Measurement and scan reproducibility of parameters of intravoxel incoherent motion in renal tumor and normal renal parenchyma: a preliminary research at 3.0 T MR

Zeitschrift:
Abdominal Radiology > Ausgabe 7/2018
Autoren:
Jingjing Pan, Hongtao Zhang, Fengyuan Man, Yanguang Shen, Yingwei Wang, Yan Zhong, Lu Ma, Haiyi Wang, Huiyi Ye

Abstract

Purpose

To prospectively estimate measurement and scan reproducibility of parameters of intravoxel incoherent motion (IVIM) in renal tumors, normal renal cortex, and medulla.

Methods

Twenty-four consecutive patients (twelve males and twelve females; median age 56.7 years, range 32–71 years) with 25 renal tumors (20 renal cell carcinomas, one urothelium carcinoma, three angiomyolipomas, and one oncocytoma) were examined twice using IVIM1 and IVIM2 with 9 and 16 b values, respectively, at 3.0 T. All the patients were re-scanned in 24–48 h. Regions of interest (ROIs) were placed in solid part of tumor, normal cortex, and medulla to derive IVIM parameters D (true diffusion coefficient), D* (pseudodiffusion coefficient), and f (perfusion fraction of pseudodiffusion). Differences in parameters between two IVIM sets and intra-observer, inter-observer, and scan–rescan differences were assessed using paired t tests. Intra-observer, inter-observer, and scan–rescan reproducibility were assessed by measuring coefficient of variation and Bland–Altman limits of agreements.

Results

Intra-observer reproducibility of renal tumors, normal renal cortex, and medulla was excellent for apparent diffusion coefficient (ADC; CV: 3.45%–5.34%, BA-LA: −14% to 18%) and D (CV: 3.65% to 6.04%, BA-LA: −18% to 19%), good for f (CV: 11.96%–16.08%, BA-LA: −76.4% to 92.1% except f of medulla with CV of 32.59% and BA-LA of −76.4% to 92.1% in IVIM1), and poor for D* (CV: 25.0% to 75.4%, BA-LA: −111% to 150%). The same order was in inter-observer reproducibility analysis. Scan–rescan reproducibility was the worst of the three parameters. Renal medulla showed worse reproducibility than renal tumors and the normal cortex. The metrics of IVIM2 had better reproducibility than IVIM1.

Conclusion

Excellent reproducibility evaluation for ADC and D, good for f, and poor for D* derived from IVIM was performed in renal tumors, normal renal cortex, and medulla. D* has limited reliability and scan–rescan reproducibility should be improved.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 7/2018

Abdominal Radiology 7/2018 Zur Ausgabe

Classics in Abdominal Imaging

Beaver tail liver

Classics in Abdominal Imaging

Spinnaker sail sign

  1. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Radiologie

Meistgelesene Bücher aus der Radiologie

2016 | Buch

Medizinische Fremdkörper in der Bildgebung

Thorax, Abdomen, Gefäße und Kinder

Dieses einzigartige Buch enthält ca. 1.600 hochwertige radiologische Abbildungen und Fotos iatrogen eingebrachter Fremdmaterialien im Röntgenbild und CT.

Herausgeber:
Dr. med. Daniela Kildal

2011 | Buch

Atlas Klinische Neuroradiologie des Gehirns

Radiologie lebt von Bildern! Der vorliegende Atlas trägt dieser Tatsache Rechnung. Sie finden zu jedem Krankheitsbild des Gehirns Referenzbilder zum Abgleichen mit eigenen Befunden.

Autoren:
Priv.-Doz. Dr. med. Jennifer Linn, Prof. Dr. med. Martin Wiesmann, Prof. Dr. med. Hartmut Brückmann

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Radiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise