Multicenter comparison of three intraoperative hemoglobin trend monitoring methods

verfasst von: Richard L. Applegate II, Patricia M. Applegate, Maxime Cannesson, Prith Peiris, Beth L. Ladlie, Klaus Torp

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 5/2020

Abstract

Transfusion decisions are guided by clinical factors and measured hemoglobin (Hb). Time required for blood sampling and analysis may cause Hb measurement to lag clinical conditions, thus continuous intraoperative Hb trend monitoring may provide useful information. This multicenter study was designed to compare three methods of determining intraoperative Hb changes (trend accuracy) to laboratory determined Hb changes. Adult surgical patients with planned arterial catheterization were studied. With each blood gas analysis performed, pulse cooximetry hemoglobin (SpHb) was recorded, and arterial blood Hb was measured by hematology (tHb), arterial blood gas cooximetry (ABGHb), and point of care (aHQHb) analyzers. Hb change was calculated and trend accuracy assessed by modified Bland–Altman analysis. Secondary measures included Hb measurement change direction agreement. Trend accuracy mean bias (95% limits of agreement; g/dl) for SpHb was 0.10 (− 1.14 to 1.35); for ABGHb was − 0.02 (− 1.06 to 1.02); and for aHQHb was 0.003 (− 0.95 to 0.95). Changes more than ± 0.5 g/dl agreed with tHb changes more than ± 0.25 g/dl in 94.2% (88.9–97.0%) SpHb changes, 98.9% (96.1–99.7%) ABGHb changes and 99.0% (96.4–99.7%) aHQHb changes. Sequential changes in SpHb, ABGHb and aHQHb exceeding ± 0.5 g/dl have similar agreement to the direction but not necessarily the magnitude of sequential tHb change. While Hb blood tests should continue to be used to inform transfusion decisions, intraoperative continuous noninvasive SpHb decreases more than − 0.5 g/dl could be a good indicator of the need to measure tHb.
Hinweise
A correction to this article is available online at https://​doi.​org/​10.​1007/​s10877-020-00566-z.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

1 Introduction

Hemoglobin (Hb) measurement informs patient-specific perioperative transfusion decisions within the context of symptoms, comorbid conditions, surgical procedure, observed bleeding and hemodynamic performance [1, 2]. Hb measurement is a key component in many parts of patient blood management bundles [3], and is recommended between transfused red blood cell units if patient stability allows [4].
Hb measurement may be performed using clinical laboratory hematology (tHb; closest to the cyanmethemoglobin standard [5]), arterial blood gas cooximetry (ABGHb), or point of care Hb analyzers. However, the time needed for blood sampling and analysis can cause Hb measurement to lag clinical situations. In surgical settings in which blood loss may not be apparent or be difficult to estimate, continuous rather than intermittent Hb monitoring could provide earlier warning of decreasing Hb.
Multiwave pulse cooximetry noninvasively determines total hemoglobin (SpHb). SpHb has been tested in volunteers [6] and in a wide range of clinical settings [7]. Prior reports indicated wide limits of agreement between SpHb and tHb that suggest caution when using SpHb alone to guide transfusion decisions [7, 8]. However, it is possible that SpHb changes could provide useful information if the direction of SpHb change accurately reflects the direction of tHb change (trend). We defined trend accuracy as agreement of sequential changes in SpHb, ABGHb, or point of care Hb with sequential tHb changes. The aim of this multicenter study was to evaluate trend accuracy of three monitoring methods in patients undergoing surgery.

2 Methods and materials

This collaborative prospective convenience sample observational study received Institutional Review Board approval at each of three USA academic medical centers prior to study initiation: Loma Linda University (LLU), Loma Linda, CA; Mayo Clinic in Florida (MCF), Jacksonville, FL; and University of California Irvine (UCI), Irvine, CA. Patient consent was obtained according to local IRB determination at each site. All study procedures were performed in accordance with the ethical standards of the institutional and research committees and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. This manuscript adheres to applicable STARD guidelines.
Adult patients were eligible for study participation if scheduled to undergo non-cardiac surgery in which arterial catheterization and expected repeated intraoperative blood gas analyzes were planned as part of patient care. Patients were excluded if pregnant or for skin abnormalities at the planned application site that would interfere with pulse oximetry (burns, scar tissue, nail polish, or infection). Attending anesthesiologists had discretion over anesthesia management, fluid administration and transfusion decisions, which were based on clinical settings (surgical considerations, ongoing bleeding, hemodynamic condition and/or any patient comorbid conditions) and institutional guidelines, which were in keeping with published guidelines [9, 10].
Multiwave disposable pulse cooximetry finger sensors (R125, Radical-7, Revision K, Masimo, Irvine, CA) were placed on the ring finger on the side of arterial cannulation with data continuously collected to computer. Oximeters were set to arterial mode to align with the source of blood samples. The SpHb algorithm continuously evaluates up to 6 min of data to calculate the displayed value. Whenever arterial blood gas analysis was performed, SpHb displayed at the time arterial blood was drawn was recorded, and arterial blood samples were obtained in appropriate collection tubes. Blood sample analysis was completed within 10 min. Each blood sample was analyzed twice using the same analyzer for tHb, ABGHb and aHQHb using:
  • Clinical Laboratory hematology analyzer Hb (tHb; LLU—Sysmex XE5000, Sysmex America Inc., Lincolnshire, IL, USA; MCF—Sysmex XE5000, Sysmex America Inc., Lincolnshire, IL, USA or Coulter AcTdiff, Beckman Coulter, Indianapolis, IN, USA. UCI—Coulter LH 750 Hematology Analyzer, Beckman Coulter, Brea, CA, USA)
  • Arterial blood gas cooximetry Hb (ABGHb; LLU—Radiometer ABL800; Radiometer, Copenhagen, Denmark. MCF—CCX or PhOX, Nova Biomedical, Waltham, MA, USA; UCI—Siemens RAPIDLab 1265, Siemens Healthcare Diagnostics, NY, USA)
  • Point of care Hb using arterial blood (aHQHb; not capillary or finger stick samples; HemoCue HB 301, HemoCue America, Brea, CA, USA)

2.1 Statistical methods

The primary outcome measure was trend accuracy using modified Bland–Altman analysis of difference between changes in SpHb, ABGHb or aHQHb and tHb changes to tHb changes as the measurement standard, to obtain bias and 95% limits of agreement between trends in SpHb, ABGHb, or aHQHb and tHb trend. Trends were defined as sequential change in Hb using results of the first analysis performed on each blood sample: Hb sample 2—Hb sample 1; Hb sample 3—Hb sample 2; etc. We included in analysis only samples for which all measures were available for trend calculations and did not exclude or separately analyze SpHb in low perfusion states.
Power and sample size calculation: A prior study of patients undergoing similar procedures [11] reported a median of 4 (up to 9) blood samples allowing an average of 3 trend calculations per patient. The mean bias of SpHb to tHb reported in prior studies in which the intended hematology analyzers were used [7] ranged from − 0.53 to 1.22, with standard deviation averaging 1.055. Using that standard deviation, setting alpha to 0.05 and power to 0.8, a sample size of 135 patients would be needed to estimate the precision of the 95% confidence interval of the difference among the 3 methods within 0.2 g/dl This is a conservative estimate of our true precision because patients would have repeated blood draws. Power analyses were performed using SAS 9.4 (SAS Institute, Cary, North Carolina, USA); statistical analyses were performed using JMP Pro version 13.2.0 (SAS Institute, Cary, North Carolina, USA) and Prism 8.1.0 (GraphPad Software, San Diego, California, USA).

3 Secondary outcome measures

3.1 Hb accuracy

Calculated Hb change could be impacted by Hb measurement accuracy. Accuracy of SpHb, ABGHb or aHQHb compared to tHb was evaluated by Bland–Altman analysis to determine bias and 95% limits of agreement. Hb measurement results were compared to clinically acceptable error as previously recommended [12].

3.2 Hb measurement repeatability

Results of the 2 analyses performed on each blood sample were compared by Bland–Altman analysis to find bias and 95% limits of agreement for tHb, ABGHb and aHQHb.

3.3 Agreement of change direction

The 95% limits of agreement obtained from Hb measurement repeatability analysis were used to define exclusion zones for change direction agreement analysis. Change direction agreement was assessed overall and for samples with tHb < 9.0 g/dl. This was reported as % agreement; 95% confidence interval.

4 Results

Each center enrolled independently with one-hundred thirty-five patients studied (Loma Linda University n = 51; Mayo Clinic in Florida n = 58; University of California Irvine n = 26). On average patients had 4 samples obtained, ranging from 2 to 13. A total of 568 blood gas samples were drawn from these patients. Of these, 5 (0.88%) samples were missing a blood gas analysis result, 5 (0.88%) were missing SpHb and 7 (1.23%) had automated data collection errors, leaving 551 samples in which all measurements were available providing 416 changes to calculate trends for comparisons. Patient and procedure characteristics are shown in Table 1.
Table 1
Patient characteristics
Patient characteristics
All
N = 135
LLU
N = 50
MCF
N = 59
UCI
N = 26
Sex # (%)
Female
70 (51.9%)
29 (58.0%)
26 (44.1%)
15 (51.9%)
Male
65 (48.1%)
21 (42.0%)
33 (55.9%)
11 (42.3%)
Age years
61
[50–69]
58.5
[44.5–68.5]
63
[54–70]
58.5
[47.8–68.3]
Weight kg
79.3
[66.1–96.2]
79.6
[64.9–98.9]
79.3
[66.6–92.6]
76.9
[64.6–96.4]
Body mass index kg m−2
28.0
[24.0–32.7]
28.0 [24.5–35.0]
28.1 [23.8–29.9]
27.9 [22.7–33.2]
Number of samples per patient
4
4
4
4
Range
[3–5]
2–13
[3–5]
2–6
[2–6]
2–13
[4–4]
2–4
1st intraoperative hemoglobin g/dl
Range
10.8
[9.1–11.9]
5.9–14.7
10.8
[9.3–11.9]
5.9–14.7
10.1
[8.6–11.7]
6.8–13.8
11.7a
[10.3–12.9]
9.1–13.8
Intraoperative hemoglobin g/dl
Range
10.1
[8.9–11.5]
4.9–14.7
10.0
[8.7–11.3]
4.9–14.7
9.5
[8.6–11.0]
5.9–14.1
11.5
[10.3–12.5]
8.4–14.3b
Sequential change in laboratory hemoglobin g/dl
Range
− 0.1
[− 0.6 to 0.4]
− 5.0 to 3.7
0.3
[− 0.8 to 0.3]
− 5.0 to 3.2d
0
[− 0.7 to 0.5]
− 3.3 to 3.7
0
[− 0.2 to 0.3]
− 3.3 to 0.9
Duration of monitoring minutes
302
[181–390]
335
[257–386]
354
[266–437]
88c
[79–101]
Surgical procedure type #
 Major abdominal
58
19
26
13
 Liver resection or transplant
25
3
17
5
 Major orthopedic
16
8
5
23
 Major urologic
16
3
11
2
 Major gynecologic
12
11
0
1
 Major vascular
6
6
0
0
 Major neurosurgical
2
0
0
2
Characteristics of patients undergoing surgery with arterial catheterization at one of 3 academic medical centers: LLU Loma Linda University; MCF Mayo Clinic in Florida; UCI University of California Irvine. Results are median [25th to 75th percentile] except sex and procedure type which are number (%)
aFirst intraoperative hemoglobin higher at UCI than LLU (Hodges Lehman difference 0.9; 0.1–0.8 g/dl p = 0.03) and MCF (Hodges Lehman difference 1.3; 0.5–2.2 g/dl p = 0.003)
bIntraoperative hemoglobin higher at UCI than LLU (Hodges Lehman difference 1.4; 1.0–1.8 g/dl) and MCF (Hodges Lehman difference 1.7; 1.3–2.1 g/dl) both p < 0.0001
cDuration of monitoring less at UCI than LLU (Hodges Lehman difference − 237; − 268 to − 201 min) and MCF (Hodges Lehman difference − 264; − 298 to − 225 min) both p < 0.0001
dSequential change more negative at LLU than MCF (Hodges Lehman L difference − 0.2; − 0.4 to 0 g/dl; p = 0.04) and UCI (Hodges Lehman difference − 0.3; − 0.4 to − 0.1 g/dl; p = 0.006)

4.1 Trend accuracy

Modified Bland–Altman analysis showed small bias with slightly wider limits of agreement for SpHb trends compared to ABGHb or aHQHb trends (Fig. 1). Mean bias (limits of agreement g/dl) for SpHb was 0.10 (− 1.14 to 1.35); for ABGHb was − 0.02 (− 1.06 to 1.02); and for aHQHb was 0.003 (− 0.95 to 0.95).

4.2 Hb accuracy

Compared to tHb, mean bias (95% limits of agreement) for SpHb was 0.24 (− 2.05 to 2.53) g/dl; for ABGHb was − 0.36 (− 1.47 to 0.78) g/dl; and for aHQHb was − 0.43 (− 1.46 to 0.60) g/dl. Hb accuracy is plotted against clinically acceptable error [12, 13] in Fig. 2. None of the results were in the zone that has been proposed to potentially expose patients to larger risks [13].

4.3 Measurement repeatability

Bias (95% limits of agreement) of results from the 2 analyses of each blood sample for tHb was 0.0055 (− 0.25 to 0.26) g/dl; for ABGHb was − 0.032 (− 0.61 to 0.54) g/dl; and for aHQHb was − 0.0004 (− 0.62 to 0.61) g/dl. These limits of agreement were used to set exclusion zones for comparing change direction agreement. A tHb change more than ± 0.25 g/dl was considered an increase or decrease, while a change more than ± 0.5 g/dl was considered an increase or decrease for SpHb, ABGHb and aHQHb.

4.4 Agreement in change direction

Using the defined exclusion zones, change direction agreed in 129 of 137 SpHb changes (94.2%; 88.9–97.0%); 179 of 181 ABGHb changes (98.9%; 96.1–99.7%); and in 195 of 197 aHQHb changes (99.0%; 96.4–99.7%). Table 2 details change direction agreement for increases and decreases. Four quadrant plots using the defined exclusion zones are shown in Fig. 3. In all samples when SpHb, ABGHb or aHQHb increased but tHb decreased the decreases in tHb were smaller than − 1 g/dl. For tHb < 9.0 g/dl changes in SpHb, ABGHb or aHQHb more than ± 0.5 g/dl agreed with tHb change direction in all but 3 instances (Fig. 4). Most samples in which SpHb, ABGHb or aHQHb change was not more than ± 0.5 g/dl were associated with tHb changes within ± 0.25 g/dl. However, when tHb was < 9.0 g/dl, one ABGHb and 3 SpHb changes not more than ± 0.5 g/dl had an associated tHb decrease more than − 1 g/dl.
Table 2
Agreement of trend direction between tHb change more than ± 0.25 g/dl and SpHb, ABGHb and aHQHb change more than ± 0.5 g/dl
 
Same direction as tHb trend
N (%; 95% CI)
Not the same direction as tHb trend
N (%; 95% CI)
Increase
 SpHb
61 (92.4%; 83.5–96.7%)
5 (7.6%; 3.3–16.5%)
 ABGHb
76 (97.4%; 91.1–99.3%)
2 (2.6%; 0.7–8.9%)
 aHQHb
84 (100%; 95.6–100%)
0
Decrease
 SpHb
68 (97.8%; 88.3–98.6%)
3 (4.2%; 1.4–11.7%)
 ABGHb
103 (100%; 96.4–100%)
0
 aHQHb
111 (98.2%; 93.8–99.5%)
2 (1.8%; 0.5–6.2%)

5 Discussion

Intraoperative Hb trend accuracy limits of agreement were approximately ± 1 g/dl for ABGHb and aHQHb and slightly larger for SpHb trends. Considering the 95% confidence intervals for agreement in change direction overlapped, our findings suggest that changes more than ± 0.5 g/dl in SpHb, ABGHb and aHQHb provide similar information regarding the direction of intraoperative tHb change, but not necessarily the magnitude of that change. The impact of trend accuracy on transfusion decision-making is more significant when tHb is less than 9.0 g/dl as current guidelines suggest that red blood cell transfusion may be reasonable [14] for tHb < 8.0 g/dl, but usually unnecessary [9] for tHb > 10.0 g/dl. Analysis of change direction agreement suggests that if continuous SpHb is < 9 g/dl then a decrease more than − 0.5 g/dl could be a good indication to obtain a blood sample for tHb measurement.
Repeatability analysis demonstrated limits of agreement that have implications for Hb accuracy and trend monitoring studies. The exclusion zones we defined based on 95% limits of agreement for ABGHb and aHQHB suggest that in our patients, a change up to ± 0.5 g/dl may not reflect a real change in circulating hemoglobin measured by tHb. Similarly, reported tHb changes of ± 0.25 g/dl may not represent real changes in circulating Hb. Our findings support use of an exclusion zone when comparing Hb analyzer trend performance and highlight the need to provide specific education regarding the exclusion zone appropriate to clinical use of any Hb monitor. This exclusion zone could be different in other centers or when using varying analyzers.
The SpHb results can be compared to prior reports of single center studies. A study of volunteers found 95.4% SpHb change agreement in 22 samples with tHb < 10.0 g/dl [15], which is similar to what we found. Using venous blood samples, a study of 70 trauma patients reported similar bias with narrower limits of agreement (− 0.05; − 0.62 to 0.51 g/dl) of SpHb change to tHb change than we report [16]. The number of patients and change calculations is smaller than included herein and reasons for the narrower limits of agreement are not clear, but may be partially explained by the multicenter source of our data. Studies of 48 vascular surgery patients [17] and of 70 patients undergoing major orthopedic surgery [18] concluded SpHb had clinically acceptable trending with ABGHb or tHb. In 49 patients undergoing spine surgery, change concordance was 85.1% when excluding SpHb with perfusion index under 1 [19]. When using the exclusion zones determined by measurement repeatability analysis but not excluding SpHb based on low perfusion index, SpHb concordance was 94.2%. Our findings are also better than reported in 69 patients undergoing spine or cancer surgery that employed a 1 g/dl exclusion zone [20]. A study of patients undergoing abdominal or pelvic surgery found 41 of 269 changes at any Hb had increased SpHb when ABGHb decreased, including 14 in which the decrease was more than − 1 g/dl [11]. Our finding of better trend direction agreement may be related to use of a newer SpHb version in the present study. We also found that Hb accuracy is improved compared to reports using earlier versions of the SpHb monitor [68, 11].
Study limitations include having only about 10% of change samples with tHb ≤ 8.0 g/dl (43 of 416), which impacts our ability to assess clinical utility at very low tHb. However, change direction agreement was good for tHb < 9.0 g/dl. The number of samples in which SpHb, ABGHb or aHQHb change were not more than ± 0.5 g/dl represents a potentially problematic grey zone for assessing Hb change. Although at tHb < 9.0 g/dl only 1 ABGHb and 3 SpHb decreases in this grey zone had associated tHb changes more than − 1.0 g/dl, potential clinical impact of these is not clear and could be evaluated in a future prospective study. We studied SpHb Rev K but newer SpHb versions could potentially have different or better trend accuracy. This could be assessed in a future study. The use of arterial blood can be seen as problematic for generalization to clinical settings in which venous or capillary blood is more commonly sampled. We used arterial blood samples as we studied patients at risk for blood loss and routinely use arterial catheters in such patients to facilitate care and blood sampling. Although venous blood may be easier to obtain in some clinical situations, setting SpHb monitors to arterial mode and using only arterial blood removed potential confounding that a mix of arterial and venous blood samples could have introduced into trend accuracy analyses. Differences between arterial and venous Hb have been reported as 0.2–0.3 g/dl [2123]. We would expect trend calculations to be similar to our findings if only venous blood samples were tested. Use of arterial blood samples to determine aHQHb likely provides different results compared to capillary finger stick blood samples. However, our patients had as many as 13 samples which would have required many finger or ear lobe punctures to allow comparison to capillary samples, thus we chose to use arterial blood samples. Compared to our results, wider limits of agreement between tHb and point of care Hb have been reported when using capillary finger or toe stick samples in patients undergoing cesarean section [24], with gastrointestinal bleeding [25], or in the emergency room [26] and correlated poorly to arterial and venous tHb in intensive care patients [27]. Finally, we did not ascertain whether blood loss was suspected by clinical conditions but the range of tHb change in sequential samples was − 5.0 to + 3.7 g/dl, which reflects blood loss as well as transfusion decisions made by the anesthesiologists caring for the patients.
Hb measurement is essential to making patient centered transfusion decisions [28] which should improve outcomes [29] and reduce transfusion requirements [3032]. Hb measurement is particularly important when blood loss is not obvious or is difficult to estimate during surgery as blood sampling to determine tHb can lag clinical situations. Surgical patients are reported to receive both unnecessary or excessive transfusion [33, 34] and anemia and transfusion can increase perioperative morbidity and mortality [3537] for most but not all patients [38, 39]. However, postoperative outcome is reportedly better using less restrictive transfusion practices following some types of surgery [40], in elderly patients [41, 42] or in patients with cardiovascular disease [43, 44].

5.1 Conclusions

We found that SpHb, ABGHb and aHQHb changes more than ± 0.5 g/dl have similar correlation to the direction but not necessarily the magnitude of tHb change during surgery. The similar agreement in trend direction suggests that clinicians can choose which to use based on availability or preference, although continuous SpHb monitoring may provide useful ongoing Hb trend information. Continuous noninvasive SpHb decreases exceeding − 0.5 g/dl may prompt a decision to obtain a confirmatory tHb measurement if low tHb is clinically suspected, but not replace blood Hb measurement in guiding transfusion decision making. Importantly, this study did not evaluate the transfusion impact of using these monitors so the transfusion impact of continuous noninvasive Hb monitoring needs to be studied prospectively.

Acknowledgements

The authors acknowledge the assistance of Tristan Grogan MS, Principal Statistician, Division of General Internal Medicine and Health Services Research, University of California Los Angeles, and the research assistance of Gerald D. Andrews MBBS and Michael-David Calderon, BS.

Compliance with ethical standards

Conflicts of interest

PP, BL, and PMA have nothing to disclose. RLA, MC and KT disclose each has been principal investigator in studies for which the Departments of Anesthesiology at their institutions received funding from Masimo Corporation and have served as consultants to Masimo Corporation scientific advisory boards. Dr. Torp holds regular shares in Masimo in an amount of less than $50,000. Dr. Cannesson is a consultant for Edwards Lifesciences and Masimo Corp, and has funded research from Edwards Lifesciences and Masimo Corp. He is also the founder of Sironis owns patents and receive royalties for closed loop hemodynamic management that have been licensed to Edwards Lifesciences. Dr. Cannesson’s Department receives funding from the NIH (R01GM117622; R01 NR013012; U54HL119893; 1R01HL144692).
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
2.
Zurück zum Zitat Carson JL, Guyatt G, Heddle NM, Grossman BJ, Cohn CS, Fung MK, Gernsheimer T, Holcomb JB, Kaplan LJ, Katz LM, Peterson N, Ramsey G, Rao SV, Roback JD, Shander A, Tobian AA. Clinical practice guidelines from the AABB: red blood cell transfusion thresholds and storage. JAMA. 2016;316(19):2025–35. https://doi.org/10.1001/jama.2016.9185.CrossRefPubMed Carson JL, Guyatt G, Heddle NM, Grossman BJ, Cohn CS, Fung MK, Gernsheimer T, Holcomb JB, Kaplan LJ, Katz LM, Peterson N, Ramsey G, Rao SV, Roback JD, Shander A, Tobian AA. Clinical practice guidelines from the AABB: red blood cell transfusion thresholds and storage. JAMA. 2016;316(19):2025–35. https://​doi.​org/​10.​1001/​jama.​2016.​9185.CrossRefPubMed
5.
Zurück zum Zitat Zwart A, van Assendelft OW, Bull BS, England JM, Lewis SM, Zijlstra WG. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1995) and specifications for international haemiglobinocyanide standard (4th edition). J Clin Pathol 1996;49 (4):271–274. https://doi.org/10.1136/jcp.49.4.271. Zwart A, van Assendelft OW, Bull BS, England JM, Lewis SM, Zijlstra WG. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1995) and specifications for international haemiglobinocyanide standard (4th edition). J Clin Pathol 1996;49 (4):271–274. https://​doi.​org/​10.​1136/​jcp.​49.​4.​271.
8.
Zurück zum Zitat Hiscock R, Kumar D, Simmons SW. Systematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7 or Pronto-7) and HemoCue(R) absorption spectrometers (B-Hemoglobin or 201 +) with laboratory haemoglobin estimation. Anaesth Intensive Care. 2015;43(3):341–50. https://doi.org/10.1177/0310057X1504300310.CrossRefPubMed Hiscock R, Kumar D, Simmons SW. Systematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7 or Pronto-7) and HemoCue(R) absorption spectrometers (B-Hemoglobin or 201 +) with laboratory haemoglobin estimation. Anaesth Intensive Care. 2015;43(3):341–50. https://​doi.​org/​10.​1177/​0310057X15043003​10.CrossRefPubMed
9.
10.
Zurück zum Zitat American Society of Anesthesiologists Task Force on Perioperative Blood T, Adjuvant T. Practice guidelines for perioperative blood transfusion and adjuvant therapies: an updated report by the American Society of Anesthesiologists Task Force on Perioperative Blood Transfusion and Adjuvant Therapies. Anesthesiology. 2006;105(1):198–208. https://doi.org/10.1097/00000542-200607000-00030.CrossRef American Society of Anesthesiologists Task Force on Perioperative Blood T, Adjuvant T. Practice guidelines for perioperative blood transfusion and adjuvant therapies: an updated report by the American Society of Anesthesiologists Task Force on Perioperative Blood Transfusion and Adjuvant Therapies. Anesthesiology. 2006;105(1):198–208. https://​doi.​org/​10.​1097/​00000542-200607000-00030.CrossRef
14.
Zurück zum Zitat Society of Thoracic Surgeons Blood Conservation Guideline Task F, Ferraris VA, Brown JR, Despotis GJ, Hammon JW, Reece TB, Saha SP, Song HK, Clough ER, Society of Cardiovascular Anesthesiologists Special Task Force on Blood T, Shore-Lesserson LJ, Goodnough LT, Mazer CD, Shander A, Stafford-Smith M, Waters J, International Consortium for Evidence Based P, Baker RA, Dickinson TA, FitzGerald DJ, Likosky DS, Shann KG. 2011 update to the Society of Thoracic Surgeons and the Society of Cardiovascular Anesthesiologists blood conservation clinical practice guidelines. Ann Thorac Surg. 2011;91(3):944–82. https://doi.org/10.1016/j.athoracsur.2010.11.078.CrossRef Society of Thoracic Surgeons Blood Conservation Guideline Task F, Ferraris VA, Brown JR, Despotis GJ, Hammon JW, Reece TB, Saha SP, Song HK, Clough ER, Society of Cardiovascular Anesthesiologists Special Task Force on Blood T, Shore-Lesserson LJ, Goodnough LT, Mazer CD, Shander A, Stafford-Smith M, Waters J, International Consortium for Evidence Based P, Baker RA, Dickinson TA, FitzGerald DJ, Likosky DS, Shann KG. 2011 update to the Society of Thoracic Surgeons and the Society of Cardiovascular Anesthesiologists blood conservation clinical practice guidelines. Ann Thorac Surg. 2011;91(3):944–82. https://​doi.​org/​10.​1016/​j.​athoracsur.​2010.​11.​078.CrossRef
28.
Zurück zum Zitat Mueller MM, Van Remoortel H, Meybohm P, Aranko K, Aubron C, Burger R, Carson JL, Cichutek K, De Buck E, Devine D, Fergusson D, Follea G, French C, Frey KP, Gammon R, Levy JH, Murphy MF, Ozier Y, Pavenski K, So-Osman C, Tiberghien P, Volmink J, Waters JH, Wood EM, Seifried E, Group IPF. Patient blood management: recommendations from the 2018 Frankfurt consensus conference. JAMA. 2019;321(10):983–97. https://doi.org/10.1001/jama.2019.0554.CrossRefPubMed Mueller MM, Van Remoortel H, Meybohm P, Aranko K, Aubron C, Burger R, Carson JL, Cichutek K, De Buck E, Devine D, Fergusson D, Follea G, French C, Frey KP, Gammon R, Levy JH, Murphy MF, Ozier Y, Pavenski K, So-Osman C, Tiberghien P, Volmink J, Waters JH, Wood EM, Seifried E, Group IPF. Patient blood management: recommendations from the 2018 Frankfurt consensus conference. JAMA. 2019;321(10):983–97. https://​doi.​org/​10.​1001/​jama.​2019.​0554.CrossRefPubMed
36.
Zurück zum Zitat Baron DM, Hochrieser H, Posch M, Metnitz B, Rhodes A, Moreno RP, Pearse RM, Metnitz P, European Surgical Outcomes Study group for Trials Groups of European Society of Intensive Care M, European Society of A. Preoperative anaemia is associated with poor clinical outcome in non-cardiac surgery patients. Br J Anaesth. 2014;113(3):416–23. https://doi.org/10.1093/bja/aeu098.CrossRefPubMed Baron DM, Hochrieser H, Posch M, Metnitz B, Rhodes A, Moreno RP, Pearse RM, Metnitz P, European Surgical Outcomes Study group for Trials Groups of European Society of Intensive Care M, European Society of A. Preoperative anaemia is associated with poor clinical outcome in non-cardiac surgery patients. Br J Anaesth. 2014;113(3):416–23. https://​doi.​org/​10.​1093/​bja/​aeu098.CrossRefPubMed
44.
Zurück zum Zitat Docherty AB, O’Donnell R, Brunskill S, Trivella M, Doree C, Holst L, Parker M, Gregersen M, Pinheiro de Almeida J, Walsh TS, Stanworth SJ. Effect of restrictive versus liberal transfusion strategies on outcomes in patients with cardiovascular disease in a non-cardiac surgery setting: systematic review and meta-analysis. BMJ. 2016;352:i1351. https://doi.org/10.1136/bmj.i1351.CrossRefPubMedPubMedCentral Docherty AB, O’Donnell R, Brunskill S, Trivella M, Doree C, Holst L, Parker M, Gregersen M, Pinheiro de Almeida J, Walsh TS, Stanworth SJ. Effect of restrictive versus liberal transfusion strategies on outcomes in patients with cardiovascular disease in a non-cardiac surgery setting: systematic review and meta-analysis. BMJ. 2016;352:i1351. https://​doi.​org/​10.​1136/​bmj.​i1351.CrossRefPubMedPubMedCentral
Metadaten
Titel
Multicenter comparison of three intraoperative hemoglobin trend monitoring methods
verfasst von
Richard L. Applegate II
Patricia M. Applegate
Maxime Cannesson
Prith Peiris
Beth L. Ladlie
Klaus Torp
Publikationsdatum
03.12.2019
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 5/2020
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00428-3

Weitere Artikel der Ausgabe 5/2020

Journal of Clinical Monitoring and Computing 5/2020 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.