1 Introduction
Hemoglobin (Hb) measurement informs patient-specific perioperative transfusion decisions within the context of symptoms, comorbid conditions, surgical procedure, observed bleeding and hemodynamic performance [
1,
2]. Hb measurement is a key component in many parts of patient blood management bundles [
3], and is recommended between transfused red blood cell units if patient stability allows [
4].
Hb measurement may be performed using clinical laboratory hematology (tHb; closest to the cyanmethemoglobin standard [
5]), arterial blood gas cooximetry (ABGHb), or point of care Hb analyzers. However, the time needed for blood sampling and analysis can cause Hb measurement to lag clinical situations. In surgical settings in which blood loss may not be apparent or be difficult to estimate, continuous rather than intermittent Hb monitoring could provide earlier warning of decreasing Hb.
Multiwave pulse cooximetry noninvasively determines total hemoglobin (SpHb). SpHb has been tested in volunteers [
6] and in a wide range of clinical settings [
7]. Prior reports indicated wide limits of agreement between SpHb and tHb that suggest caution when using SpHb alone to guide transfusion decisions [
7,
8]. However, it is possible that SpHb changes could provide useful information if the direction of SpHb change accurately reflects the direction of tHb change (trend). We defined trend accuracy as agreement of sequential changes in SpHb, ABGHb, or point of care Hb with sequential tHb changes. The aim of this multicenter study was to evaluate trend accuracy of three monitoring methods in patients undergoing surgery.
2 Methods and materials
This collaborative prospective convenience sample observational study received Institutional Review Board approval at each of three USA academic medical centers prior to study initiation: Loma Linda University (LLU), Loma Linda, CA; Mayo Clinic in Florida (MCF), Jacksonville, FL; and University of California Irvine (UCI), Irvine, CA. Patient consent was obtained according to local IRB determination at each site. All study procedures were performed in accordance with the ethical standards of the institutional and research committees and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. This manuscript adheres to applicable STARD guidelines.
Adult patients were eligible for study participation if scheduled to undergo non-cardiac surgery in which arterial catheterization and expected repeated intraoperative blood gas analyzes were planned as part of patient care. Patients were excluded if pregnant or for skin abnormalities at the planned application site that would interfere with pulse oximetry (burns, scar tissue, nail polish, or infection). Attending anesthesiologists had discretion over anesthesia management, fluid administration and transfusion decisions, which were based on clinical settings (surgical considerations, ongoing bleeding, hemodynamic condition and/or any patient comorbid conditions) and institutional guidelines, which were in keeping with published guidelines [
9,
10].
Multiwave disposable pulse cooximetry finger sensors (R125, Radical-7, Revision K, Masimo, Irvine, CA) were placed on the ring finger on the side of arterial cannulation with data continuously collected to computer. Oximeters were set to arterial mode to align with the source of blood samples. The SpHb algorithm continuously evaluates up to 6 min of data to calculate the displayed value. Whenever arterial blood gas analysis was performed, SpHb displayed at the time arterial blood was drawn was recorded, and arterial blood samples were obtained in appropriate collection tubes. Blood sample analysis was completed within 10 min. Each blood sample was analyzed twice using the same analyzer for tHb, ABGHb and aHQHb using:
-
Clinical Laboratory hematology analyzer Hb (tHb; LLU—Sysmex XE5000, Sysmex America Inc., Lincolnshire, IL, USA; MCF—Sysmex XE5000, Sysmex America Inc., Lincolnshire, IL, USA or Coulter AcTdiff, Beckman Coulter, Indianapolis, IN, USA. UCI—Coulter LH 750 Hematology Analyzer, Beckman Coulter, Brea, CA, USA)
-
Arterial blood gas cooximetry Hb (ABGHb; LLU—Radiometer ABL800; Radiometer, Copenhagen, Denmark. MCF—CCX or PhOX, Nova Biomedical, Waltham, MA, USA; UCI—Siemens RAPIDLab 1265, Siemens Healthcare Diagnostics, NY, USA)
-
Point of care Hb using arterial blood (aHQHb; not capillary or finger stick samples; HemoCue HB 301, HemoCue America, Brea, CA, USA)
2.1 Statistical methods
The primary outcome measure was trend accuracy using modified Bland–Altman analysis of difference between changes in SpHb, ABGHb or aHQHb and tHb changes to tHb changes as the measurement standard, to obtain bias and 95% limits of agreement between trends in SpHb, ABGHb, or aHQHb and tHb trend. Trends were defined as sequential change in Hb using results of the first analysis performed on each blood sample: Hb sample 2—Hb sample 1; Hb sample 3—Hb sample 2; etc. We included in analysis only samples for which all measures were available for trend calculations and did not exclude or separately analyze SpHb in low perfusion states.
Power and sample size calculation: A prior study of patients undergoing similar procedures [
11] reported a median of 4 (up to 9) blood samples allowing an average of 3 trend calculations per patient. The mean bias of SpHb to tHb reported in prior studies in which the intended hematology analyzers were used [
7] ranged from − 0.53 to 1.22, with standard deviation averaging 1.055. Using that standard deviation, setting alpha to 0.05 and power to 0.8, a sample size of 135 patients would be needed to estimate the precision of the 95% confidence interval of the difference among the 3 methods within 0.2 g/dl This is a conservative estimate of our true precision because patients would have repeated blood draws. Power analyses were performed using SAS 9.4 (SAS Institute, Cary, North Carolina, USA); statistical analyses were performed using JMP Pro version 13.2.0 (SAS Institute, Cary, North Carolina, USA) and Prism 8.1.0 (GraphPad Software, San Diego, California, USA).
5 Discussion
Intraoperative Hb trend accuracy limits of agreement were approximately ± 1 g/dl for ABGHb and aHQHb and slightly larger for SpHb trends. Considering the 95% confidence intervals for agreement in change direction overlapped, our findings suggest that changes more than ± 0.5 g/dl in SpHb, ABGHb and aHQHb provide similar information regarding the direction of intraoperative tHb change, but not necessarily the magnitude of that change. The impact of trend accuracy on transfusion decision-making is more significant when tHb is less than 9.0 g/dl as current guidelines suggest that red blood cell transfusion may be reasonable [
14] for tHb < 8.0 g/dl, but usually unnecessary [
9] for tHb > 10.0 g/dl. Analysis of change direction agreement suggests that if continuous SpHb is < 9 g/dl then a decrease more than − 0.5 g/dl could be a good indication to obtain a blood sample for tHb measurement.
Repeatability analysis demonstrated limits of agreement that have implications for Hb accuracy and trend monitoring studies. The exclusion zones we defined based on 95% limits of agreement for ABGHb and aHQHB suggest that in our patients, a change up to ± 0.5 g/dl may not reflect a real change in circulating hemoglobin measured by tHb. Similarly, reported tHb changes of ± 0.25 g/dl may not represent real changes in circulating Hb. Our findings support use of an exclusion zone when comparing Hb analyzer trend performance and highlight the need to provide specific education regarding the exclusion zone appropriate to clinical use of any Hb monitor. This exclusion zone could be different in other centers or when using varying analyzers.
The SpHb results can be compared to prior reports of single center studies. A study of volunteers found 95.4% SpHb change agreement in 22 samples with tHb < 10.0 g/dl [
15], which is similar to what we found. Using venous blood samples, a study of 70 trauma patients reported similar bias with narrower limits of agreement (− 0.05; − 0.62 to 0.51 g/dl) of SpHb change to tHb change than we report [
16]. The number of patients and change calculations is smaller than included herein and reasons for the narrower limits of agreement are not clear, but may be partially explained by the multicenter source of our data. Studies of 48 vascular surgery patients [
17] and of 70 patients undergoing major orthopedic surgery [
18] concluded SpHb had clinically acceptable trending with ABGHb or tHb. In 49 patients undergoing spine surgery, change concordance was 85.1% when excluding SpHb with perfusion index under 1 [
19]. When using the exclusion zones determined by measurement repeatability analysis but not excluding SpHb based on low perfusion index, SpHb concordance was 94.2%. Our findings are also better than reported in 69 patients undergoing spine or cancer surgery that employed a 1 g/dl exclusion zone [
20]. A study of patients undergoing abdominal or pelvic surgery found 41 of 269 changes at any Hb had increased SpHb when ABGHb decreased, including 14 in which the decrease was more than − 1 g/dl [
11]. Our finding of better trend direction agreement may be related to use of a newer SpHb version in the present study. We also found that Hb accuracy is improved compared to reports using earlier versions of the SpHb monitor [
6‐
8,
11].
Study limitations include having only about 10% of change samples with tHb ≤ 8.0 g/dl (43 of 416), which impacts our ability to assess clinical utility at very low tHb. However, change direction agreement was good for tHb < 9.0 g/dl. The number of samples in which SpHb, ABGHb or aHQHb change were not more than ± 0.5 g/dl represents a potentially problematic grey zone for assessing Hb change. Although at tHb < 9.0 g/dl only 1 ABGHb and 3 SpHb decreases in this grey zone had associated tHb changes more than − 1.0 g/dl, potential clinical impact of these is not clear and could be evaluated in a future prospective study. We studied SpHb Rev K but newer SpHb versions could potentially have different or better trend accuracy. This could be assessed in a future study. The use of arterial blood can be seen as problematic for generalization to clinical settings in which venous or capillary blood is more commonly sampled. We used arterial blood samples as we studied patients at risk for blood loss and routinely use arterial catheters in such patients to facilitate care and blood sampling. Although venous blood may be easier to obtain in some clinical situations, setting SpHb monitors to arterial mode and using only arterial blood removed potential confounding that a mix of arterial and venous blood samples could have introduced into trend accuracy analyses. Differences between arterial and venous Hb have been reported as 0.2–0.3 g/dl [
21‐
23]. We would expect trend calculations to be similar to our findings if only venous blood samples were tested. Use of arterial blood samples to determine aHQHb likely provides different results compared to capillary finger stick blood samples. However, our patients had as many as 13 samples which would have required many finger or ear lobe punctures to allow comparison to capillary samples, thus we chose to use arterial blood samples. Compared to our results, wider limits of agreement between tHb and point of care Hb have been reported when using capillary finger or toe stick samples in patients undergoing cesarean section [
24], with gastrointestinal bleeding [
25], or in the emergency room [
26] and correlated poorly to arterial and venous tHb in intensive care patients [
27]. Finally, we did not ascertain whether blood loss was suspected by clinical conditions but the range of tHb change in sequential samples was − 5.0 to + 3.7 g/dl, which reflects blood loss as well as transfusion decisions made by the anesthesiologists caring for the patients.
Hb measurement is essential to making patient centered transfusion decisions [
28] which should improve outcomes [
29] and reduce transfusion requirements [
30‐
32]. Hb measurement is particularly important when blood loss is not obvious or is difficult to estimate during surgery as blood sampling to determine tHb can lag clinical situations. Surgical patients are reported to receive both unnecessary or excessive transfusion [
33,
34] and anemia and transfusion can increase perioperative morbidity and mortality [
35‐
37] for most but not all patients [
38,
39]. However, postoperative outcome is reportedly better using less restrictive transfusion practices following some types of surgery [
40], in elderly patients [
41,
42] or in patients with cardiovascular disease [
43,
44].
5.1 Conclusions
We found that SpHb, ABGHb and aHQHb changes more than ± 0.5 g/dl have similar correlation to the direction but not necessarily the magnitude of tHb change during surgery. The similar agreement in trend direction suggests that clinicians can choose which to use based on availability or preference, although continuous SpHb monitoring may provide useful ongoing Hb trend information. Continuous noninvasive SpHb decreases exceeding − 0.5 g/dl may prompt a decision to obtain a confirmatory tHb measurement if low tHb is clinically suspected, but not replace blood Hb measurement in guiding transfusion decision making. Importantly, this study did not evaluate the transfusion impact of using these monitors so the transfusion impact of continuous noninvasive Hb monitoring needs to be studied prospectively.