Skip to main content
Erschienen in: Journal of Clinical Monitoring and Computing 6/2016

19.09.2015 | Original Research

Performance of computer simulated inhalational anesthetic uptake model in comparison with real time isoflurane concentration

verfasst von: Umeshkumar Athiraman, M. Ravishankar, Sameer Jahagirdhar

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 6/2016

Einloggen, um Zugang zu erhalten

Abstract

Gas Man software was developed to enhance our understanding of the pharmacokinetics of inhalational anaesthetics. To date the Gas Man software has not been validated in humans. In this study we compared the Gas Man software with real time measured end tidal isoflurane concentrations while using a semi closed circle system in anesthetised patients. Thirty-four ASA I and II patients 18–60 years of age were selected for the study. After a standardized induction anesthesia was maintained with N2O + O2 mixture and isoflurane using the circle system. The fresh gas flow or dial setting of Isoflurane vaporizer were changed at random. The inspired and end-tidal concentration values of isoflurane measured at 1 min intervals were retrieved from the patient monitor. Real time anesthetic settings for the patient were simultaneously simulated in the Gas Man software to generate the inspired and end-tidal concentration of isoflurane values at every minute for comparison. Varvel’s criteria have been used to assess this model. The median absolute performance error was 9.39 %, median performance error was −5.30 %, wobble was 5.16 %, and divergence was −1.82 %. All criteria were within limits of the acceptable performance of the model. The end-tidal concentration values of isoflurane in real patients are very close to those predicted by Gas Man software. The pharmocokinetics of inhalational anesthetic administration in patients can be taught accurately using Gas Man technology. This technology may also help in developing different kinetic models of inhalational agents in the body.
Fußnoten
Literatur
1.
Zurück zum Zitat Tanner G. Pharmacokinetics of inhalation anesthetics: a three-compartment linear model. Anesth Analg. 1982;61:587–94.CrossRefPubMed Tanner G. Pharmacokinetics of inhalation anesthetics: a three-compartment linear model. Anesth Analg. 1982;61:587–94.CrossRefPubMed
2.
Zurück zum Zitat Kety SS. The physiological and physical factors governing the uptake of anesthetic gases by the body. Anesthesiology. 1950;11:517–26.CrossRefPubMed Kety SS. The physiological and physical factors governing the uptake of anesthetic gases by the body. Anesthesiology. 1950;11:517–26.CrossRefPubMed
3.
Zurück zum Zitat Heffernan PB, Gibbs JM, McKinnon AE. Teaching the uptake and distribution of halothane. A computer simulation program. Anaesthesia. 1982;37:9–17.CrossRefPubMed Heffernan PB, Gibbs JM, McKinnon AE. Teaching the uptake and distribution of halothane. A computer simulation program. Anaesthesia. 1982;37:9–17.CrossRefPubMed
4.
Zurück zum Zitat Munson ES, Egar EI II, Bowers DL. Effects of anesthetic-depressed ventilation and cardiac output on anesthetic uptake: a computer nonlinear simulation. Anesthesiology. 1973;38:251–9.CrossRefPubMed Munson ES, Egar EI II, Bowers DL. Effects of anesthetic-depressed ventilation and cardiac output on anesthetic uptake: a computer nonlinear simulation. Anesthesiology. 1973;38:251–9.CrossRefPubMed
5.
Zurück zum Zitat Paskin S, Raemer DB, Garfield JM, Philip JH. Is computer simulation of anesthetic uptake and distribution an effective teaching tool for anesthesia residents. J Clin Monit. 1985;1:87–8. Paskin S, Raemer DB, Garfield JM, Philip JH. Is computer simulation of anesthetic uptake and distribution an effective teaching tool for anesthesia residents. J Clin Monit. 1985;1:87–8.
6.
Zurück zum Zitat Philip JH, Lema MJ, Raemer DB, Crocker D. Is computer simulation as effective as lecture for teaching residents anesthetic uptake and distribution? Anesthesiology. 1985;63:A503.CrossRef Philip JH, Lema MJ, Raemer DB, Crocker D. Is computer simulation as effective as lecture for teaching residents anesthetic uptake and distribution? Anesthesiology. 1985;63:A503.CrossRef
7.
Zurück zum Zitat Philip JH. Gas Man®—An example of goal oriented computer-assisted teaching which results in learning. Int J Clin Monit. 1986;3:165–73.CrossRef Philip JH. Gas Man®—An example of goal oriented computer-assisted teaching which results in learning. Int J Clin Monit. 1986;3:165–73.CrossRef
8.
Zurück zum Zitat Garfield JM, Paskin S, Philip JH. An evaluation of the effectiveness of a computer simulation of anaesthetic uptake and distribution as a teaching tool. Med Educ. 1989;23:457–62.CrossRefPubMed Garfield JM, Paskin S, Philip JH. An evaluation of the effectiveness of a computer simulation of anaesthetic uptake and distribution as a teaching tool. Med Educ. 1989;23:457–62.CrossRefPubMed
9.
Zurück zum Zitat Gage JS. Mathematics and judgment: a review of the computer simulation program GAS MAN. MD Comput. 1992;9:54–6.PubMed Gage JS. Mathematics and judgment: a review of the computer simulation program GAS MAN. MD Comput. 1992;9:54–6.PubMed
10.
Zurück zum Zitat Lerou JG, Dirksen R, Beneken KHH, Booij LH. A system model for closed-circuit inhalation anesthesia. I. Computer study. Anesthesiology. 1991;75:345–55.CrossRefPubMed Lerou JG, Dirksen R, Beneken KHH, Booij LH. A system model for closed-circuit inhalation anesthesia. I. Computer study. Anesthesiology. 1991;75:345–55.CrossRefPubMed
11.
Zurück zum Zitat Soong TT. Pharmacokinetics with uncertainties in the rate constants.s.I. Math Biosci. 1971;12:235–43.CrossRef Soong TT. Pharmacokinetics with uncertainties in the rate constants.s.I. Math Biosci. 1971;12:235–43.CrossRef
12.
Zurück zum Zitat Soong TT. Random differential equations in science and engineering: mathematics in science and engineering, vol. 103. New York: Academic Press; 1973. Soong TT. Random differential equations in science and engineering: mathematics in science and engineering, vol. 103. New York: Academic Press; 1973.
13.
Zurück zum Zitat Varvel J, Donoho D, Shafer S. Measuring the predictive performance of computer-controlled infusion pumps. J Pharmacokinet Biopharm. 1992;20:63–94.CrossRefPubMed Varvel J, Donoho D, Shafer S. Measuring the predictive performance of computer-controlled infusion pumps. J Pharmacokinet Biopharm. 1992;20:63–94.CrossRefPubMed
14.
Zurück zum Zitat Swinhoe CF, Peacock JE, Glen JB, Reilly CS. Evaluation of the predictive performance of a’Diprifusor’ TCI system. Anaesthesia. 1998;53:61–7.CrossRefPubMed Swinhoe CF, Peacock JE, Glen JB, Reilly CS. Evaluation of the predictive performance of a’Diprifusor’ TCI system. Anaesthesia. 1998;53:61–7.CrossRefPubMed
15.
Zurück zum Zitat Schuttler J, Kloos S, Schwilden H, Stoeckel H. Total intravenous anaesthesia with propofol and alfentanil by computer assisted infusion. Anaesthesia. 1988;43(Supplement):2–7.CrossRefPubMed Schuttler J, Kloos S, Schwilden H, Stoeckel H. Total intravenous anaesthesia with propofol and alfentanil by computer assisted infusion. Anaesthesia. 1988;43(Supplement):2–7.CrossRefPubMed
16.
Zurück zum Zitat Glass PJA, Jacobs JR, Reeves JG. Intravenous drug delivery. In: Milder RD, editor. Anaesthesia. 3rd ed. New York: Churchill Livingstone; 1990. p. 367–88. Glass PJA, Jacobs JR, Reeves JG. Intravenous drug delivery. In: Milder RD, editor. Anaesthesia. 3rd ed. New York: Churchill Livingstone; 1990. p. 367–88.
17.
Zurück zum Zitat Short TG, Lim TA, Tam YH. Prospective evaluation of pharmacokinetic model-controlled infusion of propofol in adult patients. Br J Anaesth. 1996;6:313–5.CrossRef Short TG, Lim TA, Tam YH. Prospective evaluation of pharmacokinetic model-controlled infusion of propofol in adult patients. Br J Anaesth. 1996;6:313–5.CrossRef
18.
Zurück zum Zitat Pandin PC, Cantraine F, Ewalenko P, et al. Predictive accuracy of target-controlled propofol and sufentanil coinfusion in longlasting surgery. Anesthesiology. 2000;93:653–61.CrossRefPubMed Pandin PC, Cantraine F, Ewalenko P, et al. Predictive accuracy of target-controlled propofol and sufentanil coinfusion in longlasting surgery. Anesthesiology. 2000;93:653–61.CrossRefPubMed
19.
Zurück zum Zitat Kennedy RR, French RA, Spencer C. Predictive accuracy of a model of volatile anesthetic uptake. Anesth Analg. 2002;95:1616–21.CrossRefPubMed Kennedy RR, French RA, Spencer C. Predictive accuracy of a model of volatile anesthetic uptake. Anesth Analg. 2002;95:1616–21.CrossRefPubMed
20.
Zurück zum Zitat Kennedy RR. Predictive performance of a model of anaesthetic uptake with desflurane. Anaesth Intensive Care. 2006;34:197–202.PubMed Kennedy RR. Predictive performance of a model of anaesthetic uptake with desflurane. Anaesth Intensive Care. 2006;34:197–202.PubMed
21.
Zurück zum Zitat Heffernan PB, Gibbs JM, McKinnon AE. Evaluation of a computer simulation program for teaching halothane uptake and distribution. Anaesthesia. 1982;37:43–6.CrossRefPubMed Heffernan PB, Gibbs JM, McKinnon AE. Evaluation of a computer simulation program for teaching halothane uptake and distribution. Anaesthesia. 1982;37:43–6.CrossRefPubMed
Metadaten
Titel
Performance of computer simulated inhalational anesthetic uptake model in comparison with real time isoflurane concentration
verfasst von
Umeshkumar Athiraman
M. Ravishankar
Sameer Jahagirdhar
Publikationsdatum
19.09.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 6/2016
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9776-6

Weitere Artikel der Ausgabe 6/2016

Journal of Clinical Monitoring and Computing 6/2016 Zur Ausgabe

Update AINS