Erschienen in: European Radiology 8/2016

12.11.2015 | Computed Tomography

Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue

verfasst von: André Euler, Zsolt Szucs-Farkas, Anna L. Falkowski, Nadine Kawel-Böhm, Luigia D’Errico, Sebastién Kopp, Jens Bremerich, Tilo Niemann

Erschienen in: European Radiology | Ausgabe 8/2016

Einloggen, um Zugang zu erhalten

Abstract

Objectives

Organ-based tube current modulation aims to reduce exposure to radiosensitive organs like the breasts by considering their anatomical location and altering tube current during rotation. Former phantom studies demonstrated a dose reduction of 20–37 %. Our study aimed to estimate the potential of dose reduction with this technique in relation to the actual location of breast tissue in a large clinical cohort.

Methods

A 1-year cohort of chest CTs of females (N=1,263) was retrospectively evaluated. To estimate the relative dose effect, breast location was analysed by measuring the angle range of glandular tissue within the different dose zones. Relative exposure compared with constant tube current was calculated. Descriptive statistics and Wilcoxon-test were applied.

Results

Only 63 % of angle range of glandular breast tissue was found inside the reduced dose zone. The estimated mean relative dose reduction was lower than observed in former phantom studies(16 % vs. 20–37 %) but still significant compared to constant tube current (p<0.0001).

Conclusions

Although organ-based tube current modulation results in a significant reduction of breast exposure compared to non-modulated irradiation, the technique cannot unfold its full potential, because breast tissue is often located outside the reduced dose zone, resulting in significantly lower dose reduction than expected.

Key Points

OBTCM results in significant dose reduction compared to constant tube current scans.
A substantial portion of glandular tissue lies outside the reduced dose zone.
Potential dose reduction using organ-based tube current modulation may be overestimated.
Literatur
1.
Zurück zum Zitat Niemann T, Zbinden I, Roser HW, Bremerich J, Remy-Jardin M, Bongartz G (2013) Computed tomography for pulmonary embolism: assessment of a 1-year cohort and estimated cancer risk associated with diagnostic irradiation. Acta Radiol 54(7):778–784. Niemann T, Zbinden I, Roser HW, Bremerich J, Remy-Jardin M, Bongartz G (2013) Computed tomography for pulmonary embolism: assessment of a 1-year cohort and estimated cancer risk associated with diagnostic irradiation. Acta Radiol 54(7):778–784.
2.
Zurück zum Zitat The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37(2-4):1–332 The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37(2-4):1–332
3.
Zurück zum Zitat Kim YK, Sung YM, Choi JH, Kim EY, Kim HS (2013) Reduced radiation exposure of the female breast during low-dose chest CT using organ-based tube current modulation and a bismuth shield: comparison of image quality and radiation dose. AJR Am J Roentgenol 200(3):537–544CrossRefPubMed Kim YK, Sung YM, Choi JH, Kim EY, Kim HS (2013) Reduced radiation exposure of the female breast during low-dose chest CT using organ-based tube current modulation and a bismuth shield: comparison of image quality and radiation dose. AJR Am J Roentgenol 200(3):537–544CrossRefPubMed
4.
Zurück zum Zitat Vollmar SV, Kalender WA (2008) Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations. Eur Radiol 18(8):1674–1682CrossRefPubMed Vollmar SV, Kalender WA (2008) Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations. Eur Radiol 18(8):1674–1682CrossRefPubMed
5.
Zurück zum Zitat McLaughlin DJ, Mooney RB (2004) Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the users' needs? Clin Radiol 59(5):446–450CrossRefPubMed McLaughlin DJ, Mooney RB (2004) Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the users' needs? Clin Radiol 59(5):446–450CrossRefPubMed
6.
Zurück zum Zitat Kubo T, Lin P-JP, Stiller W, Takahashi M, Kauczor H-U, Ohno Y et al (2008) Radiation dose reduction in chest CT: a review. Am J Roentgenol 190(2):335–343CrossRef Kubo T, Lin P-JP, Stiller W, Takahashi M, Kauczor H-U, Ohno Y et al (2008) Radiation dose reduction in chest CT: a review. Am J Roentgenol 190(2):335–343CrossRef
7.
Zurück zum Zitat Duan X, Wang J, Christner JA, Leng S, Grant KL, McCollough CH (2011) Dose reduction to anterior surfaces with organ-based tube-current modulation: evaluation of performance in a phantom study. AJR Am J Roentgenol 197(3):689–695CrossRefPubMed Duan X, Wang J, Christner JA, Leng S, Grant KL, McCollough CH (2011) Dose reduction to anterior surfaces with organ-based tube-current modulation: evaluation of performance in a phantom study. AJR Am J Roentgenol 197(3):689–695CrossRefPubMed
8.
Zurück zum Zitat Lungren MP, Yoshizumi TT, Brady SM, Toncheva G, Anderson-Evans C, Lowry C et al (2012) Radiation dose estimations to the thorax using organ-based dose modulation. AJR Am J Roentgenol 199(1):W65–W73CrossRefPubMed Lungren MP, Yoshizumi TT, Brady SM, Toncheva G, Anderson-Evans C, Lowry C et al (2012) Radiation dose estimations to the thorax using organ-based dose modulation. AJR Am J Roentgenol 199(1):W65–W73CrossRefPubMed
9.
Zurück zum Zitat Taylor S, Litmanovich DE, Shahrzad M, Bankier AA, Gevenois PA, Tack D (2015) Organ-based tube current modulation: are women's breasts positioned in the reduced-dose zone? Radiology 274(1):260–266CrossRefPubMed Taylor S, Litmanovich DE, Shahrzad M, Bankier AA, Gevenois PA, Tack D (2015) Organ-based tube current modulation: are women's breasts positioned in the reduced-dose zone? Radiology 274(1):260–266CrossRefPubMed
10.
Zurück zum Zitat Li J, Udayasankar UK, Toth TL, Seamans J, Small WC, Kalra MK (2007) Automatic patient centering for MDCT: effect on radiation dose. AJR Am J Roentgenol 188(2):547–552CrossRefPubMed Li J, Udayasankar UK, Toth TL, Seamans J, Small WC, Kalra MK (2007) Automatic patient centering for MDCT: effect on radiation dose. AJR Am J Roentgenol 188(2):547–552CrossRefPubMed
11.
Zurück zum Zitat Kaasalainen T, Palmu K, Lampinen A, Kortesniemi M (2013) Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT: phantom study. Pediatr Radiol 43(6):673–684CrossRefPubMed Kaasalainen T, Palmu K, Lampinen A, Kortesniemi M (2013) Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT: phantom study. Pediatr Radiol 43(6):673–684CrossRefPubMed
12.
Zurück zum Zitat Edyvean S (2012) Radiation Dose Metrics and the Effect of CT Scan Protocol Parameters. In: Tack D, Kalra MK, Gevenois PA (eds) Radiation Dose from Multidetector CT, vol 2nd. Springer Berlin Heidelberg, Berlin, Heidelberg, p 112 Edyvean S (2012) Radiation Dose Metrics and the Effect of CT Scan Protocol Parameters. In: Tack D, Kalra MK, Gevenois PA (eds) Radiation Dose from Multidetector CT, vol 2nd. Springer Berlin Heidelberg, Berlin, Heidelberg, p 112
13.
Zurück zum Zitat Ketelsen D, Buchgeister M, Fenchel M, Schmidt B, Flohr TG, Syha R et al (2012) Automated computed tomography dose-saving algorithm to protect radiosensitive tissues: estimation of radiation exposure and image quality considerations. Investig Radiol 47(2):148–152CrossRef Ketelsen D, Buchgeister M, Fenchel M, Schmidt B, Flohr TG, Syha R et al (2012) Automated computed tomography dose-saving algorithm to protect radiosensitive tissues: estimation of radiation exposure and image quality considerations. Investig Radiol 47(2):148–152CrossRef
14.
Zurück zum Zitat Duan X, Wang J, Christner JA, Leng S, Grant KL, McCollough CH (2011) Dose reduction to anterior surfaces with organ-based tube-current modulation: evaluation of performance in a phantom study. AJR Am J Roentgenol 197(3):689–695CrossRefPubMed Duan X, Wang J, Christner JA, Leng S, Grant KL, McCollough CH (2011) Dose reduction to anterior surfaces with organ-based tube-current modulation: evaluation of performance in a phantom study. AJR Am J Roentgenol 197(3):689–695CrossRefPubMed
15.
Zurück zum Zitat Mettler FA, Bhargavan M, Faulkner K, Gilley DB, Gray JE, Ibbott GS et al (2009) Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources: 1950-2007. Radiology 253(2):520–531CrossRefPubMed Mettler FA, Bhargavan M, Faulkner K, Gilley DB, Gray JE, Ibbott GS et al (2009) Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources: 1950-2007. Radiology 253(2):520–531CrossRefPubMed
16.
Zurück zum Zitat Mathews JD, Forsythe AV, Brady Z, Butler MW, Goergen SK, Byrnes GB et al (2013) Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ 346(may21 1):f2360CrossRefPubMedPubMedCentral Mathews JD, Forsythe AV, Brady Z, Butler MW, Goergen SK, Byrnes GB et al (2013) Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ 346(may21 1):f2360CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, Kim KP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380(9840):499–505CrossRefPubMedPubMedCentral Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, Kim KP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380(9840):499–505CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Hoang JK, Yoshizumi TT, Choudhury KR, Nguyen GB, Toncheva G, Gafton AR et al (2012) Organ-based dose current modulation and thyroid shields: techniques of radiation dose reduction for neck CT. AJR Am J Roentgenol 198(5):1132–1138CrossRefPubMed Hoang JK, Yoshizumi TT, Choudhury KR, Nguyen GB, Toncheva G, Gafton AR et al (2012) Organ-based dose current modulation and thyroid shields: techniques of radiation dose reduction for neck CT. AJR Am J Roentgenol 198(5):1132–1138CrossRefPubMed
19.
Zurück zum Zitat Preston DL, Mattsson A, Holmberg E, Shore R, Hildreth NG, Boice JD Jr (2002) Radiation effects on breast cancer risk: a pooled analysis of eight cohorts. Radiat Res 158(2):220–235CrossRefPubMed Preston DL, Mattsson A, Holmberg E, Shore R, Hildreth NG, Boice JD Jr (2002) Radiation effects on breast cancer risk: a pooled analysis of eight cohorts. Radiat Res 158(2):220–235CrossRefPubMed
20.
Zurück zum Zitat Lee AH (2005) Why is carcinoma of the breast more frequent in the upper outer quadrant? A case series based on needle core biopsy diagnoses. Breast 14(2):151–152CrossRefPubMed Lee AH (2005) Why is carcinoma of the breast more frequent in the upper outer quadrant? A case series based on needle core biopsy diagnoses. Breast 14(2):151–152CrossRefPubMed
Metadaten
Titel
Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue
verfasst von
André Euler
Zsolt Szucs-Farkas
Anna L. Falkowski
Nadine Kawel-Böhm
Luigia D’Errico
Sebastién Kopp
Jens Bremerich
Tilo Niemann
Publikationsdatum
12.11.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 8/2016
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-015-4085-5

Weitere Artikel der Ausgabe 8/2016

European Radiology 8/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.