LASER THERAPY
Online ISSN : 1884-7269
Print ISSN : 0898-5901
ISSN-L : 0898-5901
Original Article
Four different diode lasers comparison on soft tissues surgery: a preliminary ex vivo study
Carlo FornainiElisabetta MerigoMichele SozziJean-Paul RoccaFederica PoliStefano SelleriAnnamaria Cucinotta
Author information
JOURNAL FREE ACCESS

2016 Volume 25 Issue 2 Pages 105-114

Details
Abstract

Objectives: The introduction of diode lasers in dentistry had several advantages, principally consisting on the reduced size, reduced cost and possibility to beam delivering by optical fibbers. Up today only the wavelengths around 810 and 980 nm were the most utilized in oral surgery but recently more different lasers had been proposed.The aim of this study was to compare the efficacy of four diode laser wavelengths (810, 980, 1470 and 1950 nm) for the ablation of soft tissues.
Material and Methods: Specimens were surgically collected from the dorsal surface of four bovine tongues and irradiated by four different diode wavelengths. Thermal increase was measured by two thermocouples, the first at a depth of 0.5 mm, and the second at a depth of 2 mm.
Initial and final surface temperatures were recorded by IR thermometer.
Epithelial changes, connective tissue modifications, presence of vascular modification and incision morphology were histologically evaluated by two blind pathologists.
Results:The time necessary to perform the excision varied between 271 seconds (808 nm, 2W) and 112 seconds (1950 nm, 4W). Temperature increase superficial level varied from 16.3° (980 nm, 4W) and 9.2° (1950 nm, 2 W). The most significant deep temperature increase was recorded by 980 nm, 4 W (17.3°) and the lowest by 1950 nm, 2 W (9.7°). The width of epithelial tissue injuries varied between 74 μm from 1950 nm diode laser at 2 W to 540 μm for 1470 nm diode laser at 4 W.
Conclusions: The quality of incision was better and the width of overall tissue injuries was minor in the specimens obtained with higher wavelength (1950 nm) at lower power (2W).

Content from these authors
© 2016 Japan Medical Laser Laboratory
Previous article Next article
feedback
Top