trueness and precision increased in the order of Fastscan with the active triangulation technology, iTero, Trios and Zfx Intrascan with the parallel confocal technology, and E4D dentist with the optical coherence tomography technology. In a study of Nedelcu et al. on CEREC, E4D dentist, iTero and Lava COS, the better result was obtained in the order of the active triangulation, confocal microscopy and optical coherence tomography, consistent with this study.2 However, Ender et al. and Seelbach et al. studied using CEREC, iTero and Lava COS, and reported no notable difference between the active triangulation and parallel confocal scanners,5, 7 while Schaefer et al. suggested that the confocal microscopy displayed higher accuracy than the active triangulation.3 These latter studies are in disagreement with our study, and thus yet another different result may be expected in the actual clinical setting, which warrants further studies on this matter.

In studies of Ender et al. and Seelbach et al., similar accuracy was observed regardless of the use of powder while Schaefer et al. observed a significant higher accuracy in the powder free group.3, 5, 7 In contrast, Nedelcu et al. suggested that the group using powder showed higher accuracy, in consistent with our study.2 In addition, they prepared an experimental group with excessive spray coating, and investigated its impact on the accuracy, which yielded no significant difference. Rupf et al. reported that 4.6 × 106 of fine and ultrafine particles were accumulated per minute on average under conventional dental suction if scanning spray was applied, and recommended high volume evacuation for the prevention of the exposure to fine or ultrafine particles.27 Although the use of powder can increase the scanning speed and the ease of scan, inconvenience to patients may be caused during the application and removal process of powder, and the powder remaining in the oral cavity may exert a harmful effect on the body.

Schaefer et al. and Nedelcu et al. classified IOSs largely into video acquisition and still image acquisition types depending on the scanning method,2, 3 and Ender et al. analyzed the accuracy of Bluecam and Lava COS by superposing images from these scanners with those of the reference scanner, which showed that there was no significant difference in the deviations between the video acquisition and still image acquisition methods.6 Their further study on CEREC Bluecam, CEREC Omnicam, iTero and Lava COS showed similar results.7, 8 Also, the studies of Schaefer et al. and Seelbach et al. did not show any significant difference in the accuracy between the video acquisition and still image acquisition types.3, 5 Likewise, in this study, a significant difference in trueness between the two scanning methods was not observed. However, the video acquisition method has a clear advantage that scanning is convenient and images can be monitored during scanning.

A limitation of this study is that the data on the restoration itself fabricated using IOSs could not be provided. Also, as actual scanning conditions were not taken into consideration in this study, the lower trueness is expected in the actual clinical setting than in this study due to the level of patient's compliance, the skill of a practitioner, the presence of saliva during scanning, reflectivity of the tooth and the intraoral structure. Therefore, it is considered that further study using the final restoration in the clinical setting is necessary.

CONCLUSION

Except for 2 intraoral scanners, Fastscan, iTero and Trios displayed comparable levels of trueness and precision values in tested phantom model. The difference in trueness was observed depending on the restoration type, the preparation out line form, the scanning technology and the application of power.

Notes

This work was supported by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT & Future Planning (NRF-2013R1A1A1076022).

References

    1. Miyazaki T, Hotta Y, Kunii J, Kuriyama S, Tamaki Y. A review of dental CAD/CAM: current status and future perspectives from 20 years of experience. Dent Mater J 2009;28:44–56.
    1. Nedelcu RG, Persson AS. Scanning accuracy and precision in 4 intraoral scanners: an in vitro comparison based on 3-dimensional analysis. J Prosthet Dent 2014;112:1461–1471.
    1. Schaefer O, Decker M, Wittstock F, Kuepper H, Guentsch A. Impact of digital impression techniques on the adaption of ceramic partial crowns in vitro. J Dent 2014;42:677–683.
    1. Logozzo S, Zanetti EM, Franceschini G, Kilpelä A, Mäkynen A. Recent advances in dental optics – Part I: 3D intraoral scanners for restorative dentistry. Opt Laser Eng 2014;54:203–221.
    1. Seelbach P, Brueckel C, Wöstmann B. Accuracy of digital and conventional impression techniques and workflow. Clin Oral Investig 2013;17:1759–1764.
    1. Ender A, Mehl A. Full arch scans: conventional versus digital impressions-an in-vitro study. Int J Comput Dent 2011;14:11–21.
    1. Ender A, Mehl A. Influence of scanning strategies on the accuracy of digital intraoral scanning systems. Int J Comput Dent 2013;16:11–21.
    1. Ender A, Mehl A. In-vitro evaluation of the accuracy of conventional and digital methods of obtaining full-arch dental impressions. Quintessence Int 2015;46:9–17.
    1. Patzelt SB, Vonau S, Stampf S, Att W. Assessing the feasibility and accuracy of digitizing edentulous jaws. J Am Dent Assoc 2013;144:914–920.
    1. Patzelt SB, Emmanouilidi A, Stampf S, Strub JR, Att W. Accuracy of full-arch scans using intraoral scanners. Clin Oral Investig 2014;18:1687–1694.
    1. Papaspyridakos P, Gallucci GO, Chen CJ, Hanssen S, Naert I, Vandenberghe B. Digital versus conventional implant impressions for edentulous patients: accuracy outcomes. Clin Oral Implants Res 2016;27:465–472.
    1. Andriessen FS, Rijkens DR, van der Meer WJ, Wismeijer DW. Applicability and accuracy of an intraoral scanner for scanning multiple implants in edentulous mandibles: a pilot study. J Prosthet Dent 2014;111:186–194.
    1. van der Meer WJ, Andriessen FS, Wismeijer D, Ren Y. Application of intra-oral dental scanners in the digital workflow of implantology. PLoS One 2012;7:e43312.
    1. Kim SR, Lee WS, Kim WC, Kim HY, Kim JH. Digitization of dental alginate impression: Three-dimensional evaluation of point cloud. Dent Mater J 2015;34:835–840.
    1. Atzeni E, Iuliano L, Minetola P, Salmi A. Proposal of an innovative benchmark for accuracy evaluation of dental crown manufacturing. Comput Biol Med 2012;42:548–555.
    1. Paranhos LR, Lima CS, da Silva RH, Daruge Júnior E, Torres FC. Correlation between maxillary central incisor crown morphology and mandibular dental arch form in normal occlusion subjects. Braz Dent J 2012;23:149–153.
    1. Paranhos LR, Zaroni M, Carli JP, Okamoto R, Zogheib LV, Torres FC. Association between the facial type and morphology of the upper central incisor in normal occlusion subjects. J Contemp Dent Pract 2014;15:29–33.
    1. Martin CB, Chalmers EV, McIntyre GT, Cochrane H, Mossey PA. Orthodontic scanners: what's available? J Orthod 2015;42:136–143.
    1. Sannino G, Gloria F, Schiavetti R, Ottria L, Barlattani A. Dental Wings CAD/CAM system precision: an internal and marginal fit sperimental analisys. Oral Implantol (Rome) 2009;2:11–20.
    1. Mehl A, Ender A, Mörmann W, Attin T. Accuracy testing of a new intraoral 3D camera. Int J Comput Dent 2009;12:11–28.
    1. Kim SY, Kim MJ, Han JS, Yeo IS, Lim YJ, Kwon HB. Accuracy of dies captured by an intraoral digital impression system using parallel confocal imaging. Int J Prosthodont 2013;26:161–163.
    1. Ender A, Mehl A. Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. J Prosthet Dent 2013;109:121–128.
    1. An S, Kim S, Choi H, Lee JH, Moon HS. Evaluating the marginal fit of zirconia copings with digital impressions with an intraoral digital scanner. J Prosthet Dent 2014;112:1171–1175.
    1. Keul C, Stawarczyk B, Erdelt KJ, Beuer F, Edelhoff D, Güth JF. Fit of 4-unit FDPs made of zirconia and CoCr-alloy after chairside and labside digitalization-a laboratory study. Dent Mater 2014;30:400–407.
    1. Brawek PK, Wolfart S, Endres L, Kirsten A, Reich S. The clinical accuracy of single crowns exclusively fabricated by digital workflow-the comparison of two systems. Clin Oral Investig 2013;17:2119–2125.
    1. Sturdevant JR, Bayne SC, Heymann HO. Margin gap size of ceramic inlays using second-generation CAD/CAM equipment. J Esthet Dent 1999;11:206–214.
    1. Rupf S, Berger H, Buchter A, Harth V, Ong MF, Hannig M. Exposure of patient and dental staff to fine and ultrafine particles from scanning spray. Clin Oral Investig 2015;19:823–830.

Metrics
Share
Figures

1 / 4

Tables

1 / 5

ORCID IDs
Funding Information
PERMALINK