Metal implants on CT: comparison of iterative reconstruction algorithms for reduction of metal artifacts with single energy and spectral CT scanning in a phantom model

verfasst von: Jieming Fang, Da Zhang, Carol Wilcox, Benedikt Heidinger, Vassilios Raptopoulos, Alexander Brook, Olga R. Brook

Erschienen in: Abdominal Radiology | Ausgabe 3/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To assess single energy metal artifact reduction (SEMAR) and spectral energy metal artifact reduction (MARS) algorithms in reducing artifacts generated by different metal implants.

Materials and method

Phantom was scanned with and without SEMAR (Aquilion One, Toshiba) and MARS (Discovery CT750 HD, GE), with various metal implants. Images were evaluated objectively by measuring standard deviation in regions of interests and subjectively by two independent reviewers grading on a scale of 0 (no artifact) to 4 (severe artifact). Reviewers also graded new artifacts introduced by metal artifact reduction algorithms.

Results

SEMAR and MARS significantly decreased variability of the density measurement adjacent to the metal implant, with median SD (standard deviation of density measurement) of 52.1 HU without SEMAR, vs. 12.3 HU with SEMAR, p < 0.001. Median SD without MARS of 63.1 HU decreased to 25.9 HU with MARS, p < 0.001. Median SD with SEMAR is significantly lower than median SD with MARS (p = 0.0011). SEMAR improved subjective image quality with reduction in overall artifacts grading from 3.2 ± 0.7 to 1.4 ± 0.9, p < 0.001. Improvement of overall image quality by MARS has not reached statistical significance (3.2 ± 0.6 to 2.6 ± 0.8, p = 0.088). There was a significant introduction of artifacts introduced by metal artifact reduction algorithm for MARS with 2.4 ± 1.0, but minimal with SEMAR 0.4 ± 0.7, p < 0.001.

Conclusion

CT iterative reconstruction algorithms with single and spectral energy are both effective in reduction of metal artifacts. Single energy-based algorithm provides better overall image quality than spectral CT-based algorithm. Spectral metal artifact reduction algorithm introduces mild to moderate artifacts in the far field.
Literatur
1.
Zurück zum Zitat Sofue, K, Yoshikawa, T, Negi, N, et al. Abdominal CT with single-energy metal artifact reduction (SEMAR): initial experiences. Poster session presented at: European Congress of Radiology, 2014 March 6–10, Vienna, Austria Sofue, K, Yoshikawa, T, Negi, N, et al. Abdominal CT with single-energy metal artifact reduction (SEMAR): initial experiences. Poster session presented at: European Congress of Radiology, 2014 March 6–10, Vienna, Austria
2.
Zurück zum Zitat Morsbach F, Bickelhaupt S, Wanner GA, et al. (2013) Reduction of metal artifacts from hip prostheses on CT images of the pelvis: value of iterative reconstructions. Radiology 268(1):237–244CrossRefPubMed Morsbach F, Bickelhaupt S, Wanner GA, et al. (2013) Reduction of metal artifacts from hip prostheses on CT images of the pelvis: value of iterative reconstructions. Radiology 268(1):237–244CrossRefPubMed
3.
Zurück zum Zitat Brook OR, Gourtsoyianni S, Brook A, et al. (2012) Spectral CT with metal artifacts reduction software for improvement of tumor visibility in the vicinity of gold fiducial markers. Radiology 263(3):696–705CrossRefPubMed Brook OR, Gourtsoyianni S, Brook A, et al. (2012) Spectral CT with metal artifacts reduction software for improvement of tumor visibility in the vicinity of gold fiducial markers. Radiology 263(3):696–705CrossRefPubMed
4.
Zurück zum Zitat Yang Q, Peng S, Wu J, et al. (2015) Spectral CT with monochromatic imaging and metal artifacts reduction software for artifacts reduction of (125)I radioactive seeds in liver brachytherapy. Jpn J Radiol 33(11):694–705CrossRefPubMed Yang Q, Peng S, Wu J, et al. (2015) Spectral CT with monochromatic imaging and metal artifacts reduction software for artifacts reduction of (125)I radioactive seeds in liver brachytherapy. Jpn J Radiol 33(11):694–705CrossRefPubMed
5.
Zurück zum Zitat Funama Y, Taguchi K, Utsunomiya D, et al. (2015) A newly-developed metal artifact reduction algorithm improves the visibility of oral cavity lesions on 320-MDCT volume scans. Phys Med 31(1):66–71CrossRefPubMed Funama Y, Taguchi K, Utsunomiya D, et al. (2015) A newly-developed metal artifact reduction algorithm improves the visibility of oral cavity lesions on 320-MDCT volume scans. Phys Med 31(1):66–71CrossRefPubMed
6.
Zurück zum Zitat Sonoda A, Nitta N, Ushio N, et al. (2015) Evaluation of the quality of CT images acquired with the single energy metal artifact reduction (SEMAR) algorithm in patients with hip and dental prostheses and aneurysm embolization coils. Jpn J Radiol 33(11):710–716CrossRefPubMed Sonoda A, Nitta N, Ushio N, et al. (2015) Evaluation of the quality of CT images acquired with the single energy metal artifact reduction (SEMAR) algorithm in patients with hip and dental prostheses and aneurysm embolization coils. Jpn J Radiol 33(11):710–716CrossRefPubMed
7.
Zurück zum Zitat Andersson KM, Nowik P, Persliden J, et al. (2015) Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors. Br J Radiol 88(1052):20140473CrossRefPubMedPubMedCentral Andersson KM, Nowik P, Persliden J, et al. (2015) Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors. Br J Radiol 88(1052):20140473CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24(6):1679–1691CrossRefPubMed Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24(6):1679–1691CrossRefPubMed
9.
Zurück zum Zitat Yu L, Li H, Mueller J, et al. (2009) Metal artifact reduction from reformatted projections for hip prostheses in multislice helical computed tomography: techniques and initial clinical results. Investig Radiol 44(11):691–696CrossRef Yu L, Li H, Mueller J, et al. (2009) Metal artifact reduction from reformatted projections for hip prostheses in multislice helical computed tomography: techniques and initial clinical results. Investig Radiol 44(11):691–696CrossRef
10.
Zurück zum Zitat Watzke O, Kalender WA (2004) A pragmatic approach to metal artifact reduction in CT: merging of metal artifact reduced images. Eur Radiol 14(5):849–856CrossRefPubMed Watzke O, Kalender WA (2004) A pragmatic approach to metal artifact reduction in CT: merging of metal artifact reduced images. Eur Radiol 14(5):849–856CrossRefPubMed
11.
Zurück zum Zitat Boas FE, Fleischmann D (2011) Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology 259(3):894–902CrossRefPubMed Boas FE, Fleischmann D (2011) Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology 259(3):894–902CrossRefPubMed
12.
Zurück zum Zitat Meyer E, Raupach R, Lell M, et al. (2010) Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys 37(10):5482–5493CrossRefPubMed Meyer E, Raupach R, Lell M, et al. (2010) Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys 37(10):5482–5493CrossRefPubMed
13.
Zurück zum Zitat Prell D, Kyriakou Y, Kachelrie M, et al. (2010) Reducing metal artifacts in computed tomography caused by hip endoprostheses using a physics-based approach. Investig Radiol 45(11):747–754CrossRef Prell D, Kyriakou Y, Kachelrie M, et al. (2010) Reducing metal artifacts in computed tomography caused by hip endoprostheses using a physics-based approach. Investig Radiol 45(11):747–754CrossRef
Metadaten
Titel
Metal implants on CT: comparison of iterative reconstruction algorithms for reduction of metal artifacts with single energy and spectral CT scanning in a phantom model
verfasst von
Jieming Fang
Da Zhang
Carol Wilcox
Benedikt Heidinger
Vassilios Raptopoulos
Alexander Brook
Olga R. Brook
Publikationsdatum
02.01.2017
Verlag
Springer US
Erschienen in
Abdominal Radiology / Ausgabe 3/2017
Print ISSN: 2366-004X
Elektronische ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-016-1023-1

Weitere Artikel der Ausgabe 3/2017

Abdominal Radiology 3/2017 Zur Ausgabe

Classics in Abdominal Imaging

Corkscrew esophagus

Classics in Abdominal Imaging

The northern exposure sign

Classics in Abdominal Imaging

The stepladder sign

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.