Opening-wedge osteotomies of the distal femur: minor advantages for a biplanar compared to a uniplanar technique

Zeitschrift:
Knee Surgery, Sports Traumatology, Arthroscopy > Ausgabe 7/2019
Autoren:
Martin Pietsch, Markus Hochegger, Martin Winkler, Sabrina Sandriesser, Thomas Freude, Peter Augat
Wichtige Hinweise
The present study received the Michael Freeman Award 2018, which was presented at the 19th EFORT Congress in Barcelona, Spain.

Abstract

Purpose

Valgus malalignment of the distal femur may be treated with corrective osteotomy. The purpose of this study was to compare the primary stability of a lateral opening-wedge osteotomy (LOWO) using a uniplanar compared to a biplanar technique. A study was carried out to test both surgeries, with both an intact medial cortex and with a deliberate attached cut of the medial cortex simulating a fracture. The primary hypothesis was that the biplanar technique provides higher axial and torsional stiffness. It was further hypothesized that the mechanical superiority of the biplanar technique would not be affected in the case of breakage of the far medial cortex.

Methods

A LOWO was performed in ten synthetic femora (#3406 left large Femur, 4th Generation, Sawbones, Malmö, Sweden) using a lateral angle stable locking plate (NCB© Distal Femur Plate, Zimmer Biomet, Warsaw, USA). A uniplanar osteotomy was performed in five femora, and a biplanar osteotomy was performed in five femora. The femora were tested for axial and torsional loads using a servo-hydraulic testing machine (Instron 8874, Instron Structural Testing GmbH, High Wycombe, UK).

Results

Axial stiffness decreased significantly (p = 0.001) in both groups (20% in the uniplanar group and 28 % in the biplanar group) by cutting the medial cortex. The type of osteotomy had no significant effect. A slightly lower but not statistically significant axial stiffness was seen in the biplanar group both for intact and broken medial cortices. Internal torsional stiffness dropped by more than 30% for the uniplanar group and almost 24% for the biplanar group when the cortex was cut (p < 0.001). No significant change concerning internal torsional stiffness was found between the two groups. External torsional stiffness decreased by 32% for the uniplanar group and 4% for the biplanar group after the cortical cut (p = 0.029). No significant change concerning external torsional stiffness was found between the groups, but the biplanar group showed a tendency towards higher values of external torsional stiffness.

Conclusions

The axial and torsional stiffness of the implant-bone construct were not significantly affected by the type of osteotomy performed. Biplanar osteotomy tended to increase external torsional stiffness. In cases of fracture of the medial cortex, biplanar osteotomy significantly reduced the external rotation at the osteotomy and showed a significantly increased external torsional stiffness.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 7/2019

Knee Surgery, Sports Traumatology, Arthroscopy 7/2019 Zur Ausgabe
  1. Sie können e.Med Orthopädie & Unfallchirurgie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Orthopädie und Unfallchirurgie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Orthopädie und Unfallchirurgie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise