2012 年 68 巻 12 号 p. 1662-1668
Multi detector-row computed tomography with 64 data acquisition systems are widely used for coronary CT angiography with an electrocardiograph (ECG) gated helical scan (HS). Step and shoot with ECG gated non-helical scan (snap shot pulse: SSP) could reduce exposure dose but banding artifact-like discontinuity was observed between adjacent slabs on volume rendering (VR) and curved planner reconstruction (CPR). Therefore, we investigated the factors that influence continuity of VR and CPR images by calculating image properties of Z-axis direction of slab. The observer performance studies were performed for evaluating continuity of simulated blood vessels of VR and CPR images at simulated heart rates: 50, 55, 57 and 60 beat per minute (bpm). As a result, the value of SD at both slab edges in SSP were 20.5% lower than middle part of slab and differences of value of SD were up to 4.4 between adjacent slab edges. Slice thickness of both slab edges were 20.3% thinner than that of the peripheral part of slab. At the border of the adjacent slab, the position of the simulated blood vessel was shifted. VR images of SSP at 57 bpm was indicated as the highest score and HS was significantly superior to SSP at 55 and 60 bpm (p<0.05). In CPR images, there were no significant differences at all simulated heart rates. In conclusion, we considered that VR images of SSP were influenced heart rates except 57 bpm (resonance case) and there was little difference of visibility for discontinuity of both CPR images obtained by SSP and HS.