Linshom thermodynamic sensor is a reliable alternative to capnography for monitoring respiratory rate

verfasst von: David Preiss, Benjamin A. Drew, James Gosnell, Bhavani S. Kodali, James H. Philip, Richard D. Urman

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2018

Einloggen, um Zugang zu erhalten

Abstract

Monitoring ventilation accurately is a technically challenging, yet indispensable aspect of patient care in the intra- and post-procedural settings. A new prototypical device known as the Linshom Respiratory Monitoring Device (LRMD) has been recently designed to non-invasively, inexpensively, and portably measure respiratory rate. The purpose of this study was to measure the accuracy and variability of LRMD measurements of respiratory rate relative to the measurement of capnography. In this prospective study, participants were enrolled and individually fitted with a face mask monitored by the LRMD and capnography. With a baseline oxygen flow rate and digital metronome to pace their respiratory rate, the participants were instructed to breathe at 10 breaths per minute (bpm) for 3 min, 20 bpm for 3 min, 30 bpm for 3 min, 0 bpm for 30 s, and resume regular breathing for 30 s. Both sensors were connected to a computer for continuous temperature and carbon dioxide waveform recordings. The data were then retrospectively analyzed. Twenty-six healthy volunteers, mean (range) age 27.8 (23–37) and mean (range) BMI 23.1 (18.8–29.2) kg/m2 were recruited. There were 15 males (57.7%) and 11 females (42.3%). After excluding 3 subjects for technical reasons, 13,800 s of breathing and 4,140 expiratory breaths were recorded. Throughout the protocol, the average standard deviation (SD) for the LRMD and capnography was 1.11 and 1.81 bpm, respectively. The overall mean bias (±2SD) between LRMD and capnography was −0.33 (±0.1.56) bpm. At the lowest and intermediate breathing rates reflective of hypoventilation and normal ventilation, the LRMD variance was 0.55 and 1.23 respectively, compared to capnography with 5.54 and 7.47, respectively. At higher breathing rates indicative of hyperventilation, the variance of the test device was 4.52, still less than that of capnography at 5.73. This study demonstrated a promising correlation between the LRMD and capnography for use as a respiratory rate monitor. The LRMD technology may be a significant addition to monitoring vital signs because it offers a minimally intrusive opportunity to detect respiratory rate and apnea, without expensive or complex anesthetic equipment, before the need for life-saving resuscitation arises.
Literatur
1.
Zurück zum Zitat Jones JG, Sapsford DJ, Wheatley RG. Postoperative hypoxemia: mechanisms and time course. Anesthesia. 1990;45:566–73.CrossRef Jones JG, Sapsford DJ, Wheatley RG. Postoperative hypoxemia: mechanisms and time course. Anesthesia. 1990;45:566–73.CrossRef
2.
Zurück zum Zitat Chiumello D, Chevallard G, Gregoreti C. Non-invasive ventilation in postoperative patients: A systematic review. Intensive Care Med. 2011;37(6):918–29.CrossRefPubMed Chiumello D, Chevallard G, Gregoreti C. Non-invasive ventilation in postoperative patients: A systematic review. Intensive Care Med. 2011;37(6):918–29.CrossRefPubMed
3.
Zurück zum Zitat Taylor S, Kirton OC, Staff I, Kozol RA. Postoperative day one: a high risk period for respiratory events. Am J Surg. 2005;190:752–6. Taylor S, Kirton OC, Staff I, Kozol RA. Postoperative day one: a high risk period for respiratory events. Am J Surg. 2005;190:752–6.
4.
Zurück zum Zitat McAlister FA, Bertsch K, Man J, et al. Incidence of and risk factors for pulmonary complications after nonthoracic surgery. Am J Respir Crit Care Med. 2005;171:514–7.CrossRefPubMed McAlister FA, Bertsch K, Man J, et al. Incidence of and risk factors for pulmonary complications after nonthoracic surgery. Am J Respir Crit Care Med. 2005;171:514–7.CrossRefPubMed
5.
Zurück zum Zitat Hall JC, Tarala RA, Hall JL, Mander J. A multivariate analysis of the risk of pulmonary complications after laparotomy. Chest. 1991;99:923.CrossRefPubMed Hall JC, Tarala RA, Hall JL, Mander J. A multivariate analysis of the risk of pulmonary complications after laparotomy. Chest. 1991;99:923.CrossRefPubMed
6.
Zurück zum Zitat Schein RM, Hazday N, Pena M, Ruben BH, Sprung CL. Clinical antecedents to in-hospital cardiopulmonary arrest. Chest. 1990;98:1388–92.CrossRefPubMed Schein RM, Hazday N, Pena M, Ruben BH, Sprung CL. Clinical antecedents to in-hospital cardiopulmonary arrest. Chest. 1990;98:1388–92.CrossRefPubMed
7.
Zurück zum Zitat Mailey J, Digiovine B, Baillod D, Gnam G, Jordan J, Rubinfeld I. Reducing hospital standardized mortality rate with early interventions. J Trauma. Nursing. 2006;13:178–82.CrossRef Mailey J, Digiovine B, Baillod D, Gnam G, Jordan J, Rubinfeld I. Reducing hospital standardized mortality rate with early interventions. J Trauma. Nursing. 2006;13:178–82.CrossRef
8.
Zurück zum Zitat Lawrence VA, Hilsenbeck SG, Mulrow CD, et al. Incidence and hospital stay for cardiac and pulmonary complications after abdominal surgery. J Gen Intern Med. 1995;10:671.CrossRefPubMed Lawrence VA, Hilsenbeck SG, Mulrow CD, et al. Incidence and hospital stay for cardiac and pulmonary complications after abdominal surgery. J Gen Intern Med. 1995;10:671.CrossRefPubMed
9.
Zurück zum Zitat Practice guidelines for post anesthetic care. A report by the American Society of Anesthesiologists Task Force on postanesthetic care. Anesthesiology. 2002;96:742–72. Practice guidelines for post anesthetic care. A report by the American Society of Anesthesiologists Task Force on postanesthetic care. Anesthesiology. 2002;96:742–72.
10.
Zurück zum Zitat World Health Organization. Fourth Programme report, 1988–1989: ARI programme for control of acute respiratory infections. Geneva: WHO; 1990. p. 31. World Health Organization. Fourth Programme report, 1988–1989: ARI programme for control of acute respiratory infections. Geneva: WHO; 1990. p. 31.
11.
Zurück zum Zitat Semmes B, Tobin M, Snyder J, Grenvik A. Subjective and objective measurement of tidal volume in critically ill patients. Chest. 1985;78:577–9.CrossRef Semmes B, Tobin M, Snyder J, Grenvik A. Subjective and objective measurement of tidal volume in critically ill patients. Chest. 1985;78:577–9.CrossRef
12.
Zurück zum Zitat Lovett P, Buchwald J, Sturmann K, Bijur P. The vexations vital: Neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage. Ann Emerg Med. 2005;45(1):68–76.CrossRefPubMed Lovett P, Buchwald J, Sturmann K, Bijur P. The vexations vital: Neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage. Ann Emerg Med. 2005;45(1):68–76.CrossRefPubMed
13.
Zurück zum Zitat Gaucher A, Frasca D, Mimoz O, Debaene B. Accuracy of respiratory rate monitoring by capnometry using the Capnomask in extubated patients receiving supplemental oxygen after surgery. Br J Anaesth. 2012;108(2):316–20.CrossRefPubMed Gaucher A, Frasca D, Mimoz O, Debaene B. Accuracy of respiratory rate monitoring by capnometry using the Capnomask in extubated patients receiving supplemental oxygen after surgery. Br J Anaesth. 2012;108(2):316–20.CrossRefPubMed
14.
Zurück zum Zitat Hogan J. Why don’t nurses monitor the respiratory rates of patients? Br J Nurs. 2006;15(9):489–92.CrossRefPubMed Hogan J. Why don’t nurses monitor the respiratory rates of patients? Br J Nurs. 2006;15(9):489–92.CrossRefPubMed
16.
Zurück zum Zitat Vimlati L, Gilsanz F, Goldik Z. Quality and safety guidelines of postanaesthesia care: working party on post anaesthesia care (approvedby the European board and section of anaesthesiology, Union Europeenne des MedecinsSpecialistes). Eur J Anaesthesiol. 2009;26:715–21.CrossRefPubMed Vimlati L, Gilsanz F, Goldik Z. Quality and safety guidelines of postanaesthesia care: working party on post anaesthesia care (approvedby the European board and section of anaesthesiology, Union Europeenne des MedecinsSpecialistes). Eur J Anaesthesiol. 2009;26:715–21.CrossRefPubMed
17.
Zurück zum Zitat Standards for Basic Anesthetic Monitoring. ASA House of Delegates on October 21, 1986, last ammended on October 20, 2010, and last affirmed on October 28, 2015. Standards for Basic Anesthetic Monitoring. ASA House of Delegates on October 21, 1986, last ammended on October 20, 2010, and last affirmed on October 28, 2015.
18.
Zurück zum Zitat Ramsay MA, Usman M, Lagow E, Mendoza M, Untalan E, De Vol E. The accuracy, precision, and reliability of measuring ventilatory rate and detecting ventilatory pause by rainbow acoustic monitoring and capnometry. AnesthAnalg. 2013;117(1):69–75. Ramsay MA, Usman M, Lagow E, Mendoza M, Untalan E, De Vol E. The accuracy, precision, and reliability of measuring ventilatory rate and detecting ventilatory pause by rainbow acoustic monitoring and capnometry. AnesthAnalg. 2013;117(1):69–75.
19.
20.
21.
Zurück zum Zitat Mimoz O, Benard T, Gaucher A, Frasca D, Debaene B. Accuracy of respiratory rate monitoring using a non-invasive acoustic method after general anesthesia. British. J Anesth. 2012;108(5):872–5. Mimoz O, Benard T, Gaucher A, Frasca D, Debaene B. Accuracy of respiratory rate monitoring using a non-invasive acoustic method after general anesthesia. British. J Anesth. 2012;108(5):872–5.
22.
Zurück zum Zitat Jovanov E, Raskovic D, Hormigo R. Thermistor-based breathing sensor for circadian rhythm evaluation. Biomed Sci Instrument Conf Proc. 2001;37:493–7. Jovanov E, Raskovic D, Hormigo R. Thermistor-based breathing sensor for circadian rhythm evaluation. Biomed Sci Instrument Conf Proc. 2001;37:493–7.
23.
Zurück zum Zitat Nilsson L, Johansson A, Kalman S. Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique. J Clin Monit Comput. 2000;16(4):309–15.CrossRefPubMed Nilsson L, Johansson A, Kalman S. Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique. J Clin Monit Comput. 2000;16(4):309–15.CrossRefPubMed
24.
Zurück zum Zitat Ono Y, Mohamed D, Kobayashi M, Jen C. Piezoelectric membrane sensor and technique for breathing monitoring. Proceedings of IEEE Ultrasonics Symposium Conference Proceedings 2008, 795–798. Ono Y, Mohamed D, Kobayashi M, Jen C. Piezoelectric membrane sensor and technique for breathing monitoring. Proceedings of IEEE Ultrasonics Symposium Conference Proceedings 2008, 795–798.
25.
Zurück zum Zitat Chen Z, Lau D, Teo J, Ng S, Yang X, Kei P. Simultaneous measurement of breathing rate and heart rate using a microbend multimode fiber optic sensor. J Biomed Opt. 2014;19(5):057001.CrossRefPubMed Chen Z, Lau D, Teo J, Ng S, Yang X, Kei P. Simultaneous measurement of breathing rate and heart rate using a microbend multimode fiber optic sensor. J Biomed Opt. 2014;19(5):057001.CrossRefPubMed
26.
Zurück zum Zitat Boccanfuso L, O’Kane J. Remote measurement of breathing rate in real time using a high precision, single-point infrared temperature sensor. Proceedings of the IEEE RAS and EMBS International Conference on Biomedical Robotics and Biomechatronics; 2012, 1704–1709. Boccanfuso L, O’Kane J. Remote measurement of breathing rate in real time using a high precision, single-point infrared temperature sensor. Proceedings of the IEEE RAS and EMBS International Conference on Biomedical Robotics and Biomechatronics; 2012, 1704–1709.
27.
Zurück zum Zitat Zhang Z, Zheng J, Wu H, Wang W, Wang B, Liu H. Development of a respiratory inductive plethysmography module supporting multiple sensors for wearable systems. Sensors. 2012;12(10):13167–84.CrossRefPubMedPubMedCentral Zhang Z, Zheng J, Wu H, Wang W, Wang B, Liu H. Development of a respiratory inductive plethysmography module supporting multiple sensors for wearable systems. Sensors. 2012;12(10):13167–84.CrossRefPubMedPubMedCentral
28.
Zurück zum Zitat Egleston CV, Ben Aslam H, Lambert MA. Capnography for monitoring non-intubated spontaneously breathing patients in an emergency room setting. J AccidEmerg Med. 1997;14:222–4.CrossRef Egleston CV, Ben Aslam H, Lambert MA. Capnography for monitoring non-intubated spontaneously breathing patients in an emergency room setting. J AccidEmerg Med. 1997;14:222–4.CrossRef
29.
Zurück zum Zitat Hok B, Wiklund L, Henneberg S. A new respiratory rate monitor: development and initial clinical experience. J Clin Monit Comput. 1993;10(2):101–7.CrossRef Hok B, Wiklund L, Henneberg S. A new respiratory rate monitor: development and initial clinical experience. J Clin Monit Comput. 1993;10(2):101–7.CrossRef
30.
Zurück zum Zitat Manczur T, Greenough A, Hooper R, Allen K, Latham S, Price JF, Rafferty GF. Tidal breathing parameters in young children: comparison of measurement by respiratory inductance plethysmography to a facemask, pneumotachograph system. Pediatr Pulmonol. 1999;28(6):436–41.CrossRefPubMed Manczur T, Greenough A, Hooper R, Allen K, Latham S, Price JF, Rafferty GF. Tidal breathing parameters in young children: comparison of measurement by respiratory inductance plethysmography to a facemask, pneumotachograph system. Pediatr Pulmonol. 1999;28(6):436–41.CrossRefPubMed
31.
Zurück zum Zitat Larsson C, Staun P. Evaluation of a new fiber-optical monitor for respiratory rate monitoring. J Clin Monit Comput. 1999;15(5):295–8.CrossRefPubMed Larsson C, Staun P. Evaluation of a new fiber-optical monitor for respiratory rate monitoring. J Clin Monit Comput. 1999;15(5):295–8.CrossRefPubMed
32.
Zurück zum Zitat Van Loon, K. Breteler MJ, Van Wolfwinkel L, RheineckLeyssius AT, Kossen S, Kalkman CJ, Van Zaane B, Peelen LM. Wireless non-invasive continous respiratory monitoring FMCW radar: a clinical validation study. Van Loon, K. Breteler MJ, Van Wolfwinkel L, RheineckLeyssius AT, Kossen S, Kalkman CJ, Van Zaane B, Peelen LM. Wireless non-invasive continous respiratory monitoring FMCW radar: a clinical validation study.
33.
Zurück zum Zitat Hers, V, Corbugy D, Joslet I, Hermant P, Demarteau J, Delhougne B, Vandermoten G, Hermanne, JP. New Concept using Passive Infrared technolgoy for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients. J ClinMonitComput. 2013; 27(5): 521–529. Hers, V, Corbugy D, Joslet I, Hermant P, Demarteau J, Delhougne B, Vandermoten G, Hermanne, JP. New Concept using Passive Infrared technolgoy for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients. J ClinMonitComput. 2013; 27(5): 521–529.
34.
Zurück zum Zitat Sathyamoorthy M, Lerman J, Feldman D, Feldman R, Moser J, Feldman U. Linsholm: A New Respiratory Monitor. ASA Abstract. San Francisco, USA, 2013: A5032. Sathyamoorthy M, Lerman J, Feldman D, Feldman R, Moser J, Feldman U. Linsholm: A New Respiratory Monitor. ASA Abstract. San Francisco, USA, 2013: A5032.
35.
Zurück zum Zitat Lerman J, Feldman D, Feldman R, Moser J, Feldman L, Sathyamoorthy M, Deitch K, Feldman U. Linshom respiratory monitoring device: a novel temperature-based respiratory monitor. Can J Anaesth. 2016 Oct;63(10):1154–60.CrossRefPubMed Lerman J, Feldman D, Feldman R, Moser J, Feldman L, Sathyamoorthy M, Deitch K, Feldman U. Linshom respiratory monitoring device: a novel temperature-based respiratory monitor. Can J Anaesth. 2016 Oct;63(10):1154–60.CrossRefPubMed
36.
Zurück zum Zitat Standards for Basic Anesthetic Monitoring. American Society of Anesthesiologists Committee for Standards and Practice Parameters. Affirmed on 28 October 2015. Standards for Basic Anesthetic Monitoring. American Society of Anesthesiologists Committee for Standards and Practice Parameters. Affirmed on 28 October 2015.
37.
Zurück zum Zitat Practice guidelines for postanesthetic care. An updated report by the ASA Task Force on post anesthetic care. Anesthesiology. 2014;118(2):1–17. Practice guidelines for postanesthetic care. An updated report by the ASA Task Force on post anesthetic care. Anesthesiology. 2014;118(2):1–17.
38.
Zurück zum Zitat National Association of State EMS Officials Medical Directors Council. National Model EMS Clinical Guidelines, 23 Oct 2014. National Association of State EMS Officials Medical Directors Council. National Model EMS Clinical Guidelines, 23 Oct 2014.
39.
Zurück zum Zitat Blankush JM, Freeman R, McIlvane J, Tran T, Nassani S, Leitman IM. Implementation of a novel postoperative monitoring system using automated modified early warnign scores (MEWS) incorporating end-tital capnography. J Clin Monit Comput 2016. Epub ahead of print. Blankush JM, Freeman R, McIlvane J, Tran T, Nassani S, Leitman IM. Implementation of a novel postoperative monitoring system using automated modified early warnign scores (MEWS) incorporating end-tital capnography. J Clin Monit Comput 2016. Epub ahead of print.
40.
Zurück zum Zitat Helfand M, Christensen V, Anderson J. Technology Assessment: Early Sense for Monitroing vital Signs in Hospitalized Patients. VA Evidence-based Synethesis Program Evidence Briefs [Internet]. Washington (DC): Department fo Veterans Affairs (US); 2011. Helfand M, Christensen V, Anderson J. Technology Assessment: Early Sense for Monitroing vital Signs in Hospitalized Patients. VA Evidence-based Synethesis Program Evidence Briefs [Internet]. Washington (DC): Department fo Veterans Affairs (US); 2011.
41.
Zurück zum Zitat Tang K, Chiu J, Low E. Airway and ventilator equipment in field anesthesia: what’s new? Mil Med. 2004;169(5):342–8.CrossRefPubMed Tang K, Chiu J, Low E. Airway and ventilator equipment in field anesthesia: what’s new? Mil Med. 2004;169(5):342–8.CrossRefPubMed
Metadaten
Titel
Linshom thermodynamic sensor is a reliable alternative to capnography for monitoring respiratory rate
verfasst von
David Preiss
Benjamin A. Drew
James Gosnell
Bhavani S. Kodali
James H. Philip
Richard D. Urman
Publikationsdatum
22.02.2017
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2018
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-017-0004-4

Weitere Artikel der Ausgabe 1/2018

Journal of Clinical Monitoring and Computing 1/2018 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.