Pedicle screw loosening: the value of radiological imagings and the identification of risk factors assessed by extraction torque during screw removal surgery

verfasst von: Xiaoliang Wu, Jiawei Shi, Jinyan Wu, Yongquan Cheng, Kaiwen Peng, Jianting Chen, Hui Jiang

Erschienen in: Journal of Orthopaedic Surgery and Research | Ausgabe 1/2019

Abstract

Background context

Pedicle screw loosening is a common complication after spine surgeries. Traditionally, it was assessed by radiological approaches, both X-ray and CT (computed tomography) scan, while reports using mechanical method to study screw loosening after spine surgery are rare. The primary objective was to study the prevalent of pedicle screw loosening according to extraction torque during screw removal surgery and access the sensitivity and specificity of both X-ray and CT scan for diagnosing screw loosening. The second objective was to identify the risk factors for low extraction torque of pedicle screw that might lead to loosening.

Methods

Thirty-three patients who underwent pedicle screw removal surgery after at least 2 years from primary surgery were evaluated preoperatively for fixation stability by X-ray and CT scan. In total, 236 screws were taken out, and the extraction torque data was recorded and analyzed to identify the sensitivity and specificity of both imaging studies for screw loosening. Furthermore, risk factors that might contribute to low extraction torque were also studied.

Results

The mean extraction torque of removed screws was 1.55 ± 1.00 Nm; a torque force of less than 1.02 Nm was used to define a screw as loosened. According to such criterion, the loosening rate was found to be 33%. X-ray had a sensitivity of 24% and a specificity of 98%, while CT scan had a sensitivity of 22% and a specificity of 96%. Extraction torque of pedicle screws inserted in fractured vertebrae was significantly lower than those in non-fractured vertebrae (p = 0.009); meanwhile, screws of non-fusion surgery had lower extraction torque when compared to those in fusion surgery (p = 0.001). BMD (bone mineral density) and age had low but significant linear relationship with screw extraction torque (p = 0.01, R2 = 0.304; p = 0.045, R2 = 0.123).

Conclusions

Our findings showed that both X-ray and CT scan had high specificity for screw loosening detection, but their sensitivities were relatively low. Surgeons needed to be more cautious when assessing screw loosening merely according to radiological examination, and aware of that screws in fractured vertebrae or non-fusion surgery were vulnerable to loosening.
Hinweise
Xiaoliang Wu and Jiawei Shi contributed equally to this work.
Abkürzungen
BMD
Bone mineral density
CT
Computed Tomography

Background

Pedicle screw fixation is widely used in spine surgery for numbers of indications, such as degenerative disease, trauma, tumor, infection, and deformity. It reduces the range of motion of the stabilized spine, increases the fusion rate, and is generally considered to be safe with relatively low complication rate associated with the device [1, 2].
One of the typical complications widely reported in literatures is screw loosening, which may lead to fixation failure and require revision surgery [3]. The key factor regarding investigation of screw loosening is the assessment of whether a screw is loosened or not, which is traditionally based on radiological approaches [3]. The diagnostic criteria for loosening developed by X-ray include the radiolucent area (thicker than 1 mm) around screw [49] and the “double halo” [4, 10] defined as the presence of radiolucent area and radiopaque rim at the same X-ray. Nevertheless, the specific details regarding X-ray criteria of loosening were not described in most papers, suggesting that the subjective viewpoints of surgeon and radiologist played an important role. Furthermore, the sensitivity and specificity of the X-ray criteria of loosening could be confounded by many factors, such as metal image artifact, intestinal gas, and display angle. CT scan was also employed by some studies and considered the diagnostic imaging modality of choice for detection of screw loosening [4, 9, 11, 12]. Ohtori et al. used both CT scan and X-ray to assess screw loosening, and their results showed that CT scan was significantly more sensitive than X-ray [11]. Nevertheless, like X-ray, the details about how the screw loosening was evaluated by CT scan were obscure and the assessment was subjective too. In general, both X-ray and CT scan lack uniform and explicit standard.
The limitation of radiological approaches led to a great variety of screw loosening rate in literatures. Some papers showed relatively low loosening rate, less than 1% in non-osteoporotic patients evaluated by X-ray [2, 5, 13], while other studies indicated a much higher rate of loosening [8, 12, 14, 15]. Roellinghoff reported that in 64 patients treated with multilevel pedicle screw fixation, 35(54.69%) patients showed radiographic signs of screw loosening [16]. The loosening rate was expected to be even higher in osteoporotic patients [17, 18]. Therefore, the actual circumstances of pedicle screw loosening are not acknowledged due to the various diagnostic criteria and conflicting reports.
Extraction torque, as an objective mechanical indicator, has been used to evaluate the mechanical fixation of pedicle screws in animal model [19]. However, this indicator has seldom been used to study pedicle screw loosening after spine surgeries, especially regarding its relationship with radiological findings. Sanden and colleagues [20], in a cohort study of 21 patients who underwent pedicle screw removal surgery, reported that the radiolucent zones around pedicle screws was associated with lower extraction torque, but they did not employ CT scan or evaluate the risk factors for loosening according extraction torque. Several factors, such as osteoporosis or osteopenia, non-fusion surgery, and long segment fixation, were considered to be related with screws loosening based on imaging study [3]. To the best of our knowledge, no study has evaluated such risk factors according to the extraction torque in vivo. Therefore, we intended to investigate the pedicle screw loosening rate using extraction torques during instrumentation removal surgery, and compare it with X-ray and CT image findings. Meanwhile, by using extraction torque data, we analyzed the risk factors of screw loosening.

Patients and methods

This was a prospective designed single-center study. Patients who underwent pedicle screw removal surgery were screened for eligibility. The indications for screw removal included the following: (1) pedicle screw fixation for thoracolumbar fracture without fusion and imaging confirmed solid fracture union, (2) patients required screw removal that presented persistent axial para-midline back pain to palpation or abnormal foreign body sensation due to pedicle fixation with imaging confirmed solid fusion, and no other cause found, e.g., infection. Those with significant pedicle malplacement or destructive spine disorders, such as bone metabolic disease, were excluded from our study. Informed consent was obtained from all individual participants included in the study.
The titanium alloy pedicle screws (Johnson&Johnson, USA; Medtronic, USA; Stryker, USA; Kanghui Med, China; FULE, China) with diameter of 4.0 to 6.5 mm and length of 30 to 55 mm were used in the primary surgeries. Conventional lateral and anteroposterior radiographs were taken before and at 3 to 6 months after primary surgery. The same X-ray and CT scan were scheduled before screw removal surgery to evaluate fracture union and spine fusion, as well as stability of instrumentation. Bone mineral density (BMD) was tested using dual energy X-ray absorptiometry. The maximum extraction torque was recorded while unscrewing the screw using torque gauge with a range of 0.06 to 6.00 Nm (Park Tool, China). The same surgeon (J.S. senior resident) with sufficient practice made all the extraction torque recordings. The radiographs were evaluated by a senior spinal surgeon (X.W. senior attending doctor), who was blinded to the extraction torque and patient information. X-ray criteria for screw loosening were a radiolucent zone surrounding the screw thicker than 1 mm and/or the “double halo” sign. CT scan criterion of loosening was a no signal zone surrounding the whole body of screw on the CT image. Because of metal artifact, no signal zone was usually seen around screw tail, which could not be interpreted as screw loosening (Fig. 1).
All values are given as the mean ± SD. Statistical analyses were performed using the IBM SPSS Statistics 13.0 (Chicago, USA). The Chi-square test was used for categorical variables. Mann-Whitney U tests or Kruskal-Wallis test was used for continuous variables. TwoStep Cluster was used to clustered screws by torque. We constructed a linear regression model with Pearson correlations analysis to assess whether clinical data, such as age and BMD, were correlated with the extraction torque of screw. The level of statistical significance was selected to be p = 0.05.

Results

Thirty-three patients, 10 females, and 23 males, aged from 17 to 66 years (average age 38 years) at the time of implant removal, were included; patient characteristics are shown in Table 1. The minimal time interval between primary surgery to screw removal was 18 months. Two hundred thirty-six pedicle screws were extracted in total, including 86 in thoracic spine, 138 lumbar, and 12 sacral. One hundred thirty of them were polyaxial and 106 were monoaxial. Four breaking screws were detected. Details of the fixation segments are shown in Table 2. Due to inadequate exposure of screw and malfunction of screw tail, 6 screws had no reading of extraction torque. Therefore, the reading of extraction torque was taken in 226 screws. The mean torque of screws was 1.55 ± 1.00 Nm. The data distribution showed two prominent peaks and one valley. There is a dramatic decline in screw numbers between 0.90 and 1.20 Nm. The screws were clustered into three clusters based on torque by Twostep Cluster as shown in Fig. 2. The cutoff values were 1.02 and 2.22 Nm. Given such result, the torques of 36 pedicle screws in 8 patients were measured during screw insertion as normal match group. The mean torque of the newly implanted screws was 2.73 ± 0.75 Nm (95% confidence interval 1.23 to 4.22). Since 1.02 Nm was significantly less than the 95% confidence interval of newly implanted screws, the pedicle screw loosening based on torque was defined as screws with an extraction torque less than or equal to 1.02 Nm. According to this criterion, there were 74 (33%) loosening screws.
Table 1
Patients’ characteristics
Patient characteristics
Overall (n = 33)
Age (years)
38.2 (SD 13.8) (17 to 66)
Gender
 Females
10 (28%)
 Males
23 (72%)
Height (cm)
163.6 (SD 7.2)
Weight (kg)
61.9 (SD 14.3)
BMI (kg/m2)
22.9 (SD 4.0)
Diagnosis
 Facture
16
 Degenerative disc diseases
9
 Scoliosis
8
Number of patient with X-ray
33
Number of patient with CT scan
28
BMD (g/cm2)
0.98 (SD 0.13)
Osteoporosis
1
Persisting pain/foreign body sensation
18
 Recovery after removal surgery
16
Time to implant removal (months)
36.4 (18 to 77)
Surgical duration (min)
125 (SD 90)
Fusion in primary surgery
19
Hospital stay (days)
8.4 (SD 3.3)
SD standard deviation, BMI body mass index, BMD bone mineral density
Table 2
Detail of removed pedicle screws
Screw characteristics
Overall (n = 236)
Valid torque data
226 (95.8%)
Screw breakage
4 (1.7%)
Undetectable
6 (2.5%)
Thoracic spine
86 (36.4%)
 t2-t9
38 (16.1%)
 t10
6 (2.5%)
 t11
10 (4.2%)
 t12
32 (13.6%)
Lumbar spine
138 (58.5%)
 l1
33 (14.0%)
 l2
38 (16.1%)
 l3
19 (8.1%)
 l4
24 (10.2%)
 l5
24 (10.2%)
Sacral(S1)
12 (5.1%)
Polyaxial screw
130 (55.1%)
Monoaxial screw
106 (44.9%)
Short segment
36 (9 patients)
Multiple segments(≥ 3)
200 (24 patients)
Using X-ray radiographs taken the day before removal operations, radiolucent area (wider than 1 mm) around the screw and/or the double halo were detected in 20 screws. Using CT scan, the no density zones were found surrounding 17 screws. As shown in Tables 3 and 4, the X-ray criteria of loosening had a sensitivity of 24% and a specificity of 98%, while the CT scan criterion of loosening had a sensitivity of 22% and a specificity of 96%. There was no significant difference in sensitivity (p = 0.863) or specificity (p = 1.00) between X-ray and CT scan. The mean torque of screws, which were diagnosed as loosening by X-ray, was 0.53 ± 0.65 Nm, which was significantly lower (p < 0.0001) than others (1.65 ± 0.98). While the mean torque of loosened screws diagnosed by CT was 1.03 ± 1.00 Nm, which was also significantly lower (p = 0.008) than others (1.71 ± 1.00 Nm).
Table 3
Fourfold data of X-ray criteria indicated X-ray had a sensitivity of 24% and a specificity of 98%
X-ray
Loose
Not loose
p value
Positive
17 (24%)
3 (2%)
< 0.001a
Negative
53 (76%)
153 (98%)
 
Positive—radiolucent area (thicker than 1 mm) and/or the double halo around the screw
aChi-square test
Table 4
Fourfold data of CT criteria indicated CT scan had a sensitivity of 22% and a specificity of 96%
CT scanning
Loose
Not loose
p value
Positive
12 (22%)
5 (4%)
0.001a
Negative
43 (78%)
107 (96%)
 
Positive—no density zones around screw
aChi-square test
The risk factors of screw loosening were analyzed. Regarding extraction torque, there was no significant difference between polyaxial and monoaxial screws (p = 0.673) as shown in Fig. 3a. Significant difference was found between different segments (p < 0.001) as shown in Fig. 3b. The mean torque of screws placed in the lumbosacral junction (L4, L5, and S1, n = 54) was 2.14 ± 1.12 Nm, which was significantly higher than those placed in other segments. There were 26 screws placed in 13 fractured vertebrae. The mean torque of screws placed in fractured vertebrae was 1.03 ± 0.63 Nm, which was significantly lower (p = 0.009) than those in non-fractured vertebrae, as shown in Fig. 3c. The screws at the ends of implant were supposed to share more strain and be vulnerable to loosening [3]. However, in 24 patients who underwent multilevel (more than two segments) instrumentation, there was no significant difference regarding extraction torque (p = 0.437) between screws placed in the end segments (n = 88) and those in the mid segments (n = 103), as shown in Fig. 3d. Patients with fusion surgery were considered to get better stability than those without fusion; hence, screws in fused spine might share lower strain and have higher torque. As expected, extraction torques of pedicle screw after fusion surgery (n = 142) were significantly higher than those in non-fusion surgery (n = 84) (p = 0.001), as shown in Fig. 3e. There was no significant difference regarding the diameter or length of screws as shown in Fig. 3f, g.
Linear regression analysis was used to estimate the relevance between the extraction torque of pedicle screw and other clinical findings. Our results showed a low but significant linear correlation between extraction torque and BMD (p = 0.010, R2 = 0.304, F = 8.296) as well as age (p = 0.045, R2 = 0.123, F = 4.345), indicating pedicle screw in aged patients or those with low bone density may be less stable according to mechanical measurement, as shown in Figs. 4 and 5. There was no linear correlation between extraction torque and patients’ height (p = 0.848), weight (p = 0.196), BMI (p = 0.125), and time interval between the primary surgery and screw removal surgery (p = 0.965).

Discussion

Pedicle screw loosening has been widely reported as one of the concerning complications after spine instrumentation surgery, which may require revision surgery [21]. But the reported data regarding screw loosening were traditionally based on radiological observation, which could be subjective and lead to a considerable variation. Our study reported a 33% loosening rate according to mechanical measurement of extraction torque during instrumentation removal surgery. Meanwhile, we compared our extraction torque data with X-ray and CT scan findings, the result showed that the specificity of both imaging approaches were excellent, and the peri-screw osteolysis present both in X-Ray and CT scan could indicate low extraction torque of screw anchor, but their sensitivities were less than satisfactory (24% and 22% respectively), regarding detection of screw loosening. Furthermore, to our knowledge, for the first time we used extraction torque during screw removal surgery to analyze risk factors for screw loosening, our findings demonstrated that screws in non-fusion spine and fractured vertebrae had significantly lower extraction torque, while BMD and age showed low, but significant linear correlation with extraction torque.
A number of factors have been reported to be related to screw loosening. Excessive strain between the screw and bone interface is considered to be the primary cause for screw loosening [12, 22], which could be deteriorated when fusion is failure or the anterior support is inadequate. Meanwhile, stress shielding can lead to a decrease of stress transferred through the bone tissue, which can reduce bone mineral density and remodel the bone surrounding the screw. The presence of wear debris [23] was reported to induce osteolysis leading to screw loosening. The debris elicited an inflammatory cytokine-mediated particulate-induced response through increased expression of intracellular TNF-alpha, increased osteoclastic activity, and cellular apoptosis. Other factors that could cause bone loss or destruction, such as infection surrounding the implant, bone tumor, metabolic diseases, and microfracture due to excessive loading, are risk factors of pedicle screw loosening. Screw loosening may become a worsening problem due to the aging of population and the increasing number of osteopenic and osteoporotic patients. Wu et al. [17] reported higher occurrence of screw loosening in osteoporotic bone. In our study, we also found a significant linear correlation between BMD and extraction torque, indicating pedicle screws in aged patients or patients with lower BMD might be less stable due to lower extraction torque.
Our findings showed that X-ray had a sensitivity of 24% and a specificity of 98%, while CT scan had a sensitivity of 22% and a specificity of 95% regarding extraction torque as criterion of screw loosening. This result indicated that both radiological examinations were effective to confirm loosening screws; however, the low sensitivity implied a considerable number of loosening screws could be neglected by imaging study. Based on the torque data, the loosening rate was 33%, while both X-ray and CT scan only detected less than 30% of all loosening screws. Sanden [20] reported a 64% sensitivity of X-ray in 79 screws and a 35% rate of screw loosening, but their definition of a loosed screw was an extraction torque of 0.4 Nm or less as there were no screws with and extraction torque between 0.4 Nm and 0.75 Nm. We could not detect such a clear gap in our data. However, by using Twostep Cluster Analysis, we found the distribution of extraction torque data could be clustered into 3 clusters, with the cut-off value of 1.02 Nm and 2.22 Nm respectively. We also tested the average torque of newly implanted screws and found that 1.02 Nm was lower than the low limit of 95% confidence interval of newly implanted screws. Therefore, we set our cut-off torque at 1.02 Ncm for screw loosening. Although the torque values for screw loosening were different, the loosening rate of Sanden’s study based on extraction torque was similar to our findings, around 30–35%.
Ohtori employed both CT scan and X-ray and the results showed that CT scan was more sensitive than X-ray [11]. In our study, there was no significant difference between X-ray and CT scan regarding both sensitivity (p = 0.863) and specificity (p = 1.00). It was observed in one case that CT scan failed to show a clear gap around screw, even though an obvious double halo was found on X-ray and the extraction torque was 0.06 Nm (Fig. 6). This could be resulted from the metal artifact that seriously interfered CT reconstruction and led to incorrect image surrounded metal instrumentation. Therefore, based on our result, CT might not be superior to X-ray in assessment of screw loosening, especially considering its higher cost and radiation exposure.
The risk factors of screw loosening we found based on extraction torque were fixation in fractured vertebrae and non-fusion spine. In the fractured vertebrae, the continuity of cortical and structure of trabecula were damaged, which might affect the stability of screw placed in it. Meanwhile, the strain imposed on pedicle screws might significantly increase when fusion had not been obtained or anterior column support was inadequate. We also found that pedicle screws in the lumbosacral junction (L4, L5, and S1) had significant higher extraction torque than those in other segments. A possible explanation could be that surgeries performed in lumbosacral spine were mainly for degenerative disc diseases, which often required spinal fusion, while in the thoracolumbar spine, where operations were more likely for vertebral fracture, fusion were not always necessary.
As expected, linear regression analysis showed that the stability of pedicle screw correlated positively with BMD and negatively with patients’ age. The relatively low related coefficients might be due to the existence of numerous confounding factors, and these results indicated that the failure risk of instrumentation increased with age and osteopenia/osteoporosis, which was widely credited but rarely proven in vivo with mechanical measurement.
Some reports showed that increasing length and diameter could increase the stability of pedicle screw [24, 25], which had not been observed from our results. The reasons that no significant different extraction torque was found among screws with different length and diameter might be due to the relatively small sample size and narrow range of length (30–55 mm) and diameter (4.0–6.5 mm). Further study with larger simple size is needed to confirm the effect of screw length and diameter on extraction torque in the human spine. Meanwhile, there was no significant difference in extraction torque regarding the screw design (polyaxial vs. monoaxial) and location (placed at the end segment vs. at the middle segment).
There are several limitations in our study. First, although this was a prospective and blinded designed study, the relatively small number of patients’ enrollment and the heterogeneity of screw size and position might render our findings susceptible to confounding factors. Hence, further studies with larger sample size and stratified data according to different factors will be needed to better understand the prevalent of pedicle screw loosening. Second, the lack of pedicle torque during primary surgeries made it unable to observe the longitudinal change of torque and the influence of inserting torque on screw loosening.

Conclusion

In general, we found a 33% pedicle screw showed an extraction torque less than 1.02 Nm, which might be considered to be loosening according to our data distribution. Both X-ray and CT had high specificity to detect screw loosening, but their sensitivities could be overestimated. Surgeons need to be more cautious when assessing screw loosening based on radiological examination, since a considerable fraction of low extraction torque screws might have been underestimated. Pedicle screws of non-fusion surgery placed in fractured vertebrae had significantly lower extraction torque and, therefore, could be vulnerable to loosening. Pedicle screws in aged patients or patients with lower BMD might be less stable due to lower extraction torque.

Acknowledgements

The authors wish to acknowledge Dongbin Qu, MD; Jixing Wang, MD; Jianming Jiang, MD; Haiming Wang, MD; Hailong Ren, MD; Minghui Zheng; Zhaoming Zhong, MD; Bo Xu, MD; and Dehong Yang, MD, for their kind help in this study.

Funding

No funding was received for the current study.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.
This study was a prospective and clinical study and was approved by the Nanfang Hospital, Southern Medical University Institutional Review Board. Consent to participate was obtained from the participants.
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Esses SI, Sachs BL, Dreyzin V. Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members. Spine. 1993;18(15):2231–8 discussion 2238-2239.CrossRef Esses SI, Sachs BL, Dreyzin V. Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members. Spine. 1993;18(15):2231–8 discussion 2238-2239.CrossRef
2.
Zurück zum Zitat McAfee PC, Weiland DJ, Carlow JJ. Survivorship analysis of pedicle spinal instrumentation. Spine. 1991;16(8 Suppl):S422–7.PubMed McAfee PC, Weiland DJ, Carlow JJ. Survivorship analysis of pedicle spinal instrumentation. Spine. 1991;16(8 Suppl):S422–7.PubMed
3.
Zurück zum Zitat Galbusera F, Volkheimer D, Reitmaier S, Berger-Roscher N, Kienle A, Wilke HJ. Pedicle screw loosening: a clinically relevant complication? Eur Spine J. 2015;24(5):1005–16.CrossRef Galbusera F, Volkheimer D, Reitmaier S, Berger-Roscher N, Kienle A, Wilke HJ. Pedicle screw loosening: a clinically relevant complication? Eur Spine J. 2015;24(5):1005–16.CrossRef
4.
Zurück zum Zitat Ko CC, Tsai HW, Huang WC, Wu JC, Chen YC, Shih YH, Chen HC, Wu CL, Cheng H. Screw loosening in the Dynesys stabilization system: radiographic evidence and effect on outcomes. Neurosurg Focus. 2010;28(6):E10.CrossRef Ko CC, Tsai HW, Huang WC, Wu JC, Chen YC, Shih YH, Chen HC, Wu CL, Cheng H. Screw loosening in the Dynesys stabilization system: radiographic evidence and effect on outcomes. Neurosurg Focus. 2010;28(6):E10.CrossRef
5.
Zurück zum Zitat Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Posterior lumbar interbody fusion: a retrospective study of complications after facet joint excision and pedicle screw fixation in 148 cases. Acta Orthop Scand. 1999;70(4):329–34.CrossRef Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Posterior lumbar interbody fusion: a retrospective study of complications after facet joint excision and pedicle screw fixation in 148 cases. Acta Orthop Scand. 1999;70(4):329–34.CrossRef
6.
Zurück zum Zitat Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Can insertional torque predict screw loosening and related failures? An in vivo study of pedicle screw fixation augmenting posterior lumbar interbody fusion. Spine. 2000;25(7):858–64.CrossRef Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Can insertional torque predict screw loosening and related failures? An in vivo study of pedicle screw fixation augmenting posterior lumbar interbody fusion. Spine. 2000;25(7):858–64.CrossRef
7.
Zurück zum Zitat Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Influence of bone mineral density on pedicle screw fixation: a study of pedicle screw fixation augmenting posterior lumbar interbody fusion in elderly patients. Spine J. 2001;1(6):402–7.CrossRef Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Influence of bone mineral density on pedicle screw fixation: a study of pedicle screw fixation augmenting posterior lumbar interbody fusion in elderly patients. Spine J. 2001;1(6):402–7.CrossRef
8.
Zurück zum Zitat Ozawa T, Takahashi K, Yamagata M, Ohtori S, Aoki Y, Saito T, Inoue G, Ito T, Moriya H. Insertional torque of the lumbar pedicle screw during surgery. J Orthop Sci. 2005;10(2):133–6.CrossRef Ozawa T, Takahashi K, Yamagata M, Ohtori S, Aoki Y, Saito T, Inoue G, Ito T, Moriya H. Insertional torque of the lumbar pedicle screw during surgery. J Orthop Sci. 2005;10(2):133–6.CrossRef
9.
Zurück zum Zitat Wu JC, Huang WC, Tsai HW, Ko CC, Wu CL, Tu TH, Cheng H. Pedicle screw loosening in dynamic stabilization: incidence, risk, and outcome in 126 patients. Neurosurg Focus. 2011;31(4):E9.CrossRef Wu JC, Huang WC, Tsai HW, Ko CC, Wu CL, Tu TH, Cheng H. Pedicle screw loosening in dynamic stabilization: incidence, risk, and outcome in 126 patients. Neurosurg Focus. 2011;31(4):E9.CrossRef
10.
Zurück zum Zitat Dakhil-Jerew F, Jadeja H, Cohen A, Shepperd JA. Inter-observer reliability of detecting Dynesys pedicle screw using plain X-rays: a study on 50 post-operative patients. Eur Spine J. 2009;18(10):1486–93.CrossRef Dakhil-Jerew F, Jadeja H, Cohen A, Shepperd JA. Inter-observer reliability of detecting Dynesys pedicle screw using plain X-rays: a study on 50 post-operative patients. Eur Spine J. 2009;18(10):1486–93.CrossRef
11.
Zurück zum Zitat Ohtori S, Inoue G, Orita S, Yamauchi K, Eguchi Y, Ochiai N, Kishida S, Kuniyoshi K, Aoki Y, Nakamura J, et al. Comparison of teriparatide and bisphosphonate treatment to reduce pedicle screw loosening after lumbar spinal fusion surgery in postmenopausal women with osteoporosis from a bone quality perspective. Spine. 2013;38(8):E487–92.CrossRef Ohtori S, Inoue G, Orita S, Yamauchi K, Eguchi Y, Ochiai N, Kishida S, Kuniyoshi K, Aoki Y, Nakamura J, et al. Comparison of teriparatide and bisphosphonate treatment to reduce pedicle screw loosening after lumbar spinal fusion surgery in postmenopausal women with osteoporosis from a bone quality perspective. Spine. 2013;38(8):E487–92.CrossRef
12.
Zurück zum Zitat Schizas C, Tzinieris N, Tsiridis E, Kosmopoulos V. Minimally invasive versus open transforaminal lumbar interbody fusion: evaluating initial experience. Int Orthop. 2009;33(6):1683–8.CrossRef Schizas C, Tzinieris N, Tsiridis E, Kosmopoulos V. Minimally invasive versus open transforaminal lumbar interbody fusion: evaluating initial experience. Int Orthop. 2009;33(6):1683–8.CrossRef
13.
Zurück zum Zitat Faraj AA, Webb JK. Early complications of spinal pedicle screw. Eur Spine J. 1997;6(5):324–6.CrossRef Faraj AA, Webb JK. Early complications of spinal pedicle screw. Eur Spine J. 1997;6(5):324–6.CrossRef
14.
Zurück zum Zitat Glaser J, Stanley M, Sayre H, Woody J, Found E, Spratt K. A 10-year follow-up evaluation of lumbar spine fusion with pedicle screw fixation. Spine. 2003;28(13):1390–5.PubMed Glaser J, Stanley M, Sayre H, Woody J, Found E, Spratt K. A 10-year follow-up evaluation of lumbar spine fusion with pedicle screw fixation. Spine. 2003;28(13):1390–5.PubMed
15.
Zurück zum Zitat Pihlajamaki H, Myllynen P, Bostman O. Complications of transpedicular lumbosacral fixation for non-traumatic disorders. J Bone Joint Surg Br. 1997;79(2):183–9.CrossRef Pihlajamaki H, Myllynen P, Bostman O. Complications of transpedicular lumbosacral fixation for non-traumatic disorders. J Bone Joint Surg Br. 1997;79(2):183–9.CrossRef
16.
Zurück zum Zitat Rollinghoff M, Schluter-Brust K, Groos D, Sobottke R, Michael JW, Eysel P, Delank KS. Mid-range outcomes in 64 consecutive cases of multilevel fusion for degenerative diseases of the lumbar spine. Orthop Rev. 2010;2(1):e3.CrossRef Rollinghoff M, Schluter-Brust K, Groos D, Sobottke R, Michael JW, Eysel P, Delank KS. Mid-range outcomes in 64 consecutive cases of multilevel fusion for degenerative diseases of the lumbar spine. Orthop Rev. 2010;2(1):e3.CrossRef
17.
Zurück zum Zitat Wu ZX, Gong FT, Liu L, Ma ZS, Zhang Y, Zhao X, Yang M, Lei W, Sang HX. A comparative study on screw loosening in osteoporotic lumbar spine fusion between expandable and conventional pedicle screws. Arch Orthop Trauma Surg. 2012;132(4):471–6.CrossRef Wu ZX, Gong FT, Liu L, Ma ZS, Zhang Y, Zhao X, Yang M, Lei W, Sang HX. A comparative study on screw loosening in osteoporotic lumbar spine fusion between expandable and conventional pedicle screws. Arch Orthop Trauma Surg. 2012;132(4):471–6.CrossRef
18.
Zurück zum Zitat Kumano K, Hirabayashi S, Ogawa Y, Aota Y. Pedicle screws and bone mineral density. Spine. 1994;19(10):1157–61.CrossRef Kumano K, Hirabayashi S, Ogawa Y, Aota Y. Pedicle screws and bone mineral density. Spine. 1994;19(10):1157–61.CrossRef
19.
Zurück zum Zitat Christensen FB, Dalstra M, Sejling F, Overgaard S, Bunger C. Titanium-alloy enhances bone-pedicle screw fixation: mechanical and histomorphometrical results of titanium-alloy versus stainless steel. Eur Spine J. 2000;9(2):97–103.CrossRef Christensen FB, Dalstra M, Sejling F, Overgaard S, Bunger C. Titanium-alloy enhances bone-pedicle screw fixation: mechanical and histomorphometrical results of titanium-alloy versus stainless steel. Eur Spine J. 2000;9(2):97–103.CrossRef
20.
Zurück zum Zitat Sanden B, Olerud C, Petren-Mallmin M, Johansson C, Larsson S. The significance of radiolucent zones surrounding pedicle screws. Definition of screw loosening in spinal instrumentation. J Bone Joint Surg Br. 2004;86(3):457–61.CrossRef Sanden B, Olerud C, Petren-Mallmin M, Johansson C, Larsson S. The significance of radiolucent zones surrounding pedicle screws. Definition of screw loosening in spinal instrumentation. J Bone Joint Surg Br. 2004;86(3):457–61.CrossRef
21.
Zurück zum Zitat Berjano P, Bassani R, Casero G, Sinigaglia A, Cecchinato R, Lamartina C. Failures and revisions in surgery for sagittal imbalance: analysis of factors influencing failure. Eur Spine J. 2013;22(Suppl 6):S853–8.CrossRef Berjano P, Bassani R, Casero G, Sinigaglia A, Cecchinato R, Lamartina C. Failures and revisions in surgery for sagittal imbalance: analysis of factors influencing failure. Eur Spine J. 2013;22(Suppl 6):S853–8.CrossRef
22.
Zurück zum Zitat Villa T, La Barbera L, Galbusera F. Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems. Spine J. 2014;14(4):695–704.CrossRef Villa T, La Barbera L, Galbusera F. Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems. Spine J. 2014;14(4):695–704.CrossRef
23.
Zurück zum Zitat Botolin S, Merritt C, Erickson M. Aseptic loosening of pedicle screw as a result of metal wear debris in a pediatric patient. Spine. 2013;38(1):E38–42.CrossRef Botolin S, Merritt C, Erickson M. Aseptic loosening of pedicle screw as a result of metal wear debris in a pediatric patient. Spine. 2013;38(1):E38–42.CrossRef
24.
Zurück zum Zitat Brantley AG, Mayfield JK, Koeneman JB, Clark KR. The effects of pedicle screw fit. An in vitro study. Spine. 1994;19(15):1752–8.CrossRef Brantley AG, Mayfield JK, Koeneman JB, Clark KR. The effects of pedicle screw fit. An in vitro study. Spine. 1994;19(15):1752–8.CrossRef
25.
Zurück zum Zitat Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims--what is best? Spine. 1998;23(12):1374–9.CrossRef Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims--what is best? Spine. 1998;23(12):1374–9.CrossRef
Metadaten
Titel
Pedicle screw loosening: the value of radiological imagings and the identification of risk factors assessed by extraction torque during screw removal surgery
verfasst von
Xiaoliang Wu
Jiawei Shi
Jinyan Wu
Yongquan Cheng
Kaiwen Peng
Jianting Chen
Hui Jiang
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
Journal of Orthopaedic Surgery and Research / Ausgabe 1/2019
Elektronische ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-018-1046-0