Noise reduction and motion elimination in low-dose 4D myocardial computed tomography perfusion (CTP): preliminary clinical evaluation of the ASTRA4D algorithm

verfasst von: Steffen Lukas, Sarah Feger, Matthias Rief, Elke Zimmermann, Marc Dewey

Erschienen in: European Radiology | Ausgabe 9/2019

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To propose and evaluate a four-dimensional (4D) algorithm for joint motion elimination and spatiotemporal noise reduction in low-dose dynamic myocardial computed tomography perfusion (CTP).

Methods

Thirty patients with suspected or confirmed coronary artery disease were prospectively included and underwent dynamic contrast-enhanced 320-row CTP. A novel deformable image registration method based on the principal component analysis (PCA) of the ante hoc temporally smoothed voxel-wise time-attenuation curves (ASTRA4D) is presented. Quantitative (standard deviation, signal-to-noise ratio (SNR), temporal variation, volumetric deformation) and qualitative (motion, contrast, contour sharpness [1, poor; 5, excellent]) measures of CTP quality were assessed for the original and motion-compensated sequences (without and with temporal filtering, PCA/ASTRA4D). Following myocardial perfusion deficit detection by two readers, diagnostic accuracy was evaluated using magnetic resonance myocardial perfusion imaging (MR-MPI) as the reference standard in 15 patients.

Results

Registration using ASTRA4D was successful in all 30 patients and resulted in comparison with the benchmark PCA in significantly (p < 0.001) reduced noise over time (− 83%, 178.5 vs 29.9) and spatially (− 34%, 21.4 vs 14.1) as well as improved SNR (+ 47%, 3.6 vs 5.3) and subjective image quality (motion, contrast, contour sharpness [+ 1.0, + 1.0, + 0.5]). ASTRA4D had significantly improved per-segment sensitivity of 91% (58/64) and similar specificity of 96% (429/446) compared with PCA (52%, 33/64; 98%, 435/446; p = 0.011) in the visual detection of perfusion deficits.

Conclusions

The ASTRA4D registration algorithm improved the spatiotemporal noise profile and CTP sequence image quality, resulting in significantly improved sensitivity of 4D CTP in the detection of myocardial ischemia.

Key Points

• ASTRA4D combines local temporal regression and deformable image registration.
• Quantitative and qualitative measures of CTP quality are improved compared to PCA.
• Improved spatiotemporal differentiation of ischemic regions leads to an excellent perfusion deficit concordance of ASTRA4D with MRI.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Napp AE, Haase R, Laule M et al (2017) Computed tomography versus invasive coronary angiography: design and methods of the pragmatic randomised multicentre DISCHARGE trial. Eur Radiol 27:2957–2968CrossRefPubMed Napp AE, Haase R, Laule M et al (2017) Computed tomography versus invasive coronary angiography: design and methods of the pragmatic randomised multicentre DISCHARGE trial. Eur Radiol 27:2957–2968CrossRefPubMed
2.
Zurück zum Zitat Bamberg F, Becker A, Schwarz F et al (2011) Detection of hemodynamically significant coronary artery stenosis: incremental diagnostic value of dynamic CT-based myocardial perfusion imaging. Radiology 260:689–698CrossRefPubMed Bamberg F, Becker A, Schwarz F et al (2011) Detection of hemodynamically significant coronary artery stenosis: incremental diagnostic value of dynamic CT-based myocardial perfusion imaging. Radiology 260:689–698CrossRefPubMed
3.
Zurück zum Zitat George RT, Jerosch-Herold M, Silva C et al (2007) Quantification of myocardial perfusion using dynamic 64-detector computed tomography. Invest Radiol 42:815–822CrossRefPubMed George RT, Jerosch-Herold M, Silva C et al (2007) Quantification of myocardial perfusion using dynamic 64-detector computed tomography. Invest Radiol 42:815–822CrossRefPubMed
4.
Zurück zum Zitat So A, Wisenberg G, Islam A et al (2012) Non-invasive assessment of functionally relevant coronary artery stenoses with quantitative CT perfusion: preliminary clinical experiences. Eur Radiol 22:39–50CrossRefPubMed So A, Wisenberg G, Islam A et al (2012) Non-invasive assessment of functionally relevant coronary artery stenoses with quantitative CT perfusion: preliminary clinical experiences. Eur Radiol 22:39–50CrossRefPubMed
5.
Zurück zum Zitat Varga-Szemes A, Meinel FG, De Cecco CN, Fuller SR, Bayer RR 2nd, Schoepf UJ (2015) CT myocardial perfusion imaging. AJR Am J Roentgenol 204:487–497 Varga-Szemes A, Meinel FG, De Cecco CN, Fuller SR, Bayer RR 2nd, Schoepf UJ (2015) CT myocardial perfusion imaging. AJR Am J Roentgenol 204:487–497
6.
Zurück zum Zitat de Jong MC, Genders TS, van Geuns RJ, Moelker A, Hunink MG (2012) Diagnostic performance of stress myocardial perfusion imaging for coronary artery disease: a systematic review and meta-analysis. Eur Radiol 22:1881–1895CrossRefPubMedPubMedCentral de Jong MC, Genders TS, van Geuns RJ, Moelker A, Hunink MG (2012) Diagnostic performance of stress myocardial perfusion imaging for coronary artery disease: a systematic review and meta-analysis. Eur Radiol 22:1881–1895CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Rief M, Chen MY, Vavere AL et al (2018) Coronary artery disease: analysis of diagnostic performance of CT perfusion and MR perfusion imaging in comparison with quantitative coronary angiography and SPECT-multicenter prospective trial. Radiology 286:461–470CrossRefPubMed Rief M, Chen MY, Vavere AL et al (2018) Coronary artery disease: analysis of diagnostic performance of CT perfusion and MR perfusion imaging in comparison with quantitative coronary angiography and SPECT-multicenter prospective trial. Radiology 286:461–470CrossRefPubMed
8.
Zurück zum Zitat Takx RA, Blomberg BA, El Aidi H et al (2015) Diagnostic accuracy of stress myocardial perfusion imaging compared to invasive coronary angiography with fractional flow reserve meta-analysis. Circ Cardiovasc Imaging 8(1):e002666 Takx RA, Blomberg BA, El Aidi H et al (2015) Diagnostic accuracy of stress myocardial perfusion imaging compared to invasive coronary angiography with fractional flow reserve meta-analysis. Circ Cardiovasc Imaging 8(1):e002666
9.
Zurück zum Zitat Williams MC, Mirsadraee S, Dweck MR et al (2017) Computed tomography myocardial perfusion vs (15)O-water positron emission tomography and fractional flow reserve. Eur Radiol 27:1114–1124CrossRefPubMed Williams MC, Mirsadraee S, Dweck MR et al (2017) Computed tomography myocardial perfusion vs (15)O-water positron emission tomography and fractional flow reserve. Eur Radiol 27:1114–1124CrossRefPubMed
10.
Zurück zum Zitat Williams MC, Newby DE (2016) CT myocardial perfusion imaging: current status and future directions. Clin Radiol 71:739–749CrossRefPubMed Williams MC, Newby DE (2016) CT myocardial perfusion imaging: current status and future directions. Clin Radiol 71:739–749CrossRefPubMed
11.
Zurück zum Zitat Kikuchi Y, Oyama-Manabe N, Naya M et al (2014) Quantification of myocardial blood flow using dynamic 320-row multi-detector CT as compared with (1)(5)O-H(2) O PET. Eur Radiol 24:1547–1556CrossRefPubMed Kikuchi Y, Oyama-Manabe N, Naya M et al (2014) Quantification of myocardial blood flow using dynamic 320-row multi-detector CT as compared with (1)(5)O-H(2) O PET. Eur Radiol 24:1547–1556CrossRefPubMed
12.
Zurück zum Zitat Feger S, Shaban A, Lukas S et al (2017) Temporal averaging for analysis of four-dimensional whole-heart computed tomography perfusion of the myocardium: proof-of-concept study. Int J Cardiovasc Imaging 33:371–382CrossRefPubMed Feger S, Shaban A, Lukas S et al (2017) Temporal averaging for analysis of four-dimensional whole-heart computed tomography perfusion of the myocardium: proof-of-concept study. Int J Cardiovasc Imaging 33:371–382CrossRefPubMed
13.
Zurück zum Zitat Li Z, Yu L, Leng S et al (2016) A robust noise reduction technique for time resolved CT. Med Phys 43:347CrossRefPubMed Li Z, Yu L, Leng S et al (2016) A robust noise reduction technique for time resolved CT. Med Phys 43:347CrossRefPubMed
14.
Zurück zum Zitat Pisana F, Henzler T, Schönberg S, Klotz E, Schmidt B, Kachelrieß M (2017) Noise reduction and functional maps image quality improvement in dynamic CT perfusion using a new k-means clustering guided bilateral filter (KMGB). Med Phys 44:3464–3482CrossRefPubMed Pisana F, Henzler T, Schönberg S, Klotz E, Schmidt B, Kachelrieß M (2017) Noise reduction and functional maps image quality improvement in dynamic CT perfusion using a new k-means clustering guided bilateral filter (KMGB). Med Phys 44:3464–3482CrossRefPubMed
15.
16.
Zurück zum Zitat Hamy V, Dikaios N, Punwani S et al (2014) Respiratory motion correction in dynamic MRI using robust data decomposition registration—application to DCE-MRI. Med Image Anal 18:301–313CrossRefPubMed Hamy V, Dikaios N, Punwani S et al (2014) Respiratory motion correction in dynamic MRI using robust data decomposition registration—application to DCE-MRI. Med Image Anal 18:301–313CrossRefPubMed
17.
Zurück zum Zitat Huizinga W, Poot DH, Guyader JM et al (2016) PCA-based groupwise image registration for quantitative MRI. Med Image Anal 29:65–78CrossRefPubMed Huizinga W, Poot DH, Guyader JM et al (2016) PCA-based groupwise image registration for quantitative MRI. Med Image Anal 29:65–78CrossRefPubMed
18.
Zurück zum Zitat Melbourne A, Atkinson D, White MJ, Collins D, Leach M, Hawkes D (2007) Registration of dynamic contrast-enhanced MRI using a progressive principal component registration (PPCR). Phys Med Biol 52:5147–5156CrossRefPubMed Melbourne A, Atkinson D, White MJ, Collins D, Leach M, Hawkes D (2007) Registration of dynamic contrast-enhanced MRI using a progressive principal component registration (PPCR). Phys Med Biol 52:5147–5156CrossRefPubMed
19.
Zurück zum Zitat Wollny G, Kellman P, Santos A, Ledesma-Carbayo MJ (2012) Automatic motion compensation of free breathing acquired myocardial perfusion data by using independent component analysis. Med Image Anal 16:1015–1028CrossRefPubMedPubMedCentral Wollny G, Kellman P, Santos A, Ledesma-Carbayo MJ (2012) Automatic motion compensation of free breathing acquired myocardial perfusion data by using independent component analysis. Med Image Anal 16:1015–1028CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Mihai G, Ding Y, Xue H et al (2012) Non-rigid registration and KLT filter to improve SNR and CNR in GRE-EPI myocardial perfusion imaging. J Biomed Sci Eng 5:871–877CrossRefPubMedPubMedCentral Mihai G, Ding Y, Xue H et al (2012) Non-rigid registration and KLT filter to improve SNR and CNR in GRE-EPI myocardial perfusion imaging. J Biomed Sci Eng 5:871–877CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Muenzel D, Kabus S, Gramer B et al (2013) Dynamic CT perfusion imaging of the myocardium: a technical note on improvement of image quality. PLoS One 8(10):e75263 Muenzel D, Kabus S, Gramer B et al (2013) Dynamic CT perfusion imaging of the myocardium: a technical note on improvement of image quality. PLoS One 8(10):e75263
22.
Zurück zum Zitat Feger S, Rief M, Zimmermann E et al (2015) The impact of different levels of adaptive iterative dose reduction 3D on image quality of 320-row coronary CT angiography: a clinical trial. PLoS One 10:e0125943CrossRefPubMedPubMedCentral Feger S, Rief M, Zimmermann E et al (2015) The impact of different levels of adaptive iterative dose reduction 3D on image quality of 320-row coronary CT angiography: a clinical trial. PLoS One 10:e0125943CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Techasith T, Cury RC (2011) Stress myocardial CT perfusion: an update and future perspective. JACC Cardiovasc Imaging 4:905–916CrossRefPubMed Techasith T, Cury RC (2011) Stress myocardial CT perfusion: an update and future perspective. JACC Cardiovasc Imaging 4:905–916CrossRefPubMed
24.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V et al (2002) Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Int J Cardiovasc Imaging 18:539–542PubMed Cerqueira MD, Weissman NJ, Dilsizian V et al (2002) Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Int J Cardiovasc Imaging 18:539–542PubMed
25.
Zurück zum Zitat Bamberg F, Marcus RP, Becker A et al (2014) Dynamic myocardial CT perfusion imaging for evaluation of myocardial ischemia as determined by MR imaging. JACC Cardiovasc Imaging 7:267–277CrossRefPubMed Bamberg F, Marcus RP, Becker A et al (2014) Dynamic myocardial CT perfusion imaging for evaluation of myocardial ischemia as determined by MR imaging. JACC Cardiovasc Imaging 7:267–277CrossRefPubMed
26.
Zurück zum Zitat Bischoff B, Bamberg F, Marcus R et al (2013) Optimal timing for first-pass stress CT myocardial perfusion imaging. Int J Cardiovasc Imaging 29:435–442CrossRefPubMed Bischoff B, Bamberg F, Marcus R et al (2013) Optimal timing for first-pass stress CT myocardial perfusion imaging. Int J Cardiovasc Imaging 29:435–442CrossRefPubMed
27.
Zurück zum Zitat Yang Z, Silver MD (2015) Denoising method and system for preserving clinically significant structures in reconstructed images using adaptively weighted anisotropic diffusion filter. Google Patents Yang Z, Silver MD (2015) Denoising method and system for preserving clinically significant structures in reconstructed images using adaptively weighted anisotropic diffusion filter. Google Patents
28.
Zurück zum Zitat So A, Imai Y, Nett B et al (2016) Technical note: evaluation of a 160-mm/256-row CT scanner for whole-heart quantitative myocardial perfusion imaging. Med Phys 43:4821CrossRefPubMed So A, Imai Y, Nett B et al (2016) Technical note: evaluation of a 160-mm/256-row CT scanner for whole-heart quantitative myocardial perfusion imaging. Med Phys 43:4821CrossRefPubMed
29.
Zurück zum Zitat Modgil D, Bindschadler MD, Alessio AM, La Rivière PJ (2017) Variable temporal sampling and tube current modulation for myocardial blood flow estimation from dose-reduced dynamic computed tomography. J Med Imaging (Bellingham) 4:026002CrossRef Modgil D, Bindschadler MD, Alessio AM, La Rivière PJ (2017) Variable temporal sampling and tube current modulation for myocardial blood flow estimation from dose-reduced dynamic computed tomography. J Med Imaging (Bellingham) 4:026002CrossRef
Metadaten
Titel
Noise reduction and motion elimination in low-dose 4D myocardial computed tomography perfusion (CTP): preliminary clinical evaluation of the ASTRA4D algorithm
verfasst von
Steffen Lukas
Sarah Feger
Matthias Rief
Elke Zimmermann
Marc Dewey
Publikationsdatum
04.02.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 9/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5899-8

Weitere Artikel der Ausgabe 9/2019

European Radiology 9/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.