Nonenhanced MR angiography of the pulmonary arteries using single-shot radial quiescent-interval slice-selective (QISS): a technical feasibility study

verfasst von: Robert R. Edelman, Robert I. Silvers, Kiran H. Thakrar, Mark D. Metzl, Jose Nazari, Shivraman Giri, Ioannis Koktzoglou

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2017

Abstract

Background

For evaluation of the pulmonary arteries in patients suspected of pulmonary embolism, CT angiography (CTA) is the first-line imaging test with contrast-enhanced MR angiography (CEMRA) a potential alternative. Disadvantages of CTA include exposure to ionizing radiation and an iodinated contrast agent, while CEMRA is sensitive to respiratory motion and requires a gadolinium-based contrast agent. The primary goal of our technical feasibility study was to evaluate pulmonary arterial conspicuity using breath-hold and free-breathing implementations of a recently-developed nonenhanced approach, single-shot radial quiescent-interval slice-selective (QISS) MRA.

Methods

Breath-hold and free-breathing, navigator-gated versions of radial QISS MRA were evaluated at 1.5 Tesla in three healthy subjects and 11 patients without pulmonary embolism or arterial occlusion by CTA. Images were scored by three readers for conspicuity of the pulmonary arteries through the level of the segmental branches. In addition, one patient with pulmonary embolism was imaged.

Results

Scan time for a 54-slice acquisition spanning the pulmonary arteries was less than 2 minutes for breath-hold QISS, and less than 3.4 min using free-breathing QISS. Pulmonary artery branches through the segmental level were conspicuous with either approach. Free-breathing scans showed only mild blurring compared with breath-hold scans. For both readers, less than 1% of pulmonary arterial segments were rated as “not seen” for breath-hold and navigator-gated QISS, respectively. In subjects with atrial fibrillation, single-shot radial QISS consistently depicted the pulmonary artery branches, whereas navigator-gated 3D balanced steady-state free precession showed motion artifacts. In one patient with pulmonary embolism, radial QISS demonstrated central pulmonary emboli comparably to CEMRA and CTA. The thrombi were highly conspicuous on radial QISS images, but appeared subtle and were not prospectively identified on scout images acquired using a single-shot bSSFP acquisition.

Conclusions

In this technical feasibility study, both breath-hold and free-breathing single-shot radial QISS MRA enabled rapid, consistent demonstration of the pulmonary arteries through the level of the segmental branches, with only minimal artifacts from respiratory motion and cardiac arrhythmias. Based on these promising initial results, further evaluation in patients with suspected pulmonary embolism appears warranted.
Abkürzungen
bSSFP
Balanced steady-state free precession
CE
Contrast enhanced
CTA
Computed tomography angiography
ECG
Electrocardiogram
MRA
Magnetic resonance angiography
PE
Pulmonary embolism
QISS
Quiescent-interval slice-selective

Background

Patients presenting with suspected pulmonary embolism (PE) are routinely evaluated by computed tomography angiography (CTA) [1]. The risk of having a pulmonary embolism doubles for each 10 years after age 60 [https://​www.​nhlbi.​nih.​gov/​health/​health-topics/​topics/​pe/​atrisk]. Given the nearly 40% prevalence of chronic kidney disease among patients over the age of 60 years [2] and the potentially nephrotoxic effects of iodinated contrast agents, there is a substantial need for a safer imaging option in this patient group. First-pass contrast-enhanced MR angiography (CEMRA) of the pulmonary arteries is a promising alternative to CTA that avoids exposure to iodinated contrast agents and ionizing radiation [3]. However, potential challenges include the requirement for breath-holding, which may be problematic for dyspneic patients, and incomplete pulmonary artery opacification due to mistiming of the data acquisition with respect to the gadolinium bolus. Moreover, gadolinium-based contrast agents are absolutely contraindicated in pregnant patients and relatively contraindicated in those with severe renal impairment [4]. For patients with contraindications to CTA or CEMRA, a nonenhanced MRA (NEMRA) alternative would be useful.
Recently, a radial quiescent-interval slice-selective (QISS) technique was described for breath-hold imaging of the coronary arteries [5]. Radial k-space trajectories provide several advantages over Cartesian trajectories including reduced motion sensitivity, more flexible control over temporal and spatial resolution, and higher undersampling factors [6, 7]. For coronary artery QISS, two or more shots are typically required to achieve sufficiently high temporal resolution, on the order of 150 msec or less, to minimize blurring from coronary motion. Such high temporal resolution is not needed for imaging of the pulmonary arteries, which allows the use of a more efficient single-shot radial QISS acquisition. Not only does the use of a single-shot acquisition at least double the number of slices that can be acquired in each breath-hold, it avoids artifacts from shot-to-shot signal variations caused by respiratory motion or cardiac arrhythmias. Given its high imaging efficiency and resistance to motion artifacts, single-shot radial QISS could have potential utility as a nonenhanced option for evaluating patients with suspected pulmonary embolism. As an initial step, we performed a technical feasibility study to evaluate pulmonary arterial conspicuity using breath-hold and free-breathing implementations of single-shot radial QISS.

Methods

This investigational review board (IRB)-approved study was conducted on a 1.5 Tesla scanner (MAGNETOM Avanto, Siemens Healthcare, Erlangen, Germany). Two groups of subjects were studied: (1) 3 healthy subjects; and (2) 11 patients (10 male, age range 50–73 years) who were scheduled for pulmonary vein isolation due to recurrent or persistent atrial fibrillation and had recently undergone CTA for anatomical evaluation of the pulmonary veins and heart. This patient cohort additionally provided the opportunity to evaluate the robustness of the radial QISS technique with respect to cardiac arrhythmias, since six patients were in atrial fibrillation at the time of the MR exam. Retrospective approval from the IRB was obtained for one additional patient: a 68-year-old male with shortness of breath and suspected peri-valvular leak following mitral valve repair, who underwent cardiac MR which revealed clinically unsuspected central pulmonary emboli. This patient subsequently underwent a chest CTA that confirmed the MR findings.
Imaging parameters for radial QISS were developed empirically from prior studies in volunteers and patients. Typical QISS imaging parameters included: electrocardiographic gating (ECG), one slice acquired per RR interval, radial balanced steady-state free precession (bSSFP) readout with 98 views, flip angle for the bSSFP RF excitation = 120 degrees, chemical shift-selective fat suppression, in-plane inversion using a frequency offset corrected inversion (FOCI) RF pulse with thickness = 4.5 mm, TI ~ 600 msec, in-plane resolution ~1.6-mm (or 0.8-mm after interpolation), field of view ~260-mm, slice thickness ~ 3.1-mm with 30% slice overlap, number of slices ~54, equidistant azimuthal view angle increment, readout bandwidth = 1359 Hz/pixel, bSSFP repetition time (TR) ~ 2.9 msec. Breath-hold scans were acquired over 3 breath-holds (~18 slices per breath-hold) with 10 to12 seconds between breath-holds. Free-breathing radial QISS used navigator gating with a 3-mm acceptance window and both leading and trailing cross-pair navigators. In total, all 14 subjects were imaged with breath-hold QISS, while only 13 of these subjects were imaged with navigator-gated QISS. Scans were acquired in tilted coronal and axial planes.
Navigator-gated, T2-prepared fat-saturated 3D bSSFP scans were obtained using a Cartesian k-space trajectory, bSSFP RF excitation flip angle = 90 degrees, ~72 3-mm thick slices (1.5-mm after interpolation), in-plane resolution ~1–2-mm (before interpolation), 25 segments, readout bandwidth = 1313 Hz/pixel, bSSFP TR ~ 3.1 msec. Due to time constraints, navigator-gated 3D scans were only obtained in eight subjects (1 volunteer, 7 patients).
First-pass CEMRA used a breath-hold fluoro-triggered technique with scan time = 17 s, flip angle = 23 degrees, TR = 2.7 msec, TE = 1.0 msec, spatial resolution = 2-mm × 1.1-mm × 1.0-mm, ipat acceleration factor = 4, 6/8 partial Fourier in slice and phase directions. CTA was done using a standard high-pitch retrospective-ECG gated spiral protocol on a dual-source scanner (Siemens MAGNETOM Flash, Erlangen, Germany) with 0.6-mm slice thickness.
Image Analysis: Radial QISS source images and thin multi-planar reconstructions from the three healthy volunteers and 11 patients were evaluated by three radiologists, each with more than 5 years’ experience in the interpretation of pulmonary CTA and body MRA. The pulmonary arterial tree was divided into 25 segments per Jackson and Huber [8], and vessel conspicuity was rated as: 1 = vessel seen, sharp margins, negligible artifacts; 2 = vessel seen, mildly blurred margins, mild artifacts; 3 = vessel seen, markedly blurred margins, moderate artifacts; 4 = vessel not seen, severe artifacts. To determine which pulmonary artery segments were evaluable for the image analysis, CEMRA was the reference in healthy volunteers and CTA the reference in patients. Subsegmental arterial branches, as well as segments not evaluable by CEMRA (in volunteers) or CTA (in patients) were excluded from the analysis.
Statistical Analysis: Differences in image quality ratings between the three nonenhanced protocols were assessed using Friedman tests. Gwet’s AC1 was used to assess inter-reader agreement for vessels proximal to the segmental arteries, and for the segmental branches. Bonferroni-corrected P values <0.05 were considered statistically significant. Analyses were performed in R software (version 3.3.2, R Foundation for Statistical Computing, Vienna, Austria).

Results

Typical breath-hold times were ~15–20 s for an 18-slice radial QISS scan, depending on the heart rate. A 54-slice, three breath-hold acquisition spanning the pulmonary arteries was completed in less than 2 min in all subjects. Scan times for a 54-slice navigator-gated QISS scan ranged from 2.0 to 3.4 min. Pulmonary artery branches were conspicuous using either breath-hold or navigator-gated radial QISS, with only mild if any blurring apparent on the free-breathing images (Figs. 1 and 2). Compared with navigator-gated 3D, vessels generally appeared sharper with navigator-gated radial QISS and there was less signal from pericardial fluid (Fig. 3). In patients with an irregular cardiac rhythm due to atrial fibrillation, the pulmonary artery branches were conspicuous on radial QISS images, but were degraded with navigator-gated 3D (Fig. 4).
Contrast-enhanced MRA in the volunteers and CTA in the patients demonstrated patency of all scored pulmonary arterial segments, with no evidence of pulmonary embolism or pulmonary artery occlusion down to the segmental level. The mean conspicuity of evaluable pulmonary arterial segments was rated in the acceptable range (1 to 3) by all three readers for both breath-hold and navigator-gated single-shot radial QISS (Table 1). Across the three readers, only 0.57% (2/350) and 0.62% (2/325) of pulmonary arterial segments were rated as “not seen” for breath-hold and navigator-gated QISS, respectively. By comparison, 31.0% (62/200) pulmonary arterial segments were rated as “not seen” during navigator-gated, T2-prepared fat-saturated 3D bSSFP. Friedman testing revealed significant differences in image quality scores between the three nonenhanced protocols (P < 0.05). Mean image quality scores for breath-hold QISS, navigator-gated QISS and T2-prepared fat-saturated 3D bSSFP were 2.1, 2.3 and 2.7 respectively (P < 0.001 across nonenhanced protocols), and 2.0 for CEMRA. In general, breath-hold QISS provided better image quality than CEMRA for the pulmonary arteries through the level of the lobar branches, while the converse was true for the segmental branches. However, the number of subjects with CEMRA was too small for statistically meaningful comparison.
Table 1
Image quality ratings
Segment
Nonenhanced  QISS BH
Nonenhanced  QISS Nav
Nonenhanced  3D Nav bSSFP
Friedman  P-value
CEMRA
Main Pulmonary Artery
1.0
1.1
1.6
NS
1.4
Right Pulmonary Artery
1.1
1.2
1.6
NS
1.4
Left Pulmonary Artery
1.1
1.2
1.6
NS
1.4
Right Upper Lobar Artery
1.3
1.6
2.0
NS
1.6
Right Lower Lobar Artery
1.3
1.5
2.0
NS
1.7
Left Upper Lobar Artery
1.7
1.7
2.3
NS
1.4
Left Lower Lobar Artery
1.3
1.7
2.0
NS
1.4
RUL-Apical
2.5
2.7
3.0
<0.01
1.9
RUL-Anterior
2.6
2.7
3.0
NS
2.1
RUL-Posterior
2.5
2.7
3.0
<0.01
2.1
RML-Lateral
2.6
2.7
3.1
<0.01
2.1
RML-Medial
2.6
2.7
3.0
<0.01
2.1
RLL-Superior
2.6
2.7
3.0
<0.01
2.1
RLL-Medial Basal
2.5
2.7
3.0
<0.05
2.1
RLL-Anterior Basal
2.5
2.7
3.0
<0.01
2.1
RLL-Lateral Basal
2.6
2.7
3.1
<0.05
2.1
RLL-Posterior Basal
2.4
2.7
3.1
<0.01
2.1
LUL-Apicoposterior
2.4
2.7
3.0
<0.05
2.1
LUL-Anterior
2.4
2.7
3.0
<0.05
2.2
LUL-Superior Lingular
2.5
2.7
3.0
<0.01
2.4
LUL-Inferior Lingular
2.5
2.7
3.0
<0.05
2.4
LLL-Superior
2.5
2.7
3.0
<0.01
2.4
LLL-Anteromedial Basal
2.5
2.7
3.0
<0.01
2.4
LLL-Lateral Basal
2.4
2.6
3.1
<0.01
2.2
LLL-Posterior Basal
2.2
2.4
3.1
<0.001
2.2
All Segments
2.1
2.3
2.7
<0.001
2.0
Values are means across the three readers. BH breath-hold, Nav navigator-gated, bSSFP balanced steady-state free precession, RUL right upper lobe, RML right middle lobe, RLL right lower lobe, LUL left upper lobe, LLL left lower lobe, NS not significant
For the pulmonary arteries through the level of the lobar branches, inter-reader agreement was substantial for breath-hold QISS (AC1 = 0.62, P < 0.001), moderate for navigator-gated QISS (AC1 = 0.47, P < 0.001) and T2-prepared fat-saturated 3D bSSFP (AC1 = 0.45, P < 0.001), and poor for CEMRA (AC1 = 0.15, P < 0.01). For segmental branches, inter-reader agreement was fair for breath-hold QISS (AC1 = 0.24, P < 0.001), navigator-gated QISS (AC1 = 0.37, P < 0.001), T2-prepared fat-saturated 3D bSSFP (AC1 = 0.22, P < 0.001), and CEMRA (AC1 = 0.25, P < 0.001).
In one patient with clinically unsuspected pulmonary embolism, radial QISS demonstrated the central pulmonary thrombi comparably to CEMRA and CTA (Fig. 5). The thrombi were highly conspicuous on radial QISS images, but appeared subtle and were not prospectively identified on scout images acquired using an ECG-gated single-shot bSSFP acquisition.

Discussion

Nonenhanced MRA provides a risk-free imaging alternative to CTA and CEMRA. For instance, navigator-gated 3D bSSFP is a well-described free-breathing technique that has been used to image the coronary arteries, pulmonary arteries and other great vessels in the chest [912]. However, no NEMRA technique has yet provided the combination of image quality, spatial resolution, scan speed, and resistance to artifacts from cardiac and respiratory motion needed to reliably evaluate the pulmonary arteries. To date, NEMRA has shown only modest sensitivity and specificity in clinical trials for pulmonary embolism [13, 14].
In this prospective technical feasibility study, breath-hold single-shot radial QISS MRA demonstrated pulmonary arterial anatomy from the main pulmonary artery through the level of the segmental branches in less than 2 min. Navigator-gated acquisitions were only slightly slower with scan times of 3.4 min or less, despite the use of both leading and trailing navigators to minimize blurring from respiration. Navigator-gated QISS tended to show slight blurring compared with breath-hold scans, but all pulmonary artery segments were adequately visualized.
In one patient, single-shot QISS clearly demonstrated multiple pulmonary emboli, whereas single-shot bSSFP showed poor conspicuity of the emboli. This case highlights that QISS, unlike bSSFP, is a highly flow-dependent imaging technique. A potential concern with QISS is saturation of in-plane flow, which is a common source of artifacts with conventional 2D time-of-flight MRA. However, we did not observe a significant degree of in-plane saturation in the pulmonary arteries through the segmental level. To maximize flow contrast, the QISS technique applies an in-plane inversion pulse prior to a quiescent interval of a few hundred milliseconds, which is then followed by a single-shot bSSFP readout. QISS differs from 2D time-of-flight MRA in that rapid systolic flow during the quiescent interval will tend to wash out saturated spins, even when the slice and vessel orientations are substantially aligned.
Cardiac arrhythmias are common in older patients due to hypertension and other co-morbidities [15]. We found that single-shot radial QISS, in contrast to navigator-gated 3D bSSFP, is resistant to motion artifacts from atrial fibrillation. The sub-second (~284 msec) readout duration for each single-shot radial QISS image effectively freezes cardiac motion and thereby avoids image artifacts. On the other hand, a navigator-gated 3D bSSFP acquisition accumulates data over dozens of cardiac cycles, so that beat-to-beat variations in cardiac dimensions and tissue signal cause image artifacts.
Compared with CEMRA, radial QISS eliminates the possibility of artifacts from imaging too early or late during the contrast infusion. The ability to repeat the scan and tailor scan planes and spatial resolution as needed, along with the capability for free-breathing acquisitions, represent other potential advantages over CEMRA, where flexibility is limited by the need to acquire all data during the first pass of contrast agent.
Further improvements in radial QISS image quality should be readily obtainable. For instance, given that these images are sparse and undersampled, iterative reconstruction techniques such as non-Cartesian SENSE or compressed sensing can be used to improve image quality [16, 17]. Simultaneous multi-slice imaging techniques can further shorten scan time, although to date most efforts have been directed to Cartesian rather than radial imaging [18].
There are several limitations to our study design. While radial QISS consistently demonstrated the pulmonary arteries down to the segmental level, imaging of smaller subsegmental branches could prove challenging due to the impact of off-resonance effects within the lung parenchyma relating to the use of a bSSFP readout. As more powerful gradient systems become available, it should be possible to shorten the bSSFP TR, thereby reducing off-resonance artifacts and improving the conspicuity of distal pulmonary arterial branches. The quality of the B0 shim is more critical with QISS, which uses a bSSFP readout, than with CEMRA, which uses a short-TE 3D spoiled gradient-echo readout. Only a small number of subjects were evaluated, and only one had significant pulmonary arterial pathology. While pulmonary emboli were anecdotally demonstrated using QISS in one patient, no assumptions can be made about the diagnostic accuracy of the technique.

Conclusions

In summary, this technical feasibility study has demonstrated that both breath-hold and free-breathing radial QISS can rapidly and consistently depict normal pulmonary arterial anatomy down to the segmental level with an exam time on the order of a few minutes. The technique unambiguously detected central pulmonary emboli in one patient, with much improved contrast compared with standard bSSFP. Further study will be required to determine the accuracy and utility of the radial QISS technique in patients suspected of pulmonary embolism.

Acknowledgements

We would like to thank Dr. Wei Li for assisting with data collection and analysis.

Funding

Research support, NIH grants R01 HL130093 and R21 HL126015. Research support, Siemens Healthcare. Research support, Department of Radiology, NorthShore University HealthSystem.

Availability of data and materials

Not applicable.

Authors’ contributions

RE participated in all aspects of the study and is the guarantor of study integrity. RS assisted with image analysis and manuscript review. KT assisted with image analysis and manuscript review. MM assisted with patient recruitment and manuscript review. JN assisted with patient recruitment and manuscript review. SG assisted with pulse sequence implementation and manuscript review. IK assisted with pulse sequence implementation, statistical analysis and manuscript review. All authors read and approved the manuscript.

Authors’ information

None.

Competing interests

RE: Research support and invention licensing agreement, Siemens Healthcare.
SG: Employee, Siemens Healthcare.
There were no non-financial conflicts of interest for any of the authors.
Consent for publication of their individual details and accompanying images in this manuscript was prospectively provided in 13 subjects, the need for consent was retrospectively waived by the IRB in one patient. No protected health information for any subject is given in this manuscript.
The prospective volunteer/patient study (n = 13) was approved by the NorthShore University HealthSystem Institutional Review Board (IRB). For the retrospective study (n = 1), written consent was waived by the IRB.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
2.
Zurück zum Zitat Coresh J, Astor BC, Greene T, Eknoyan G, Levey AS. Prevalence of chronic kidney disease and decreased kidney function in the adult US population. Third National Health and nutrition examination survey. Am J Kidney Dis. 2003;41:1–12.CrossRefPubMed Coresh J, Astor BC, Greene T, Eknoyan G, Levey AS. Prevalence of chronic kidney disease and decreased kidney function in the adult US population. Third National Health and nutrition examination survey. Am J Kidney Dis. 2003;41:1–12.CrossRefPubMed
3.
Zurück zum Zitat Nagle SK, Schiebler ML, Repplinger MD, François CJ, Vigen KK, Yarlagadda R, et al. Contrast enhanced pulmonary magnetic resonance angiography for pulmonary embolism: building a successful program. Eur J Radiol. 2016;85(3):553–63. doi:10.1016/j.ejrad.2015.12.018. Epub 2015 Dec 29CrossRefPubMed Nagle SK, Schiebler ML, Repplinger MD, François CJ, Vigen KK, Yarlagadda R, et al. Contrast enhanced pulmonary magnetic resonance angiography for pulmonary embolism: building a successful program. Eur J Radiol. 2016;85(3):553–63. doi:10.​1016/​j.​ejrad.​2015.​12.​018. Epub 2015 Dec 29CrossRefPubMed
5.
Zurück zum Zitat Edelman RR, Giri S, Pursnani A, Botelho MPF, Li W, Koktzoglou I. Breath-hold imaging of the coronary arteries using quiescent-interval slice-selective (QISS) magnetic resonance angiography: pilot study at 1.5 Tesla and 3 Tesla. J Cardiovasc Magn Reson. 2015;17:101. PMID 26597281CrossRefPubMedPubMedCentral Edelman RR, Giri S, Pursnani A, Botelho MPF, Li W, Koktzoglou I. Breath-hold imaging of the coronary arteries using quiescent-interval slice-selective (QISS) magnetic resonance angiography: pilot study at 1.5 Tesla and 3 Tesla. J Cardiovasc Magn Reson. 2015;17:101. PMID 26597281CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Glover GH, Pauly JM. Projection reconstruction techniques for reduction of motion effects in MRI. Magn Reson Med. 1992;28:275–89.CrossRefPubMed Glover GH, Pauly JM. Projection reconstruction techniques for reduction of motion effects in MRI. Magn Reson Med. 1992;28:275–89.CrossRefPubMed
7.
Zurück zum Zitat Edelman RR, Botelho M, Pursnani A, Giri S, Koktzoglou I. Improved dark blood imaging of the heart using radial balanced steady-state free precession. J Cardiovasc Magn Reson. 2016 Oct 19;18(1):69.CrossRefPubMedPubMedCentral Edelman RR, Botelho M, Pursnani A, Giri S, Koktzoglou I. Improved dark blood imaging of the heart using radial balanced steady-state free precession. J Cardiovasc Magn Reson. 2016 Oct 19;18(1):69.CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Jackson CL, Huber JF. Correlated applied anatomy of the bronchial tree and lungs with a system of nomenclature. Dis Chest. 1943;9:319–26.CrossRef Jackson CL, Huber JF. Correlated applied anatomy of the bronchial tree and lungs with a system of nomenclature. Dis Chest. 1943;9:319–26.CrossRef
9.
Zurück zum Zitat François CJ, Tuite D, Deshpande V, Jerecic R, Weale P, Carr JC. Pulmonary vein imaging with unenhanced three-dimensional balanced steady-state free precession MR angiography: initial clinical evaluation. Radiology. 2009;250(3):932–9. doi:10.1148/radiol.2502072137. Epub 2009 Jan 22CrossRefPubMed François CJ, Tuite D, Deshpande V, Jerecic R, Weale P, Carr JC. Pulmonary vein imaging with unenhanced three-dimensional balanced steady-state free precession MR angiography: initial clinical evaluation. Radiology. 2009;250(3):932–9. doi:10.​1148/​radiol.​2502072137. Epub 2009 Jan 22CrossRefPubMed
10.
Zurück zum Zitat Hui BK, Noga ML, Gan KD, Wilman AH. Navigator-gated three-dimensional MR angiography of the pulmonary arteries using steady-state free precession. J Magn Reson Imaging. 2005 Jun;21(6):831–5.CrossRefPubMed Hui BK, Noga ML, Gan KD, Wilman AH. Navigator-gated three-dimensional MR angiography of the pulmonary arteries using steady-state free precession. J Magn Reson Imaging. 2005 Jun;21(6):831–5.CrossRefPubMed
11.
Zurück zum Zitat Gebker R, Gomaa O, Schnackenburg B, Rebakowski J, Fleck E, Nagel E. Comparison of different MRI techniques for the assessment of thoracic aortic pathology: 3D contrast enhanced MR angiography, turbo spin echo and balanced steady state free precession. Int J Cardiovasc Imaging. 2007 Dec;23(6):747–56.CrossRefPubMed Gebker R, Gomaa O, Schnackenburg B, Rebakowski J, Fleck E, Nagel E. Comparison of different MRI techniques for the assessment of thoracic aortic pathology: 3D contrast enhanced MR angiography, turbo spin echo and balanced steady state free precession. Int J Cardiovasc Imaging. 2007 Dec;23(6):747–56.CrossRefPubMed
12.
Zurück zum Zitat Tomasian A, Lohan DG, Laub G, Singhal A, Finn JP, Krishnam MS. Noncontrast 3D steady state free precession magnetic resonance angiography of the thoracic central veins using nonselective radiofrequency excitation over a large field of view. Investig Radiol. 2008;43:306–13. PubMed: 18424951 CrossRef Tomasian A, Lohan DG, Laub G, Singhal A, Finn JP, Krishnam MS. Noncontrast 3D steady state free precession magnetic resonance angiography of the thoracic central veins using nonselective radiofrequency excitation over a large field of view. Investig Radiol. 2008;43:306–13. PubMed: 18424951 CrossRef
13.
Zurück zum Zitat Kluge A, Mueller C, Strunk J, Lange U, Bachmann G. Experience in 207 combined MRI examinations for acute pulmonary embolism and deep vein thrombosis. AJR Am J Roentgenol. 2006 Jun;186(6):1686–96.CrossRefPubMed Kluge A, Mueller C, Strunk J, Lange U, Bachmann G. Experience in 207 combined MRI examinations for acute pulmonary embolism and deep vein thrombosis. AJR Am J Roentgenol. 2006 Jun;186(6):1686–96.CrossRefPubMed
14.
Zurück zum Zitat Kalb B, Sharma P, Tigges S, et al. MR imaging of pulmonary embolism: diagnostic accuracy of contrast-enhanced 3D MR pulmonary angiography, contrast-enhanced low-flip angle 3D GRE, and nonenhanced free-induction FISP sequences. Radiology. 2012 Apr;263(1):271–8. doi:10.1148/radiol.12110224.CrossRefPubMed Kalb B, Sharma P, Tigges S, et al. MR imaging of pulmonary embolism: diagnostic accuracy of contrast-enhanced 3D MR pulmonary angiography, contrast-enhanced low-flip angle 3D GRE, and nonenhanced free-induction FISP sequences. Radiology. 2012 Apr;263(1):271–8. doi:10.​1148/​radiol.​12110224.CrossRefPubMed
15.
Metadaten
Titel
Nonenhanced MR angiography of the pulmonary arteries using single-shot radial quiescent-interval slice-selective (QISS): a technical feasibility study
verfasst von
Robert R. Edelman
Robert I. Silvers
Kiran H. Thakrar
Mark D. Metzl
Jose Nazari
Shivraman Giri
Ioannis Koktzoglou
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2017
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-017-0365-3

Weitere Artikel der Ausgabe 1/2017

Journal of Cardiovascular Magnetic Resonance 1/2017 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.