冠動脈CTにおける Step&Shoot overlap reconstruction(SSOR)の有用性に関する検討
蝶野 大樹平野 透井上 健徳安 真一樋口 江
著者情報
ジャーナル フリー

2011 年 67 巻 8 号 p. 905-911

詳細
抄録

Background: Step&Shoot cardiac computed tomography (CT) provides the benefit of significant reduction in radiation dose compared to helical cardiac CT acquisitions. We think that a difference occurs in image quality by presence of overlap reconstruction (fractionated sentence, attempted to clarify). Purpose: We studied the utility of Step&Shoot overlap reconstruction (SSOR). Methods: We evaluated image quality of SSOR by comparing z-axis spatial resolution on various s in scanners that SSOR is possible and for those that are impossible. Results: SSOR indicated better z-axis spatial resolution and less variation over the field of view (FOV) (in z & x-y directions) as compared with Step&Shoot without overlap reconstruction (SS). In addition, SS showed inadequate image reproducibility due to aliasing error in z-direction (lack of sampling interval). Conclusions: SSOR would contribute to improvement of the image quality of Step&Shoot cardiac CT.