Modification of chest radiography exposure parameters using a neonatal chest phantom

Zeitschrift:
Pediatric Radiology > Ausgabe 1/2020
Autoren:
Stefan B. Schäfer, Sabine Papst, Martin Fiebich, Claudia Rudolph, Jan de Laffolie, Gabriele A. Krombach
Wichtige Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abstract

Background

The acquisition of chest radiographs in neonates is of critical importance in diagnostics because of the risk of respiratory distress syndrome and pneumothorax in preterm infants.

Objective

To achieve a dose reduction while preserving a diagnostic image quality for chest radiographs of neonates.

Materials and methods

All radiographs, generated on a fully digital X-ray unit by using a neonatal chest phantom, were evaluated under variation of the tube voltage (40–70 kV) and mAs levels (1–10.2 mAs) with and without an additional 0.1-mm copper (Cu) filtration. Noise, contrast and contrast-to-noise ratio for bronchus, heart, lungs and vessels were determined. Visual assessment of the image quality was carried out by three radiologists using a Likert scale. To evaluate a maximally possible dose reduction, the dose of the radiographs with still acceptable image quality at a minimal dose was compared to the dose of the radiographs with the standard settings used in clinical routine.

Results

The noise showed decreasing values with increasing dose, while the contrast values were increased. For the contrast-to-noise ratio, a digressive course of the values as a function of the tube voltage was found. The visual evaluation of image quality showed the best evaluation of the structures at the lowest possible dose in the settings (44 kV, 3.36 mAs) with copper filtration and in the settings (44 kV, 1.56 mAs) without copper filtration. A maximum dose reduction from 8.29 μSv to 2.21 μSv (about 73%) was obtained.

Conclusion

A dose reduction while preserving diagnostic image quality in a digital X-ray system is generally possible by reducing the tube voltage and simultaneous adaptation of the mAs settings.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag als Mediziner

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Alle e.Med Abos bis 30. April 2021 zum halben Preis!

Jetzt e.Med zum Sonderpreis bestellen!

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 1/2020

Pediatric Radiology 1/2020 Zur Ausgabe
  1. Sie können e.Med Pädiatrie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

  2. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Radiologie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Radiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise