Skip to main content
Erschienen in:

14.05.2024 | Original Article

Optimizing hook implantation angle of the clavicular hook plate: a cadaveric study

verfasst von: Pamok Saengsirinavin, Chumnarn Iamsirikulmit, Peerapong Piyapittayanun, Vajarin Phiphobmongkol, Thanawan Jongthanakamol, Rosarin Ratanalekha

Erschienen in: European Journal of Orthopaedic Surgery & Traumatology | Ausgabe 5/2024

Einloggen, um Zugang zu erhalten

Abstract

Purpose

While Clavicle hook plates have demonstrated favorable results regarding bone and shoulder function, their design can potentially lead to complications due to pressure concentration at the plate's tip. This study aims to investigate the impact of different hook implantation angles on the contact surface area between the hook plate and acromion, with the goal of minimizing mismatch and maximizing contact surface area.

Methods

Twenty soft shoulder cadavers were included in the study, and the contact surface area of the hook plate was measured in different positions based on the hook implantation angle.

Results

The results showed variations in compatibility, width, and length of the contact surface area depending on the hook implantation angle and the medial or lateral row placement. The lateral row generally demonstrated superior compatibility (84.0% vs 46.67%, p-value < 0.001), with a broader contact area (3.55 ± 0.08 mm vs 3.09 ± 0.10 mm, p-value = 0.004) and a longer contact area (7.36 ± 0.19 mm vs 5.10 ± 0.23 mm, p-value < 0.001) at specific angles. A detailed analysis of the lateral position revealed that the zero angle of implantation resulted in the greatest contact surface area, measuring 3.91 ± 0.70 mm in width (p value = 0.083) and 8.85 ± 1.24 mm in length (p value < 0.001).

Conclusion

Placing the hook laterally and at the zero position according to the hook implantation angle can maximize contact surface area, may reduce stress concentration, and minimize complications in hook plate fixation. Further research and consideration of anatomical variations are warranted to refine the placement technique and enhance patient outcomes.

Level of Evidence

Level V evidence.
Literatur
2.
Zurück zum Zitat Asadollahi S, Bucknill A (2019) Hook plate fixation for acute unstable distal clavicle fracture: a systematic review and meta-analysis. Orthop Trauma 33(8):417–422 Asadollahi S, Bucknill A (2019) Hook plate fixation for acute unstable distal clavicle fracture: a systematic review and meta-analysis. Orthop Trauma 33(8):417–422
3.
Zurück zum Zitat Tiren D, Bemmel AJ, Swank DJ, Linden FM (2012) Hook plate fixation of acute Displaced lateral clavicle fracture: mid-term results and a brief literature overview. J Orthop Surgery Res 7(2):1–8 Tiren D, Bemmel AJ, Swank DJ, Linden FM (2012) Hook plate fixation of acute Displaced lateral clavicle fracture: mid-term results and a brief literature overview. J Orthop Surgery Res 7(2):1–8
4.
Zurück zum Zitat Wu X et al (2020) Digital technology combined with 3D printing to evaluate the matching performance of AO clavicular hook plates. Indian J Orthop 54:141–147CrossRefPubMedPubMedCentral Wu X et al (2020) Digital technology combined with 3D printing to evaluate the matching performance of AO clavicular hook plates. Indian J Orthop 54:141–147CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Wu X et al (2020) 3D printed model used as preoperative tool for treating acromioclavicular joint dislocation with pre-contoured clavicle hook plate: technical note. OrthopUnfall 158:221–226 Wu X et al (2020) 3D printed model used as preoperative tool for treating acromioclavicular joint dislocation with pre-contoured clavicle hook plate: technical note. OrthopUnfall 158:221–226
6.
Zurück zum Zitat ElMaraghy AW, Devereaux MW, Ravichandiran K, Agur AM (2010) Subacromial morphometric assessment of the clavicle hook plate injury. Int J Care injured 41:613–619CrossRef ElMaraghy AW, Devereaux MW, Ravichandiran K, Agur AM (2010) Subacromial morphometric assessment of the clavicle hook plate injury. Int J Care injured 41:613–619CrossRef
7.
Zurück zum Zitat Yoon JP, Lee YS, Song GS, Oh JH (2017) Morphological analysis of acromion and hook plate for the fixation of acromioclavicular joint dislocation. Knee Surg Sports Traumatol Arthrosc 25:980–986CrossRefPubMed Yoon JP, Lee YS, Song GS, Oh JH (2017) Morphological analysis of acromion and hook plate for the fixation of acromioclavicular joint dislocation. Knee Surg Sports Traumatol Arthrosc 25:980–986CrossRefPubMed
8.
Zurück zum Zitat Liu J et al (2021) Biomechanical analysis of a novel clavicular hook plate for the treatment of acromioclavicular joint dislocation: a finite element analysis. Comput Biol Med 133:1–7CrossRef Liu J et al (2021) Biomechanical analysis of a novel clavicular hook plate for the treatment of acromioclavicular joint dislocation: a finite element analysis. Comput Biol Med 133:1–7CrossRef
9.
Zurück zum Zitat Shih CM, Huang KC, Pan CC, Lee CH, Su KC (2015) Biomechanical analysis of acromioclavicular joint dislocation treated with clavicle hook plates in different lengths. Int Orthop 39:2239–2244CrossRefPubMed Shih CM, Huang KC, Pan CC, Lee CH, Su KC (2015) Biomechanical analysis of acromioclavicular joint dislocation treated with clavicle hook plates in different lengths. Int Orthop 39:2239–2244CrossRefPubMed
10.
Zurück zum Zitat Hung LK, Su KC, Lu WH, Lee CH (2017) Biomechanical analysis of clavicle hook plate implantation with different hook angles in the acromioclavicular joint. Int Orthop 41:1663–1669CrossRefPubMed Hung LK, Su KC, Lu WH, Lee CH (2017) Biomechanical analysis of clavicle hook plate implantation with different hook angles in the acromioclavicular joint. Int Orthop 41:1663–1669CrossRefPubMed
Metadaten
Titel
Optimizing hook implantation angle of the clavicular hook plate: a cadaveric study
verfasst von
Pamok Saengsirinavin
Chumnarn Iamsirikulmit
Peerapong Piyapittayanun
Vajarin Phiphobmongkol
Thanawan Jongthanakamol
Rosarin Ratanalekha
Publikationsdatum
14.05.2024
Verlag
Springer Paris
Erschienen in
European Journal of Orthopaedic Surgery & Traumatology / Ausgabe 5/2024
Print ISSN: 1633-8065
Elektronische ISSN: 1432-1068
DOI
https://doi.org/10.1007/s00590-024-03981-z

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie erweitert durch Fallbeispiele, Videos und Abbildungen. Zur Fortbildung und Wissenserweiterung, verfasst und geprüft von Expertinnen und Experten der Gesellschaft für Arthroskopie und Gelenkchirurgie (AGA).


Jetzt entdecken!

Neu im Fachgebiet Orthopädie und Unfallchirurgie

Viele Versäumnisse bei Psoriasis-Arthritis

Menschen mit Psoriasis-Arthritis (PsA) müssen länger auf die Diagnose warten und werden zögerlicher behandelt als an rheumatoider Arthritis (RA) Erkrankte. Diese Defizite hat eine Untersuchung in Großbritannien aufgedeckt.

Yoga gegen Kniearthrose nicht schlechter als Krafttraining

Menschen mit Gonarthrose profitieren von Yogaübungen nicht weniger als von gezielten Kräftigungsübungen für die lädierten Knie. In einer Vergleichsstudie haben sich für Yogis und Yoginis sogar einige Vorteile ergeben.

Muskelrelaxanzien wohl nur bei akuten Kreuzschmerzen hilfreich

Bei akuten Rückenschmerzen können Muskelrelaxanzien, eventuell in Kombination mit NSAR, zur Schmerzlinderung beitragen. Wegen der Nebenwirkungen wird jedoch empfohlen, die Medikamente nur über wenige Tage einzusetzen.

Thoracic-Outlet-Syndrom nur in Ausnahmefällen operieren!

Das Thoracic-Outlet-Syndrom erfordert nur in ganz bestimmten Fällen ein operatives Vorgehen. Beim DCK wurde vor schwerwiegenden Komplikationen des anspruchsvollen Eingriffs gewarnt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.