Outflow graft anastomosis site design could be correlated to aortic valve regurgitation under left ventricular assist device support
verfasst von:
Kei Iizuka, Tomohiro Nishinaka, Yuki Ichihara, Takuma Miyamoto, Kenji Yamazaki
Erschienen in:
Journal of Artificial Organs
|
Ausgabe 2/2018
Einloggen, um Zugang zu erhalten
Abstract
Aortic valve regurgitation (AR) is a critical complication during circulatory support with a left ventricular assist device (LVAD). The time-course of AR and related factors, including outflow graft anastomosis site design, were investigated. Twenty-three patients who had continuous-flow LVAD implantation and were supported for more than 6 months were investigated. AR grade (none, 0; trivial, 0.5; mild, 1; mild-moderate, 1.5; moderate, 2; moderate-severe, 2.5; severe, 3) and aortic valve opening were evaluated with echocardiography. Computed tomography was performed to all the patients postoperatively. The angle of the outflow graft to the aorta (O-A angle, parallel 0; tangent 90°, 0–180°), aortic diameter at the anastomosis site, sino-tubular junction (STJ) diameter, distance between the STJ and the anastomosis site, and distance between the anastomosis site and the brachiocephalic artery were measured. The patients’ age was 38 ± 11 years. Support duration was 686 ± 354 days. Mean AR grade after continuous-flow LVAD implantation was increased to around mild and was maintained thereafter. No patient needed any intervention to the aortic valve. The aortic valves of 82.6% of patients were closed continuously. The O-A angle (83 ± 14) was positively correlated with maximum AR grade (p = 0.0095). The O-A angle was significantly smaller in patients with maximum AR grade of 1 or less (77 ± 9°) than in those with 1.5 or greater (94 ± 15°, p = 0.021). The other CT measurements had no correlation with AR grade. In conclusion, the O-A angle was correlated with AR grade progression. The O-A angle appears to be one of the important factors related to AR under continuous-flow LVAD support.