Open Access 01.03.2023 | ORIGINAL RESEARCH

Patients with Intermittent Exotropia and Exophoria Exhibit Non-aggravated Lens Decentration After Orthokeratology Application: The Nanjing Strabismus Cohort

verfasst von: Tianxi Li, Xiaoxia Zuo, Tong Zhang, Lei Liu, Zhongzheng Wang, Lin Han, Hu Liu, Zijin Wang

Erschienen in: Ophthalmology and Therapy | Ausgabe 3/2023

Abstract

Introduction

There is a high prevalence of intermittent exotropia and exophoria in myopic populations, and orthokeratology is one of the effective interventions to control myopia progression in children. However, it is still obscure whether intermittent exotropia and exophoria children could wear orthokeratology without experiencing aggravated lens decentration.

Methods

This was a multi-center, prospective cohort study. A total of 123 myopic participants aged 8–14 years were recruited, where conditions of deviation included intermittent exotropia, exophoria, and orthophoria. Uncorrected visual acuity and corneal topography data were obtained at baseline and after 1 month of wearing orthokeratology lens. Lens decentration was analyzed in a MATLAB program. Magnitude of deviation and refractive errors were evaluated prior to orthokeratology treatment. Fisher’s exact test, ANOVA test, and univariate and multivariate linear regression models were established to evaluate the role of magnitude of deviation in lens decentration.

Results

There was no significant difference in magnitude and direction of lens decentration among three groups (magnitude: F = 1.25, P = 0.289; direction: Fisher = 9.91, P = 0.078). According to scale division of decentration, 1 (2.6%) intermittent exotropia subject, 2 (3.8%) exophoria subjects, and 1 (3.0%) orthophoria subject experienced severe decentration (Fisher = 1.10, P = 0.947). Inferotemporal decentration was most common among all subjects (intermittent exotropia 50.0%, exophoria 76.9%, orthophoria 72.7%). Univariate and multivariate linear regression analyses revealed that magnitude of deviation was not an independent risk factor for lens decentration [β = −0.00, 95% confidence interval (CI) −0.01–0.00, P = 0.180], while surface asymmetry index (SAI) (β = 0.21, 95% CI 0.02–0.40, P = 0.028) and surface regularity index (SRI) (β = −0.39, 95% CI −0.66 to −0.13, P = 0.004) had significant correlation with polar decentration.

Conclusion

Patients with intermittent exotropia and exophoria exhibit non-aggravated lens decentration after orthokeratology application. Thus, lens decentration is not the concern for orthokeratology prescription.
Hinweise
Tianxi Li, Xiaoxia Zuo and Tong Zhang contributed equally to the study.
Key Summary Points
Why carry out this study?
There is a high prevalence of intermittent exotropia and exophoria among myopic children. Though orthokeratology has been proved effective in myopia control in recent years, it is still unclear whether children with intermittent exotropia could wear orthokeratology without experiencing increased adverse effects.
What was learned from the study?
We compared lens decentration and uncorrected visual acuity after orthokeratology application among intermittent exotropia, exophoria, and orthophoria populations. Subjects with intermittent exotropia and exophoria manifested good response with non-aggravated lens decentration and clear vision.
We confirmed that lens decentration is not a concern for orthokeratology prescription in subjects with intermittent exotropia.

Introduction

Exophoria (XPH) is a tendency of oculomotor deviation which can be compensated by fusion to maintain alignment and binocular vision. Intermittent exotropia (IXT) is a transitional strabismus between XPH and constant exotropia, where fusion is only able to control or motor fuse the exodeviation for some time. It is quite common that myopia coexists with XPH or IXT. A population-based study of Australian school children confirmed that 69–75% of children with myopia presented with XPH when near, while the proportion was 9–17% at distance [1]. A retrospective study estimated that myopia occurs in more than 90% of IXT by 20 years of age, which is much higher than that in non-strabismus populations [2]. The increase of accommodation and accommodative convergence to compensate for the insufficiency of fusional convergence in IXT was recognized as one of the most important explanations for the correlation [3, 4]. Therefore, IXT was identified as an influencing factor for myopia. On the other hand, myopia was regarded as an independent indicator for concomitant exotropia. It has been reported that children with myopia had 2.85- to 10.27-fold increase in risk to develop exotropia than those without significant ametropia [57]. In recent years, studies regarding natural history of IXT reminded ophthalmologists that surgery is not the only therapeutic option for patients with IXT. Regular follow-up and observation could be preferable treatment modalities under some conditions [8]. Due to the common coexistence and possible mutual promotion between myopia and IXT, it is necessary to pay attention to myopia control in patients with IXT and XPH under observation.
Orthokeratology (Ortho-K), a kind of optical correction method, is one of the most effective interventions to slow down the progression of myopia in children [9]. It is characterized by specially designed reverse geometry and gas-permeable overnight contact lens, temporarily reshaping the cornea by flattening the central zone and steepening the peripheral zone [10]. Peripheral hyperopic defocus and higher-order aberrations were proved to produce marked effect [11, 12]. Lens decentration (LD) refers to the deviation from pupil center to the treatment zone (TZ) center after Ortho-K application. Severe LD could give rise to a variety of vision disturbances (including glare, ghost images, double vision, or reduction in contrast sensitivity), by inducing unexpected astigmatism and higher-order aberrations [13, 14]. In vision shaping treatment (VST) of Ortho-K, alignment curve (AC) is designed to maintain and guarantee the correction of lens fixing and position [15]. However, due to inevitable asymmetry of corneal surface and eyelid pressure, LD remains the most common fitting challenge of Ortho-K.
Fusional vergence, or motor fusion, is an optomotor reflex that aligns images of the two eyes globally. It Is generally accepted that fusion would be less exerted during sleep, thus exodeviation would manifest in patients with XPH and IXT [16, 17]. Due to the possible divergent eye position in IXT during sleep, it is speculated that asymmetry of eyelid pressure may arise and produce adverse effect on the stability of lens location. In the opinion of many ophthalmologists and optometrists, patients with IXT are more likely to experience severe LD after Ortho-K treatment and become less appropriate for wearing Ortho-K. However, excluding the important procedure of Ortho-K, on the basis of empirical speculation for IXT complicated by patients with myopia, lacks evidence, and whether Ortho-k may produce enough followability during sleep for these patients triggers curiosity. This study was designed to determine whether patients with XPH or IXT would experience aggravated LD after Ortho-K application compared with orthophoria in the Nanjing Strabismus Cohort.

Methods

The Nanjing Strabismus Cohort aims to systematically integrate clinical and population resources to achieve strabismus risk assessment and prediction, individualized prevention, precise diagnosis and treatment, and efficacy monitoring in Nanjing, Jiangsu, China. The present study was conducted in the First Affiliated Hospital with Nanjing Medical University and Nanjing Orange Optometry Clinic. The study was approved by the institutional review board in the First Affiliated Hospital with Nanjing Medical University (2019-SR-134) and was conducted in accordance with the tenets of the Declaration of Helsinki. Written consent was obtained from the guardians of all children.

Subjects

Inclusion criteria: (1) age between 8 and 14 years old, (2) myopia between −0.75 and −6.00 D, (3) astigmatism less than or equal to 1.50 D, (4) equal visual acuity (best corrected visual acuity, BCVA) of 0.20 logMAR or better in each eye, and (5) choosing α ORTHO-K.
Exclusion criteria: (1) constant exotropia, (2) prior history of wearing contact lenses, (3) ocular injury or prior surgery, (4) Ortho-K contraindications, and (5) simultaneous application of other measures for myopia control.
Subjects included three groups as follows: (1) IXT group: intermittent or constant exotropia at distance, and XPH or IXT when near, magnitude of deviation ≥ 15 prism diopters (PD) at distance or near measurements by the prism and alternate cover test (PACT), and near/distance deviation difference of ≤ 10PD. (2) Orthophoria group: absolute magnitude of deviation ≤ 2PD both at distance and when near. (3) XPH group: manifest between orthophoria and IXT.
All enrolled participants were fitted with α ORTHO-K (Alpha Corporation, Japan) made of Boston material according to the manufacturer’s guidelines. The optical center thickness was 0.22 mm. The width of base curve (BC) was determined by spherical errors, Jessen factor of 0.75, and flattest corneal curvature. The total diameter was determined by subtraction of 0.1 mm from the horizontal visible iris diameter (HVID), with modifications made to the width of AC. A typical bull’s eye pattern observed by corneal topography on the second day confirmed a satisfactory fit. Subjects enrolled were instructed about the wearing, removing, and caring procedures and wore their Ortho-K lenses on both eyes every night for at least 7 consecutive hours. All follow-up examinations were performed within 2 h after removal of the Ortho-K lens.

Parameters Measured

Baseline tests included measurements of uncorrected visual acuity (UCVA), BCVA, magnitude of deviation at distance and when near, refraction, axial length, and corneal topographic parameters. Magnitude of deviation was estimated at 33 cm and 6 m using PACT after performing a 1-h occlusion on the non-dominant eye. Magnitude of deviation at distance was included in the analysis. A previous study has shown that the corneal topography usually stabilizes after 1 month of Ortho-K wearing [18]. Therefore, all the measurements were repeated after 1 month of wearing. All the evaluations were conducted by one professional examiner.
The Medmont E300 corneal topographer (Medmont Studio 4 software version 4.14; Medmont International Pty Ltd, Victoria, Australia) was used to measure corneal topography. Maps with an accuracy of ≥ 95% were brought into analyses. Flat k (k1), Steep k (k2), flat eccentricity (e1), steep eccentricity (e2), surface asymmetry index (SAI), surface regularity index (SRI), and pupil diameter (PD) were recorded, while e ratio (e1/e2) and corneal toricity (k2 − k1) were calculated.
After Ortho-K treatment, a difference map was obtained by subtracting the pre-Ortho-K tangential curvature map from the 1-month post-Ortho-K tangential curvature map. LD was computed in a MATLAB program (Fig. 1). Modified from previous studies [1922], we used regiongrowing and imfill function to obtain the outline of TZ. Then, coordinates of the outline were extracted and accessed into circle fit function to calculate the best-fitting circle. The center of the circle was defined as the center of TZ, and its distance from the pupil center by automatic recognition was defined as the magnitude of LD. As LD was moderately correlated between right and left eyes (r = 0.553, P < 0.001), analyses were conducted on right eyes only.

Statistical Analysis

Data analyses were performed using the IBM Statistical Package for the Social Sciences program statistical package V21.0 (SPSS Inc, Chicago, IL, USA). All statistical tests were two-sided, and P < 0.05 was considered statistically significant. The Kolmogorov–Smirnov test was used to assess the normality of the data distribution. Descriptive statistics were presented as mean ± standard deviation or median (interquartile range, IQR) based on normality test, and as percentage for the categorical measures. Pearson correlation coefficient was used to determine the correlation of LD between two eyes. LD among groups was compared using one-way ANOVA or Kruskal–Wallis H test based on Kolmogorov–Smirnov test for continuous variables, and χ2 test or Fisher's exact test (if the expected count in any cell of the contingency table was < 5) for categorical variables. Univariate and multivariate linear regression models were used to explore the influencing factors of LD.

Results

A total of 123 subjects were included in the analyses, comprised by 38 IXT, 52 XPH, and 33 orthophoria. Table 1 shows the baseline characteristics of IXT, XPH, and orthophoria groups before treatment. Subjects among groups did not show significant difference at baseline (P > 0.05). None of the Ortho-K wearers complained of discomfort, low vision, or obvious adverse events (e.g., severe corneal staining graded on the Oxford scheme [23], corneal abrasion, conjunctivitis, contact lens-induced peripheral ulcer). After 1 month of Ortho-K wearing, the median (IQR) of UCVA was 0.08 (0.00, 0.08) in orthophoria group, 0.04 (0.00, 0.08) in XPH group, and 0.08 (0.00, 0.08) in IXT group. There was no statistically significant difference in UCVA among three groups (H = 0.35, P = 0.841).
Table 1
Baseline characteristics of intermittent exotropia, exophoria, and orthophoria groups
Variables
Intermittent exotropia (n = 38)
Exophoria (n = 52)
Orthophoria (n = 33)
F/H/χ2
P value
Age
10.00 (9.00, 11.50)
9.50 (9.00, 11.00)
9.00 (8.00, 12.00)
2.42
0.298
Sex
   
1.80
0.407
 Male
16 (42.1%)
25 (48.1%)
11 (33.3%)
  
 Female
22 (57.9%)
27 (51.9%)
22 (66.7%)
  
Lens diameter (mm)
10.60 (10.60, 10.65)
10.60 (10.50, 10.60)
10.60 (10.60, 10.60)
0.86
0.650
Steep k (D)
44.16 ± 1.30
44.25 ± 1.31
44.02 ± 1.30
0.31
0.731
Flat k (D)
42.89 ± 1.23
43.06 ± 1.14
42.78 ± 1.36
0.58
0.563
Corneal toricity (D)
1.27 ± 0.48
1.18 ± 0.47
1.24 ± 0.50
0.39
0.677
Steep e
0.50 ± 0.16
0.55 ± 0.21
0.57 ± 0.14
1.47
0.233
Flat e
0.45 ± 0.13
0.44 ± 0.13
0.42 ± 0.09
0.39
0.679
e ratio
1.20 ± 0.49
1.31 ± 0.52
1.40 ± 0.42
1.60
0.206
SAI
0.55 ± 0.25
0.59 ± 0.04
0.55 ± 0.16
0.53
0.589
SRI
0.48 ± 0.15
0.51 ± 0.19
0.50 ± 0.17
0.29
0.751
PD (mm)
4.69 ± 1.00
4.76 ± 0.77
4.57 ± 1.20
0.36
0.699
Magnitude of deviation (PD)
20.34 ± 5.57
8.94 ± 2.17
0.60 ± 0.72
280.06
0.000
AL (mm)
24.53 ± 0.90
24.39 ± 0.78
24.48 ± 0.84
0.34
0.713
SE (D)
− 2.79 ± 1.30
− 2.13 ± 1.38
− 2.47 ± 1.36
2.65
0.075
n number of subjects, SAI surface asymmetry index, SRI surface regularity index, PD pupil diameters, AL axial length, SE sphere equivalent
The magnitude and direction of LD after 1 month Ortho-K treatment were presented in Table 2 and Fig. 2. For magnitude of LD, a negative value means temporal and inferior shift, while a positive value stands for nasal and superior shift. No significant difference was found in magnitude of polar, horizontal, or vertical decentration among 3 groups (polar decentration: F = 1.25, P = 0.289; horizontal decentration: F = 1.46, P = 0.236; vertical decentration: F = 0.57, P = 0.567). Besides, 1 (2.6%) in IXT group, 2 (3.8%) in XPH group, and 1 (3.0%) in orthophoria group experienced severe decentration (polar decentration > 1 mm; Fisher = 1.10, P = 0.947). There was no significant difference in the direction of LD (Fisher = 9.91, P = 0.078), with 19 (50.0%) of IXT, 40 (76.9%) of XPH, and 24 (72.7%) of orthophoria experiencing inferotemporal decentration.
Table 2
Magnitude and direction of lens decentration in intermittent exotropia, exophoria and orthophoria groups
 
Intermittent exotropia
Exophoria
Orthophoria
F/Fisher
P value
Magnitude of treatment zone decentration
 Polar decentration (mm)
0.51 ± 0.24
0.49 ± 0.24
0.51 ± 0.24
1.25
0.289
 Horizonal decentration (mm)
− 0.28 ± 0.34
− 0.34 ± 0.23
− 0.40 ± 0.30
1.46
0.236
 Vertical decentration (mm)
− 0.14 ± 0.33
− 0.21 ± 0.29
− 0.18 ± 0.33
0.57
0.567
Quadrant of treatment zone decentration
   
9.91
0.078
 Supernasal
1 (2.6%)
1 (1.9%)
0 (0.0%)
  
 Superotemporal
12 (31.6%)
9 (17.3%)
8 (24.2%)
  
 Inferotemporal
19 (50.0%)
40 (76.9%)
24 (72.7%)
  
 Inferonasal
6 (15.8%)
2 (3.8%)
1 (3.0%)
  
One-way ANOVA analysis and Fisher’s exact test were applied corresponding to magnitude and direction of lens decentration
Univariate linear regression analysis revealed that SRI was significantly negatively associated with polar LD. Though SAI was not significantly associated with LD in the univariate analysis, we put it into multivariate analysis, as positive association was found in a previous study. Whether deviation affects LD was our concern. Thus, apart from SRI and SAI, we included magnitude of deviation into multivariate linear regression model. However, magnitude of deviation was not significantly associated with LD (β = −0.00, 95% CI −0.01 to 0.00, P = 0.180). SAI (β = 0.21, 95% CI 0.02–0.40, P = 0.028) and SRI (β = −0.39, 95% CI −0.66 to −0.13, P = 0.004) were still significantly associated with LD (Table 3).
Table 3
Linear regression analyses of the association of baseline characteristics with polar decentration
 
Univariate model
Multivariate model
Beta (95% CI)
P value
Beta (95% CI)
P value
Lens diameter
0.05 (−0.21, 0.30)
0.714
Steep k (D)
−0.02 (−0.05, 0.01)
0.248
Flat k (D)
−0.03 (−0.06, 0.01)
0.106
Corneal toricity (D)
0.05 (−0.04, 0.14)
0.317
Steep e
−0.17 (−0.42, 0.08)
0.175
Flat e
−0.13 (−0.49, 0.24)
0.485
e ratio
−0.04 (−0.13, 0.05)
0.423
SAI
0.11 (−0.07, 0.29)
0.228
0.21 (0.02, 0.40)
0.028
SRI
−0.30 (−0.55, −0.04)
0.022
−0.39 (−0.66, −0.13)
0.004
PD
0.01 (−0.04, 0.06)
0.684
Magnitude of deviation
−0.00 (−0.01, 0.00)
0.253
−0.00 (−0.01, 0.00)
0.180
AL
0.04 (−0.01, 0.09)
0.113
SAI surface asymmetry index, SRI surface regularity index, PD pupil diameters, SE sphere equivalent, AL axial length

Discussion

In this study, we compared LD and UCVA after Ortho-K application among IXT, XPH, and orthophoria populations and found no significant difference.
With the deepening of insight into the natural course of IXT [8, 24], regular follow-up visits and observation have been the first choice for many patients with IXT. As we have known, there is a high prevalence of IXT and XPH among myopic children [1, 6]. For these children, the need of myopia control should not be ignored, and should even be addressed due to the possibility of mutual promotion between IXT and myopia. However, controversy still remains whether IXT children could choose myopia control procedures without experiencing increased adverse effects.
Ortho-K is a kind of eye contact lenses that is worn overnight, to correct myopia and to slow myopia progression. Due to asymmetry of eyelid pressure and corneal topographical parameters [20] (paracentral corneal asymmetry, corneal elevation, the diameter of the cornea [25], and Q value [26]), LD is common and ineluctable in Ortho-K application. Consistent with a previous study [21], the most common direction of LD was found to be inferonasal. This might result from the steeper corneal shape in temporal and inferior areas, and the tendency of lenses drifting to the steeper side [25]. Similarly, according to linear regression models, we found SAI as an independent risk factor for LD and SRI as a protective factor. SAI stands for the asymmetry of the corneal surface, which is a centrally weighted summation of differences in corneal power between corresponding points in opposite directions on 128 equally separated meridians. On the contrary, SRI represents the regularity of corneal surface, obtained by a summation of local corneal power along 256 equally separated meridians [27].
In the opinion of many doctors and optometrists, Ortho-K has been recommended less for patients with IXT due to the concern for the possibility of aggravated LD. In our study, among IXT, XPH, and orthophoria groups, we did not find any statistical difference either in magnitude or direction of LD. The reasons might include the following. Firstly, favorable fitting of Ortho-K lenses depends on proper matching of the lens sag height with the corneal sag height along the flatter corneal meridian, to create a water-tight space filled with tear [20]. According to hydrodynamics ductility principle, when wearing an eye contact lens, once “seesaw” (two sides of the lens become tilted) occurs in any corneal meridian, lens will decenter to get access to another water-tight location. Even during rapid eye movement (REM), where bursts of eye movements are present and characterized as ocular rotation at a speed greater than 30° per second, well-matched Ortho-K lens could fit well on the central part of cornea [28]. Secondly, sleep–wake cycles include four periods: awake, non-rapid eye movement (NREM) sleep, period of transition to REM sleep, and REM sleep [29], with NREM occupying most of the sleep time [30]. During non-REM sleep, the two eyes slowly rotate upwards and in the abducting direction, producing a tonic divergent position [31]. Thus, it is rational to speculate that the three groups shared similar divergent eye position during NREM sleep [32], which makes for comparable magnitude and direction of LD after overnight wearing of Ortho-K lenses.
Average magnitude of decentration in orthophoria group in this study was 0.51 mm and in accordance with previous reports [19, 20]. It is generally regarded that, decentration less than 0.5 mm (mild decentration) is optimal while 0.5–1.0 mm (moderate decentration) is acceptable, and greater than 1.0 mm (severe decentration) should be avoided [33]. Only 1 in 38 of the IXT group experienced severe decentration, with a vertical decentration of −1.17 mm, which was comparable to other groups, produced by distinct eyelid pressure and inappropriate fit tightness. Also, none of the IXT or XPH subjects experienced low vision or under-correction. Therefore, LD should not be a main concern when prescribing overnight Ortho-K for patients with XPH or IXT.
To our knowledge, this is the first study to evaluate the application of Ortho-K in patients with IXT. However, one limitation is that patients with IXT included in this study manifested a deviation that ranged from 0 to 37 PD, and the application of Ortho-K in IXT with larger angles of deviation needs further study. Secondly, we only confirmed the non-aggravated LD of Ortho-K in IXT and XPH subjects. It is promising to evaluate the effect of Ortho-K in myopia control, binocular function, and exodeviation control for patients with IXT in the future.

Conclusion

In conclusion, we did not find aggravated LD after Ortho-K treatment in patients with IXT or XPH. LD is not a concern for Ortho-K prescription.

Acknowledgements

We thank the children, their parents or legal guardians, all the members of Department of Ophthalmology, The First Affiliated Hospital with Nanjing Medical University, Nanjing, China, and Nanjing Orange Optometry Clinic, Nanjing, China for their helpful advice and support.

Funding

This work was supported by the National Natural Science Foundation of China (Grant No.82273159); Jiangsu Province's Science and Technology Project (Grant No. BE2020722). The study sponsors are also funding the journal’s Rapid Service Fee, but had no role in the design or conduct of this research.

Author Contributions

Zijin Wang and Hu Liu designed the study and revised this article. Tianxi Li contributed to writing and revision of this article. Tianxi Li and Xiaoxia Zuo performed data interpretation and analysis. Tong Zhang, Lei Liu, Zhongzheng Wang and Lin Han performed the ocular examinations.

Disclosures

Tianxi Li, Xiaoxia Zuo, Tong Zhang, Lei Liu, Zhongzheng Wang, Lin Han, Hu Liu and Zijin Wang have nothing to disclose.

Compliance with Ethics Guidelines

The study was approved by the institutional review board in the First Affiliated Hospital with Nanjing Medical University (2019-SR-134) and was conducted in accordance with the tenets of the Declaration of Helsinki. Written consent was obtained from the guardians of all children.

Data Availability

The data used in this study are not publicly available because they contain information that could compromise research participants’ privacy, but may be available from the corresponding author upon reasonable request.
Open AccessThis article is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License, which permits any non-commercial use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by-nc/​4.​0/​.
Literatur
1.
Zurück zum Zitat Leone JF, Cornell E, Morgan IG, et al. Prevalence of heterophoria and associations with refractive error, heterotropia and ethnicity in Australian school children. Br J Ophthalmol. 2010;94(5):542–6.CrossRefPubMed Leone JF, Cornell E, Morgan IG, et al. Prevalence of heterophoria and associations with refractive error, heterotropia and ethnicity in Australian school children. Br J Ophthalmol. 2010;94(5):542–6.CrossRefPubMed
2.
Zurück zum Zitat Ekdawi NS, Nusz KJ, Diehl NN, et al. The development of myopia among children with intermittent exotropia. Am J Ophthalmol. 2010;149(3):503–7.CrossRefPubMedPubMedCentral Ekdawi NS, Nusz KJ, Diehl NN, et al. The development of myopia among children with intermittent exotropia. Am J Ophthalmol. 2010;149(3):503–7.CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Ahn SJ, Yang HK, Hwang JM. Binocular visual acuity in intermittent exotropia: role of accommodative convergence. Am J Ophthalmol. 2012;154(6):981-986 e3.CrossRefPubMed Ahn SJ, Yang HK, Hwang JM. Binocular visual acuity in intermittent exotropia: role of accommodative convergence. Am J Ophthalmol. 2012;154(6):981-986 e3.CrossRefPubMed
4.
Zurück zum Zitat Ha S-G, Jang S-M, Cho YA, et al. Clinical exhibition of increased accommodative loads for binocular fusion in patients with basic intermittent exotropia. BMC Ophthalmol. 2016;16(1):77.CrossRefPubMedPubMedCentral Ha S-G, Jang S-M, Cho YA, et al. Clinical exhibition of increased accommodative loads for binocular fusion in patients with basic intermittent exotropia. BMC Ophthalmol. 2016;16(1):77.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Robaei D, Kifley A, Mitchell P. Factors associated with a previous diagnosis of strabismus in a population-based sample of 12-year-old Australian children. Am J Ophthalmol. 2006;142(6):1085–8.CrossRefPubMed Robaei D, Kifley A, Mitchell P. Factors associated with a previous diagnosis of strabismus in a population-based sample of 12-year-old Australian children. Am J Ophthalmol. 2006;142(6):1085–8.CrossRefPubMed
6.
Zurück zum Zitat Zhu H, Yu JJ, Yu RB, et al. Association between childhood strabismus and refractive error in Chinese preschool children. PLoS ONE. 2015;10(3): e0120720.CrossRefPubMedPubMedCentral Zhu H, Yu JJ, Yu RB, et al. Association between childhood strabismus and refractive error in Chinese preschool children. PLoS ONE. 2015;10(3): e0120720.CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Cotter SA, Varma R, Tarczy-Hornoch K, et al. Risk factors associated with childhood strabismus: the multi-ethnic pediatric eye disease and Baltimore pediatric eye disease studies. Ophthalmology. 2011;118(11):2251–61.CrossRefPubMed Cotter SA, Varma R, Tarczy-Hornoch K, et al. Risk factors associated with childhood strabismus: the multi-ethnic pediatric eye disease and Baltimore pediatric eye disease studies. Ophthalmology. 2011;118(11):2251–61.CrossRefPubMed
8.
Zurück zum Zitat Pediatric Eye Disease Investigator G, Writing C, Mohney BG, et al. Three-year observation of children 3 to 10 years of age with untreated intermittent exotropia. Ophthalmology. 2019;126(9):1249–60.CrossRef Pediatric Eye Disease Investigator G, Writing C, Mohney BG, et al. Three-year observation of children 3 to 10 years of age with untreated intermittent exotropia. Ophthalmology. 2019;126(9):1249–60.CrossRef
9.
Zurück zum Zitat Huang J, Wen D, Wang Q, et al. Efficacy comparison of 16 interventions for myopia control in children: a network meta-analysis. Ophthalmology. 2016;123(4):697–708.CrossRefPubMed Huang J, Wen D, Wang Q, et al. Efficacy comparison of 16 interventions for myopia control in children: a network meta-analysis. Ophthalmology. 2016;123(4):697–708.CrossRefPubMed
11.
Zurück zum Zitat Hiraoka T, Kakita T, Okamoto F, et al. Influence of ocular wavefront aberrations on axial length elongation in myopic children treated with overnight orthokeratology. Ophthalmology. 2015;122(1):93–100.CrossRefPubMed Hiraoka T, Kakita T, Okamoto F, et al. Influence of ocular wavefront aberrations on axial length elongation in myopic children treated with overnight orthokeratology. Ophthalmology. 2015;122(1):93–100.CrossRefPubMed
12.
Zurück zum Zitat VanderVeen DK, Kraker RT, Pineles SL, et al. Use of orthokeratology for the prevention of myopic progression in children: a report by the American Academy of Ophthalmology. Ophthalmology. 2019;126(4):623–36.CrossRefPubMed VanderVeen DK, Kraker RT, Pineles SL, et al. Use of orthokeratology for the prevention of myopic progression in children: a report by the American Academy of Ophthalmology. Ophthalmology. 2019;126(4):623–36.CrossRefPubMed
13.
Zurück zum Zitat Hiraoka T, Furuya A, Matsumoto Y, et al. Quantitative evaluation of regular and irregular corneal astigmatism in patients having overnight orthokeratology. J Cataract Refract Surg. 2004;30(7):1425–9.CrossRefPubMed Hiraoka T, Furuya A, Matsumoto Y, et al. Quantitative evaluation of regular and irregular corneal astigmatism in patients having overnight orthokeratology. J Cataract Refract Surg. 2004;30(7):1425–9.CrossRefPubMed
14.
Zurück zum Zitat Hiraoka T, Okamoto C, Ishii Y, et al. Contrast sensitivity function and ocular higher-order aberrations following overnight orthokeratology. Invest Ophthalmol Vis Sci. 2007;48(2):550–6.CrossRefPubMed Hiraoka T, Okamoto C, Ishii Y, et al. Contrast sensitivity function and ocular higher-order aberrations following overnight orthokeratology. Invest Ophthalmol Vis Sci. 2007;48(2):550–6.CrossRefPubMed
15.
Zurück zum Zitat Nti AN, Berntsen DA. Optical changes and visual performance with orthokeratology. Clin Exp Optom. 2020;103(1):44–54.CrossRefPubMed Nti AN, Berntsen DA. Optical changes and visual performance with orthokeratology. Clin Exp Optom. 2020;103(1):44–54.CrossRefPubMed
16.
Zurück zum Zitat He G. A contribution to the pathogeny of concomitant squinting. Trans Ophthalmol Soc UK. 1890;10:1. He G. A contribution to the pathogeny of concomitant squinting. Trans Ophthalmol Soc UK. 1890;10:1.
17.
18.
Zurück zum Zitat Alharbi A, Swarbrick HA. The effects of overnight orthokeratology lens wear on corneal thickness. Invest Ophthalmol Vis Sci. 2003;44(6):2518–23.CrossRefPubMed Alharbi A, Swarbrick HA. The effects of overnight orthokeratology lens wear on corneal thickness. Invest Ophthalmol Vis Sci. 2003;44(6):2518–23.CrossRefPubMed
19.
Zurück zum Zitat Chen Z, Xue F, Zhou J, et al. Prediction of orthokeratology lens decentration with corneal elevation. Optom Vis Sci. 2017;94(9):903–7.CrossRefPubMed Chen Z, Xue F, Zhou J, et al. Prediction of orthokeratology lens decentration with corneal elevation. Optom Vis Sci. 2017;94(9):903–7.CrossRefPubMed
20.
Zurück zum Zitat Maseedupally VK, Gifford P, Lum E, et al. Treatment zone decentration during orthokeratology on eyes with corneal toricity. Optom Vis Sci. 2016;93(9):1101–11.CrossRefPubMed Maseedupally VK, Gifford P, Lum E, et al. Treatment zone decentration during orthokeratology on eyes with corneal toricity. Optom Vis Sci. 2016;93(9):1101–11.CrossRefPubMed
21.
Zurück zum Zitat Gu T, Gong B, Lu D, et al. Influence of corneal topographic parameters in the decentration of orthokeratology. Eye Contact Lens. 2019;45(6):372–6.CrossRefPubMed Gu T, Gong B, Lu D, et al. Influence of corneal topographic parameters in the decentration of orthokeratology. Eye Contact Lens. 2019;45(6):372–6.CrossRefPubMed
22.
Zurück zum Zitat Chen R, Chen Y, Lipson M, et al. The effect of treatment zone decentration on myopic progression during Or-thokeratology. Curr Eye Res. 2020;45(5):645–51.CrossRefPubMed Chen R, Chen Y, Lipson M, et al. The effect of treatment zone decentration on myopic progression during Or-thokeratology. Curr Eye Res. 2020;45(5):645–51.CrossRefPubMed
23.
Zurück zum Zitat Bron AJ, Evans VE, Smith JA. Grading of corneal and conjunctival staining in the context of other dry eye tests. Cornea. 2003;22(7):640–50.CrossRefPubMed Bron AJ, Evans VE, Smith JA. Grading of corneal and conjunctival staining in the context of other dry eye tests. Cornea. 2003;22(7):640–50.CrossRefPubMed
24.
Zurück zum Zitat Cotter SA, Mohney BG, Chandler DL, et al. Three-year observation of children 12 to 35 months old with untreated intermittent exotropia. Ophthalmic Physiol Opt. 2020;40(2):202–15.CrossRefPubMedPubMedCentral Cotter SA, Mohney BG, Chandler DL, et al. Three-year observation of children 12 to 35 months old with untreated intermittent exotropia. Ophthalmic Physiol Opt. 2020;40(2):202–15.CrossRefPubMedPubMedCentral
25.
Zurück zum Zitat Li Z, Cui D, Long W, et al. Predictive role of paracentral corneal toricity using elevation data for treatment zone decentration during orthokeratology. Curr Eye Res. 2018;43(9):1083–9.CrossRefPubMed Li Z, Cui D, Long W, et al. Predictive role of paracentral corneal toricity using elevation data for treatment zone decentration during orthokeratology. Curr Eye Res. 2018;43(9):1083–9.CrossRefPubMed
26.
Zurück zum Zitat Li J, Yang C, Xie W, et al. Predictive role of corneal Q-value differences between nasal-temporal and superior-inferior quadrants in orthokeratology lens decentration. Medicine (Baltimore). 2017;96(2): e5837.CrossRefPubMed Li J, Yang C, Xie W, et al. Predictive role of corneal Q-value differences between nasal-temporal and superior-inferior quadrants in orthokeratology lens decentration. Medicine (Baltimore). 2017;96(2): e5837.CrossRefPubMed
27.
Zurück zum Zitat Wilson SE, Klyce SD. Advances in the analysis of corneal topography. Surv Ophthalmol. 1991;35(4):269–77.CrossRefPubMed Wilson SE, Klyce SD. Advances in the analysis of corneal topography. Surv Ophthalmol. 1991;35(4):269–77.CrossRefPubMed
28.
Zurück zum Zitat Takahashi K, Atsumi Y. Precise measurement of individual rapid eye movements in REM sleep of humans. Sleep. 1997;20(9):743–52.CrossRefPubMed Takahashi K, Atsumi Y. Precise measurement of individual rapid eye movements in REM sleep of humans. Sleep. 1997;20(9):743–52.CrossRefPubMed
30.
Zurück zum Zitat Le Bon O. Relationships between REM and NREM in the NREM-REM sleep cycle: a review on competing concepts. Sleep Med. 2020;70:6–16.CrossRefPubMed Le Bon O. Relationships between REM and NREM in the NREM-REM sleep cycle: a review on competing concepts. Sleep Med. 2020;70:6–16.CrossRefPubMed
31.
32.
Zurück zum Zitat Ku HC, Lee SY, Lee YC. Change of eye position in patients with orthophoria and horizontal strabismus under general anesthesia. Korean J Ophthalmol. 2005;19(1):55–61.CrossRefPubMed Ku HC, Lee SY, Lee YC. Change of eye position in patients with orthophoria and horizontal strabismus under general anesthesia. Korean J Ophthalmol. 2005;19(1):55–61.CrossRefPubMed
33.
Zurück zum Zitat Tsai Y-Y, Lin J-M. Ablation centration after active eye-tracker-assisted photorefractive keratectomy and laser in situ keratomileusis. J Cataract Refract Surg. 2000;26(1):28–34.CrossRefPubMed Tsai Y-Y, Lin J-M. Ablation centration after active eye-tracker-assisted photorefractive keratectomy and laser in situ keratomileusis. J Cataract Refract Surg. 2000;26(1):28–34.CrossRefPubMed
Metadaten
Titel
Patients with Intermittent Exotropia and Exophoria Exhibit Non-aggravated Lens Decentration After Orthokeratology Application: The Nanjing Strabismus Cohort
verfasst von
Tianxi Li
Xiaoxia Zuo
Tong Zhang
Lei Liu
Zhongzheng Wang
Lin Han
Hu Liu
Zijin Wang
Publikationsdatum
01.03.2023
Verlag
Springer Healthcare
Erschienen in
Ophthalmology and Therapy / Ausgabe 3/2023
Print ISSN: 2193-8245
Elektronische ISSN: 2193-6528
DOI
https://doi.org/10.1007/s40123-023-00685-1

Weitere Artikel der Ausgabe 3/2023

Ophthalmology and Therapy 3/2023 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Semaglutid bei Herzinsuffizienz: Wie erklärt sich die Wirksamkeit?

17.05.2024 Herzinsuffizienz Nachrichten

Bei adipösen Patienten mit Herzinsuffizienz des HFpEF-Phänotyps ist Semaglutid von symptomatischem Nutzen. Resultiert dieser Benefit allein aus der Gewichtsreduktion oder auch aus spezifischen Effekten auf die Herzinsuffizienz-Pathogenese? Eine neue Analyse gibt Aufschluss.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.