Erschienen in: European Journal of Trauma and Emergency Surgery 5/2020

31.10.2019 | Original Article

PFNA and DHS for AO/OTA 31-A2 fractures: radiographic measurements, morbidity and mortality

verfasst von: Franz Müller, Matthias Doblinger, Tanja Kottmann, Bernd Füchtmeier

Erschienen in: European Journal of Trauma and Emergency Surgery | Ausgabe 5/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

For this retrospective cohort study, we assessed pertrochanteric fracture types AO/OTA 31-A2. PFNA and DHS were the devices used. We determined both devices in relation to peri-operative variables, postoperative radiographic measurements, implant-related complications and mortality up to 2 years. The null hypothesis was no effect between the two devices.

Methods

This single-centre study was conducted based on our computerized data. The treatment period ranged from 2006 to 2015. Only patients with type AO/OTA 31-A2 fractures and an age ≥ 65 years were included. Apart from descriptive variables, the following measurements were assessed: (1) duration of surgery, (2) blood loss, (3) transfusion, (4) hospitalization, (5) tip-apex distance (TAD), (6) fracture reduction, (7) screw position, (8) implant-related complications, and (9) mortality. The follow-up was 2 years for each living patient. Missing data were evaluated by telephone call.

Results

A total of 375 consecutive patients were enrolled into three groups: (1) 75 patients treated with DHS and antirotation screw (ARS); (2); 100 patients treated with DHS + ARS + TSP (trochanteric stabilization plate); and (3) 200 patients treated with PFNA. Apart from dementia, the descriptive data (e.g., age and BMI) demonstrated no effects between the three groups. Compared to PFNA, DHS with or without TSP was adversely affected by a longer operation time, higher blood loss, increase in transfusion, and more implant-related complications including cut-out, infection and failure. The rate of cut-out was significantly higher in TAD ≥ 25 mm (p = 0.005), and PFNA demonstrated significantly better TAD measurements (p = 0.001), better fracture reduction (0.002), more central-central screw positions (p = 0.014), and less poor screw placement (p = 0.001). The mortality rate was without effect between the three groups (log rank 0.698).

Conclusions

DHS with or without TSP was associated with significantly higher rates of implant-related complications based on inferior radiographic measurements. Therefore, we only recommend PFNA for the treatment of proximal type AO/OTA 31-A2 femoral fractures.

Level of evidence

Therapeutic level III.
Literatur
1.
Zurück zum Zitat Marsh JL, Slongo TF, Agel J, et al. Fracture and dislocation classification compendium—2007: Orthopaedic Trauma Association Classification, Database and Outcomes Committee. J Orthop Trauma. 2007;21(Supplement 10):S1–S163.CrossRefPubMed Marsh JL, Slongo TF, Agel J, et al. Fracture and dislocation classification compendium—2007: Orthopaedic Trauma Association Classification, Database and Outcomes Committee. J Orthop Trauma. 2007;21(Supplement 10):S1–S163.CrossRefPubMed
2.
Zurück zum Zitat Bhandari M, Schemitsch E, Jönsson A, et al. Gamma nails revisited: gamma nails versus compression hip screws in the management of intertrochanteric fractures of the hip: a meta-analysis. J Orthop Trauma. 2009;23:460–4.CrossRefPubMed Bhandari M, Schemitsch E, Jönsson A, et al. Gamma nails revisited: gamma nails versus compression hip screws in the management of intertrochanteric fractures of the hip: a meta-analysis. J Orthop Trauma. 2009;23:460–4.CrossRefPubMed
3.
Zurück zum Zitat Shen L, Zhang Y, Shen Y, et al. Antirotation proximal femoral nail versus dynamic hip screw for intertrochanteric fractures: a meta-analysis of randomized controlled studies. Orthop Traumatol Surg Res. 2013;99:377–83.CrossRefPubMed Shen L, Zhang Y, Shen Y, et al. Antirotation proximal femoral nail versus dynamic hip screw for intertrochanteric fractures: a meta-analysis of randomized controlled studies. Orthop Traumatol Surg Res. 2013;99:377–83.CrossRefPubMed
4.
Zurück zum Zitat Tucker A, Donnelly KJ, Rowan C, et al. Is the best plate a nail? A review of 3230 unstable intertrochanteric fractures of the proximal femur. J Orthop Trauma. 2018;32:53–60.CrossRefPubMed Tucker A, Donnelly KJ, Rowan C, et al. Is the best plate a nail? A review of 3230 unstable intertrochanteric fractures of the proximal femur. J Orthop Trauma. 2018;32:53–60.CrossRefPubMed
5.
Zurück zum Zitat Barton TM, Gleeson R, Topliss C, et al. A comparison of the long nail with the sliding hip screw for the treatment of AO/OTA 31–A2 fractures of the proximal part of the femur: a prospective randomized trial. J Bone Jt Surg Am. 2010;92:792–8.CrossRef Barton TM, Gleeson R, Topliss C, et al. A comparison of the long nail with the sliding hip screw for the treatment of AO/OTA 31–A2 fractures of the proximal part of the femur: a prospective randomized trial. J Bone Jt Surg Am. 2010;92:792–8.CrossRef
6.
Zurück zum Zitat Jonnes C, Sm S, Najimudeen S. Type II intertrochanteric fractures: proximal femoral nailing (PFN) versus dynamic hip screw (DHS). Arch Bone Jt Surg. 2016;4:23–8.PubMedPubMedCentral Jonnes C, Sm S, Najimudeen S. Type II intertrochanteric fractures: proximal femoral nailing (PFN) versus dynamic hip screw (DHS). Arch Bone Jt Surg. 2016;4:23–8.PubMedPubMedCentral
7.
Zurück zum Zitat Zehir S, Zehir R, Zehir S, et al. Proximal femoral nail antirotation against dynamic hip screw for unstable trochanteric fractures; a prospective randomized comparison. Eur J Trauma Emerg Surg. 2015;41:393–400.CrossRefPubMed Zehir S, Zehir R, Zehir S, et al. Proximal femoral nail antirotation against dynamic hip screw for unstable trochanteric fractures; a prospective randomized comparison. Eur J Trauma Emerg Surg. 2015;41:393–400.CrossRefPubMed
8.
Zurück zum Zitat Zhang K, Zhang S, Yang J, et al. Proximal femoral nail vs. dynamic hip screw in treatment of intertrochanteric fractures: a meta-analysis. Med Sci Monit. 2014;12:1628–33. Zhang K, Zhang S, Yang J, et al. Proximal femoral nail vs. dynamic hip screw in treatment of intertrochanteric fractures: a meta-analysis. Med Sci Monit. 2014;12:1628–33.
9.
Zurück zum Zitat Ma KL, Wang X, Luan FJ, et al. Proximal femoral nails antirotation, Gamma nails, and dynamic hip screws for fixation of intertrochanteric fractures of femur: a meta-analysis. Orthop Traumatol Surg Res. 2014;100:859–66.CrossRefPubMed Ma KL, Wang X, Luan FJ, et al. Proximal femoral nails antirotation, Gamma nails, and dynamic hip screws for fixation of intertrochanteric fractures of femur: a meta-analysis. Orthop Traumatol Surg Res. 2014;100:859–66.CrossRefPubMed
10.
Zurück zum Zitat Zhu Q, Xu X, Yang X, et al. Intramedullary nails versus sliding hip screws for AO/OTA 31-A2 trochanteric fractures in adults: a meta-analysis. Int J Surg. 2017;43:67–74.CrossRefPubMed Zhu Q, Xu X, Yang X, et al. Intramedullary nails versus sliding hip screws for AO/OTA 31-A2 trochanteric fractures in adults: a meta-analysis. Int J Surg. 2017;43:67–74.CrossRefPubMed
11.
Zurück zum Zitat Liu Y, Tao R, Liu F, et al. Mid-term outcomes after intramedullary fixation of peritrochanteric femoral fractures using the new proximal femoral nail antirotation (PFNA). Injury. 2010;41:810–7.CrossRefPubMed Liu Y, Tao R, Liu F, et al. Mid-term outcomes after intramedullary fixation of peritrochanteric femoral fractures using the new proximal femoral nail antirotation (PFNA). Injury. 2010;41:810–7.CrossRefPubMed
12.
Zurück zum Zitat Boldin C, Seibert FJ, Fankhauser F, et al. The proximal femoral nail (PFN) - a minimal invasive treatment of unstable proximal femoral fractures: a prospective study of 55 patients with a follow-up of 15 months. Acta Orthop Scand. 2003;74:53–8.CrossRefPubMed Boldin C, Seibert FJ, Fankhauser F, et al. The proximal femoral nail (PFN) - a minimal invasive treatment of unstable proximal femoral fractures: a prospective study of 55 patients with a follow-up of 15 months. Acta Orthop Scand. 2003;74:53–8.CrossRefPubMed
13.
Zurück zum Zitat Kumar R, Singh RN, Singh BN. Comparative prospective study of proximal femoral nail and dynamic hip screw in treatment of intertrochanteric fracture femur. J Clin Orthop Trauma. 2012;3:28–36.CrossRefPubMedPubMedCentral Kumar R, Singh RN, Singh BN. Comparative prospective study of proximal femoral nail and dynamic hip screw in treatment of intertrochanteric fracture femur. J Clin Orthop Trauma. 2012;3:28–36.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Saudan M, Lübbeke A, Sadowski C, et al. Pertrochanteric fractures: is there an advantage to an intramedullary nail? A randomized, prospective study of 206 patients comparing the dynamic hip screw and proximal femoral nail. J Orthop Trauma. 2002;16:386–93.CrossRefPubMed Saudan M, Lübbeke A, Sadowski C, et al. Pertrochanteric fractures: is there an advantage to an intramedullary nail? A randomized, prospective study of 206 patients comparing the dynamic hip screw and proximal femoral nail. J Orthop Trauma. 2002;16:386–93.CrossRefPubMed
15.
Zurück zum Zitat Reindl R, Harvey EJ, Berry GK, Canadian Orthopaedic Trauma Society (COTS), et al. Intramedullary verus extramedullary fixation for unstable intertrochenteric fractures: a prospective randomized controlled trial. J Bone Jt Surg Am. 2015;97:1905–12.CrossRef Reindl R, Harvey EJ, Berry GK, Canadian Orthopaedic Trauma Society (COTS), et al. Intramedullary verus extramedullary fixation for unstable intertrochenteric fractures: a prospective randomized controlled trial. J Bone Jt Surg Am. 2015;97:1905–12.CrossRef
16.
Zurück zum Zitat Zou J, Xu Y, Yang H. A comparison of proximal femoral nail antirotation and dynamic hip screw devices in trochanteric fractures. J Int Med Res. 2009;37:1057–64.CrossRefPubMed Zou J, Xu Y, Yang H. A comparison of proximal femoral nail antirotation and dynamic hip screw devices in trochanteric fractures. J Int Med Res. 2009;37:1057–64.CrossRefPubMed
17.
Zurück zum Zitat Simmermacher RK, Ljungqvist J, Bail H, AO-PFNA studygroup, et al. The new proximal femoral nail antirotation (PFNA) in daily practice: results of a multicentre clinical study. Injury. 2008;39:932–9.CrossRefPubMed Simmermacher RK, Ljungqvist J, Bail H, AO-PFNA studygroup, et al. The new proximal femoral nail antirotation (PFNA) in daily practice: results of a multicentre clinical study. Injury. 2008;39:932–9.CrossRefPubMed
18.
Zurück zum Zitat Baumgaertner MR, Curtin SL, Lindskog DM. Intramedullary versus extramedullary fixation for the treatment of intertrochanteric hip fractures. Clin Orthop Relat Res. 1998;348:87–94.CrossRef Baumgaertner MR, Curtin SL, Lindskog DM. Intramedullary versus extramedullary fixation for the treatment of intertrochanteric hip fractures. Clin Orthop Relat Res. 1998;348:87–94.CrossRef
19.
Zurück zum Zitat Cleveland M, Bosworth DM, Thompson FR, et al. A 10-year analysis of intertrochanteric fractures of the femur. J Bone Jt Surg Am. 1959;41:1399–408.CrossRef Cleveland M, Bosworth DM, Thompson FR, et al. A 10-year analysis of intertrochanteric fractures of the femur. J Bone Jt Surg Am. 1959;41:1399–408.CrossRef
20.
Zurück zum Zitat Parvizi J, Zmistowski B, Berbari EF, et al. New definition for periprosthetic joint infection: from the Workgroup of the Musculoskeletal Infection Society. Clin Orthop Relat Res. 2011;469:2992–4.CrossRefPubMedPubMedCentral Parvizi J, Zmistowski B, Berbari EF, et al. New definition for periprosthetic joint infection: from the Workgroup of the Musculoskeletal Infection Society. Clin Orthop Relat Res. 2011;469:2992–4.CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Parker MJ. Cutting-out of the dynamic hip screw related to its position. J Bone Jt Surg Br. 1992;74:625.CrossRef Parker MJ. Cutting-out of the dynamic hip screw related to its position. J Bone Jt Surg Br. 1992;74:625.CrossRef
22.
Zurück zum Zitat Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Statist Assoc. 1958;53:457–81.CrossRef Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Statist Assoc. 1958;53:457–81.CrossRef
23.
Zurück zum Zitat American Society of Anesthesiology. New classification of physical status. Anesthesiology. 1963;24:111–4. American Society of Anesthesiology. New classification of physical status. Anesthesiology. 1963;24:111–4.
24.
Zurück zum Zitat Lenich A, Vester H, Nerlich M, et al. Clinical comparison of the second and third generation of intramedullary devices for trochanteric fractures of the hip—blade vs screw. Injury. 2010;41:1292–6.CrossRefPubMed Lenich A, Vester H, Nerlich M, et al. Clinical comparison of the second and third generation of intramedullary devices for trochanteric fractures of the hip—blade vs screw. Injury. 2010;41:1292–6.CrossRefPubMed
25.
Zurück zum Zitat Yu W, Zhang X, Zhu X, et al. Proximal femoral nails anti-rotation versus dynamic hip screws for treatment of stable intertrochanteric femur fractures: an outcome analyses with a minimum 4 years of follow-up. BMC Musculoskelet Disord. 2016;21(17):222.CrossRef Yu W, Zhang X, Zhu X, et al. Proximal femoral nails anti-rotation versus dynamic hip screws for treatment of stable intertrochanteric femur fractures: an outcome analyses with a minimum 4 years of follow-up. BMC Musculoskelet Disord. 2016;21(17):222.CrossRef
26.
Zurück zum Zitat Galler M, Zellner M, Roll C, et al. A prospective study with 10 years follow-up of two-hundred patients with proximal femoral fracture. Injury. 2018;49:841–5.CrossRefPubMed Galler M, Zellner M, Roll C, et al. A prospective study with 10 years follow-up of two-hundred patients with proximal femoral fracture. Injury. 2018;49:841–5.CrossRefPubMed
27.
Zurück zum Zitat Mereddy P, Kamath S, Ramakrishnan M, et al. The AO/ASIF proximal femoral nail antirotation (PFNA): a new design for the treatment of unstable proximal femoral fractures. Injury. 2009;40:428–32.CrossRefPubMed Mereddy P, Kamath S, Ramakrishnan M, et al. The AO/ASIF proximal femoral nail antirotation (PFNA): a new design for the treatment of unstable proximal femoral fractures. Injury. 2009;40:428–32.CrossRefPubMed
28.
Zurück zum Zitat Mundi S, Chaudhry H, Bhandari M. Systemic review on the inclusion of patients with cognitive impairment in hip fractures trials: a missed opportunity? Can J Surg. 2014;57:E141–E145145.CrossRefPubMedPubMedCentral Mundi S, Chaudhry H, Bhandari M. Systemic review on the inclusion of patients with cognitive impairment in hip fractures trials: a missed opportunity? Can J Surg. 2014;57:E141–E145145.CrossRefPubMedPubMedCentral
29.
Zurück zum Zitat Hsu CE, Chiu YC, Tsai SH, et al. Trochanter stabilising plate improves treatment outcomes in AO/OTA 31-A2 intertrochanteric fractures with critical thin femoral lateral walls. Injury. 2015;46:1047–53.CrossRefPubMed Hsu CE, Chiu YC, Tsai SH, et al. Trochanter stabilising plate improves treatment outcomes in AO/OTA 31-A2 intertrochanteric fractures with critical thin femoral lateral walls. Injury. 2015;46:1047–53.CrossRefPubMed
30.
Zurück zum Zitat Puram C, Pradhan C, Patil A, et al. Outcome of dynamic hip screw augmented with trochanteric wiring for tretement of unstable type A2 intertrochanteric femur fractures. Injury. 2017;48(Suppl 2):S72–S7777.CrossRefPubMed Puram C, Pradhan C, Patil A, et al. Outcome of dynamic hip screw augmented with trochanteric wiring for tretement of unstable type A2 intertrochanteric femur fractures. Injury. 2017;48(Suppl 2):S72–S7777.CrossRefPubMed
31.
Zurück zum Zitat Geller JA, Saifi C, Morrison TA, Macaulay W. Tip-apex distance of intramedullary devices as a predictor of cut-out failure in the treatment of pertrochanteric elderly hip fractures. Int Orthop. 2010;34:719–22.CrossRefPubMed Geller JA, Saifi C, Morrison TA, Macaulay W. Tip-apex distance of intramedullary devices as a predictor of cut-out failure in the treatment of pertrochanteric elderly hip fractures. Int Orthop. 2010;34:719–22.CrossRefPubMed
32.
Zurück zum Zitat Fujii T, Nakayama S, Hara M, et al. Tip-apex distance is most important of six predictors of screw cutout after internal fixation of intertrochanteric fractures in woman. JBJS Open Access. 2017;2:e0022.CrossRefPubMedPubMedCentral Fujii T, Nakayama S, Hara M, et al. Tip-apex distance is most important of six predictors of screw cutout after internal fixation of intertrochanteric fractures in woman. JBJS Open Access. 2017;2:e0022.CrossRefPubMedPubMedCentral
33.
Zurück zum Zitat Lenich A, Bachmeier S, Prantl L, et al. Is the rotation of the femoral head a potential initiation for cuting out? A theoretical and experimental approach. BMC Musculoskelet Disord. 2011;12:79.CrossRefPubMedPubMedCentral Lenich A, Bachmeier S, Prantl L, et al. Is the rotation of the femoral head a potential initiation for cuting out? A theoretical and experimental approach. BMC Musculoskelet Disord. 2011;12:79.CrossRefPubMedPubMedCentral
Metadaten
Titel
PFNA and DHS for AO/OTA 31-A2 fractures: radiographic measurements, morbidity and mortality
verfasst von
Franz Müller
Matthias Doblinger
Tanja Kottmann
Bernd Füchtmeier
Publikationsdatum
31.10.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Trauma and Emergency Surgery / Ausgabe 5/2020
Print ISSN: 1863-9933
Elektronische ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-019-01251-w

Weitere Artikel der Ausgabe 5/2020

European Journal of Trauma and Emergency Surgery 5/2020 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.