Local motion correction for lung tumours in PET/CT—first results

verfasst von: Ralph A. Bundschuh, Axel Martínez-Möller, Markus Essler, Stephan G. Nekolla, Sibylle I. Ziegler, Markus Schwaiger

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 11/2008

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Respiratory motion of lung lesions is a limiting factor of quantification of positron emission tomography (PET) data. As some important applications of PET such as therapy monitoring and radiation therapy treatment planning require precise quantification, it is necessary to correct PET data for motion artefacts.

Methods

The method is based on list-mode data. First, the motion of the lesion was detected by a centre of mass approach. In the second step, data were sorted corresponding to the breathing state. A volume of interest (VOI) around the lesion was defined manually, and the motion of the lesion in this VOI was measured with reference to the end-expiration image. Then, all voxels in the VOI were shifted according to the measured lesion motion. After optimisation of parameters and verification of the method using a computer-controlled motion phantom, it was applied to nine patients with solitary lesions of the lung.

Results

Fifty percent difference in measured lesion volume and 26% in mean activity concentration were found comparing PET data before and after applying the correction algorithm when simulating a motion amplitude of 28 mm in phantom studies. For patients, maximum changes of 27% in volume and 13% in mean standardised uptake values (SUV) were found.

Conclusion

As respiratory motion is affecting quantification of PET images, correction algorithms are essential for applications that require precise quantification. We described a method which improves the quantification of moving lesions by a local motion correction using list-mode data without increasing acquisition time or reduced signal-to-noise ratio of the images.
Literatur
1.
Zurück zum Zitat Bradley JD, Perez CA, Dehdashti F, Siegel BA. Implementing biologic target volumes in radiation treatment planning for non-small-cell lung cancer. J Nucl Med 2004;45:96S–101S.PubMed Bradley JD, Perez CA, Dehdashti F, Siegel BA. Implementing biologic target volumes in radiation treatment planning for non-small-cell lung cancer. J Nucl Med 2004;45:96S–101S.PubMed
2.
Zurück zum Zitat Baardwijk Av, Baumert BG, Bosmans G, Kroonenburgh Mv, Stroobants S, Gregoire V, et al. The current status of FDG-PET in tumour volume definition in radiotherapy treatment planning. Cancer Treat Rev 2006;32:245–60.PubMedCrossRef Baardwijk Av, Baumert BG, Bosmans G, Kroonenburgh Mv, Stroobants S, Gregoire V, et al. The current status of FDG-PET in tumour volume definition in radiotherapy treatment planning. Cancer Treat Rev 2006;32:245–60.PubMedCrossRef
3.
Zurück zum Zitat Hoekstra CJ, Hoekstra OS, Stroobants SG, Vansteenkiste J, Nuyts J, Smit EF, et al. Methods to monitor response to chemotherapy in non-small-cell lung cancer with 18F-FDG PET. J Nucl Med 2002;43:1304–09.PubMed Hoekstra CJ, Hoekstra OS, Stroobants SG, Vansteenkiste J, Nuyts J, Smit EF, et al. Methods to monitor response to chemotherapy in non-small-cell lung cancer with 18F-FDG PET. J Nucl Med 2002;43:1304–09.PubMed
4.
Zurück zum Zitat Schwarz JD, Bader M, Jenicke L, Hemminger G, Jänicke F, Avril N. Early prediction of response to chemotherapy in metastatic breast cancer using sequential 18F-FDG PET. J Nucl Med 2005;46:1144–50. Schwarz JD, Bader M, Jenicke L, Hemminger G, Jänicke F, Avril N. Early prediction of response to chemotherapy in metastatic breast cancer using sequential 18F-FDG PET. J Nucl Med 2005;46:1144–50.
5.
Zurück zum Zitat Ott K, Weber WA, Lordick F, Becker K, Busch R, Herrmann K, et al. Metabolic imaging predicts response, survival, and recurrence in adenocarcinomas of the esophagogastric junction. J Clin Oncol 2006;24:4692–8.PubMedCrossRef Ott K, Weber WA, Lordick F, Becker K, Busch R, Herrmann K, et al. Metabolic imaging predicts response, survival, and recurrence in adenocarcinomas of the esophagogastric junction. J Clin Oncol 2006;24:4692–8.PubMedCrossRef
6.
Zurück zum Zitat Nestle U, Kremp S, Schaefer-Schuler A, Sebastian-Welsch C, Hellwig D, Rübe C, et al. Comparison of different methods for delineation of 18F-FDG PET-positive tissue for target volume definition in radiotherapy of patients with non-small cell lung cancer. J Nucl Med 2005;46:1342–48.PubMed Nestle U, Kremp S, Schaefer-Schuler A, Sebastian-Welsch C, Hellwig D, Rübe C, et al. Comparison of different methods for delineation of 18F-FDG PET-positive tissue for target volume definition in radiotherapy of patients with non-small cell lung cancer. J Nucl Med 2005;46:1342–48.PubMed
7.
Zurück zum Zitat Goerres GW, Kamel E, Seifert B, Burger C, Buck A, Hany TF, et al. Accuracy of image coregistration of pulmonary lesions in patients with non-small cell lung cancer using an integrated PET/CT system. J Nucl Med 2002;43:1469–75.PubMed Goerres GW, Kamel E, Seifert B, Burger C, Buck A, Hany TF, et al. Accuracy of image coregistration of pulmonary lesions in patients with non-small cell lung cancer using an integrated PET/CT system. J Nucl Med 2002;43:1469–75.PubMed
8.
Zurück zum Zitat Ashamalla H, Rafla S, Parikh K, Mokhtar B, Goswami G, Kambam S, et al. The contribution of integrated PET/CT to the evolving definition of treatment volumes in radiation treatment planning in lung cancer. Int J Radiat Oncol Biol Phys 2005;63:1016–23.PubMed Ashamalla H, Rafla S, Parikh K, Mokhtar B, Goswami G, Kambam S, et al. The contribution of integrated PET/CT to the evolving definition of treatment volumes in radiation treatment planning in lung cancer. Int J Radiat Oncol Biol Phys 2005;63:1016–23.PubMed
9.
Zurück zum Zitat Plathow C, Ley S, Fink C, Puderbach M, Hosch W, Schmähl A, et al. Analysis of intrathoracic tumor mobility during whole breathing cycle by dynamic MRI. Int J Radiat Oncol Biol Phys 2004;59:952–59.PubMedCrossRef Plathow C, Ley S, Fink C, Puderbach M, Hosch W, Schmähl A, et al. Analysis of intrathoracic tumor mobility during whole breathing cycle by dynamic MRI. Int J Radiat Oncol Biol Phys 2004;59:952–59.PubMedCrossRef
10.
Zurück zum Zitat Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Schoder H, Larson SM, et al. Effects of respiratory gating on quantifying PET images of lung cancer. J Nucl Med 2002;43:876–81.PubMed Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Schoder H, Larson SM, et al. Effects of respiratory gating on quantifying PET images of lung cancer. J Nucl Med 2002;43:876–81.PubMed
11.
Zurück zum Zitat Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Squire OD, Braban LE, et al. Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer. Med Phys 2002;29:366–71.PubMedCrossRef Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Squire OD, Braban LE, et al. Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer. Med Phys 2002;29:366–71.PubMedCrossRef
12.
Zurück zum Zitat Visvikis D, Barret O, Fryer TD, Lamare F, Turzo A, Biazis Y, et al. Evaluation of respiratory motion effects in comparison with other parameters affecting PET image quality. IEEE Nuclear Science Symposium; 2004. Visvikis D, Barret O, Fryer TD, Lamare F, Turzo A, Biazis Y, et al. Evaluation of respiratory motion effects in comparison with other parameters affecting PET image quality. IEEE Nuclear Science Symposium; 2004.
13.
Zurück zum Zitat Martinez-Moller A, Zikic D, Botnar RM, Bundschuh RA, Howe W, Ziegler SI, et al. Dual cardiac-respiratory gated PET: implementation and results from a feasibility study. Eur J Nucl Med Mol Imaging 2007;34:1447–54.PubMedCrossRef Martinez-Moller A, Zikic D, Botnar RM, Bundschuh RA, Howe W, Ziegler SI, et al. Dual cardiac-respiratory gated PET: implementation and results from a feasibility study. Eur J Nucl Med Mol Imaging 2007;34:1447–54.PubMedCrossRef
14.
Zurück zum Zitat Detorie NC, Kesner AL, Solberg TD, Dahlbom M. Evaluation of image noise in respiratory gated PET. IEEE Trans Nucl Sci 2007;54:66–70.CrossRef Detorie NC, Kesner AL, Solberg TD, Dahlbom M. Evaluation of image noise in respiratory gated PET. IEEE Trans Nucl Sci 2007;54:66–70.CrossRef
15.
Zurück zum Zitat Hutton BF, Kyme AZ, Lau YH, Skerrett DW, Fulton RR. Hybrid 3-D reconstruction/registration algorithm for correction of head motion in emission tomography. IEEE Trans Nucl Sci 2002;49:188–94.CrossRef Hutton BF, Kyme AZ, Lau YH, Skerrett DW, Fulton RR. Hybrid 3-D reconstruction/registration algorithm for correction of head motion in emission tomography. IEEE Trans Nucl Sci 2002;49:188–94.CrossRef
16.
Zurück zum Zitat Livieratos L, Stegger L, Bloomfield PM, Schafers K, Bailey DL, Camici PG. Rigid-body transformation of list-mode projection data for respiratory motion correction in cardiac PET. Phys Med Biol 2005;50:3313–22.PubMedCrossRef Livieratos L, Stegger L, Bloomfield PM, Schafers K, Bailey DL, Camici PG. Rigid-body transformation of list-mode projection data for respiratory motion correction in cardiac PET. Phys Med Biol 2005;50:3313–22.PubMedCrossRef
17.
Zurück zum Zitat Schäfers KP, Dawood M, Lang N, Büther F, Schäfers M, Schober O. Motion correction in PET/CT. Nuklearmedizin 2005;5a:S46–S50. Schäfers KP, Dawood M, Lang N, Büther F, Schäfers M, Schober O. Motion correction in PET/CT. Nuklearmedizin 2005;5a:S46–S50.
18.
Zurück zum Zitat Lamare F, Cresson T, Savean J, Cheze Le Rest C, Reader AJ, Visvikis D. Respiratory motion correction for PET oncology applications using affine transformation of list mode data. Phys Med Biol 2007;52:121–40.PubMedCrossRef Lamare F, Cresson T, Savean J, Cheze Le Rest C, Reader AJ, Visvikis D. Respiratory motion correction for PET oncology applications using affine transformation of list mode data. Phys Med Biol 2007;52:121–40.PubMedCrossRef
19.
Zurück zum Zitat Martinez MJ, Bercier Y, Schwaiger M, Ziegler SI. PET/CT biograph sensation 16: performance improvement using faster electronics. Nuklearmedizin 2006;3:126–33. Martinez MJ, Bercier Y, Schwaiger M, Ziegler SI. PET/CT biograph sensation 16: performance improvement using faster electronics. Nuklearmedizin 2006;3:126–33.
20.
Zurück zum Zitat Bundschuh RA, Martinez-Moeller A, Essler M, Martinez MJ, Nekolla SG, Ziegler SI, et al. Postacquisition detection of tumor motion in the lung and upper abdomen using list-mode PET data: a feasibility study. J Nucl Med 2007;48:758–63.PubMedCrossRef Bundschuh RA, Martinez-Moeller A, Essler M, Martinez MJ, Nekolla SG, Ziegler SI, et al. Postacquisition detection of tumor motion in the lung and upper abdomen using list-mode PET data: a feasibility study. J Nucl Med 2007;48:758–63.PubMedCrossRef
21.
Zurück zum Zitat Erdi YE, Mawlawi O, Larson SM, Imbriaco M, Yeung H, Finn R, et al. Segmentation of lung lesion volume by adaptive positron emission tomography image thresholding. Cancer 1997;80:2505–9.PubMedCrossRef Erdi YE, Mawlawi O, Larson SM, Imbriaco M, Yeung H, Finn R, et al. Segmentation of lung lesion volume by adaptive positron emission tomography image thresholding. Cancer 1997;80:2505–9.PubMedCrossRef
22.
Zurück zum Zitat Martinez-Möller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions. J Nucl Med 2007;48:188–93.PubMed Martinez-Möller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions. J Nucl Med 2007;48:188–93.PubMed
Metadaten
Titel
Local motion correction for lung tumours in PET/CT—first results
verfasst von
Ralph A. Bundschuh
Axel Martínez-Möller
Markus Essler
Stephan G. Nekolla
Sibylle I. Ziegler
Markus Schwaiger
Publikationsdatum
01.11.2008
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 11/2008
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-008-0868-0

Weitere Artikel der Ausgabe 11/2008

European Journal of Nuclear Medicine and Molecular Imaging 11/2008 Zur Ausgabe