Mechanical testing of small fracture implants for comparison of insertion and failure torques

verfasst von: Craig R. Glauser, Z. Maria Oden, Catherine G. Ambrose, Mark B. Willits, Kevin J. Coupe

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 8/2003

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Small fracture screws are among the most commonly used implants in the field of orthopedic surgery. The goal of this study was to compare the insertion and failure torques of three screw types: cortical, partially threaded cancellous, and fully threaded cancellous from three manufacturers: Zimmer, Richards, and Synthes.

Materials and methods

Each type of screw was subjected to biomechanical tests to determine the insertion (n=6/group) and failure (n=10/group) torques.

Results

Two-factor ANOVA tests were run to determine whether the insertion or failure torques were different for the different screw types and manufacturers. In the case of insertion torques, neither the screw nor the manufacturer had any significant effect. In the case of failure torque, significant differences were found based on both the screw type and the manufacturer, with the cortical screws manufactured by Zimmer being the strongest. Although there were strength differences, the most important comparison clinically is between the failure torque and the insertion torque of each screw. In all cases, the failure torques were approximately 20 times larger than the insertion torques, and therefore no failures should occur if only torsional loads are applied during insertion. This comparison shows that factors other than screw strength and manufacturing processes may be involved in cases of screw failure during insertion.

Conclusion

All three screw types from all three manufacturers appear to be mechanically reliable, with the proper insertion to failure torque ratio. The surgeon's choice of implant should be based on other considerations.
Literatur
1.
Zurück zum Zitat Ansell RH, Scales JT (1968) A study of some factors which affect the strength of screws and their insertion and holding power in bone. J Biomech 1:279–302 Ansell RH, Scales JT (1968) A study of some factors which affect the strength of screws and their insertion and holding power in bone. J Biomech 1:279–302
2.
Zurück zum Zitat ASTM F138–92 (1995) Standard specification for stainless steel bar and wire for surgical implants (special quality). Annual Book of ASTM Standards, Philadelphia, Vol 13.01 ASTM F138–92 (1995) Standard specification for stainless steel bar and wire for surgical implants (special quality). Annual Book of ASTM Standards, Philadelphia, Vol 13.01
3.
Zurück zum Zitat Collinge CA, Cordes SD, Lautenschlager, E, Stern SH (2000) The mechanical properties of small fragment orthopedic screws. Clin Orthop 373:277–284PubMed Collinge CA, Cordes SD, Lautenschlager, E, Stern SH (2000) The mechanical properties of small fragment orthopedic screws. Clin Orthop 373:277–284PubMed
4.
Zurück zum Zitat Gerlach KL, Grobel KH, Heidemann W, Kollner HG (1998) Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws. J Craniomaxillofac Surg 26:50–55PubMed Gerlach KL, Grobel KH, Heidemann W, Kollner HG (1998) Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws. J Craniomaxillofac Surg 26:50–55PubMed
5.
Zurück zum Zitat Hutton WC, York MJ (1996) Treatment of screw hole defects using bone graft materials: a histologic and biomechanic study. J South Orthop Assoc 5:5–12PubMed Hutton WC, York MJ (1996) Treatment of screw hole defects using bone graft materials: a histologic and biomechanic study. J South Orthop Assoc 5:5–12PubMed
6.
Zurück zum Zitat ISO 5832–1 (1987) Implants for surgery, metallic materials: wrought stainless steel. International Organization for Standardization ISO 5832–1 (1987) Implants for surgery, metallic materials: wrought stainless steel. International Organization for Standardization
7.
Zurück zum Zitat Smith & Nephew (1998) TC-100 screw and plating system: a technical summary. Smith & Nephew (1998) TC-100 screw and plating system: a technical summary.
8.
Zurück zum Zitat Trader JE, Johnson RP, Kalbfleisch JH (1979) Bone-mineral content, surface hardness and mechanical fixation in the human radius. J Bone Joint Surg Am 61:1217–1220PubMed Trader JE, Johnson RP, Kalbfleisch JH (1979) Bone-mineral content, surface hardness and mechanical fixation in the human radius. J Bone Joint Surg Am 61:1217–1220PubMed
Metadaten
Titel
Mechanical testing of small fracture implants for comparison of insertion and failure torques
verfasst von
Craig R. Glauser
Z. Maria Oden
Catherine G. Ambrose
Mark B. Willits
Kevin J. Coupe
Publikationsdatum
01.10.2003
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 8/2003
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-003-0581-1

Weitere Artikel der Ausgabe 8/2003

Archives of Orthopaedic and Trauma Surgery 8/2003 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.