Multicentre analysis of PET SUV using vendor-neutral software: the Japanese Harmonization Technology (J-Hart) study

verfasst von: Yuji Tsutsui, Hiromitsu Daisaki, Go Akamatsu, Takuro Umeda, Matsuyoshi Ogawa, Hironori Kajiwara, Shigeto Kawase, Minoru Sakurai, Hiroyuki Nishida, Keiichi Magota, Kazuaki Mori, Masayuki Sasaki, J-Hart study group

Erschienen in: EJNMMI Research | Ausgabe 1/2018

Abstract

Background

Recent developments in hardware and software for PET technologies have resulted in wide variations in basic performance. Multicentre studies require a standard imaging protocol and SUV harmonization to reduce inter- and intra-scanner variability in the SUV. The Japanese standardised uptake value (SUV) Harmonization Technology (J-Hart) study aimed to determine the applicability of vendor-neutral software on the SUV derived from positron emission tomography (PET) images. The effects of SUV harmonization were evaluated based on the reproducibility of several scanners and the repeatability of an individual scanner.
Images were acquired from 12 PET scanners at nine institutions. PET images were acquired over a period of 30 min from a National Electrical Manufacturers Association (NEMA) International Electrotechnical Commission (IEC) body phantom containing six spheres of different diameters and an 18F solution with a background activity of 2.65 kBq/mL and a sphere-to-background ratio of 4. The images were reconstructed to determine parameters for harmonization and to evaluate reproducibility. PET images with 2-min acquisition × 15 contiguous frames were reconstructed to evaluate repeatability. Various Gaussian filters (GFs) with full-width at half maximum (FWHM) values ranging from 1 to 15 mm in 1-mm increments were also applied using vendor-neutral software. The SUVmax of spheres was compared with the reference range proposed by the Japanese Society of Nuclear Medicine (JSNM) and the digital reference object (DRO) of the NEMA phantom. The coefficient of variation (CV) of the SUVmax determined using 12 PET scanners (CVrepro) was measured to evaluate reproducibility. The CV of the SUVmax determined from 15 frames (CVrepeat) per PET scanner was measured to determine repeatability.

Results

Three PET scanners did not require an additional GF for harmonization, whereas the other nine required additional FWHM values of GF ranging from 5 to 9 mm. The pre- and post-harmonization CVrepro of six spheres were (means ± SD) 9.45% ± 4.69% (range, 3.83–15.3%) and 6.05% ± 3.61% (range, 2.30–10.7%), respectively. Harmonization significantly improved reproducibility of PET SUVmax (P = 0.0055). The pre- and post-harmonization CVrepeat of nine scanners were (means ± SD) 6.59% ± 1.29% (range, 5.00–8.98%) and 4.88% ± 1.64% (range, 2.65–6.72%), respectively. Harmonization also significantly improved the repeatability of PET SUVmax (P < 0.0001).

Conclusions

Harmonizing SUV using vendor-neutral software produced SUVmax for 12 scanners that fell within the JSNM reference range of a NEMA body phantom and improved SUVmax reproducibility and repeatability.
Abkürzungen
CV
Coefficient of variation
DICOM
Digital Imaging and Communications in Medicine
DRO
Digital reference object
FDG
Fluorodeoxyglucose
FWHM
Full-width at half maximum
GF
Gaussian filter
IEC
International Electrotechnical Commission
J-Hart study
Japanese SUV Harmonization Technology study
JSNM
Japanese Society of Nuclear Medicine
NEMA
National Electrical Manufacturers Association
OSEM
Ordered subset expectation maximization
PERCIST
PET Response Criteria in Solid Tumours
PET/CT
Positron emission tomography/computed tomography
PSF
Point spread function
QIBA
Quantitative Imaging Biomarker Alliance
RC
Recovery coefficients
RMSE
Root mean square error
ROI
Region of interest
RSNA
Radiological Society of North America
SD
Standard deviation
SUV
Standardised uptake value
TOF
Time of flight

Background

18F-fluoro-2-deoxy-2-d-glucose (18F-FDG) positron emission tomography (PET) is a valuable imaging tool for the diagnosis, staging and assessment of the responses of various malignancies to therapy [15]. Reports indicate that 18F-FDG PET is a more effective biomarker of treatment responses than morphological information because it measures glucose metabolism [69]. The standardised uptake value (SUV) has served as a semi-quantitative metric of 18F-FDG uptake by tumours.
The SUV must be highly reproducible and repeatable to serve as a reliable biomarker in multicentre trials using various PET scanners, because SUV variability among scanners can contribute to uncertainty in results. The Quantitative Imaging Biomarker Alliance (QIBA), which was organised by the Radiological Society of North America (RSNA), recommended definitions for reproducibility and repeatability [10, 11]. In short, reproducibility refers to the consistency of values derived from repeated tests of one individual by different operators, different scanners, software, or at different sites and times. Repeatability refers to the consistency of values derived from repeated tests of one individual by a single operator using the same scanner and software. That is, reproducibility refers to inter-scanner variability, whereas repeatability refers to intra-scanner variability.
Recent developments in hardware and software for PET technologies have resulted in wide variations in basic performance. The SUV considerably varies due to biological and technical factors involving the model of the PET scanner, acquisition protocols, reconstruction algorithm and parameters [1215]. For example, ordered subset expectation maximization (OSEM) with a point spread function (PSF) algorithm improves the spatial resolution, but it causes edge artifacts as overestimation [16, 17]. Therefore, multicentre studies require a standard imaging protocol and SUV harmonization to reduce inter- and intra-scanner variability in the SUV.
Several organizations have published guidelines to standardise 18F-FDG PET imaging protocols and harmonize SUV harmonization [14, 1820]. In terms of SUV harmonization, the European Association of Nuclear Medicine (EANM) and the Japanese Society of Nuclear Medicine (JSNM) have proposed a specified range of recovery coefficients (RC) as a function of sphere size [21]. “Harmonization” is defined herein as a PET image smoothed with an additional Gaussian filter (GF) so that the RC of each sphere falls within the specified range of the body phantom. This harmonization strategy minimises variations in SUV measurements during scan acquisition and processing [2224] and reduces reconstruction-dependent variability in the PET Response Criteria in Solid Tumours (PERCIST) classification [25].
The usefulness of commercially available software for harmonization was first determined using EQ.PET software [23]. However, a costly, software-specific workstation provided by the Siemens Healthineers is required. The recently commercially available GI-PET software (AZE VirtualPlace Hayabusa, Tokyo, Japan) adjusts the RC of PET images into a reference RC using an additional GF. This vendor-neutral quantitative software can be installed in general personal computers.
We designed the Japanese SUV Harmonization Technology (J-Hart) study based on this background to determine the applicability of vendor-neutral software to multicentre PET SUV harmonization. We also assessed the effects of SUV harmonization on SUV reproducibility and repeatability.

Methods

Phantom

We used a NEMA 2001 International Electrotechnical Commission (IEC) body phantom (Data Spectrum Corp., Durham, NC, USA), consisting of six spheres (Model PET/IEC–BODY/P) of 10, 13, 17, 22, 28 and 37 mm in diameter, with a wall thickness of 1 mm in a quasi-cylindrical cavity (280 × 210 × 180 mm). All spheres and the background were filled with solutions containing 10.6 and 2.65 kBq/mL (at the midpoint of 30-min acquisitions), respectively, of 18F-FDG to obtain a sphere-to-background ratio of 4.

PET/CT scanners

Data were acquired at nine Japanese institutions using 12 popular PET systems (Table 1): Discovery ST Elite Performance (DSTEP), Discovery ST Elite (DSTE), Discovery 600 Motion (D600), Discovery 690 (D690) and Discovery IQ (DIQ) scanners (General Electric Medical Systems, Milwaukee, WI, USA); Biograph 64 True Point (BioTP), Biograph mCT 3ring (Bio3R) and Biograph mCT flow 4ring (Bio4R) scanners (Siemens Healthineers, Erlangen, Germany); GEMINI-TF16 (GTF), Gemini GXL (GXL) scanners (Philips Medical Systems, Cleveland, OH, USA); Aquiduo (Aquiduo) and Celesteion PCA-9000A (Celesteion) scanners (Toshiba Medical Systems, Otawara, Japan). The collaborating institutions comprised five university hospitals, three cancer centres and one research institution that normally conduct many oncological PET studies. The dose calibrator and each PET scanner of all evaluated PET/CT systems were cross-calibrated before data acquisition from the phantom according to the guidelines [19].
Table 1
Clinical parameters of PET scanners
 
Aquiduo
(Aquiduo)
TOSHIBA
Biograph mCT 3 ring
(Bio3R)
SIEMENS
Biograph mCT Flow 4 ring
(Bio4R)
SIEMENS
Biograph 64 True Point
(BioTP)
SIEMENS
Celesteion PCA-9000A
(Celesteion)
TOSHIBA
Discovery IQ
(DIQ)
GE
Discovery ST Elite
(DSTE)
GE
Discovery STEP
(DSTEP)
GE
Discovery 600 Motion
(D600)
GE
Discovery 690
(D690)
GE
GEMINI TF16
(GTF)
Philips
Gemini GXL
(GXL)
Philips
Cross-calibration (background SUV)
1.03
1.03
0.96
1.01
1.00
1.04
0.96
1.03
1.04
1.01
1.04
1.03
PET reconstruction
 Reconstruction
FORE
OSEM
3D-OSEM
PSF
TOF
3D-OSEM
PSF
TOF
3D-OSEM
PSF
3D-OSEM
3D-OSEM
PSF
3D-OSEM
3D-OSEM
3D-OSEM
3D-OSEM
TOF
Full list mode
TOF
3D-OSEM
LOR
RAMLA
 Iterations
4
2
2
2
3
4
2
2
3
3
3
2
 Subsets
14
21
21
21
10
12
28
21
16
8
33
 Smoothing
Gaussian
Gaussian
Gaussian
Gaussian
Gaussian
Gaussian
Gaussian
Gaussian
Gaussian
Gaussian
Smooth/sharp; smooth A
Smooth/sharp; normal
 FWHM of filter (mm)
8
6
4
4
6
5
6
5.14
4
4
FORE Fourier Rebinning, OSEM ordered subset expectation maximization, PSF point spread function, TOF time of flight, LOR line of response, RAMLA row action maximum likelihood algorithm, FWHM full-width at half maximum

Region of interest

The SUVmax of the six hot spheres was determined from circular regions of interest (ROIs) placed on the centre slice of images of each sphere. The diameter of each ROI was equal to that of each hot sphere (10, 13, 17, 22, 28 and 37 mm). Twelve circular ROIs with a diameter of 10 mm were placed on the background region at the centre slice of the PET image; the average SUV was calculated from twelve ROIs as the SUVmean of the background. The acceptance criterion for the SUVmean of the background region was 1.00 ± 0.05. Table 1 shows the average SUVmean of the background region after cross-calibration.

Digital reference object

The QIBA suggested that computational imaging models or phantom data might play an important role in reproducibility assessment [26] and thus developed an 18F-FDG PET/CT digital reference object (DRO), which is a synthetic test object in Digital Imaging and Communications in Medicine (DICOM) format [27]. We obtained the DRO from the QIBA [28]. The mathematically developed DRO of the NEMA body phantom is an ideal object with a uniform background with a transaxial diameter of 20 cm that simulates a human abdominal cross section with an SUV of 1.00. Six spheres with diameters of 10, 13, 17, 22, 28 and 37 mm contained an 18F-FDG solution with an SUV of 4.00. The SUV of a central cylinder with a diameter of 5 cm was 0.00. A three-dimensional (3-D) GF was applied to the DRO to simulate a PET image with relatively low spatial resolution. The RC of the DRO was included in the JSNM reference range for a GF with a full-width at half maximum (FWHM) value of 10–13 mm, and the RC was highest at a 10-mm FWHM of GF (Fig. 1). We adopted DRO10mm as the reference SUV (SUVref) for harmonization.

Data acquisition and processing

Emission data were acquired from the phantom over a period of 30 min using list mode with the bed in position 1. A PET image with 30-min acquisition was reconstructed to determine the parameters for harmonization and to evaluate reproducibility. PET images with 2-min acquisition × 15 contiguous frames were reconstructed to evaluate the repeatability. The reconstruction parameters were individually determined according to the clinical setting at each institution (Table 1). All reconstructed PET images were output in DICOM format.

Harmonization

We used GI-PET software on a personal computer for harmonization processing. Stand-alone GI-PET software can quantify PET images and it is typically used to adjust spatial resolution to harmonize PET images using a 3-D GF.
The harmonization procedure of GI-PET was as follows: Firstly, reconstructed PET images with 30-min acquisitions in DICOM format were loaded into the GI-PET software. The original voxels were converted into isotropic voxels of an equivalent size in the x, y and z directions using the bi-linear method because the 3D-GF of GI-PET requires isotropic data. A 3-D GF with various FWHM values of 1–15 mm in 1-mm increments was applied to the PET images, and then, a graph of the SUVmax of the spheres was generated as a function of the diameter. The FWHM of the GF that provided an SUVmax within the reference range proposed by the JSNM [21] was determined. The reference ranges of the lower to the upper limits of the SUVmax on 10-, 13-, 17-, 22-, 28- and 37-mm hot spheres were 1.19–2.00, 1.52–3.04, 2.58–3.71, 3.25–4.09, 3.56–4.21 and 3.82–4.17, respectively. If only one FWHM provided an SUVmax within the reference range, then that FWHM was taken as the optimal parameter of GF for harmonization. On the other hand, if several FWHM provided SUVmax within the reference range, we compared the root mean square error (RMSE) with DRO10mm. The RMSE is the square root of the variance in the SUVmax of the six spheres between PET images and DRO10mm and is calculated as:
$$ \mathrm{RMSE}=\sqrt{\frac{1}{6}\sum \limits_{i=10,13,17,22,28,37\ \mathrm{mm}}{\left({\mathrm{SUV}}_{\max\ \mathrm{in}\ \mathrm{PET}\ \mathrm{scanner},i}-{\mathrm{SUV}}_{\max\ \mathrm{in}\ \mathrm{DRO}10\mathrm{mm},i}\right)}^2} $$
Finally, we determined the optimal FWHM that provided the smallest RMSE (Fig. 2). RSME is a simple mathematical measure that has been used extensively in nuclear medicine and molecular imaging for several years. Combining RSME with DRO is reportedly useful to create appropriate reconstruction conditions for other types of nuclear medicine image [29].

Reproducibility and repeatability

We calculated the coefficients of variation (CV) of the SUVmax on images of spheres with different diameters acquired over a period of 30 min by 12 PET scanners to evaluate the reproducibility of quantitation and defined the CV across the scanners as CVrepro. We then compared CVrepro between the pre- and post-harmonized images calculated as the mean ± standard deviation (SD). The CV is the ratio of the standard deviation (SD) of the mean calculated as:
$$ \mathrm{CV}=\raisebox{1ex}{$\mathrm{SD}$}\!\left/ \!\raisebox{-1ex}{$\mathrm{mean}$}\right.\times 100\ \left(\%\right) $$
We calculated the CV of SUVmax on PET images with 2-min acquisitions × contiguous 15 frames for each PET scanner and each sphere size to evaluate the repeatability of quantitation and then defined the CV across the frames as CVrepeat. We then compared CVrepeat between the pre- and post-harmonized images calculated as the mean ± SD.

Statistical analyses

Differences in CVrepro and CVrepeat were compared among six and 54 pairs, respectively, between pre- and post-harmonization by Wilcoxon signed-rank tests using JMP® 13 (SAS Institute Inc., Cary, NC, USA). P values < 0.05 were considered statistically significant.

Results

Harmonization

The SUVmax fell within the JSNM reference range in three PET scanners (DSTE, DSTEP and GXL) without adding a GF (Table 2). Thus, adjustment was not required for harmonization. The FWHM values of the additional GF that resulted in an SUVmax within the JSNM reference range were 5–7, 5–9, 8–10, 9, 6–10, 8–9, 6–9, 6–9 and 5–7 mm for the Aquiduo, Bio3R, Bio4R, BioTP, Celesteion, DIQ, D600, D690 and GTF, respectively. The RMSE compared with DRO10mm was the smallest at FWHM values of 5, 5, 8, 9, 6, 8, 6, 6 and 5 mm, respectively (Table 2). The SUVmax of these nine scanners fell within the JSNM reference range only after harmonization. Figure 3 shows the pre- and post-harmonization RC of 12 PET scanners.
Table 2
Full-width at half maximum for SUV harmonization of PET scanners
PET scanner
Range of FWHM for GF for inclusion in JSNM reference range (mm)
Optimum FWHM for GF to obtain smallest RMSE compared with DRO10mm (mm)
Aquiduo
5–7
5
Bio3R
5–9
5
Bio4R
8–10
8
BioTP
9
9
Celesteion
6–10
6
DIQ
8, 9
8
DSTE
No filter
No filter
DSTEP
No filter
No filter
D600
6–9
6
D690
6–9
6
GTF
5–7
5
GXL
No filter
No filter
DRO digital reference object, FWHM full-width at half maximum, GF Gaussian filter, RMSE root mean square error

Reproducibility

Figure 4 shows the pre- and post-harmonization CVrepro of the 30-min PET images among 12 PET scanners. The (means ± SD) pre- and post-harmonization CVrepro of six spheres were 9.45% ± 4.69% (range, 3.83–15.3%) and 6.05% ± 3.61% (range, 2.30–10.7%), respectively. Harmonization significantly improved the reproducibility of the PET SUVmax (P = 0.0055).

Repeatability

Figure 5 shows the CVrepeat of images across 15 frames acquired from each PET scanner. The post-harmonization CVrepeat of the nine PET scanners that required an additional GF for harmonization (Aquiduo, Bio3R, Bio4R, BioTP, Celesteion, DIQ, D600, D690 and GTF) was lower than that at pre-harmonization (Table 3). The pre- and post-harmonization CVrepeat (means ± SD) of nine scanners were 6.59% ± 1.29% (range, 5.00–8.98%) and 4.88% ± 1.64% (range, 2.65–6.72%), respectively. Harmonization significantly improved the repeatability of the PET SUVmax (P < 0.0001).
Table 3
Pre- and post-harmonization coefficients of variation across 15 frames of 12 PET scanners
 
CVrepeat (%)
Pre-harmonization
Post-harmonization
Aquiduo
6.03 ± 1.54
5.15 ± 1.21
Bio3R
7.54 ± 0.95
6.72 ± 0.77
Bio4R
7.06 ± 1.48
5.89 ± 1.01
BioTP
8.98 ± 1.83
6.64 ± 0.69
Celesteion
5.00 ± 2.28
3.09 ± 1.32
DIQ
6.20 ± 2.33
2.65 ± 0.92
DSTE
7.51 ± 2.13
DSTEP
6.30 ± 0.87
D600
5.66 ± 1.65
3.99 ± 0.84
D690
5.31 ± 1.47
3.33 ± 0.90
GTF
7.57 ± 1.06
6.48 ± 0.69
GXL
4.86 ± 1.35
CV coefficient of variation

Discussion

The J-Hart study examined whether a vendor-neutral software could harmonize SUV across 12 PET scanners from various manufacturers. We obtained an appropriate FWHM of GF for harmonization by comparing the SUVmax of 12 PET scanners at nine institutions with both the JSNM reference range and the DRO10mm. Harmonization significantly improved the reproducibility and repeatability of the SUVmax.
Our harmonization process was based on the simple down-smoothing method proposed by Boellaard et al. [30]. This strategy has been widely applied to SUV harmonization [15, 31] and provides a pair of PET images with relatively high and low resolution. With a comparable SUV across different scanners as well as in EQ.PET software, not only high-resolution images for lesion detection but also harmonized images can be obtained [23]. The EANM provides the strategy for harmonization applied to RC [19], whereas JSNM recommends harmonization applied to SUVmax that can detect errors such as cross-calibration. Therefore, SUVmax data are presented in graphs of the present study.
Quak et al. validated a proprietary software tool (EQ.PET) to harmonize SUV across three PET systems independently of the applied reconstruction algorithm [23]. They reported that all RC of SUVmax fell within EANM curve after applying the EQ.PET filter. The present study confirmed that vendor-neutral GI-PET software could harmonize the SUVmax across 12 PET scanners. As a result, all SUVmax fell within in the JSNM reference range on GI-PET after adding a GF. Therefore, we consider that GI-PET and EQ.PET are equally useful for harmonization. The post-harmonization SUVmax of various PET scanners became comparable based on the same standard, which is relevant to multicentre and follow-up studies that use various PET scanners. E Q.PET is a proprietary software that can only be installed on its own dedicated workstation with PET system. In contrast, GI-PET is a vendor-neutral software that can be applied at any institution using only a personal computer, regardless of the PET system. Although the present study used the JSNM reference range of SUVmax, the EANM reference range can be used in vendor-neutral software.
Harmonization of the SUV improved the CVrepro of SUVmax across 12 scanners in the present study. Takahashi et al. reported a maximum difference of 45.7% in the SUVmax with a 27-mm sphere across five different PET scanners [32]. Such a substantial variation is a problem for reproducibility. Quak et al. evaluated harmonization across three PET scanners using EQ.PET for 1380 FDG-avid tumours [23]. They reported that the mean ratio with and without PSF of the SUVmax for 1380 tumour lesions was 1.46 (95%CI 0.86–2.06) before and 1.02 (95%CI 0.88–1.16) after harmonization. Because the mean value, + 1.96 × SD, was the upper limit of the CI, the SD before harmonization was 0.31, thus yielding a CV of 21%. The SD after harmonization was 0.07, thus yielding a CV of 7%. The present study found that GI-PET improved the range of CVrepro across 12 PET scanners from 3.83–15.3% to 2.30–10.7%, suggesting that inter-scanner variability was improved equally well by GI-PET and EQ.PET. Harmonization is thought to decrease differences in variations among acquisition protocols (2D, 3D, TOF, etc.), machine specificities (such as type of crystal) and other elements (clock synchronisation, activity preparation and injection). However, their study included 517 patients with tumours who were examined using three PET scanners, whereas we examined a NEMA IEC body phantom using 12 PET scanners. The difference may also be affected by the biological features of tumours and patients. Patients should ideally be clinically assessed using the same scanner. However, patients often undergo PET evaluations with different scanners within the same institution. Multicentre studies usually have different scanners, and thus, therapeutic effects can be assessed using various scanners, and patients are often transferred to other hospitals for advanced specialist care.
Harmonization of the SUV also improved the CVrepeat of SUVmax across 15 frames in each scanner. Velasquez et al. evaluated SUV repeatability in a study of 18F-FDG PET and found that the intra-subject CV for SUVmax was 10.7% [33]. Doot et al. evaluated the influence of the FWHM of a GF on the repeatability of phantom data and found that increase of smoothing parameter decreased the standard deviation (SD) of the RC [34]. Kelly et al. evaluated the effects of harmonization on quantitative variation in the SUVmax relative to the reconstruction protocol. They found that the CV of the SUVmax across 15 repeat scans pre-harmonization was 2.81%, 3.25% and 4.69% for OSEM, OSEM+PSF and OSEM+PSF+ time of flight (TOF), respectively, whereas those values after harmonization were 2.28%, 2.00% and 2.58%, respectively [31]. Harmonization is considered to improve the repeatability of any reconstruction protocol.
The CVrepeat of all scanners and spheres was < 10% except for that with the 10-mm sphere using the DSTE. With respect to CV repeatability, Lodge et al. reported that the within-subject CV of a tumour SUV was 10% when acquired with careful attention to the protocol [35]. Data acquisition with the DSTE for clinical examinations requires 3 min per bed position, whereas our phantom study acquired data for 2 min per frame in the evaluation of repeatability. This might be why the CVrepeat of the 10-mm sphere using DSTE was > 10%.
The SUVmax of all PET scanners in the present study was above the lower limit of the reference SUV range. Lasnon et al. examined the RC using Biograph TrueV and found that the RC for a 10-mm sphere on OSEM images was slightly below the lower limit of the EANM proposed range [15]. Harmonization using a GF is available when the RC is above the lower limit of the reference range, and it cannot proceed when the RC is below this limit. However, since some older PET and PET/CT findings of low spatial resolution might have an SUVmax below the reference range, the reconstruction settings (including reconstruction methods, matrix, pixel size, iterations, subsets and post-filter) should be carefully considered to obtain an RC within the reference range.
Several limitations are associated with the present study. The basic performance of PET/CT has improved in recent years as the technology has progressed. Therefore, the reference range of the SUV might change with future advancements in PET/CT. We only assessed the SUVmax, although other metrics, such as metabolic tumour volume and total lesion glycolysis, are also useful [36]. The reproducibility and repeatability of new metrics that might be reliable biomarkers should be evaluated. Differences in the voxel size depending on the reconstruction protocol require consideration. GI-PET automatically converts the original voxel size to the isotropic voxel size, which might result in the SUV differing from the value obtained on a proprietary workstation. A vendor-neutral software for SUV harmonization should be able to apply a 3D-GF to original (non-isotropic voxel) data. The present study is evaluated using only the phantom. Therefore, the feasibility of vendor-neutral software in clinical practice requires validation by further clinical studies.

Conclusions

We harmonized quantitative values among 12 PET scanners using a commercial vendor-neutral software. This harmonization strategy based on the simple down-smoothing method improved the reproducibility and the repeatability of the SUV determined from a hot lesion. Therefore, this software might enable comparison of SUV directly across different scanners and facilitate multicentre oncology PET studies.

Acknowledgements

The authors thank the staff of Kyushu University Hospital, National Institute of Radiological Sciences, Cancer Institute Hospital, Yokohama City University Hospital, National Center for Global Health and Medicine, Kyoto University Hospital, Clinical Imaging Center for Healthcare, Nippon Medical School, Institute of Biomedical Research and Innovation, Hokkaido University Hospital and Toranomon Hospital for their valuable support.

Availability of data and materials

Please contact the author for data requests.

Ethics approval

Not applicable.
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Literatur
1.
Zurück zum Zitat Fletcher JW, Djulbegovic B, Soares HP, Siegel BA, Lowe VJ, Lyman GH, et al. Recommendations on the use of 18F-FDG PET in oncology. J Nucl Med. 2008;49:480–508.CrossRefPubMed Fletcher JW, Djulbegovic B, Soares HP, Siegel BA, Lowe VJ, Lyman GH, et al. Recommendations on the use of 18F-FDG PET in oncology. J Nucl Med. 2008;49:480–508.CrossRefPubMed
2.
Zurück zum Zitat Ben-Haim S, Ell P. 18F-FDG PET and PET/CT in the evaluation of cancer treatment response. J Nucl Med. 2009;50:88–99.CrossRefPubMed Ben-Haim S, Ell P. 18F-FDG PET and PET/CT in the evaluation of cancer treatment response. J Nucl Med. 2009;50:88–99.CrossRefPubMed
3.
Zurück zum Zitat Gupta T, Master Z, Kannan S, Agarwal JP, Ghsoh-Laskar S, Rangarajan V, et al. Diagnostic performance of post-treatment FDG PET or FDG PET/CT imaging in head and neck cancer: a systematic review and meta-analysis. Eur J Nucl Med Mol Imaging. 2011;38:2083–95.CrossRefPubMed Gupta T, Master Z, Kannan S, Agarwal JP, Ghsoh-Laskar S, Rangarajan V, et al. Diagnostic performance of post-treatment FDG PET or FDG PET/CT imaging in head and neck cancer: a systematic review and meta-analysis. Eur J Nucl Med Mol Imaging. 2011;38:2083–95.CrossRefPubMed
4.
Zurück zum Zitat Bengtsson T, Hicks RJ, Peterson A, Port RE. 18F-FDG PET as a surrogate biomarker in non-small cell lung cancer treated with erlotinib: newly identified lesions are more informative than standardized uptake value. J Nucl Med. 2012;53:530–7.CrossRefPubMed Bengtsson T, Hicks RJ, Peterson A, Port RE. 18F-FDG PET as a surrogate biomarker in non-small cell lung cancer treated with erlotinib: newly identified lesions are more informative than standardized uptake value. J Nucl Med. 2012;53:530–7.CrossRefPubMed
5.
Zurück zum Zitat Hicks RJ. Role of 18F-FDG PET in assessment of response in non-small cell lung cancer. J Nucl Med. 2009;50(Suppl 1):31S–42S.CrossRefPubMed Hicks RJ. Role of 18F-FDG PET in assessment of response in non-small cell lung cancer. J Nucl Med. 2009;50(Suppl 1):31S–42S.CrossRefPubMed
6.
Zurück zum Zitat Hellwig D, Graeter TP, Ukena D, Groeschel A, Sybrecht GW, Schaefers HJ, et al. 18F-FDG PET for mediastinal staging of lung cancer: which SUV threshold makes sense? J Nucl Med. 2007;48:1761–6.CrossRefPubMed Hellwig D, Graeter TP, Ukena D, Groeschel A, Sybrecht GW, Schaefers HJ, et al. 18F-FDG PET for mediastinal staging of lung cancer: which SUV threshold makes sense? J Nucl Med. 2007;48:1761–6.CrossRefPubMed
7.
Zurück zum Zitat Dijkman BG, Schuurbiers OCJ, Vriens D, Looijen-Salamon M, Bussink J, Timmer-Bonte JNH, et al. The role of (18)F-FDG PET in the differentiation between lung metastases and synchronous second primary lung tumours. Eur J Nucl Med Mol Imaging. 2010;37:2037–47.CrossRefPubMedPubMedCentral Dijkman BG, Schuurbiers OCJ, Vriens D, Looijen-Salamon M, Bussink J, Timmer-Bonte JNH, et al. The role of (18)F-FDG PET in the differentiation between lung metastases and synchronous second primary lung tumours. Eur J Nucl Med Mol Imaging. 2010;37:2037–47.CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Agarwal M, Brahmanday G, Bajaj SK, Ravikrishnan KP, Wong C-YO. Revisiting the prognostic value of preoperative (18)F-fluoro-2-deoxyglucose ( (18)F-FDG) positron emission tomography (PET) in early-stage (I & II) non-small cell lung cancers (NSCLC). Eur J Nucl Med Mol Imaging. 2010;37:691–8.CrossRefPubMed Agarwal M, Brahmanday G, Bajaj SK, Ravikrishnan KP, Wong C-YO. Revisiting the prognostic value of preoperative (18)F-fluoro-2-deoxyglucose ( (18)F-FDG) positron emission tomography (PET) in early-stage (I & II) non-small cell lung cancers (NSCLC). Eur J Nucl Med Mol Imaging. 2010;37:691–8.CrossRefPubMed
9.
Zurück zum Zitat Liao S, Penney BC, Wroblewski K, Zhang H, Simon CA, Kampalath R, et al. Prognostic value of metabolic tumor burden on 18F-FDG PET in nonsurgical patients with non-small cell lung cancer. Eur J Nucl Med Mol Imaging. 2012;39:27–38.CrossRefPubMed Liao S, Penney BC, Wroblewski K, Zhang H, Simon CA, Kampalath R, et al. Prognostic value of metabolic tumor burden on 18F-FDG PET in nonsurgical patients with non-small cell lung cancer. Eur J Nucl Med Mol Imaging. 2012;39:27–38.CrossRefPubMed
10.
Zurück zum Zitat Sullivan DC, Obuchowski NA, Kessler LG, Raunig DL, Gatsonis C, Huang EP, et al. Metrology standards for quantitative imaging biomarkers. Radiology. 2015;277:813–25.CrossRefPubMedPubMedCentral Sullivan DC, Obuchowski NA, Kessler LG, Raunig DL, Gatsonis C, Huang EP, et al. Metrology standards for quantitative imaging biomarkers. Radiology. 2015;277:813–25.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat O’Connor JPB, Aboagye EO, Adams JE, Aerts HJWL, Barrington SF, Beer AJ, et al. Imaging biomarker roadmap for cancer studies. Nat Rev Clin Oncol. 2017;14:169–86. Nature Publishing GroupCrossRefPubMed O’Connor JPB, Aboagye EO, Adams JE, Aerts HJWL, Barrington SF, Beer AJ, et al. Imaging biomarker roadmap for cancer studies. Nat Rev Clin Oncol. 2017;14:169–86. Nature Publishing GroupCrossRefPubMed
12.
Zurück zum Zitat Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med. 2004;45:1519–27.PubMed Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med. 2004;45:1519–27.PubMed
13.
Zurück zum Zitat Westerterp M, Pruim J, Oyen W, Hoekstra O, Paans A, Visser E, et al. Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parameters. Eur J Nucl Med Mol Imaging. 2007;34:392–404.CrossRefPubMed Westerterp M, Pruim J, Oyen W, Hoekstra O, Paans A, Visser E, et al. Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parameters. Eur J Nucl Med Mol Imaging. 2007;34:392–404.CrossRefPubMed
14.
Zurück zum Zitat Sunderland JJ, Christian PE. Quantitative PET/CT scanner performance characterization based upon the Society of Nuclear Medicine and Molecular Imaging Clinical Trials Network oncology clinical simulator phantom. J Nucl Med. 2015;56:145–52.CrossRefPubMed Sunderland JJ, Christian PE. Quantitative PET/CT scanner performance characterization based upon the Society of Nuclear Medicine and Molecular Imaging Clinical Trials Network oncology clinical simulator phantom. J Nucl Med. 2015;56:145–52.CrossRefPubMed
15.
Zurück zum Zitat Lasnon C, Desmonts C, Quak E, Gervais R, Do P, Dubos-Arvis C, et al. Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patients. Eur J Nucl Med Mol Imaging. 2013;40:985–96.CrossRefPubMedPubMedCentral Lasnon C, Desmonts C, Quak E, Gervais R, Do P, Dubos-Arvis C, et al. Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patients. Eur J Nucl Med Mol Imaging. 2013;40:985–96.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Rahmim A, Qi J, Sossi V. Resolution modeling in PET imaging: theory, practice, benefits, and pitfalls. Med Phys. 2013;40:64301.CrossRef Rahmim A, Qi J, Sossi V. Resolution modeling in PET imaging: theory, practice, benefits, and pitfalls. Med Phys. 2013;40:64301.CrossRef
17.
Zurück zum Zitat Kidera D, Kihara K, Akamatsu G, Mikasa S, Taniguchi T, Tsutsui Y, et al. The edge artifact in the point-spread function-based PET reconstruction at different sphere-to-background ratios of radioactivity. Ann Nucl Med. 2016;30:97–103.CrossRefPubMed Kidera D, Kihara K, Akamatsu G, Mikasa S, Taniguchi T, Tsutsui Y, et al. The edge artifact in the point-spread function-based PET reconstruction at different sphere-to-background ratios of radioactivity. Ann Nucl Med. 2016;30:97–103.CrossRefPubMed
18.
Zurück zum Zitat Graham MM, Wahl RL, Hoffman JM, Yap JT, Sunderland JJ, Boellaard R, et al. Summary of the UPICT protocol for 18F-FDG PET/CT imaging in oncology clinical trials. J Nucl Med. 2015;56:955–61.CrossRefPubMedPubMedCentral Graham MM, Wahl RL, Hoffman JM, Yap JT, Sunderland JJ, Boellaard R, et al. Summary of the UPICT protocol for 18F-FDG PET/CT imaging in oncology clinical trials. J Nucl Med. 2015;56:955–61.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Boellaard R, Delgado-Bolton R, Oyen WJG, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42:328–54. Boellaard R, Delgado-Bolton R, Oyen WJG, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42:328–54.
20.
Zurück zum Zitat Fukukita H, Suzuki K, Matsumoto K, Terauchi T, Daisaki H, Ikari Y, et al. Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of version 2.0. Ann. Nucl Med. 2014;28:693–705.CrossRefPubMedPubMedCentral Fukukita H, Suzuki K, Matsumoto K, Terauchi T, Daisaki H, Ikari Y, et al. Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of version 2.0. Ann. Nucl Med. 2014;28:693–705.CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Quak E, Le Roux P-Y, Hofman MS, Robin P, Bourhis D, Callahan J, et al. Harmonizing FDG PET quantification while maintaining optimal lesion detection: prospective multicentre validation in 517 oncology patients. Eur J Nucl Med Mol Imaging. 2015;42:2072–82.CrossRefPubMedPubMedCentral Quak E, Le Roux P-Y, Hofman MS, Robin P, Bourhis D, Callahan J, et al. Harmonizing FDG PET quantification while maintaining optimal lesion detection: prospective multicentre validation in 517 oncology patients. Eur J Nucl Med Mol Imaging. 2015;42:2072–82.CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Makris NE, Huisman MC, Kinahan PE, Lammertsma AA, Boellaard R. Evaluation of strategies towards harmonization of FDG PET/CT studies in multicentre trials: comparison of scanner validation phantoms and data analysis procedures. Eur J Nucl Med Mol Imaging. 2013;40:1507–15.CrossRefPubMed Makris NE, Huisman MC, Kinahan PE, Lammertsma AA, Boellaard R. Evaluation of strategies towards harmonization of FDG PET/CT studies in multicentre trials: comparison of scanner validation phantoms and data analysis procedures. Eur J Nucl Med Mol Imaging. 2013;40:1507–15.CrossRefPubMed
25.
Zurück zum Zitat Quak E, Le Roux P-Y, Lasnon C, Robin P, Hofman MS, Bourhis D, et al. Does PET SUV harmonization affect PERCIST response classification? J Nucl Med. 2016;57:1699–706.CrossRefPubMed Quak E, Le Roux P-Y, Lasnon C, Robin P, Hofman MS, Bourhis D, et al. Does PET SUV harmonization affect PERCIST response classification? J Nucl Med. 2016;57:1699–706.CrossRefPubMed
26.
Zurück zum Zitat Raunig DL, McShane LM, Pennello G, Gatsonis C, Carson PL, Voyvodic JT, et al. Quantitative imaging biomarkers: a review of statistical methods for technical performance assessment. Stat Methods Med Res. 2015;24:27–67.CrossRefPubMed Raunig DL, McShane LM, Pennello G, Gatsonis C, Carson PL, Voyvodic JT, et al. Quantitative imaging biomarkers: a review of statistical methods for technical performance assessment. Stat Methods Med Res. 2015;24:27–67.CrossRefPubMed
27.
Zurück zum Zitat Pierce LA, Elston BF, Clunie DA, Nelson D, Kinahan PE. A digital reference object to analyze calculation accuracy of PET standardized uptake value. Radiology. 2015;277:538–45.CrossRefPubMedPubMedCentral Pierce LA, Elston BF, Clunie DA, Nelson D, Kinahan PE. A digital reference object to analyze calculation accuracy of PET standardized uptake value. Radiology. 2015;277:538–45.CrossRefPubMedPubMedCentral
29.
Zurück zum Zitat Nakahara T, Daisaki H, Yamamoto Y, Iimori T, Miyagawa K, Okamoto T, et al. Use of a digital phantom developed by QIBA for harmonizing SUVs obtained from the state-of-the-art SPECT/CT systems: a multicenter study. EJNMMI Res. 2017;7:53.CrossRefPubMedPubMedCentral Nakahara T, Daisaki H, Yamamoto Y, Iimori T, Miyagawa K, Okamoto T, et al. Use of a digital phantom developed by QIBA for harmonizing SUVs obtained from the state-of-the-art SPECT/CT systems: a multicenter study. EJNMMI Res. 2017;7:53.CrossRefPubMedPubMedCentral
30.
Zurück zum Zitat Boellaard R. Standards for PET image acquisition and quantitative data analysis. J Nucl Med. 2009;50(Suppl 1):11S–20S.CrossRefPubMed Boellaard R. Standards for PET image acquisition and quantitative data analysis. J Nucl Med. 2009;50(Suppl 1):11S–20S.CrossRefPubMed
32.
Zurück zum Zitat Takahashi Y, Oriuchi N, Otake H, Endo K, Murase K. Variability of lesion detectability and standardized uptake value according to the acquisition procedure and reconstruction among five PET scanners. Ann Nucl Med. 2008;22:543–8.CrossRefPubMed Takahashi Y, Oriuchi N, Otake H, Endo K, Murase K. Variability of lesion detectability and standardized uptake value according to the acquisition procedure and reconstruction among five PET scanners. Ann Nucl Med. 2008;22:543–8.CrossRefPubMed
33.
Zurück zum Zitat Velasquez LM, Boellaard R, Kollia G, Hayes W, Hoekstra OS, Lammertsma AA, et al. Repeatability of 18F-FDG PET in a multicenter phase I study of patients with advanced gastrointestinal malignancies. J Nucl Med. 2009;50:1646–54.CrossRefPubMed Velasquez LM, Boellaard R, Kollia G, Hayes W, Hoekstra OS, Lammertsma AA, et al. Repeatability of 18F-FDG PET in a multicenter phase I study of patients with advanced gastrointestinal malignancies. J Nucl Med. 2009;50:1646–54.CrossRefPubMed
34.
Zurück zum Zitat Doot RK, Scheuermann JS, Christian PE, Karp JS, Kinahan PE. Instrumentation factors affecting variance and bias of quantifying tracer uptake with PET/CT. Med Phys. 2010;37:6035–46.CrossRefPubMedPubMedCentral Doot RK, Scheuermann JS, Christian PE, Karp JS, Kinahan PE. Instrumentation factors affecting variance and bias of quantifying tracer uptake with PET/CT. Med Phys. 2010;37:6035–46.CrossRefPubMedPubMedCentral
36.
Zurück zum Zitat Lim R, Eaton A, Lee NY, Setton J, Ohri N, Rao S, et al. 18F-FDG PET/CT metabolic tumor volume and total lesion glycolysis predict outcome in oropharyngeal squamous cell carcinoma. J Nucl Med. 2012;53:1506–13.CrossRefPubMed Lim R, Eaton A, Lee NY, Setton J, Ohri N, Rao S, et al. 18F-FDG PET/CT metabolic tumor volume and total lesion glycolysis predict outcome in oropharyngeal squamous cell carcinoma. J Nucl Med. 2012;53:1506–13.CrossRefPubMed
Metadaten
Titel
Multicentre analysis of PET SUV using vendor-neutral software: the Japanese Harmonization Technology (J-Hart) study
verfasst von
Yuji Tsutsui
Hiromitsu Daisaki
Go Akamatsu
Takuro Umeda
Matsuyoshi Ogawa
Hironori Kajiwara
Shigeto Kawase
Minoru Sakurai
Hiroyuki Nishida
Keiichi Magota
Kazuaki Mori
Masayuki Sasaki
J-Hart study group
Publikationsdatum
01.12.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
EJNMMI Research / Ausgabe 1/2018
Elektronische ISSN: 2191-219X
DOI
https://doi.org/10.1186/s13550-018-0438-9

Weitere Artikel der Ausgabe 1/2018

EJNMMI Research 1/2018 Zur Ausgabe