01.12.2014 | MULTIMEDIA REPORT | Ausgabe 3/2014

Journal of Interventional Cardiac Electrophysiology 3/2014

Longitudinal follow-up of Riata leads reveals high annual incidence of new conductor externalization and electrical failure

Zeitschrift:
Journal of Interventional Cardiac Electrophysiology > Ausgabe 3/2014
Autoren:
Christian Steinberg, Jean-François Sarrazin, François Philippon, Jean Champagne, Marc-André Bouchard, Franck Molin, Isabelle Nault, Louis Blier, Gilles O’Hara
Wichtige Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1007/​s10840-014-9951-6) contains supplementary material, which is available to authorized users.

Abstract

Background and purpose

RiataTM defibrillation leads are susceptible to conductor externalization. The point prevalence of insulation defect in RiataTM leads is up to 33 %, but prospective data concerning incidence of new lead abnormalities are lacking. The purpose of our study was to determine the annual incidence of new conductor externalizations and electrical lead failure.

Methods

A prospective observational study was conducted at a single tertiary center. One hundred forty-one patients were followed over 12 months. A posterior-anterior (PA)/lateral chest x-ray (CXR) with zooming was performed at baseline and at 12 months to screen for conductor externalization. Electrical abnormalities and clinical outcome were also assessed.

Results

The overall incidence of new insulation defects was 8.5 % at 12 months. High-risk leads for new conductor externalization were lead models 1580, 1582, and 1590 with an annual rate of 11.9, 11.1, and 10 %, respectively. New conductor externalizations were three times more common in 8 Fr leads compared to 7 Fr leads. The overall incidence of new electrical dysfunction was 6.4 % at 12 months. Electrical dysfunction was significantly higher in abnormal leads (25 % [3/12], 4.7 % [6/129]; p = 0.03) and mostly driven by high ventricular pacing thresholds. There was no difference in inappropriate shock or failure of high-voltage therapy.

Conclusion

The annual incidence of new insulation defects in RiataTM leads is much higher than previously reported. Lead models 1580, 1582, and 1590 are at highest risk for new conductor externalization. Electrical dysfunction in RiataTM leads is also much higher than reported and is associated with conductor externalization.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Zusatzmaterial
ESM 1 (DOCX 38 kb)
10840_2014_9951_MOESM1_ESM.docx
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 3/2014

Journal of Interventional Cardiac Electrophysiology 3/2014 Zur Ausgabe
  1. Das kostenlose Testabonnement läuft nach 14 Tagen automatisch und formlos aus. Dieses Abonnement kann nur einmal getestet werden.

Neu im Fachgebiet Kardiologie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Kardiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise