Skip to main content
Erschienen in: Journal of Clinical Monitoring and Computing 1/2021

Open Access 17.03.2020 | Review Paper

Metrology part 1: definition of quality criteria

verfasst von: Pierre Squara, Thomas W. L. Scheeren, Hollmann D. Aya, Jan Bakker, Maurizio Cecconi, Sharon Einav, Manu L. N. G. Malbrain, Xavier Monnet, Daniel A. Reuter, Iwan C. C. van der Horst, Bernd Saugel

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2021

Abstract

Any measurement is always afflicted with some degree of uncertainty. A correct understanding of the different types of uncertainty, their naming, and their definition is of crucial importance for an appropriate use of measuring instruments. However, in perioperative and intensive care medicine, the metrological requirements for measuring instruments are poorly defined and often used spuriously. The correct use of metrological terms is also of crucial importance in validation studies. The European Union published a new directive on medical devices, mentioning that in the case of devices with a measuring function, the notified body is involved in all aspects relating to the conformity of the device with the metrological requirements. It is therefore the task of the scientific societies to establish the standards in their area of expertise. Adopting the same understandings and definitions among clinicians and scientists is obviously the first step. In this metrologic review (part 1), we list and explain the most important terms defined by the International Bureau of Weights and Measures regarding quantities and units, properties of measurements, devices for measurement, properties of measuring devices, and measurement standards, with specific examples from perioperative and intensive care medicine.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
VIM
International Vocabulary of Metrology
σ
Standard deviation
σ2
Variance
SEM
Standard error of the mean
BIMP
International Bureau of Weights and Measures (Bureau International des Poids et Mesures)
SI
International standard
CO
Cardiac output
BP
Blood pressure

1 Introduction

“Metrology is the science of measurement, embracing both experimental and theoretical determinations at any level of uncertainty, in any field of science and technology” [1]. Metrology is therefore of key importance not only for engineers and scientists but also for any clinician using tools to measure, assess, or estimate physiological variables. Understanding metrological concepts and recognizing their limitations and constraints is a prerequisite for the interpretation of data obtained within clinical practice and research, especially research on the validation of new medical devices.
Unfortunately, there is still a lot of confusion regarding the exact definitions of metrological terms in many papers reporting the results of method comparison studies. For example, most papers claim to report the “accuracy” of a measuring instrument despite the fact that accuracy qualifies a single measurement and cannot qualify a measuring instrument.
In addition, the new directive on medical devices of the European Union [2] mentioned that the notified body needs to be involved in all aspects related to the conformity of the device with the metrological requirements. Therefore, physicians need to adopt the appropriate terms and definitions as a first step before determining a set of minimum metrologic requirements for any measuring instrument used in perioperative and intensive care medicine.
Consequently, physicians must rigorously share the same terms and definitions with other scientists. This is of particular importance in perioperative and intensive care medicine because clinical decision-making considers, or even completely relies on a variety of variables, measured using medical devices including advanced hemodynamic and respiratory monitoring.
The consensual metrological list of terms of the "International Vocabulary of Metrology (VIM)" is divided into five main headings: (1) quantities and units, (2) measurement, (3) devices for measurement, (4) properties of measuring devices, and (5) measurement standards (Etalons) [1]. The complete list can be found in a guidance document of the Joint Committee for Guides in Metrology [1].
In the present document (part 1), we describe and define terms used to qualify and quantify medical measurements to provide a framework for a common and standardized way of describing, reporting, and discussing measurements in perioperative and intensive care medicine.

2 Quantities and units

A quantity is a property of a phenomenon, body, or substance, to which a magnitude is attributed, that can be expressed as a number and a reference (a measurement unit, a measurement procedure, a reference material, or a combination of such). A quantity is, therefore, characterized by a dimension, a unit, and a value. There are seven base quantities from which all quantities of the international system (SI) are derived. They are listed in Table 1 together with five derived quantities frequently used in medicine. The complete list can be found in documents released by the intergovernmental organization International Bureau of Weight and Measures (BIPM) [1]. In assessing quantities, the VIM distinguishes facts (measurements) and methods (instruments).

3 Measurement

A measurement is a process of experimentally obtaining one or more values that can reasonably be attributed to a quantity [1]. Since the true value of a quantity is necessarily unknown, a measurement result is generally expressed as a value and a measurement uncertainty. The measurand is the quantity to be measured [1]. A measurement method is based on a measurement principle, i.e., a physical, chemical, or biological phenomenon serving as the basis of a measurement [1]. A reference measurement procedure is a measurement procedure that provides measurement results fitting for their intended use [1]. Although it has no international definition, a criterion standard (often referred to as gold standard) is supposed to be the best practically available reference method. A measurement error is the difference between a single measurement and a reference quantity value.
The uncertainty of a measurement is characterized by different components listed below and schematized in Fig. 1.

3.1 Measurement precision

The precision is the closeness of agreement between measured values obtained by replicate measurements on the same or similar quantities under specified and stable conditions [1, 3]. In other words, precision describes the variability of replicate measurements of a given quantity value, without reference to a true or reference value (Fig. 1). Precision is a quality and should not be expressed as a numerical value but is generally assessed by the random measurement error. The random measurement error can be expressed as a number by the standard deviation (σ) or variance (σ2) of the repeated measurements and assuming a mean random error of zero (Figs. 1, 2). The coefficient of variation (2σ/mean) can also be expressed as variability in %. The specified conditions of precision assessment may add variabilities of different kinds [3]. Repeatability is the precision under conditions that include the same measurement procedure, same operators, same measuring system, same operation conditions, same location, and replicate measurements on the same or similar objects over a short period of time [1]. Reproducibility is the precision under a set of conditions that include different locations, operators, measuring systems, and replicate measurements on the same or similar objects [1]. Between repeatability and reproducibility, intermediate precision is the precision under a set of intermediate conditions of a measurement (Fig. 2) [1].
Example
If the systolic blood pressure of a patient, measured by a pressure transducer connected to a radial arterial catheter, is constant for 20 min (showing a stable reference quantity value irrespective of its value), and if 20 consecutive oscillometric upper-arm cuff measurements fluctuate during the same time between 110 and 130 mmHg (σ = 12.5 mmHg), the oscillometric upper-arm cuff measurement can be described as being “precise” or “imprecise”, according to the intended use if this level of variability is considered as excessive or acceptable.

3.2 Measurement trueness

The trueness is the closeness of agreement between the average of an infinite number of replicate measured quantity values and the true or reference quantity value of the measurand [1]. Trueness is a quality and cannot be expressed as a numerical value but is generally assessed by the systematic measurement error [1]. Since the mean random error of an infinite number of replicates is zero, the difference between the averaged measured value and the reference value (also called measurement bias) is, therefore, an estimate of the systematic measurement error (Fig. 1). Consequently, a measurement with a small systematic measurement error is considered to be true [1]. A correction can be applied to compensate for a known systematic error.
Example
If the radial arterial catheter-derived systolic blood pressure of a patient is constantly measured at 120 mmHg over 20 min (stable reference quantity value) and if averaging 20 consecutive oscillometric upper-arm cuff measurements during the same time give a mean value of 120 mmHg, the systematic measurement error (measurement bias) is therefore 0. The oscillometric measurement mean value can be described as being true, irrespective of its variability.

3.3 Measurement accuracy

The measurement accuracy is the closeness of agreement between a single measured value and a true or reference value of the measurand [1]. Accuracy is a quality and cannot be expressed as a numerical value but is generally assessed by a measurement error (Fig. 1) [1]. A measurement with a small measurement error is considered accurate [1]. A measurement error can, therefore, be the result of a random measurement error (σ; qualifying the imprecision), a systematic measurement error (bias, qualifying the untrueness), or both [1].
Example
If a single radial arterial catheter-derived systolic blood pressure measurement is 120 mmHg (reference quantity value) and if a single simultaneously obtained oscillometric upper-arm cuff measurement is 150 mmHg, the oscillometric upper-arm cuff measurement error is 30 mmHg and combines random and systematic measurement errors. This measurement can be described as being "inaccurate” according to the intended use and combines untrueness and imprecision.

3.4 Measurement uncertainty

In the error approach (traditional approach, see above) the measurement error adds systematic and random errors, but no rule can be derived on how they combine for any given measurement. The uncertainty approach aims at characterizing the dispersion (pattern of distribution) of the values being attributed to a measurand, based on the information used [1]. This concept is broader than precision and may add systematic effects including uncertainty due to the reference method, time drift, definitional uncertainty, and other uncertainties. The objective of measurement in the uncertainty approach is not to determine a true value as closely as possible, but to reduce the range of values that can reasonably be attributed to the measurand [1].
Note
The translation from one language to another may be another source of confusion. For example, the VIM [1] is written in French and English. The French translation of “precision” is “fidélité” whereas “fidelity” in English is not mentioned in the document and usually refers to the degree of exactness with which something is copied or reproduced.
One solution would be to ban these quality concepts (accuracy, trueness, and precision) for which no specified numerical values are given and to be descriptive, speaking of “measurement error”, “systematic measurement error”, and “random measurement error”.
This—for two main reasons—is especially the case when using Bland-Altman analysis: First, the Bland-Altman plot has been proposed to compare two measuring instruments “when neither provides an unequivocally correct measurement” [4]. The second reason is that one important condition for estimating systematic and random errors is to average replicate measurements of the same quantity. Therefore, when several intra- or inter-patient measurements are done under different conditions, these estimations are strictly speaking impossible [3], and the criterium that is studied is the systematic discordance (or difference in agreement) between the two measuring instruments and its variability under different conditions.
An appropriate use of the Bland-Altman analysis to estimate the measurement trueness and precision would require: (1) a reference method and (2) replicate measurements of the same quantity, for example replicate measurements of the same measurand in the same patient in steady-state conditions (see part 2).

4 Methods (instruments for measurements)

A measuring instrument is a device used for making quantity measurements, alone or in conjunction with one or more supplementary devices (measuring system) [1]. A measuring instrument is frequently a transducer, i.e., a device that provides an output quantity (most often an electric signal) having a specific relation with an input quantity (most often a physiological signal). The physiological signal is collected by a sensor defined as an element of a measuring system that is directly affected by a phenomenon, body, or substance carrying a quantity to be measured, or less frequently by a detector defined as a device or substance that indicates the presence of a phenomenon, body, or substance when a threshold value of an associated quantity is exceeded [1].

5 Properties of measuring instruments (or devices)

An indication is a value provided by a measuring instrument [1]. An indication may result from many elementary measurements followed by a mathematic and/or algorithmic treatment. The measuring interval (or measuring range) is the set of values of the same kind that can be measured by a given instrument with specified instrumental uncertainty, under defined conditions [1]. A measuring instrument/system is characterized by different properties. The three main qualities of measurements seen before (precision, trueness, and accuracy) are obviously linked to instrumental properties, however, although measurements are facts that cannot be changed, instrumental properties are methods that can be improved by specific interventions (Table 2). Yet, the daily use may create confusion even in scientific documents. This is another reason for better being descriptive.
Table 1
International system of units
Quantity
Dimension
Unit
Symbol
Length
L
Meter
m
Mass
M
Kilogram
kg
Time
T
Second
s
Current
I
Ampere
A
Temperature
Θ
Kelvin
K
Amount of substance
N
Mole
mol
Luminous intensity
J
Candela
cd
Force
ML T2
Newton
N
Pressure
ML− 1T− 2
Pascal
Pa
Work or energy
ML2T− 2
Joule
J
Power
ML2T− 3
Watt
W
Frequency
T− 1
Hertz
Hz
All other quantities can be derived from these base quantities such as flow = volume (length3) / time (L3T-1) or hydraulic resistance = pressure/flow (ML−4T−1)
Table 2
Summary of measurement qualities and instrumental properties
Measurements (facts)
Quality
Quantity
Numerical value
Correction
Measurements precision
Random error
MV: σ, 2σ, 2σ/mean
Measurements trueness
Systematic error
Bias = AMV - R
Measurement accuracy
Measurement error
MV-R
Instruments (methods)
Property
Quantity
Numerical value
Correction
Instrumental precision
Random error
IV: σ, 2σ, 2σ/mean
Signal/noise
Systematic error
Bias = AIV-R
Zero, offset
Sensitivity
ΔIV/ΔR
Signal, gain
Linearity
ΔIV/ΔR = constant
Signal, gain
Resolution/step time response
Linked to SEM of IV
Signal, gain
MV measurement value, IV  indication value, R  reference value, A average, Δ change, SEM  standard error of the mean

5.1 Instrumental precision (sometimes called precision of method)

In analogy to the measurement precision, the instrumental precision is the closeness of agreement between indications obtained by replicate measurements on the same or similar quantities under specified and stable conditions [1, 3]. Although this is incorrect, the quality “instrumental precision” is often confounded with its linked “quantity”, the variability of indications.
Example
If the real blood flow of a patient is stable and equal to 5 L/min (as produced by a calibrated pump, or measured using a reference method such as an internal flow probe) and if, at the same time, 20 consecutive indications of a measuring device vary from 4.7 to 5.3 L/min, the random error of the indications can be estimated by σ = 0.25 L/min, 2σ = 0.5 L/min, or 2σ/mean value = 10%. Whether it can be said that the instrument precision is acceptable or not depends on the intended use. Strictly speaking, it should not be concluded that the instrumental precision is 10%.

5.2 Instrumental bias

In analogy to the measurement bias, the instrumental bias is the average of replicate indications minus a reference quantity value [1]. It estimates the systematic error provided by the measuring device. There is no quality linked to the instrumental bias, such as “instrumental trueness” in the VIM. Since accuracy is qualifying one single measurement, this quality cannot be used to describe an instrument. However, the term “accuracy class” is used to qualify measuring instruments that meet stated metrologic requirements.
Example
If the real blood flow of a patient is stable at 5.0 L/min (as measured in the example in 5.1), and if, at the same time, the average of 20 consecutive indications of a measuring device is 6.0 L/min, the tested measuring device has an instrumental bias of 1.0 L/min.

5.3 Sensitivity

The sensitivity is the quotient of the change in an indication and the corresponding change in a measurand [1]. The change considered must be large compared with the resolution (defined below, Fig. 3) [1]. The metrological sensitivity should not be confounded with the statistical sensitivity. Being a quotient between two changes, sensitivity is mathematically a regression slope, ideally = 1. Linearity, which is not a metrological but a mathematical property, illustrates the property of maintaining the sensitivity constant over the measuring interval. In other words, the linearity is also the capability of maintaining the instrumental bias constant. Preferably, the regression line should be close to the identity line (y = x, bias = 0 on the measuring interval). When the slope is not on the identity line (y = ax; a ≠ 1), it shows a constant but poor sensibility. When the slope formula is (y = ax + b; b ≠ 0), the sensibility can be good in a part of the measuring interval but not on the whole as exemplified in Fig. 3.
Example
If the real blood flow (as measured in the example in 5.1) is changing from 4.0 to 6.0 L/min, and if, at the same time, the indications of a measuring device change from 4.5 to 5.5 L/min, although the mean values are comparable, the tested measuring device is not sensitive.

5.4 Selectivity

The selectivity is a property, used with a specified measurement instrument, whereby it provides indications for one or more measurands such that the indications of each measurand are independent of other measurands or other quantities being investigated (Fig. 4) [1].
Example
If the real blood flow is stable at 5.0 L/min (as measured in the example in 5.1), and if, at the same time, the indications of a measuring device change from 5.0 to 6.0 L/min while blood pressure is increasing, the tested measuring device is not selective and may be dependent on the blood pressure.

5.5 Resolution

The resolution is the smallest change in a measurand that causes a perceptible change in the corresponding indication [1]. The concept of resolution is linked to the discrimination threshold, the largest change in the measurand that causes no detectable change in the corresponding indication, and to the dead band, which is the maximum interval through which a measurand can change in both directions without producing a detectable change in the corresponding indication [1]. Resolution may be linked to the physical granularity of the measurand (pixels, bits, quanta) often coming from the digitalization, but for most physiologic signals the smallest change in a measurand is limited by the standard error of the mean (SEM) of the corresponding indication. The change in the indication could be due to random errors. The SEM is proportional to the variability and to the number (n) of the elementary measurements used to display the indication \(2{\rm{~SEM~}} = {\rm{~}}2\sigma {\rm{~}}/{\rm{~}}\sqrt n\). Therefore, resolution, discriminating threshold, and dead band are linked to the random errors of elementary measurements (instrumental precision). A prescribed resolution can be reached by decreasing the random error, or by averaging more elementary measurements to give an indication (Fig. 5). The concept of least significant change (2\(\sqrt{2}\) SEM) that can be considered as statistically significant is linked to resolution.
Example
A blood flow measuring instrument connected to a bench giving a constant flow of 10 L/min provides the elementary measurements every second with a mean value = 10 L/min and a variability σ = 5 L/min. If an indication is to be given by the measuring instrument every second, the smallest perceptible change in the bench signal must exceed 10 L/min to be indicated (2SEM = 2σ/\({\sqrt{n}}\) = 10/1). If less, the change in the indication could be due to noise (random errors). Then, the resolution would be 10 L/min. If the manufacturer wanted to reach a resolution of 1 L/min, there were only two solutions: first decreasing the noise (variability) from 5 to 0.5 (2 σ/\(\sqrt{n}\) = 1/1), second increasing the number of measurements from 1 to 100 (2 σ /\(\sqrt{n}\) = 10/10), therefore giving an indication only every 100 s.

5.6 Step response time

The step response time is the duration between the instant when a measurand is subjected to an abrupt change and the instant when the corresponding indication of a measuring instrument settles within specified limits around its final steady value (Fig. 6) [1]. The way by which the final steady value is determined may be different: for example, the inflection point between two regression curves, or the first point of a flat curve slope, or the first point where the σ becomes below specified limits. Step response time is also linked to the measurement precision since low precision increases the number of indications needed to establish the final steady state.
Example
If the real blood flow changes from 5.0 to 6.0 L/min in 10 s (as measured in the example in 5.1), and if the indications of a measuring device changed from 5.0 to 6.0 L/min, in 10 min, the step response time of the tested measuring device is close to 10 min.

5.7 Stability

The stability is the property of a measuring instrument, whereby its metrological properties remain constant in time [1]. An instrumental drift is a continuous or incremental change over time of the indication due to change in at least one metrological property.
Example
If the real blood flow (as measured in the example in 5.1) is 5.0 L/min, and if the indications of a measuring device changed from 5.0 to 6.0 L/min within a certain time period (e.g. 12 h), the tested measuring device is not stable.

5.8 Maximum permissible measurement error or limit of errors

The maximum permissible measurement error (or limits of errors) is the extreme value of measurement error permitted by specifications or regulations for a given measurement, measuring instrument, or measuring system [1]. The term tolerance (not defined in the VIM), should not be used to designate the maximum permissible error [1]. Tolerance most often includes the true value ± the maximum permissible error of a fixed physical property.
Example
In the preceding example, if the clinical requirements allow a maximum permissible error of 20%, the instability becomes unacceptable after 12 h and the device needs recalibration.

6 Measurement standards (etalon)

Any measurement requires a measurement standard (etalon), which is the embodiment of the definition of a given quantity, with stated quantity value and associated measurement uncertainty, used as the reference [1]. This definition shows that the uncertainty with the measurement standard contributes to the combined measurement uncertainty since values that result from the measurement process are in reality ratios between the measured values and the measurement standard, expressed in the same units. In November 2018, the BIPM has changed the definitions of the international standards. All definitions are now based on atomic constants to minimize uncertainties [5]. These changes came into force on May 20th, 2019 [5].
Example
The kilogram was defined until now by the mass of a cylinder alloy (90% platinum and 10% iridium) manufactured in 1889, stored at the BIPM, with official copies sent in 40 national metrologic centers worldwide. Although carefully stored, these etalons diverge from the original by ≈ 50 μg per century. Therefore, the kilogram is now defined by the Planck constant set at = 6.62607015 × 10 − 34 m2 kg s− 1 and practically obtained from a Kibble balance.

6.1 Calibration

A measurement standard is the prerequisite of any calibration, which is the operation that, in a first step (in specified conditions) establishes a relation between a device indication and the corresponding quantity values provided by a measurement standard (with known uncertainty) and, in a second step, uses this information for obtaining a measurement result (with appropriate units) from an indication [1]. Strictly speaking, calibration is just a comparison [1]. However, in general use, the term calibration also refers to a second step that is using these initial steps for (1) the verification that the test device meets the prescribed standards, and if not, (2) for the adjustment of a measuring system, sometimes improperly called “auto-calibration”, which is the set of operations (zero, offset, and span or gain adjustment) carried out on a measuring system so that it provides prescribed indications corresponding to given values of a measurand (Fig. 7) [1].
Example
If the real blood flow is 4.0, 5.0, and 6.0 L/min during a given maneuver, and if the corresponding indications of a measuring device are 5.0, 6.0, and 7.0 L/min, the measuring device needs a recalibration (zero and offset). If the corresponding indications of another measuring device are 4.0, 4.5, and 5.0 L/min, the second measuring device needs a recalibration (offset and gain).

6.2 Metrological traceability

The metrological traceability is the property of a measurement result whereby the result can be related to a reference through an unbroken chain of calibrations, each contributing to the measurement uncertainty [1]. An unbroken chain of calibrations means that SI units are determined. The working instrument is then compared (calibrated) with the best practically available reference method. This reference method is compared to a higher standard (a standard with less uncertainty) again and again, and the chain is documented through calibration certificates.

7 Conclusion

In perioperative and intensive care medicine, the metrological requirements for measurements (facts) and measuring instruments (methods) are poorly defined. One of the reasons may be the lack of consensus among physicians and scientific societies on which are the minimum quality criteria. Full transparency is needed in the validation of new measuring devices [6]. Adopting the same understandings and definitions among physicians and scientists is obviously the first step.

Acknowledgements

This project was endorsed by the European Society of Intensive Care Medicine. All definitions are reproduced from the reference 1 with permission. However, the only authentic versions are those of the documents of the Joint Committee for Guides in Metrology.

Compliance with ethical standards

Conflict of interest

BS has received honoraria for consulting, honoraria for giving lectures, and refunds of travel expenses from Edwards Lifesciences Inc. (Irvine, CA, USA). BS has received honoraria for consulting, institutional restricted research grants, honoraria for giving lectures, and refunds of travel expenses from Pulsion Medical Systems SE (Feldkirchen, Germany). BS has received institutional restricted research grants, honoraria for giving lectures, and refunds of travel expenses from CNSystems Medizintechnik GmbH (Graz, Austria). BS has received institutional restricted research grants from Retia Medical LLC. (Valhalla, NY, USA). BS has received honoraria for giving lectures from Philips Medizin Systeme Böblingen GmbH (Böblingen, Germany). BS has received honoraria for consulting, institutional restricted research grants, and refunds of travel expenses from Tensys Medical Inc. (San Diego, CA, USA). XM is a member of the medical advisory board of Pulsion Medical Systems, member of Getinge, and gave some lectures for Cheetah medical. MLNGM is member of the medical advisory Board of Getinge (former Pulsion Medical Systems) and Serenno Medical. He consults for Baxter, Maltron, ConvaTec, Acelity, Spiegelberg and Holtech Medical. TWLS received research grants and honoraria from Edwards Lifesciences (Irvine, CA, USA) and Masimo Inc. (Irvine, CA, USA) for consulting and lecturing and from Pulsion Medical Systems SE (Feldkirchen, Germany) for lecturing. TWLS is associate editor of the Journal of Clinical Monitoring and Computing but had no role in the handling of this paper. PS received grants from Sorin for patents fees and was reimbursed for patent fees maintenance by Medtronic. He received consultant honorarium and was reimbursed for travel expenses from Medtronic. The other authors declared no conflicts of interest.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
2.
Zurück zum Zitat Regulation (EU) 2017/745 of the European Parliament and of the council on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC. Official Journal of the European Union. 2017. http://data.europa.eu/eli/reg/2017/2745/oj. Regulation (EU) 2017/745 of the European Parliament and of the council on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC. Official Journal of the European Union. 2017. http://​data.​europa.​eu/​eli/​reg/​2017/​2745/​oj.
4.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.CrossRef
5.
Zurück zum Zitat BIPM's Member States. On the revision of the SI. Paris: International bureau for weight and measures (BIPM); 2018. BIPM's Member States. On the revision of the SI. Paris: International bureau for weight and measures (BIPM); 2018.
Metadaten
Titel
Metrology part 1: definition of quality criteria
verfasst von
Pierre Squara
Thomas W. L. Scheeren
Hollmann D. Aya
Jan Bakker
Maurizio Cecconi
Sharon Einav
Manu L. N. G. Malbrain
Xavier Monnet
Daniel A. Reuter
Iwan C. C. van der Horst
Bernd Saugel
Publikationsdatum
17.03.2020
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2021
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-020-00494-y

Weitere Artikel der Ausgabe 1/2021

Journal of Clinical Monitoring and Computing 1/2021 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.