Background
Components of the human body such as collagen, nicotinamide-adenine dinucleotide phosphate (NADP), and flavin-adenine dinucleotide (FAD), emit fluorescence when irradiated with light of an appropriate excitation wavelength [
1,
2]. Normal human bronchial epithelial tissue emits green autofluorescence at a wavelength of ca. 540 nm due to NADP and FAD when excited with 405-nm blue light. This green autofluorescence is less intense in cancer lesions due to thickening of the epithelium, reductions in the levels of the source materials, and absorption of the fluorescence within the lesion. Therefore, cancerous lesions of the bronchus will be demonstrated by a reduction in the intensity of green autofluorescence when the lesions are observed using autofluorescence endoscopy.
Several endoscopy systems have been developed for use in early detection of cancer lesions in the human bronchus. These systems include the LIFE lung [
3,
4] (Xillix, Richmond, Canada), SAFE-3000 [
5,
6] (Asahi Optical, Tokyo, Japan), D-Light AF [
7] (Storz, Tuttlingen, Germany), and AFI (Olympus, Tokyo, Japan). Superior rates of early bronchial carcinoma detection using autofluorescence bronchoscopy (AFB) have been reported in meta-analyses that included data from our study [
8,
9]. Although, LIFE lung and SAFE 3000 can detect both red and green fluorescence, only a decrease in the intensity of green autofluorescence in the cancer lesion is detectable using the above-mentioned systems, because their sensitivity is too low to permit visualization of color autofluorescence from human bronchial tissue and because a black and white charged coupled device (CCD) is used in the AFI system [
10].
We developed a color fluorescence endoscopy system (PDS-2000 [
11,
12]; Hamamatsu Photonics, Hamamatsu, Japan) to observe autofluorescence emanating from human tissues. This system detects both green autofluorescence from normal human organs as well as red autofluorescence from the accumulation of administered porphyrin derivatives. We compared the sensitivity of detection for bronchial cancers and precancerous lesions using this system and found that rate of lesion detection increased significantly, from 54.1 to 89.2%, when AFB was combined with white-light bronchoscopy [
13]. During the above clinical study, we detected red autofluorescence emanating from cancer lesions, contact bleeding sites, and blood vessels, and we reported that the red to green autofluorescence ratio (R/G ratio) was significantly higher in the cancer lesions [
13].
The accumulation of de novo porphyrin derivatives in cancer tissue, including the accumulation of protoporphyrin IX, has been reported [
14,
15]. However, previous reports were based on the results of spectral analyses of resected tumor and drawn blood samples [
16‐
18]. We observed red autofluorescence in human cancer lesions, contact bleeding sites, and the blood vessels of the bronchial wall using a color AFB system. The wavelength of the observed red autofluorescence differed from that reported in previous studies. In the present study, we measured the wavelength of red autofluorescence in order to determine the fluorescent component. This is the first report describing the origin of red autofluorescence observed in human cancer tissues, blood vessels, and contact bleeding sites in living patients using autofluorescence endoscopy.
Discussion
Detection of red autofluorescence in cancers of the bladder, stomach, and lung has been reported. High-performance liquid chromatography (HPLC) analysis of tissues from patients with these cancers revealed substances emitting faint red fluorescence [
17,
18]. The source of this red fluorescence has been attributed to the de novo accumulation of porphyrins [
16,
24]. However, this hypothesis has not been confirmed. We observed bronchogenic cancer lesions using a color fluorescence endoscopy system and found an increase in the R/G ratio in the cancer lesions [
13]. Kluftunger et al. [
25] reported increase of R/G ratio greater than 1.5 times the control, fluorescence imaging correctly identified areas of hyperplasia, dysplasia, CIS and invasive cancer using DMBA-induced hamster cheek pouch model. In our previous study, R/G ratio in bronchogenic cancer was significantly greater than those in normal bronchial wall due to decrease of green fluorescence and increase of red fluorescence in the cancer lesions [
13]. This red fluorescence was also observed in blood vessels as well as in fresh contact bleeding sites in the bronchial wall. We found that the wavelength of the red fluorescence was 617.7 nm, and the source of the red fluorescence in the present study was identified as Zn-photoprotoporphyrin. Zn-photoprotoporphyrin seems to be formed from Zn-protoporphyrin following irradiation with 405-nm blue light. However, it was difficult to extract porphyrin analogues from small biopsy specimen from the bronchial wall.
De novo protoporphyrin IX has been implicated as a source of the red autofluorescence associated with cancerous tissues. Moesta et al. reported the emission of red fluorescence from colorectal cancers [
18]. They analyzed chemical extracts of involved lymph nodes using reversed-phase HPLC and found a substance emitting 630-nm fluorescence. They concluded that protoporphyrin IX was the source of the red autofluorescence in these involved lymph nodes. Croce et al. reported naturally occurring porphyrins in a spontaneous tumor-bearing mouse model [
17]. They reported substantial levels of protoporphyrin IX in tumor, spleen, liver, and plasma samples.
Protoporphyrin IX is formed from 5-aminolevulinic acid; however, its concentration in normal human tissues is low [
26]. In addition, the wavelength of protoporphyrin IX fluorescence is 635 nm when excited with 405-nm light [
27]. These data suggest that the 617.7-nm autofluorescence emanating from cancer lesions, blood, and blood vessels in the present study was from a source other than protoporphyrin IX. The human body must therefore naturally contain a substance that emits strong, red autofluorescence. The present study was conducted to identify the source of the 617.7-nm red autofluorescence observed in previous studies.
The major porphyrin derivatives found in normal human blood are uroporphyrin, coproporphyrin, and Zn-protoporphyrin. Normal blood levels of porphyrins are 0–1.0 μg/dl for total porphyrin, <2 μg/dl for coproporphyrin, 16–60 μg/dl for protoporphyrin, <2 μg/dl for uroporphyrin [
28] and 23 μg/dl for Zn-protoporphyrin [
29]. Zn-protoporphyrin reportedly emits fluorescence at 585 nm, but our synthesized Zn-protoporphyrin examined emitted 578-nm red fluorescence. This wavelength differed from the 617.7-nm fluorescence observed in bronchial cancer lesions, blood vessels, and contact bleeding sites. We then examined our synthesized Zn-photoprotoporphyrin and photoprotoporphyrin by dissolving them in 5% albumin solution to mimic the conditions of the human body, and fluorescence from both of these porphyrin derivatives was detected. Zn-photoprotoporphyrin and photoprotoporphyrin emitted fluorescence at 625.5 and 664.0 nm, respectively, and we therefore concluded that the red fluorescence emanating from bronchial cancer lesions, blood vessels, and contact bleeding sites in the present study was associated with Zn-photoprotoporphyrin. This conclusion is plausible, as the difference in wavelengths was acceptable, considering the measurement method and the in vivo and in vitro conditions. In the human body, Zn-protoporphyrin (emitting 578-nm red fluorescence) seems to become Zn-photoprotoporphyrin (emitting 625.5-nm red fluorescence) following irradiation with 405-nm excitation light via photooxidation [
30]. It is known that cancer lesions emit bi-phasic red fluorescence during photodynamic therapy (PDT) forming protoporphyrin photoproducts [
31,
32]. This bi-phasic red fluorescence is emitted by protoporphyrin IX, which emits 636 nm red fluorescence, and photoprotoporphyrin, which emits 674 nm red fluorescence in case of PDT using 5-ALA [
32]. In PDT, protoporphyrin becomes photoprotoporphyrin upon laser irradiation. Our present report is the first to describe the origin of red autofluorescence emanating from cancer lesions, blood vessels and fresh contact bleeding sites in living patients.
Autofluorescence endoscopy revealed a decrease in the intensity of the green fluorescence emanating from normal human tissue. Autofluorescence endoscopy typically utilizes AFI and D-light AF systems. However, it is difficult to detect the red autofluorescence that emanates from cancer lesions, blood, and blood vessels using either system. This has led some researchers to conclude that human blood and blood vessels do not emit autofluorescence or emit only weak autofluorescence associated with hemoglobin. However, we found that red autofluorescence could be clearly detected using a sensitive autofluorescence endoscopy system such as the PDS-2000. We have observed red autofluorescence not only in bronchogenic carcinoma but also in tumors metastasized from breast, colon, and pancreatic cancers. We developed a new autofluorescence endoscopy system using an EM-CCD, PDS-TriMode (FLOVEL, Tachikawa, Japan), based on the PDS-2000 technology. The PDS-TriMode is a high-vision system, and its sensitivity is greater than that of the PDS-2000. The PDS-TriMode is capable of clearly detecting not only decreases in green autofluorescence but also abnormal red autofluorescence emanating from cancer lesions, blood, and blood vessels.
Analysis of the wavelength of red fluorescence can provide very important information. When 5-aminolevulinic acid (5-ALA) is orally administered, levels of protoporphyrin IX (which emits 635-nm red fluorescence when excited with 405-nm light) increase in cancer tissues [
33,
34]. Photofrin® and Lazerphyrin® have been approved and are currently used in PDT in Japan. In cancer tissues, Photofrin® and Lazerphyrin® emit 640- and 664-nm red fluorescence, respectively. These drugs are also used to detect cancerous tissue in photodynamic diagnosis (PDD). It is obviously difficult to differentiate 617.7-nm red autofluorescence emanating from the blood from 635-, 640-, and 664-nm red fluorescence using a sensitive color CCD camera. Attempts to do so could lead to false results in PDD. Our results indicate that reduction in the intensity of 617.7-nm red autofluorescence emanating from the blood is necessary for reliable PDD using porphyrin derivatives and 5-ALA.