Long-term longevity and clinical outcomes of Linox S/SD implantable cardioverter-defibrillator leads: a single-center experience

verfasst von: Mohammad Alasti, Colin Machado, Sam Mirzaee, Stewart Healy, Logan Bittinger, David Adam, Emily Kotschet, Jack Krafchek, Jeffrey Alison

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 1/2021

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Since the introduction of the Biotronik Linox S/SD leads in 2006, there have been multiple reports of premature lead failure. The purpose of this study was to investigate the longevity of the Linox S/SD leads and to identify the possible predictors of lead failure in a single tertiary implant center.

Methods

We retrospectively reviewed patients who underwent implantation of Linox S/SD leads or Sorin Vigila 1CR/2CR leads (the same Linox S/SD leads marketed by Sorin) at our center. The cumulative lead survival was estimated using the Kaplan-Meier curve, and variables associated with lead failure were assessed by Cox proportional hazard model.

Results

A total of 187 patients (154 (82%) male) underwent Linox S/SD or Vigila 1CR/2CR implantation between 2007 and 2013. During follow-up with a median time of 75 months, nine lead failures were identified (4.8%). The mean and median times from lead implantation to lead failure were 70.7 ± 21 months and 64 (45–111) months, respectively. The cumulative survival probability for the Linox S/SD at 5 years was 97.1% and at 12 years was 90.3%. Non-physiological high-rate sensing was the most common type of lead failure in patients. In two-thirds of these patients, this led to inappropriate shock. We did not find any significant relationships between patients’ clinical and procedural characteristics and lead failure.

Conclusions

At our center, the 5-year lead survival of the Linox S/SD has been better than reports from other centers. The majority of lead failures presented as non-physiological high-rate sensing with subsequent inappropriate therapy.
Literatur
1.
Zurück zum Zitat Al-Khatib SM, Stevenson WG, Ackerman MJ, Bryant WJ, Callans DJ, Curtis AB, et al. 2017 AHA/ACC/HRS Guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2018;138:e272–391.PubMed Al-Khatib SM, Stevenson WG, Ackerman MJ, Bryant WJ, Callans DJ, Curtis AB, et al. 2017 AHA/ACC/HRS Guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2018;138:e272–391.PubMed
2.
Zurück zum Zitat De Maria E, Borghi A, Bonetti L, Fontana PL, Cappelli S. Externalized conductors and insulation failure in Biotronik defibrillator leads: history repeating or a false alarm? World J Clin Cases. 2017;5(2):27–34.CrossRef De Maria E, Borghi A, Bonetti L, Fontana PL, Cappelli S. Externalized conductors and insulation failure in Biotronik defibrillator leads: history repeating or a false alarm? World J Clin Cases. 2017;5(2):27–34.CrossRef
3.
Zurück zum Zitat Lam A, Buehler S, Goulouti E, Sweda R, Haeberlin A, Medeiros-Domingo A, et al. Comparison of lead failure manifestation of Biotronik Linox with St. Jude Medical Riata and Medtronic Sprint Fidelis lead. J Interv Card Electrophysiol. 2019;54:161–70.CrossRef Lam A, Buehler S, Goulouti E, Sweda R, Haeberlin A, Medeiros-Domingo A, et al. Comparison of lead failure manifestation of Biotronik Linox with St. Jude Medical Riata and Medtronic Sprint Fidelis lead. J Interv Card Electrophysiol. 2019;54:161–70.CrossRef
5.
Zurück zum Zitat Noti F, Lam A, Klossner N, Seiler J, Servatius H, Medeiros-Domingo A, et al. Failure rate and conductor externalization in the Biotronik Linox/Sorin Vigila implantable cardioverter-defibrillator lead. Heart Rhythm. 2016;13:1075–82.CrossRef Noti F, Lam A, Klossner N, Seiler J, Servatius H, Medeiros-Domingo A, et al. Failure rate and conductor externalization in the Biotronik Linox/Sorin Vigila implantable cardioverter-defibrillator lead. Heart Rhythm. 2016;13:1075–82.CrossRef
6.
Zurück zum Zitat Padfield GJ, Steinberg C, Karim SS, Tung S, Bennett MT, Le Maitre JP, et al. Early failure of the Biotronik Linox implantable cardioverter-defibrillator lead. J Cardiovasc Electrophysiol. 2015;26:274–81.CrossRef Padfield GJ, Steinberg C, Karim SS, Tung S, Bennett MT, Le Maitre JP, et al. Early failure of the Biotronik Linox implantable cardioverter-defibrillator lead. J Cardiovasc Electrophysiol. 2015;26:274–81.CrossRef
7.
Zurück zum Zitat Howe AJ, McKeag NA, Wilson CM, Ashfield KP, Roberts MJ. Insulation failure of the Linox defibrillator lead: a case report and retrospective review of a single center experience. J Cardiovasc Electrophysiol. 2015;26:686–9.CrossRef Howe AJ, McKeag NA, Wilson CM, Ashfield KP, Roberts MJ. Insulation failure of the Linox defibrillator lead: a case report and retrospective review of a single center experience. J Cardiovasc Electrophysiol. 2015;26:686–9.CrossRef
8.
Zurück zum Zitat van Malderen SCH, Szili-Torok T, Yap SC, Hoeks SE, Zijlstra F, Theuns DAMJ. Comparative study of the failure rates among 3 implantable defibrillator leads. Heart Rhythm. 2016;13:2299–305.CrossRef van Malderen SCH, Szili-Torok T, Yap SC, Hoeks SE, Zijlstra F, Theuns DAMJ. Comparative study of the failure rates among 3 implantable defibrillator leads. Heart Rhythm. 2016;13:2299–305.CrossRef
9.
Zurück zum Zitat Manfredi JA, Smithgall SM, Kircher CM, Lollis MA. Insulation failure with externalized conductor of a Linox SD lead: a case report. J Cardiovasc Electrophysiol. 2014;25:440–1.CrossRef Manfredi JA, Smithgall SM, Kircher CM, Lollis MA. Insulation failure with externalized conductor of a Linox SD lead: a case report. J Cardiovasc Electrophysiol. 2014;25:440–1.CrossRef
10.
Zurück zum Zitat Kawada S, Nishii N, Morimoto Y, Miyoshi A, Tachibana M, Sugiyama H, et al. Comparison of longevity and clinical outcomes of implantable cardioverter-defibrillator leads among manufactures. Heart Rhythm. 2017;14:1496–503.CrossRef Kawada S, Nishii N, Morimoto Y, Miyoshi A, Tachibana M, Sugiyama H, et al. Comparison of longevity and clinical outcomes of implantable cardioverter-defibrillator leads among manufactures. Heart Rhythm. 2017;14:1496–503.CrossRef
11.
Zurück zum Zitat O’Connor M, Hooks D, Webber M, Shi B, Morrison S, Harding S, et al. Long-term single-center comparison of ICD lead survival: evidence for premature Linox lead failure. J Cardiovasc Electrophysiol. 2018;29(7):1024–31.CrossRef O’Connor M, Hooks D, Webber M, Shi B, Morrison S, Harding S, et al. Long-term single-center comparison of ICD lead survival: evidence for premature Linox lead failure. J Cardiovasc Electrophysiol. 2018;29(7):1024–31.CrossRef
12.
Zurück zum Zitat Marai I, Milman A, Nof E, Gurevitz O, Barlev D, Lipchenca I, et al. Performance of the Linox implantable cardioverter defibrillator leads: a single-center experience. Pacing Clin Electrophysiol. 2019;42(12):1524–8.CrossRef Marai I, Milman A, Nof E, Gurevitz O, Barlev D, Lipchenca I, et al. Performance of the Linox implantable cardioverter defibrillator leads: a single-center experience. Pacing Clin Electrophysiol. 2019;42(12):1524–8.CrossRef
13.
Zurück zum Zitat McKeag NA, Noad RL, Ashfield K, Wilson CM, McEneaney DJ, Roberts MJD. Prospective assessment of Linox implantable cardioverter defibrillator leads for structural or electrical abnormalities. Adv Ther. 2018;35(5):666–70.CrossRef McKeag NA, Noad RL, Ashfield K, Wilson CM, McEneaney DJ, Roberts MJD. Prospective assessment of Linox implantable cardioverter defibrillator leads for structural or electrical abnormalities. Adv Ther. 2018;35(5):666–70.CrossRef
14.
Zurück zum Zitat Good ED, Cakulev I, Orlov MV, Hirsh D, Simeles J, Mohr K, et al. Long-term evaluation of Biotronik Linox and Linox (smart) implantable cardioverter defibrillator leads. J Cardiovasc Electrophysiol. 2016;27(6):735–42.CrossRef Good ED, Cakulev I, Orlov MV, Hirsh D, Simeles J, Mohr K, et al. Long-term evaluation of Biotronik Linox and Linox (smart) implantable cardioverter defibrillator leads. J Cardiovasc Electrophysiol. 2016;27(6):735–42.CrossRef
15.
Zurück zum Zitat Díez DP, Rubín JM, Cuervo DC, Iglesias DG, De La Tassa CM. Analysis of early failure of Biotronik Linox Smart implantable cardioverter-defibrillator leads: a comparative study of three defibrillator leads. Pacing Clin Electrophysiol. 2018;41(9):1165–70.CrossRef Díez DP, Rubín JM, Cuervo DC, Iglesias DG, De La Tassa CM. Analysis of early failure of Biotronik Linox Smart implantable cardioverter-defibrillator leads: a comparative study of three defibrillator leads. Pacing Clin Electrophysiol. 2018;41(9):1165–70.CrossRef
16.
Zurück zum Zitat Guédon-Moreau L, Lacroix D, Sadoul N, Clémenty J, Kouakam C, Hermida JS, et al. A randomized study of remote follow-up of implantable cardioverter defibrillators: safety and efficacy report of the ECOST trial. Eur Heart J. 2013;34(8):605–14.CrossRef Guédon-Moreau L, Lacroix D, Sadoul N, Clémenty J, Kouakam C, Hermida JS, et al. A randomized study of remote follow-up of implantable cardioverter defibrillators: safety and efficacy report of the ECOST trial. Eur Heart J. 2013;34(8):605–14.CrossRef
17.
Zurück zum Zitat Varma N, Michalski J, Epstein AE, Schweikert E. Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial. Circ Arrhythm Electrophysiol. 2010;3(5):428–36.CrossRef Varma N, Michalski J, Epstein AE, Schweikert E. Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial. Circ Arrhythm Electrophysiol. 2010;3(5):428–36.CrossRef
18.
Zurück zum Zitat Varma N. Remote monitoring for advisories: automatic early detection of silent lead failure. PACE. 2009;32(4):525–7.CrossRef Varma N. Remote monitoring for advisories: automatic early detection of silent lead failure. PACE. 2009;32(4):525–7.CrossRef
Metadaten
Titel
Long-term longevity and clinical outcomes of Linox S/SD implantable cardioverter-defibrillator leads: a single-center experience
verfasst von
Mohammad Alasti
Colin Machado
Sam Mirzaee
Stewart Healy
Logan Bittinger
David Adam
Emily Kotschet
Jack Krafchek
Jeffrey Alison
Publikationsdatum
02.06.2020
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 1/2021
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-020-00787-x

Weitere Artikel der Ausgabe 1/2021

Journal of Interventional Cardiac Electrophysiology 1/2021 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.