Background
Long-term survival of total hip arthroplasty (THA) requires excellent fixation between the host bone and the prosthesis. This fixation can be achieved via cement, bone in-growth, or bone on-growth. The latter can be promoted through the use of bone growth promoting substances, such as hydroxyapatite (HA). HA is a naturally occurring mineral found within bone composition, and its presence on the femoral component supports osseointegration of the stem [
1,
2]. While primary stability of the femoral stem in cementless THA models depends on a press fit between the bone and femoral component, long-term stability requires osseointegration of cancellous bone to the HA-coated stem [
3]. Hydroxyapatite-coated femoral stems demonstrate high clinical success rates [
1,
4]; however, some concerns remain over the resorption of HA and delamination of HA from the stem which can lead to loosening of the implant [
5].
In metal-on-metal (MoM) hip arthroplasty, bearing surfaces of the femoral head and acetabular liner are manufactured with high precision to allow fluid-film lubrication to minimize wear. Despite this design, both patient and implant-specific factors have been identified that can lead to significant bearing wear with production of large quantities of cobalt (Co) and chromium (Cr) wear particles and ions that can have local and systemic effects [
6]. Metal wear particles induce an inflammatory response which can cause damage to surrounding soft tissue and bone [
7,
8] resulting in pain, loss of function, and often revision arthroplasty [
9]. Beyond local reactions, high systemic metal ion concentrations have been associated with poor osseointegration of the femoral stem. Shah et al. showed that high Co and Cr ions decrease osteoblastic activity and alkaline phosphatase activity, both important processes of bone growth and development [
10]. Metal ions also have a direct negative effect on HA by binding into the HA crystal in place of calcium and inhibiting the mineralization process [
11]. These deleterious effects of metal ions reportedly discourage initial osseointegration processes; however, the mid- to long-term effects on bone-HA fixation are not known and numerous confounding factors may also be of direct influence.
As a result of the poor clinical performance of MoM THAs, most manufacturers have eliminated MoM as a bearing option corresponding to steep decline in their use since 2008 [
12,
13]. However, many tens of thousands of patients still have functioning MoM THAs and may yet be affected by the long-term effects of these generally poor performing implants.
Several HA-coated femoral stems were identified through the implant retrieval analysis program at the Orthopaedic Innovation Centre which had clinically loosened after several years of appearing radiographically osseointegrated. We hypothesized that these cases may be associated with metal particle generation from the MoM articulation or from corrosion of the head-neck taper and subsequent depletion of the HA coating on the femoral stem. Other possible factors of mid-term HA-coated femoral stem loosening, such as stress shielding, were not examined in this study.
This study was undertaken to examine the possible causes and associated risk factors for mid-term (3–7 years post-surgery) aseptic loosening of HA-coated femoral stems and to detail their clinical manifestation through radiographic and implant retrieval analysis.
Discussion
Despite improvements in bearing technology, aseptic loosening remains a leading cause for revision hip arthroplasty [
18]. Promising wear simulator data for MoM has not translated clinically into improved survivorship, and results of the DePuy Pinnacle MoM THA have been varied. Some authors have noted good mid-term survivorship from 97.0–99.4% at 5–7 years [
19‐
22]. Others have found much worse survival ranging from 92.8% at 7 years [
23], 88.9% at 8 years [
24], to 83.6–86% at 9 years [
25,
26]. Ten-year cumulative percentage probability of revision reported in the Australian and UK national joint registries for the Corail stem and Pinnacle cup is 3.2–4.9% for conventional bearings vs 11.7–14.6% for MoM. Our results for Corail Pinnacle MoM suggest taper corrosion is a factor in the loosening of a fully HA-coated stem that has previously shown excellent survivorship with other bearings such as MoP, CoP, and CoC [
27,
28].
A recent publication by Buttaro et al. described a similar phenomenon with this same stem design presenting with “metaphyseal debonding” at a mean of 36 months from surgery. Their case-control study found association with increased BMI and the use of hard-on-hard bearing surfaces (both ceramic-on-ceramic and MoM) [
29]. In contrast, our study did not find patient variables, including BMI, to be associated with loss of HA coating, nor were ceramic-on-ceramic bearings found to undergo HA coating loss. Buttaro et al. were only able to assess 6 implants of the 18 cases identified, all six of which were either ceramic-on-ceramic or ceramic-on-polyethylene. Four of the 6 retrieved stems demonstrated loss of the proximal HA coating, but no analysis of the condition of the taper was discussed in the publication.
We found the presence and observed severity of taper corrosion to be associated with aseptic loosening of HA-coated femoral stems in a relatively small number of samples. The predominant mechanism of taper corrosion is thought to be mechanically assisted crevice corrosion [
15]. Other contributing factors include use of dissimilar alloys [
30], taper size and design [
31], head size and length [
32], stem offset [
33], flexural rigidity of the stem neck [
34], intraoperative contamination with blood or fluid, assembly force [
35], and length of implantation [
33,
36]. The Corail taper with its short, roughened surface may be particularly prone to corrosion [
37]. Interestingly, MoP and CoC had no difference in corrosion scores. Dissimilar metals alone fail to account for the difference seen among the MoM as the taper, and CoCr head of the MoP group had little corrosion, although this may be explained by their shorter implantation time [
32].
Corrosion score was also positively correlated with proximal HA loss. All retrieved MoM stems in our study had significant proximal HA loss (grade 2 or 3), with six out of seven stems having failed for aseptic loosening also showing a similar degree of HA loss. The proposed mechanisms of HA loss in the literature include delamination from the metal substrate, dissolution into extracellular fluid, and cell-mediated resorption. Based on an autopsy retrieval study of HA-coated stems, Bauer et al. suggested osteoclastic resorption was the primary mode of HA resorption, assisted by enzymatic breakdown common to bone remodeling [
38]. This process may be enhanced by corrosion byproducts that act as potent activators of macrophage-induced osteolysis [
39]. Additionally, delayed-type hypersensitivity reactions have been implicated in aseptic osteolysis seen in MoM due to metal-induced lymphocytes releasing cytokines to shift the balance in favor of osteoclastic resorption [
40]. Three of the MoM cases that failed for aseptic loosening in our series had tissue pathology results available. The tissue analyses were consistent with mild to moderate hypersensitivity reaction with moderate ALVAL scores (6, 7, and 7), suggesting this may have contributed to stem loosening.
Another possible explanation relating proximal HA loss to taper corrosion is that HA can be dissolved in acidic solutions, as demonstrated in dentistry literature [
41]. While the pH of the joint space is more neutral compared to that of the mouth, it is feasible that it is locally altered by corrosion and wear products. Cobalt chromium and titanium alloys are biocompatible metals primarily because when in contact with oxygen, they form a protective surface film of oxidized metal rendering it non-reactive. Mechanical removal of this layer either by taper micromotion or bearing wear results in immediate reformation of the oxide layer with nearby oxygen atoms from water molecules. This oxidation reaction creates free H
+ ions which increase the acidity of the surrounding environment. Milosev et al. measured the pH of osteoarthritic synovial fluid, compared it to those with metal hip and knee arthroplasty components, and noted a small but significant decrease in pH in the arthroplasty group [
42]. Inflammatory conditions such as rheumatoid arthritis and septic arthritis have also been found to lower synovial fluid pH. Willert et al. measured the pH at the interface of 17 titanium alloy stems at time of revision noted to have signs of corrosion and found 7 stems had sub-physiologic levels dropping as low as 2.2 [
43]. Lowering of local pH by continual reoxidation of CoCr alloy on various surfaces of the THA, and specifically the taper, can therefore potentially contribute to dissolution of the HA coating and aseptic loosening.
A further explanation for stem loosening is increased particulate debris from bearing wear, which has been found in some MoM implants that have higher risk of mid-term aseptic loosening. When functioning well, MoM bearings are extremely wear resistant; however, they are less tolerant for implant malposition and the smaller size and subsequent greater surface area of wear debris has high bioactivity and cytotoxicity. Further, smaller cup size has been shown to have a higher risk of wear in MoM bearings as the corresponding metal insert is thinner with higher likelihood of deformation when engaged in the shell taper. The 50 mm and 52 mm outside diameters of 36 mm liners (wall thickness of 7 mm and 8 mm) have demonstrated significantly increased rates of revision in the literature [
25]. In our series, liner wall thickness ranged 7 mm up to 12 mm, with one sample at 7 mm and two samples at 8 mm. Insufficient bearing clearance has also been implicated as a possible cause for premature failure of MoM THAs [
25]. Reduced clearance increases likelihood and severity of bearing wear on MoM THAs but may also increase the toggling torque on the head-neck taper. Micromotion of the head-neck taper initiates mechanically-assisted crevice corrosion, a well-described process of taper corrosion and subsequent metal debris release [
15]. While increased torque-induced stem micromotion may lead to failure of initial press fit and osseointegration as suggested in Pilliar et al., [
44], it does not seem to explain the progressive loosening of previously well-fixed stems seen in the current series.
Higher cup inclination angle in MoM bearings [
45] has been shown to influence wear; however, this was unlikely a factor in our study as mean cup inclination of the 7 MoM hips that failed for aseptic loosening with available x-rays was 42°(range 39–47°). Reinisch et al. found no evidence of increased bearing wear among 22 single manufacturer MoM implants with 28 mm heads revised for early (12–59 months) aseptic loosening, suggesting this process can occur under normal bearing wear settings [
46]. Langton et al. also found 6 cases of failed MoM with well-functioning bearing surfaces and described significantly worn taper junctions as the cause [
47]. This finding has been confirmed in a retrospective cohort study examining mid-term revisions of the pinnacle MoM bearing paired with Corail or S-ROM stems showing failure from the taper junction was more common than at the bearing surface [
25]. The MoM bearing wear in our study was similarly found to be low and within the normal range of reported values for this articulation [
25].
Among the nine MoM hips that had 2 or more x-rays available for review, radiolucency progressed from proximal to distal at a mean rate of 0.9 mm/year. None of the cases saw progression stop until time of revision. If radiolucency progression is related to ongoing taper corrosion that predictably leads to mid-term aseptic loosening, then this represents both an early warning sign and a potential therapeutic window. If recognized early when the stem remains well fixed, removing the source of corrosion may salvage the stem, limit soft tissue damage, and avoid the morbidity of more extensive revision. A recent systematic review identified nine studies examining the treatment of hip implant failures secondary to taper corrosion with the majority being treated with stem retention and exchange to a ceramic head with titanium sleeve [
48]. Although follow-up was short (1–36 months), the majority of patients had resolution of their symptoms.
Limitations of this study include the small sample size available, x-ray availability in only a subset, varying demographics between the patient groups (age, demand on the implant, and potentially activity level), variation in implant dimensions and designs, and longer implantation time in the MoM group which biases radiolucency, HA coating removal, and taper corrosion. The observed mid-term loosening, however, is unexpected in a stem with good, lengthy clinical track record.