OSL線量計を用いたMDCT装置のCT室内散乱線分布の測定
富田 博信諸澄 邦彦
著者情報
ジャーナル フリー

2004 年 60 巻 11 号 p. 1550-1554

詳細
抄録

The recent introduction of multi-detector row computed tomography (MDCT) has made it possible to scan the entire abdomen within approximately 10 sec in procedures such as interventional radiology computed tomography (IVRCT), which are associated with operator exposure. Therefore, anxious patients and patients who are not able to remain still can be examined with an assistant. In the present study, radiation exposure to the assistant was estimated, and the distribution of scattered radiation near the gantry was measured using an optically stimulated luminescence (OSL) dosimeter. Simultaneous measurements were obtained using a direction storage (DIS) dosimeter for reference. The maximum value of 1.47 mSv per examination was obtained at the point closest to the gantry's center (50 cm from the center at a height of 150 cm above the floor). In addition, scattered radiation decreased as the measurement point was moved further from the gantry's center, falling below the limit of detection (0.1 mSv or less) at 200 cm and at the sides of the gantry. OSL dosimeters are also employed as personal dosimeters, permitting reliable values to be obtained easily. They were found to be an effective tool for the measurement of scattered radiation, as in the present study, helping to provide better understanding of the distribution of scattered radiation within the CT room.

著者関連情報
© 2004 公益社団法人 日本放射線技術学会