Skip to main content
main-content

01.04.2015 | Original Article | Ausgabe 3/2015

Annals of Nuclear Medicine 3/2015

Optimization of attenuation and scatter corrections in sentinel lymph node scintigraphy using SPECT/CT systems

Zeitschrift:
Annals of Nuclear Medicine > Ausgabe 3/2015
Autoren:
Hiroto Yoneyama, Hiroyuki Tsushima, Masahisa Onoguchi, Takahiro Konishi, Kenichi Nakajima, Shinro Matsuo, Daiki Kayano, Hiroshi Wakabayashi, Anri Inaki, Seigo Kinuya

Abstract

Objective

Although SPECT/CT systems have been used for sentinel lymph node (SLN) imaging, few studies have focused on optimization of attenuation correction (AC) and scatter correction (SC). While SLNs could be detected in conventional planar images, they sometimes do not appear in SPECT/CT images. The purpose of this study was to investigate the optimal AC and SC and to improve the detectability of SLNs in examinations using SPECT/CT systems.

Materials and methods

The study group consisted of 56 female patients with breast cancer. In SPECT/CT imaging, four kinds of images were created with and without AC and SC; namely, AC−SC−, AC+SC−, AC−SC+ and AC+SC+. Five nuclear medicine physicians interpreted the planar and SPECT/CT images with five grades of confidence levels (1–5). The detection rate was calculated as the number of patients whose average confidence levels of interpretation were more than 4, divided by the total number of patients.

Results

The confidence level of interpretation and the detection rate provided by the planar images were 4.76 ± 0.49 and 94.6 %, respectively. In SPECT/CT imaging, the AC+SC− provided the best detection rate (confidence level of interpretation, 4.81 ± 0.38; detection rate, 98.2 %), followed by the AC−SC− (4.70 ± 0.55, 89.3 %), and the AC−SC+ (4.39 ± 1.2, 78.6 %). The lowest values were obtained for the AC+SC+ (4.36 ± 1.22, 78.6 %). Regarding the confidence levels of interpretation, significant differences were observed between AC+SC− and AC−SC−, AC+SC− and AC+SC+, AC+SC− and AC−SC+, and between planar images and AC+SC+ (P = 0.0021, 0.0009, 0.0013, and 0.0056, respectively).

Conclusions

When SPECT/CT was used, AC improved the detection of SLNs. SC caused disappearance of a faint SLN in some cases and should not be performed.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 3/2015

Annals of Nuclear Medicine 3/2015 Zur Ausgabe
  1. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

  2. Das kostenlose Testabonnement läuft nach 14 Tagen automatisch und formlos aus. Dieses Abonnement kann nur einmal getestet werden.