Peak deflection index as a predictor of a free-wall implantation of contemporary leadless pacemakers

verfasst von: Takatsugu Kajiyama, Yusuke Kondo, Masahiro Nakano, Kazuo Miyazawa, Miyo Nakano, Tomohiko Hayashi, Ryo Ito, Haruhiro Takahira, Mari Kitagawa, Yoshio Kobayashi

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 2/2021

Einloggen, um Zugang zu erhalten

Abstract

Background

Leadless pacemakers are an effective treatment for bradycardia. However, some cases exhibit pericardial effusions, presumably associated with device implantations on the right ventricular free-wall. The present study was carried out to find the ECG features during ventricular pacing with a Micra, which enabled distinguishing free-wall implantations from septal implantations without using imaging modalities.

Methods

Thirty-one consecutive patients who received Micra implantations in our facility were enrolled. The location of the device in the right ventricle was evaluated using echocardiography or computed tomography in order to determine whether the device was implanted on the septum (Sep group), apex (Apex group), or free-wall (FW group). The differences in the 12-lead ECG during ventricular pacing by the Micra were analyzed between the Sep and FW groups.

Results

The body of the Micra was clearly identifiable in 22 patients. The location of the device was classified into Sep in 12 patients, Apex in 4, and FW in 6. The mean age was highest in the FW and lowest in the Sep group (82.7 ± 6.6 vs. 72.8 ± 8.7 years, p = 0.027). The peak deflection index (PDI) was significantly larger in the FW group than Sep/Apex group in lead V1 (Sep: 0.505 ± 0.010, Apex: 0.402 ± 0.052, FW: 0.617 ± 0.043, p = 0.004) and lead V2 (Sep: 0.450 ± 0.066, Apex: 0.409 ± 0.037, FW: 0.521 ± 0.030, p = 0.011), whereas there was no difference in the QRS duration, transitional zone, and QRS notching.

Conclusion

The PDI in V1 could be useful for predicting implantations of Micra devices on the free-wall and may potentially stratify the risk of postprocedural pericardial effusions.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Baddour LM, Epstein AE, Erickson CC, Knight BP, Levison ME, Lockhart PB, et al. Circulation. 2010;121:458–77.CrossRef Baddour LM, Epstein AE, Erickson CC, Knight BP, Levison ME, Lockhart PB, et al. Circulation. 2010;121:458–77.CrossRef
2.
Zurück zum Zitat Olsen T, Jorgensen OD, Nielsen JC, Thogersen AM, Philbert BT, Johansen JB. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982-2018). Eur Heart J. 2019;40:1862–9.CrossRef Olsen T, Jorgensen OD, Nielsen JC, Thogersen AM, Philbert BT, Johansen JB. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982-2018). Eur Heart J. 2019;40:1862–9.CrossRef
3.
Zurück zum Zitat Reddy VY, Exner DV, Cantillon DJ, Doshi R, Bunch TJ, Tomassoni GF, et al. Percutaneous implantation of an entirely Intracardiac leadless pacemaker. N Engl J Med. 2015;373:1125–35.CrossRef Reddy VY, Exner DV, Cantillon DJ, Doshi R, Bunch TJ, Tomassoni GF, et al. Percutaneous implantation of an entirely Intracardiac leadless pacemaker. N Engl J Med. 2015;373:1125–35.CrossRef
4.
Zurück zum Zitat Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al. A leadless Intracardiac Transcatheter pacing system. N Engl J Med. 2016;374:533–41.CrossRef Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al. A leadless Intracardiac Transcatheter pacing system. N Engl J Med. 2016;374:533–41.CrossRef
5.
Zurück zum Zitat Roberts PR, Clementy N, Al Samadi F, Garweg C, Martinez-Sande JL, Iacopino S, et al. A leadless pacemaker in the real-world setting: the Micra Transcatheter pacing system post-approval registry. Heart Rhythm. 2017;14:1375–9.CrossRef Roberts PR, Clementy N, Al Samadi F, Garweg C, Martinez-Sande JL, Iacopino S, et al. A leadless pacemaker in the real-world setting: the Micra Transcatheter pacing system post-approval registry. Heart Rhythm. 2017;14:1375–9.CrossRef
6.
Zurück zum Zitat Hai JJ, Fang J, Tam CC, Wong CK, Un KC, Siu CW, et al. Safety and feasibility of a midseptal implantation technique of a leadless pacemaker. Heart Rhythm. 2019;16:896–902.CrossRef Hai JJ, Fang J, Tam CC, Wong CK, Un KC, Siu CW, et al. Safety and feasibility of a midseptal implantation technique of a leadless pacemaker. Heart Rhythm. 2019;16:896–902.CrossRef
7.
Zurück zum Zitat McGavigan AD, Roberts-Thomson KC, Hillock RJ, Stevenson IH, Mond HG. Right ventricular outflow tract pacing: radiographic and electrocardiographic correlates of lead position. Pacing Clin Electrophysiol. 2006;29:1063–8.CrossRef McGavigan AD, Roberts-Thomson KC, Hillock RJ, Stevenson IH, Mond HG. Right ventricular outflow tract pacing: radiographic and electrocardiographic correlates of lead position. Pacing Clin Electrophysiol. 2006;29:1063–8.CrossRef
8.
Zurück zum Zitat Pang BJ, Joshi SB, Lui EH, Tacey MA, Ling LH, Alison J, et al. Validation of conventional fluoroscopic and ECG criteria for right ventricular pacemaker lead position using cardiac computed tomography. Pacing Clin Electrophysiol. 2014;37:495–504.CrossRef Pang BJ, Joshi SB, Lui EH, Tacey MA, Ling LH, Alison J, et al. Validation of conventional fluoroscopic and ECG criteria for right ventricular pacemaker lead position using cardiac computed tomography. Pacing Clin Electrophysiol. 2014;37:495–504.CrossRef
9.
Zurück zum Zitat van Oosterom A, Oostendorp TF. ECGSIM: an interactive tool for studying the genesis of QRST waveforms. Heart. 2004;90:165–8.CrossRef van Oosterom A, Oostendorp TF. ECGSIM: an interactive tool for studying the genesis of QRST waveforms. Heart. 2004;90:165–8.CrossRef
11.
Zurück zum Zitat Hachiya H, Hirao K, Sasaki T, Higuchi K, Hayashi T, Tanaka Y, et al. Novel ECG predictor of difficult cases of outflow tract ventricular tachycardia: peak deflection index on an inferior lead. Circ J. 2010;74:256–61.CrossRef Hachiya H, Hirao K, Sasaki T, Higuchi K, Hayashi T, Tanaka Y, et al. Novel ECG predictor of difficult cases of outflow tract ventricular tachycardia: peak deflection index on an inferior lead. Circ J. 2010;74:256–61.CrossRef
12.
Zurück zum Zitat Liaw A, Wiener M. Classification and regression by randomForest. R News. 2002;2(3):18–22. Liaw A, Wiener M. Classification and regression by randomForest. R News. 2002;2(3):18–22.
13.
Zurück zum Zitat El-Chami MF, Roberts PR, Kypta A, Omdahl P, Bonner MD, Kowal RC, et al. How to implant a leadless pacemaker with a tine-based fixation. J Cardiovasc Electrophysiol. 2016;27:1495–501.CrossRef El-Chami MF, Roberts PR, Kypta A, Omdahl P, Bonner MD, Kowal RC, et al. How to implant a leadless pacemaker with a tine-based fixation. J Cardiovasc Electrophysiol. 2016;27:1495–501.CrossRef
14.
Zurück zum Zitat Ritter P, Duray GZ, Steinwender C, Soejima K, Omar R, Mont L, et al. Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study. Eur Heart J. 2015;36:2510–9.CrossRef Ritter P, Duray GZ, Steinwender C, Soejima K, Omar R, Mont L, et al. Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study. Eur Heart J. 2015;36:2510–9.CrossRef
15.
Zurück zum Zitat Piccini JP, Stromberg K, Jackson KP, Laager V, Duray GZ, El-Chami M, et al. Long-term outcomes in leadless Micra transcatheter pacemakers with elevated thresholds at implantation: Results from the Micra Transcatheter Pacing System Global Clinical Trial. Heart Rhythm. 2017;14:685–91.CrossRef Piccini JP, Stromberg K, Jackson KP, Laager V, Duray GZ, El-Chami M, et al. Long-term outcomes in leadless Micra transcatheter pacemakers with elevated thresholds at implantation: Results from the Micra Transcatheter Pacing System Global Clinical Trial. Heart Rhythm. 2017;14:685–91.CrossRef
Metadaten
Titel
Peak deflection index as a predictor of a free-wall implantation of contemporary leadless pacemakers
verfasst von
Takatsugu Kajiyama
Yusuke Kondo
Masahiro Nakano
Kazuo Miyazawa
Miyo Nakano
Tomohiko Hayashi
Ryo Ito
Haruhiro Takahira
Mari Kitagawa
Yoshio Kobayashi
Publikationsdatum
02.04.2020
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 2/2021
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-020-00724-y

Weitere Artikel der Ausgabe 2/2021

Journal of Interventional Cardiac Electrophysiology 2/2021 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.