Open Access 01.12.2017 | Research article

Normal limits of the electrocardiogram derived from a large database of Brazilian primary care patients

verfasst von: Daniel M. F. Palhares, Milena S. Marcolino, Thales M. M. Santos, José L. P. da Silva, Paulo R. Gomes, Leonardo B. Ribeiro, Peter W. Macfarlane, Antonio L. P. Ribeiro

Erschienen in: BMC Cardiovascular Disorders | Ausgabe 1/2017

Abstract

Background

Knowledge of the normal limits of the electrocardiogram (ECG) is mandatory for establishing which patients have abnormal ECGs. No studies have assessed the reference standards for a Latin American population. Our aim was to establish the normal ranges of the ECG for pediatric and adult Brazilian primary care patients.

Methods

This retrospective observational study assessed all the consecutive 12-lead digital electrocardiograms of primary care patients at least 1 year old in Minas Gerais state, Brazil, recorded between 2010 and 2015. ECGs were excluded if there were technical problems, selected abnormalities were present or patients with selected self-declared comorbidities or on drug therapy. Only the first ECG from patients with multiple ECGs was accepted. The University of Glasgow ECG analysis program was used to automatically interpret the ECGs. For each variable, the 1st, 2nd, 50th, 98th and 99th percentiles were determined and results were compared to selected studies.

Results

A total of 1,493,905 ECGs were recorded. 1,007,891 were excluded and 486.014 were analyzed. This large study provided normal values for heart rate, P, QRS and T frontal axis, P and QRS overall duration, PR and QT overall intervals and QTc corrected by Hodges, Bazett, Fridericia and Framingham formulae. Overall, the results were similar to those from other studies performed in different populations but there were differences in extreme ages and specific measurements.

Conclusions

This study has provided reference values for Latinos of both sexes older than 1 year. Our results are comparable to studies performed in different populations.
Abkürzungen
ECG
Electrocardiogram
MC
Minnesota code
QTc
Corrected QT
QTi
QT index
TNMG
Telehealth Network of Minas Gerais
Uni-G
The University of Glasgow ECG Analysis Program

Background

The electrocardiogram (ECG) is a noninvasive, easy to perform, low cost test of wide clinical utility for investigation of the cardiac electrical activity with established diagnostic significance [1]. Knowledge of the normal ranges of measurements of intervals and axes of the pediatric and adult ECG is mandatory for establishing which patients have abnormal ECGs and who may therefore need special medical management. The use of computerized programs for automated ECG interpretation has shown good accuracy levels for ECG interval measurements, with benefits in saving time and money, and thus its use has increased for ECG analysis in epidemiological studies [25]. Different authors have studied digital ECGs using automated interpretation, and some have developed the reference values in different populations: Chinese, Caucasian, Blacks, South Asians and others [613].
Despite their importance, certain studies appear to have some gaps that require further investigation to complement current knowledge. For instance, some studies included only a small number of age groups or had small sample sizes for subjects of extreme ages or did not study reference values in children. Other studies did not contain all possible ECG variables. Of importance is the fact that there is lack of data regarding the normal limits specific to a Latin American population. Given the frequent immigration of this population into North America and Europe and the known ECG variations in different populations [14], the study of the normal limits of the ECG in Latinos is of importance for medical staff around the world.
Therefore, the aim of this study was to establish the normal limits of ECG measurements in apparently healthy Brazilians by using a large sample of pediatric and adult primary care patients in whom ECGs were recorded with a modern digital electrocardiograph. Measurements were obtained with an internationally known well-validated ECG program [5, 15]. In addition, the software had the capability to undertake Minnesota Coding [16] using automated techniques [17].

Methods

Study population

This retrospective observational study assessed all the 12-lead digital electrocardiograms of primary care patients of at least 1 year old in the state of Minas Gerais, Brazil, whose exams were sent to the Telehealth Network of Minas Gerais (TNMG) between 1st January 2010 and 21th January 2015. TNMG is a public telehealth service that was created in 2005 to provide support to the poorest cities of the state. After successive expansions, this service now assists the primary care professionals in 750 of the 853 cities. Minas Gerais is a special Brazilian state that can be considered representative in comparison to the rest of the country because of two main reasons. Firstly, while the north and northeast of Minas Gerais has HDI and poverty rates similar to the poorest states of Brazil, the south and central region of the state resembles the richest states of the country. Secondly, Minas Gerais is located in the middle of the country (southeast) and is the fourth largest state (586,521 Km2) with the second largest number of inhabitants in the country (20,869,101 [1, 1820]).
ECGs were excluded from the study if any one of the following criteria was met:
  • There was interference, artifacts or electrode placement errors (Minnesota Code [MC] 9.8.1 or 9.8.2);
  • The ECG had an established abnormality: old myocardial infarction (major Q wave abnormalities [MC 1.1.x or 1.2.x]), possible old myocardial infarction (minor Q wave abnormalities plus ST-T abnormalities [MC 1.3 plus 4.1, 4.2, 5.1 or 5.2]), major isolated ST-T abnormalities (MC 4.1, 4.2, 5.1 or 5.2) complete or intermittent intraventricular blocks (MC 7.1, 7.2, 7.4 or 7.8), left ventricular hypertrophy plus ST-T abnormalities (MC 3.1 plus 4.1.x, 4.2, 5.1 or 5.2), major prolonged uncorrected QT interval (QTi ≥ 116%), major atrioventricular conduction abnormalities (MC 6.1, 6.2.x, 6.4, 6.8, 8.6.1 or 8.6.2), atrial fibrillation or flutter (MC 8–3-x), supraventricular tachycardia (MC 8.4.2 or 8–4-1 with heart rate > 140 bpm), wandering atrial pacemaker (MC 8.1.4), supraventricular rhythm persistent (MC 8.4.1 plus heart rate ≤ 140), high amplitude P wave (MC 9.3), asystole or ventricular fibrillation (MC 8.2);
  • Repeated exams: for the purpose of this analysis, only the first ECG from patients with multiple ECGs was analyzed.
  • Patient had these self-declared comorbidities or cardiovascular risk factors: arterial hypertension, diabetes, smoking, dyslipidemia, personal history of myocardial infarction, personal history of coronary revascularization, Chagas disease and chronic pulmonary disease;
  • Patient was receiving any kind of drug therapy (diuretics, digitalis, beta-blockers, angiotensin-converting-enzyme inhibitors, amiodarone, calcium channel blockers or any drug listed in the “others” field).
The study population was divided into 14 age groups: 01–02, 03–04, 05–07, 08–11, 12–15, 16–19, 20–29, 30–39, 40–49, 50–59, 60–69, 70–79, 80–89 years, and 90 years and older.

Data acquisition

Each primary care center received one of the two available digital electrocardiographs to record the digital 12-lead ECGs, ErgoPC 13 (MICROMED, Brazil) or ECG PC (TEB, Brazil), and one specific software that allowed entry of clinical information based on patient’s self-report and local medical record. The data acquisition rate of this two devices were 500 Hz and 600 Hz at 5 and 3.9 μV resolution, respectively. Thereafter, the ECGs with all personal information attached, were sent through the internet to the analysis center of the TNMG and were saved in a database that was used for data collection in the present study. All the patients’ data were collected by the primary care practitioners and represent self-declared information.

Database

For each patient, the database contains:
  • patient identification: full name, gender, birth date and city of domicile;
  • comorbidities and cardiovascular risk factors: arterial hypertension, obesity, diabetes, smoking, dyslipidemia, personal history of myocardial infarction, personal history of coronary revascularization, family history of coronary disease, Chagas disease and chronic pulmonary disease;
  • medications: diuretics, digitalis, beta-blockers, angiotensin-converting-enzyme inhibitors, amiodarone and calcium channel blockers) or free text field (“others”);
  • symptoms: chest pain and any equivalent;
  • anthropometric measures: height and weight;
  • 12 lead ECG recordings;
  • administrative information about the ECG: date and city in which it was performed.
Patients’ information (name, birth date and city) was standardized to avoid errors in identification, which could cause the inclusion of the same patient twice. Due to the frequent miscegenation in the Brazilian population, data from race was not considered in this study. Also, other information like pregnancy was not available and the healthcare professional could not fill out the form with blood pressure at the time the ECG was recorded.

The Glasgow program

The University of Glasgow (Uni-G) ECG analysis program (release 28.5, issued on January 2014) is internationally well-recognized computer software that was used to automatically interpret the ECGs in the TNMG database [3]. It has been in continuous development for over 25 years and it is applicable to neonates as well as adults [5, 15]. This program has been extensively evaluated, meeting the requirements of IEC 60601–2-25 and is used routinely world-wide. It provides all the standard amplitude, duration and axes measurements as well as a rhythm analysis and diagnostic interpretation [2123]. It is well suited for epidemiological studies [35, 15].
The Uni-G program uses strict standards of electrocardiographic interpretation and allows the export of two types of diagnostic statements, long and short:
□ Quantitative description: average heart rate; P, QRS, and T axes; P and QRS durations; PR and QT intervals; and corrected QT (QTc) by the methods of Framingham, Hodges, Bazett and Fridericia;
□ Qualitative description: the software uses the quantitative description to automatically classify the electrocardiographic abnormalities according to the Minnesota Code.
All the ECGs had their quantitative data analyzed, and reference values were established for each variable for the study population. QT index (QTi) was calculated according the formulae QTi = (QT / 656) x (heart rate + 100) [24, 25] and its prevalence was demonstrated in Fig. 1.

Comparison with other studies

A systematic review of available literature was undertaken to identify relevant publications in English about electrocardiographic reference values for heart rate, axis and measurements that were derived from digital ECGs and a sample size of at least one thousand patients. A comprehensive search was conducted in the electronic database PubMed (covering all dates up to July 30, 2015), using the following keywords and their combinations: “digital”, “electrocardiogram”, “electrocardiographic”, “ecg”, “ekg”, “reference values”, “reference ranges”, “normal limits”, “normal values” and “epidemiology”. Additional strategies included reviews of journals or periodicals not indexed in the above mentioned electronic database. From all the available studies, we chose four that included patients with age groups, percentiles and variables similar to those that were included in the present study and we compared all the results according age and sex based groups [7, 12, 13, 26]. Only one of them included data for percentiles 1 and 99 [13], but all included percentiles 2, 50 and 98 (Table 1). Because not all studies included all of the variables which we studied, some comparisons could not be undertaken among all the studies. Different age groups were not compared, except when we considered that the difference was irrelevant eg. 90-99 y versus age ≥ 90 y.
Table 1
Selected studies of normal ECG intervals
Study
Population
Year
Patients (n)
ECG acquisition
Sampling rate
Analytic program
Percentiles
Rijnbeek et al.
0 to 16 y; Dutch
2001
Male (944); Female (968)
Cardio Control (Delft, Netherlands)
1200 Hz
MEANSa
2nd, 50th and 98th
Wu et al.
18 to ≥ 60 y; Chinesese
2003
Male (3614); Female (1746)
Cardio Control (Delft, Netherlands)
1200 Hz
MEANSa
2nd, 50th and 98th
Mason et al.
0 to 99 years; Northern America (70%), Europe (21%), Africa (3%), Latin America (3%), and Asia (2%) and Oceania (2%).
2007
Male (14,297); Female (12,201)
MTX-2 (CCSSI); MAC 1200 (GE Medical Systems)
500 Hz
CCSSI
1st, 2nd, 50th, 98th, 99th
Rijnbeek et al.
16 to 89 y; Dutch
2014
Male (7326); Female (6028)
ACTA (ESAOTE, Florence, Italy); Cardio Perfect equipment (Welch Allyn Cardio Control, USA); Megacart (Siemens, Erlangen, Germany)
500 Hz
MEANSa
2nd, 50th and 98th
aModular ECG Analysis System (MEANS)

Statistical analysis and estimation of normal values

Descriptive statistics were computed for the whole database. Categorical data were reported as counts and percentages; continuous variables were reported as mean and standard deviation or median and percentiles, as appropriate. For each ECG variable, the median plus the 1st, 2nd, 98th and 99th percentiles of the measurement distribution per age and gender were determined. The 2nd percentile was taken as the lower limit and the 98th percentile as the upper limit of the normal range.
For analysis purposes, age was rounded to the nearest integer. Then, for each integer age, the reference percentile of the quantitative ECG variable was computed. Subsequently, the locally weighted polynomial regression method (LOESS) was used to graphically display a smoothed relationship for this reference percentile as a function of age for patients from 1 to 90 years old [27]. The span (smoothing) parameter was chosen after a visual inspection and was set to 0.40.
Graphs with the percentiles 2nd, 50th and 98th for each age group and both sexes were created to compare the result of different studies.
Data management and statistical computations were performed with IBM SPSS Statistics for Windows version 20.0 (IBM Corp. Released 2011. Armonk, NY: IBM Corp) and R statistical computing software version 3.2.0 with foreign and plyr packages. Loess curves were calculated using the R loess function [27].
This study was approved by the Research Ethics Committee of the Universidade Federal de Minas Gerais.

Results

Characteristics of the study population

A total of 1,493,905 ECGs were recorded during the study period. The exclusion criteria were applied in consecutive steps (Fig. 2). First, 18,619 ECGs with technical problems were excluded. From the 1,475,286 remaining ECGs, a total of 599,390 patients on various medicines or with selected ECG abnormalities or self-declared comorbidities were excluded. Finally, 389,882 ECGs from patients who had more than one ECG recorded during study period were ruled out. After all the exclusion criteria were applied, 486,014 ECGs were analyzed (patients’ mean age 42.2 ± 18.6 years, 58.8% females). Figure 3 and Table 2 describe the study population.
Table 2
Age and sex distribution of the study population
Age group (years)
Male
Female
Total
n
n
n
%
1–2
259
216
475
0.1%
3–4
840
667
1507
0.3%
5–7
2486
1947
4433
0.9%
8–11
6802
4601
11,403
2.4%
12–15
11,812
10,709
22,521
4.6%
16–19
11,295
14,453
25,748
5.3%
20–29
26,063
43,281
69,344
14.3%
30–39
31,138
55,671
86,809
17.9%
40–49
34,414
57,942
92,356
19.0%
50–59
32,988
46,563
79,551
16.4%
60–69
24,149
28,620
52,769
10.9%
70–79
13,499
15,280
28,779
5.9%
80–89
4120
4998
9118
1.9%
> = 90
482
719
1201
0.2%
Total
200,347
285,667
486,014
100%

Electrocardiographic findings

Global measurements

Tables 3 and 4 show percentiles 1st,2nd, 50th, 98th and 99th for heart rate, P, QRS and T frontal axis, overall P and QRS duration, overall P and QT interval and QTc corrected by Hodges, Bazett, Fridericia and Framingham methods for different age groups and both sexes. All global parameters showed age trends and sex differences [2831].
Table 3
ECG parameters for patients from 1 to 29 years-old: (1st percentile; 2nd percentile; 50th percentile; 98th percentile; 99th percentile)
Parameter
Sex
1–2
3–4
5–7
8–11
12–15
16–19
20–29
Heart rate (bpm)
Male
(52; 59; 107; 166; 170)
(67; 70; 97; 132; 139)
(59; 61; 86; 120; 126)
(53; 55; 76; 108; 114)
(47; 49; 69; 101; 107)
(44; 46; 65; 98; 105)
(44; 46; 65; 98; 104)
Female
(58; 66; 108; 163; 199)
(64; 69; 98; 138; 141)
(62; 64; 88; 124; 133)
(56; 58; 81; 115; 121)
(52; 54; 75; 107; 114)
(51; 53; 73; 105; 112)
(51; 53; 73; 104; 111)
P Frontal Axis (°)
Male
(−95; −10; 53; 85; 116)
(−30; −14; 50; 79; 96)
(−34; −22; 44; 74; 81)
(−35; −21; 41; 74; 79)
(−33; −20; 44; 77; 82)
(−32; −21; 54; 81; 85)
(−22; −13; 54; 80; 85)
Female
(−12; −6; 53; 78; 85)
(−30; −10; 51; 76; 90)
(−19; −11; 49; 74; 83)
(−21; −12; 45; 74; 78)
(−21; −13; 48; 76; 80)
(−22; −13; 51; 77; 82)
(−18; −11; 51; 78; 81)
QRS Frontal Axis (°)
Male
(−46; −26; 61; 124; 132)
(−21; −9; 61; 99; 111)
(−29; −10; 64; 101; 111)
(−31; −8; 64; 99; 107)
(−28; −11; 66; 101; 107)
(−41; −18; 68; 100; 107)
(−37; −20; 59; 96; 101)
Female
(−49; −20; 62; 102; 136)
(−22; −3; 66; 102; 108)
(−9; 0; 67; 100; 109)
(−14; 0; 66; 98; 104)
(−8; 2; 65; 95; 100)
(−13; 0; 63; 95; 99)
(−16; −5; 56; 89; 95)
T Frontal Axis (°)
Male
(−9; 2; 43; 70; 73)
(5; 9; 43; 65; 69)
(7; 13; 45; 67; 70)
(7; 12; 46; 68; 71)
(2; 9; 48; 71; 73)
(−3; 3; 47; 72; 75)
(−12; −5; 40; 71; 75)
Female
(8; 12; 43; 73; 77)
(−4; 0; 42; 68; 70)
(4; 9; 44; 68; 71)
(1; 8; 45; 69; 72)
(0; 4; 44; 69; 73)
(−2; 3; 42; 70; 73)
(−7; −1; 38; 68; 72)
Overall P Duration (ms)
Male
(63; 64; 84; 124; 142)
(64; 70; 86; 112; 124)
(64; 68; 88; 114; 128)
(68; 72; 94; 118; 132)
(68; 72; 98; 124; 134)
(68; 74; 102; 126; 134)
(72; 78; 106; 132; 138)
Female
(52; 63; 84; 115; 128)
(64; 68; 86; 115; 127)
(68; 72; 90; 116; 128)
(70; 72; 94; 116; 130)
(72; 76; 98; 124; 132)
(72; 78; 100; 128; 134)
(74; 78; 104; 132; 138)
Overall PR Interval (ms)
Male
(81; 86; 118; 184; 192)
(90; 92; 122; 170; 186)
(90; 94; 124; 174; 188)
(96; 100; 132; 180; 190)
(98; 104; 136; 188; 202)
(102; 106; 142; 196; 206)
(104; 110; 148; 200; 212)
Female
(84; 87; 118; 171; 178)
(88; 90; 120; 166; 171)
(90; 94; 124; 166; 173)
(94; 98; 128; 178; 188)
(98; 102; 136; 184; 196)
(98; 104; 138; 188; 198)
(102; 106; 142; 192; 202)
Overall QRS Duration (ms)
Male
(60; 62; 74; 98; 101)
(64; 66; 76; 92; 96)
(66; 66; 80; 99; 102)
(68; 70; 84; 102; 106)
(72; 74; 90; 108; 112)
(74; 76; 94; 114; 116)
(76; 78; 96; 114; 116)
Female
(58; 60; 72; 97; 100)
(62; 62; 74; 90; 94)
(62; 64; 78; 96; 98)
(66; 68; 80; 98; 100)
(68; 70; 84; 102; 106)
(70; 70; 86; 104; 108)
(70; 72; 88; 106; 110)
QT Interval (ms)
Male
(239; 250; 314; 412; 426)
(276; 281; 332; 396; 405)
(294; 300; 356; 412; 423)
(314; 322; 378; 436; 446)
(324; 332; 390; 452; 460)
(320; 328; 390; 454; 462)
(324; 332; 390; 456; 464)
Female
(224; 255; 314; 419; 454)
(271; 278; 332; 396; 412)
(286; 296; 350; 402; 408)
(306; 314; 368; 426; 434)
(314; 322; 380; 440; 450)
(316; 326; 382; 442; 450)
(320; 328; 386; 446; 454)
QTc Hodges (ms)
Male
(367; 370; 405; 444; 455)
(364; 369; 399; 430; 437)
(370; 375; 403; 436; 441)
(374; 377; 408; 440; 445)
(370; 374; 408; 446; 452)
(364; 368; 400; 442; 450)
(365; 368; 400; 442; 449)
Female
(363; 367; 404; 459; 470)
(369; 371; 401; 436; 440)
(371; 375; 403; 431; 437)
(373; 378; 407; 438; 442)
(375; 379; 409; 443; 449)
(373; 377; 408; 443; 449)
(375; 379; 411; 447; 452)
QTc Bazett (ms)
Male
(377; 384; 421; 459; 469)
(377; 384; 421; 460; 467)
(381; 388; 425; 463; 469)
(381; 386; 425; 463; 469)
(368; 374; 419; 462; 468)
(358; 364; 406; 452; 458)
(359; 364; 406; 451; 459)
Female
(380; 383; 421; 467; 479)
(382; 388; 424; 464; 470)
(385; 390; 425; 461; 467)
(384; 389; 427; 465; 471)
(382; 387; 426; 466; 471)
(377; 383; 423; 463; 469)
(379; 385; 426; 466; 471)
QTc Fridericia (ms)
Male
(332; 342; 381; 429; 434)
(349; 358; 389; 427; 432)
(361; 367; 400; 439; 443)
(371; 376; 408; 442; 447)
(368; 372; 408; 447; 453)
(362; 366; 400; 441; 448)
(363; 367; 400; 439; 446)
Female
(332; 339; 383; 445; 453)
(354; 358; 391; 431; 438)
(361; 366; 398; 433; 436)
(369; 374; 406; 440; 444)
(372; 377; 410; 446; 451)
(371; 375; 408; 444; 450)
(374; 378; 411; 448; 454)
QTc Framingham (ms)
Male
(336; 344; 382; 429; 435)
(355; 361; 390; 426; 430)
(364; 371; 401; 436; 441)
(374; 378; 409; 442; 446)
(368; 373; 409; 446; 452)
(359; 365; 400; 440; 447)
(361; 365; 400; 438; 444)
Female
(332; 345; 382; 441; 448)
(358; 359; 390; 428; 435)
(363; 369; 399; 430; 434)
(372; 376; 406; 439; 444)
(375; 379; 410; 444; 450)
(373; 377; 409; 444; 449)
(375; 380; 412; 447; 453)
Table 4
ECG parameters for patients older than 30 years-old: (1st percentile; 2nd percentile; 50th percentile; 98th percentile; 99th percentile)
Parameter
Sex
30–39
40–49
50–59
60–69
70–79
80–89
> = 90
Heart rate (bpm)
Male
(44; 46; 65; 99; 105)
(44; 46; 66; 98; 104)
(44; 46; 66; 98; 104)
(44; 46; 66; 99; 104)
(45; 47; 66; 99; 105)
(45; 47; 68; 104; 111)
(47; 49; 70; 109; 118)
Female
(50; 52; 72; 102; 107)
(49; 52; 71; 101; 107)
(48; 50; 70; 99; 105)
(48; 50; 70; 101; 106)
(48; 50; 71; 102; 106)
(48; 51; 73; 105; 111)
(50; 52; 76; 111; 120)
P Frontal Axis (°)
Male
(−15; −8; 53; 80; 85)
(−13; −5; 55; 80; 85)
(−12; −3; 57; 82; 87)
(−11; −3; 58; 84; 93)
(−12; −5; 60; 86; 95)
(−25; −9; 61; 94; 103)
(−35; −23; 61; 95; 101)
Female
(−15; −8; 52; 78; 82)
(−13; −5; 53; 79; 83)
(−12; −4; 53; 79; 83)
(−12; −5; 54; 80; 85)
(−13; −5; 57; 84; 94)
(−15; −7; 58; 85; 97)
(−45; −13; 59; 85; 99)
QRS Frontal Axis (°)
Male
(−41; −28; 47; 90; 96)
(−48; −36; 38; 85; 91)
(−56; −45; 31; 82; 85)
(−61; −51; 23; 80; 84)
(−65; −57; 18; 79; 84)
(−70; −62; 15; 78; 83)
(−69; −62; 15; 84; 98)
Female
(−24; −14; 46; 84; 87)
(−32; −23; 37; 80; 85)
(−41; −33; 28; 77; 82)
(−48; −40; 21; 74; 80)
(−53; −46; 17; 73; 79)
(−58; −51; 15; 72; 78)
(−63; −55; 13; 77; 80)
T Frontal Axis (°)
Male
(−18; −10; 35; 71; 75)
(−20; −12; 35; 73; 77)
(−23; −13; 39; 77; 83)
(−26; −15; 43; 81; 88)
(−27; −15; 48; 85; 95)
(−28; −18; 52; 90; 102)
(−37; −14; 55; 95; 99)
Female
(−12; −4; 36; 69; 72)
(−16; −8; 36; 72; 76)
(−18; −9; 38; 77; 83)
(−18; −8; 42; 81; 90)
(−21; −10; 46; 85; 95)
(−23; −12; 50; 89; 99)
(−26; −11; 52; 97; 104)
Overall P Duration (ms)
Male
(76; 82; 110; 136; 140)
(78; 84; 110; 136; 140)
(78; 84; 112; 138; 144)
(74; 82; 114; 140; 146)
(72; 78; 114; 142; 146)
(68; 74; 114; 146; 150)
(62; 72; 114; 146; 150)
Female
(74; 80; 106; 134; 138)
(76; 80; 106; 136; 140)
(76; 82; 108; 136; 142)
(76; 80; 110; 138; 144)
(72; 78; 112; 142; 146)
(68; 74; 112; 144; 150)
(68; 72; 112; 144; 148)
Overall PR Interval (ms)
Male
(108; 112; 150; 204; 216)
(108; 112; 152; 204; 216)
(106; 114; 152; 208; 220)
(108; 114; 154; 216; 228)
(108; 114; 158; 230; 246)
(108; 114; 164; 246; 268)
(106; 115; 166; 253; 289)
Female
(102; 106; 144; 194; 206)
(102; 106; 144; 194; 204)
(104; 110; 148; 200; 210)
(104; 110; 150; 208; 218)
(106; 110; 154; 214; 230)
(106; 112; 156; 232; 244)
(106; 108; 160; 241; 251)
Overall QRS Duration (ms)
Male
(76; 78; 96; 114; 116)
(76; 78; 94; 114; 116)
(74; 76; 94; 114; 116)
(74; 76; 94; 114; 116)
(74; 76; 94; 114; 116)
(72; 76; 94; 116; 116)
(70; 72; 92; 116; 118)
Female
(70; 72; 88; 108; 110)
(72; 72; 88; 108; 112)
(72; 74; 90; 110; 112)
(72; 74; 90; 110; 114)
(70; 72; 90; 112; 114)
(70; 72; 88; 112; 116)
(66; 70; 88; 110; 112)
QT Interval (ms)
Male
(324; 334; 392; 458; 468)
(328; 336; 396; 460; 472)
(328; 336; 398; 466; 478)
(328; 338; 402; 472; 484)
(326; 336; 404; 474; 486)
(320; 332; 406; 476; 488)
(307; 319; 400; 480; 491)
Female
(326; 334; 392; 452; 460)
(328; 338; 396; 458; 468)
(332; 340; 402; 466; 474)
(330; 338; 402; 470; 480)
(328; 338; 402; 470; 478)
(320; 332; 400; 470; 480)
(309; 319; 394; 473; 480)
QTc Hodges (ms)
Male
(368; 372; 404; 445; 451)
(371; 375; 407; 450; 456)
(374; 378; 411; 454; 461)
(375; 380; 414; 459; 466)
(377; 382; 417; 462; 468)
(378; 383; 421; 465; 470)
(380; 385; 420; 471; 478)
Female
(379; 383; 414; 452; 458)
(381; 385; 418; 456; 462)
(382; 387; 421; 460; 466)
(383; 388; 423; 464; 470)
(383; 388; 424; 464; 469)
(382; 388; 425; 465; 470)
(381; 385; 423; 465; 470)
QTc Bazett (ms)
Male
(363; 369; 411; 457; 463)
(367; 372; 414; 462; 468)
(370; 376; 418; 465; 471)
(373; 379; 421; 468; 474)
(375; 382; 426; 472; 478)
(379; 385; 431; 475; 479)
(375; 386; 434; 479; 484)
Female
(382; 388; 429; 470; 476)
(385; 391; 432; 473; 478)
(386; 392; 435; 475; 480)
(386; 393; 437; 476; 481)
(389; 395; 439; 478; 481)
(389; 396; 441; 480; 483)
(383; 397; 442; 483; 485)
QTc Fridericia (ms)
Male
(367; 371; 404; 443; 449)
(370; 374; 407; 448; 454)
(372; 377; 411; 452; 458)
(374; 378; 414; 456; 461)
(374; 380; 418; 461; 465)
(376; 381; 421; 463; 467)
(377; 380; 420; 471; 473)
Female
(377; 382; 416; 454; 459)
(379; 384; 419; 458; 463)
(381; 386; 423; 461; 465)
(381; 387; 425; 463; 468)
(380; 386; 426; 464; 467)
(377; 384; 427; 465; 469)
(371; 382; 425; 468; 470)
QTc Framingham (ms)
Male
(365; 370; 404; 442; 448)
(368; 373; 407; 447; 453)
(371; 376; 410; 451; 457)
(373; 378; 414; 455; 460)
(375; 380; 417; 460; 464)
(376; 382; 420; 462; 466)
(375; 380; 420; 470; 473)
Female
(379; 383; 416; 453; 458)
(381; 386; 419; 457; 462)
(382; 387; 423; 460; 464)
(383; 388; 424; 463; 467)
(381; 388; 426; 463; 467)
(380; 386; 426; 465; 468)
(375; 381; 424; 465; 469)

Heart rate

Figure 4 shows our reference values for heart rate. In males, it decreases during childhood and adolescence, reaches a median value of 65 bpm by the age of 16, stabilizes around 65–66 bpm from 20 to 79 years and increases to 70 bpm after 90 years-old. In females, the heart seems to beat faster. It also decreases during the first part of life, but reaches a median value of 73 bpm at age 16, then fluctuates around 70 to 73 bpm until 79 years old and increases to 76 bpm in women older than 90 years-old.
The percentiles for heart rate were compared to those of Mason [13], Wu [7] and Rijnbeek [12, 26] studies (Fig. 5), and we considered that our results were similar for most age groups in both sexes. However, some major differences were seen in results for men from 1-2 y, 16-19 y, 80-89 y and ≥90 y, and women from 1-2y, 3-4 y, 80-89 y and ≥90 y.

P, QRS and T frontal axis

Figures 6, 7 and 8 show our results for the normal limits of the P, QRS and T-wave frontal axes, respectively.
With respect to the P wave, the male P axis is generally, after 8 years of age, slightly more orientated in the inferior direction compared to that of females. For both males and females, the median P-wave frontal axis shifts superiorly during childhood, until 8–11 years, and then turns inferiorly with advancing age. Notwithstanding these small changes, the P wave axis is relatively constant throughout life.
It is well known that the median QRS axis shifts counterclockwise in the neonate and infant in the weeks and months after birth [32]. In our study, where the youngest participant was aged 1 year, the median QRS wave frontal axis shifted inferiorly for both sexes during childhood and by the age of 12–15 for females and 16–19 for males, it shifted superiorly, reaching 15 and 13 degrees for men and women, respectively, after 90 years-old. Although the median value is very similar for both sexes, men have a wider reference range during the whole of life.
The T wave frontal axis is very similar for men and women. The median varies from 43 to 38 and 43 to 45 degrees for men and women, respectively, until 20 years of age. In adult life, it rotates superiorly to 35–40 degrees but then moves inferiorly after 50 years of age for both sexes. The upper and lower limits tend to diverge progressively with aging.
Figures 9, 10 and 11 compares our results to the others. The median P frontal axis was quite similar for all ages and sexes to that of Rijnbeek [12, 26], although the percentiles 2 and 98 varied a little for males and females younger than 4 years and older than 80 years.
Although there were similar values for P frontal axis, Rijnbeek’s [12, 26] and Wu’s [7] studies showed larger differences for the QRS and T frontal axes, and in these studies, the axes were seen to be orientated more inferiorly than ours. This difference happened in both sexes, but was more important for males than females. For the QRS frontal axis, our values were more comparable to those of Mason [13] than the others.

Overall P and QRS duration

Figure 12 shows our results for the normal limits of the overall P duration, which is slightly greater for men than for women. The median value increases from about 85 ms for both sexes in childhood to 114 ms in men and 112 ms in women after 90 years-old.
Figure 13 shows our results for the overall QRS duration. The median duration has an important rise from the early phase of life until early adulthood in both sexes. Then, the median QRS duration remains stable around 92-96 ms and 88–90 ms for men and women, respectively. The median P and QRS durations are higher for men in almost all age groups.
Figures 14 and 15 compares our results to the other studies. The median values of overall P duration for both sexes and all ages were similar to those of to Rijnbeek [12, 26] and Wu [7]. However, some differences were observed for the 2nd percentile from older than 60 years and for the 98th percentile from 1 to 4 years, in males and females.
The comparison of overall QRS duration showed similar results for the first, second and fifth percentiles in all studies, for both sexes and all age groups. However, percentiles 98 and 99 appeared to be slightly different for men older than 90 years in Mason’s study and also the percentile 98 for men from 16 to 29 years and 70–89 years in Rijnbeek’s [12, 26] study. For women, a small difference was seen for the 2nd percentile in subjects older than 90 years in Mason’s [13] study and for the 98th percentile from 1 to 2 years in comparison with Rijnbeek’s [12, 26] study.

Overall PR intervals

Figure 16 shows our results for the normal limits of the overall PR intervals. The median value is slightly higher for men in all age groups and increases with age in both sexes. In age groups 30–39 and 80–89 years, the medians are, respectively, 150 ms and 164 ms for men and 144 ms and 156 ms for women.
The median overall PR intervals were compared in all studies (Fig. 17) and they appeared to have similar values. However, some discrepant values were seen in extreme age groups in both sexes. Some discrepancies were seen in the 2nd percentile in men older than 90 years in Mason’s [13] study and in the 98th percentile for children from 1 to 7 years and men older than 80 years in Rijnbeek’s [12, 26] studies and in men older than 70 years and women older than 90 years old in Mason’s [13] study.

QT interval

Figure 18 shows our results for the normal limits of the QT interval. It increases progressively from 1 to 89 years old in men and from 1 to 79 in women and reduces a little in the last phase of life. The values are quite similar in both sexes for each age group.
Figure 19 compares it to the other studies. For men, the values of QT interval in our study were quite similar to those of Rijnbeek [12], while the values from Mason [13] and Wu [7] were longer than ours. For women, all the studies had similar values, except the percentiles 1, 2, 98 and 99 for subjects older than 90 years in Mason’s [13] study, which appeared to have a wide range.

QTc Hodges, Bazett, Fridericia, Framingham [2831]

Figures 20, 21, 22 and 23 show our normal limits of the QTc using four different methods of correction. Women have higher medians, upper and lower limits than men in almost all age groups. The Bazett method showed higher values in both sexes for all ages in comparison to the other methods, whereas Hodges, Fridericia and Framingham showed values similar to each other. In the Bazett method, the median and 2nd and 98th percentiles have a fast decrease during childhood and adolescence, followed by a progressive increase until they reach a median 421 ms and 437 ms for men and women, respectively, from 60 to 69 years-old. Hodges, Fridericia and Framingham have similar behaviors: their median and percentiles initially describe an upward concavity during childhood. After adolescence, they have a linear increase with aging. The median for men and women from 60 to 69 years-old is, respectively, 421 ms and 437 ms (Bazett), 414 ms and 423 ms (Hodges), 414 ms and 425 ms (Fridericia) and 414 ms and 424 ms (Framingham).
The corrected QT intervals were compared to other studies as shown in Figs. 24, 25, 26 and 27. The QTc Hodges and QTc Framingham were very similar in our study and in the study of Rijnbeek [12], and the only large difference occurred in the 98th percentile for men and the 2nd and 98th percentiles for women from 80 to 90 years old. QTc Bazett and QTc Fridericia were compared to the results of Mason [13] and Rijnbeek [12, 26]. Overall, the results of Rijnbeek’s [7, 12, 26] study were quite similar to ours, whereas Mason’s [13] study had higher values than ours, for all percentiles, irrespective of age and sex.

Discussion

We studied the normal limits of the ECG in almost 500,000 pediatric and adult patients from primary care centers in the state of Minas Gerais Brazil, using automated analysis by an internationally recognized computer program [5]. To our knowledge, this is the largest study that has derived ECG reference values and it is the first performed exclusively in Latinos. Other authors have undertaken similar studies in different populations, but using smaller sample sizes, without including all the variables and age groups that we did [613].
Our population was large enough to permit very reliable measures from all age groups for both sexes, even at the extremes of age. In addition, the large sample size was important to create graphs with minimal smoothing, which would not be possible in smaller populations. For example, we created 14 age groups with an average of 34,715 patients in each group (ranging from 475 to 92,356) and included more than 39,000 patients in their seventies or older. We divided the patients into age groups similar to other studies and chose percentiles that had previously been adopted to allow comparison of reference values. Although the Brazilian population has a high level of miscegenation, we did not split up this population according to race.
Overall, our results were similar to those of other studies. Some larger differences were observed in extreme ages and extreme percentiles, which might have occurred mainly because of the small number of patients of extreme age in other studies but also due to differences in study population (race; inclusion or exclusion criteria).
Our study included 39,098 patients older than 70 years old while the studies of Mason [13] and Rijnbeek [12] included 5139 and 942 patients, respectively, for the same age group. In addition, for children, we included 1982 patients from 1 to 4 years old, while Rijnbeek [26] studied only 363. As we had a considerably greater number of patients in extreme age groups, we consider that for young children and seniors our results might be more stable than the data from other studies.
Most previous studies recruited healthy volunteers from the community with a normal cardiovascular system and included a clinical examination in order to exclude those with evidence of any illness likely to affect the cardiovascular system and possibly the ECG. As we used a database that contained each study participant’s ECG and clinical information based on a patient’s self-report and medical record, we excluded individuals with known clinical conditions likely to affect the cardiovascular system, those on any kind of drugs and/or with frankly abnormal Minnesota Codes.. We opted to maintain the 6454 (1.3%) patients with premature beats (ventricular, supraventricular or combined) since it might happened in the “normal” population [8]. In summary, we have excluded patients with conditions that might affect the cardiovascular system and we accepted that the remaining population was an “apparently healthy population”. However, since the fulfillment of patient data was undertaken by professionals in primary care, some diagnostic errors or underreporting of latent illness may have occurred, and this is one of our study limitations. On the other hand, the extremely large number of study participants should ensure that any measurements outliers due to such shortcomings will not have significantly affected the normal 2nd to 98th percentile normal ranges.
In broad terms, our results for QTc are very similar to the detailed paper by Luo et al. [33] who showed that QT corrected by Bazett was generally out of line with QTc corrected by the Hodges, Fridericia and Framingham methods.
Results of PR interval measurements are also of considerable interest. An upper limit of normal of 220 ms is often suggested but our data show two points. First of all, the large sample size indicates that males have a longer PR interval than females by around 20 ms which may be accounted for by heart size and secondly the upper limit of normal increases rapidly above 70 years of age from 220 ms to 260 ms at age 90 in men and from 200 ms to 240 ms in women.
Once all the measures were made using computers, the reference values might be valid for computerized analyses and not manual measurements on paper. One other factor that has to be acknowledged is that the automated approach itself is subject to what might be termed inter-program variation analogous to differences in measurements between two or more cardiologists, i.e. inter observer variation. In a comparison of 4 major algorithms using the same ECGs, it was shown that the mean QT interval varied by 8 ms in a group of 200 normal subjects [21]. Although the sample size was small, the inference is clear that some variation in QT measurements between studies is due to the different algorithms used to derive the results. On this subject, it should be noted that the other four studies used different hardware and software as well as different analysis programs, which might be a confounding factor for accurately compare to our data. Also, their sample selection methods and populations sampled were different with the potential to affect direct measurement comparisons.
Although the ECG intervals are influenced by sex and age, simpler reference values like normal heart rate 60-100 bpm or QRS duration <100 ms for all ages and both sexes, are commonly used in primary care centers, and even in medical education. We recommend that those old normal ranges should be discarded and replaced by age and sex-specific values. In that respect, we consider that graphs and tables with the percentiles may help general practitioners undertaking ECG analysis.
Finally, this paper has concentrated on the comparison of ECG interval and axes measurements in different populations. In due course, it is hoped to produce a detailed comparison for the major ECG wave component amplitudes.

Conclusion

This study contributes to the knowledge of the reference values in Latinos. Although we have seen small differences in comparison to studies in other populations, reference values for electrocardiogram intervals and axes in Latinos are in general comparable to those obtained in those other populations.

Acknowledgments

The authors thank the entire team of the Telehealth Network of Minas Gerais, specially Dr. Maria Beatriz M. Alkmim.

Funding

This study was funded by FAPEMIG (Fundação de Amparo à Pesquisa de Minas Gerais, RED-00004-14) and CNPq (Conselho Nacional de Desenvolvimento Científico e Tecnológico, 465518/2014-1) and publication fees were supported by the Pró-Reitoria de Pesquisa e Pós-Graduação em Ciências da Saúde: Infectologia e Medicina Tropical da Universidade Federal de Minas Gerais. The Telehealth Network of Minas Gerais is mainly funded by the State Government of Minas Gerais, by its Health Department (Secretaria de Estado da Saúde de Minas Gerais), but also by the Ministry of Health, Ministry of Science and Technology, and Brazilian research agencies FINEP (Financiadora de Estudos e Projetos, Projeto SIGTEL— Convênio 01.10.0717.00) FAPEMIG and CNPq. ALPR is recipient of scholarships from CNPq (Bolsa de Produtividade em Pesquisa) and FAPEMIG (Pesquisador Mineiro).

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Authors’ contributions

DMFP, PWM, MSM and ALPR contributed to the conception and study design. DMFP, MSM, TMMS, JLPS, PRG, LBR, PWM and ALPR were involved in the creation of the database, in drafting and revising the manuscript; DMFP, MSM, TMMS, JLPS, and ALPR were involved in the analysis of the data. Additionally, MSM and ALPR were responsible to coordinate and guide all the steps. All authors read and approved the final manuscript.

Competing interests

This is an independent study. The authors declare that they have no competing interests.
Not applicable.
This study was approved by the ethics committee of the Universidade Federal de Minas Gerais (COEP-UFMG): CAAE – 37,600,914.7.0000.5149. The data from the Telehealth Network of Minas Gerais was used with the written permission of the coordinator of the Telehealth Network of Minas Gerais, Dr. Maria Beatriz Moreira Alkmim. Due to the retrospective nature of this research and considering the great number of involved patients, the authors could not have the informed consent from each patient. Therefore, the data was used without identifying or exposing any patient and with the permission of the ethics committee of the Universidade Federal de Minas Gerais which approved the study. The Professor Peter W. Macfarlane, who is one of the developers of the University of Glasgow (Uni-G) ECG analysis program, allowed the TNMG to use that program for the purpose of this study.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Ribeiro AL, Alkmim MB, Cardoso CS, Carvalho GG, Caiaffa WT, Andrade MV, da Cunha DF, Antunes AP, Resende AGA, Resende ES. Implementation of a telecardiology system in the state of Minas Gerais: the minas Telecardio project. Arq Bras Cardiol. 2010;95(1):70–8.CrossRefPubMed Ribeiro AL, Alkmim MB, Cardoso CS, Carvalho GG, Caiaffa WT, Andrade MV, da Cunha DF, Antunes AP, Resende AGA, Resende ES. Implementation of a telecardiology system in the state of Minas Gerais: the minas Telecardio project. Arq Bras Cardiol. 2010;95(1):70–8.CrossRefPubMed
2.
Zurück zum Zitat Kors JA, van Herpen G. The coming of age of computerized ECG processing: can it replace the cardiologist in epidemiological studies and clinical trials? Stud Health Technol Inform. 2001;84(Pt 2):1161–7.PubMed Kors JA, van Herpen G. The coming of age of computerized ECG processing: can it replace the cardiologist in epidemiological studies and clinical trials? Stud Health Technol Inform. 2001;84(Pt 2):1161–7.PubMed
3.
Zurück zum Zitat Macfarlane PW, Devine B, Clark E. The university of glasgow (Uni-G) ECG analysis program. Comput Cardiol, 2005; 2005 25-28 Sept. 2005. Macfarlane PW, Devine B, Clark E. The university of glasgow (Uni-G) ECG analysis program. Comput Cardiol, 2005; 2005 25-28 Sept. 2005.
4.
Zurück zum Zitat Salerno SM, Alguire PC, Waxman HS. Competency in interpretation of 12-lead electrocardiograms: a summary and appraisal of published evidence. Ann Intern Med. 2003;138(9):751–60.CrossRefPubMed Salerno SM, Alguire PC, Waxman HS. Competency in interpretation of 12-lead electrocardiograms: a summary and appraisal of published evidence. Ann Intern Med. 2003;138(9):751–60.CrossRefPubMed
5.
Zurück zum Zitat Willems JL, Abreu-Lima C, Arnaud P, van Bemmel JH, Brohet C, Degani R, Denis B, Gehring J, Graham I, van Herpen G, Machado H, Macfarlane PW, Michaelis J, Moulopoulos SD, Rubel P, Zywietz C. The diagnostic performance of computer programs for the interpretation of electrocardiograms. N Engl J Med. 1991;325(25):1767–73.CrossRefPubMed Willems JL, Abreu-Lima C, Arnaud P, van Bemmel JH, Brohet C, Degani R, Denis B, Gehring J, Graham I, van Herpen G, Machado H, Macfarlane PW, Michaelis J, Moulopoulos SD, Rubel P, Zywietz C. The diagnostic performance of computer programs for the interpretation of electrocardiograms. N Engl J Med. 1991;325(25):1767–73.CrossRefPubMed
6.
Zurück zum Zitat Chen CY, Chiang BN, Macfarlane PW. Normal limits of the electrocardiogram in a Chinese population. J Electrocardiol. 1989;22(1):1–15.CrossRefPubMed Chen CY, Chiang BN, Macfarlane PW. Normal limits of the electrocardiogram in a Chinese population. J Electrocardiol. 1989;22(1):1–15.CrossRefPubMed
7.
Zurück zum Zitat Wu J, Kors JA, Rijnbeek PR, van Herpen G, Lu Z, Xu C. Normal limits of the electrocardiogram in Chinese subjects. Int J Cardiol. 2003;87(1):37–51.CrossRefPubMed Wu J, Kors JA, Rijnbeek PR, van Herpen G, Lu Z, Xu C. Normal limits of the electrocardiogram in Chinese subjects. Int J Cardiol. 2003;87(1):37–51.CrossRefPubMed
8.
Zurück zum Zitat Macfarlane PW, van Oosterom A, Pahlm O, Kligfield P, Janse J, Camm J. Comprehensive electrocardiology. London: Springer-Verlag; 2011. Macfarlane PW, van Oosterom A, Pahlm O, Kligfield P, Janse J, Camm J. Comprehensive electrocardiology. London: Springer-Verlag; 2011.
9.
Zurück zum Zitat Katibi I, Clark EN, Devine B, Lloyd SM, Macfarlane PW. Normal limits of the electrocardiogram in Nigerians. J Electrocardiol. 2013;46(4):289–95.CrossRefPubMed Katibi I, Clark EN, Devine B, Lloyd SM, Macfarlane PW. Normal limits of the electrocardiogram in Nigerians. J Electrocardiol. 2013;46(4):289–95.CrossRefPubMed
10.
Zurück zum Zitat Dewhurst MJ, Di Marco LY, Dewhurst F, Adams PC, Murray A, Orega GP, Mwita JC, Walker RW, Langley P. Electrocardiographic reference values for a population of older adults in sub-Saharan Africa. Ann Noninvasive Electrocardiol. 2014;19(1):34–42.CrossRefPubMed Dewhurst MJ, Di Marco LY, Dewhurst F, Adams PC, Murray A, Orega GP, Mwita JC, Walker RW, Langley P. Electrocardiographic reference values for a population of older adults in sub-Saharan Africa. Ann Noninvasive Electrocardiol. 2014;19(1):34–42.CrossRefPubMed
11.
Zurück zum Zitat Macfarlane PW, Lloyd SM, Singh D, Hamde S, Clark E, Devine B, et al. Normal limits of the electrocardiogram in Indians. J Electrocardiol. 2015;48(4):652–68.CrossRefPubMed Macfarlane PW, Lloyd SM, Singh D, Hamde S, Clark E, Devine B, et al. Normal limits of the electrocardiogram in Indians. J Electrocardiol. 2015;48(4):652–68.CrossRefPubMed
12.
Zurück zum Zitat Rijnbeek PR, van Herpen G, Bots ML, Man S, Verweij N, Hofman A, et al. Normal values of the electrocardiogram for ages 16–90 years. J Electrocardiol. 2014;47(6):914–21.CrossRefPubMed Rijnbeek PR, van Herpen G, Bots ML, Man S, Verweij N, Hofman A, et al. Normal values of the electrocardiogram for ages 16–90 years. J Electrocardiol. 2014;47(6):914–21.CrossRefPubMed
13.
Zurück zum Zitat Mason JW, Ramseth DJ, Chanter DO, Moon TE, Goodman DB, Mendzelevski B. Electrocardiographic reference ranges derived from 79,743 ambulatory subjects. J Electrocardiol. 2007;40(3):228–34.CrossRefPubMed Mason JW, Ramseth DJ, Chanter DO, Moon TE, Goodman DB, Mendzelevski B. Electrocardiographic reference ranges derived from 79,743 ambulatory subjects. J Electrocardiol. 2007;40(3):228–34.CrossRefPubMed
14.
Zurück zum Zitat Macfarlane PW, Katibi IA, Hamde ST, Singh D, Clark E, Devine B, Francq BG, Lloyd S, Kumar V. Racial differences in the ECG--selected aspects. J Electrocardiol. 2014;47(6):809–14.CrossRefPubMed Macfarlane PW, Katibi IA, Hamde ST, Singh D, Clark E, Devine B, Francq BG, Lloyd S, Kumar V. Racial differences in the ECG--selected aspects. J Electrocardiol. 2014;47(6):809–14.CrossRefPubMed
16.
Zurück zum Zitat Prineas RJ, Crow RS, Blackburn H. The Minnesota code manual of electrocardiographic findings. Wright, Bristol: Standards and Procedures for Measurement and Classification; 1982. Prineas RJ, Crow RS, Blackburn H. The Minnesota code manual of electrocardiographic findings. Wright, Bristol: Standards and Procedures for Measurement and Classification; 1982.
17.
Zurück zum Zitat Macfarlane PW, Latif S. Automated serial ECG comparison based on the Minnesota code. J Electrocardiol. 1996;29 Suppl:29–34. Macfarlane PW, Latif S. Automated serial ECG comparison based on the Minnesota code. J Electrocardiol. 1996;29 Suppl:29–34.
18.
Zurück zum Zitat Marcolino MS, Alkmim MB, Assis TGP, Palhares DMF, Silva GAC, Cunha LR, Sousa L, Abreu MP, Figueira RM, Ribeiro AL. A Rede de Teleassistência de Minas Gerais e suas contribuições para atingir os princípios de universalidade, equidade e integralidade do SUS - relato de experiência. R Eletr de Com Inf Inov Saúde 2013 2013;7(2). Marcolino MS, Alkmim MB, Assis TGP, Palhares DMF, Silva GAC, Cunha LR, Sousa L, Abreu MP, Figueira RM, Ribeiro AL. A Rede de Teleassistência de Minas Gerais e suas contribuições para atingir os princípios de universalidade, equidade e integralidade do SUS - relato de experiência. R Eletr de Com Inf Inov Saúde 2013 2013;7(2).
19.
Zurück zum Zitat Alkmim MB, Figueira RM, Marcolino MS, Cardoso CS. Pena de Abreu M, Cunha LR, da Cunha DF, Antunes AP, Resende AG, Resende ES, Ribeiro AL. Improving patient access to specialized health care: the Telehealth Network of Minas Gerais, Brazil. Bull World Health Organ. 2012;90(5):373–8.CrossRefPubMedPubMedCentral Alkmim MB, Figueira RM, Marcolino MS, Cardoso CS. Pena de Abreu M, Cunha LR, da Cunha DF, Antunes AP, Resende AG, Resende ES, Ribeiro AL. Improving patient access to specialized health care: the Telehealth Network of Minas Gerais, Brazil. Bull World Health Organ. 2012;90(5):373–8.CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Kligfield P, Badilini F, Rowlandson I, Xue J, Clark E, Devine B, de Bie J, Mortara D, Babaeizadeh S, Gregg R, Helfenbein ED, Green CL. Comparison of automated measurements of electrocardiographic intervals and durations by computer-based algorithms of digital electrocardiographs. Am Heart J. 2014;167(2):150–9. e1CrossRefPubMed Kligfield P, Badilini F, Rowlandson I, Xue J, Clark E, Devine B, de Bie J, Mortara D, Babaeizadeh S, Gregg R, Helfenbein ED, Green CL. Comparison of automated measurements of electrocardiographic intervals and durations by computer-based algorithms of digital electrocardiographs. Am Heart J. 2014;167(2):150–9. e1CrossRefPubMed
22.
Zurück zum Zitat Macfarlane PW, Devine B, Latif S, McLaughlin S, Shoat DB, Watts MP. Methodology of ECG interpretation in the Glasgow program. Methods Inf Med. 1990;29(4):354–61.PubMed Macfarlane PW, Devine B, Latif S, McLaughlin S, Shoat DB, Watts MP. Methodology of ECG interpretation in the Glasgow program. Methods Inf Med. 1990;29(4):354–61.PubMed
23.
Zurück zum Zitat Macfarlane PW. Evolution of the Glasgow program for computer-assisted reporting of electrocardiograms--1964/1998. Acta Cardiol. 1998;53(2):117–20.PubMed Macfarlane PW. Evolution of the Glasgow program for computer-assisted reporting of electrocardiograms--1964/1998. Acta Cardiol. 1998;53(2):117–20.PubMed
24.
Zurück zum Zitat Rautaharju PM, Warren JW, Calhoun HP. Estimation of QT prolongation. A persistent, avoidable error in computer electrocardiography. J Electrocardiol. 1990;23 Suppl:111–117. Rautaharju PM, Warren JW, Calhoun HP. Estimation of QT prolongation. A persistent, avoidable error in computer electrocardiography. J Electrocardiol. 1990;23 Suppl:111–117.
25.
Zurück zum Zitat Rautaharju PM, Zhou SH, Wong S, Calhoun HP, Berenson GS, Prineas R, Davignon A. Sex differences in the evolution of the electrocardiographic QT interval with age. Can J Cardiol. 1992;8(7):690–5.PubMed Rautaharju PM, Zhou SH, Wong S, Calhoun HP, Berenson GS, Prineas R, Davignon A. Sex differences in the evolution of the electrocardiographic QT interval with age. Can J Cardiol. 1992;8(7):690–5.PubMed
26.
Zurück zum Zitat Rijnbeek PR, Witsenburg M, Schrama E, Hess J, Kors JA. New normal limits for the paediatric electrocardiogram. Eur Heart J. 2001;22(8):702–11.CrossRefPubMed Rijnbeek PR, Witsenburg M, Schrama E, Hess J, Kors JA. New normal limits for the paediatric electrocardiogram. Eur Heart J. 2001;22(8):702–11.CrossRefPubMed
27.
Zurück zum Zitat R Development Core Team (2011), R: A Language and Environment for Statistical Computing. Vienna, Austria: the R Foundation for Statistical Computing. ISBN: 3–900051–07-0. Available online at http://www.R-project.org/. R Development Core Team (2011), R: A Language and Environment for Statistical Computing. Vienna, Austria: the R Foundation for Statistical Computing. ISBN: 3–900051–07-0. Available online at http://​www.​R-project.​org/​.
28.
Zurück zum Zitat Sagie A, Larson MG, Goldberg RJ, Bengtson JR, Levy D. An improved method for adjusting the QT interval for heart rate (the Framingham heart study). Am J Cardiol. 1992;70(7):797–801.CrossRefPubMed Sagie A, Larson MG, Goldberg RJ, Bengtson JR, Levy D. An improved method for adjusting the QT interval for heart rate (the Framingham heart study). Am J Cardiol. 1992;70(7):797–801.CrossRefPubMed
29.
Zurück zum Zitat Fridericia LS. Die Systolendauer im Elektrokardiogramm bei normalen Menschen und bei Herzkranken. Acta Med Scan. 1920;53:469–86.CrossRef Fridericia LS. Die Systolendauer im Elektrokardiogramm bei normalen Menschen und bei Herzkranken. Acta Med Scan. 1920;53:469–86.CrossRef
30.
Zurück zum Zitat Bazett HC. An analysis of the time relations of electrocardiograms. Heart. 1920;7:353–70. Bazett HC. An analysis of the time relations of electrocardiograms. Heart. 1920;7:353–70.
31.
Zurück zum Zitat Hodges M, Salerno D, Erlien D. Bazett’s QT correction reviewed. Evidence that a linear QT correction for heart is better. J Am Coll Cardiol. 1983;1:694. Hodges M, Salerno D, Erlien D. Bazett’s QT correction reviewed. Evidence that a linear QT correction for heart is better. J Am Coll Cardiol. 1983;1:694.
32.
Zurück zum Zitat RP DA, Boisselle E, Soumis F, Megelas M, Choquette A. Normal ECG standards for infants and children. Pediatr Cardiol. 1979;1:123–52. RP DA, Boisselle E, Soumis F, Megelas M, Choquette A. Normal ECG standards for infants and children. Pediatr Cardiol. 1979;1:123–52.
33.
Zurück zum Zitat Luo S, Michler K, Johnston P, Macfarlane PW. A comparison of commonly used QT correction formulae: the effect of heart rate on the QTc of normal ECGs. J Electrocardiol. 2004;37 Suppl: 81-90. Luo S, Michler K, Johnston P, Macfarlane PW. A comparison of commonly used QT correction formulae: the effect of heart rate on the QTc of normal ECGs. J Electrocardiol. 2004;37 Suppl: 81-90.
Metadaten
Titel
Normal limits of the electrocardiogram derived from a large database of Brazilian primary care patients
verfasst von
Daniel M. F. Palhares
Milena S. Marcolino
Thales M. M. Santos
José L. P. da Silva
Paulo R. Gomes
Leonardo B. Ribeiro
Peter W. Macfarlane
Antonio L. P. Ribeiro
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
BMC Cardiovascular Disorders / Ausgabe 1/2017
Elektronische ISSN: 1471-2261
DOI
https://doi.org/10.1186/s12872-017-0572-8

Weitere Artikel der Ausgabe 1/2017

BMC Cardiovascular Disorders 1/2017 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.