MR装置の安全管理に関する実態調査の報告 —思った以上に事故は起こっている—
土'井 司山谷 裕哉上山 毅錦 成郎小倉 明夫川光 秀昭土橋 俊男奥秋 知幸松田 豪熊代 正行
著者情報
ジャーナル フリー

2011 年 67 巻 8 号 p. 895-904

詳細
抄録

Using a questionnaire, we surveyed 2,500 facilities in Japan to clarify medical accidents concerning the magnetic resonance device and its environment. Data derived from 1,319 valid responses (52.8%), allowed us to analyze the situation of (or the reason for) the occurrence of the accidents and their environmental factors. Five hundred and nine facilities (39% of all facilities) had the experience of magnetically induced displacement of the large ferromagnetic material. Intravenous (I.V.) drip stands were involved the largest number of them: 31% (228 cases). Oxygen bottles had the second largest number of incidents: 20%. There were also many incidents involving various materials brought in by non-medical staff (e.g. stepladder for construction). About 20% of the accidents occurred outside of working hours. Patients in 12% of the facilities (154 facilities) experienced burns. In 39 of the cases, burns were received to the inside of the thighs. In 38 of the cases, patients received burns from an electrical cable touching the skin. There were also frequent incidents of burning regarding the boa. We received reports of burns and pain from the halo vest even though it’s required to be worn for MR safety. Regarding incidents of contraindications, 280 patients with pacemakers were brought into the magnetic resonance (MR) inspection room. Twelve percent of the facilities experienced natural quench. Lack of training for the staff who introduce and operate high magnetic field devices are considered involving frequently occurring accidents of attractions and burns at hospitals with over 500 beds caused by carrying in materials.

著者関連情報
© 2011 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top