Optical and material analysis of opacified hydrophilic intraocular lenses after explantation: a laboratory study

verfasst von: Tamer Tandogan, Ramin Khoramnia, Chul Young Choi, Alexander Scheuerle, Martin Wenzel, Philipp Hugger, Gerd U. Auffarth

Erschienen in: BMC Ophthalmology | Ausgabe 1/2015

Abstract

Background

The opacification of hydrophilic intraocular lenses (IOLs) is a very rare complication in terms of absolute numbers. We report on the analyses of opacified Euromaxx ALI313Y and ALI313 IOLs (Argonoptics, Germany) using light and scanning electron microscopy, X-ray spectroscopy and optical bench analysis.

Methods

Opacified Euromaxx ALI313Y and ALI313 IOLs were explanted after patients presented with a decrease in visual acuity. The explants were sent to our laboratory and examined using light and scanning electron microscopy. The composition of the deposits was analysed using X-ray spectroscopy. The optical quality of the intraocular lens (IOL) was assessed using the OptiSpheric IOL PRO optical bench (Trioptics GmbH Wedel, Germany). Modulation transfer function (MTF) was measured at all spatial frequencies and United States Air Force (USAF) 1951 resolution target pictures were documented.

Results

Macroscopically, the entire optic was opacified in all IOLs. Light and scanning electron microscopy revealed numerous fine, granular, crystalline-like deposits, which were always distributed in a line parallel to the anterior and posterior surfaces of the IOLs. X-ray spectroscopy could prove the deposits consisted of Calcium and Phosphate. Measurements in the optical bench showed deterioration of MTF values at all spatial frequencies and the USAF target pictures demonstrated a significant reduction of brightness as well as resolution with the opacified IOLs.

Conclusions

The calcification of hydrophilic IOLs only occurs rarely. The exact chemical composition of the deposits can be assessed by means of X-ray spectroscopy. Optical quality analysis of the explanted Euromaxx ALI313Y and ALI313 IOLs showed significant reduction of MTF values, which was confirmed by USAF target pictures.
Hinweise

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

TT, RK and GUA were responsible for the conception and design of this study. TT, RK, CYC, AS, MW, PH and GUA acquired the data. TT, RK, CYC and GUA analysed and interpreted the data. TT, RK and CYC drafted the manuscript. GUA, AS, MW and PH revised the manuscript critically for important intellectual content. All authors have read and approved the final manuscript and agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.
Abkürzungen
IOL
intraocular lens
IOLs
intraocular lenses
MTF
Modulation transfer function
USAF target
United States Air Force target
SEM
scanning electron microscopy

Background

The calcification of hydrophilic IOLs is a very rare complication [1]. The intracameral injection of rtPA, gas or air seems to increase the risk of opacification [2, 3]. In most cases, however, the exact cause cannot be revealed [4]. As some systemic diseases might play a role (e.g., diabetes mellitus), several processing steps during manufacturing as well as packaging material may also facilitate opacification of the optic material [4]. Impurities by silicon particles are also considered to be another potential causative factor [5].
Opacities of the optic region of the intraocular lens (IOL) can cause a significant reduction of visual acuity, contrast sensitivity or glare that can make an IOL exchange necessary. The surgical removal of an opacified IOL is often related with an increased complication rate [6]. Prior to explantation the IOL is usually cut into two or more pieces. As a consequence, the optical properties of these IOLs cannot be assessed. A material analysis is however possible in sectioned IOLs using staining methods and light and scanning electron microscopy as well as X-ray spectroscopy.
We report on the examination of six opacified and explanted Euromaxx IOLs from Argonoptics, Germany (five Euromaxx ALI313Y and one Euromaxx ALI313); using light microscopy, scanning electron microscopy and X-ray spectroscopy. The IOLs were all explanted due to a reduction of visual acuity which was associated with discoloration of the optic component of the lens (Fig. 1).
Optical properties of IOLs can be analysed using Modulation Transfer Function (MTF) measurements [7, 8]. IOL metrology allows qualitative and quantitative analyses of the imaging ability of a lens. We measured MTF values in IOLs which were explanted with intact optic, using IOL metrology. Furthermore, we documented United States Air Force (USAF) 1951 resolution target images of these opacified but intact lenses (Fig. 2).

Methods

Five calcified Euromaxx ALI313Y and one ALI313 IOLs were explanted after patients complained of a decrease in visual acuity. These IOLs were sent to our laboratory and examined using light microscopy and scanning electron microscopy. Chemical composition of the deposits was analyzed by X-ray spectroscopy (EDX). The optical quality was assessed using the Trioptics OptiSpheric IOL PRO optical bench (Trioptics GmbH, Wedel, Germany).
The IOLs were examined and photographed under an Olympus BX50 light microscope with an attached Olympus C-7070 camera (Olympus Optical Co. Ltd., Tokyo, Japan).
Prior to microscopic analysis the optical quality was evaluated exemplary for one explant, which had an undamaged optic zone, using the optical bench. The Optispheric IOL-PRO optical bench (Trioptics, Wedel, Germany) was used to evaluate the optical quality parameters of IOL including Modulation Transfer Function (MTF). The MTF describing the resolution and performance of an optical system is the ratio of relative image contrast divided by relative object contrast. This technology is suggested in the ISO 11-979-2 describing the test of IOLs as the most suitable method [8]. The instrument features different types of targets as objects which are projected to the infinity through a collimator. The lens under test thus gives an image of the target at its focal plane. The measurement head composed of a microscope objective and an imaging system conjugated with a CCD camera scans through the imaging zone to find the best focus image created by the IOL tested. This image detected by the camera is then used for further analysis. The light source illuminating the target is a broad band visible spectrum light source associated with a narrow band interferential filter at 546 nm as required in the ISO norm. The image obtained via the IOL is collected by a microscope and is analyzed by the software. After image processing, an MTF curve is obtained. Using a United States Air Force (USAF) 1951 Resolution Target it is also possible to document and compare the quality of the image at a focus plane.
After optical measurement, each IOL specimen was cut in halves. One half of the IOL was stained (Alizarin Red and von Kossa stains) to determine the chemical presence nature of the deposits within the IOL. For this purpose the IOL material was treated for 2 min with 4 % buffered Formaldehyde solution and rinsed with distilled water. Subsequently the IOL-halves were placed for 3 min in an Alizarin-Red solution of 1 %. Afterwards they were rinsed again and examined by the light microscope. The specimens were then dehydrated and fixed in Paraffin, cut in 5 μm thick sagittal slices along the centre of the IOL optic and dyed according to the von Kossa method to show calcium particles. The slices of the optic material were freed of Paraffin, rehydrated with water, incubated in 5 % Silver Nitrate solution, treated with UV-light for 30 min, rinsed repeatedly with distilled water, reacted with 5 % Sodium thiosulphate and washed again before being analysed under the light microscope.
The other halves of the IOLs were used for scanning electron microscopy (SEM). For the SEM examinations the IOLs were prepared by ultramicrotomy (UCT, Leica, Germany) obtaining a very smooth cross section through the lens using a 35° diamond knife (Diatome, Switzerland). Subsequently, we examined this cross section by SEM under low voltage (<1 kV) conditions using a SU8000 microscope (Hitachi, Japan) without further coating of the sample. For a local chemical analysis, Energy Dispersive X-ray Spectroscopy (EDS) was performed using a Quantax 400 EDS detector (Bruker, Germany) attached to the SEM to detect any possible exogenous chemical elements within the IOL besides the material from which it is made.
Our laboratory study only involves analyses of IOL explants. We did not perform any additional examinations or procedures on humans. Therefore, informed consent and ethics committee approval were not required.

Results and discussion

Gross examination of all IOLs showed an opacification of the entire optic area (Fig. 2). The opacifications stained with Alizarin-Red were visible during light microscopy (Fig. 3). Under Light microscopy, the cross section of all IOLs (with von Kossa staining) revealed numerous fine, granular, crystalline-like deposits distributed in a line parallel to the anterior and posterior surface of all IOLs (Fig. 4). These findings were confirmed by scanning electron microscopy in all IOLs (Fig. 5). X-ray spectroscopy could prove the deposits to always consist of Calcium phosphate (Fig. 6). Element mapping supported these results (Fig. 7).
The MTF measurements of the IOL with an intact optic showed a significant decrease in optical quality (Fig. 8). MTF values had deteriorated at all spatial frequencies. The USAF target analysis showed a significant reduction in brightness. After adjusting the light, however, a quite acceptable resolution could be achieved (Fig. 9). This suggests that patients may experience the impact of this type of opacification in the IOL optic as a reduction of resolution and light.
Calcification in hydrophilic IOLs has been previously reported but this is, to the best of our knowledge, the first publication about calcification of this IOL model and about the assessment of optical quality in opacified IOLs using IOL metrology [4, 5]. Different types of hydrophilic acrylic IOL materials are not equally susceptible to opacification. Therefore, material related factors seem to have a significant impact on the development of calcifications. However, the possibility of patient-related factors, such as metabolic imbalance, cannot be fully ruled out.
Izak et al., described this type of deposit in Hydroview (Bausch and Lomb, Rochester, USA) and SC60B-OUV IOLs (Medical Developmental Research Inc., Clearwater, USA) [4]. The Hydroview IOL showed granular deposits on the surface of the optic [4]. The opacification in the SC60B-OUV IOL, however, were similar to a nuclear cataract. The surface of the optic as well as the haptic region was free of Calcium phosphate deposits. Numerous granular, fine, crystalline-like deposits could be detected that were distributed in a line parallel just below the IOL surface [4, 5].
In opacified Aqua-Sense IOLs (Ophthalmic Innovations International Inc., Ontario, California) light microscopy and scanning electron microscopy analyses showed typical granular deposits on the surface the IOL [1, 46]. Additionally granular deposits of different sizes lined parallel to the surface of the IOL optic were found using Alizarin Red as well as the von Kossa method for staining [1, 4]. The deposits were generally located in the optic as well as the haptic region [46].
Aside from calcifications, other alterations of the IOL material are also described in the literature, for example IOL schisis, which is a gap formation within the IOL material [9, 10]. We did not observe this in the Euromaxx explants.
Glistenings should also be considered as a differential diagnosis in the early stage of a calcification. There are many reports in the literature about glistenings in hydrophobic acrylic IOLs, especially for the AcrySof IOL (Alcon Laboratories, Fort Worth, Texas, USA) [4, 11, 12]. However, glistenings can also be detected in other IOL materials [11]. Glistenings can be observed as early as one week after the implantation of the IOL [13]. Morphologically glistenings correlate to microvacuoles within the IOL material.
They seem to reduce the contrast sensitivity [11, 13]. However, in contrast to calcifications, a reduction of visual acuity is rare [11, 13, 14].
The explantation of the IOL is the only therapeutic option in symptomatic patients with calcified IOLs. The incidence of IOL explantation because of calcification is reported to be low. Izak et al., for example, reported in three different IOL-models with opacifcations (Hydroview, SC60B-OUV and AquaSense) the need for explantation in less than 1 % in each of the three groups [4]. The explantation of IOLs involves the risk of intra- and postoperative complications. In a series of 25 eyes with opacified Aqua-Sense IOLs, Dagres et al. reported complications (e.g., zonular dehiscence, posterior capsular rupture, corneal decompensation) in 48 % of cases [6]. Especially a strong attachment of the IOL haptics to the lens capsule makes the surgical procedure more difficult. Therefore, an indication for a surgical removal of the IOL should require thorough informed patient consent and be considered only in the case of severe complaints by the patient. It is also important not to perform a Nd:YAG-Laser capsulotomy in eyes with an opacified IOL as that can increase the complication rate during an IOL exchange procedure.
In this study, unfortunately we could not evaluate all IOLs using optical quality assessment, because some IOLs were explanted after bisection during surgery. Additional studies are needed to understand the relationship between the clinical characteristics of opacification and the results of analyses on optical bench systems as well as to determine possible patient-related factors (e.g., metabolic disorders) as causes for the opacification of this IOL model.

Conclusions

The calcification of hydrophilic IOLs is a rare complication. Granular deposits under the surface of the Euromaxx ALI313Y and ALI313 IOL lenses can cause a degradation of lens optical quality and reduction of visual acuity.
The calcium deposits can be demonstrated using the von Kossa staining-method or staining with Alizarin-Red. The exact chemical composition of the deposits can be assessed using X-ray spectroscopy.
The Argonoptics company has reported that it uses hydrophilic acrylic blanks from a different polymer manufacturer since January 2012. According to Argonoptics, there has been no case of opacification with the new material so far (Personal communication).

Availability of supporting data

The data on which the conclusion of the paper relies are included in the manuscript and presented in form of figures.

Acknowledgements

The authors thank Dr. I. Lieberwirth at the Max Planck Institute for Polymer Research Mainz Germany for his help with scanning electron microscopy and X-ray spectroscopy.

Grant support

Klaus Tschira Stiftung, Heidelberg, Germany
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

TT, RK and GUA were responsible for the conception and design of this study. TT, RK, CYC, AS, MW, PH and GUA acquired the data. TT, RK, CYC and GUA analysed and interpreted the data. TT, RK and CYC drafted the manuscript. GUA, AS, MW and PH revised the manuscript critically for important intellectual content. All authors have read and approved the final manuscript and agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.
Literatur
1.
Zurück zum Zitat Khoramnia R, Salgado JP, Auffarth GU, Schmidt S, Wegner A, Kobuch KA, et al. [Opacification of a hydrophilic intraocular lens 4 years after cataract surgery. A biomaterial analysis]. Ophthalmologe. 2012;109(5):483–6.CrossRefPubMed Khoramnia R, Salgado JP, Auffarth GU, Schmidt S, Wegner A, Kobuch KA, et al. [Opacification of a hydrophilic intraocular lens 4 years after cataract surgery. A biomaterial analysis]. Ophthalmologe. 2012;109(5):483–6.CrossRefPubMed
2.
Zurück zum Zitat Fung SS, Sykakis E, Islam NM, Zambarakji HJ, Khoramnia R, Auffarth GU, et al. Intraocular Lens Opacification following Intracameral Injection of Recombinant Tissue Plasminogen Activator to Treat Inflammatory Membranes after Cataract Surgery. J Ophthalmol. 2015;2015:975075.PubMedCentralCrossRefPubMed Fung SS, Sykakis E, Islam NM, Zambarakji HJ, Khoramnia R, Auffarth GU, et al. Intraocular Lens Opacification following Intracameral Injection of Recombinant Tissue Plasminogen Activator to Treat Inflammatory Membranes after Cataract Surgery. J Ophthalmol. 2015;2015:975075.PubMedCentralCrossRefPubMed
3.
Zurück zum Zitat Schmidinger G, Pemp B, Werner L. [Opacification of an intraocular lens: calcification of hydrophilic intraocular lenses after gas tamponade of the anterior chamber]. Ophthalmologe. 2013;110(11):1066–8.CrossRefPubMed Schmidinger G, Pemp B, Werner L. [Opacification of an intraocular lens: calcification of hydrophilic intraocular lenses after gas tamponade of the anterior chamber]. Ophthalmologe. 2013;110(11):1066–8.CrossRefPubMed
4.
Zurück zum Zitat Izak AM, Werner L, Pandey SK, Apple DJ. Calcification of modern foldable hydrogel intraocular lens designs. Eye. 2003;17(3):393–406.CrossRefPubMed Izak AM, Werner L, Pandey SK, Apple DJ. Calcification of modern foldable hydrogel intraocular lens designs. Eye. 2003;17(3):393–406.CrossRefPubMed
5.
Zurück zum Zitat Werner L, Hunter B, Stevens S, Chew JJ, Mamalis N. Role of silicon contamination on calcification of hydrophilic acrylic intraocular lenses. Am J Ophthalmol. 2006;141(1):35–43.CrossRefPubMed Werner L, Hunter B, Stevens S, Chew JJ, Mamalis N. Role of silicon contamination on calcification of hydrophilic acrylic intraocular lenses. Am J Ophthalmol. 2006;141(1):35–43.CrossRefPubMed
6.
Zurück zum Zitat Dagres E, Khan MA, Kyle GM, Clark D. Perioperative complications of intraocular lens exchange in patients with opacified Aqua-Sense lenses. J Cataract Refract Surg. 2004;30(12):2569–73.CrossRefPubMed Dagres E, Khan MA, Kyle GM, Clark D. Perioperative complications of intraocular lens exchange in patients with opacified Aqua-Sense lenses. J Cataract Refract Surg. 2004;30(12):2569–73.CrossRefPubMed
7.
Zurück zum Zitat Madrid-Costa D, Ruiz-Alcocer J, Ferrer-Blasco T, Garcia-Lazaro S, Montes-Mico R. Optical quality differences between three multifocal intraocular lenses: bifocal low add, bifocal moderate add, and trifocal. J Refract Surg. 2013;29(11):749–54.CrossRefPubMed Madrid-Costa D, Ruiz-Alcocer J, Ferrer-Blasco T, Garcia-Lazaro S, Montes-Mico R. Optical quality differences between three multifocal intraocular lenses: bifocal low add, bifocal moderate add, and trifocal. J Refract Surg. 2013;29(11):749–54.CrossRefPubMed
8.
Zurück zum Zitat Gatinel D, Houbrechts Y. Comparison of bifocal and trifocal diffractive and refractive intraocular lenses using an optical bench. J Cataract Refract Surg. 2013;39(7):1093–9.CrossRefPubMed Gatinel D, Houbrechts Y. Comparison of bifocal and trifocal diffractive and refractive intraocular lenses using an optical bench. J Cataract Refract Surg. 2013;39(7):1093–9.CrossRefPubMed
9.
Zurück zum Zitat Neuhann I, Lege B, Sommer B, Neuhann T. [Conspicuous intraocular lens]. Ophthalmologe. 2008;105(12):1154–6.CrossRefPubMed Neuhann I, Lege B, Sommer B, Neuhann T. [Conspicuous intraocular lens]. Ophthalmologe. 2008;105(12):1154–6.CrossRefPubMed
10.
Zurück zum Zitat Neuhann IM, Neuhann TF, Kleinmann G, Apple DJ. Clinical presentation of intraocular lens schisis. J Cataract Refract Surg. 2006;32(9):1503–8.CrossRefPubMed Neuhann IM, Neuhann TF, Kleinmann G, Apple DJ. Clinical presentation of intraocular lens schisis. J Cataract Refract Surg. 2006;32(9):1503–8.CrossRefPubMed
11.
Zurück zum Zitat Werner L. Glistenings and surface light scattering in intraocular lenses. J Cataract Refract Surg. 2010;36(8):1398–420.CrossRefPubMed Werner L. Glistenings and surface light scattering in intraocular lenses. J Cataract Refract Surg. 2010;36(8):1398–420.CrossRefPubMed
12.
Zurück zum Zitat Moreno-Montanes J, Alvarez A, Rodriguez-Conde R, Fernandez-Hortelano A. Clinical factors related to the frequency and intensity of glistenings in AcrySof intraocular lenses. J Cataract Refract Surg. 2003;29(10):1980–4.CrossRefPubMed Moreno-Montanes J, Alvarez A, Rodriguez-Conde R, Fernandez-Hortelano A. Clinical factors related to the frequency and intensity of glistenings in AcrySof intraocular lenses. J Cataract Refract Surg. 2003;29(10):1980–4.CrossRefPubMed
13.
Zurück zum Zitat Dhaliwal DK, Mamalis N, Olson RJ, Crandall AS, Zimmerman P, Alldredge OC, et al. Visual significance of glistenings seen in the AcrySof intraocular lens. J Cataract Refract Surg. 1996;22(4):452–7.CrossRefPubMed Dhaliwal DK, Mamalis N, Olson RJ, Crandall AS, Zimmerman P, Alldredge OC, et al. Visual significance of glistenings seen in the AcrySof intraocular lens. J Cataract Refract Surg. 1996;22(4):452–7.CrossRefPubMed
14.
Zurück zum Zitat Allers A, Baumeister M, Steinkamp GW, Ohrloff C, Kohnen T. Intra-individual comparison of intraocular lenses of highly refractive silicone (Allergan SI40NB) and hydrophobic acrylate (Alcon Acrysof MA60BM). 1-year follow-up. Ophthalmologe. 2000;97(10):669–75.CrossRefPubMed Allers A, Baumeister M, Steinkamp GW, Ohrloff C, Kohnen T. Intra-individual comparison of intraocular lenses of highly refractive silicone (Allergan SI40NB) and hydrophobic acrylate (Alcon Acrysof MA60BM). 1-year follow-up. Ophthalmologe. 2000;97(10):669–75.CrossRefPubMed
Metadaten
Titel
Optical and material analysis of opacified hydrophilic intraocular lenses after explantation: a laboratory study
verfasst von
Tamer Tandogan
Ramin Khoramnia
Chul Young Choi
Alexander Scheuerle
Martin Wenzel
Philipp Hugger
Gerd U. Auffarth
Publikationsdatum
01.12.2015
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2015
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-015-0149-1

Weitere Artikel der Ausgabe 1/2015

BMC Ophthalmology 1/2015 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.