Aim

To detect early glaucoma, Porciatti and Ventura suggested in 2004 the pattern electroretinogram (PERG) protocol “PERGLA” with the following features: (1) skin electrodes, (2) steady-state reversal (15 rps) of horizontal 1.6 cpd gratings and (3) Fourier-based analysis. We compared this to our “PERG Ratio” protocol which uses (1) corneal (DTL) electrodes, (2) nearly the same reversal rate, but 2 check sizes of 0.8° and 16° and (3) Fourier-based analysis using the ratio of amplitudes to the 2 check sizes.

Methods

We examined 16 eyes with glaucoma (age 64 ± 7 years) and a control group (n = 13, age 59 ± 8 years). Responses to all stimuli were simultaneously recorded with both electrode types using PERGLA-type gratings and checkerboards as necessary for the PERG Ratio.

Results

The median intra-individual test–retest coefficient of variation in normals pooled across stimuli for skin was 10.0 (95 % CI 1–85 %) and for DTL 9.95 (0.5–49 %), and they are thus nearly identical. The amplitudes obtained from skin were on average 30 % of those with DTL; the skin’s signal-to-noise ratio (SNR) was 66 % of DTL electrodes. Glaucoma detection was assessed via receiver operating characteristics (ROCs). Using skin electrodes, ROC area-under-curve (AUC) was 72/76/72 % using gratings (PERGLA), checks or the PERG Ratio, respectively. Using DTL electrodes, the ROC areas were 60/67/77 %; the last value corresponds to the PERG Ratio protocol.

Conclusions

Our results suggest that skin electrodes are a valid alternative to corneal electrodes: their advantages being no direct eye contact and smaller normal amplitude range compared to DTL; disadvantages: amplitude reduced to 30 % of DTL and SNR ratio reduced to 66 % of DTL. Efficacy in detecting glaucoma was a little higher with the PERG Ratio protocol (ROC AUC: 77 %, PERGLA: 72 %), but not significantly so.
Literatur
1.
Zurück zum Zitat Groneberg A, Teping C (1980) Topodiagnostik von Sehstörungen durch Ableitung retinaler und kortikaler Antworten auf Umkehr-Kontrastmuster. Ber Dtsch Ophthalmol Ges 77:409–415CrossRef Groneberg A, Teping C (1980) Topodiagnostik von Sehstörungen durch Ableitung retinaler und kortikaler Antworten auf Umkehr-Kontrastmuster. Ber Dtsch Ophthalmol Ges 77:409–415CrossRef
2.
Zurück zum Zitat Maffei L, Fiorentini A (1981) Electroretinographic responses to alternating gratings before and after section of the optic nerve. Science 211:953–954CrossRef Maffei L, Fiorentini A (1981) Electroretinographic responses to alternating gratings before and after section of the optic nerve. Science 211:953–954CrossRef
3.
Zurück zum Zitat Sieving PA, Steinberg RH (1987) Proximal retinal contributions to the intraretinal 8-Hz pattern ERG of cat. J Neurophysiol 57:104–120PubMed Sieving PA, Steinberg RH (1987) Proximal retinal contributions to the intraretinal 8-Hz pattern ERG of cat. J Neurophysiol 57:104–120PubMed
4.
Zurück zum Zitat Zrenner E (1989) Physiological basis of the pattern electroretinogram. Prog Retin Res 9:427–464CrossRef Zrenner E (1989) Physiological basis of the pattern electroretinogram. Prog Retin Res 9:427–464CrossRef
5.
Zurück zum Zitat Bach M, Gerling J, Geiger K (1992) Optic atrophy reduces the pattern-electroretinogram for both fine and coarse stimulus patterns. Clin Vis Sci 7:327–333 Bach M, Gerling J, Geiger K (1992) Optic atrophy reduces the pattern-electroretinogram for both fine and coarse stimulus patterns. Clin Vis Sci 7:327–333
6.
Zurück zum Zitat Bach M, Hoffmann MB (2006) The origin of the pattern electroretinogram (PERG). In: Heckenlively J, Arden G (eds) Principles and practice of clinical electrophysiology of vision. MIT Press, Cambridge, pp 185–196 Bach M, Hoffmann MB (2006) The origin of the pattern electroretinogram (PERG). In: Heckenlively J, Arden G (eds) Principles and practice of clinical electrophysiology of vision. MIT Press, Cambridge, pp 185–196
7.
Zurück zum Zitat Bach M, Hoffmann MB (2008) Update on the pattern electroretinogram in glaucoma. Optom Vis Sci 85:386–395CrossRefPubMed Bach M, Hoffmann MB (2008) Update on the pattern electroretinogram in glaucoma. Optom Vis Sci 85:386–395CrossRefPubMed
8.
Zurück zum Zitat Holder GE (2001) Pattern electroretinography (PERG) and an integrated approach to visual pathway diagnosis. Prog Retin Eye Res 20:531–561CrossRefPubMed Holder GE (2001) Pattern electroretinography (PERG) and an integrated approach to visual pathway diagnosis. Prog Retin Eye Res 20:531–561CrossRefPubMed
9.
Zurück zum Zitat Holder GE (2001) The pattern electroretinogram. In: Fishman GA, Birch DG, Holder GE, Brigell MG (eds) Electrophysiologic testing in disorders of the retina, optic nerve and visual pathway. American Academy of Ophthalmology, San Francisco, pp 197–235 Holder GE (2001) The pattern electroretinogram. In: Fishman GA, Birch DG, Holder GE, Brigell MG (eds) Electrophysiologic testing in disorders of the retina, optic nerve and visual pathway. American Academy of Ophthalmology, San Francisco, pp 197–235
10.
Zurück zum Zitat Porciatti V, Falsini B, Brunori S, Colotto A, Moretti G (1987) Pattern electroretinogram as a function of spatial frequency in ocular hypertension and early glaucoma. Doc Ophthalmol 65:349–355CrossRefPubMed Porciatti V, Falsini B, Brunori S, Colotto A, Moretti G (1987) Pattern electroretinogram as a function of spatial frequency in ocular hypertension and early glaucoma. Doc Ophthalmol 65:349–355CrossRefPubMed
11.
Zurück zum Zitat Bach M, Hiss P, Röver J (1988) Check-size specific changes of pattern electroretinogram in patients with early open-angle glaucoma. Doc Ophthalmol 69:315–322CrossRefPubMed Bach M, Hiss P, Röver J (1988) Check-size specific changes of pattern electroretinogram in patients with early open-angle glaucoma. Doc Ophthalmol 69:315–322CrossRefPubMed
12.
Zurück zum Zitat Bayer AU, Maag KP, Erb C (2002) Detection of optic neuropathy in glaucomatous eyes with normal standard visual fields using a test battery of short-wavelength automated perimetry and pattern electroretinography. Ophthalmology 109:1350–1361CrossRefPubMed Bayer AU, Maag KP, Erb C (2002) Detection of optic neuropathy in glaucomatous eyes with normal standard visual fields using a test battery of short-wavelength automated perimetry and pattern electroretinography. Ophthalmology 109:1350–1361CrossRefPubMed
13.
Zurück zum Zitat Bach M (2001) Electrophysiological approaches for early detection of glaucoma. Eur J Ophthalmol 11(Suppl 2):S41–S49PubMed Bach M (2001) Electrophysiological approaches for early detection of glaucoma. Eur J Ophthalmol 11(Suppl 2):S41–S49PubMed
14.
Zurück zum Zitat Arai M, Yoshimura N, Sakaue H, Chihara E, Honda Y (1993) A 3-year follow-up study of ocular hypertension by pattern electroretinogram. Ophthalmologica 207:187–195CrossRefPubMed Arai M, Yoshimura N, Sakaue H, Chihara E, Honda Y (1993) A 3-year follow-up study of ocular hypertension by pattern electroretinogram. Ophthalmologica 207:187–195CrossRefPubMed
15.
Zurück zum Zitat Pfeiffer N, Tillmon B, Bach M (1993) Predictive value of the pattern-electroretinogram in high-risk ocular hypertension. Invest Ophthalmol Vis Sci 34:1710–1715PubMed Pfeiffer N, Tillmon B, Bach M (1993) Predictive value of the pattern-electroretinogram in high-risk ocular hypertension. Invest Ophthalmol Vis Sci 34:1710–1715PubMed
16.
Zurück zum Zitat Gonzalvo Ibanez FJ, Fernandez Tirado FJ, Almarcegui Lafita C, Polo Llorens V, Sanchez Perez A, Honrubia Lopez FM (2001) Predictive value of the pattern-electroretinogram in glaucoma. Arch Soc Esp Oftalmol 76:485–491PubMed Gonzalvo Ibanez FJ, Fernandez Tirado FJ, Almarcegui Lafita C, Polo Llorens V, Sanchez Perez A, Honrubia Lopez FM (2001) Predictive value of the pattern-electroretinogram in glaucoma. Arch Soc Esp Oftalmol 76:485–491PubMed
17.
Zurück zum Zitat Philippin H, Unsoeld A, Maier P, Walter S, Bach M, Funk J (2006) Ten-year results: detection of long-term progressive optic disc changes with confocal laser tomography. Graefes Arch Clin Exp Ophthalmol 244:460–464CrossRefPubMed Philippin H, Unsoeld A, Maier P, Walter S, Bach M, Funk J (2006) Ten-year results: detection of long-term progressive optic disc changes with confocal laser tomography. Graefes Arch Clin Exp Ophthalmol 244:460–464CrossRefPubMed
18.
Zurück zum Zitat Bach M, Unsoeld AS, Philippin H, Staubach F, Maier P, Walter HS, Bomer TG, Funk J (2006) Pattern ERG as an early glaucoma indicator in ocular hypertension: a long-term, prospective study. Invest Ophthalmol Vis Sci 47:4881–4887CrossRefPubMed Bach M, Unsoeld AS, Philippin H, Staubach F, Maier P, Walter HS, Bomer TG, Funk J (2006) Pattern ERG as an early glaucoma indicator in ocular hypertension: a long-term, prospective study. Invest Ophthalmol Vis Sci 47:4881–4887CrossRefPubMed
19.
Zurück zum Zitat Bode SF, Jehle T, Bach M (2011) Pattern electroretinogram (PERG) in glaucoma suspects—new findings from a longitudinal study. Invest Ophthalmol Vis Sci 52:4300–4306CrossRefPubMed Bode SF, Jehle T, Bach M (2011) Pattern electroretinogram (PERG) in glaucoma suspects—new findings from a longitudinal study. Invest Ophthalmol Vis Sci 52:4300–4306CrossRefPubMed
20.
Zurück zum Zitat Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK, Wilson MR, Gordon MO (2002) The Ocular Hypertension Treatment Study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 120:701–713; discussion 829–830CrossRefPubMed Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK, Wilson MR, Gordon MO (2002) The Ocular Hypertension Treatment Study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 120:701–713; discussion 829–830CrossRefPubMed
21.
Zurück zum Zitat Heijl A, Leske MC, Bengtsson B, Hyman L, Hussein M (2002) Reduction of intraocular pressure and glaucoma progression: results from the early manifest glaucoma trial. Arch Ophthalmol 120:1268–1279CrossRefPubMed Heijl A, Leske MC, Bengtsson B, Hyman L, Hussein M (2002) Reduction of intraocular pressure and glaucoma progression: results from the early manifest glaucoma trial. Arch Ophthalmol 120:1268–1279CrossRefPubMed
22.
Zurück zum Zitat Van den Berg TJ, Boltjes B (1987) The point-spread function of the eye from 0 degrees to 100 degrees and the pattern electroretinogram. Doc Ophthalmol 67:347–354CrossRefPubMed Van den Berg TJ, Boltjes B (1987) The point-spread function of the eye from 0 degrees to 100 degrees and the pattern electroretinogram. Doc Ophthalmol 67:347–354CrossRefPubMed
23.
Zurück zum Zitat Leipert KP, Gottlob I (1987) Pattern electroretinogram: effects of miosis, accommodation, and defocus. Doc Ophthalmol 67:335–346CrossRefPubMed Leipert KP, Gottlob I (1987) Pattern electroretinogram: effects of miosis, accommodation, and defocus. Doc Ophthalmol 67:335–346CrossRefPubMed
24.
Zurück zum Zitat Bach M, Mathieu M (2004) Different effect of dioptric defocus vs. light scatter on the pattern electroretinogram (PERG). Doc Ophthalmol 108:99–106CrossRefPubMed Bach M, Mathieu M (2004) Different effect of dioptric defocus vs. light scatter on the pattern electroretinogram (PERG). Doc Ophthalmol 108:99–106CrossRefPubMed
25.
Zurück zum Zitat Ver Hoeve JN, Danilov YP, Kim CB, Spear PD (1999) VEP and PERG acuity in anesthetized young adult rhesus monkeys. Vis Neurosci 16:607–617PubMed Ver Hoeve JN, Danilov YP, Kim CB, Spear PD (1999) VEP and PERG acuity in anesthetized young adult rhesus monkeys. Vis Neurosci 16:607–617PubMed
26.
Zurück zum Zitat Porciatti V, Ventura LM (2004) Normative data for a user-friendly paradigm for pattern electroretinogram recording. Ophthalmology 111:161–168CrossRefPubMed Porciatti V, Ventura LM (2004) Normative data for a user-friendly paradigm for pattern electroretinogram recording. Ophthalmology 111:161–168CrossRefPubMed
27.
Zurück zum Zitat Ventura LM, Porciatti V, Ishida K, Feuer WJ, Parrish RK (2005) Pattern electroretinogram abnormality and glaucoma. Ophthalmology 112:10–19CrossRefPubMed Ventura LM, Porciatti V, Ishida K, Feuer WJ, Parrish RK (2005) Pattern electroretinogram abnormality and glaucoma. Ophthalmology 112:10–19CrossRefPubMed
28.
Zurück zum Zitat Ventura LM, Porciatti V (2006) Pattern electroretinogram in glaucoma. Curr Opin Ophthalmol 17:196–202CrossRefPubMed Ventura LM, Porciatti V (2006) Pattern electroretinogram in glaucoma. Curr Opin Ophthalmol 17:196–202CrossRefPubMed
29.
Zurück zum Zitat Holder GE, Brigell MG, Hawlina M, Meigen T, Vaegan (2007) ISCEV standard for clinical pattern electroretinography—2007 update. Doc Ophthalmol 114:111–116CrossRefPubMed Holder GE, Brigell MG, Hawlina M, Meigen T, Vaegan (2007) ISCEV standard for clinical pattern electroretinography—2007 update. Doc Ophthalmol 114:111–116CrossRefPubMed
30.
Zurück zum Zitat Kakisu Y, Mizota A, Adachi E (1986) Clinical application of the pattern electroretinogram with lid skin electrode. Doc Ophthalmol 63:187–194CrossRefPubMed Kakisu Y, Mizota A, Adachi E (1986) Clinical application of the pattern electroretinogram with lid skin electrode. Doc Ophthalmol 63:187–194CrossRefPubMed
31.
Zurück zum Zitat Coupland SG, Janaky M (1989) ERG electrode in pediatric patients: comparison of DTL fiber, PVA-gel, and non-corneal skin electrodes. Doc Ophthalmol 71:427–433CrossRefPubMed Coupland SG, Janaky M (1989) ERG electrode in pediatric patients: comparison of DTL fiber, PVA-gel, and non-corneal skin electrodes. Doc Ophthalmol 71:427–433CrossRefPubMed
32.
Zurück zum Zitat McCulloch DL, Van Boemel GB, Borchert MS (1997) Comparisons of contact lens, foil, fiber and skin electrodes for patterns electroretinograms. Doc Ophthalmol 94:327–340CrossRefPubMed McCulloch DL, Van Boemel GB, Borchert MS (1997) Comparisons of contact lens, foil, fiber and skin electrodes for patterns electroretinograms. Doc Ophthalmol 94:327–340CrossRefPubMed
33.
Zurück zum Zitat Sehi M, Grewal DS, Goodkin ML, Greenfield DS (2010) Reversal of retinal ganglion cell dysfunction after surgical reduction of intraocular pressure. Ophthalmology 117:2329–2336CrossRefPubMed Sehi M, Grewal DS, Goodkin ML, Greenfield DS (2010) Reversal of retinal ganglion cell dysfunction after surgical reduction of intraocular pressure. Ophthalmology 117:2329–2336CrossRefPubMed
34.
Zurück zum Zitat Bowd C, Tafreshi A, Zangwill LM, Medeiros FA, Sample PA, Weinreb RN (2011) Pattern electroretinogram association with spectral domain-OCT structural measurements in glaucoma. Eye 25:224–232CrossRefPubMed Bowd C, Tafreshi A, Zangwill LM, Medeiros FA, Sample PA, Weinreb RN (2011) Pattern electroretinogram association with spectral domain-OCT structural measurements in glaucoma. Eye 25:224–232CrossRefPubMed
35.
Zurück zum Zitat Forte R, Ambrosio L, Bonavolontà P, Ambrosio G (2010) Pattern electroretinogram optimized for glaucoma screening (PERGLA) and retinal nerve fiber thickness in suspected glaucoma and ocular hypertension. Doc Ophthalmol 120:187–192CrossRefPubMed Forte R, Ambrosio L, Bonavolontà P, Ambrosio G (2010) Pattern electroretinogram optimized for glaucoma screening (PERGLA) and retinal nerve fiber thickness in suspected glaucoma and ocular hypertension. Doc Ophthalmol 120:187–192CrossRefPubMed
36.
Zurück zum Zitat Tafreshi A, Racette L, Weinreb RN, Sample PA, Zangwill LM, Medeiros FA, Bowd C (2010) Pattern electroretinogram and psychophysical tests of visual function for discriminating between healthy and glaucoma eyes. Am J Ophthalmol 149:488–495CrossRefPubMed Tafreshi A, Racette L, Weinreb RN, Sample PA, Zangwill LM, Medeiros FA, Bowd C (2010) Pattern electroretinogram and psychophysical tests of visual function for discriminating between healthy and glaucoma eyes. Am J Ophthalmol 149:488–495CrossRefPubMed
37.
Zurück zum Zitat Banitt MR, Ventura LM, Feuer WJ, Savatovsky E, Luna G, Shif O, Bosse B, Porciatti V (2013) Progressive loss of retinal ganglion cell function precedes structural loss by several years in glaucoma suspects. Invest Ophthalmol Vis Sci 54:2346–2352CrossRefPubMed Banitt MR, Ventura LM, Feuer WJ, Savatovsky E, Luna G, Shif O, Bosse B, Porciatti V (2013) Progressive loss of retinal ganglion cell function precedes structural loss by several years in glaucoma suspects. Invest Ophthalmol Vis Sci 54:2346–2352CrossRefPubMed
38.
Zurück zum Zitat Ventura LM, Golubev I, Feuer WJ, Porciatti V (2013) Pattern electroretinogram progression in glaucoma suspects. J Glaucoma 22:219–225CrossRefPubMed Ventura LM, Golubev I, Feuer WJ, Porciatti V (2013) Pattern electroretinogram progression in glaucoma suspects. J Glaucoma 22:219–225CrossRefPubMed
39.
Zurück zum Zitat Hiss P, Fahl G (1991) Changes in the pattern electroretinogram in glaucoma and ocular hypertension are dependent on stimulus frequency. Fortschr Ophthalmol 88:562–565PubMed Hiss P, Fahl G (1991) Changes in the pattern electroretinogram in glaucoma and ocular hypertension are dependent on stimulus frequency. Fortschr Ophthalmol 88:562–565PubMed
40.
Zurück zum Zitat Bach M, Speidel-Fiaux A (1989) Pattern electroretinogram in glaucoma and ocular hypertension. Doc Ophthalmol 73:173–181CrossRefPubMed Bach M, Speidel-Fiaux A (1989) Pattern electroretinogram in glaucoma and ocular hypertension. Doc Ophthalmol 73:173–181CrossRefPubMed
41.
Zurück zum Zitat World Medical Association (2000) Declaration of Helsinki: ethical principles for medical research involving human subjects. J Am Med Assoc 284:3043–3045CrossRef World Medical Association (2000) Declaration of Helsinki: ethical principles for medical research involving human subjects. J Am Med Assoc 284:3043–3045CrossRef
42.
Zurück zum Zitat Holopigian K, Bach M (2010) A primer on common statistical errors in clinical ophthalmology. Doc Ophthalmol 121:215–222CrossRefPubMed Holopigian K, Bach M (2010) A primer on common statistical errors in clinical ophthalmology. Doc Ophthalmol 121:215–222CrossRefPubMed
43.
Zurück zum Zitat Aulhorn E, Karmeyer H (1976) Frequency distribution in early glaucomatous visual field defects. Doc Ophthalmol Proc Ser 14:75–83 Aulhorn E, Karmeyer H (1976) Frequency distribution in early glaucomatous visual field defects. Doc Ophthalmol Proc Ser 14:75–83
45.
Zurück zum Zitat Bach M (2007) The Freiburg visual acuity test-variability unchanged by post hoc re-analysis. Graefes Arch Clin Exp Ophthalmol 245:965–971CrossRefPubMed Bach M (2007) The Freiburg visual acuity test-variability unchanged by post hoc re-analysis. Graefes Arch Clin Exp Ophthalmol 245:965–971CrossRefPubMed
47.
Zurück zum Zitat Bach M, Meigen T, Strasburger H (1997) Raster-scan cathode-ray tubes for vision research—limits of resolution in space, time and intensity, and some solutions. Spat Vis 10:403–414CrossRefPubMed Bach M, Meigen T, Strasburger H (1997) Raster-scan cathode-ray tubes for vision research—limits of resolution in space, time and intensity, and some solutions. Spat Vis 10:403–414CrossRefPubMed
48.
Zurück zum Zitat Fahle M, Bach M (2006) Basics of the VEP. In: Heckenlively J, Arden G (eds) Principles and practice of clinical electrophysiology of vision. MIT Press, Cambridge, pp 207–234 Fahle M, Bach M (2006) Basics of the VEP. In: Heckenlively J, Arden G (eds) Principles and practice of clinical electrophysiology of vision. MIT Press, Cambridge, pp 207–234
49.
Zurück zum Zitat Bach M, Meigen T (1999) Do’s and don’ts in Fourier analysis of steady-state potentials. Doc Ophthalmol 99:69–82CrossRefPubMed Bach M, Meigen T (1999) Do’s and don’ts in Fourier analysis of steady-state potentials. Doc Ophthalmol 99:69–82CrossRefPubMed
50.
Zurück zum Zitat Meigen T, Bach M (1999) On the statistical significance of electrophysiological steady-state responses. Doc Ophthalmol 98:207–232CrossRefPubMed Meigen T, Bach M (1999) On the statistical significance of electrophysiological steady-state responses. Doc Ophthalmol 98:207–232CrossRefPubMed
51.
Zurück zum Zitat Norcia AM, Tyler CW, Hamer RD, Wesemann W (1989) Measurement of spatial contrast sensitivity with the swept contrast VEP. Vis Res 29:627–637CrossRefPubMed Norcia AM, Tyler CW, Hamer RD, Wesemann W (1989) Measurement of spatial contrast sensitivity with the swept contrast VEP. Vis Res 29:627–637CrossRefPubMed
53.
Zurück zum Zitat Swets JA (1973) The relative operating characteristic in psychology A technique for isolating effects of response bias finds wide use in the study of perception and cognition. Science 182:990–1000CrossRefPubMed Swets JA (1973) The relative operating characteristic in psychology A technique for isolating effects of response bias finds wide use in the study of perception and cognition. Science 182:990–1000CrossRefPubMed
54.
Zurück zum Zitat Robin X, Turck N, Hainard A, Tiberti N, Lisacek F, Sanchez J-C, Müller M (2011) pROC: an open-source package for R and S+ to analyze and compare ROC curves. BMC Bioinformatics 12:77CrossRefPubMed Robin X, Turck N, Hainard A, Tiberti N, Lisacek F, Sanchez J-C, Müller M (2011) pROC: an open-source package for R and S+ to analyze and compare ROC curves. BMC Bioinformatics 12:77CrossRefPubMed
55.
Zurück zum Zitat Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160CrossRefPubMed Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160CrossRefPubMed
56.
Zurück zum Zitat Ventura LM, Sorokac N, De Los Santos R, Feuer WJ, Porciatti V (2006) The relationship between retinal ganglion cell function and retinal nerve fiber thickness in early glaucoma. Invest Ophthalmol Vis Sci 47:3904–3911CrossRefPubMed Ventura LM, Sorokac N, De Los Santos R, Feuer WJ, Porciatti V (2006) The relationship between retinal ganglion cell function and retinal nerve fiber thickness in early glaucoma. Invest Ophthalmol Vis Sci 47:3904–3911CrossRefPubMed
57.
Zurück zum Zitat Vizzeri G, Tafreshi A, Weinreb RN, Bowd C (2010) Effect of operator and optical defocus on the variability of pattern electroretinogram optimized for glaucoma detection (PERGLA). J Glaucoma 19:77–82PubMed Vizzeri G, Tafreshi A, Weinreb RN, Bowd C (2010) Effect of operator and optical defocus on the variability of pattern electroretinogram optimized for glaucoma detection (PERGLA). J Glaucoma 19:77–82PubMed
58.
Zurück zum Zitat Bradshaw K, Hansen R, Fulton A (2004) Comparison of ERGs recorded with skin and corneal-contact electrodes in normal children and adults. Doc Ophthalmol 109:43–55CrossRefPubMed Bradshaw K, Hansen R, Fulton A (2004) Comparison of ERGs recorded with skin and corneal-contact electrodes in normal children and adults. Doc Ophthalmol 109:43–55CrossRefPubMed
59.
Zurück zum Zitat Hennessy MP, Vaegan (1995) Amplitude scaling relationships of Burian–Allen, gold foil and Dawson, Trick and Litzkow electrodes. Doc Ophthalmol 89:235–248CrossRefPubMed Hennessy MP, Vaegan (1995) Amplitude scaling relationships of Burian–Allen, gold foil and Dawson, Trick and Litzkow electrodes. Doc Ophthalmol 89:235–248CrossRefPubMed
60.
Zurück zum Zitat Otto T, Bach M (1996) Retest variability and diurnal effects in the pattern electroretinogram. Doc Ophthalmol 92:311–323CrossRefPubMed Otto T, Bach M (1996) Retest variability and diurnal effects in the pattern electroretinogram. Doc Ophthalmol 92:311–323CrossRefPubMed
61.
Zurück zum Zitat Bartel P, Becker P, Robinson E (1991) The intrasession repeatability of pattern electroretinograms and the effects of digital filtering. Doc Ophthalmol 76:351–358CrossRefPubMed Bartel P, Becker P, Robinson E (1991) The intrasession repeatability of pattern electroretinograms and the effects of digital filtering. Doc Ophthalmol 76:351–358CrossRefPubMed
62.
Zurück zum Zitat Odom JV, Holder GE, Feghali JG, Cavender S (1992) Pattern electroretinogram intrasession reliability: a two center comparison. Clin Vis Sci 7:263–281 Odom JV, Holder GE, Feghali JG, Cavender S (1992) Pattern electroretinogram intrasession reliability: a two center comparison. Clin Vis Sci 7:263–281
63.
Zurück zum Zitat Heinrich TS, Bach M (2001) Contrast adaptation in human retina and cortex. Invest Ophthalmol Vis Sci 42:2721–2727PubMed Heinrich TS, Bach M (2001) Contrast adaptation in human retina and cortex. Invest Ophthalmol Vis Sci 42:2721–2727PubMed
64.
Zurück zum Zitat Heinrich TS, Bach M (2002) Contrast adaptation in retinal and cortical evoked potentials: no adaptation to low spatial frequencies. Vis Neurosci 19:645–650PubMed Heinrich TS, Bach M (2002) Contrast adaptation in retinal and cortical evoked potentials: no adaptation to low spatial frequencies. Vis Neurosci 19:645–650PubMed
65.
Zurück zum Zitat Porciatti V, Sorokac N, Buchser W (2005) Habituation of retinal ganglion cell activity in response to steady state pattern visual stimuli in normal subjects. Invest Ophthalmol Vis Sci 46:1296–1302CrossRefPubMed Porciatti V, Sorokac N, Buchser W (2005) Habituation of retinal ganglion cell activity in response to steady state pattern visual stimuli in normal subjects. Invest Ophthalmol Vis Sci 46:1296–1302CrossRefPubMed
66.
Zurück zum Zitat Bach M, Brigell MG, Hawlina M, Holder GE, Johnson MA, McCulloch DL, Meigen T, Viswanathan S (2013) ISCEV standard for clinical pattern electroretinography (PERG): 2012 update. Doc Ophthalmol 124:1–13CrossRef Bach M, Brigell MG, Hawlina M, Holder GE, Johnson MA, McCulloch DL, Meigen T, Viswanathan S (2013) ISCEV standard for clinical pattern electroretinography (PERG): 2012 update. Doc Ophthalmol 124:1–13CrossRef
67.
Zurück zum Zitat Bach M, Schumacher M (2002) The influence of ambient room lighting on the pattern electroretinogram (PERG). Doc Ophthalmol 105:281–289CrossRefPubMed Bach M, Schumacher M (2002) The influence of ambient room lighting on the pattern electroretinogram (PERG). Doc Ophthalmol 105:281–289CrossRefPubMed
68.
Zurück zum Zitat Hess RF, Baker CL (1984) Human pattern-evoked electroretinogram. J Neurophysiol 51:939–951PubMed Hess RF, Baker CL (1984) Human pattern-evoked electroretinogram. J Neurophysiol 51:939–951PubMed
69.
Zurück zum Zitat Bach M, Holder GE (1996) Check size tuning of the pattern-ERG: a reappraisal. Doc Ophthalmol 92:193–202CrossRefPubMed Bach M, Holder GE (1996) Check size tuning of the pattern-ERG: a reappraisal. Doc Ophthalmol 92:193–202CrossRefPubMed
70.
Zurück zum Zitat Johnson MA, Drum BA, Quigley HA, Sanchez RM, Dunkelberger GR (1989) Pattern-evoked potentials and optic nerve fiber loss in monocular laser-induced glaucoma. Invest Ophthalmol Vis Sci 30:897–907PubMed Johnson MA, Drum BA, Quigley HA, Sanchez RM, Dunkelberger GR (1989) Pattern-evoked potentials and optic nerve fiber loss in monocular laser-induced glaucoma. Invest Ophthalmol Vis Sci 30:897–907PubMed
71.
Zurück zum Zitat Preiser D, Lagrèze WA, Bach M, Poloschek CM (2013) Photopic negative response versus pattern electroretinogram in early glaucoma. Invest Ophthalmol Vis Sci 54:1182–1191CrossRefPubMed Preiser D, Lagrèze WA, Bach M, Poloschek CM (2013) Photopic negative response versus pattern electroretinogram in early glaucoma. Invest Ophthalmol Vis Sci 54:1182–1191CrossRefPubMed
Metadaten
Titel
Pattern electroretinogram to detect glaucoma: comparing the PERGLA and the PERG Ratio protocols
verfasst von
Michael Bach
Anke Ramharter-Sereinig
Publikationsdatum
01.12.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
Documenta Ophthalmologica / Ausgabe 3/2013
Print ISSN: 0012-4486
Elektronische ISSN: 1573-2622
DOI
https://doi.org/10.1007/s10633-013-9412-z

Weitere Artikel der Ausgabe 3/2013

Documenta Ophthalmologica 3/2013 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.